HVAD ® System Written Exam for Clinical Staff 1. Match the word on the right with the correct function: a) This icon, located at the front of the controller, lights up when one or more alarms occurs __c __ AC Adapter b) This button silences the low and medium audible alarms __a__ Alarm Indicator c) This adapter/cable transfers power from an electric outlet to run the controller __b__ Alarm Mute Button 2. Label the cable connections to the controller: a. Power Source Connection b. Power Source Connection c. Driveline Connection d. Monitor Cable (Data)/ Alarm Adapter Connection Choose True or False 3. The flashing red visual alarm on the controller requires immediate action True False 4. It is ok to have moisture, cracks, tears or punctures in the driveline True False Page 1 of 9 GL1091 Rev01 11/14 a b d c

Upload: others

Post on 11-Jan-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1:  · Web viewMatch the word on the right with the correct function: This icon, located at the front of the controller, lights up when one or more alarms occurs __c__ AC Adapter

HVAD® System

Written Exam for Clinical Staff

1. Match the word on the right with the correct function:

a) This icon, located at the front of the controller, lights up when one or more alarms occurs

__c__ AC Adapter

b) This button silences the low and medium audible alarms

__a__ Alarm Indicator

c) This adapter/cable transfers power from an electric outlet to run the controller __b__ Alarm Mute Button

2. Label the cable connections to the controller:

a. Power Source Connectionb. Power Source Connection

c. Driveline Connectiond. Monitor Cable (Data)/ Alarm Adapter

Connection

Choose True or False

3. The flashing red visual alarm on the controller requires immediate action True False

4. It is ok to have moisture, cracks, tears or punctures in the driveline True False

5. It is safe for a patient to take a tub bath or swim True False

Page 1 of 7

GL1091 Rev01 11/14

a b

dc

Page 2:  · Web viewMatch the word on the right with the correct function: This icon, located at the front of the controller, lights up when one or more alarms occurs __c__ AC Adapter

HVAD® System

Written Exam for Clinical Staff

6. This button is used to see all active alarms and pump information on the controller display

True False

7. Pushing the ALARM MUTE BUTTON to silence an alarm does not resolve the alarm condition

True False

8. A controller with a blank display and no audible alarm should be replaced True False

9. The controller needs to be disconnected from the driveline before defibrillation or cardioversion

True False

10. A properly functioning battery charger may take up to 10 hours to fully charge a depleted battery

True False

11. The controller AC/DC indicator will be green if the AC Adapter or DC Adapter (car

adapter) is powering the controller

True False

12. The HeartWare® Monitor has an internal battery that can be used to power the monitor while transporting a patient

True False

13. HVAD pump flow is preload dependent and afterload sensitive True False

14. List 2 items the patient should take when he/she leaves the house:_________________________________ Answer: charged batteries, back up controller, emergency contact sheet, cell phone

15. How long does 1 fully charged battery usually provide power? ____________________________________________Answer: 4 to 6 hours

16. Why do LVAD patients take anticoagulation medicine?_________________________________ Answer: To reduce the risk of clots in blood and pump

Page 2 of 7

GL1091 Rev01 11/14

Page 3:  · Web viewMatch the word on the right with the correct function: This icon, located at the front of the controller, lights up when one or more alarms occurs __c__ AC Adapter

HVAD® System

Written Exam for Clinical Staff

17. When looking at the controller, how much power is remaining in a battery if there are:

4 green lights ______ % Answer: 75-100%

3 green lights ______ % Answer: 50-74%

2 yellow lights ______ % Answer: 25-49%

1 red light ______ % Answer: < 24%

18. Place a check beside activities that will cause the HVAD pump to stop:

___ Disconnecting the AC adapter while only one fully charged battery remains connected to the controller

__ Disconnecting both batteries at the same time

__ Disconnecting the driveline from the controller

__ Unplugging the DC adapter without a back-up battery connected to the controller

19. Two power sources should be connected to the controller at all times. What are the 3 different power options?

Battery and ____________________ Answer: BatteryBattery and ____________________ Answer: AC AdapterBattery and ____________________ Answer: DC Adapter

20. Place a check beside activities that are not allowed while on the HVAD pump:

___ Chest x-ray

__ MRI scan

___ CT scan

___ Cardiac catheterization

__ Swimming

___ Walking

21. What is the recommended speed range for operation of the HVAD pump? ___________ RPM to __________ RPMAnswer: 2400 to 3200

22. Waveforms normally have a positive deflection. During normal operation there should be _________L/min of pulsatility (from trough to peak) and the trough level should be at least __________L/min.

Answer: >2 L/min and >2L/min

Page 3 of 7

GL1091 Rev01 11/14

Page 4:  · Web viewMatch the word on the right with the correct function: This icon, located at the front of the controller, lights up when one or more alarms occurs __c__ AC Adapter

HVAD® System

Written Exam for Clinical Staff

23. Please put the steps below in the correct order (1-9) to demonstrate the sequence for changing the controller:

4 Pull back the white driveline cover from the original controller’s silver connector

1 Have the patient sit or lie down

5 Disconnect the driveline from the original controller (this will stop the pump)

7 Insert the red alarm adapter into the blue connector on the original controller

2 Place the new controller within easy reach

3 Connect back-up power sources to the new controller

9 Slide the white driveline cover up to cover new controller’s silver connector

6 Connect the driveline to the new controller (this will restart the pump)

8 Disconnect both power sources from the original controller

24. Put the following actions in the correct order to switch from battery operation to tethered AC power.

_3__ Connect AC power by lining up the solid white arrow on the cable connector with the white dot on the controller

_4__ Confirm that the power cable is properly locked to the controller by gently pulling on the cable near the connector

_1__ Plug the AC adapter into an electrical outlet

_2__ Disconnect the battery with the least remaining charge

25. One power source indicator will light up based on which port is providing primary power. Which statements are correct for the image below? Check all that apply

____ Power source #2 is the port providing primary power

___ The AC or DC adapter is providing power

___ Power source #1 is the port providing primary power

Page 4 of 7

GL1091 Rev01 11/14

Page 5:  · Web viewMatch the word on the right with the correct function: This icon, located at the front of the controller, lights up when one or more alarms occurs __c__ AC Adapter

HVAD® System

Written Exam for Clinical Staff

26. What is the procedure to clear a resolved, inactive medium priority alarm from remaining on the display screen?

___ Hold down the alarm mute button for 5 seconds to clear the alarm___ No action is needed, alarm clears itself___ Change out the controller__ Press the scroll to clear the alarm from the controller

27. Match the battery charger light to the correct battery status:

Green Battery ready for use and fully charged

Yellow Battery too cold or too hot, waiting to charge

Flashing Yellow Defective battery. Do not use

Red Battery not charging. Check connections or switch battery to another slot. If persists do

not use

Flashing Red Battery being charged, not ready for use

28. To minimize the risk of interference (electrostatic discharge) to a controller due to static electricity, which of the following should patients avoid? Check all that apply

____ Putting controller in a silk bag

_____ Using dryer sheets when doing laundry

_____ Using a humidifier in the home

____ Touching electronic devices (TV screens, microwaves in operation, laptop/computer screens)

____ Operating on one battery while going to the restroom in the middle of the night

_____ Ensuring that the driveline cover is firmly positioned against the controller

Page 5 of 7

GL1091 Rev01 11/14

Page 6:  · Web viewMatch the word on the right with the correct function: This icon, located at the front of the controller, lights up when one or more alarms occurs __c__ AC Adapter

HVAD® System

Written Exam for Clinical Staff

29. Your patient reports a sudden drop in battery capacity. They noticed the battery indicator on the controller went from 4 lights to 1 light. Circle the appropriate action?

a. Replace the abnormally behaving battery and remove from service

b. Monitor the battery over time to see if it does it again

c. Do nothing, this can happen from time to time

30. Which of the following conditions may lead to a decrease in preload? Check all that apply.

____ Compromised right ventricular volume (e.g. RVF, Tamponade)____ Low intravascular volume _____ Increased system vascular resistance ____ Increase in pump speed (a volume/speed mismatch)

31. Check the statements that are correct about an electrical fault (EF) alarm, which when active, the controller will display [ELECTRICAL FAULT, CALL], with a flashing yellow alarm indicator and.audible tone Check all that apply.

___ An EF alarm indicates that the pump is running on a single stator____ The EF alarm cannot be silenced___ Do not change the controller unless instructed to on the controller____ Inspection of equipment for defects (e.g. controller, driveline) is not necessary

32. Which waveform represents a suction event?

Page 6 of 7

GL1091 Rev01 11/14

Page 7:  · Web viewMatch the word on the right with the correct function: This icon, located at the front of the controller, lights up when one or more alarms occurs __c__ AC Adapter

HVAD® System

Written Exam for Clinical Staff

________

_______

______

WARNING: Serious and life-threatening adverse events, including stroke, have been associated with use of this device. A user must fully consider the risks of this device with that of other treatment modalities before deciding to proceed with device implantation.

For full prescribing information please see the Instructions for Use (IFU). The IFU can be found at www.heartware.com/clinicians/intructions -use .

In the USA the HVAD system is intended for use as a bridge to cardiac transplantation in patients who are at risk of death from refractory end-stage left ventricular heart failure.

CAUTION: Federal law (USA) restricts this device to sale by or on the order of a physician. Refer to the “Instructions for Use” for complete indications for Use, Contraindications, Warnings, Precautions, Adverse Events and Instructions prior to using this device.

Page 7 of 7

GL1091 Rev01 11/14