www.chemometec.com counting of somatic cells in raw milk nucleocounter ® scc-100 tm
TRANSCRIPT
![Page 1: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/1.jpg)
www.chemometec.com
Counting of Somatic Cells in Raw Milk
NucleoCounter®SCC-100TM
![Page 2: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/2.jpg)
www.chemometec.com
NucleoCounter SCC-100
Stand-alone fluorescence microscope with built in computer for image analysis
Disposable sampling and measuring device, preloaded with nuclei staining dye
![Page 3: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/3.jpg)
www.chemometec.com
Easy operation
1. Lyse 2. Load 3. Count
![Page 4: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/4.jpg)
www.chemometec.com
Fast Result
+ = 30 seconds
= 30 seconds
------------------
max 1 minute
![Page 5: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/5.jpg)
www.chemometec.com
Base technology of ChemoMetec
Fluorescence Microscopy
![Page 6: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/6.jpg)
www.chemometec.com
Integrated Fluorescence Microscope
Lenses: magnification <1X.
Excitation filter: sorts out green light.
Emission filter: sorts out long wave length including red light.
LensesCCD
Emission Filter
LED’s
Excitation Filter
fixture
![Page 7: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/7.jpg)
www.chemometec.com
NucleoCassette
![Page 8: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/8.jpg)
www.chemometec.com
Propidium iodide (PI)
•Immobilized in the interior of the NucleoCassette, 2.8 ug in each
•A fluorescence dye, which intercalates with DNA of cell nuclei
•Cell membrane permeable
![Page 9: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/9.jpg)
www.chemometec.com
1 dye molecule per 4-5 bp
Genome between 2.5 x 109 – 4.0 x 109 bp. Propidium Iodide
DNA base pair (bp):
A: Adenine
T: Thymine
G: Guanine
C: Cytosine
C27H34N42+.2I- Mr 668.4
![Page 10: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/10.jpg)
www.chemometec.com
Reagent C-100• Disrupts the plasma membranes of the mammalian cells, rendering the nuclei
susceptible to staining with PI
![Page 11: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/11.jpg)
www.chemometec.com
Total cell count
Viable Non-viable
Free nucleus
PI
PI-stained nucleus
Lysis buffer
(Reagents C)
![Page 12: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/12.jpg)
www.chemometec.com
Sensitive parts,
moving parts and
“wear and tear” parts
have been taken out of the instrument
and integrated in the disposable NucleoCassette
- Pump- Tubing’s- Mixer- Precision optics- Chemicals- Calibration- Waste container
The NucleoCassette
![Page 13: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/13.jpg)
www.chemometec.com
Calibrated SCC-cassettes
W
T (
co
de
)
Dots for thickness code (T)
TAF0.01TAreaTWLVolume
![Page 14: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/14.jpg)
www.chemometec.com
Highly reliable LEDs
Thermal drift in optical intensity is approximately 0.2 %/°C
Optical intesity versus number of Number of analysis for the NucleoCounter SCC-100
0.90
0.95
1.00
1.05
1.10
0 250000 500000 750000 1000000
Number of analysis
Optica
l inte
nsi
ty (
a.u
.) v
Test conditions:t = 60°C, If(LED) = 100mASCC-100 Operating conditions max:t = 40°C, If(LED) = 30mA
![Page 15: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/15.jpg)
www.chemometec.com
Key Advantages•Simple Operation
–Ready in 5 seconds–5 minutes operator training–No cleaning
•Calibration Free–LED's, CCD and fixed optics assure stable factory calibration–Individually calibrated cassettes assure long term stability
•Maintenance Free–Robust and reliable parts used in design–2 years product guarantee
•Operator Independent–Objective cell counting
•Safe sample handling–PI immobilized inside NucleoCassette
![Page 16: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/16.jpg)
www.chemometec.com
Good reproducibility•Results from different NucleoCounters can be directly compared:
–All NucleoCounters are calibrated to respond as the same Master NucleoCounter at ChemoMetec.
–The initial calibration will maintain throughout the lifetime of a NucleoCounter.
![Page 17: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/17.jpg)
www.chemometec.com
Repeatability•Repeatability error dominated by count statistics (Poisson distribution) •CV = 1/ x 100%
CV% - NucleoCounter SCC-100
0.0%
2.0%
4.0%
6.0%
8.0%
10.0%
12.0%
14.0%
16.0%
18.0%
0 200 400 600 800 1000 1200
SCC
Nucle
oC
ounte
r SC
C-1
00 C
V%
NucleoCounter SCC-100 Poisson
![Page 18: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/18.jpg)
www.chemometec.com
Repeatability
•Repeatability*:
• - CV < 8% at 150,000 cells/ml• - CV < 6% at 300,000 cells/ml• - CV < 4% at 500,000 cells/ml• - CV < 3% at 1,000,000 cells/ml
•CV = Repeatability error Coefficient of Variance
![Page 19: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/19.jpg)
www.chemometec.com
Reproducibility•Comparison of two instruments from different production series - Factory Settings!
NucleoCounter SCC-100 B vs A
y = 1.0047x - 0.0285
0
200
400
600
800
1000
1200
0 200 400 600 800 1000 1200
NucleoCounter SCC-100 Inst. A
Nucle
oC
ounte
r SC
C-1
00 I
nst.
B
![Page 20: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/20.jpg)
www.chemometec.com
Reproducibility
•Only significant method contribution is repeatability–In test more than 95% of the intra-laboratory reproducibility explained by repeatability
![Page 21: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/21.jpg)
www.chemometec.com
Routine Milk Samples
![Page 22: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/22.jpg)
www.chemometec.com
Milk Application• SCC Stability
– The inherent stability of the NucleoCounter SCC-100 offers tools to monitor and assure stability of SCC results obtained by routine instruments
• Pilot Samples– The NucleoCounter SCC-100 allows fast and reliable determination of SCC pilot samples
![Page 23: Www.chemometec.com Counting of Somatic Cells in Raw Milk NucleoCounter ® SCC-100 TM](https://reader033.vdocuments.us/reader033/viewer/2022051401/56649db45503460f94aa4d8e/html5/thumbnails/23.jpg)
www.chemometec.com
Somatic Cell Counting• NucleoCounter SCC-100
– Stand-alone fluorescence microscope– Maintenance free– Calibration free
• NucleoCassette– Pre-loaded with nuclei staining dye– Analysed volume 1µl (working factor 1,000)– Calibrated measurement chamber (Dot code)
• Cell Counting– Operator independent– Reproducibility - Appr. Equal to Repeatability!– Repeatability Error - Appr. Poisson statistics
• (≈5% @ 500,000/ml)– Sample throughput – About 100 samples/hour– SCC - 10,000/ml to 1,000,000/ml