wavewin sniffer 24 manual-4

Upload: vasek

Post on 07-Aug-2018

240 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    1/86

     

    WAVEWIN SNIFFER 24

    Disturbance Monitor  

    Operation & MaintenanceGuide 

    SOFTSTUF, INC.SOFTWARE STRUCTURES FOR UNLIMITED FUNCTIONALITY 

    P.O. BOX 40245PHILADELPHIA, PA 19106-0245

    1-800-818-3463 · 215-922-6880www.softstuf.com www.wavewin.net

    TM

    http://www.softstuf.com/http://www.softstuf.com/

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    2/86

     

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    3/86

     

    Copyright © 1992-2012 by SoftStuf, Inc. All rights reserved.Printed in the United States of America.

    Last Update: 05 / 24 / 2012

     Acknowledgments

    Wavewin is a registered trademark of SoftStuf Incorporated.Windows is a registered trademark of Microsoft Corporation.

     All other products and brand names are trademarks or registered trademarks of their respective holders.

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    4/86

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    5/86

     

    PrefaceThis operations and maintenance guide provides detailed descriptive information and user instructions foroperating the Wavewin Sniffer 24 Disturbance Monitoring (Sniffer) device. This document is intended foruse by individuals working in protection, engineering, and system operations.

     A detailed technical description of the Sniffer and operational examples are included.

    The manual is divided into the following Chapters:

    Chapter 1: Introduction General Information

    Chapter 2: Specifications Hardware Specifications

    Chapter 3: Installation Installing the Hardware and Software

    Chapter 4: System Software Software Applications Description

    Chapter 5: System Settings General and Channel Settings Information

     Appendix A: System Keys Shortcut Keys and Button Functionality

     Appendix B: Hardware Specifications Sensor Specifications

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    6/86

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    7/86

     

    Table of Contents 

    CHAPTER 1 ............................................................................................................................... 1 

    Introduction ........................................................................................................................... 1 Overview ............................................................................................................................. 1

     

    Sensors ............................................................................................................................... 3 Event Reports ..................................................................................................................... 3

     

    Measurements and Calculations ......................................................................................... 4 

    Waveform Captures and Triggers ........................................................................................ 5 

     Applications ......................................................................................................................... 5 Features .............................................................................................................................. 5 

    CHAPTER 2 ............................................................................................................................... 7 

    Specifications ....................................................................................................................... 7 Processor Specifications ..................................................................................................... 7 Recorder Specifications ....................................................................................................... 7 Optional Equipment ............................................................................................................. 7

     

    GPS Satellite Clock: .......................................................................................................................... 7 

    IRIG-B ............................................................................................................................................... 8 

    Surge Protector ................................................................................................................................. 8 Monitor / Keyboard: ........................................................................................................................... 8 Current Sensors ................................................................................................................................ 8 Voltage Sensors ................................................................................................................................ 9 Other Sensors ................................................................................................................................. 10

     

    CHAPTER 3 ............................................................................................................................. 11 

    Installation ........................................................................................................................... 11 Installing the Recorder ....................................................................................................... 11 Connecting the Sensors .................................................................................................... 11 Upgrading the Software ..................................................................................................... 11

     

    Running the Software ........................................................................................................ 13 

    Version Information ........................................................................................................... 13 

    Technical Support ............................................................................................................. 14 

    CHAPTER 4 ............................................................................................................................. 15 

    System Software ................................................................................................................. 15 Configuration and Polling .................................................................................................. 15 

    Device Configuration ....................................................................................................................... 15 

    IEEE Long File Naming Format ...................................................................................................... 16 

    Channel Settings ............................................................................................................................. 17 

    Channel Titles ................................................................................................................................. 19 

    Configuration Tab ........................................................................................................................... 19 

    Triggering Tab ................................................................................................................................. 20 

    Calibration Tab ................................................................................................................................ 21 

    Data Monitor.................................................................................................................................... 22 Configuration File ............................................................................................................................ 23 File Properties ................................................................................................................................. 23 

    File management and Analysis .......................................................................................... 25 Event and SOE Formats ................................................................................................................. 25

     

    File Table ........................................................................................................................................ 26 

    SOE Files ........................................................................................................................................ 27 

    Event Summary .............................................................................................................................. 27 

    Event Analysis ................................................................................................................................ 29 

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    8/86

    Navigating ....................................................................................................................................... 30 Setting the Cursor Bars ................................................................................................................... 31 Data Bar .......................................................................................................................................... 31

     

    Reference Bar ................................................................................................................................. 31 

    RMS Bar .......................................................................................................................................... 31 

    Fault Bar .......................................................................................................................................... 32 

    Horizontal Bars ............................................................................................................................... 32 

    Marking, Deleting, and Restoring Channels ................................................................................... 32 

    Scaling Analog Channels ................................................................................................................ 32 Zooming Channels .......................................................................................................................... 33 

    Viewing Analog Data ....................................................................................................................... 33 Viewing Digital Data ........................................................................................................................ 36 

    CHAPTER 5 ............................................................................................................................. 37 

    System Settings .................................................................................................................. 37 General Device Settings .................................................................................................... 37 Calibrating the Sensors ..................................................................................................... 38 

    Manual Calibration .......................................................................................................................... 39 

     Automatic Calibration ...................................................................................................................... 41 

    Calibration Fields ............................................................................................................................ 42 

    Setting Triggers ................................................................................................................. 43 

    Current Detection ............................................................................................................................ 44 

    APPENDIX-A ............................................................................................................................. 1 

    System Keys ......................................................................................................................... 1 Sniffer Configuration & Polling ............................................................................................. 1

     

    File Manager ....................................................................................................................... 2 

    Query Fields ........................................................................................................................ 3  Analysis ............................................................................................................................... 3 

    APPENDIX-B ............................................................................................................................. 1 

    HARDWARE SPECIFICATIONS ............................................................................................ 1 Clothespin Sensor (CS-HE-CPL) ......................................................................................... 1

     

    Output Signals .................................................................................................................................. 1 

    Specifications .................................................................................................................................... 1 

    Split-Core CT Sensor (CS-SC-200) ..................................................................................... 3 

    Output Signals .................................................................................................................................. 3 Specifications .................................................................................................................................... 3 

    Solid-Ring CT Sensor (CS-SR-150) .................................................................................... 5 Output Signals .................................................................................................................................. 5

     

    Specifications .................................................................................................................................... 5 

     AC/DC Voltage Sensor (VS-OA-500) .................................................................................. 7 

    Output Signals .................................................................................................................................. 7 

    Specifications .................................................................................................................................... 7 

     AC Voltage Sensor (VS-VT-600) ......................................................................................... 9 

    Output Signals .................................................................................................................................. 9 

    Specifications .................................................................................................................................... 9 

    Temperature Sensor (TS-LM-212)..................................................................................... 11 Output Signals ................................................................................................................................ 11

     

    Specifications .................................................................................................................................. 11 

    Dry Contact Sensor (DCS-01/5) ........................................................................................ 13 

    Output Signals ................................................................................................................................ 13 

    Specifications .................................................................................................................................. 13 

    IRIG-B (CBL-IRIG-B) ......................................................................................................... 15 Output Signals ................................................................................................................................ 15 

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    9/86

     

    Specifications .................................................................................................................................. 15 

    LIST OF FIGURES

    FIGURE 1.1  – PORTABLE UNIT  ....................................................................................................................... 1 

    FIGURE 1.2  – R ACK MOUNT UNIT ................................................................................................................... 1 

    FIGURE 1.3  – DIFFERENTIAL ANALOG INPUTS WIRING DIAGRAM ...................................................................... 2 

    FIGURE 1.4  – REAR P ANEL  ............................................................................................................................ 2 

    FIGURE 1.5  – TRIGGER PROPERTIES .............................................................................................................. 4 

    FIGURE 3.1  – RJ45 CONNECTORS ............................................................................................................... 11 

    FIGURE 3.2  – START SOFTWARE INSTALLATION ............................................................................................ 12 

    FIGURE 3.3  – CREATE INSTALL P ATH  ........................................................................................................... 12 

    FIGURE 3.4  – FINISH SOFTWARE INSTALLATION ............................................................................................ 13 

    FIGURE 3.5  –  ABOUT WINDOW ..................................................................................................................... 14 

    FIGURE 4.1  – CONFIGURATION AND POLLING SOFTWARE .............................................................................. 15 

    FIGURE 4.2  – DEVICE CONFIGURATION FIELDS ............................................................................................. 16 

    FIGURE 4.3  – CHANNEL SETTINGS  ............................................................................................................... 17 

    FIGURE 4.4  – CONFIGURATION T AB .............................................................................................................. 20 

    FIGURE 4.5  – TRIGGERING T AB .................................................................................................................... 21 

    FIGURE 4.6  – C ALIBRATION T AB ................................................................................................................... 22 

    FIGURE 4.7  – D ATA MONITOR  ...................................................................................................................... 23 

    FIGURE 4.8  – S AVE CHANGES  ..................................................................................................................... 23 

    FIGURE 4.9  – FILE PROPERTIES ................................................................................................................... 24 

    FIGURE 4.10  – FILE M ANAGEMENT AND ANALYSIS APPLICATION (W AVEWIN) .................................................. 26 

    FIGURE 4.11  – FILE T ABLE PROPERTIES DIALOG .......................................................................................... 27 

    FIGURE 4.12  – SOE FILE  ............................................................................................................................ 27 

    FIGURE 4.13  – EVENT SUMMARY  ................................................................................................................. 28 

    FIGURE 4.14  – EVENT ANALYSIS .................................................................................................................. 30 

    FIGURE 5.1  – GENERAL DEVICE SETTINGS ................................................................................................... 37 

    FIGURE 5.2  – FILE PROPERTIES DIALOG....................................................................................................... 38 

    FIGURE 5.3  – C ALIBRATION T AB ................................................................................................................... 39 

    FIGURE 5.4  – C ALIBRATION SESNORS .......................................................................................................... 39 

    FIGURE 5.5  – M ANUALLY C ALIBRATE SENSORS – STEP 1 .............................................................................. 40 

    FIGURE 5.6  – M ANUALLY C ALIBRATE SENSORS – STEP 2 .............................................................................. 40 

    FIGURE 5.7  – M ANUALLY C ALIBRATE SENSORS – STEP 3 .............................................................................. 41 

    FIGURE 5.8  – TRIGGER T AB ......................................................................................................................... 43 

    FIGURE 5.9  – TRIGGER REGIONS  ................................................................................................................. 44 

    FIGURE 5.10  – DC +  AC SIGNAL  ................................................................................................................. 44 

    FIGURE 5.11  – TRIGGER SETTINGS .............................................................................................................. 44 

    FIGURE 5.12  – HYSTERESIS REGION ............................................................................................................ 45 

    FIGURE B.1  – CLOTHESPIN CURRENT SENSOR (MODEL# CS-HE-CPL) ...................................................... B.1 

    FIGURE B.2  – SPLIT-CORE CT CURRENT SENSOR (MODEL# CS-SC-200) .................................................. B.3 

    FIGURE B.3  – SOLID-RING CT CURRENT SENSOR (MODEL# CS-SR-150) ................................................... B.5 

    FIGURE B.4  –  AC/DC VOLTAGE SENSOR (MODEL# VS-OA-500) ................................................................ B.7 

    FIGURE B.5  –  AC VOLTAGE SENSOR (MODEL# VS-VT-600) ...................................................................... B.9 

    FIGURE B.6  – TEMPERATURE SENSOR (MODEL# TS-LM-212) .................................................................. B.11 

    FIGURE B.7  – DRY CONTACT SENSOR (MODEL# DCS-01/5)..................................................................... B.13 

    FIGURE B.8  – IRIG-B (MODEL#  CBL-IRIG-B) ......................................................................................... B.15 

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    10/86

    LIST OF TABLES

    T ABLE 1.1  – SENSORS LIST  ........................................................................................................................... 3 

    T ABLE 4.1  – DEVICE CONFIGURATION FIELDS ............................................................................................... 16 

    T ABLE 4.2  – LONG FILE N AMING FORMAT ..................................................................................................... 16 

    T ABLE 4.3  – CHANNEL SETTINGS ................................................................................................................. 19 

    T ABLE 4.4  – FILE PROPERTIES FIELDS ......................................................................................................... 25 

    T ABLE 4.5  – EVENT SUMMARY ..................................................................................................................... 29 

    T ABLE 4.6  – PEAK TO PEAK ANALOG T ABLE COLUMNS ................................................................................. 34 

    T ABLE 4.7  – SINUSOIDAL RMS C ALIBRATED ANALOG T ABLE COLUMNS ......................................................... 35 

    T ABLE 4.8  – LOAD FILES ANALOG T ABLE COLUMNS ...................................................................................... 35 

    T ABLE 4.9  – DIGITAL T ABLE COLUMNS ......................................................................................................... 36 

    T ABLE 5.1  – C ALIBRATION FIELDS ................................................................................................................ 42 

    T ABLE 5.2  – TRIGGER REGIONS ................................................................................................................... 43 

    T ABLE 5.3  – TRIGGER FIELDS  ...................................................................................................................... 45 

    T ABLE B.1  – CLOTHESPIN M ALE CONNECTOR  ............................................................................................ B.1 

    T ABLE B.2  – CLOTHESPIN SPECS............................................................................................................... B.2 

    T ABLE B.3  – SPLIT-CORE CT M ALE CONNECTOR  ....................................................................................... B.3 

    T ABLE B.4  – SPLIT-CORE CT SPECS  ......................................................................................................... B.3 

    T ABLE B.5  – SOLID-RING CT M ALE CONNECTOR ........................................................................................ B.5 

    T ABLE B.6  – SOLID-RING CT SPECS  .......................................................................................................... B.5 

    T ABLE B.7  –  AC/DC VOLTAGE SENSOR FEMALE CONNECTOR .................................................................... B.7 

    T ABLE B.8  –  AC/DC VOLTAGE SENSOR SPECS  .......................................................................................... B.8 

    T ABLE B.9  –  AC VOLTAGE SENSOR FEMALE CONNECTOR  .......................................................................... B.9 

    T ABLE B.10  –  AC VOLTAGE SENSOR SPECS  .............................................................................................. B.9 

    T ABLE B.11  – TEMPERATURE SENSOR M ALE CONNECTOR  ....................................................................... B.11 

    T ABLE B.12  – TEMPERATURE SENSOR SPECS .......................................................................................... B.11 

    T ABLE B.13  – DRY CONTACT SENSOR  ..................................................................................................... B.13 

    T ABLE B.14  – DRY CONTACT SENSOR SPECS  .......................................................................................... B.13 T ABLE B.15  – IRIG-B M ALE CONNECTOR  ................................................................................................ B.15 

    T ABLE B.16  – IRIG-B SPECS  .................................................................................................................. B.15 

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    11/86

    C H A P T E

    Introduction

    OVERVIEW 

    The Sniffer is a high accuracy, highanalog input channels with 16-bit resamples/cycle). Refer to figure 1.3. Iin a portable case. Refer to figures

    Figure 1.1 – Port

    Figure 1.2 –

    The Sniffer is rugged, with a fan-lesmultiple sensor types on any inputinvestigative tool for troubleshootindisturbance monitor with Ethernet nchannels.

    Ch

    R 1

    speed analog to digital recorder. The recorder hassolution and up to 21.6 KHz per channel sampling frIt is enclosed in a 19 inch rack mount case that can.1 and 1.2.

    ble Unit

    ack Mount Unit

    s processor at an operating temperature of -30°C toort to monitor AC and/or DC Voltages and Currents., timing and monitoring. I t may also be used as a peetwork connectivity. It has user defined triggers on a

    apter 1: Introduction

    4 differentialequency (360also be mounted

    +70° C. It accepts. It is anrmanentny or all

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    12/86

    Chapter 1: Introduction

    Figure 1.3 – Differential Ana

    Utilizing the system sensors it can swithout removing equipment from sevents or static testing. Users may

    Figure 1.4 – Rear Panel

    85 - 265VAC /120 - 370VDC

    Reset Switch

    Etherne

    log Inputs Wiring Diagram

    afely be connected to equipment without any circuitrvice. It is ideally suited for deployment and captureonnect an optional monitor, keyboard and mouse.

    t Port

    Status LEDs

    4 COM Ports 2 USB Po

     

    interruption orof transientefer to figure 1.4.

    rts

    ouse andKeyboard

    Ports

    Monitor Port

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    13/86

    Chapter 1: Introduction

    The Sniffer uses the Wavewin software for data capture, file management, and waveform analysis. Serialports are provided for connecting to and polling other IEDs (refer to the Wavewin Device Managermanual).

    SENSORS 

     A variety of sensors can be used to monitor AC and/or DC Voltages and Currents. The sensors are non-

    intrusive which does not require circuit interruption or removing equipment from service. The sensors areconnected to the Sniffer via RJ45 connectors. The RJ45 connectors minimize installation andreplacement time and reduce repair time in case the sensors need to be switched out. The ability toquickly replace sensors simplifies bench testing allowing for the sensors to be replaced in only a fewminutes. Below is a list of available sensors:

    Sensor Part Number Range Accuracy

    AC Split-Core CT Current CS-SC-200 0.01 to 200 amps 0.5%

    AC Solid-Ring CT Current CS-SR-150 0.01 to 150 Amps 0.25%

    AC/DC Clothespin Current CS-HE-CPL 0.5 to 50 Amps 2%

    AC/DC Voltage VS-OA-500  0 to 500 Volts 0.5%

    AC Voltage VS-VT-600  0 to 600 Volts AC 0.5%

    Temperature TS-LM-212  -40 to 212oF +/- 35.6

    oF

    Dry Contact DCS-01/5 0 to 5 Volts 0 to 90%

    IRIG-B CBL-IRIG-B --- ---

    Table 1.1 – Sensors List

    Refer to Appendix B for detailed sensor specifications.

    EVENT

    REPORTS

     The Sniffer generates unfiltered events when a trigger is encountered on any of the 24 programmedchannels. Each channel can be programmed to trigger on RMS, Instantaneous, Magnitude or Angle. Thetrigger level is specified by the programming engineer. A digital channel is added when a trigger occurson an analog channel. The digital channel displays the trigger start and end time along with the length ofthe trigger state.

    The duration of the event file (in cycles) is also specified by the user. In the Sniffer’s File Properties dialogenter the number of pre-fault and post-fault cycles. Refer to figure 1.5. It is advisable not to exceed 3seconds for a file’s duration (180 cycle’s total). Any file over 3 seconds becomes difficult to analyze. Thedefault values for pre-fault = 6 cycles and post-fault = 54 cycles (saving 1 second files).

    The sampling frequency is specified by the user. The default value is 3000 Hz.

     A text based SOE report is also generated for each triggered event. The SOE report contains thefollowing fields:

    •  Trigger Date

    •  Trigger Time

    •  Duration of Trigger in milliseconds

    •  Maximum Value during the Trigger State

    •  Primary / Secondary Scale

    •  Primary / Secondary Ratio Order of Trigger

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    14/86

    Chapter 1: Introduction

    •  Channel Title

    •  Device Name

    •  Group Name (Station)

     All event files and SOE files are saved in the user defined trigger path. The trigger path is defaulted toc:\Wavewin\Triggers. The Sniffer has a built in processor with 16 Gigabytes of flash drive. The number ofevents stored in the Sniffer is limited to the space available on the 16 Gigabyte flash drive.

    Figure 1.5 – Trigger Properties

    MEASUREMENTS AND CALCULATIONS 

      Voltages: A, B, C line to neutral phases•  Currents: A, B, C phases and neutral

    •  Temperature: Fahrenheit, Kelvin, Celsius

    •  Targets: DC signatures (relay trip)

    •  Frequency: Cyclic, delta T, instantaneous

    •  Components: Negative, zero, positive sequences

    •  Location: Single/Double ended

    •  Power: Active, reactive, power factor

    •  Keys: Zero crossings/peaks

    •  Duration: Up to 22k scans w/user settings

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    15/86

    Chapter 1: Introduction

    •  Rate: Up to 360 samples/cycle

    •  Standards: C37.111, C37.114, C37.118, C37.232

    •  Timing: IRIG-B/PPS (0.2 millisecond accuracy)

    WAVEFORM CAPTURES AND TRIGGERS 

    •  Disturbance: Continuous capture (1 - 5 minutes)

    •  Periodic: 1 second - 15 minute averages (daily)

    •  Triggers: Instantaneous (spike filters)RMS (recursive, 1 cycle seed)Peak (recursive DFT)Phase (user defined reference)Harmonic/Frequency (custom)

    •  Settings: Less than (raw or absolute)Equal to (raw or absolute)Greater than (raw or absolute)Rate of change (for any trigger)

    •  Hysteresis: Persistence (in milliseconds)Boundary (upper and lower)

    APPLICATIONS 

    •  Dynamic circuit breaker timing 

    •  Control circuit timing 

    •  Dynamic protective relay operation timing 

    •  Capture switching transients 

    •  Transformer inrush 

    •  Tap changer smooth voltage progression

    •  Troubleshooting

    •  Unattended event monitoring 

    •  Automation process control •  Connects to various DFRs, SERs and IEDs

    •  Geomagnetic disturbance monitoring 

    FEATURES 

    •  AC / DC current monitoring

    •  AC / DC voltage monitoring

    •  Up to 21.6 KHz sampling frequency

    •  Up to 360 samples per cycle

    •  24 analog input channels

      Dry contact monitor•  Non intrusive current and voltage sensors

    •  User defined triggers on all channels

    •  Unattended data capture and file storage

    •  Universal power supply 85-265VAC, 120-370VDC

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    16/86

    Chapter 1: Introduction

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    17/86

    Chapter 2: Specifications

    C H A P T E R 2

    SpecificationsThis chapter lists the specifications for the processor, recorder and available sensors.

    PROCESSOR SPECIFICATIONS 

    The processor specifications are listed below.

    Specifications:SBC: Intel

    ® ATOM

    ™ N450 (1.66 GHz)

    RAM: 1GBUSB Ports: (2) USB 2.0Serial Ports: (4) RS-232 (2) RS 232/485Networking: 10/100/1000 MbpsMemory: 16GB flash driveInput power: 85-265VAC, 120-370VDC

    Power: 20WOperating Temp: -40°C to +85°CShock: IEC 68-2-27Vibration: MIL-STD-810E 514.4Dimensions (inches): 19” x 15” x 3.5”Weight (portable): 20 lbs

    RECORDER SPECIFICATIONS 

    The recorder specifications are listed below.

    Recorder Specifications:

    Inputs: 24 channels (differential)Resolution: 16-bit A/D resolution (216

     or 65,536 steps)Skew Factor: 0.5

    o max (parallel sampling)

    Sample Rate: Up to 21.6 KHz / channel (360 samples/sec)Input mode: DifferentialInput ranges: ±10 V, ±5 V, ±2 V, ±1 V, ±0.5 V, ±0.2 V, ±0.1 V

     Accuracy:

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    18/86

    Chapter 2: Specifications

    IRIG-B

    It transmits the IRIG-B protocol for time synchronization from the GPS Satellite Controlled Clock to theSniffer. The IRIG-B cable is part coaxial cable and part twisted pair cable. The coax cable segment isterminated with a BNC Male connector and the twisted pair cable is terminated with an RJ45 maleconnector to carry the un-modulated IRIG-B signals to the Sniffer. Assembled in 1 ft segments using BNCT-Type adapters the cable can transmit IRIG-B signals to multiple Sniffers.

    Coaxial Cable: Stranded CopperCoaxial Connector: BNC Male Twist OnCoaxial Cable Length: 1 ft SegmentMating Connector: BNC T-type AdapterOutput Cable: Stranded Twisted PairOutput Cable Conn.: RJ45 MaleOutput Cable Length: 1 ft Segment

    Surge Protector

    Optional Surge Protectors

    •  125 VDC Surge Protector

    •  120 VAC Surge Protector

    Monitor / Keyboard:

    Ruggedized portable metal suite case.

    •  3mm impact-resistant LCD screen glass

    •  101 key keyboard touch pad

    •  6’ KVM cable

    •  Connectors:

    •  1 x D-Sub 15 pin (female) for VGA

    •  1 x P/S 2 for Keyboard

      1 x P/S 2 for Mouse•  Construction: Aluminum alloy portable suite case with rubber corner bumpers

    •  LCD Panel Size: 17"

    •  Max. Resolution: 1280x1024

    •  Brightness (cd/m^): 200 nit

    •  Contrast Ratio: 350:1

    •  Response Time: 35 ms Universal Power Supply 85-265Vac, 120-370Vdc

    •  LCD MTBF: 50,000 hours

    •  Colors: 16M

    •  Power Supply : 85V~264V AC input

    •  Color: Black

    •  Operation Temperature: 0 to 55 degrees C (32 to 131 degrees F)

    •  Weight: 20lb

    •  Dimensions: 17.25"(W) x 3.30"(D) x 13.38"(H)

    Current Sensors

    Below is a list of the current sensors available. Refer to Appendix B for more detailed sensor information.

    Clothespin Current Sensor

    •  Non-Intrusive, Small, Clamp On AC/DC Sensor

    •  Hall-effect Technology

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    19/86

    Chapter 2: Specifications

    •  Less than 10 Microseconds Response Time

    •  Locking Mechanism

    •  Very High Sensitivity

    •  Very Low Noise

    •  Shielded Enclosure (curved mu-metal strip)

    •  Single Cable for Power and Output Signals

    Split-Core CT Current Sensor

    •  Clamp On Current Sensor

    •  Current Transformer Technology

    •  Non-Intrusive Installation

    •  Sensitive to AC Currents (0.1 to 200 Amps)

    •  Secure Hinge and Snap Locking Mechanism

    •  Very Low Noise 

    •  High Sensitivity 

    •  Analog Output Voltage 

    Solid-Ring CT Current Sensor

    •  Precision CT for Metering

    •  Current Transformer Technology

      Compact Epoxy Molding•  Sensitive to AC Currents (0.1 to 150 Amps)

    •  Confirms to ANSI C12xx and IEC 62053.22

    •  Very Low Noise 

    •  High Sensitivity 

    •  Analog Output Voltage

    Voltage Sensors

    Below is a list of the voltage sensors available. Refer to Appendix B for more detailed sensor information.

     AC/DC Voltage Sensor

    •  Small, Rugged, Voltage Sensor•  +/- 500 V Peak Max Voltage Input

    •  Accurate AC/DC Voltage Measurements

    •  DC to 5 KHz Bandwidth

    •  High Accuracy +/- 0.05%

    •  Single Cable for Power and Output Signals

    •  DIN Rail Mounts

    •  UL, CUL Certified

    •  Fused Input

     AC Voltage Sensor

    •  Small, Rugged, Voltage Sensor

    •  600 V Max Input Voltage

    •  Accurate AC Voltage Measurements

    •  50 to 400 Hz Bandwidth

    •  Accuracy 0.5%

    •  Panel Mounts

    •  UL, CUL Certified

    •  Fused Input

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    20/86

    Chapter 2: Specifications

    10 

    Other Sensors

    Below is a list of other sensors available.

    Temperature Sensor

    •  2 Inch Stainless Steel Probe

    •  Water Proof Enclosure

    •  Wide Operating Range

    •  Analog Output Voltage

    •  Directly Calibrated in Kelvin

    •  Single Cable for Power and Output Signals

    •  Powered from the Receiver (no additional power needed)

    Dry Contact Sensor

    •  Small, Lightweight Sensor

    •  Rugged Binding Posts

    •  Single Cable for Output Signals

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    21/86

    Chapter 3: Installation

    11 

    C H A P T E R 3

    InstallationThis chapter explains installation of the Wavewin Sniffer 24 Disturbance Monitor (Sniffer), connecting thesensors, upgrading and running the software.

    INSTALLING THE RECORDER 

    To install and connect the Sniffer safely the user must be familiar with substation equipment installation.

    Rack MountSelect a position on the rack mount frame to allow access to all the circuits that will be monitored. Securethe device with four rack mount screws, two on each side on the front of the device.

    Portable UnitPosition the portable unit to allow access to all monitored circuits. Remove the front and back covers.

    Sniffer Startup Procedure 1 (Configuration - first time): follow the steps below.

    1. Connect the LCD and the PS\2 keyboard-mouse before connecting power to the Sniffer.2. Connect the power (AC power cable provided). There is no on/off switch, the unit will boot up and

    the Sniffer configuration software will automatically run when power is connected.3. Before configuring the Sniffer software, click “End Recorder” to stop the device software from

    polling. Connect the sensors, and then configure the software.4. Optional: (remove the LCD and the keyboard-mouse after configuration is complete) Power down

    the Sniffer, turn off the LCD and remove both LCD and keyboard-mouse.5. Connect power to the sniffer.

    Sniffer Startup Procedure 2 (Reconnecting LCD and keyboard-mouse): follow the steps below.

    1. In case the LCD and keyboard-mouse were removed after configuring the Sniffer. Beforereconnecting them, power down the Sniffer first, then reconnect the LCD and keyboard-mouse.

    2. Power up the Sniffer and press F8 at startup to boot in Safe Mode.3. When in safe mode, run the video controller software and follow the steps on the screen.4. Power down the Sniffer, turn LCD off then reconnect power to the Sniffer and turn the LCD on.

    CONNECTING THE SENSORS 

    The sensors are connected to the recorder using RJ45 ports. Refer to figure 3.1.

    Figure 3.1 – RJ45 Connectors

    UPGRADING THE SOFTWARE 

    In the event an upgrade to the Sniffer software in needed follow the instructions for upgrading with a USBKey drive.

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    22/86

    Chapter 3: Installation

    12 

    USB Key: To install the software using a USB Key drive, place the USB Key into one of the two USBports. Open Windows Explorer (right click on the Start menu and select the Explorer menu option)navigate to the USB drive select the folder where the Wavewin Sniffer 24 software is located and doubleclick on the install.exe application.

    The installation instructions are listed in the following:

    Figure 3.2 – Start Software Installation

    The default destination folder path is C:\Wavewin. To change the default path either type in a new installpath or click on the browse button to select an existing directory.

    The destination folder is the location where all the files are to be copied.

    Click “Next” to start the installation.

    Figure 3.3 – Create Install Path

    Click “Yes” to create the path.

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    23/86

    Chapter 3: Installation

    13 

    Figure 3.4 – Finish Software Installation

    The software upgrade is now complete click “Finish” to end the installation.

    To restart the software double click on the Wavewin Sniffer 24 icon installed on the desktop.

    RUNNING THE SOFTWARE 

     An icon for the configuration and polling software (Wavewin Sniffer 24) is installed on the desktop. To runthe configuration and polling software click on the Wavewin Sniffer 24 icon. To run the Wavewin softwareclick on the “View Files” button located at the bottom of the configuration and polling window.

    VERSION INFORMATION 

    To view the current version number and release date of the Wavewin Sniffer 24 configuration and pollingsoftware, click on the “About” button.

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    24/86

    Chapter 3: Installation

    14 

    Figure 3.5 – About Window

    TECHNICAL SUPPORT 

     Although this system is easy to use and understand, at some point you may encounter a technicalquestion, feel that the system has improperly operated, or have suggestions for future improvements. Ineither case, contact SoftStuf using one of the following methods:

    Phone: 215-922-6880 • 800-818-3463, hours are from 9:00 a.m. to 6:00 p.m. Mon- Fri, (EST).Fax: 215-625-2497, response time is 24 hours.E-mail: [email protected], response time 24 hours.

    mailto:[email protected]:[email protected]:[email protected]

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    25/86

    Chapter 4: System Software

    15 

    C H A P T E R 4

    System SoftwareThis chapter describes the 2 software applications pre-installed on the Sniffer, the configuration andpolling software (Wavewin Sniffer 24) and the file management and analysis software (Wavewin). The

    following sections explain each application.

    CONFIGURATION AND POLLING 

    The configuration and polling software is used to configure the device and for starting and stopping thepolling process. There are 2 main sections, the device configuration and the channel settings. Refer tofigure 4.1.

    Figure 4.1 – Configuration and Polling Software

    Device Configuration

    The 4 fields located in the device configuration section are used to set the general configuration for thedevice. Refer to figure 4.2. Enter the station where the device is located, the name of the device, thecompany that owns the device and the time code where the device is located. When entering informationinto these 4 fields do not include the following characters:

    : ? “ / \ < > * | @ #

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    26/86

    Chapter 4: System Software

    16 

    These 4 fields are used to compose the IEEE long file name for the event and SOE files. Refer to “IEEELong File Naming Format” section for more information. Microsoft does not permit the above charactersin a filename.

    Figure 4.2 – Device Configuration Fields

    The table below defines each field in the device configuration section.

    Field  Description  Default 

    Station Name of the substation where the device is installed. SUBSTATION

    Device Name of the installed device.  DEVICE

    Company Name of the company that owns the device. COMPANY

    Time Code Time code where the device is installed. Time is offset fromGreenwich Mean Time (GMT).

    -5

    Table 4.1 – Device Configuration Fields

    IEEE Long File Naming Format

     All event and SOE files are saved using the IEEE long file naming format. The IEEE long file namingformat is a PSRC format used to name time sequenced data files. The file name contains the following 10fields stored in a comma-delimited fashion:

    Example: 000112,123433234,-5S,South Arkey,T4 Breaker,Sun Power,3000.TIS

    Field Definitions:

    Field Example Displayed Definition

    Date 040112 01/12/2004 The Date field defines the start date of the file.The date fields are stored as: year (2characters), month and day.

    Time 123433234 12:34:33.234 The time field defines the start time of the file.The Time fields are defined as: hour, minutes,seconds and milliseconds.

    Tcode -5S -5S The Time Code defines the time offset fromGMT time. -5s would be specified for USEastern Standard Time. If the start time isexpressed in UT, this field is coded 0z.Note: GMT is the international abbreviation forGreenwich Mean Time.

    Substation South Arkey South Arkey The substation name or code where theoriginating device is located.

    Device T4 Breaker T4 Breaker The device name or code that generated thefile.

    Company Sun Power Sun Power The company of the specifed substation.

    SamplingFrequency

    3000 Hz 3000 The files sampling frequency.

    Table 4.2 – Long File Naming Format

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    27/86

    Chapter 4: System Software

    17 

    Channel Settings

    The analog channel settings section defines all the fields needed to configure the connected sensors. Thefields are divided into 5 sections: Channel Titles, Configuration, Triggering, Calibration and the DataMonitor. Refer to figure 4.3.

    Figure 4.3 – Channel Settings

    The first 8 channels are displayed on the first page and channels 9 through 16 are on the second pageand the last page contains channels 17 through 24. To navigate from field to field use the tab key. Tonavigate between rows in an edit field use the up and down arrow keys or use the page up and pagedown keys. Click on the scroll bar to move up or down the channels.

    The table below defines each field in the channel settings section.

    Header Description Default

    Channel Titles The name of the channel. By default, the channel title isnamed “Unused”. This field is saved in the first line of thedata file. Click on the header to default all titles to“Unused”. (Required)

    Unused

    Configuration - Channel Settings 

    Sensor The Sensor field indicates the type of sensor connectedto that port. From the drop down list, select the type ofsensor to be polled. Click on the sensor header to defaultboth the sensor and unit columns to “None”.

    None

    Unit The Unit field is directly related to the type of sensorbeing polled. The units are automatically associated withthe type of sensor selected.

    None

    P/S Ratio The primary to secondary ratios for AC measurements.DC primary ratio is always 1/1. Click on the header todefault the column to 1/1.

    1/1

    Calculate Mag/Ang Calculate the Discrete Fourier Transform (DFT) anddisplay the Magnitude and Angle values in the datamonitor section. Checked = On. Click on the header toturn all on or off.

    Checked (On)

    Reference Angle Defines the DFT reference angle for displaying the angle 1st Channel

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    28/86

    Chapter 4: System Software

    18 

    Header Description Default

    in the data monitor section. Only one channel can beselected as the reference angle. Click on the header todefault channel one as the referenced angle.

    Triggering - Channel Settings 

    Trig Value The trigger value to initiate recording. This field along withthe following 6 fields defines when to save an event file to

    disk. Trigger values can be defined for monitoring RMS,Instantaneous, Magnitude or Angle values. Click on theheader to clear all trigger value fields.

    Blank

    Trig Type The type of values to monitor for the entered triggervalue, RMS, Instantaneous, Magnitude or Angle. Click onthe drop down list and select the desired trigger type.Click on the header to default all trigger types to Inst.

    Inst

    Duration (ms)  The duration of the trigger before recording an event file.The duration is measured in milliseconds (ms). Click onthe header to default the column to 0.

    0

    Upper Hysteresis(Upper)

    Upper offset for the trigger level(Trigger Level + Upper Hysteresis = Upper Trigger Level).Click on the header to default the column to 0.

    0

    Lower Hysteresis(Lower) 

    Lower offset for trigger level(Trigger Level - Lower Hysteresis = Lower Trigger Level).Click on the header to default the column to 0.

    0

    Absolute Values Take the absolute value of the samples before comparingthem to the entered trigger value. This option is useful incase the sensor was mistakenly mounted in the reversepolarity direction. Unchecked = Off. Click on the header toturn all on or off.

    Unchecked (Off)

    Operator The logic to use when determining if a trigger level isactive. There are 4 types of options available: greaterthan (>), less than ().

    >

    Calibration - Channel Settings 

    Offset The Offset is a measure of the distance to the zeroreference axis. To set the offset refer to the calibratingsection. Click on the header to default all offset values to0.

    0

    Scale Factor The Scale Factor is a real number used to scale the rawdata. This field is automatically populated with a defaultvalue when a sensor type is selected. The default valuesare listed in the calibrating section. To calibrate the scalefactor refer to the calibrating section. Set this field to 1before calibrating the sensors that need calibrating. Click

    on the header to default the column to the default scalefactor for the selected sensor type.

    1

    Calibrate Value The known value for the calibration process. Enter theexpected amount of current injected into the wire duringthe calibration process. This field is automaticallypopulated with a default value when a sensor type isselected.

    2

    Calibrate The calibrate check box is used to indicate if the sensorneeds to be calibrated. If checked, the channel will becalibrated. To calibrate, refer to the “Calibrating theSensor” section. Click on the header to make all channels

    Checked(Active)

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    29/86

    Chapter 4: System Software

    19 

    Header Description Default

    active or inactive.

    DC Cancellation(DC Cancel) 

    Defines if the software should automatically calculate theoffset value for the sensors connected when polling. TheHall-effect clamp on current sensor may drift off the zeroreference point over time causing the sample values tobe incorrect. Click on the header to turn all on or off.

    Checked (On)

    Range Voltage Range of the analog channels. It increases the

    magnitude of an input signal. Range options are:±10 Volts±5 Volts±2 Volts±1 Volts±0.5 Volts±0.2 Volts±0.1 Volts.Useful when measuring very low current levels (below 1amp). Click on the header to default the column to ±2Volts.

    ±2 Volts

    Data Monitor - Channel Settings 

    RMS The RMS value calculated over a one cycle window. Thevales are displayed when polling begins and clearedwhen polling is stopped.

    Blank

    Instantaneous  Amplitude of the input signal at a particular instant. It isthe raw values received from the device minus the offsetvalue, and multiplied by the scale factor ((raw values-offset)*scale factor). The vales are displayed when pollingbegins and cleared when polling is stopped.

    Blank

    Magnitude Magnitude of the input signal. The values are displayedwhen the “Calculate Mag/Ang” check box is checked. Thevales are displayed when polling begins and clearedwhen polling is stopped.

    Blank

    Angle  Angle of the input signal. The values are displayed when

    the “Calculate Mag/Ang” check box is checked. The valesare displayed when polling begins and cleared whenpolling is stopped.

    Blank

    Sampling Frequency Enter the desired sampling frequency. 3000

    Start/Stop Capture Click on the Start Capture button to being the pollingprocess, once the polling starts the button caption willchange to Stop Capture. Click to stop the polling process.If any errors are encountered a message box will bedisplayed.

    Start Capture

    Table 4.3 – Channel Settings

    Channel Titles

    The channel titles are listed on the left side of the tabs. The first 8 channels are listed on the first pageand the next 8 channels 9 through 16 are listed on the second page and the last page contains channels17 through 24. Use the page up and page down keys or the scroll bar to navigate through the channels.Enter a title for each sensor connected. The channel titles are defaulted to unused.

    Configuration Tab

    The configuration tab lists the general settings for the sensors connected. Refer to figure 4.4. The sensorcolumn defines what type of sensor is connected on each port. The sensor drop down list is composed

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    30/86

    Chapter 4: System Software

    20 

    from the TIS.INI file located in the Sniffer’s install path. When a sensor is selected the unit, offset, scalefactor, calibrate value, DC cancelation and P/S ratio fields are automatically populated from the factorysettings stored in the TIS.INI file.

    Enter the unit and P/S ratio for each channel if different from factory settings. The “Cal Mag/Ang” columndefines if the Discrete Fourier Transform (DFT) calculation will be performed on the incoming samples.The reference angle column defines the channel that all other channel angles will be referenced from.

    Figure 4.4 – Configuration Tab

    Triggering Tab

    The trigger tab lists all the fields needed for defining a channel’s trigger. Refer to figure 4.5. The triggerfields define when to initiate the recording of an event file. Enter the trigger value to monitor for the triggertype selected in the “Trig Type” drop down list. The available trigger types are RMS, Instantaneous,Magnitude or Angle. The duration column defines how long the trigger must persist before recording anevent. The upper and lower columns define the channel’s hysteresis region. Refer to the “SettingTriggers” section in Chapter 5 for more detailed information.

    The abs column will check all samples in absolute value and the last column; operator defines the type ofoperation to use when checking for trigger levels.

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    31/86

    Chapter 4: System Software

    21 

    Figure 4.5 – Triggering Tab

    Calibration Tab

    The calibration tab is used for manually calibrating the sensors. A complete description is defined in the“Calibrating the Sensors” section in Chapter 5. The offset and scale factor columns are used whenscaling the raw samples for display in the data monitor section and when plotting the event files.

    The calibrate value column defines the value to calibrate the sensors at. When calibrating differentsensors you may what to calibrate sensors at different calibration values. The calibrate column allows forturning on or off the calibration process. Checked indicates to calibrate the channel. The DC cancelationcolumn continually calculates and updates the Offset column. The range column allows for setting thevoltage range for each sensor. The available ranges are: ±10 Volts, ±5 Volts, ±2 Volts, ±1 Volts, ±0.5Volts, ±0.2 Volts and ±0.1 Volts.

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    32/86

    Chapter 4: System Software

    22 

    Figure 4.6 – Calibration Tab

    Data Monitor

    The data monitor section displays the samples and calculated values when the polling process is active.The values are displayed once per second. The angle values are displayed with respect to the activereferenced channel defined in the configuration tab.

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    33/86

    Chapter 4: System Software

    23 

    Figure 4.7 – Data Monitor

    Configuration File

     All the information entered into the device configuration and the channel settings sections are saved to an ASCII text file located in the install path (c:\Wavewin). The text file is named: WavewinSniffer24.txt. Savethe current settings using the “Save” button. If changes were made when exiting the application amessage box will be displayed asking to save the changes. Click “Yes” to save or “No” to discard thechanges. Refer to figure 4.8.

    Figure 4.8 – Save Changes

    File Properties

    This section defines the fields in the File Properties dialog. This dialog allows the user to define the savepaths for the events and SOE files (Trigger) and for creating snap shot files (Continuous). Refer to figure4.9. The trigger path is defaulted to c:\Wavewin\Triggers and the continuous save path is defaulted toc:\Wavewin\Continuous. To select an existing path, click on the folder button to the right of the path fields.

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    34/86

    Chapter 4: System Software

    24 

    The number of pre-fault and post-fault cycles saved to the event files are defined under the TriggerProperties section. The number of pre-fault cycles is defaulted to 6 and the number of post-fault cycles isdefaulted to 54.

    The IRIG-B Properties section allows the user to define the time setting on the connected IRIG-B clock.

    Click on the “Files Properties” button to open the dialog.

    Figure 4.9 – File Properties

    The following table defines the fields available in the File Properties dialog.

    Field Description

    Save Continuous Continuously save all monitored data to create a snap shot file. All filesare assigned the “.TIS” extension. Click on the “Save Continuous” checkbox to save all monitored data. Uncheck this box when only capturingevent trigger files. By default this box is unchecked.

    File Duration The time duration of the data file measured in minutes. The software willautomatically save a new file when the maximum file duration isreached. Save options are (1 to 5) minutes. By default the duration is setto 1 minute.

    Continuous Path The save path for the snap shot files. The file duration option defines thelength of each file. Enter a path or click on the “Folder” button to browse

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    35/86

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    36/86

    Chapter 4: System Software

    26 

    Device NameGroup Name (Station Name)

    File Table

    To view the events and SOE files click on the “File Properties” button. The Wavewin file management andanalysis application will be displayed with the Trigger Path as the active path. The Wavewin file

    management and analysis display contains a folder tree to the left of the window and the file table to theright. The folder tree lists the available drives and folders on the device. The file table lists all the folderand files in the active path. The event files are colored black and the SOE files are colored gray.

    Figure 4.10 – File Management and Analysis Application (Wavewin)

    The file table lists the fields in the IEEE long file name in separate columns. The trigger date and timefields are listed under fault date and time columns. The device name is listed under the device column,the group name is listed under the substation column and the company name is listed under the companycolumn. The sampling frequency is listed under the first optional column (Optional-1).

    The columns displayed in the table can be repositioned through the “Display” feature under the “Options”menu. Use the Move Up and Move Down buttons to change the position of a column. The table columnscan also be resized. Position the mouse over the column separator and drag the mouse to the desiredlocation or double click on the table separator to resize the column to the largest display.

    The size of the font displayed in the table can also be changed. Use the “Table Font Size” drop down listto select the desired font size. Also, to use the Window’s default f ile marking, select the “Windows” optionfrom the “File Marking” drop down list. To mark files with a single mouse click select the “Single MouseClick” option.

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    37/86

    Chapter 4: System Software

    27 

    Figure 4.11 – File Table Properties Dialog

    SOE Files

    The SOE files are a text report that lists all the triggers that were encountered during an event file. Thetriggers are listed in a table. Each trigger has a trigger date and time, the duration of the trigger inmilliseconds, the maximum value encountered during the trigger, the primary and secondary scale andratio at the time of the trigger and the channel and device information.

    The number listed before the channel title is the order the trigger was detected. In the example belowSensor Three first encountered a trigger at 00:30:54.319 for 41.68 milliseconds then Sensor Onedetected a trigger then Sensor Two. Sensor One and Two’s trigger lasted 46.13 milliseconds.

    Figure 4.12 – SOE File

    Event Summary

     An ASCII summary of an event file can be displayed from the file table. To display the summary for anevent file move the table cursor to the desired event file, open the “Options” menu, select the “WaveformSummaries” sub menu then select the “View Selected File Summary” menu option. The summary engineextracts key information from the file and saves it to a small summary file. The summary includes general

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    38/86

    Chapter 4: System Software

    28 

    event information, a summary table for the analog channels and a sequence of events for the digitalchannels.

    Figure 4.13 – Event Summary

    The summary includes the following fields:

    Event Information

    Field Description

    Station  Name of the Station associated with the event file. Filename  Name of the event file. 

    File Size  File size (displayed in kilobytes). 

    Prefault-Time  Date and time of the first prefault sample. 

    Fault-Time  Date and time of the first fault sample. 

    Save-Time  Date and time the file was saved to disk. 

    Process-Time Date and time the file summary was processed.

    Start Date andTime

    Date and time of the first sample in the file.

    End Date andTime

    Date and time of the last sample in the file.

    File Duration Duration of the file measured in days, hours, seconds, milliseconds and/ormicroseconds, depending on the type of file.

    SamplingFrequency

    Sampling frequency and the time between each sample.

    Line Frequency Line Frequency defined in the file.

    Maximum/Minimum Analog Summary

    Field Description

    Max-Inst  Instantaneous maximum values.

    Min-Inst  Instantaneous minimum values.

    Max-RMS  RMS maximum values.

    Min-RMS  RMS minimum values.

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    39/86

    Chapter 4: System Software

    29 

    Field Description

    OneBit  Channel's full-scale value divide by the channel's resolution.

    Inst-Diff The difference between the Max-Inst and Min-Inst values.

    RMS-Diff The difference between the Max-RMS and Min-RMs values.

    pU  Channel prefix and unit.

    Description  Channel title and number.

    Events/Sensors Activity Summary

    Field DescriptionFst  State at which the channel started, A=alarm and N=normal. 

    Lst  State at which the channel ended, A=alarm and N=normal. 

    Fst-Change  Date and time the channel first changed state. 

    Lst-Change  Date and time the channel last changed state. 

    Changes  Number of times the channel changed state. 

    Description  Channel title and number.

    Events/Sensors Activity Log

    Field Description

    State  State of the channel at the triggered time, A=alarm and N=normal. 

    Trigger Time  Time the channel-changed state. 

    Description  Channel title and number. 

    Table 4.5 – Event Summary

    The xx:xx:xx.xxx displayed in the “Fst-Change” and/or “Lst-Change” fields of the Events/Sensors ActivitySummary indicates that the digital channel’s state did not change from the initial state (Fst).

    Event Analysis

    To open an event file for analyzing double click on the file or move the table cursor to the desired file andpress enter. The analog and digital channels are displayed in a graphical interface. The analog channelsare displayed in the top portion of the window along with the analog values table and the phasor diagram.Only the active analog channels are displayed. Analog channels that have a title of “Unused” are

    considered inactive channels.

    The digital channels are displayed in the bottom portion of the window along with the digital summarytable.

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    40/86

    Chapter 4: System Software

    30 

    Figure 4.14 – Event Analysis

    Displayed channels can be marked,superimposed, scaled, numericallywindows can be opened at one tim

    The display contains two sections: tchannels. The analog table displaysinstantaneous, maximum, and minidigital view plots the digital channeltime of the first change, time of the l

    Navigating

    Use the up and down arrow keys ortoggles between the analog and dig

    The analog and digital values are dibutton down and drag the table sepvalues use the following navigation

    •  Left and right arrow keys t

    •  Ctrl+left or ctrl+right keys

    •  Shift+ctrl+left or shift+ctrl

    •  Home and end keys to dis

    •  Triangle at the bottom

    •  Page up and page down

    merged, appended, moved, zoomed, removed, resrocessed, exported and summarized. A maximum.

    he analog view and the digital view. The analog vievalues such as the channel’s highest peak, RMS, pum values. The cursor bars are used to view the dand displays the channel’s original state, the chan

    ast change, and the number of times the channel ch

    the vertical scroll bar to browse the analog channelital views.

    splayed in a table to the right of the channel traces.arator bars to resize the viewing area. To view the atools:

    o navigate sample by sample

    o peak navigate

    right keys to cycle hop

    play the channel’s first and last samples

    f the data bar to drag the data bar through the sam

      eys to page through the samples

    ored,f ten analysis

    plots the analoghase, reference,ta values. Theel’s final state,anged state.

    . The tab key

    Hold the mousenalog sample

    les

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    41/86

    •  Left button displayed tthe position of the first sa

    •  Left button displayed tto the position of the data

    Click the left/right arrow button (right/left arrows to scroll through thefield descriptions.

    NOTE: If no channels are marked tfirst channel’s data.

    Setting the Cursor Bars

    Four vertical cursor bars are displabar, the black solid line represents tdotted line represents the fault posithorizontal bars displayed when the “

    Data Bar

    The data bar is displayed as a blaautomatically displayed at the enmove the data bar use the left ankeys to peak hop, use the Shift-Cdown keys to move one page updrag the triangle to scroll throughdisplays the sample number at thsample at the data bar is displayeare displayed to the right of the tr

    Reference Bar

    The reference bar is displayed asspecified in the file. To move the r

    Bar to Data Bar" option inside theClick the opposite mouse button tfield Delta X in the status bar at thseconds) between the referencetwo bars.

    RMS Bar

    The RMS bar is displayed as a bldata bar, except when the data bcalculating the RMS value displayis calculated using all of the sampthe RMS bar to the position of the

    Bar" option inside the “View” men

    Chapte

    the left of the data scroll bar to move the sample aple displayed

    the right of the data scroll bar to move the trace anbar

    located to the right of the analog table headers) or uanalog table columns. Refer to the “Viewing Analo

    hen the peak navigate and cycle hop features navig

    ed in the analog view. The blue dotted line represenhe data bar, the black dotted line represents the RMion defined in the file’s configuration information. ThHorizontal Bars” menu option under the “View” men

    ck solid line with a white triangle below the line. Theof the first cycle in the data window when it is firstright arrow keys to move one sample, use the Ctrl-

    rl-left and Shift-Ctrl-right keys to cycle hop, use ther down or left click the mouse to move to any positi

    the data. When the mouse is held over the triangledata bar and the delta time from the first sample.

    d in the second status bar field. The channel valuesces in the analog channel information table.

    a blue dotted line. The reference bar is defaulted toeference bar to the position of the data bar use the

    “View” menu or press Ctrl-A or click the SetRef  meo move the reference bar to any position in the datae bottom of the screen shows the time difference (inar and the data bar. It also shows how many cycles

    ck dotted line. The RMS bar is defaulted to one cycr is positioned at the beginning of the data. This bared in the analog information table. The RMS value ile values displayed between the data bar and the Rreference bar (blue dotted line) use the "Move RM

    u or press Ctrl-Z or click the SetRMS  menu but

    4: System Software

    31 

    the data bar to

    d table separator

    se the shift-  Data” section for

    te through the

    ts the referenceS bar and the redere are also twou is checked.

    data bar isisplayed. Toleft and Ctrl-rightage up and pagen in the data orhint message

    he time of theat the data bar

    the fault timeMove Reference

    u button .area. The statusmilliseconds orare between the

    le away from theis used for

    n the analog tableMS bar. To move

    Bar to Reference

    ton.

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    42/86

    Chapter 4: System Software

    32 

    Fault Bar

    The fault bar is displayed as a reddefined in the file’s configuration i“Yes” or “No” options for the “ShoSettings” tab.

    Horizontal Bars

    When the “Horizontal Bars” menudisplayed. The solid black line follThe bars will be positioned at themarked, then they are positionedshows the difference between the

    To automatically resize the RMS sliopen the “Set Sliding Window Size”window click the opposite mouse b

    data bar position then click the Set

    position. The Delta X field displayeddifference (in milliseconds or seconbetween the two bars. Use the left,the horizontal scroll bar to move the

    Marking, Deleting, and Rest

    To mark or unmark a channel, moumark a group of channels click on tare displayed in red.

    To mark/unmark all analog and digioption under the “Channels” menuchannels will be marked. If any chamark/unmark all the analog channel“Channels” menu. To mark/unmarkoption under the “Channels” menu.

    Channels must be marked to deletechannels and the Insert key restore

    Scaling Analog Channels

    When the analysis display is initiallythe channels according to the maximenu button. This option toggles adisplayed in the “AS” status field. E

    •  ON – The On view plots treference line.

    •  OFF – The Off view plots

    •  ++ – The ++ view plots thchannel. The highest valu

    dotted line. The fault bar is fixed and positioned atnformation. The fault bar can be shown or hidden b

    Vertical Fault Bar” field in the properties dialog un

    option under the “View” menu is checked two horizows the data bar and the dotted blue line follows thefirst marked analog channel (displayed in red), if noat the first displayed channel. The Delta Y field in thtwo bars.

    ing window click on the Resize Sliding Window mdialog from the “View” menu. To manually resize thtton to set the reference position and the mouse but

    MS  menu button. The RMS bar is moved to th

    in the status bar at the bottom of the screen showss) between the date bar and reference bar and theright, ctrl+left, and ctrl+right, shift+ctrl+left, and shiftsliding window.

    oring Channels

    e click on the channel ID or channel title, or use thee first channel then shift click on the last channel. M

    al channels press the F8 key or select the “Mark/Unption. If no channels are marked, all of the analog anels are marked, all of the channels will be unmark

    s select the “Analog Mark/Unmark All” menu optionall of the digital channels select the “Digital Mark/Un

    them from the display window. The Delete key remall the deleted channels.

    opened, all the analog channels are scaled to oneum space allocated for display, press F6 or click thong the three views: On, Off and ++. The active auch view is defined below:

    e channel data scaled to the maximum value alloca

    all of the channels that are scaled to the maximum v

    signal using the number of maximum pixels allocate is plotted at the maximum position and the smalle

    he fault timeselecting theer the “Display

    ntal bars will bereference bar.

    channels arestatus bar

    enu button orRMS sliding

    ton to set the

    e reference

    the timenumber of cyclesctrl+right keys or

    space bar. Toarked channels

    Mark All” menund digitaled. Tounder themark All” menu

    ves the marked

    alue. To scalee AutoScale o scale state is

    ted along the zero

    alue in the file.

    ed for thet value is plotted

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    43/86

    at the lowest position. Thichannel, a Vdc channel a

    To increase or decrease a channel’channels then click the AmpUp The auto scale multiplier (ASM) is uwhen the amplitude increases the Athe amplitude decreases the ASM v

     ASM value, select the “Properties”tab, enter a number and click OK. T

    analog channels amplitude, click thscroll bar. To increase/decrease onl

    buttons located in the phasor

    To increase or decrease the channor press the ctrl+page up and ctrl+p

    NOTE: If no channels are marked

    Zooming ChannelsTo zoom in on specific analog or diViewMrks menu button. The unmarked channels are rescaled to fitthe key, or click the V

     When returning to the original viewadditional channels for a new view.

    Viewing Analog Data

    The values displayed in the analogconcentrated form (combination viecan only be displayed if there is eno

    navigate through the analog table ctable headers) or the shift-right/leftthe header. Valid analog channelsinformation in the center table. An aany of the following strings in the be

    •  UNUSE

    •  UNDEF

    •  NOT D

    •  NOT U

    •  NOT I•  NAT A

    •  UNDEF

    •  {

    •  N/A

    •  ANALOG INPUT

    •  ANALOG CHANNEL

    •  EXTERNAL INPUT

    •  EVENT CHANNEL

    •  CHANNEL

    Chapte

    feature was added to clearly show the profile of a f d load data channels.

    amplitude, along with the phasor and circular chartor AmpDn menus buttons or use the ctrl+up/dosed to amplify or attenuate the channel’s data valueSM value is multiplied by the channel’s current “Pixalue is divided by the channel’s current “Pixsdisp”.

    enu option under the “File” menu then click the “Dihis value is initially defaulted to 2.00. To increase/d

    up and down arrow buttons located to the riy the phasor magnitude or circular chart click the up

    /circular chart header.

    l’s time scale, click the Condense  or Expandage down keys.

    ll the visible channels are scaled accordingly.

    ital channels, first mark the channels then press or click theow and thethe key,

    quick selection of

    or in aoncentrated vieweach channel. To

    of the analogbutton located inanalog

    y, or it contains

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    44/86

    Chapter 4: System Software

    34 

    •  DIGITAL TRACE #

    •  SPARE

    To show an invalid channel use the“Show/Hide Channel Title” section f

     A maximum of 256 analog channelstable and combination view are des

    Analog Table View:

    The analog table view is the defaultnavigate through the columns of thefollowing data types:

    •  Peak to Peak data

    •  RMS Calibrated data

    •  Log files.

     All of the display drivers in the systechange the settings for a driver sele"Driver Data Type" tab and select th

    File’s data type cannot be changed. The following tables describe the asinusoidal RMS data types:

    Field Description

    Title The analog c

    RMS The TrueRMall the samplthe data bar.the two bars

    InstPeak The highestreference cro

    Phase The phase aInstVal The sample

    RefVal The sample

    MaxPeak The maximu

    MinPeak The minimu

    Units The analog c

    PixsDisp The number

    DFT Peak The DFT Maand the data

    Crest The DFTMag

    Table 4.6 – Peak to Peak Anal

    Field Description

    Title The analog cRMS The RMS Val

    between theSince the datsquare root o

    InstPeak The square rtwo referenc

    Phase The phase a

    InstVal The RMS sa2.

    “Show/Hide Channel Title” dialog in the file manage  or usage.

    can be displayed in one window. The values displacribed below.

    view. Use the view button or the shift-right/left atable. The original sample values are plotted accor

    m are defaulted to peak to peak except the predefinct the "Window Properties" option under the "File"e type from the "Data Multiplier Type" drop down lis

    alog data for the sinusoidal peak-to-peak, non-sinu

    hannel titles.

    value is calculated by taking the summation of thevalues that are between the RMS bar (black dotted

    The result is divided by the total number of samplesnd takes the square root of that result.

    bsolute value of all of the samples between the twossings surrounding the data bar (black solid line).

    gle of each channel.alue at the data bar (black solid line).

    alue at the reference bar (blue dotted line).

    peak value of the channel.

    peak value of the channel.

    hannels prefix and units.

    f pixels allocated for displaying the trace.

    nitude calculated between the RMS bar (black dottbar (solid data bar).

    column divided by the RMS column.

    g Table Columns

    hannel titles.column calculates an RMS value for all of the samMS bar (black dotted line) and the data bar (black

    a is RMS calibrated each sample value is multipliedf 2 before it is squared.

    ot of 2 multiplied by the peak value measured betwcrossings surrounding the data bar (black solid line

      gle of each channel.

    ple value at the data bar (black solid line) multiplie

    r. Refer to the

    ed in the analog

    rrow keys toing to one of the

    ed log drivers. Toenu. Click on the. Periodic Log

    oidal, and

    square ofline) andbetween

    zero

    d line)

    lesolid line).by the

    een the).

    by Root

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    45/86

    Chapter 4: System Software

    35 

    Field Description

    RefVal The RMS sample value at the reference bar (blue dotted line) multiplied byRoot 2.

    MaxPeak The RMS maximum peak value of the channel multiplied by Root 2.

    MinPeak The RMS minimum peak value of the channel multiplied by Root 2.

    Units The analog channels prefix and units.

    PixsDisp The number of pixels allocated for displaying each trace.

    DFT Peak The DFT Magnitude calculated between the RMS bar (black dotted line)

    and the data bar (solid data bar).Crest The DFTMag column divided by the RMS column.

    Table 4.7 – Sinusoidal RMS Calibrated Analog Table Columns

    Field Description

    Title The analog channel titles.

    MaxWin The absolute maximum value between the sliding window bar (black dottedline) and the data bar (black solid line).

    InstVal The sample value at the data bar (black solid line).

    RefVal The sample value at the reference bar (blue dotted line).

    MaxVal The maximum value of the channel.

    MinVal The minimum value of the channel.

    Units The analog channels prefix and units.

    PixsDisp The number of pixels allocated for displaying the trace.

     AvgWin The average value of all of the samples between the sliding window bar(black dotted line) and the data bar (black solid line)

    Table 4.8 – Load Files Analog Table Columns

    Combination View:The combination view shows all of the channel information in a signal view. This view is only available ifthere is sufficient room between analog channels to display two or more lines of text.

    Default Display format:

    Peak to Peak:

    Channel Title

    RMS MaxPeak RefValInstVal MinPeak Units ASV

    RMS Calibrated:

    Channel Title

    RMS MaxPeak RefVal

    InstVal MinPeak Units ASV

    Load Files:

    Channel Title

    MaxWin MaxVal RefVal

    InstVal MinVal Units ASV

    The peak sample values are displayed in red when the data bar is on the channel’s maximum value anddisplayed in blue when the data bar is on the channel’s minimum value. Use the Tab key to togglebetween the analog and digital channels. To hide the channel information, select the "ChannelInformation" menu option from the "View" menu.

    The analog table and combination views can be resized by selecting the vertical separator bar anddragging it to the right or left. The mouse icon changes to the vertical resize cursor when the mouse ispositioned over the separator bar.

  • 8/20/2019 Wavewin Sniffer 24 Manual-4

    46/86

    Chapter 4: System Software

    36 

    To change how the analog data is d

    "Properties" menu option from the "speed bar. The “Analog Table” tabof the analog information window.

    Some of the functions of the “Propecolumns; changing the data position

    colors; and for changing the drivers

    Viewing Digital Data

    The default digital view consists ofof the screen. To view all of the digiDigital Channels” from the “View” m

    The digital trace is displayed as a thin the displayed format and is displaoriginal state. The Cursor State coluthe SEL defined state. These value

    state, “A” = different than original st

    The digital information is displayed i

    Column Number Descriptio

    1 – Cursor State The digital

    2 – Title The chann

    3 – Fst The digital

    4 – Lst The digital

    5 – Fst-Change The time th

    6 – Lst-Change The time th

    7 – Changes The numb

    Table 4.9 – Digital Table Colum

     Use the scroll bar or the up and dokey to toggle between the analog a

    For more information on the analysi

    isplayed in the analog table and combination view s

    ile" menu or click on the “Properties” menu bnd the “Analog Combination” tab allows for changin

    rties” dialog are reordering, hiding, and showing thes in the combination view; changing the backgroun

    data type and trace/phasor scale multipliers.

    nly the triggered digital channels, which are displayal channels including the unused channels press Fenu.

    in black line when the sample value equals the origiyed as a thick green line when the sample value diffmn in the