you can order this book in one of two formats: hardcover ... .pdf · report no. 151 (ncrp 2005b)....

23
Here is a sample chapter from Shielding Techniques for Radiation Oncology Facilities © . This sample chapter is copyrighted and made available for personal use only . No part of this chap- ter may be reproduced or distributed in any form or by any means without the prior written permis- sion of Medical Physics Publishing. You can order this book in one of two formats: Hardcover: ISBN 978-1-951134-00-6 eBook: ISBN 978-1-944838-01-3 To order by phone, call MPP at 1-800-442-5778. 3RD EDITION MELISSA MARTIN PATTON H. MC GINLEY SHIELDING TECHNIQUES FOR RADIATION ONCOLOGY FACILITIES

Upload: others

Post on 19-Oct-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

Here is a sample chapter from Shielding Techniques for Radiation Oncology Facilities©.

This sample chapter is copyrighted and made available for personal use only. No part of this chap-ter may be reproduced or distributed in any form or by any means without the prior written permis-sion of Medical Physics Publishing.

You can order this book in one of two formats:

Hardcover: ISBN 978-1-951134-00-6eBook: ISBN 978-1-944838-01-3

To order by phone, call MPP at 1-800-442-5778.

3RD EDITION

MELISSA MARTIN PATTON H. MCGINLEY

SHIELDING TECHNIQUES FOR RADIATION ONCOLOGY FACILITIES

Page 2: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

9

2

Conventional Shielding

Design

Shielding Design Goals and Occupancy

The purpose of radiation shielding is to reduce the effective dose ofradiation produced by radiation therapy equipment to a sufficientlylow level outside the room. The effective dose level required isdetermined by the local or national regulatory bodies. The requireddose level is typically different for public occupancy (uncontrolledaccess) vs. occupational occupancy (controlled access).

The dose rate reaching a protected location is directly propor-tional to the workload (W), a measure of the radiation produced bythe machine. For a linear accelerator, workload at isocenter is theabsorbed dose-rate at isocenter, determined at the depth of the maxi-mum absorbed dose in water, specified in gray (Gy) per some inter-val of time (e.g., per hour, week, or year) (NCRP 2005b). Theworkload at isocenter is then normalized to 1 meter from the x-raytarget (if the distance from the x-ray target to isocenter is not1 meter) to produce the workload (W) used in the shielding calcula-tions.

In addition to workload, the required shielding is also a functionof the machine energies (MVs); the distance from the x-ray target, orisocenter, to the point being shielded; the fraction of time that thebeam is oriented in that particular direction; and the fraction of timethat the space under consideration is considered to be occupied.

Page 3: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

Shielding Techniques for Radiation Oncology Facilities10

The energy rating of a linear accelerator x-ray beam, as stated bythe manufacturer, refers to the accelerating voltage and the end pointenergy of the bremsstrahlung spectrum produced by the linear accel-erator. The energy rating of a linear accelerator beam is commonlyreferred to simply as MV (for megavoltage). Precise definitions ofMV are provided by British Journal of Radiology (BJR) Supple-ments No. 11 (BJR 1972) and 17 (BJR 1983). The BJR 11 definitionof MV is used as the basis for the parameters included in this book.The term “MeV” is also used throughout this book to refer to mono-energetic photons or particles, not the energy rating of the linearaccelerator.

A protected location is a location outside the treatment room thatmay be occupied. This is customarily considered to be no closer toisocenter than 300 mm beyond a barrier. The shielding design goal(P) is a practical reference limit (or dose constraint) on the dose rateat a protected location. A controlled area is a limited-access area inwhich the occupational exposure of personnel to radiation or radio-active material is under the supervision of an individual in charge ofradiation protection.

NCRP 151 recommended a 5-mSievert (mSv)-per-year shieldingdesign goal for controlled areas and a 1-mSv-per-year shieldingdesign goal for uncontrolled areas. These shielding design goals mayalso be expressed as the yearly average dose per week (0.1 mSv/week for controlled areas and 0.02 mSv/week for uncontrolledareas) assuming 50 work weeks per year. In this case workload is theannual workload expressed as a per-week average.

Note NCRP 116 (NCRP 1993) recommended a public dose con-straint of 0.25 mSv per year for any individual facility (vs. 1 mSv/year), a discrepancy that delayed publication of NCRP 151 and thecorresponding NCRP 147 (NCRP 2004b; diagnostic shielding stan-dard). The discrepancy was resolved by NCRP Statement 10 (NCRP2004a), which noted the shielding methodology is sufficiently con-servative that the actual exposure of the maximally exposed individ-ual would be expected to be at least a factor of four less than theshielding design goal. Adherence to the NCRP 151 methodology istherefore important independent of any anticipated survey measure-ments. Specifically, for primary barriers, attenuation provided by apatient or phantom at isocenter must not be included in the shieldingcalculations or survey measurements. The use factor (U) for primarybarrier calculations also should be conservative compared with thefraction of clinical workload directed toward the location with three-dimensional (3D) treatment planning. For secondary barriers, it wascommon knowledge that the linear accelerators in use when NCRP

Page 4: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

2. Conventional Shielding Design 11

151 was published had a leakage fraction on the order of a factor offour less than the 0.1%. Using a similarly conservative x-ray leakagefraction is therefore necessary to maintain the intent of NCRP 151.Scatter calculations should be based on the maximum field dimen-sions, with typical dimensions anticipated to be at least a factor offour less in area.

Although International Commission on Radiological Protection(ICRP) Publication 103 (ICRP 2007) also recommends 1 mSv peryear as the individual dose limit for the public, regulatory bodiesmay impose a lower dose constraint (e.g., 0.3 mSv/year in theUnited Kingdom). Although there is no biological basis for therequirement, some countries may also impose a limit on instanta-neous dose rate to simplify survey measurements.

Occupancy factor (T) reflects the fraction of time a protectedlocation may be occupied. The maximum permissible value ofshielded dose rate at a location with occupancy T is given by P/T.The time-averaged dose rate (TADR) for an individual at a protectedlocation is the total shielded dose rate at that location multiplied bythe occupancy T. The use of TADR allows direct comparison withthe shielding design goal. Comparing shielded dose rate with P/T orTADR with P are equally acceptable ways to express compliancewith the shielding design goal.

The occupancy factor guidelines from NCRP 151 are given inTable 2-1. The appropriate occupancy factor at the entrance willdepend on the proximity to nearby high occupancy locations.Assuming T = 1/8 would be appropriate at an isolated entrance loca-

Table 2-1 Suggested occupancy factors [NCRP 151 Table B.1]

Occupancy Factor Location

1 Full occupancy areas (areas occupied full-time by an individual), e.g., administrative or clerical offices, treatment planning areas, treatment control rooms, nurse stations, receptionist areas, attended waiting rooms, and occupied space in nearby buildings

0.5 Adjacent treatment room and patient examination room adjacent to shielded vault

0.2 Corridors, employee lounges, and staff rest rooms

0.125 Treatment vault doors (if at isolated location)

0.05 Public toilets, unattended vending rooms, storage areas, outdoor areas with seating, unattended waiting rooms, patient holding areas, attics, and janitors’ closets

0.025 Outdoor areas with only transient pedestrian or vehicular traffic, unattended parking lots, vehicular drop-off areas (unattended), stairways, and unattended elevators

Page 5: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

Shielding Techniques for Radiation Oncology Facilities12

tion normally occupied only when entering and leaving the treatmentroom. A corridor immediately beyond an entrance would normallybe designated at least T = 1/2 if not T = 1 depending on the proxim-ity of the control area or other high-occupancy location.

The United States Nuclear Regulatory Commission (NRC)requires the shielded dose rate at any unrestricted location to notexceed 0.02 mSv in any hour. This is not an instantaneous dose raterequirement, but it is interpreted by NCRP 151 to be the shieldeddose rate resulting from the maximum number of patient treatmentsanticipated per hour using the yearly average workload per patient.NCRP 151 Equation 3.14 effectively places a limit on the maximumvalue of P/T as shown in Equation (2-1), where Nmax is the maximumnumber of patient treatments per hour and Nwk is the average numberof patient treatments per week. This in turn can then be viewed as aconstraint on the minimum occupancy for uncontrolled access loca-tions (P = 0.02 mSv/week) as given in Equation (2-2). Although T =1/40 is in theory permitted for uncontrolled access locations withonly transient occupancy, the NRC 0.02-mSv-per-hour requirementwill typically imply a somewhat higher minimum occupancy.

Example: minimum occupancy40 patient treatments per dayNmax = 6 patient treatments per hourNwk = 5 × 40 = 200 patient treatments per weekP/Tmax = 0.02 × 200 / 6 = 0.667 mSv/weekTmin = 6 / 200 = 0.030

The procedure described above can be adapted for other countriesthat have a specific requirement for radiation exposure at a locationin an hour with the maximum clinical workload per hour (e.g.,0.0075 mSv/hour in the United Kingdom).

Calculation Methods

The shielding design methods described here are based on NCRPReport No. 151 (NCRP 2005b). Figure 2-1 shows a typical roomplan for a high-energy (15 MeV) medical accelerator. Two types ofradiation barriers or shields are considered—primary and second-ary. The primary barrier is irradiated by the x-ray beam produced by

(2-1)

(2-2)P T mSv wk mSv hr N N

T N Nwk

wk

/ ( / ) . / /

/max max

min max

0 02

Page 6: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

2. Conventional Shielding Design 13

the accelerator, and the secondary barrier receives only radiationscattered by the patient and the surfaces of the treatment room andradiation transmitted through the accelerator shielding (head leak-age). Primary radiation is limited in direction by the placement ofthe accelerator in the treatment room, and the maximum beam sizeis used to determine the portion of the walls, ceiling, and floor thatwill be designated as primary barriers. Secondary radiation, how-ever, is emitted in all directions and covers all of the treatment roomsurfaces.

If the linear accelerator operates at multiple energies, separatecalculations should be performed at each energy for primary barrier(protecting locations A and B) and secondary barrier calculations(protecting Locations C through G). The total shielded dose rate isthe sum of the shielded dose rate values calculated for each energy.The shielding evaluation at the entrance (Location H) must alsoinclude not only secondary radiation, but also scattered radiationreaching the entrance through the maze.

As illustrated in Figure 2-2, the ceiling of the treatment room alsorequires shielding with primary and secondary barriers. Even for a

H

G

F

D

E EB

CC A

Entrance

X-ray Target Rotational Plane

Maze

Primary barriers typically 2.1 to 2.4 m thick

Secondary barriers typically 0.9 to 1.2 m thick

Thickness required varies with workload, MV,

occupancy, and distance

Typically swing door at entrance

May be no door at entrance for long maze with additional bends

Figure 2-1 Simplified drawing of treatment room with maze. (Letters in the figure are referred to in the text.)

ABIsocenter

X-ray Target*1 m

Ceiling primary barrier typically 1.8 to 2.4 m thick depending primarily on occupancy

2.1 to 2.4 m

2.1 to 2.4 mFigure 2-2 Simplified

section drawing of treatment room.

Page 7: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

Shielding Techniques for Radiation Oncology Facilities14

single-story facility, it is preferable to provide sufficient shieldingfor low occupancy on the roof. Restricting access to the roof canreduce, but not eliminate, ceiling shielding because radiationthrough the ceiling will scatter to surrounding locations. Skyshinecalculations must be performed if no primary barrier is included inthe ceiling.

As illustrated in Figure 2-3, a treatment room with no mazerequires substantially less space. However, the door at the entranceto the treatment room must provide shielding comparable to theadjacent secondary barrier wall, making it far more expensive.Given space constraints faced by medical facilities, treatment roomswith no maze are increasingly common in the United States.Improvements in the design and operation of sliding doors has madethe no-maze design more appealing.

Primary Barrier Calculations

The traditional primary barrier approach uses Equation (1-4) to cal-culate the maximum primary barrier transmission for a given work-load and MV. The maximum transmission can then be used todetermine the minimum barrier thickness. The more generalapproach is to develop an expression for the total shielded dose ratebeyond the primary barrier, potentially including multiple MVs andboth photons and neutrons. The primary barrier composition andthickness(es) are then selected to ensure this total shielded dose rateis suitably less than the required P/T. This more general approach isdescribed here.

D

E EB

CC A

G

F

A

Entrance

X-ray Target Rotational Plane

Primary barriers typically 2.1 to 2.4 m thick

Secondary barriers typically 0.9 to 1.2 m thick

Thickness required varies with workload, MV,

occupancy, and distance

Variety of door types at entrance

(including sliding, swinging, bi-parting)

Figure 2-3 Simplified drawing of treatment room with no maze.

Page 8: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

2. Conventional Shielding Design 15

Primary barrier calculations use the geometry illustrated in Fig-ure 2-4. The unshielded dose rate (HUX) in sievert per week is givenby Equation (2-3).

In Equation (2-3) dpri is the distance from the x - r ay target to thepoint protected in meters; W is the workload or dose in gray perweek at 1 m from the x-ray target; and U is the use factor, or frac-tion of the workload directed toward the primary barrier.

Example: primary barrier unshielded dose rateWorkload at 15 MV: 230 Gy/weekU = 0.25Distance from target to protected location (dpri) = 6.5 m

HUX = (230)(0.25) / 6.52 = 1.36 × 100 Sv/week = 1.36 × 103 mSv/week

The shielded x-ray dose rate (Htr) (i.e., the calculated transmittedx-ray dose equivalent rate) is given by Equation (2-4).

In Equation (2-4), B is the x-ray transmission factor for the barrier.The transmission factor is calculated separately for each layer ofmaterial in the barrier, with B then calculated by multiplying thetransmission factor for these layers together. The transmission factorfor layer i is given by Equation (2-5).

Figure 2-4 Primary barrier geometry.

(2-3)HWU

dUX

pri2

(2-4)H H Btr UX

(2-5)B t TVL t TVLi i ie e10 101 1( / ) ( / )

Page 9: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

Shielding Techniques for Radiation Oncology Facilities16

where TVL1 is the first tenth-value layer (TVL) thickness, TVLe isthe thickness for each subsequent TVL, ti is the thickness of layer iof the barrier, ti1 is the amount of ti (if any) that falls within the firstTVL of the barrier, and tie is ti – ti1. Primary TVLs for customaryshielding materials are provided in Table A-1. See also Appendix B,which describes the physical properties of commonly used barriermaterials as well as addressing various material-related issues thatmay arise during design and construction.

Example: primary barrier transmission and shielded dose rateBarrier: 220 cm normal weight concreteTVL1 = 440 mm, TVLe = 410 mm from Table A-1

B = 10^(–440/440) × 10^[–(2200 – 440)/410] = 5.10 × 10–6

Htr = (1.36 × 103 mSv/week)(5.10 × 10–6) = 0.0069 mSv/week

The total shielded dose rate is the sum of the shielded dose rate ascalculated above for all the linear accelerator MVs. The barrier mustbe sufficient for the total shielded dose rate to be less than P/T. Alter-natively, the shielded dose rate can be multiplied by the occupancy(T) to calculate the TADR for the location, with the TADR thencompared directly with the shielding design goal (P) instead of P/T.

If the material thicknesses and densities are consistent with theassumptions, the measured x-ray dose rate is typically very close tothe dose rate calculated using Equation (2-4). In practice, slight vari-ation in the construction is to be expected (e.g., concrete density maybe slightly lower than anticipated). This may result in the measuredx-ray dose rate being slightly higher than the calculated x-ray doserate. With good construction practice, the measured dose rate shouldbe no larger than the calculated x-ray dose rate multiplied by the rec-ommended x-ray dose rate margin. A factor of two to three margin istypically appropriate for cast-in-place concrete, with lower marginpotentially acceptable for manufactured shielding material (becausethe density is more closely controlled). Where survey results areavailable for existing construction, it may be unnecessary to includeany margin (i.e., a factor of one margin).

With the current United States regulatory requirements, primarybarriers are typically the equivalent of 6.5- to 8-foot-thick normal-weight concrete. The thickness varies with location, depending pri-marily on occupancy and workload. The 6.5-foot thickness wouldtypically be suitable only for an exterior wall with no nearby highoccupancy anticipated. The 8-foot thickness is typically compatiblewith full occupancy uncontrolled access. This thickness also may be

Page 10: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

2. Conventional Shielding Design 17

used in a standardized design because it could typically be usedwithout modification. A typical primary barrier thickness is 7 to 7.5feet, which would typically be compatible with a location designatedP = 0.1, T = 1 or P = 0.2, T = 0.2, which are the most common com-binations of shielding design goal and occupancy adjacent to a treat-ment room.

At energies above 6 MV, photoneutron generation also must beconsidered if the primary barrier includes metal. This is addressed inChapter 6.

Secondary Barrier Calculations

Secondary radiation comes from two sources: leakage radiation fromthe linear accelerator head and scattered radiation from the patient.For energies above 6 MV, both x-ray and neutron leakage must beconsidered. The unshielded dose rate for secondary barriers is calcu-lated using the geometry in Figure 2-5.

The x-ray leakage unshielded dose rate due to leakage from thelinear accelerator head (HUL) in sievert per week is given by Equa-tion (2-6).

The secondary distance to the protected point (dsec) in Equation (2-6)is customarily measured from isocenter, since this is the average lin-ear accelerator head location. If the amount of shielding varies sig-nificantly with gantry orientation, it may be appropriate to measuredistance from the linear accelerator head instead, with separate cal-culations performed as a function of gantry angle. The x-ray leakage

* X-ray targetIsocenter

t

ScatterAngle

0.3 m

dsec

dsca

(A) Geometry for side wall immediately adjacent to primary barrier

* X-ray target

Isocenter

t

dsec

0.3 m

dsca

(B) Geometry for front or back wall parallel to plane of rotation

Figure 2-5 Secondary barrier geometry.

(2-6)HW f

dUL

L L

sec2

Page 11: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

Shielding Techniques for Radiation Oncology Facilities18

fraction (fL) is customarily 10–3, corresponding to the standard Inter-national Electrotechnical Commission (IEC) recommendation thatthe average x-ray leakage outside the beam must be less than 0.1%of the dose rate at isocenter. Note that measured x-ray leakage fortraditional linear accelerators is typically at least a factor of 4 lowerthan 0.1% (one of the factors considered in NCRP Statement 10).Using a leakage fraction less than 0.1% is therefore appropriate onlywith manufacturer data illustrating typical leakage of at least a factorof 4 less than the assumed leakage fraction.

The leakage workload (WLi) for a given energy (or MV) and treat-ment delivery category is the conventional workload multiplied bythe intensity-modulated radiation therapy (IMRT) ratio (CIi), asshown in Equation (2-7). Typical values of the IMRT ratio rangefrom 3 to 15 depending on the technique used for treatment.

As shown in Equation (2-8), the x-ray leakage workload (WL) fora given MV is then given by the sum of the leakage workload foreach treatment delivery category.

Example: x-ray leakage unshielded dose rate2 stereotactic body radiation therapy (SBRT) patient treatments

per day at 10 Gy each with CI = 320 Volumetric Modulated Arc Therapy (VMAT)IMRT patient treatments per day at 3 Gy each with CI = 312 3D patient treatments per day at 3 Gy each with CI = 1Leakage workload: 5 × (2 × 10 × 3 + 20 × 3 × 3 + 12 × 3 × 1) =

1380 Gy/weekLeakage fraction: 0.001Distance from isocenter to protected location: 7.5 mHUL = (1380)(10–3) / 7.52 = 2.45 × 10–2 Sv/week = 2.45 ×

101 mSv/week

The shielded x-ray leakage dose rate (HL) is given by Equation(2-9).

Here BL is the leakage transmission calculated from the barrier thick-ness and x-ray leakage tenth-value layers TVL1 and TVLe using the

(2-7)W W CLi i Ii

(2-8)W WL Lii

(2-9)H H BL UL L

Page 12: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

2. Conventional Shielding Design 19

same procedure used to calculate the primary barrier transmission.Leakage TVL values for customary shielding materials are given inTable A-2.

Example: x-ray leakage barrier transmission and shieldeddose rate

Barrier: 1067 mm normal weight concreteTVL1 = 340 mmTVLe = 290 mm at 6 MV from Table A-2

B = 10^(–340/340) × 10^[–(1067 – 340)/290] = 3.11 × 10–4

HUL = 2.45 × 101 mSv/week

HL = (2.45 × 101 mSv/week)(3.11 × 10–4) = 0.0076 mSv/week

As illustrated in Figure 2-5, the path to the protected location maytraverse the barrier at an angle, termed the slant angle ( ). In thiscase, the barrier thickness t is multiplied by the slant factor 1/cos( )to get the effective barrier thickness used to calculate barrier trans-mission. The effective barrier thickness is typically referred to asslant thickness.

Example: slant thicknessBarrier thickness: 1067 mm normal weight concreteSlant angle = 26°cos(26°) = 0.899Slant thickness = 1067 mm / 0.899 = 1187 mm

For energies above 6 MV, neutron leakage is included in the cal-culated secondary shielded dose rate. Neutron leakage is calculatedusing the same approach as for x-ray leakage—Equations (2-5)through (2-8)—except with using neutron leakage fraction (seeTable A-3) and neutron TVLs (see Table A-6). A lower neutronleakage workload may be assumed for a dual energy accelerator inwhich only the higher energy is used for the shielding calculationsbecause much of the actual workload will be typically performed at6 MV. If a separate calculation is performed for each MV, the neu-tron leakage workload is the same as the x-ray leakage workload.

The unshielded patient scatter dose rate (HUPS) in sievert perweek is given by Equation (2-10).

(2-10)Ha W U F

d dUPS

sca

( ) ( / )

sec

4002 2

Page 13: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

Shielding Techniques for Radiation Oncology Facilities20

As illustrated in Figure 2-5, dsca (meters) is the target to isocenterdistance (typically 1 m), dsec (meters) is the secondary distance fromisocenter to the point protected, U is use factor, and W is the work-load in gray per week at 1 m from the x-ray target, where

a( ) = scatter fraction or fraction of the primary beam absorbed dose at 1 m from the x-ray target that scatters from the patient at scatter angle at 1-m distance from isocenter for a 400-cm2 field area (Table A-4), and

and

F = field area in square centimeters at isocenter (e.g., 1600 cm2 for 40 × 40-cm field).

If dsca is not 1 m, it may be more straightforward to normalize Fto 1 m from the x-ray target instead of isocenter and remove the dsca

term from Equation (2-10). Assuming U = 1 for scatter is recom-mended. For secondary barriers immediately adjacent to a primarybarrier, applying the primary barrier use factor (typically 0.25) toscatter is permissible, particularly for existing construction. Note theleakage fraction will exceed 10–3 near the central axis, so assumingU = 1 for scatter compensates for this to some extent.

Note that the average field area in clinical use may be substan-tially less than the maximum field area (e.g., 20 × 20 cm2 might beconsidered typical). This conservatism is one of the reasons NCRPStatement 10 permitted NCRP 151 to use 1 mSv per year as theshielding design goal; hence, using the maximum field area is rec-ommended for shielding calculations, especially for new construc-tion.

Example: patient scatter unshielded dose rate2 SBRT patient treatments per day at 10 Gy each20 VMAT IMRT patient treatments per day at 3 Gy each12 3D patient treatments per day at 3 Gy eachWorkload: (5)(2 × 10 + 20 × 3 + 12 × 3) = 580 Gy/weekUse factor = 1 assumed.Scatter fraction at 26° scatter angle (6 MV): 3.95 × 10–3 (interpo-

lated from Table A-4 as the log of scatter fraction vs. linearangle)

Field area at 1 m dsca: 1600 cm2

Distance from isocenter to protected location: 6.2 mHUPS = (3.95 × 10–3)(580)(1)(1600/400) / 6.22 = 2.38 × 10–1 Sv/

week = 2.38×102 mSv/week

Page 14: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

2. Conventional Shielding Design 21

The patient scatter shielded dose rate (HPS) is given by Equation(2-11).

Here BPS is the patient scatter transmission calculated from thepatient scatter TVL1 and TVLe values, using the same procedure usedto calculate the primary barrier transmission. The patient scatterTVLs, which are given in Tables A-5, vary both with machine MVand scatter angle. The TVL values in Tables A-5 are based on NCRP151 Tables B.5 and B.6 and NCRP 151 Figure A.1. For values ofmachine MV or scatter angle outside the region included in NCRP151 Table B.5, the TVL values are interpolated from Figure A.1based on scatter energy from NCRP 151 Table B.6. Note the scatterenergy used to determine TVLs from NCRP 151 Figure A.1 areincreased slightly to make them consistent with the NCRP 151 TableB.5 scatter TVLs for lead.

Example: patient scatter shielded dose rateHUPS = 2.38 × 102 mSv/weekBarrier thickness: 1067 mm normal weight concreteSlant angle: 26°, cos(26°) = 0.899Slant thickness = 1067 mm / 0.899 = 1187 mmPatient scatter TVL = 281.3-mm linear interpolation from Table

A-5aB = 10^(–281.3/281.3) × 10^[–(1187 – 281.3)/281.3] = 6.03 × 10–5

HPS = (2.38 × 102 mSv/week)(6.03 × 10–5) = 0.0144 mSv/weekNote: the calculated HPS final digit may vary depending on num-

ber of significant figures retained in the intermediate calcula-tions.

The secondary shielded dose rate is the sum of HL (includingx-ray and neutron leakage) and HPS. The total secondary shieldeddose rate is the sum of the secondary shielded dose rate for all themachine MVs. The barrier must be sufficient for the total secondaryshielded dose rate to be less than P/T.

With the current U. S. regulatory requirements, secondary barri-ers for conventional linear accelerators are typically the equivalentof 3- to 4-feet-thick normal weight concrete. The thickness varieswith location, depending primarily on occupancy and workload. The3-foot thickness would typically be suitable only for an exterior backwall, with no nearby high occupancy anticipated. The 4-foot thick-ness is typically compatible with full-occupancy uncontrolled

(2-11)H H BPS UPS PS

Page 15: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

Shielding Techniques for Radiation Oncology Facilities22

access. The 4-foot thickness may also be used in a standardizeddesign because it could typically be used without modification. Atypical secondary barrier thickness is 3.5 feet, which would typicallybe compatible with a location designated P = 0.1, T = 1 or P = 0.2,T = 0.2, the most common combinations of shielding design goaland occupancy adjacent to a treatment room. For a control area, 4-foot wall thickness is preferred instead of 3.5-foot wall thicknessbecause of the high occupancy and the other potential sources ofscattered radiation (e.g., treatment room entrance and duct penetra-tion above) present at that location. For linear accelerators with avery high leakage workload, the secondary barrier would need to beproportionally thicker, with 5-foot or more thickness possibly appro-priate.

Flattening Filter Free (FFF) Accelerators

Linear accelerators have traditionally used flattening filters to pro-vide a relatively uniform dose rate within the treatment field. Theadvent of sophisticated 3D treatment planning software reduced theimportance of having uniform intensity in the field because the beamshape could be incorporated into the treatment planning software.The desire to provide higher instantaneous dose rate, along with theenabling treatment planning software, led to the development of lin-ear accelerators that have a so-called flattening filter free (FFF)mode, with higher instantaneous dose rates. This is of particularimportance for modalities, such as stereotactic radiosurgery (SRS)and Stereotactic Body Radiation Therapy (SBRT), to reduce treat-ment duration associated with the larger treatment fractions.

Except for countries that have regulations limiting the maximuminstantaneous dose rate, the FFF mode does not directly impactshielding calculations. However, FFF can impact the shielding cal-culations indirectly through enabling higher average workloads perhour, week, or year. To mitigate the impact on shielding required, itis now common to distribute the workload used in the shielding cal-culations over all the accelerator energies available, as addressed inthe next section, rather than simply assuming the entire workload atthe maximum energy. There is an insignificant difference betweenthe TVLs with and without the FFF mode at a given energy, so theworkload does not need to distinguish between the flattening filter(FF) and FFF modes.

However, the apparent concrete primary TVL at 10 MV, which iscommonly used for many FFF treatments, now appears to be largerthan the value provided in NCRP 151. Consequently, the 389 mm

Page 16: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

2. Conventional Shielding Design 23

TVL value from Institute of Physics and Engineering in Medicine(IPEM) 75 (IPEM 2002) is recommended as the TVLe in Table A-1because this provides closer agreement with recent survey data thanthe NCRP 151 TVLe value (370 mm).

Workload Examples

Most linear accelerators operate at multiple energies. This requiresthe primary and secondary barrier calculations to be performed sepa-rately for each energy, with the total shielded dose rate the sum ofthe shielded dose rates calculated for each energy.

Table 2-2 illustrates the workload described in NCRP 151 Sec-tion 7.1. NCRP 151 noted that the shielded dose rate contributionfrom the 6-MV workload is fairly minor in this example. NCRP 151indicated shielding calculations based solely on the 18-MV work-load would be acceptable to simplify the shielding calculations.

Following the example of NCRP 151, an absorbed dose of 3 Gy isused for the conventional treatment fraction in Table 2-2. For a tis-sue-maximum ratio in the 0.7 to 0.8 range, this is equivalent to apatient fraction in the 210 to 240 cGy range versus a more typicalpatient fraction on the order of 200 cGy. Note, however, that theactual workload will include machine usage not associated withadministering patient treatment (e.g., quality assurance [QA], main-tenance, and physics developmental activities). The 3-Gy-per-patient treatment recommended by NCRP 151 is intentionally higherthan the typical average patient workload to allow for such machineusage.

Leakage workload (WL) differs from workload at isocenterbecause it depends on the number of monitor units (MUs) required

Table 2-2 NCRP 151 Section 7.1 workload example

Energy(MV)

Patientsper day

Workload(Gy/pt)

Workload(Gy/wk)

MU/cGyRatio

Leakage Workload(Gy/wk)

18 IMRT 12 3 180 5 900

18 3D 18 3 270 1 270

18 30 450 1170 Total

6 IMRT 12 3 180 5 900

6 3D 3 3 45 1 45

6 15 225 945 Total

45 Total patient treatments per day

Page 17: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

Shielding Techniques for Radiation Oncology Facilities24

to deliver the dose at isocenter at a given energy. IMRT requires anincrease in MUs compared with conventional radiation therapy. TheIMRT ratio is the ratio of MU to cGy at isocenter. The IMRT ratiowill vary with the treatment delivery and with MV.

Table 2-3 provides an example of workload-per-patient treatmentand IMRT ratio for a variety of treatment modalities. The workloadused in the shielding calculations should be selected by the facilitystaff to provide an upper bound on the anticipated total workload.The treatment modalities are included as an aid to estimating thetotal workload. As noted in NCRP 151, the entire workload can beassumed to be performed at the maximum MV. However, the trendtoward increasing workload (caused by higher instantaneous doserates, hypofractionation, and more extensive use of IMRT) mayrequire unnecessary shielding that can be avoided by using a morerealistic workload in the shielding calculations. For example, it iscommon for a large fraction of the leakage workload to be deliveredat 6 MV. Performing shielding calculations at the highest MV for theworkload actually delivered at 6 MV would likely require an unnec-essary amount of neutron shielding in the door.

Primary Barrier Width Calculation

The primary barrier width (w) is recommended to have sufficientwidth to ensure at least a 0.3-m margin on either side of a field of

Table 2-3 Workload example for linear accelerator with three energies

Energy(MV)

Patientsper day

Workload(Gy/pt)

Workload(Gy/wk)

MU/cGyRatio

Leakage Workload (Gy/wk)

15 VMAT 2 3 30 3 90

15 3D 8 3 120 1 20

15 10 150 210 Total

10 SRS 1 20 100 3 300

10 SBRT 2 10 100 3 300

10 VMAT 5 3 75 3 225

10 3D 5 3 75 1 75

10 13 350 900 Total

6 SRS 2 20 200 3 600

6 SBRT 5 10 250 3 750

6 VMAT 15 3 225 3 675

6 22 675 2025 Total

45 Total patient treatments per day

Page 18: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

2. Conventional Shielding Design 25

maximum size rotated 45 degrees. For a square f × f-cm field at1-meter distance from the target, w is given by:

where dN is the distance from the target to the far side of the narrow-est part of the barrier (also in meters). Figure 2-6 illustrates the typi-cal locations for the narrowest point of the barrier. The square-rootof two factor is in included in Equation (2-12) to account for the45-degree clocking of the field. If the linear accelerator does notallow clocking of the field, the square root of two factor should beremoved from Equation (2-12).

The maximum field is typically a 40 × 40-cm square. If the cor-ners of the field are clipped, which is typically the case, the effectivefield size to be used in Equation (2-6) is smaller than 40 cm. Figure2-7 illustrates this with primary barrier width typically calculatedwith a 35-cm effective field size.

Note the diagonal measurement of the 35-cm effective field sizeis approximately 50 cm. If the 50-cm diagonal field size is used asthe basis to calculate the minimum primary barrier width, the squareroot of two factor should be removed from Equation (2-12), sinceclocking of the field has already been included in the 50-cm value.

(2-12)w f dN( / ) .100 2 0 6cm meters

* Target

Isocenter

0.3 m

dN

w

Figure a

* Target

Isocenter

w

Figure b

dN

* Target

Isocenter

Figure c

dN

w

Metal

0.3 m

0.3 m 0.3 m 0.3 m

Figure 2-6 Primary barrier width.

Figure 2-7 Effective field size.

Page 19: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

Shielding Techniques for Radiation Oncology Facilities26

NCRP 151 recommended that the distance from the target to thenarrow point of the primary barrier is measured at the top of the bar-rier, not at the same height as isocenter. This recommendation is notphysically meaningful when the primary barrier extends outwardfrom the secondary barrier (Figure 2-6a); otherwise it is an appropri-ate recommendation to follow for new construction. If the top of thebarrier is h meters above isocenter and the distance from the target tothe narrow point of the barrier measured horizontally is dNH meters,then dN is calculated as illustrated in Figure 2-8.

For existing construction, it is generally impractical to increasethe primary barrier width if the barrier is slightly too narrow, as cal-culated with the beam pointing at the top of the wall. Evaluating pri-mary barrier width with the beam pointed horizontally (for walls) orvertically (for ceiling or floor if applicable) is therefore pragmaticfor existing construction. The main criterion that determines whetherthe primary barrier width is adequate is the dose rate beyond theadjacent secondary barrier. A retrofit (if required) for existing con-struction is typically achieved by adding shielding to the adjacentsecondary barrier, not increasing primary barrier width.

Example: recommended primary barrier widthFigure 2-6a, with dN = 6.2 m

w = (35/100) × × 6.2 + 0.6 = 3.67 m

A slightly wider 3.96-m primary barrier width is recommended.

Figure 2-6b, new construction with dNH = 5.4 m, h = 2 m

dN = = 5.83 m

w = (35/100) × × 5.83 + 0.6 = 3.49 m

A slightly wider 3.66-m primary barrier width is recommended.

1)1( 22NHN dhd

Figure 2-8 Distance measured to top of primary barrier.

2

2 5 4 12 2( . )

2

Page 20: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

2. Conventional Shielding Design 27

However, situations requiring increased primary barrier width dooccur if the field extends very close to (or beyond) the edge of theprimary barrier. Examples include a change from the original treat-ment room design for isocenter or the plane of rotation. In such situ-ations, in addition to evaluating the adjacent secondary barrier, aprimary barrier calculation at the edge of the field may be appropri-ate. The next section provides a model that can be used in situationsin which the primary barrier shielding at the edge of the primary bar-rier is less than provided along the central axis.

Primary Beam Dose Rate vs. Beam Angle

Normally a primary barrier has constant thickness over the entirefield, with the highest dose rate beyond the barrier occurring on thecentral axis of the beam. If the barrier thickness or material changesnear the edge of the field, a separate calculation may be appropriateat that location. Table 2-4 provides a model for the primary beamdose rate relative to the central axis dose rate for use in this calcula-tion. The normal primary TVLs are conservatively assumed to apply.

The model in Table 2-4 is based on an 18-MV 21EX Varian lin-ear accelerator with a 40 × 40-cm field clocked at 45 degrees, withmeasurements made out to 22 degrees off the central axis (Potts2007). The data are then extrapolated out to 45 degrees, the mini-

Table 2-4 Primary dose rate relative central axis vs. beam angle

Angle Ratio

0 1.000

8 0.950

11 0.900

13 0.800

14 0.550

16 0.250

18 0.150

20 0.100

22 0.070

24 0.050

26 0.030

28 0.020

32 0.010

38 0.003

45 0.001

180 0.001

Page 21: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

Shielding Techniques for Radiation Oncology Facilities28

mum angle from the central axis for which the 0.1% leakage fractionis specified. As with other primary barrier calculations, the dose ratein Table 2-4 assumes no phantom (or patient) is located at isocenter.

Tapered Primary Barrier Ceiling

When the space above the accelerator vault may be occupied, theceiling must be shielded as a primary barrier. The weight and thick-ness required are more challenging for a ceiling than for a wallbecause of structural requirements. A tapered primary barrierreduces this weight by exploiting the fact that the required barrierthickness decreases with distance away from the point immediatelyabove isocenter. The reduced thickness is primarily caused by slantthickness, but it is also partially caused by the lower use factor (U =0.1) appropriate for gantry angles that are not a multiple of 90degrees. Typically tapering is recommended only if the primary bar-rier includes high-density shielding material to decrease its thick-ness.

In Figure 2-9, the “width” and “length” dimensions of individualceiling barrier sections are defined to be consistent with wall pri-mary barrier width. The term width refers to the dimension perpen-dicular to the plane of rotation. The term length refers to the

111

1 /tan vhR dd 11 cos/1 RPLR dd

Figure 2-9 Tapered primary ceiling barrier.

Page 22: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

2. Conventional Shielding Design 29

horizontal dimension in the plane of rotation. The thickness requiredfor the central section immediately above isocenter is identical to theconventional primary calculation described previously. The centralsection length (L0) is based on the beam being within plus or minusthe angle for central section length from vertical (i.e., near the verti-cal gantry angle for which U = 0.25 is applied). The minimum widthof the central section W0 must provide a 30-cm margin on each sidewith the beam pointed at the boundary between the central sectionand the first side section.

Typically, the width of the concrete in the primary ceiling barrierwill be selected to match the wall primary barrier, with any metalembedded in the concrete possibly having a narrower width. Thenext length LR1 can be arbitrarily selected based on the dimensionsof available material. An arbitrary number of segments can be used.The ceiling primary barrier must extend a sufficient distance toensure the wall provides adequate shielding with the beam pointedjust below the ceiling shielding. This may require the ceiling toextend to the far side of the wall primary barrier.

The shielded dose rate calculations for the side sections use thesame method as the central section, except with a potentially loweruse factor (e.g., U = 1/10) and increased distance dR1, dR2, etc. Inaddition, the barrier transmission calculations are based on the slantthickness of the barrier material (vertical material thicknesses alongthe path divided by cos( R1), cos( R2), etc.). As illustrated in Figure2-9, the angles i are calculated to provide the path with the mini-mum slant angle for a given section of material.

Obliquity Factor

For primary barriers, scatter can cause the barrier transmission to behigher than predicted based on slant factor alone. To compensate forthis, the slant thickness [t / cos( )] is divided by an obliquity factor(see Table A-7, with the result termed the oblique thickness. Theobliquity factors in Table A-7 are based on Biggs (1996), with val-ues at 45 degrees modified to comply with the NCRP 151 recom-mendation to add 2 HVL at low MV and 1 HVL at high MV for slantangles of 45 degrees and higher. Interpolation is used at MVsbetween the values given in Table A-7.

Groundshine Calculations

If a primary barrier is extremely thin (e.g., only of lead), scatteredradiation through a normal weight concrete floor may exceed thedirect radiation through the wall when the beam is pointed at the

Page 23: You can order this book in one of two formats: Hardcover ... .pdf · Report No. 151 (NCRP 2005b). Figure 2-1 shows a typical room plan for a high-energy (15 MeV) medical accelerator

©Medical Physics Publishing • For Personal Use Only

©Medical Physics Publishing • For Personal Use Only

Shielding Techniques for Radiation Oncology Facilities30

base of the wall. This radiation is referred to as groundshine. Asdepicted in Figure 2-10, groundshine is modeled by first determiningthe dose rate at Location X in the concrete floor beneath the wallalong a path from the x-ray target passing beneath the base of thewall. A primary barrier calculation is used to determine the dose rateat Location X. The protected location (denoted Z in Figure 2-10) isconsidered to be 0.3 m above floor level and 0.3 m beyond the wall,with the dose rate at Location Z calculated from the dose rate atLocation X using a secondary barrier scatter calculation.

The groundshine scatter calculation should use the maximumfield size, which covers a significant area beneath the wall, not a spe-cific point such as X or Y depicted in Figure 2-10. In reality thetransmission through the floor varies significantly over the field. Toassess the impact of the varying geometry over the field, ground-shine should be calculated at a number of Location X elevations cor-responding to the vertical extent of the field below the wall. Theaverage of these calculations is the calculated groundshine. LocationY (the point yielding the same calculated dose rate as the averageover the field) can then be used to illustrate the calculation.

Neutrons should not be a significant source of groundshinebecause the concrete floor should provide adequate neutron shield-ing.

0.3 m* Isocenter

0.3 m

Figure 2-10 Basic groundshine geometry.