UOP212 METHOD APPLICATION
SYSTEM
1
PCmicrocomputer
Electronic interface
Potentiometer
ElectrodeSystem
(indicator/reference)
Sample
Gas lineWet Gas
MeterDreshelBottle
Dilutionglassware
5/11/2018 2009-06-15 Troubleshooting Titration Method - slidepdf.com
http://slidepdf.com/reader/full/2009-06-15-troubleshooting-titration-method 2/7
WHAT COULD LEAD TO WRONG RESULTS?
Battery Errors Communication Errors Custom Method Errors Delay Time Errors Dilution Errors Dispense Errors Drive Unit Errors Method Errors Priming Errors Syringe Errors Titration Errors
2
5/11/2018 2009-06-15 Troubleshooting Titration Method - slidepdf.com
http://slidepdf.com/reader/full/2009-06-15-troubleshooting-titration-method 3/7
BATTERY ERRORBATTERY FAILURE
THE ENTIRE METHOD, OR PART OF THE METHOD(CALCULATOR) HAD BEEN LOST DUE TOBATTERY PRESERVING IT RUN OFF OR ISRUNNING LOW.
3
5/11/2018 2009-06-15 Troubleshooting Titration Method - slidepdf.com
http://slidepdf.com/reader/full/2009-06-15-troubleshooting-titration-method 4/7
COMMUNICATION ERROR The connection to the drive unit is Not as good as required
and/or The protocol and baud rate settings are Not correct.
The echo received from the drive unit was not correct. The drive unit responded to the last command with either
too little or too much data. A serial I/O character was not received within the expected
amount of time.
A physical serial I/O error occurred while attempting tosend a command to the drive unit. An invalid response was received from the drive unit for the
last command sent.
4
5/11/2018 2009-06-15 Troubleshooting Titration Method - slidepdf.com
http://slidepdf.com/reader/full/2009-06-15-troubleshooting-titration-method 5/7
METHOD ERROR A titration method parameters cannot be set or copied
properly as a custom/user method.
At least one operation must be specified in a custommethod to have the validity of operation. (cannot beempty)
Mechanical/Physical limitation: This method will not
operate as expected because more volume has beenaspirated or filled than will fit in the syringe.
5
5/11/2018 2009-06-15 Troubleshooting Titration Method - slidepdf.com
http://slidepdf.com/reader/full/2009-06-15-troubleshooting-titration-method 6/7
DELAY TIME ERROR Range : Delay time is out of range.
Resolution: Delay time resolution is not as per possible
configuration.
6
5/11/2018 2009-06-15 Troubleshooting Titration Method - slidepdf.com
http://slidepdf.com/reader/full/2009-06-15-troubleshooting-titration-method 7/7
DILUTION ERROR The dilution or ratio and a volume, or any 2 volume values
must be entered for a dilution method. The dilution specified cannot be achieved using the syringe
size(s) configured. Try one of the following: Use different dilution values. Use different sized syringes. Consult the user's manual to determine appropriate values.
A dilution of one or a ratio of zero cannot be accomplished.
A dilution greater than one or a ratio greater than zeromust be used. A dilution cannot be performed when any of the dilution
volumes is zero. Dilution volumes of greater than zeromust be used.
7