© 2006 msa vaporlab™ 700 gas analyzer. © 2006 msa product description rack mount, turn-key gas...
TRANSCRIPT
© 2006 MSA
VaporLab™ 700 Gas Analyzer
VaporLab™ 700 Gas Analyzer
© 2006 MSA
Product DescriptionProduct Description
Rack mount, turn-key gas analyzer Analyzes a complex matrix for
trace quantities of toxic gases Gas chromatography is used to
separate the gas of interest from background gases
© 2006 MSA
Product DescriptionProduct Description
Rugged, long life Metal Oxide Detector
Specific chemical analysis Accurate and reproducible
measurements High Sensitivity - Detection
capability down to 5 ppb
© 2006 MSA
Product DescriptionProduct Description
Gas Chromatography Made Simple No support gas cylinders required
Scrubbed room air used as support gas No hazardous source or detector gas
Solid-state Metal-Oxide sensor No complex calculations or
adjustments Optional auto-calibrator
Ensures greatest accuracy without any customer intervention
© 2006 MSA
Product DescriptionProduct Description
Uses the power of Gas Chromatographic Analysis without the complex operational or support requirements associated with this technology
Low Cost of Operation & Maintenance
© 2006 MSA
VaporLab 700 Flow Diagram
VaporLab 700 Flow Diagram
CO
NC
EN
TRATO
R
VALVE #2
PRESSURE SENSOR
N.O
.
N.C.
CO
M
AMPLIFIERCARRIER
PRESSURE SIGNAL
VENTCARRIER
PUMPSCRUBBER
CARRIER INLET
ANALYTICAL COLUMN
VENT
OVEN HEATER
THERMISTOR
COLUMN TEMP.
SIG.
DETECTOR
GC DETECTOR
SIG.
#3N.O
.
N.C
.C
OM
N.O.
N.C.COM
VALVE #4N
.O.
N.C.
CO
M
SAMPLE PUMP
N.O
.
N.C.
CO
M
N.O
.
N.C.
CO
M
VALVE #5
VALVE #6
SAMPLE INLET
CAL GAS INLET
#1
FAN
MANIFOLD HEATER
VENT
CALIBRATION GAS DILUTION ACCESSORY
N.O.
N.C.COM
N.O.
N.C.COM
FLOW CONTROLLER
ZERO AIR
CAL GAS
MIXER
FLOW CONTROLLER
100% TO 0.1% DILUTED CALIBRATION GAS
(P> 1 ATM)
VENT
N.C.
N.C.
© 2006 MSA
Auto-Calibration/Dilutor Option
Auto-Calibration/Dilutor Option
Allows self-calibration in the field Built into the unit Reduces cost of operation
Less expensive gas standards at ppb levels
Allows for simpler field maintenanceRecalibration of the detector is easy!Multi-point calibration improves accuracy
and dynamic range
© 2006 MSA
How Does It WorkHow Does It Work
Sample is introduced into the unit
Pre-concentrator to provide greatest sensitivity
Pre-concentrator is heated to produce a plug of gas on analytical column
Gas of interest is separated from background gases on GC column
After a reproducible period of time, gas of interest is detected with solid state detector
© 2006 MSA
Optional DisplayOptional Display
Menu driven keypad
Provides easy access to menus including system status, calibration and alarms
LED’s provide quick indication of system status
© 2006 MSA
ApplicationsApplications
Industrial Monitoring Process Control Monitoring OHSA Occupational Exposure
Monitoring Perimeter Monitoring (EPA)
© 2006 MSA
BTEX DetectionBTEX Detection
BTEX = Benzene, Toluene, Ethyl Benzene and Xylenes
Oil and Gas Industry Process and storage applications
Food and Beverage Industry Approved for all ISBT Method 12 criteria
for selective measurement of BTEX in CO2
© 2006 MSA
BTEX SeparationBTEX SeparationBTEX @ 1000ppb
20000
30000
40000
50000
60000
70000
80000
90000
0 100 200 300 400
Time (Seconds)
Sig
nal
(H
z)
Benzene 33353 Hz
Toluene50726 Hz
Ethyl Benzene39361 Hz
o-Xylene26394 Hz
Retention TimesBenzene=92 secToluene=142Ethyl Benzene=206O-Xylene=316
© 2006 MSA
BTEX SeparationBTEX Separation
BTEX @ <10ppb
24000
24500
25000
25500
26000
26500
0 100 200 300 400
Time (Seconds)
Sig
na
l (H
z)Benzene831 Hz
Toluene1335 Hz
Ethyl Benzene941 Hz
o-Xylene608 Hz
Water
Injection Peak
Retention TimesBenzene=92 secToluene=142Ethyl Benzene=206O-Xylene=316
© 2006 MSA
VaporLab 700 Gas Analyzer
VaporLab 700 Gas Analyzer