R E L E A S E N O T E S 716002505, Rev. A Waters is a registered trademark of Waters Corporation. All other trademarks are the sole property of their respective owners. Copyright 2008 Waters Corporation Page 1 of 32 Waters ® Empower2 Feature Release 5 Empower2 Feature Release 5 contains software changes, new features, and enhancements. These release notes contain the following information and instructions: New features and enhancements Problems fixed in Feature Release 5 Known issues in Feature Release 5 GxP compliance requirements Installing the busLAC/E driver Removing Empower 2 Feature Release 5 Feature Release 5 program file updates Test configurations used to develop and evaluate Feature Release 5 Use these release notes in conjunction with release notes that were issued for all previous Empower service packs and feature releases. Unless superseded, all previous release note and product support information apply to this software release. Restriction: Empower 2 Feature Release 5 supports only one database instance. Therefore, do not upgrade servers that contain more than one database instance to Empower 2 Feature Release 5. See: The following guides for information on minimum system and software requirements, and for detailed instructions on installing, configuring, upgrading, and administering Empower 2 software: Empower 2 Feature Release 5 Upgrade Guide Empower 2 Installation and Configuration Guide, Revision B Requirement: Before you install Empower 2 Feature Release 5, if your Administrator-level user account contains any spaces in the name, such as rel lab, you must first add this account name to the ORA_DBA Windows User Group. (Known issue PCS 43239) Empower 2 System Administrators Guide, Revision B Empower 2 Installation and Configuration Guide, Revision A. (Refer to this guide if you are upgrading from earlier versions of the software prior to Empower 2 Feature Release 3.)

Upload: truongbao

Post on 26-Apr-2018

273 views

Category:

Documents


6 download

TRANSCRIPT

Page 1: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

R E L E A S E N O T E S

716002505, Rev. A Waters is a registered trademark of Waters Corporation.

All other trademarks are the sole property of their respective owners. Copyright 2008 Waters Corporation

Page 1 of 32

Waters® Empower™ 2 Feature Release 5

Empower™ 2 Feature Release 5 contains software changes, new features, and enhancements. These release notes contain the following information and instructions:

New features and enhancements

Problems fixed in Feature Release 5

Known issues in Feature Release 5

GxP compliance requirements

Installing the busLAC/E driver

Removing Empower 2 Feature Release 5

Feature Release 5 program file updates

Test configurations used to develop and evaluate Feature Release 5

Use these release notes in conjunction with release notes that were issued for all previous Empower service packs and feature releases. Unless superseded, all previous release note and product support information apply to this software release.

Restriction: Empower 2 Feature Release 5 supports only one database instance. Therefore, do not upgrade servers that contain more than one database instance to Empower 2 Feature Release 5.

See: The following guides for information on minimum system and software requirements, and for detailed instructions on installing, configuring, upgrading, and administering Empower 2 software:

Empower 2 Feature Release 5 Upgrade Guide

Empower 2 Installation and Configuration Guide, Revision B

Requirement: Before you install Empower 2 Feature Release 5, if your Administrator-level user account contains any spaces in the name, such as “rel lab”, you must first add this account name to the ORA_DBA Windows User Group. (Known issue PCS 43239)

Empower 2 System Administrator’s Guide, Revision B

Empower 2 Installation and Configuration Guide, Revision A. (Refer to this guide if you are upgrading from earlier versions of the software prior to Empower 2 Feature Release 3.)

Page 2: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 2 of 32

New features and enhancements

Empower 2 FR5 includes the following new features and enhancements:

New feature/enhancement Description

Support for Windows® Vista™ operating system and a new version of Oracle

Windows® Vista Business Service Pack 1

Windows Vista Enterprise Service Pack 1

Oracle® Enterprise Edition version 10.2.0.3

Microsoft Active Directory LDAP (Lightweight Directory Access Protocol) support

Administrators can manage user accounts using either Local or LDAP authentication. Either form of authentication requires an Empower user name and password to log onto Empower, sign off results, and unlock a locked session.

Waters License Management system A new Web-based license management system that provides an easy-to-use interface to register, activate, deactivate, and view licenses for Empower 2 base software and software options that you purchased.

Requirement: On Windows Vista computers, you must use the new Web-based License Management utility. If you attempt to install a key disk on a Vista computer, the Program Compatibility Assistant displays a message recommending that you install using a program that is compatible with the current version of Windows.

On a Windows Server 2003 or a Windows XP client, you can use either key disks or the new license management system, but not both types.

On a Windows 2000 Server, you must use key disk licensing. Alternatively, you can use the new license management system on a Windows XP client or a Windows Vista client connected to the Windows 2000 Server.

You can use only one type of license management system.

Updated PDF file generator Compatible with Adobe® Acrobat® 8.

Page 3: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 3 of 32

New feature/enhancement Description

Empower 2 Documentation Bookshelf CD

The new Empower 2 Documentation Bookshelf CD (Revision A) contains new and revised documentation for Empower 2 Feature Release 5, as well as existing documentation for Empower 2, in English. The System Administrator’s Guide, revised for Feature Release 5, is also available in Chinese, Japanese, and Korean.

As other language versions of the Empower 2 documentation become available, they will be included on this Multilanguage documentation CD.

Limitation: On Vista only, when you insert the documentation bookshelf CD into the CD-ROM drive, it does not automatically run the bookshelf interface. There is a known issue with Autorun.ini files on Vista. To run the documentation bookshelf interface, navigate to the CD top-level folder, and double-click the Empower2Bookshelf.chm file.

Requirement: To view the Asian language pages of the Empower 2 Documentation Bookshelf on Windows XP, you must install the Asian languages. To do so, from the Start menu, click Settings > Control Panel > Regional and Language Options > Languages tab > Install Files for Asian Languages > click OK.

Method Validation Manager Various enhancements, including those listed below:

Enable multiple sample preparations and injections

Calculate %RF, as well as RF (Response Factor).

New calculated fields in the Supportable Estimation Table for the Intermediate Precision test.

New Apply Table Preferences toolbar in the Sample Set Method Editor, Run Samples, QuickStart, and Alter Samples with default preferences to filter only the fields needed for each type of method validation test. Although initially conceived for Method Validation, this capability is available to all users.

Gas Chromatography enhancements New GC control parameters for the 6890-series gas chromatograph (6890A, 6890 Plus, or 6890N)

See also: Instrument Method Editor > Auxiliary, Events, or Valves tab > Press F1 on your keyboard for context-sensitive Help.

Empower 2 Help improvements You can now access all available Help topics from the Search tab in the Empower 2 Online Information system. To access the Empower 2 Online Information System, from any Empower Help menu, click Empower Help Topics.

The application-specific Help menu items that began with the words “Help on”, such as Help on Configuration Manager, have been removed from the Empower Help menus, as these Help topics are now included in a new Reference section in the

Page 4: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 4 of 32

New feature/enhancement Description

Empower 2 Online Information system.

System Suitability Calculation Empower includes a new formula to calculate plate count as specified by Japanese Pharmacopoeia 15 (JP15), and retains the current plate count calculation.

Sample plate enhancement You can now import plate type definitions from a text file into Empower.

Quick Start preferences You can now save Quick Start window preferences.

Problems fixed in Feature Release 5

The following table contains a list of the software problems fixed in Feature Release 5. The PCS (Product Change System) numbers identify software issues that Waters personnel monitor within a system change request tracking tool.

Problems fixed in Feature Release 5:

PCS number Description

21060 6890 - Added FID and FPD Lit Offset parameters.

35145 For 6890 GC instruments, the application appeared to round off the Split Ratio value calculated from the entered flow values and then adjusted the Split Flow to be consistent with the new, rounded off Split Ratio value. This was corrected by adding more digits of precision to the inlet flow parameters (to avoid round-off errors in flow) and allow a more exact value for gas flow in GC methods.

36107 Batch processing did not always extract and process all channels of a method set for large sample sets with large data files.

Improved memory handling fixes this issue.

However, if the processing server runs out of memory while processing a large batch job, the Message Center now displays one or more memory mapping messages or messages about fetching a method, followed by this final message: “Background processing job aborted due to a memory allocation failure”.

37031 EMD 1000 – Edit method parameters and Open Access masses were not acquired properly when the method set was run using certain instruments. Updates to the XML formatting logic corrected this issue.

37151 5890 GC – The purge valve switched to an off state during instrument setup. The functionality was updated so that if the valve is enabled, its position is not changed at setup unless you program a specific event.

37272 5890 GC – Attempting to run a sample set with multiple injections on the same line resulted in an error message. Updates to the error handling logic corrected this issue.

Page 5: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 5 of 32

Problems fixed in Feature Release 5:

PCS number Description

37318 A maximum column precision of 14 is now available in the Component Editor Amounts columns in Alter Sample, the Default Amounts tab of the Processing Method, and in the Sample Weight and Dilution fields. Rounding was corrected in the few cases where numbers with increased precision were rounding incorrectly.

37716 The “View Sample Values Used in Calculations” field showed multiple values. This condition no longer occurs.

37751 Duplicate results for the same channel were created in a Results Set without notification.

To determine whether duplicate results were superseded with newer results, add the new “Results Superseded” column to the Result table.

37887 6890 GC with G1512 injector controller. If the “next inject delay” option was not used, the instrument failed after the first injection. This has been corrected such that the use of the “next inject delay” option is no longer required.

38047 6890A and 6890 Plus GC injection volume was incorrect when a nanoliter adapter was not installed. This has been corrected.

To set syringe size parameters for the particular injector you are using, click the “GC Sample Set Info” button.

See also: PCS number 41880

38495 To run dual-tower injections on a GC instrument, the instrument required two inlets, two detectors, and two injectors. You can now use third-party injectors to run dual-tower injections on the GC instrument, but the instrument still requires two inlets and two detectors.

38520 The solvent composition of an isocratic instrument method sometimes changed after saving the method.

This condition no longer occurs.

38596 Users who had the Alter User privilege (user account properties) could alter an account to a user type with more privileges.

The list of available User Types will now only reflect User Types that have privileges equal to or a subset of those of the account that is accessing the list.

A new privilege called “Assign User Types to Users” has been added. Users who have the “Assign User Types to Users” privilege can assign any user type (except Admin) to user accounts.

See also: PCS number 40857

38665 It was difficult to display the correct fields in the Sample Set Method Editor/Alter Sample for the validation test(s) to which the sample set applies.

A new Apply Table Preferences toolbar enables you to display the fields required to complete a validation test.

Page 6: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 6 of 32

Problems fixed in Feature Release 5:

PCS number Description

38748 Some systems experienced Server Busy messages, MFC errors, and required LAC/E module reboots because the number of Oracle processes had been set to 150.

The software now sets the minimum number of processes to 500 at installation or upgrade.

39070 Empower/SDMS Restore window sometimes displayed an incorrect project list. In order to resolve this problem, a new toolkit call was added for use in a future release of SDMS.

39085 In a global project, when creating a view filter for injections, channels, or results, you cannot specify a sample name condition such as =PQ* or =#. The moment you apply a newly created view filter with a condition for sample name, the sample name column disappears from the view. However, the default view filter includes the sample name column.

39097 In Method Validation Manager, in the Method Validation Linearity Test, %Response Factor was not calculated.

The Linearity Test now calculates the %Response Factor and %Response Factor Mean.

39173 Changes to the W600 pump gradient program were not applied when it was run. Flow from a previous method version was used between injections.

The software now sets the flow rate and percentages in the isocratic fields to always match the percentages in the first line of the gradient table.

40091 Windows Server 2003 on a Citrix Server failed the Verify Files check, indicating that the operating system was unsupported.

This has been corrected for new installations only.

40106 Added a new formula for plate count calculation as specified by Japanese Pharmacopoeia 15 (JP15) while retaining current plate count calculation.

40151 Sign Off 1 user was presented with a full list of reasons despite System Policy restrictions.

Now, when users enter an invalid reason for Sign Off 1, they will receive the following message:

“This reason is not in the list of Valid Signoff 1 Reasons required to proceed with Signoff 2. Do you wish to continue with this sign off?”

Users must enter a valid Signoff 1 reason in order to assure a successful Signoff 2.

See also: PCS numbers 41566 and 41977

40326 6890 GC - could not select the detector makeup gas. Makeup gas types are now available for FID, FPD, NPD, and TCD.

Page 7: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 7 of 32

Problems fixed in Feature Release 5:

PCS number Description

40365 The error message “Connection to chromatographic system lost, eventID 54” was caused by instrument problems that would generate an error condition that the busLAC/E driver software could not handle.

New busLAC/E driver software is now installed to handle the error condition. A new capability is also available to report Instrument Server events into a log file, which you can activate as part of a troubleshooting process.

After installing Empower 2 Feature Release 5, you must manually install the busLAC/E driver.

See: The section in this document entitled “Installing the busLAC/E driver”.

40408 The Summarize Custom Fields function does not work in Empower 2 as it did in Empower 1154. The Summarize Custom Fields function will continue to work as it currently does in Empower 2.

The new Summarize Custom Fields Incrementally function provides the capability to summarize custom fields in the way that the Summarize Custom Fields function did in Empower 1154. These two additional functions were also added to vary the summarize custom field conditions:

Summarize Custom Fields Incrementally (Exclude Faulted)

Summarize Custom Fields (Exclude Faulted)

40487 Percent Recovery calculation needed more flexibility to accommodate "corrected" spike/recovery experiments.

You can now set the assessed field in the accuracy test to either % Recovery (the default) or a user-created real peak custom field.

40522

In Method Validation Manager, the Exceeds Threshold flag in the Filter Validation and Stability tests was being incorrectly set when a reference was used.

This field now works as it should.

As a result of this correction, when a reference filter is used in the Filter Validation Test, or a reference storage condition is used in the Stability Test, the values in the following fields will be calculated correctly:

In the Data Table: “Total Area Exceeding Threshold”, “Corrected %Area”, “Relative Peaks Exceeding Threshold”, and “Faulted Relative Peaks Exceeding Threshold”.

In the Relative Peaks Table: “Corrected %Area”

40638 The AutoArchive utility might crash when a project is queued and then deleted.

AutoArchive now checks first to determine whether the project exists before it runs project integrity. If the project does not exist, it logs a message in the Message Center for the project.

40649 The DBProject.BackupProject toolkit function failed in some cases due to an incorrect Oracle_Home setting.

This issue has been corrected.

Page 8: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 8 of 32

Problems fixed in Feature Release 5:

PCS number Description

40654 In Method Validation Manager, the Accuracy Test did not allow the use of a custom field.

For the Accuracy test, you can now set the assessed field to %Recovery (the default) or to a user-created real peak custom field.

40725 When using expected mass, manually adjusting the end point of the last peak in a result produced an MFC error in review.

This error no longer occurs.

40754 In Method Validation Manager, Dilution and Sample Weight fields were not working correctly.

These fields now work as they should in Method Validation calculations.

40816 In Method Validation Manager, field names in the Supportable Estimation Table were unclear.

The fields in the Supportable Estimation Table have been renamed.

40848 The Dilution field did not input or display certain values correctly.

The software now restricts the display of all double precision numbers (both user entered and calculated) to a maximum of 15 significant figures. Rounding was corrected in the few cases where numbers with increased precision were rounding incorrectly.

40857 Non-Administrative users who have privileges to create and alter user accounts could not see the entire list of "Allowed User Types" (excluding the Admin user type).

For users who create and alter user accounts, to view the entire list of “Allowed User Types”, you must have both the new privilege “Assign User Types to Users” and the existing privileges “Alter User” or “Create User”.

See also: PCS number 38596

40861 In Method Validation Manager, additional calculations were needed in the Supportable Estimation table of the Intermediate Precision Test.

The Supportable Estimation Table now includes the fields: Total DF, Total SD LCL, Total SD UCL, Total %RSD LCL, and Total %RSD UCL.

40884 The Dissolution plot overlay could be set to draw lines to the origin. This has been corrected.

See also: PCS number 41056

41016 In Method Validation Manager, the use of the term “Experiment” for a column name in the Supportable Estimation Table was incorrect.

The table column was renamed to “Points/Experiment” to improve clarity.

41054 In Method Validation Manager, you could not specify multiple preps and injections.

Page 9: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 9 of 32

Problems fixed in Feature Release 5:

PCS number Description

Rather than having to override the red X, you can now set a checkbox in the test parameters to allow the use of both multiple preps and injections.

41056 The software did not include the capability to extrapolate plot lines from the origin to the first dissolution point of a dissolution plot overlay.

This capability is now provided by setting the new Line to Origin field in the Dissolution Plot properties to True. Note that the dissolution assay should be validated over the range of values plotted.

See also: PCS number 40884

41116 The capability to save layout preferences in the QuickStart User Interface was missing.

This capability is now available.

41119 Under certain circumstances, when using exponential notation, the Component editor might incorrectly round values.

Rounding was corrected in the few cases where numbers with increased precision were rounded incorrectly.

41151 A user logged on with a Guest user type was able to sign off results when this user was also assigned a user type with the privilege to sign off results.

The software now grants you only those privileges assigned to the user type that you select when you log on. If you have multiple user types, such as Administrator and Guest, when you log on as Guest, you have only Guest privileges.

41163 Plate definitions could be exported as a text file, but the capability to import plate definition text files into Empower was not available.

You can now import plate type definition text files into Empower, so long as you have the Create Plate Type privilege.

41327 When creating an instrument method with a gradient using any composition and flow rate, if you changed the flow rate within the instrument method, the software downloaded the initial condition flow rate, not the newly specified flow rate.

To resolve this problem, the software sets the flow rate and percentages in the isocratic fields to always match the flow rate and percentages in the first line of the gradient table.

41336 The message “Unable to store Post-Run report” prevented the upload of buffered data.

When the size of a post-run report now exceeds 32K during buffering, the software will recover the data, but it cannot save the post-run report because it is too large. The software displays a message stating this condition.

41343 Confirm ID for User Changes can be bypassed by cancelling.

This condition no longer occurs.

Page 10: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 10 of 32

Problems fixed in Feature Release 5:

PCS number Description

41346 “Custom Field Type" can be changed by overwriting.

To resolve this issue, you cannot copy a custom field that has the same name as an existing custom field, when the two fields have different types.

41426 Large batch processing of Mass Spectrometer data did not extract and process all channels of a sample set.

Improved memory handling resolved this issue. Large batch processing now runs successfully.

41541 6890 Method parameters were not downloaded to an instrument if components are not enabled.

This problem no longer occurs. The software sends the initial conditions to the instrument even when the column is not selected for use.

41566 Free text was permitted for Sign Off 1 despite System Policy restrictions.

A message now appears when signoff 1 is invalid, indicating that a valid reason for signoff 1 is required in order to complete signoff 2.

See also: PCS number 40151 and 41977

41567 Locked channels could be unlocked without requiring a comment and without requiring an audit trail message.

This condition no longer occurs.

All lock channel and unlock channel actions are now recorded in audit trail.

41590 Under certain circumstances, Review might crash when attempting to manually add a drop line in the region of a deleted peak.

This condition no longer occurs.

41659 6890 Inlet purge valve did not hold pressure of zero in Constant Pressure mode.

The instrument control logic was modified so that the instrument will be left in its current state (such as ‘off’) when both purge On Time and purge Flow Rate are set to zero.

41662 When producing a summary report for bracketing with overlap, each standard from an overlapping bracket produced two results (duplicate results).

This issue was fixed by adding the “Include Duplicate Results” checkbox on the properties for the Table report group. Select or clear this checkbox to specify whether you want to include or exclude duplicate results.

41685 When User A was logged into Empower, User B with a valid user name and password was able to sign off results in User A’s session.

Signoff authority is now verified against the logged-in user.

Page 11: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 11 of 32

Problems fixed in Feature Release 5:

PCS number Description

41756 Toolkit: When a user other than Administrator used DBProject, an invalid privilege error was thrown on project backup.

This condition no longer occurs.

41820 Users sometimes experienced problems saving the archive file during an Archive and Remove Audit Trail process because a temporary audit trail table space was not deleted.

This condition might have occurred when attempting to save the audit trail to a mapped drive path containing spaces in the name.

Now, if you attempt to save the audit trail to a mapped drive name containing spaces, the error message that appears now clearly states that mapped drive paths cannot contain spaces, and the temporary audit trail table space will be deleted.

41858 After running System Precision Signal to Noise with certain instruments, a ProcMon MFC Application Error might occur.

This condition no longer occurs.

41880 6890A and 6890Plus GC injection volume was incorrect with G1512A controller.

This has been corrected. To set the syringe size parameters for the injector you are using, click the GC Sample Set Info button.

See also: PCS number 38047

41917 Percent Amount was not calculating results correctly.

The Amount values used to calculate %Amount are no longer rounded to 3 decimal places prior to performing the calculation of %Amount.

41967 User authentication was tracked per client instead of per user.

The software now tracks authentication attempts (and failures) for each user account.

41977 When modifying the default string for Signoff 1, the failed signoff was not reported until signoff 2.

If a default string is modified, it must now be selected again as a valid reason for signoff.

See also: PCS number 40151 and 41566

41996 Setting a password on the Email Center for Empower 2 in Japanese, Chinese, or Korean prevented further access.

This condition no longer occurs.

42002 After installing e-mail center in Japanese, messages are garbled.

Implemented the character sets and encoding for Japanese, Chinese, and Korean.

Page 12: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 12 of 32

Problems fixed in Feature Release 5:

PCS number Description

42003 Setting a password on AutoArchive Properties for Empower 2 Japanese, Chinese, or Korean prevented further access.

This condition no longer occurs.

42015 Method Differences did not show instrument parameters or parameter values for COM instruments.

This condition no longer occurs due to Empower software; some instruments may require updated instrument drivers in order to fully show all method differences.

42043 On clients connected to UNIX (Sun Solaris), the Method Validation Review Data Table did not sort by Result Id.

The Data Table now sorts by Result ID on clients connected to UNIX (Sun Solaris), as it currently does on Windows platforms.

42079 GPC data with an exponential skim was only shown in the chromatogram view.

This condition no longer occurs. All skims are taken into account when plotting a distribution.

42084 The ability to create a custom field that is the product of the peak fields in a chromatogram was missing.

You can now determine the product of peak fields using the new multiplication summary function PROD that works similarly to SUM, AVG, and other summary functions.

42106 The system suitability summary plot group worked correctly for peak type fields, but it did not work correctly for result type fields.

This problem has been fixed as follows: The Result, Calibration, and Cal curve fields now use the Result field “Channel Name” to separate plots for each unique channel. Channel Name is also used as the plot title. All other fields plotted on the Y axis that apply to channels will continue to use the Chromatogram field “Channel”.

42138 The requirement to view all Empower 2154 signoff pages could be bypassed by printing the reports.

This issue no longer occurs.

42158 Instrument Server would hang with rapid abort/inject selections with 2489.

This issue no longer occurs.

42166 The ability to suppress legends on Spectrum plot was needed.

The Spectra tab in the Plot Properties dialog box includes a new checkbox called “Include Channel/Spectra Description”. Use this checkbox to enable/disable the display of channel/spectra descriptions. In Report Publisher, the default font size for legends has been changed to 8 point.

Page 13: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 13 of 32

Problems fixed in Feature Release 5:

PCS number Description

42355 Empower 2 displayed wrong unit label for micro-ECD GC data.

This issue no longer occurs.

42594 Setup error occurred when using a 6890 GC Gas Sampling valve (GSV) for injections.

Empower software includes two new controls, one for each possible GSV to support GSV injection capability. A GSV injection and a tower injection cannot occur for the same injection.

42595 Method Validation Manager: The Validation Test Data table showed duplicate results with multiple signoffs.

This condition no longer occurs.

42617 MVM - Japanese translation for Q-Factor was incorrect.

Translations for “Sample Preparation” and “Q factor” have been corrected.

42618 MVM - Japanese translation for busLAC/E Scan was incorrect.

The translation has been corrected.

42620 MVM - Japanese translation for Sample Preparation. The translation was changed to match the correct translation in Feature Release 3.

Online Help defects

New information, enhancements, and corrections added within the Empower 2 online Help.

See also: The Empower Software Cumulative Release Notes.pdf file in the root directory on the Empower 2 Feature Release 5 disk. This document contains a cumulative list of changes previously implemented in Empower 2 service packs and feature releases.

Known issues in Empower 2 Feature Release 5

Known issues and workarounds:

PCS number Description of known issues

37122 For Japanese, Chinese, and Korean, the Empower 2 installer does not place the 'Manage Email Center' shortcut on the workstation desktop (Windows XP only).

Workaround: To activate email notification using the Manage Email Center, navigate to the <drive:>\Empower\Bin folder, and double-click the SetEmailDatabase.exe file.

Page 14: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 14 of 32

Known issues and workarounds:

PCS number Description of known issues

41811

When performing an archive and remove, a message displays: “Archive Failure – could not create archive file. The most likely cause is access or free space issues at the destination.”

When selecting “Archive and Remove System Audit Trail,” if you are archiving to a network location, you must use a mapped drive.

41969 When viewing PDA Libraries in Configuration Manager using the following procedure, the spectra no longer appeared in the selected library:

1. Expand the Libraries tree, and then select Libraries.

2. In the Libraries list, right-click a library, and then click Open.

3. While the library is open, collapse the Library tree view, and then expand it again.

4. In the Libraries list, right-click a library, and then click Open.

Result: The library appears empty (no components listed) because the value of “Max Rows” changed to 0.

To fix this problem, change the value of Max Rows to 1000, and update the view. The spectra now appear in the library.

42113 The newly implemented Wash Needle function does not work with v1.00 of the 2707 Instrument Control Software (ICS). A new version of the ICS will resolve this issue.

42492 When using Empower 2 FR5 on Windows Vista, and opening the TQD Tune page Diagnostics tab, if the information below the Vacuum, Power Supply, and RF fields is not visible on the screen and resizing the window does not correct the problem, increase your screen resolution settings until you see the entire window.

42493 The default screen resolution might not be sufficient to view the Function Details of an MS function.

To increase your screen resolution settings, right-click Desktop > Properties > Settings tab, and increase the settings until you can see the entire screen.

42510 The amount values in the Component Editor for Sample Set Method History are incorrectly set to default precision.

Empower 2 now saves changes for real numbers in samples, sample sets, and component amounts, when those changes increase the number past the default precision (maximum of 14 digits). However the sample history and sample set history continue to format the numbers with the default precision. Therefore, if the changes increase the number past the default precision of 14 digits, and the before and after numbers format to identical strings, the sample history states that the number was changed to the same number.

Page 15: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 15 of 32

Known issues and workarounds:

PCS number Description of known issues

42599 With Empower 2 FR5 and a ZQ instrument driver are installed on Vista Enterprise, if you configure IIS (Internet Information Services) and FTP settings using this procedure, Windows crashes: Computer Management > Services and Applications > Internet Information Services > FTP Sites > Default FTP Site.

Workaround:

Set up FTP and IIS (required):

1. From Control Panel, select Programs and Features, and then select the option Turn Windows Features On or Off.

2. In the Windows Features dialog box, select Internet Information Services, and turn on FTP Publishing Service, FTP Management Console, and FTP Server.

3. Under Web Management Tools, turn on ISS Management Console.

4. Accept these changes, and reboot the computer, as necessary.

Configure IIS and FTP:

1. Access Computer Management > Services and Applications > Internet Information Services (IIS) Manager (not Internet Information Services IIS 6.0 Manager).

2. In the center window of IIS Manager, click the connection > FTP Sites.

3. In the right window, click the item “Click here to launch”.

4. In the Internet Information Services (IIS) 6.0 Manager window, expand the tree, and right-click Default FTP Site, and then click Properties.

5. In the Default FTP Site Properties dialog box, configure properties, as required.

42600 On Vista SP1, when installing IIS and FTP for the ZQ or EMD mass detectors, the related FTP Service does not start automatically when you reboot the operating system, as it should.

To resolve this problem, configure the FTP Publishing Service to start automatically. To do so, from Computer Management > Services, in the list of services, right-click FTP Publishing Services, and then click Properties. In the Start up Type box, select Automatic.

42610 On Windows Vista SP1, problems might be encountered when establishing communications with a ZQ or EMD detector.

You must add the FTP Server to the list of Windows Firewall exceptions, and then you can reboot and successfully establish communications with the ZQ or EMD mass detectors.

Page 16: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 16 of 32

Known issues and workarounds:

PCS number Description of known issues

42628 When using Empower on a Vista computer, when you start a new application window, such as Validation Manager from the Project window, the Project window remains the active window, and the Validation Manager window opens behind the Project window.

Workaround:

In the Windows registry, modify the ForegroundLockTimeout registry key to be Decimal, and enter a value of 0 (Microsoft KB886217).

See also: www.support.microsoft.com/kb/886217 for details.

42682 When creating a project in situations when the user tablespace cannot auto extend, Empower might display the message “Error copying from project xxx”.

42800 When controlling an SQ, TQ, or 3100 mass detector on a workstation or acquisition client, you must install Empower 2 (typical or custom installation) on the C drive of that PC.

If you install Empower 2 on a drive other than the C drive, and you start an acquisition, errors occur.

42850 Although the WebServer feature and references to it are active in Empower software, this feature is unsupported in Feature Release 5.

42882 Using either LDAP or local authentication, if a user logs into Empower and receives the message “Authentication Failed”, an incorrect password is the most common reason for this problem. However, in LDAP mode, authentication can also fail for any of these and other reasons:

The user account was disabled or no longer exists on the LDAP server.

The minimum number of password characters changed on the LDAP server.

The user account underwent a department change, resulting in a change to the user’s Fully Qualified Domain Name (FQDN).

42894 The “Database/Import MassLynx Data” option from the Empower Project window is a legacy application that is not intended for importing MS data from mass detectors, such as the SQ, TQ, and 3100 mass detectors, and other newer-model mass detectors.

There are other data import capabilities available for importing more modern MS data.

Page 17: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 17 of 32

Known issues and workarounds:

PCS number Description of known issues

42978 In Method Validation Manager, the Korean translation of the word “Preparations” changed from “Jun-bi” to “Si-yak”. This translation change impacts following features of Method Validation:

When you create a new Method Validation Manager project in Empower 2 FR5, do not copy any custom fields from a project created in a previous version of Empower 2. Doing so copies the old translation of the custom field name. If this happens, you can manually change the translation of these values in the table in the Project Properties Custom Fields tab.

Although the Valid_OQMVM project contains the old translation for the Sample Preparation custom field, the OQ runs correctly. However, the name of the custom field and the custom field name that appears in the Method Validation UI will differ.

43020 In Empower 2 Feature Release 5, when connecting an Empower node to a TQ mass detector for the first time, the error message “An unnamed file contains an invalid path” appeared. Click OK to close the message.

Empower establishes communication with the instrument and the message no longer occurs.

43048 During an Empower 2 Feature Release 5 Client installation, the Windows Security Alert message “Windows Firewall has blocked some features of this program” might appear.

Click Unblock to continue the installation.

43072 When installing Waters Email Service on a Windows Vista Enterprise SP1 computer, the Email Center registry keys were not installed correctly.

When installing the Waters Email Service, you must log in locally as Administrator. On the program executable file, right-click, and then select Run as Administrator.

43125 In Method Validation online Help topics, the null hypothesis for Linearity Regression ANOVA and Residual Linearity ANOVA is incorrect.

Correction:

The null hypothesis for Regression ANOVA is that the model error is zero as compared to residual error, or the linear regression is adequate.

The null hypothesis for Residual ANOVA (lack-of-fit) is that the lack-of-fit error is zero as compared to pure (random) error in replicates at each level, or the residual error is pure (random) error.

43134 During the installation of Empower 2 FR5, multiple DOS windows that display information might appear, and then disappear from the desktop.

When these windows display, do not press any keyboard keys or take any action in these windows. Doing so might cause the installation to pause or slow down. If you do any of these actions by mistake, press the Esc key on your keyboard.

Page 18: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 18 of 32

Known issues and workarounds:

PCS number Description of known issues

43136 When you configure a 6890 instrument using both a GSV and an injector to inject in the same position, the instrument method instructs the system to use the GSV to inject.

43138 If you used Japanese characters in the Default (name) From field in the E-Mail Policies, the Japanese Email Center did not send any email messages.

To prevent this problem, in the Configuration Manager Empower System Policies dialog box, on the E-Mail Policies tab, if you specify an outgoing email server that does not support Chinese, Japanese, Korean, or other such characters in a normal email address or alias, do not use these characters in the Default Form Name field. Otherwise, the Email Center might not deliver your email message correctly, or it might not deliver it at all.

43160 You cannot run sample sets that specify Acquity vial position V:1 in the plate/well column. Empower displays an error message indicating an invalid plate number.

43196 In the Chinese, Japanese, and Korean versions of Empower 2, the Waters Email Service installation does not create the desktop shortcut correctly, using a partially translated path to “Windows Desktop”.

43202 During installation, the Empower 2 Client Installer does not display its progress in the Progress bar.

43239 To prevent database access problems, before you install Empower 2 Feature Release 5, if your Administrator-level account contains any spaces in the name, such as “rel lab”, you must add this user account name to the ORA_DBA Windows User Group.

43316 A sample set using a ZQ instrument set with the Japanese version of Empower 2 Feature Release 5 in a Windows Vista environment might not run.

To resolve this problem, in Windows, modify the long date format to use the English long date format “dddd, mm, dd, yyyy”.

43325 The ZQ2000, ZQ4000, and EMD1000 instrument calibration will not operate properly on Windows Vista Japanese until the instrument driver can be revised.

43387 After upgrading from previous versions of Empower 2 to Empower 2 Feature Release 5, when you log into Empower 2 Feature Release 5 for the first time, the Message Center might display a warning message indicating that Oracle Error 00904 has occurred on the “INVALIDPASSATTEMPT” column. The Empower 2 Feature Release 5 upgrade process creates this column later in the process. This message will not occur on subsequent logins.

43413 On a Vista Enterprise 32 bit system, when installing or removing an Acquity ICOP (instrument control option pack), the progress bar indicator begins to fill, but then appears to stop, and the Wizard page also appears as though it is not responding.

Do not cancel the operation because in most cases, the installation or removal process is running. Check the status bar at the bottom of the window for activity, and give the installation or removal process time to complete.

Page 19: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 19 of 32

Known issues and workarounds:

PCS number Description of known issues

43430 After performing a server upgrade, when you run the batch script file that moves the Index and Users files from the old database instance to the new database instance, if these files already exist in the new database instance, A message appears, informing you that these files already exist.

When this happens, you must manually delete the Index and Users files from the new database instance, and then run the batch script file again to move the two files from the old instance to the new instance.

43443 Some topics in the online Help contain references to an incorrect book title, due to changes in the documentation set that occurred after the online Help was released.

Replace all references to the Empower 2 Upgrade and Configuration Guide, Rev. B with the Empower 2 Feature Release 5 Upgrade Guide.

43496 When upgrading to FR5, during the database update, the time zone is missing from the date/time stamps of Audit Trail entries recorded before the “Created Empower Node” record. The time zone is also missing from the date/time stamp for the database update messages in the Message Center.

43502 When you have multiple channels in Run Samples, then use Customize Channels and deselect all of the channels, when you use the mouse to resize the channel plot area, an MFC error occurs.

To avoid this problem, you must enable at least one real-time channel plot before resizing any of the plot windows in Run Samples or in QuickStart.

43515 On Windows Server 2003, firewall settings are not automatically set for the Watersservice installer.

Refer to the white paper entitled "Waters Configuration Notes for Empower 2 Enterprise Server on Windows 2003 Server SP1 and R2", Revision A, dated Aug 22, 2006.

43523 The Empower 2 Online Help does not include information for specifying the minimum number of data points required to calculate any of the noise/drift results.

A minimum number of 30 data points must be included in the noise interval. When examining the minimum interval necessary for data collected at 1 Hz to report an Average Peak to Peak Noise value, 30 seconds is the smallest interval that will generate a noise result.

You must specify a minimum number of points to avoid biasing the apparent noise to low values by allowing the choice to result in the linear regression doing little more than "connecting the dots" within the noise interval.

43525 When using Empower 2 Feature Release 5 on a Personal configuration running Windows Vista Business with SP1 or a client configuration running Windows Vista Enterprise, the Waters DHCP Server service randomly stopped working, and displayed a message indicating that it stopped working and was closed.

You can restart the service, as needed.

Page 20: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 20 of 32

Known issues and workarounds:

PCS number Description of known issues

43545 In Empower 2 Feature Release 5, when printing a report for a 6890 instrument method, the System Information report group reports incorrect values for “Valve”.

This reporting error has no impact on the ability to acquire data nor the validity of the data acquired.

43546 If the Personal installation process fails to load the UpdateDBUtils.dll when preparing to import the exported TransportableTablesspaces, a message box appears containing the resource ID identifier and this string:

W_RESTOREBACKUP_RETRY

The message box should display the following information:

The upgrade of your database to Empower 2 Feature Release 5 has not completed successfully. Please permit the Empower 2 FR5 installation wizard to run to completion. You will then need to uninstall Empower 2 FR5 using Add/Remove programs. Once this is complete, you will need to restore your Empower database backup into the appropriate version of Empower 2 and then upgrade to Empower 2 FR5.

43547 Windows Firewall exceptions might appear twice in the Windows Firewall Exceptions list, after installing Empower 2 FR5 and Instrument Control Option Packs (ICOP).

This behavior is expected as a result of the fix for PCS 43485, which automatically enables and maintains general Windows Firewall Exceptions for Empower 2 and Ethernet instruments.

Do not remove the duplicate entries, as doing so might remove the required Firewall Exceptions.

43548 In the Korean and Chinese versions of Empower 2 Feature Release 5 on a Vista Client, when background printing a result using a default report method within the Defaults project, some report fields either displayed small squares or appeared blank. Information appeared correctly in Preview/Publisher.

Tip: This behavior occurred only on Empower 2 Workgroup and Enterprise. It did not occur on Empower 2 Personal. This behavior also did not occur when selecting the Report Method option “Use a Report Method that is appropriate for the data”.

This problem happens because the report method used for "background printing" called "Default Individual Report" in the "Defaults" project uses the wrong font.

Workaround: Edit the Report method and change the English font to the correct font for your language.

43554 In Empower 2 Feature Release 5, when you restore multiple files in one operation while the Confirm Identify policy is set for projects, and you cancel the Confirm Identity dialog box for one or more of those projects, the restored projects might not contain any data or audit trail information.

You can delete these projects from Configuration Manager.

Page 21: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 21 of 32

Known issues and workarounds:

PCS number Description of known issues

43557 When installing instrument control software (ICS) from the Empower 2 Feature Release disk, the ICOP installation failed with an "Error - 0" error.

To successfully install the ICS, on the ICOP Utility disk, navigate to the Instrument folder, and double-click setup.exe.

43559 In Empower 2 Feature Release 5 on a Citrix client, when copying and pasting a library from the Libraries view in Configuration Manager, Configuration Manager crashed. When copying a library from networked drive on the Citrix client, the message "The user must have an LDAP user DN was not found" appeared. After closing the message box, a "Copy in Progress" toolbar appeared, but nothing happened.

You can import a library successfully from a Citrix client.

43564 After upgrading an Empower 2 Enterprise system to Empower 2 Feature Release 5, the Waters Licensing Wizard might display the message “Invalid Option 23”.

This message is an artifact of the previous licensing strategy, and does not affect system operations or the validity of installed licenses.

43565 Before upgrading an Empower 2 database server (Feature Release 4 and Service Pack E) to Empower 2 Feature Release 5, used the Enterprise Edition key disk to activate the Dissolution option. After completing the upgrade, used the Waters Licensing Wizard to view the installed options. The Licensing Wizard listed the Dissolution option as an Empower Workgroup instead of Enterprise Edition.

This problem is a display problem only that happens when viewing licenses in the Licensing Wizard that were previously enabled using key disks.

43588 After upgrading an Empower acquisition client computer with installed drivers for the ZQ2000, ZQ4000, EMD, and Trinity SQ, Trinity TQ, and 3100 mass detectors to Empower 2 Feature Release 5, the Waters Service automatically starts and runs.

In order for the mass detectors to run correctly, you must disable the Waters Service from within Windows Services. After you disable the Waters Service, reboot both the computer and the embedded PC (EPC) in the mass detector.

Page 22: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 22 of 32

GxP compliance requirements

When you install Empower Software for the first time, or when you uninstall software, Waters recommends that you perform software qualification to meet GxP compliance requirements. Any time you install or uninstall software in a GxP environment, Waters recommends that you follow your company’s management-approved software change management standard operating procedures. To assist you in making software change management decisions, refer to the section in this document that describes the program file updates contained in this feature release.

To verify that the installation of this feature release does not affect any data, reprocess a data set before and after you install the software, and then verify that the results are equivalent.

Waters recommends that you run either the Verify Files utility or the ConnectionsAQT for Empower IQ. After you run one of these utilities, review the resulting file for an entry that states “No installation changes were detected.”

Installing the busLAC/E driver

After you install Empower 2 FR 5, install and configure the busLAC/E driver. The procedure differs slightly between Windows XP and Windows Vista.

Tip: If the Found New Hardware Wizard dialog appears automatically, start with step 5. If not, follow the procedures to access the Wizard manually.

To install and configure the busLAC/E driver on Windows XP:

1. Right-click My Computer, and then select Manage.

2. On the Computer Management page, click Computer Management (Local) > System Tools > Device Manager. In the right-hand pane, locate Other devices > PCI Device.

3. Right-click PCI Device, and then select Properties.

4. In the General tab on the PCI Device Properties page, click the Reinstall Driver button. The Found New Hardware Wizard page appears.

5. On the Found New Hardware Wizard page, select “Yes, this time only” and then click Next.

6. Select “Install from a list or specific location (Advanced)” and then click Next.

7. Select “Search for the best driver in these locations” and select the checkbox “Include this location in the search.”

8. Click the Browse button and include the following directory in the search: C:\Empower\BuslaceDrivers.

9. Click Next.

Result: The BusLAC/E Driver installs. It appears as “Buslace PCI KMDF Device Driver” under BusLAC/E Drivers in Device Manager.

Page 23: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 23 of 32

To install and configure the busLAC/E driver on Windows Vista:

1. Right-click Computer, and then select Manage.

2. On the Computer Management page, click Computer Management (Local) > System Tools > Device Manager. In the right-hand pane, locate Other devices> PCI Device.

3. Right-click PCI Device, and then select Properties.

4. In the General tab on the PCI Device Properties page, click the Reinstall Driver button. The Found New Hardware - PCI Device page appears.

5. On the Found New Hardware - PCI Device page, select "Browse my computer for driver software (advanced)."

6. Under “Browse for driver software on your computer”, click the Browse button and select the directory C:\Empower\BuslaceDrivers.

7. Select the checkbox “Include subfolders” and then click Next.

8. When the Windows Security message “Would you like to install this device software?” appears, click Install.

Result: The BusLAC/E Driver installs. It appears as “Buslace PCI KMDF Device Driver” under BusLAC/E Drivers in Device Manager.

Removing Empower 2 Feature Release 5

To remove Empower 2 Feature Release 5, you must uninstall Empower. Before you uninstall Empower, Waters recommends that you first complete these tasks in the order listed:

Back up all Empower 2 projects, library information, and/or databases.

Important: If you do not backup projects, library information, and databases, this information will be lost during the uninstall process.

Remove Waters Instrument Control Software from the Windows Control Panel > Add/Remove Programs.

Uninstall all installed options and licenses.

To remove Empower 2 Feature Release 5 from your system: 1. Reboot your computer.

2. Log onto the computer with a user account that has the local Administrator privilege.

3. From the Windows Control Panel, click the selection based on your Windows operating system:

On Windows XP, click Add/Remove Programs

On Windows Vista, click Programs and Features

4. From the program list, select Empower, and then click Remove.

5. Close the Control Panel.

6. In the File System Windows directory, manually delete the milerror.log file, which also prevents it from appearing in Verify files and the error log.

Page 24: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 24 of 32

Requirement: Whenever you install or uninstall Empower software or any software option, you must re-qualify your Empower software.

Feature Release 5 program file updates

In Feature Release 5, some program files have changed, while other program files have not changed, but have a new date only.

The first table lists the program files that have changed. The second table lists the program files that have not changed, but have a new file date.

Program files that contain changes for Feature Release 5:

File Name Purpose

AutoArchive.exe The AutoArchive application.

AutoLogon.exe The Automatic Log On application.

cmgr.exe The Configuration Manager application.

Compare.exe The Compare application.

dconvert.exe The application that restores pre-Millennium 3.0 data.

Empower.exe The main Empower application.

ImmdSrvr.exe Contains immediate background processing reporting and exporting support.

InEditor.exe The Instrument Editor application.

InstrumentServer.exe The Instrument Service application.

ManTune.exe The Manual Tune application.

msdiag.exe The TMD Diagnostics application.

mseditor.exe The Method Set Editor application.

msgcenter.exe The Message Center application.

Open Access.exe The Open Access application.

OQ.exe The Operations Qualification application.

Pdacalib.exe The PDA Calibration application.

Pdadiag.exe The PDA Diagnostic application.

pmeditor.exe The Processing Method Editor application.

procmon.exe The Processing Monitor application.

ProcSrvr.exe The Processing Server application.

project.exe The Project application.

quickset.exe The Run Samples application.

QuickStart.exe The QuickStart application.

Page 25: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 25 of 32

Program files that contain changes for Feature Release 5:

File Name Purpose

Report Publisher.exe The Report Publisher application.

restoredb.exe The Restore Database application.

Review.exe The Review application.

RptViewer.exe The Report Viewer application.

setarchdatabase.exe The Set AutoArchive database application.

SetEmailDatabase.exe The Set E-Mail database application.

Sseditor.exe The Sample Set Editor application.

StoreSrv.exe The Storage Server application.

SysMon.exe The System Monitor application.

Update2487.exe Firmware updates for the 2487 Detector.

ValidationManager.exe The Validation Manager application.

ValidationReview.exe The Validation Review application.

VerifyOpt.exe The Verify Options application.

WatersEMailService.exe The Waters E-Mail Service application.

WatersService.exe The Waters network communication network.

WDC.exe The Waters Data Converter application.

andims.dll Contains ANDI mass spectrometry library objects.

Cmgrlib.dll Contains the Configuration Manager library.

EthernetIeee.dll Contains all the control objects necessary to connect and use the National Instruments Ethernet to IEEE converter box.

impexp.dll Import and export objects.

IPalisade.dll Contains the Palisade mass spectrometry library.

mil0.dll Contains base objects which implement database connection, security, user and user type objects, etc.

mil1.dll Contains base objects which implement higher level objects such as methods, results, etc.

mil2.dll Contains base objects which implement higher level objects including data processing, integration, calibration, etc.

mil3.dll Contains objects which implement higher level objects such as reporting groups and display graphics.

milado.dll Contains objects which implement higher level objects such as instrument method editor display graphics.

MillServices.dll Contains definitions for the Millennium user interface.

Page 26: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 26 of 32

Program files that contain changes for Feature Release 5:

File Name Purpose

milp.dll Contains base objects which implement higher level objects including data processing, integration, calibration, etc.

milres.dll Contains base objects which implement test and graphic based common resources.

MilTK.dll Contains basic Toolkit objects.

MilTkEx1.dll Contains extended Toolkit objects.

MtkCtrl.dll Contains Toolkit control objects.

Netcdf.dll Contains low level processor format objects.

OptInt.dll Contains licensing and option objects.

StoreSrv.dll Contains Storage Server classes.

StoreSrvProxy.dll Contains Storage Server proxy stubs.

WDC_AC.dll Waters Data Converter module handling AccessChrom data format

WDC_CS.dll Waters Data Converter module handling ChemStation data format

WDC_MassLynx.dll Waters Data Converter module handling some MassLynx data format

New program files:

File Name Purpose

Interop.OPTINTLib.dll Contains Interoperability objects.

LicensingApp.config The specification plug-in for the Licensing Wizard.

Waters.Licensing.Wizard.exe The licensing application.

Empower Licensing Plugin.dll Used for the Waters Licensing Wizard.

Waters.UI.WatStyle.dll The UI style objects.

Waters.UI.WatTreeView.dll The tree view style objects.

Waters.UI.WatTreeView.xml The tree view style xml.

Waters.UI.WatValidator.dll The UI validation objects.

Waters.UI.WatValidator.xml The UI validation xml.

NG71SECF.dll Used for NuGenesis connectivity.

NG71 Toolkit Login Key.exe Used for NuGenesis connectivity.

Page 27: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 27 of 32

Program files unchanged – new date only:

File Name Purpose

InstrumentServerps.dll Proxy stub for Instrument Server executable.

memlib32.dll Contains library matching support objects.

pbmlib32.dll Contains mass spectrometry and PDA library matching objects.

pltlib32.dll Contains mass spectrometry and PDA library matching objects.

reflib32.dll Contains mass spectrometry and PDA library matching objects.

ulblib32.dll Contains mass spectrometry and PDA library matching objects.

WatersServiceps.dll Contains the Waters Service proxy stubs.

wlylib32.dll Contains Wiley mass spectrometry library objects.

Quickset.tlb The Quickset application type library.

Mtkctrl.tlb The Tool Kit event control type library.

MilTk.tlb The Tool Kit type library.

MilServices.tlb The Millenium environnent information type Library.

MilTkEx1.tlb The Tool Kit extension type library.

COMMethodCompare.xsl Contains COM object (XML) objects.

Page 28: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 28 of 32

Test configurations used to develop and evaluate Feature Release 5

The following test configurations were used during the development and evaluation of Empower Feature Release 5.

Empower 2 Feature Release 5 was tested on servers containing a single database instance only.

Personal workstation test configuration

Operating system and service pack: Windows XP, SP2

Microsoft Windows hotfixes: KB928366, .NET 2.0 SP1, KB927978, KB936181, KB933579, KB892130, KB893803, KB933566, KB938127, KB942615, KB947864, KB923789, KB890046, KB890859, KB891781, KB893756, KB894391, KB896344, KB896358, KB896422, KB896423, KB896424, KB896428, KB898461, KB899587, KB899589, KB899591, KB900485, KB900930, KB901017, KB901214, KB902400, KB904706, KB904942, KB905414, KB905749, KB908519, KB908531, KB910437, KB911280, KB911562, KB911567, KB911927, KB912919, KB913580, KB873339, KB914389, KB885835, KB888113, KB885836, KB886185, KB917953, KB887472, KB918439, KB887742, KB892130, KB888302, KB916595, KB920670, KB920683, KB917422, KB920214, KB921398, KB914388, KB921883, KB922616, KB917159, KB920872, KB920685, KB919007, KB922582, KB923191, KB923980, KB924191, KB924270, KB922819, KB923414, KB920213, KB920342, KB925876, KB925398, KB918118, KB926255, KB926436, KB927779, KB927802, KB923694, KB928255, KB928255, KB928843, KB923689, KB931866, KB924667, KB929338, KB931261, KB925720, KB930916, KB929399, KB932168, KB914440, KB935839, KB935840, KB929123, KB927891, KB926239, KB830178, Windows Imaging Component, KB931784, KB925902, KB885884, KB936357, KB921503, KB941568, KB942763, KB936021, KB943460, KB933729, KB941569, KB944653, KB941202, KB936782, KB938829, KB937894, KB939683, KB905474, KB938828, KB946026, KB945553, KB943485, KB943055, KB941693, KB941644, KB948590, KB948881

Operating system and service pack: Windows Vista Business SP1

KB952287, KB941693, KB948590, KB950582, KB950762, KB950974, KB951376, KB953733, KB953838, KB853839, KB947562, KB951072, KB951618, KB952709, KB955020, Windows Search 4.0, KB905866, KB938371, KB951978, KB951066, KB951698, KB936330, KB946041, KB932471, KB929735, KB931099, KB931573, KB933928, KB936824, KB936825, KB937077, KB938194, KB938979, KB941649, KB941651, KB943302, KB943899, KB937287, KB925902, KB929123, KB929916, KB930178, KB931213, KB933579, KB933729, KB935807, KB936021, KB936782, KB938123, KB938127, KB941202, KB941568, KB941569, KB942624, KB943055, KB943078, KB943411, KB944533, KB946026, KB946456, KB929399, KB930857, KB936357, KB939159, KB942763

Operating system and service pack: Windows XP SP3

KB938127, KB941569, KB944533, KB953838, Q282784, KB811113, KB936929, KB938464, KB946648, KB950762, KB950974, KB951066, KB951072-v2, KB951376-v2, KB951698, KB951748, KB951978, KB952287, KB952954, KB953839

Page 29: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 29 of 32

Windows Database Server 2003

Operating system: Windows Enterprise 2003 (English), R2 SP2

Windows hotfixes: KB925673, KB947864, KB944533, KB938127, KB950759, KB926122, KB927891, KB929123, KB930178, KB931784, KB932168, KB933729, KB933854, KB935839, KB935840, KB936021, KB925902, KB936782, KB941202, KB941568, KB941569, KB941644, KB941693, KB942763, KB943055, KB943460, KB943485, KB944338, KB944653, KB945553, KB946026, KB948496, KB948590, KB948881, KB924667-v2, KB925398, KB936357, Windows Imaging Component, KB950762, KB951748, KB950760, KB951698

Oracle version: 10.2.0.3

Windows 2000 Database Server

Operating system: Windows 2000 Enterprise (English), SP4

Windows hotfixes: KB927779, KB911564, KB925398, Update Rollup 1, KB842773, KB893756, KB896358, KB896422, KB896423, KB899587, KB899589, KB899591, KB900725, KB901017, KB901214, KB905414, KB905495, KB905749, KB908519, KB908531, KB911280, KB913580, KB914388, KB914389, KB917008, KB917537, KB918118, KB920213, KB920670, KB920683, KB920685, KB921398, KB923191, KB923414, KB923810, KB923980, KB924270, KB924667, KB925902, KB926122, KB926247, KB926436, KB927891, KB928843, KB930178, KB931784, KB933729, KB935839, KB935840, KB936021, KB937894, KB938127, KB938827, KB941202, KB941693, KB942831, KB943055, KB943485, KB944338, KB945553, KB948590, KB950749, KB950759, KB950760, KB951698, KB951748, KB893803, Q828026

Oracle version: 10.2.0.3

Raw Data File Server operating system and hotfixes

Operating system: Windows Server 2003 R2 SP2

Microsoft Windows hotfixes: KB911564, KB929969, KB938127, KB942615, KB947864, KB620759, KB925398, KB926122, KB927891, KB929123, KB930178, KB931784, KB932168, KB933729, KB933854, KB925902, KB935840, KB936021, KB936357, KB936782, KB941202, KB941568, KB941569, KB921503, KB935839, KB942763, KB924667-v2, 943460, KB944653, KB943485, KB945553, KB946026, KB948496, KB948590, KB948881, KB941693, KB941644, KB943055, KB950760, KB950762, KB951698

Chinese Database Server operating system and hotfixes

Operating System: Windows Enterprise 2003 (Chinese), R2 SP2

Microsoft Windows hotfixes: KB905759, KB944533, KB947864, KB938127, KB925902, KB926122, KB927891, KB929123, KB930178, KB931784, KB932168, KB933729, KB933854, KB935839, KB935840, KB936021, KB925398, KB936782, KB941202, KB941569, KB941644, KB941693, KB942763, KB924667-V2, KB951698, KB943055, KB943460,KB943485, KB944338, KB944653, KB945553, KB946026, KB948496, KB948590, KB950759, KB950760, KB950762, KB936357, KB942830, KB942831

Oracle version: 10.2.0.3

Page 30: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 30 of 32

Korean Database Server operating system and hotfixes

Operating System: Windows Enterprise 2003 (Korean), R2 SP2

Microsoft Windows hotfixes: KB950759, KB938127, KB925398, KB925902, KB926122, KB927891, KB929123, KB930178, KB931784, KB932168, KB933729, KB933854, KB935839, KB935840, KB936021, KB936357, KB936782, KB924667-V2, KB941568, KB941569, KB941644, 921503, KB942615, KB942763, KB944653, KB943485, KB942840,KB943460, KB941202, KB921503, KB942831, KB942830, KB945553, KB946026, KB948496, KB948590, KB950760, KB950762, KB951698, KB951748, KB941693, KB943055

Oracle version: 10.2.0.3

Japanese Database Server operating system and hotfixes

Operating System: Windows Enterprise 2003 (Japanese), R2 SP2

Microsoft Windows hotfixes: KB924667-V2, KB925398, KB925902, KB926122, KB927891, KB929123, KB930178, KB931784, KB932168, KB933729, KB921503, KB935839, KB935840, KB936021, KB936357, KB936782, KB938127, KB941202, KB941568, KB941569, KB941644, KB942615, KB942763, KB942840, KB943460, KB943485, KB944653, KB933854

Oracle version: 10.2.0.3

UNIX Database Server operating system and hotfixes

Operating system: Solaris 10

Sun Solaris patch set: Solaris 10 (last OS update: 10_Recommended Patch Set from May 2008)

Oracle version : 10.2.0.3 with the April 2008 Oracle CPU patch

Citrix Servers

Operating system: Windows Server 2003 (English), R2 SP2

Citrix version: Presentation Server 4.5 (Enterprise Edition)

Citrix Client version: 10.150

Citrix hotfixes: Citrix Hot fix Roll up Pack PSE450W2K3R02

Operating system: Windows Server 2003 R2 SP2 (English)

Microsoft Windows hotfixes: KB925673, KB938127, KB939653, KB942615, KB927891, KB924667-v2, KB930178, KB931784, KB931836, KB932168, KB925902, KB933566, KB929123, KB935840, KB935839, KB933854, KB936021, KB936357, KB936782, KB933729, KB921503, KB926122, KB925398, KB933360, KB941202, KB943485, KB944653, KB941568, KB942763, KB941644, KB941569, KB943460, Windows Imaging Component

Page 31: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 31 of 32

Citrix Web Interface Server

Operating system and service pack: Windows Server 2003 R2 SP2 (Enterprise Edition)

Internet Explorer version: 7.0 (20061027.15806)

Microsoft Windows hotfixes: KB931768, KB933566, KB938127, KB939653, KB942615, KB947864, KB932168, KB931784, KB930178, KB925902, KB931836, KB929123, KB927891, KB935840, KB935839, KB924667-v2, 933360, KB936021, KB936357, KB936782, KB941202, KB921503, KB933854, KB933729, KB926122, KB925398, KB942763, KB941568, KB941569, KB941644, KB943460, KB943485, KB944653, KB945553, KB943055, KB948496, KB948590, KB948881, KB941693, KB942831, KB942830, KB946026

Client and LAC/E32 test configurations

Client test configuration

Operating system and service pack: Windows XP, SP2

Microsoft Windows hotfixes: KB928366, .NET 2.0 SP1, KB927978, KB936181, KB933579, KB892130, KB893803, KB933566, KB938127, KB942615, KB947864, KB923789, KB890046, KB890859, KB891781, KB893756, KB894391, KB896344, KB896358, KB896422, KB896423, KB896424, KB896428, KB898461, KB899587, KB899589, KB899591, KB900485, KB900930, KB901017, KB901214, KB902400, KB904706, KB904942, KB905414, KB905749, KB908519, KB908531, KB910437, KB911280, KB911562, KB911567, KB911927, KB912919, KB913580, KB873339, KB914389, KB885835, KB888113, KB885836, KB886185, KB917953, KB887472, KB918439, KB887742, KB892130, KB888302, KB916595, KB920670, KB920683, KB917422, KB920214, KB921398, KB914388, KB921883, KB922616, KB917159, KB920872, KB920685, KB919007, KB922582, KB923191, KB923980, KB924191, KB924270, KB922819, KB923414, KB920213, KB920342, KB925876, KB925398, KB918118, KB926255, KB926436, KB927779, KB927802, KB923694, KB928255, KB928255, KB928843, KB923689, KB931866, KB924667, KB929338, KB931261, KB925720, KB930916, KB929399, KB932168, KB914440, KB935839, KB935840, KB929123, KB927891, KB926239, KB830178, Windows Imaging Component, KB931784, KB925902, KB885884, KB936357, KB921503, KB941568, KB942763, KB936021, KB943460, KB933729, KB941569, KB944653, KB941202, KB936782, KB938829, KB937894, KB939683, KB905474, KB938828, KB946026, KB945553, KB943485, KB943055, KB941693, KB941644, KB948590, KB948881

Operating system and service pack: Windows XP, SP3

Microsoft Windows hotfixes: KB928366, KB927978, KB936181, KB933579, KB892130, KB933566, KB938127, KB942615, KB947864, KB953838, KB923789, Windows Imaging Component, KB929399, KB939683, KB941569, KB936782, KB905474, KB953839, KB951376-v2, KB951698, KB952287, KB951066, KB950974, KB950762, KB946648, KB951978, KB951072-v2, KB951748, KB952954

Operating system and service pack: Windows Vista Enterprise, SP1

Microsoft Windows hotfixes: KB952287, KB950582, KB950762, KB950974, KB951376, KB953733, KB953838, KB905866, KB951072, KB951618, KB952709, KB955020, Windows Search 4.0, KB853839, KB947562, KB951978, KB951066, KB951698, KB936330, KB937954, KB941693, KB947864, KB948590, KB948881, KB938371, KB945553, KB946041, KB937287, KB929735, KB932471, KB925902, KB929916, KB930178, KB931213, KB933579, KB933729, KB935807, KB936021, KB936782, KB938123, KB938127, KB941202, KB941568, KB941569, KB942624, KB943055, KB943078, KB943411, KB944533, KB946026, KB946456, KB929399, KB930857, KB931099, KB931573, KB933928, KB935509, KB936357, KB936824, KB936825, KB937077, KB938194, KB938952, KB938979, KB939159, KB941649, KB941651, KB942763, KB943302, KB943899

Page 32: Waters® Empower 2 Feature Release 5„¢ 2 Feature Release 5 contains software changes, ... from earlier versions of the software prior to Empower 2 Feature Release 3 ... Empower

716002505, Rev. A

Page 32 of 32

LAC/E32 test configuration

Operating system and service pack: Windows XP SP2

Microsoft Windows hotfixes: KB888111, KB898458, KB923723, KB927978, KB936181, KB927977, KB892130, KB893803, KB947864, KB944533, KB885835, KB885836, KB886185, KB887472, KB887742, KB888113, KB888302, KB890047, KB890175, KB884020, KB891781, KB923789, KB873339, KB885250, KB873333, KB896422, KB896423, KB896424, KB896428, KB898461, KB899587, KB899588, KB899589, KB899591, KB900485, KB901017, KB901214, KB902400, KB904706, KB896358, KB905414, KB905749, KB908519, KB908531, KB910437, KB911280, KB911562, KB911564, KB911567, KB911927, KB912919, KB894391, KB893756, KB914388, KB914389, KB893086, KB916595, KB917159, KB917422, KB917537, KB917734, KB917953, KB918118, KB918439, KB919007, KB920213, KB920214, KB920670, KB920683, KB920685, KB920685, KB920872, KB921398, KB921883, KB922582, KB922616, KB922819, KB923191, KB923414, KB923689, KB923694, KB893066, KB923980, KB924191, KB924270, KB924667, KB925398, KB890859, KB926255, KB926436, KB927779, KB927802, KB890046, KB928255, KB928843, KB913446, KB929338, KB931836, KB913580, KB892130, KB905474, KB930178, KB932168, KB933729, KB935839, KB935840, KB935840, KB936021, KB936357, KB936782, KB937894, KB938828, KB941202, KB941568, KB931784, KB941644, KB941693, KB942763, KB943055, KB943460, KB943485, KB944338, KB944653, KB945553, KB946026, KB948590, KB948881, KB950749, KB931261, KB941569, Windows Imaging Component, KB929123, KB927891, KB925902, KB904942, KB914440, KB930916, Q282784

Operating system and service pack: Windows XP SP3

Microsoft Windows hotfixes: Internet Explorer 7, KB923689, KB941569, KB953838-IE7, Q282784, KB888111WXPSP2, KB936929, KB946648, KB950762, KB950974, KB951066, KB951072-v2, KB951376-v2, KB951698, KB951748, KB951978, KB952287, KB952954, KB953839