Ultra-cryogenics for advanced resonant

detectors (C2)

G.Frossati

Kamerlingh Onnes LaboratoryLeiden, The Netherlands

Second ILIAS-GW Meeting October 24-25 2005Palma de Mallorca

www.minigrail.nl

Sensitivity

Minimum detectable energy

Teff=T/Q +2TN

N B NE = k T =1 h with = 3KHz

-7NT 1.5×10 K

Need~0.05; Q~2x106: T~30 mK

www.minigrail.nl

SQUID noise decreases with T

0.38 0

www.minigrail.nl

ROG SQUID

• Double SQUID made of a sensor d.c. SQUID, developed by the Institute of Photonic and Nanotechnologies of CNR, and a commercial Quantum Design d.c. SQUID as a preamplifier.

• Open input and in open loop configuration: 28hbar at 4.2 K and 5.5hbar at 0.9 K.

• Feedback network :37hbar in closed-loop open-input 4.2K. 

• With a high-Q resonant input load :70hbar at 2 K.

www.minigrail.nl

MiniGRAILCryogenics

www.minigrail.nlNext: 33 ton sphere ?Project together with the Roma

group to be submitted to INFN

hQL ≈4x10-24/√Hz

hQL≈4x10-23/√Hz

www.minigrail.nl

Cooling from 300K

Fill liquid helium

www.minigrail.nl

Forced flow

MiniGRAIL 1.3 +1 tonN2 1200 L800 L He

www.minigrail.nl

ROG-Leiden-Geneva project: to be submitted to INFN

Use 4 GM cryocoolers for pre-cooling from 300K to 20K

Use 4 PT410 for cooling to 4K and for the

dilution refrigerator

GM600 (Cryomech) 1200W at 250KPT400 60W at 60K and 1W at 4.2k

Heat to remove ~70J/gm ~2.8x109JCool-down time ~2 weeks to 4k

SFERA33 +7 ton20000 N2

14000 L He !

www.minigrail.nl

www.minigrail.nl

www.minigrail.nl

www.minigrail.nl

www.minigrail.nl

Dilution Refrigeration; Schematic view

CIRCULATIONSYSTEM

STILL

HEAT EXCHANGER

MIXING CHAMBER

STILLQ

HEAT ABSORPTION REGION

1k POT

HEAT FLOW

MCQ

STILLQ

TMC

www.minigrail.nl

Pulsed tube dilutionrefrigerator

Joule Thompson heat exchanger replaces the 1K pot: No need for Helium bath

JT Heat exchanger

PTR

Mixing Chamber

50K

4K

600 mK

60 mK

10 mK

www.minigrail.nl

www.minigrail.nl

www.minigrail.nlStrain sensitivity

2m CuAl6% sphere

f=1000 Hz

f=100 Hz

f=10 Hz

MiniGRAIL