www.anton-paar.com product presentation c-ptd 180/air/qc air-cooled peltier temperature system...

15
www.anton-paar.com Product Presentation C-PTD 180/AIR/QC Air-cooled Peltier Temperature System 28.02.2013

Upload: kaley-lippincott

Post on 01-Apr-2015

329 views

Category:

Documents


4 download

TRANSCRIPT

Page 1: Www.anton-paar.com Product Presentation C-PTD 180/AIR/QC Air-cooled Peltier Temperature System 28.02.2013

www.anton-paar.com

Product Presentation C-PTD 180/AIR/QC

Air-cooled Peltier Temperature System

28.02.2013

Page 2: Www.anton-paar.com Product Presentation C-PTD 180/AIR/QC Air-cooled Peltier Temperature System 28.02.2013

www.anton-paar.com

Content

▸Pictures▸Features and Benefits▸Specification▸Performance▸Measurements▸Special Features▸Article Number and Price▸Configuration Example▸Applications and Markets▸Competition▸Summary

28.02.2013

Page 3: Www.anton-paar.com Product Presentation C-PTD 180/AIR/QC Air-cooled Peltier Temperature System 28.02.2013

www.anton-paar.com

Pictures P-CTD 180/AIR/QC

28.02.2013

Page 4: Www.anton-paar.com Product Presentation C-PTD 180/AIR/QC Air-cooled Peltier Temperature System 28.02.2013

www.anton-paar.com

Features and Benefits C-PTD 180/AIR/QC

▸ Temperature Range Complete range: 0 °C to 180 °C

▸ Measuring Systems Concentric Cylinder Measuring Systems

Disposable Measuring Systems

Stirrers and Spindles

▸ Air-cooled Peltier system No counter cooling necessary

▸ Typical Measurements Isothermal temperature tests

Temperature ramps

Temperature swing tests

28.02.2013

Page 5: Www.anton-paar.com Product Presentation C-PTD 180/AIR/QC Air-cooled Peltier Temperature System 28.02.2013

www.anton-paar.com

C-PTD 180/AIR/QC - Specification

• Temperature range 0 °C to 180 °C• Measuring systems Standard systems, DG42• Beaker Standard beaker, max. CC39 or DG42• General Good cleanability

Radiation system for counter cooling

Surrounding conditions• Temperature range 5 °C to 35 °C• Relative humidity, not condensing < 80 %• Altitude Up to 2000 m• Pollution Pollution degree 2 according to EN 61010• Surrounding Laboratory, indoor us only

Supply• Supply voltage 85 to 265 V• Frequency 50 … 60 Hz• Overvoltage Overvoltage category II according to

EN 6100

28.02.2013

Page 6: Www.anton-paar.com Product Presentation C-PTD 180/AIR/QC Air-cooled Peltier Temperature System 28.02.2013

www.anton-paar.com

Technical Specification – Heating Rates

28.02.2013

0

10

20

30

40

50

60

70

80

90

100

°C

T

10

100

1,000

10,000

mPa·s

0 5 10 15 20 25 30 35 40minTime t

C-LTD 0 - 80 °C (max) 1

T Temperature

Viscosity

C-PTD 0 - 80 °C (max) 2

T Temperature

Viscosity

Comparison Peltier / Liquid Temperature Control

Heating rate (Peltier): 8.0 °C/min

Heating rate (Liquid): 5.0 °C/min

Page 7: Www.anton-paar.com Product Presentation C-PTD 180/AIR/QC Air-cooled Peltier Temperature System 28.02.2013

www.anton-paar.com

Technical Specification – Cooling Rates

28.02.2013

-10

0

10

20

30

40

50

60

70

80

90

100

°C

T

10

100

1,000

10,000

mPa·s

0 20 40 60 80 100 120 140 160 180 220minTime t

C-LTD 80 - 0 °C (max) 1

T Temperature

Viscosity

C-PTD 80 - 0 °C (max) 3

T Temperature

Viscosity

Comparison Peltier / Liquid Temperature Control

Cooling rate (Peltier): 4.0 °C/min

Cooling rate (Liquid): 0.5 °C/min

Page 8: Www.anton-paar.com Product Presentation C-PTD 180/AIR/QC Air-cooled Peltier Temperature System 28.02.2013

www.anton-paar.com

Examples - Heating

-20

-10

0

10

20

30

40

50

60

70

80

90

100

110

120

130

140

150

160

170

180

200

°C

T

10-3

10-2

10-1

100

101

Pa·s

0 2 4 6 8 10 12 14 16 18 20minTime t

Anton Paar GmbH

Jump to 40 °C 2

CC39-SN5666; d=0 mm

T

Jump to 120 °C 2

CC39-SN5666; d=0 mm

T

Jump to 180 °C 2

CC39-SN5666; d=0 mm

T

Stable Conditions after approx. 15 min

28.02.2013

Page 9: Www.anton-paar.com Product Presentation C-PTD 180/AIR/QC Air-cooled Peltier Temperature System 28.02.2013

www.anton-paar.com

Examples - Cooling

-20

-10

0

10

20

30

40

50

60

70

80

90

100

110

120

130

140

150

160

170

180

200

°C

T

10-3

10-2

10-1

100

Pa·s

0 2 4 6 8 10 12 14 16 18 20minTime t

Anton Paar GmbH

Jump to 160 °C 2

CC39-SN5666; d=0 mm

T

Jump to 120 °C 6

CC39-SN5666; d=0 mm

T

Jump to 40 °C 3

CC39-SN5666; d=0 mm

T

Stable Conditions after approx. 15 min

28.02.2013

Page 10: Www.anton-paar.com Product Presentation C-PTD 180/AIR/QC Air-cooled Peltier Temperature System 28.02.2013

www.anton-paar.com

Special Features

• Wide temperature range: 0 °C to 180 °C

• Air cooled Peltier temperature system

• Rapid changes in temperature

• No external controller necessary

• No counter cooling by an external circulator

• Easy to install

• Space saving design

• Noise reduced

28.02.2013

Page 11: Www.anton-paar.com Product Presentation C-PTD 180/AIR/QC Air-cooled Peltier Temperature System 28.02.2013

www.anton-paar.com

Article Number

Item-no.108642

C-PTD 180/AIR/QCPeltier Temp. Control SystemTemp. range: 0 to 180 °C

28.02.2013

Page 12: Www.anton-paar.com Product Presentation C-PTD 180/AIR/QC Air-cooled Peltier Temperature System 28.02.2013

www.anton-paar.com

Standard Configuration

18318 Basic Instrument RheolabQC

108642 C-PTD 180/AIR/QC

20654 Disposable Measuring System CC39/D/QC-LTD

20655 Disposable Measuring System CC27/D/QC-LTD

Option:

18385 Software Rheoplus

28.02.2013

Page 13: Www.anton-paar.com Product Presentation C-PTD 180/AIR/QC Air-cooled Peltier Temperature System 28.02.2013

www.anton-paar.com

Applications & Markets

Coatings:Flow Curve, Yield Point,

3ITT (Thixotropy)

Food:Flow Curve, Yield Point,

Temperature Test,

3ITT (Thixotropy)BuildingMaterials:Flow Curve, Yield Point,

3ITT (Thixotropy)

28.02.2013

Page 14: Www.anton-paar.com Product Presentation C-PTD 180/AIR/QC Air-cooled Peltier Temperature System 28.02.2013

www.anton-paar.com

Applications & Markets

Adhesives: Flow Curve, 3ITT (Thixotropy),

Temperature Test

Cosmetics: Flow Curve, Yield Point,

Temperature Swing Test

Pharma: Flow Curve, Yield Point,

Temperature Swing Test

Soft Gels: Yield Point, 3ITT (Thixotropy)

Lubricants,Oils:

Flow Curve, Temperature Test

Asphalt: Flow Curve, Temperature Test

28.02.2013

Page 15: Www.anton-paar.com Product Presentation C-PTD 180/AIR/QC Air-cooled Peltier Temperature System 28.02.2013

www.anton-paar.com

Summary

C-PTD 180/AIR/QC is the new Peltier temperature system for the RheolabQC with:

▸ Wide temperature range: 0 °C to 180 °C▸ Air cooled Peltier temperature system▸ Cheaper configuration compared with the liquid temp. control systems▸ Compact design, no counter cooling by an external circulator

▸ Easy handling▸ Small footprint▸ Noise reduced compared to a fluid circulator ▸ Low power consumption▸ Fast and accurate temperature setting▸ no thermal fluids/oil necessary

▸ Firmware V1.26 for the RheolabQC required

28.02.2013