7/30/2019 14 Medical Ultrasonography
1/23
MEDICAL
ULTRASONOGRAPHY
CHAPTER 13
-
7/30/2019 14 Medical Ultrasonography
2/23
Introduction
Ultrasonography is a noninvasive technique
that uses a skin probe emitting sound waves
(ultrasound) .
Thus, visual images of the internal organs, e.g.,
the urinary tract, can be obtained for the
purpose of assessing its position .
-
7/30/2019 14 Medical Ultrasonography
3/23
Audible Sound
Humans can hear only a limited range of
frequencies that are called the audible spectrum.
The frequncy range of audible sound is
appropriately 20 Hz to 20 kHz
-
7/30/2019 14 Medical Ultrasonography
4/23
Ultrasound
Frequncies higher than the 20KHz are calledultraound.
Ultrasound is sound with a frequency greater
than the upper limit of human hearing,approximately 20 kilohertz/20,000 Hertz.
Some animals, such as dogs, dolphins, bats, and
mice have an upper limit that is greater than thatof the human ear and thus can hear ultrasound.
http://en.wikipedia.org/wiki/Soundhttp://en.wikipedia.org/wiki/Frequencyhttp://en.wikipedia.org/wiki/Hertzhttp://en.wikipedia.org/wiki/Doghttp://en.wikipedia.org/wiki/Dolphinshttp://en.wikipedia.org/wiki/Bathttp://en.wikipedia.org/wiki/Micehttp://en.wikipedia.org/wiki/Earhttp://en.wikipedia.org/wiki/Earhttp://en.wikipedia.org/wiki/Micehttp://en.wikipedia.org/wiki/Bathttp://en.wikipedia.org/wiki/Dolphinshttp://en.wikipedia.org/wiki/Doghttp://en.wikipedia.org/wiki/Hertzhttp://en.wikipedia.org/wiki/Frequencyhttp://en.wikipedia.org/wiki/Sound -
7/30/2019 14 Medical Ultrasonography
5/23
Sonography
Medical ultrasonography (sonography) is an
ultrasound-based diagnostic imaging technique used to
visualize internal organs, their size, structure and any
pathological lesions.
http://en.wikipedia.org/wiki/Ultrasoundhttp://en.wikipedia.org/wiki/Medical_imaginghttp://en.wikipedia.org/wiki/Lesionhttp://en.wikipedia.org/wiki/Image:Baby_in_ultrasound.jpghttp://en.wikipedia.org/wiki/Lesionhttp://en.wikipedia.org/wiki/Medical_imaginghttp://en.wikipedia.org/wiki/Ultrasound -
7/30/2019 14 Medical Ultrasonography
6/23
Working
Ultrasonography (sonography) uses a probecontaining one or more acoustic transducers tosend pulses of sound into a material.
Whenever a sound wave encounters a materialwith a different acoustical impedance, part ofthe sound wave is reflected, which the probedetects as an echo.
http://en.wikipedia.org/wiki/Sonographyhttp://en.wikipedia.org/wiki/Transducerhttp://en.wikipedia.org/wiki/Transducerhttp://en.wikipedia.org/wiki/Sonography -
7/30/2019 14 Medical Ultrasonography
7/23
Working
The time it takes for the echo to travel back tothe probe is measured and used to calculate thedepth of the tissue interface causing the echo.
The greater the difference between acousticimpedences, the larger the echo is.
The difference between gases and solids is sogreat that most of the acoustic energy is
reflected, and so imaging of objects beyond thatregion is not possible.
http://en.wikipedia.org/wiki/Echohttp://en.wikipedia.org/wiki/Echo -
7/30/2019 14 Medical Ultrasonography
8/23
To generate a 2D-image
The probe is swivelled, either mechanically or
electronically through a phased array of acoustic
transducers.
The data is analyzed by computer and used to
construct the image.
In a similar way, 3D images can be generated
by computer using a specialised probe.
http://en.wikipedia.org/wiki/2Dhttp://en.wikipedia.org/wiki/Phased_arrayhttp://en.wikipedia.org/wiki/3D_computer_graphicshttp://en.wikipedia.org/wiki/3D_computer_graphicshttp://en.wikipedia.org/wiki/Phased_arrayhttp://en.wikipedia.org/wiki/2D -
7/30/2019 14 Medical Ultrasonography
9/23
Display mode
A- mode
B- mode
M- Mode
-
7/30/2019 14 Medical Ultrasonography
10/23
A mode :
-
7/30/2019 14 Medical Ultrasonography
11/23
B MODE
The B mode stands for brightness.
Shifting the position of ultrasound transmision& reception results in scanning.
Echo signals received undergo intensitymodulation, so that a cross sectional image canbe diaplyed on the CRT.
Using this display method, we can see internalstructures in patient safety .
-
7/30/2019 14 Medical Ultrasonography
12/23
M MODE
M mode stands for motion.
This display mode is useful for studying moving
tragets such as the anterior mitral valves.
The transducer position remain fixed when using thismode while the display moves across the monitor
screen ina movemt called scroling.
Scrolling allows movement of a vale to be displayedas an image
-
7/30/2019 14 Medical Ultrasonography
13/23
ULTRASOUND EQUIPMENT
-
7/30/2019 14 Medical Ultrasonography
14/23
PRINCIPLES OF ULTRASOUND EQUIPMENT
Ultrasound Diagnostic equipment uses a pulsereflection system.
The pulse generation circuit generates high volatgepulses.
Then the T delay line circuit applies appropiate delaytime for each transmitted pulse for electronicfocussing for the ultrasound beam.
When the ultarsound probe touches the patients skin
surface, ultrsound waves are transmittable to thepatinet.
-
7/30/2019 14 Medical Ultrasonography
15/23
PRINCIPLES OF ULTRASOUND EQUIPMENT
The delayed high volateg pulses are applied to thepiezo electric transducer elements
They produce the ultarsound waves, which will enter
to the patiinet & will be reflected back by internalstructures. The reflected echo signals are received bythe same probe.
Then theay are input the R delay line circuit to
compensate for the trasnmission delay factor of thepulses & mix the echo signals
-
7/30/2019 14 Medical Ultrasonography
16/23
PRINCIPLES OF ULTRASOUND EQUIPMENT
Then the receiver circuits ampliier the mixedecho signasl from R delay line
In new equipment, the echo signals are
produced using a digital scan converter so thattehimage data is stored into the frame memory& read sequentilaly out in a format for displayon the TV monitor.
The result is real time imaging .
-
7/30/2019 14 Medical Ultrasonography
17/23
DOPPLER SONOGRAPHY
Ultrasonography can be enhanced with Dopplermeasurements, which employ the Doppler effectto assess whether structures (usually blood) aremoving towards or away from the probe, and itsrelative velocity.
By calculating the frequency shift of aparticular sample volume, for example a jet of
blood flow over a heart valve, its speed anddirection can be determined and visualised.
http://en.wikipedia.org/wiki/Doppler_effecthttp://en.wikipedia.org/wiki/Doppler_effect -
7/30/2019 14 Medical Ultrasonography
18/23
DOPPLER SONOGRAPHY
This is particularly useful in cardiovascularstudies (ultrasonography of the vasculature andheart) and essential in many areas such asdetermining reverse blood flow in the livervasculature in portal hypertension.
The Doppler information is displayedgraphically using spectral Doppler, or as an
image using colour Doppler or power Doppler. It is often presented audibly using stereo speakers: this
produces a very distinctive, although synthetic, sound.
http://en.wikipedia.org/wiki/Portal_hypertensionhttp://en.wikipedia.org/wiki/Portal_hypertension -
7/30/2019 14 Medical Ultrasonography
19/23
APPLICATIONS:
It images muscle and soft tissue very well and is particularlyuseful for delineating the interfaces between solid and fluid-filled spaces.
It renders "live" images, where the operator can dynamically
select the most useful section for diagnosing and documentingchanges, often enabling rapid diagnoses.
It shows the structure as well as some aspects of the function oforgans.
It has no known long-term side effects and rarely causes anydiscomfort to the patient.
http://en.wikipedia.org/wiki/Musclehttp://en.wikipedia.org/wiki/Soft_tissuehttp://en.wikipedia.org/wiki/Soft_tissuehttp://en.wikipedia.org/wiki/Muscle -
7/30/2019 14 Medical Ultrasonography
20/23
APPLICATIONS:
Equipment is widely available andcomparatively flexible.
Small, easily carried scanners are available;
examinations can be performed at the bedside. Relatively inexpensive compared to other
modes of investigation (e.g. computed X-ray
tomography, DEXA or magnetic resonanceimaging).
http://en.wikipedia.org/wiki/Computed_tomographyhttp://en.wikipedia.org/wiki/Computed_tomographyhttp://en.wikipedia.org/wiki/Dual_energy_X-ray_absorptiometryhttp://en.wikipedia.org/wiki/Magnetic_resonance_imaginghttp://en.wikipedia.org/wiki/Magnetic_resonance_imaginghttp://en.wikipedia.org/wiki/Magnetic_resonance_imaginghttp://en.wikipedia.org/wiki/Magnetic_resonance_imaginghttp://en.wikipedia.org/wiki/Dual_energy_X-ray_absorptiometryhttp://en.wikipedia.org/wiki/Computed_tomographyhttp://en.wikipedia.org/wiki/Computed_tomographyhttp://en.wikipedia.org/wiki/Computed_tomographyhttp://en.wikipedia.org/wiki/Computed_tomography -
7/30/2019 14 Medical Ultrasonography
21/23
DISADVANTAGES
Classical ultrasound devices have troublepenetrating bone but current research onultrasound bone imaging will make it possible
with dedicated devices in the future. Ultrasound performs very poorly when there is
a gas between the scan head and the organ ofinterest, due to the extreme differences inacoustical impedance .
http://en.wikipedia.org/wiki/Bonehttp://en.wikipedia.org/w/index.php?title=Ultrasound_bone_imaging&action=edithttp://en.wikipedia.org/w/index.php?title=Ultrasound_bone_imaging&action=edithttp://en.wikipedia.org/wiki/Bone -
7/30/2019 14 Medical Ultrasonography
22/23
DISADVANTAGES
Even in the absence of bone or air, the depth
penetration of ultrasound is limited, making it
difficult to image structures that are far removed
from the body surface, especially in obese patients. The method is operator-dependent. A high level of
skill and experience is needed to acquire good-quality
images and make accurate diagnoses.
-
7/30/2019 14 Medical Ultrasonography
23/23
MAINTENANCE
Make sure to check the system daily before starting to examinepatinets
Daily check will help dicover a malfunction at the earliestpossible stage.
Before deciding on the malfucntion , check whethr the power isturned ON, whether the chracters are displaying. Etc.
If charcaters are diaplyed & no ultrasound image then set thegain knob t themaximum positions.
If this operation do not correct the problem, confirm that theprobe is connected correctly.