word: neurological criticool operating...
TRANSCRIPT
NEUROLOGICAL: CRITICOOL: OPERATING PROCEDURE
DOCUMENT TYPE: PROCEDURESite Applicability
Management of Hypoxic Ischemic Encephalopathy (HIE) occurs in the NICU
Guideline StatementsThe Physician orders CritiCool™ as a primary route of therapeutic hypothermia or therapeutic normothermia for the treatment of infants with Hypoxic Ischemic Encephalopathy (HIE).
Procedure Procedure Table of Contents:
Pg. 1: PreparationPg. 4: ApplicationPg. 6: X-RayPg. 7: Re-WarmingPg. 8: MRIPg. 9: Post ProcedurePg. 9: Documentation
Gather Equipment: CritiCool™ with CliniLogger (with or without splitter attached) Hydrocolloid dressing Waterproof tape Lubricating jelly (water soluble) Rectal temperature probe #9 Fr Skin temperature sensor Thermowrap jacket Sterile water (6 litres) If using CritiCool™ with splitter: Philips monitor temperature cable
Preparation Note1. Obtain CritiCool™ device with water hoses
and clinilogger attached Clinilogger should be plugged in at the back of the
CritiCool™
2. Bring CritiCool™ and appropriately sized Thermowrap jacket to bedside
3. Plug in CritiCool™ and splitter power
See table above
C-06-12-60207 Published Date: 20-Jun-2017Page 1 of 10 Review Date: 20-Jun-2020 This is a controlled document for BCCH& BCW internal use only – see Disclaimer at the end of the document. Refer to online version as the print copy may not be current.
NEUROLOGICAL: CRITICOOL: OPERATING PROCEDURE
DOCUMENT TYPE: PROCEDURE
cables
4. Fill the reservoir with 6 litres of sterile water
Do not use tap water Do not over fill Fill to level between two red markings
5. Plug Green sensor into Surface socket of CritiCool™ (if not already plugged in)
A).If not using splitter: Plug GRAY sensor into Core
socket of CritiCool™B).If using splitter:
Plug GRAY sensor into Core socket on splitter
Plug temperature monitoring cable into Philips monitor and set temperature limits to 33.0 and 37.5 degrees Celsius
Change Philips monitor screen to CritiCool™ mode
Criticool Splitter:
6. Position the Thermowrap in the bed so the water tubing connectors are positioned above the infants head.
Connect water tubes to the Thermowrap.
The tubing/clamps need to lay flat on the mattress to prevent kinking
Put the tubing through the slot at the head of the bed, not through the side ports to prevent kinking
Check clamps are open The Thermowrap will automatically fill once the rectal
probe is in place
7. Turn on Machine will alarm for self-test and
errors
8. Select “COOLING” for both Therapeutic Hypothermia and Therapeutic Normothermia by pressing ok
The water level may fall below the two red lines Do not top up water By using the “cooling” mode the variance of
temperature will be +/- 0.2 ºC of the set point
9. Perform safety check prior to use to ensure settings are correct
1. Press Menu2. Press Settings3. Press Enter
Clinilogger should be plugged in at the back of the CritiCool™
C-06-12-60207 Published Date: 20-Jun-2017Page 2 of 10 Review Date: 20-Jun-2020 This is a controlled document for BCCH& BCW internal use only – see Disclaimer at the end of the document. Refer to online version as the print copy may not be current.
GRAY sensor
Temperature monitoring cable
NEUROLOGICAL: CRITICOOL: OPERATING PROCEDURE
DOCUMENT TYPE: PROCEDURE
4. Ensure values are entered as below:Patient *NeonatalAdult SP *33.0Delta T *0.45. Press Exit
The ‘COOLING’ Mode set temperature is 33.5ºC
10. For Therapeutic Hypothermia pre-set temperature is 33.5ºC
11. For Therapeutic Normothermia set temperature on CritiCool™ to 36.7ºC by using the up and down arrows
Check hourly that the set point is at the desired temperature (33.5 ºC for therapeutic hypothermia and 36.7 ºC for therapeutic normothermia)
12. Place CritiCool™ in ‘standby’ until infant arrives
Select ‘menu’ Scroll down and select ‘standby’ Select ‘enter’
C-06-12-60207 Published Date: 20-Jun-2017Page 3 of 10 Review Date: 20-Jun-2020 This is a controlled document for BCCH& BCW internal use only – see Disclaimer at the end of the document. Refer to online version as the print copy may not be current.
NEUROLOGICAL: CRITICOOL: OPERATING PROCEDURE
DOCUMENT TYPE: PROCEDURE
Application Note13. Lay infant, wearing only a diaper, supine
on Thermowrap Align cut outs for arms with the axilla Align lower flaps with the upper thighs Do not wrap infant yet
Infant must be repositioned according to skin risk assessment
Prone position is contra-indicated
14. Obtain rectal temperature probe (#9 Fr) and measure 5 cm from tip and mark with pen
15. Lubricate rectal temperature probe and insert 5cm into rectum. Connect to Gray sensor
Use water soluble jelly Secure at the rectum with hydrocolloid dressing
and waterproof tape
16. Secure skin temperature sensor and connect to Green sensor
Use skin temperature sensor Choose a site as close to core of infant as possible This sensor must not come in contact with the
Thermowrap (e.g. upper chest, thigh) Clean site with water, dry skin and apply sensor
17. Take CritiCool™ off ‘standby’ mode Select ‘Menu’ Select ‘Operate’
Check temperature is set at (33.5 ºC for Therapeutic Hypothermia and 36.7 ºC for Therapeutic Normothermia)
18. Monitor as the Thermowrap automatically fills
Correct any alarm conditions identified Correct any kinking or leaks Allow Thermowrap to fill with water prior to
wrapping around the infant
19. Once Thermowrap is filled, wrap around the infant’s torso and upper thighs
*Do not use safety pins on any part of the thermowrap: extra securing tapes are available if needed*
20. Fold Thermowrap as needed to expose areas of infant
Ensure there is no kinking obstructing water flow in the system
Folding the Thermowrap may be done to expose certain areas of the infant (e.g. umbilical lines)
If Thermowrap is kinked and obstructing water flow alarm ‘Check Water Connections’ (see troubleshooting guide)
21. Check the radiant warmer is off
C-06-12-60207 Published Date: 20-Jun-2017Page 4 of 10 Review Date: 20-Jun-2020 This is a controlled document for BCCH& BCW internal use only – see Disclaimer at the end of the document. Refer to online version as the print copy may not be current.
NEUROLOGICAL: CRITICOOL: OPERATING PROCEDURE
DOCUMENT TYPE: PROCEDURE
22. Assess skin integrity Utilize Skin Risk Assessment and provide corresponding care activities based on risk identified
Cooled infants are at increased risk for fat necrosis and require frequent position changes and skin assessment
Monitor carefully for moisture between the skin and Thermowrap (moisture can contribute to increased risk)
If mildly soiled clean Thermowrap with water and dry If extremely soiled or punctured, change Thermowrap
23. Parents may hold infant with Thermowrap in place to maintain cooling
C-06-12-60207 Published Date: 20-Jun-2017Page 5 of 10 Review Date: 20-Jun-2020 This is a controlled document for BCCH& BCW internal use only – see Disclaimer at the end of the document. Refer to online version as the print copy may not be current.
NEUROLOGICAL: CRITICOOL: OPERATING PROCEDURE
DOCUMENT TYPE: PROCEDURE
X-ray Note24. Loosen the wrap and slip the x-ray plate
between the infant and wrap The Thermowrap shows shadows on any x-ray
C-06-12-60207 Published Date: 20-Jun-2017Page 6 of 10 Review Date: 20-Jun-2020 This is a controlled document for BCCH& BCW internal use only – see Disclaimer at the end of the document. Refer to online version as the print copy may not be current.
NEUROLOGICAL: CRITICOOL: OPERATING PROCEDURE
DOCUMENT TYPE: PROCEDURE
Re-warming Notes25. To re-warm:
Press menu key Scroll to “MODE SELECT” Press enter Select “CONTROLLED REWARM (for
slow rewarming)” Press “OK” Machine makes audible tone Message “Switching to Re-Warming”
While the screen is visible rewarming has not started (see picture on right)
Check rectal probe is correctly placed Press “Operate”. Ensure set
temperature is 36.5 Screen will read “NEONATAL
REWARMING”
CritiCool™ will increase the Set Point Temperature by 0.40C every 60 minutes
Re-warming to 36.50C will take 7-8 hours
26. Continue therapy on the CritiCool™ device for 24 hours after the infant’s temperature has reached 36.50C
Once the infant’s temperature reaches 36.50C, the mode display will stay in “NEONATAL REWARMING”.
C-06-12-60207 Published Date: 20-Jun-2017Page 7 of 10 Review Date: 20-Jun-2020 This is a controlled document for BCCH& BCW internal use only – see Disclaimer at the end of the document. Refer to online version as the print copy may not be current.
NEUROLOGICAL: CRITICOOL: OPERATING PROCEDURE
DOCUMENT TYPE: PROCEDURE
MRI Note27. Put CritiCool™ on standby Refer to MRI checklist
The Thermowrap will automatically drain
28. Allow the Thermowrap to drain and clamp water tubes
Draining takes approximately 5 minutes
29. Disconnect water tubes from Thermowrap
30. Remove skin probe from infant Leave connected to CritiCool™ for reapplication after MRI
31. Without splitter: Disconnect rectal probe from GRAY CritiCool™ core cable and connect to Philips monitor temperature cable and plug into X2 transport monitor
32. With splitter: Disconnect rectal probe from GRAY CritiCool™ core cable on splitter and connect to Philips monitor temperature cable
33. Once infant is transported to MRI remove rectal probe
Probes are not MRI compatible Refer to MRI checklist for thermomanagement
34. Unwrap infant from Thermowrap and transfer care to MRI team
35. Upon return to NICU, reconnect water tubes to Thermowrap
36. Unclamp water tubes
37. Take CritiCool™ off ‘standby’ mode Select ‘Menu’ Select ‘Operate’
38. Verify correct mode: For infants being actively cooled
place back in ‘COOLING’ mode For infants who have completed
rewarming place back in ‘NEONATAL REWARMING’ mode.
CritiCool™ will start from the infants current core temperature (i.e. if putting CritiCool™ back in the re-warm mode after MRI the infant will be warmed from its current temperature to reach the set temperature)
Infants must complete 24 hours in ‘NEONATAL REWARMING’ mode after they have been rewarmed.
C-06-12-60207 Published Date: 20-Jun-2017Page 8 of 10 Review Date: 20-Jun-2020 This is a controlled document for BCCH& BCW internal use only – see Disclaimer at the end of the document. Refer to online version as the print copy may not be current.
NEUROLOGICAL: CRITICOOL: OPERATING PROCEDURE
DOCUMENT TYPE: PROCEDURE
Post Procedure Note39. Turn the CritiCool™ Device to ‘standby’ Wait while the water drains back into the CritiCool™
Disconnecting the unit prematurely will cause water leakage
40. Close the clamps on the Thermowrap
41. Turn off CritiCool™
42. Disconnect the water tubes from the Thermowrap.
43. Unwrap the infant and discard Thermowrap
Thermowraps are individual use items and are discarded after each infant
44. Assess infant’s skin after removal of Thermowrap
A pattern resembling the Thermowrap may appear for a short period of time on the infant’s skin
45. Remove the rectal and skin temperature probes and disconnect from the CritiCool™ unit
Gray sensor and Green sensor are not disposable
46. Save stored data from CliniLogger Remove CliniLogger and ensure labelled with patient label
Leave in CNL office at CRN desk
47. Return the CritiCool™ unit, with all cables and tubing’s connected, to the aides room
The grey and green cables and water hoses are NOT disposable
Aides will empty and clean CritiCool™ machine
DocumentationNursing Flowsheet:
Hourly skin and rectal temperature Hourly CritiCool™ ‘set point’ Skin integrity Interruptions to cooling
Physician’s Orders: Date and time cooling/ re-warming started
Documentation Nursing FlowsheetBedside Information Tool (BIT)Troubleshooting GuidelinePhysician’s Orders – Therapeutic Normothermia AdmissionPhysician’s Orders – Therapeutic Hypothermia Admission
References CritiCool™ Operating Manual
C-06-12-60207 Published Date: 20-Jun-2017Page 9 of 10 Review Date: 20-Jun-2020 This is a controlled document for BCCH& BCW internal use only – see Disclaimer at the end of the document. Refer to online version as the print copy may not be current.
NEUROLOGICAL: CRITICOOL: OPERATING PROCEDURE
DOCUMENT TYPE: PROCEDURE
Version HistoryDATE DOCUMENT NUMBER and TITLE ACTION TAKEN16-May-2019
C-06-12-60207 Neurological: Criticool: Operating Procedure
Updated by: Neonatal Critical Care Committee Temperature for normothermia adjusted from 36.0 to 36.7
04-May-2020
“ Updated by Neonatal Program Senior Practice Leader Rationale in Step 7 Page 2 removed due to software update
DisclaimerThis document is intended for use within BC Children’s and BC Women’s Hospitals only. Any other use or reliance is at your sole risk. The content does not constitute and is not in substitution of professional medical advice. Provincial Health Services Authority (PHSA) assumes no liability arising from use or reliance on this document. This document is protected by copyright and may only be reprinted in whole or in part with the prior written approval of PHSA.
C-06-12-60207 Published Date: 20-Jun-2017Page 10 of 10 Review Date: 20-Jun-2020
This is a controlled document for BCCH& BCW internal use only – see Disclaimer at the end of the document. Refer to online version as the print copy may not be current.