Page 1: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

Vista™ basic

950787 Rev 6/02

Instructions for Use

Rx Only

Page 2: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

WARRANTY

Vista™ basic Instructions for Use 1/25950787 Rev 6/02

WARRRANTYB. Braun provides 24 monthswarranty as of the date of deliveryfor every Vista™ basic InfusionPump. This includes the repair orreplacement of damaged parts asa result of design/manufacturingerrors or material defects.Modifications or repairs to the unitby the owner or third partiesinvalidate the warranty.

The warranty does not cover theelimination of faults attributable toincorrect/inexpert handling,normal wear and tear, orrechargeable batteries.

The CE mark confirms compliancewith the “Council Directive onMedical Products 93/42/EEC”dated the 14th of June, 1993.

B. Braun Melsungen AG

RESPONSIBILITY OFMANUFACTURERManufacturer, assembly, andinstallation personnel or instructorscan only be held responsible for anyeffects on device safety, reliability,and performance if♦ Installation, expansion work,

readjustments, modifications, orrepairs are carried out bypersonnel authorized by theabove,

♦ The electrical wiring for AC powersatisfies the requirements of VDE0100, 0107, and/or the IECpublications, or regional and/ornational variations,

♦ The device is operated in line withthe Instructions for Use, and

♦ Regular technical inspections(Preventative Maintenance) arecarried out every 2 years asoutlined in the Service Manual.

INSPECTION ON DELIVERYDespite careful packaging, the riskof transport damage cannot beentirely excluded. Upon delivery,please check that nothing ismissing. Do not use a damageddevice! If the device is damaged,contact the Service Department.See “Maintenance, Storage, &Service.”

Packaging: The pump packaging isreusable.

Package Contents: Vista basicInfusion Pump, drop sensor, powercable, pole clamp, Instructions forUse.

Page 3: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

Vista™ basic Contents

Vista™ basic Instructions for Use 2/25950787 Rev 6/02

CONTENTSWarranty..................................................................................... 1Warnings & Cautions.................................................................. 3Introduction ................................................................................ 5Preparing the Infusion ................................................................ 7Continuous Mode ....................................................................... 8Options: Piggyback.......................................................................... 11

Ramp/Taper Mode ............................................................ 12Program Mode ............................................................ 14Delay A Mode ................................................................... 15Dosage Calculation........................................................... 16Bolus Mode, Interval Bolus (Intermittent Mode) ................ 17Total Infused, Standby................................................. 18Occlusion Pressure, Drop Sensor, Battery ....................... 19Data Lock, Contrast Selection .......................................... 20Adjust Clock, Loudness .................................................... 21

Alarms ........................................................................................ 22AC, 12V, Battery, & CC Interface ............................................... 23Maintenance, Storage, & Service............................................... 24Specifications ............................................................................. 25

The Vista basic is a volumetric infusionpump intended to provide accurate andcontinuous flow of parenteral andenteral fluids.

Qualified medical staff should decidehow the device should be used basedon its features and specifications.

For more details, please read theInstructions for Use.

Rx only.

Page 4: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

WARNINGS & CAUTIONS

Vista™ basic Instructions for Use 3/25950787 Rev 6/02

The following should be observed to avoid patient/user injury. This Instructionsfor Use manual contains detailed instructions and warnings on the use of theVista™ basic Infusion Pump. Please read it completely prior to using this device.This manual is intended to reinforce the teaching given to the user by a trainedhealth care professional or an authorized B. Braun representative.

WARNINGS♦ B. Braun will assume no

responsibility for incidents whichmay occur if the product is not usedin accordance with product labeling.

♦ Make sure the pump is stable, eitherfastened securely to the IV pole orresting on a flat surface.

♦ Prior to use, always check theaudible and visual alarms (and staffcall, where applicable).

♦ Connection of pump tubing to thepatient should only be made afterthe tubing is loaded and the unit isturned on. Disconnect tubing frompatient when changing solutioncontainers to avoid the danger ofincorrect dosage.

♦ Ensure there are no kinks in thetubing.

♦ Always read and follow theinstructions that accompany thesource container and IVadministration sets you are using.Carefully follow the instructions forloading, removing, and reloading theset, as well as the recommended setchange intervals.

♦ Change pumping section on tubingevery 24 hours. Do not use tubingfor more than 72 hours or percurrent facility protocol.

♦ Always check pump data prior tostarting infusion.

♦ Do not operate this device inenvironments where there is anexplosion hazard! A possibleexplosion hazard exists if the pumpis used in the presence offlammable anesthetics.

♦ The EMC-limits (Electro-magneticcompatibility) according to IEC/EN60601-1-2 and IEC/EN 60601-2-24are maintained. If the equipment isoperated near other equipmentwhich may cause high levels ofinterference (e.g. HF surgicalequipment, nuclear spin tomographyunits, mobile telephones, etc.)maintain the recommendedprotective distances for thesedevices. Under certain conditions,malfunctions may occur which leadto a device alarm with permanentalarm tone. Interference may occure.g. at electro-magnetic fields > 10[V/m] resp. electro-magneticdischarges > [8 kV].

♦ Monitoring of the patient andinfusion status is necessary toensure the infusion is beingdelivered as anticipated. The Vistabasic has been designed to stopfluid flow to the patient under certainalarm conditions. Also, prior tostarting an infusion, verify that nodrops are falling in the drip chamberand the programmed information iscorrect.

Page 5: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

WARNINGS & CAUTIONS

Vista™ basic Instructions for Use 4/25950787 Rev 6/02

♦ After starting an infusion, make suredrops are falling in the drip chamber.If no drops are falling, make sure theroller clamp is open. If the rollerclamp is open and still no drops arefalling, replace and dispose of theset.

♦ The air detector cannot recognizethe introduction of air at 3-waystopcocks, infusion ports, and otherlines/tubes below the pump.

♦ See section entitled “Drop Sensor”for additional information on usingthe device without a drop sensor.

♦ To avoid mechanical or electronicdamage, do not steam autoclave orimmerse the pump in any fluids orcleaning solutions. Alwaysdisconnect electrical power cordfrom outlet before cleaning toprevent electrical shock.

CAUTIONS♦ Rx only. ♦ Do not attempt to infuse two fluids

simultaneously using the Vista™

basic.♦ Variations in pressure, e.g. caused

by change in level of fluid container,may affect the accuracy of thedevice.

♦ If several pieces of infusionequipment are connected together,it is possible there may be mutualinfluence. Possible incompatibilitiescan be found in the instructions foruse for the drug or for the otherappliances.

♦ Use only equipment, accessories,parts, and disposables that areintended for use with the Vistabasic.

♦ Do not clean, disinfect, or sterilizeany part of the device by autoclavingor with ethylene oxide gas. Doingso may damage the device and voidthe warranty. Only external parts ofthe pump should be disinfected.See “Maintenance, Storage, &Service” for suggested cleaningsolutions.

♦ Technical specifications are notguaranteed with any other tubingthan that approved for use with theVista basic.

♦ Any analog and digital componentsconnected must demonstrate thatthey meet EN specifications (i.e. EN60950 for data processingequipment and EN 60601 forelectrical medical equipment).

♦ Only an authorized systemconfiguration technician responsiblefor meeting the specifications ofstandard EN 60601-1-1 mayconnect additional devices to thesignal inputs or outputs.

SAFETY STANDARDThe Vista basic meets all safetystandards for medical electrical devices,corresponding to IEC 60601-1 and IEC60601-2-4. NOTE: IEC 60601-1 corresponds toEuropean Standard EN 60601.

Page 6: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

INTRODUCTION

Vista™ basic Instructions for Use 5/25950787 Rev 6/02

Handle for carrying

DisplayAll essential information isdisplayed in plain language.Green background lightingilluminates only whenconnected to AC power, or, inbattery operation, only when akey is pressed.

KeyboardAllows data entry

Clear KeyResets avalue tozero

Alarm LightLights up during an alarm.Display will also showflashing AAA.A.

Infusing LightIndicates pump is infusing

Page 7: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

INTRODUCTION

Vista™ basic Instructions for Use 6/25950787 Rev 6/02

Door Open Button

Aluminum CaseEasy upkeep, drip proof, andresistant to disinfectants

Function KeysAllow choices of YES, NO, OK,INFO, OPTS, etc.

ON/OFF (Power) KeyAlarm Silence KeySilences alarm tone for 2minutes

START/STOP Key

Page 8: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

PREPARING THE INFUSION

Vista™ basic Instructions for Use 7/25950787 Rev 6/02

Prepare the Pump1. Check for cleanliness, completeness, and damage. If unit appears

damaged, do not use it.2. Attach the pump to an IV pole by turning the knob on the pole clamp

clockwise, or place the pump on a flat, stable surface.3. Plug the pump’s power cord into an electrical outlet.

Turn on the Pump1. Press (ON/OFF).2. Listen for 3 loud beeps.3. Check screen for completeness of all pixels on the display. The Alarm ( ) and Infusing ( ) lights will blink once.

Load the Set1. Prepare the solution container and prime the administration set according to

the instructions provided on the set package. Close the roller clamp.2. Check that the drip chamber is about 1/3 full

and there is no air in the set.3. Open the pump door by pressing the Door

Open Button on the top right corner of thepump.

4. Push and hold open the clamp at the top ofthe fluid path.

5. Insert tubing in clamp and release clamp. 6. Thread the tubing down the fluid path, making

sure the tubing is pressed into the Air-in-Linedetector at the bottom of the fluid path. Theroller clamp should be below the pump.

7. Close the door and open the roller clamp.Check that no drops are falling in the dripchamber.

8. Place drop sensor on drip chamber (optional).9. Connect set to patient site.

Turning off the Pump1. Press to stop the infusion.2. Close the roller clamp.3. Press and hold for 2 seconds.

Page 9: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

Continuous Mode

Vista™ basic Instructions for Use 8/25950787 Rev 6/02

When the pump is turned on, it may display: Tubing Type VISTA Pump SetConfirm you are using a Vista™ Pump Set, then press YES.

The pump may then display: Is This a New Therapy?YES

Begins in Continuous Mode with alldata cleared.

NOReturns to the beginning of the lastprogram with all data retained.

The pump may then display: Prime Set?

YESPress YES to begin priming set.• Press STOP to end priming. • Press EXIT to begin Continuous

Mode.

NOBegins in Continuous Mode with alldata cleared.Press the green key below youranswer.

YESPress YES to begin priming set.• Press STOP to end priming.• Press EXIT to return to the

beginning of the last program withall data retained.

NOReturns to the beginning of the lastprogram with all data retained.

CONTINUOUS MODEThis section explains how to program a Continuous Mode infusion. Fordirections on how to program other Modes and Options, refer to those sectionslater in this manual.

There are three ways to enter infusion data in Continuous Mode:

RATE and VTBD (Volume To Be Delivered)1. Enter rate in ml/hr.2. Press VTBD (Volume To Be Delivered).3. Enter bag volume in ml.4. Press OK. (To leave screen without saving changes, press EXIT.)5. Press to begin the infusion. Check for drops in the drip chamber.

Page 10: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

Continuous Mode

Vista™ basic Instructions for Use 9/25950787 Rev 6/02

VTBD and TIME1. Press VTBD.2. Enter bag volume in ml.3. Press OK. (To leave screen without saving changes, press EXIT.)4. Press TIME.5. Enter infusion time in hours and minutes.

NOTE: For a 2 hour infusion, enter 2, 0, 0. The pump will show: Time: 02h00m6. Press OK. (To leave screen without saving changes, press EXIT.) The

pump calculates the rate and asks: Accept Rate? 7. Press OK.8. Press to begin the infusion. Check for drops in the drip chamber.

RATE and TIME 1. Enter rate in ml/hr.2. Press TIME.3. Enter infusion time in hours and minutes. 4. Press OK. (To leave screen without saving changes, press EXIT.)5. Press to begin the infusion. Check for drops in the drip chamber.

To Change the Rate1. Press to stop the infusion.2. Press . The display will reset to 000.0 ml/h.3. Enter a new rate.4. Press to begin the infusion. Check for drops in the drip chamber.

To Change the Rate While Infusing1. Enter a new rate.2. Press OK.

NOTE: If the OK key is not pressed within 10 seconds of the change, the infusion willcontinue at the old rate.

Page 11: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

Continuous Mode

Vista™ basic Instructions for Use 10/25950787 Rev 6/02

KEEP VEIN OPEN RATEAfter an infusion is complete, the pump will alarm and begin pumping at 1.0 ml/hror the current infusion rate, whichever is less.

1. Stop the infusion.2. Turn off the pump or reset the program.

INFORMATION KEYThis key is only available while the pump is infusing.

1. Press INFO.2. The Total Infused value and the Total Time of the infusion completed are

displayed.3. Press INFO again and the Battery Left and Total Operation time of the pump

is displayed.4. The pump will return to the main infusion screen after 10 seconds.

Page 12: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

OPTIONS Piggyback

Vista™ basic Instructions for Use 11/25950787 Rev 6/02

The Vista™ basic contains many options to allow custom programming. Toaccess the various options, press OPTS while the pump is stopped. PressingOPTS each time will move you to the next option. Eventually, the OPTS key willbring you back to the mode from which you began.

Depending on the model, some modes/options may have been disabled and willnot appear on the display when the OPTS key is pressed.

PIGGYBACKThe Piggyback feature allows programming of a secondary (Piggyback) infusion.The pump automatically switches from the Piggyback to the Continuous Modeinfusion when the Piggyback program is complete. The option is only available ifthere is data in the Continuous Mode.

1. Connect the Piggyback set to the injection site on thetubing above the pump on the primary set. Open theroller clamp on the Piggyback set.

2. Hang the Piggyback solution container at least 8 inchesabove the primary solution container.

3. Program the continuous infusion in the Continuous Mode(see “Continuous Mode”).

4. Press OPTS.5. Press ON to turn the Piggyback feature on.6. Press END. PIGY will now be displayed in the lower left corner.7. Press PIGY.8. Enter the Piggyback infusion data as you would Continuous Mode data.9. Press to start the Piggyback infusion. 10. Check that drops are falling in the Piggyback drip chamber.

CAUTION: Do not attempt to infuse both Continuous and Piggyback fluids at the same time. Useprimary sets with check valves.

NOTE: To prevent incorrect delivery, the programmed Piggyback volume to be delivered must equalthe amount of fluid in the Piggyback container. Automatic Piggyback is not intended for the infusionof fluids requiring flushing before and after administration.

NOTE: If the Piggyback container is not empty, the remaining Piggyback fluid will be delivered at theContinuous Mode rate once the pump transitions back to the Continuous Mode.

Page 13: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

OPTIONS Piggyback, Ramp/Taper Mode

Vista™ basic Instructions for Use 12/25950787 Rev 6/02

To Exit Piggyback

1. Press to stop the infusion.2. Press END.3. Press OPTS.4. Press OFF to turn Piggyback off5. Press END to return to Continuous Mode.

RAMP/TAPER MODEThe Ramp/Taper Mode is designed to deliver infusions, such as Total ParenteralNutrition (TPN), which require gradual ramp up and taper down rates. The pumpautomatically calculates the rate changes required to match the volume, totaltime, and ramp/taper time parameters.

1. Press OPTS until display shows RAMP TAPER MODE.2. Press ON.3. Enter the total bag VOLUME to be delivered.4. Press OK.5. Enter the RAMP UP time, in hours and minutes.

NOTE: The pump defaults to 01h00m. To change, press , and enter the time.6. Press OK.7. Enter the TAPER DOWN time, in hours and minutes.

NOTE: The pump defaults to 01h00m. To change, press , and enter the time.8. Press OK.9. Enter the TOTAL Time for the infusion. 10. Press OK

Note: The pump calculates the maximum rate and asks: Accept Rate?11. Press OK.12. Press to start the infusion. Check for drops in the drip chamber.

To Review or Change Ramp/Taper Mode Data1. Press OPTS until display shows RAMP TAPER MODE.2. Press ON.3. Review the data on each screen and press OK if data is correct or press

to clear the current data and re-enter new data. Press OK to acceptnew data.

Page 14: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

OPTIONS Ramp/Taper Mode

Vista™ basic Instructions for Use 13/25950787 Rev 6/02

To Clear Ramp/Taper Mode1. Press OPTS until display shows RAMP TAPER MODE.2. Press CLR.3. Press YES to clear all the data in Ramp/Taper Mode.4. Press ON and continue to program a new Ramp/Taper program.

To Exit Ramp/Taper Mode1. When infusion is complete, press to stop the KVO.2. Press EXIT.3. Press END to return to Continuous Mode.

Fast Taper The infusion can be interrupted to begin an early taper down process during theRamp/Taper Mode Infusion.

1. Stop the infusion.2. Press FAST.3. Enter the FAST Taper Time in hours and minutes. 4. Press OK.5. Press to start the infusion. Check for drops in the drip chamber.

Page 15: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

OPTIONS Program Mode

Vista™ basic Instructions for Use 14/25950787 Rev 6/02

PROGRAM MODEProgram Mode is for infusions requiring up to 9 separate rates and volumes to bedelivered one after another. This is designed to deliver infusions for steppedoncology drugs, circadian medication delivery, and other custom programs.

1. Press OPTS until display shows PROGRAM MODE.2. Press ON.3. Enter the rate of the first Step.4. Press OK.5. Enter the volume of the first Step.6. Press OK.7. Continue to enter the rates and volumes for Steps 2-9. 8. If less than 9 Steps are needed, press OK with a rate of zero.9. Press OK to accept the total Program Volume.

NOTE: Pressing CHNG on this screen returns to Step 1 for review and changes.10. Press to start the infusion. Check for drops in the drip chamber.

To Review or Change Program Mode data1. Press OPTS until display shows PROGRAM MODE.

Note: If Program Mode is infusing, stop the infusion and press EXIT.2. Press ON.3. Review the data on each screen and press OK if data is correct or press

to clear the current data and re-enter new data. Press OK to acceptnew data.

4. OR, press CHNG on Accept Program Volume screen (see #9 NOTE above).

To Clear Program Mode1. Press OPTS until display shows PROGRAM MODE.

Note: If Program Mode is infusing, stop the infusion and press EXIT.2. Press CLR.3. Press YES to clear all the data in Program Mode.4. Press ON and continue to program a new Program.

To Exit Program Mode1. When infusion is complete, press to stop the KVO.2. Press EXIT.3. Press END to return to Continuous Mode.

Page 16: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

OPTIONS Delay A Mode

Vista™ basic Instructions for Use 15/25950787 Rev 6/02

DELAY A MODEThe start of Continuous, Piggyback, Ramp/Taper, and Program Modes may allbe delayed. The Vista™ basic also provides an option to extend the InfusionComplete alarm at the end of the infusion.

1. Program the infusion.2. Press OPTS until display shows DELAY A MODE.3. Press ON.4. Enter in how many hours and minutes the infusion should begin.

NOTE: This is NOT a clock time like 12:30 pm.5. Press OK.6. Enter how long, in hours and minutes, after the end of the infusion the pump

should wait before alarming.7. Press OK.8. Press YES if the pump should run at KVO during the delay times. Press NO

if the pump should not infuse during the delay times.9. Review the data and press OK. To change the data, press EXIT and begin

with #3 above.10. Press to begin the delayed infusion.

To Exit Delay A Mode1. Stop the infusion.2. Press OPTS until display shows DELAY A MODE.3. Press OFF.4. Press END to return to the previously active mode.

Page 17: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

OPTIONS Dosage Calculation

Vista™ basic Instructions for Use 16/25950787 Rev 6/02

DOSAGE CALCULATIONThis feature allows a rate to be calculated given the concentration and dose.

1. Press OPTS until display shows DOSAGE CALC.2. Press ON.3. If concentration unit desired is other than the one displayed, press button

under the mcg? until units desired appear in the CONCENTRAT. (0*/ml) line.4. Enter the concentration of the solution. If the concentration of the solution is

not known, go to #6 below. Press C to clear current data and re-enter newdata.

5. Press OK.6. To calculate the concentration of the solution, use ← to move the * to the mg

field. Enter the amount of drug in the solution. Then press → to move the *to the ml field. Enter the amount of solution.

7. Press OK.8. Enter the body weight if part of the dosing parameters, then press OK. If

body weight is not part of the dosing parameters, press OK with 0 in the bodyweight field.

9. If time unit desired is other than the one displayed, press button under the/h? until time unit desired appears in the first line. Enter the dose or use tomove the * to the rate field and enter the rate.

10. Press OK.11. Press to begin the infusion.

To Review Data During Infusion1. Press INFO twice.

To Exit Dosage Calculation1. Stop the infusion.2. Press OPTS until display shows DOSAGE CALC.3. Press OFF.4. Press END.

Page 18: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

OPTIONS Bolus Mode, Interval Bolus (Intermittent Mode)

Vista™ basic Instructions for Use 17/25950787 Rev 6/02

BOLUS MODEThis feature allows a bolus to be given during a Continuous Mode infusion.

1. Press OPTS until display shows BOLUS MODE.2. Press ON.3. Press RATE.4. Enter the rate at which the bolus should infuse, then press OK.5. Press END.6. To give a bolus during the infusion, press BOL.7. Enter the amount of bolus to be given in ml.8. Press OK.

Bolus Without Entering a Volume1. During the infusion, press BOL and hold down until a second BOL is

displayed to the right. 2. Continue to hold down the first BOL while also pressing the second BOL.

A bolus will be given as long as both BOL keys are held down. The pumpwill display the amount of bolus being administered and will beep once aftereach milliliter delivered.

INTERVAL BOLUS (INTERMITTENT MODE)The basal/KVO rate and entire bag volume must be entered in CONTINUOUSMODE first.1. Press OPTS until display shows BOLUS MODE.2. Press ON.3. Press TIME.4. Enter the time between the start of each dose, then press OK.5. Press RATE.6. Enter the rate at which the dose is to be given, then press OK.7. Press VTBD.8. Enter the volume of each dose, then press OK.9. Press END.10. Press . The pump will deliver at the basal/KVO rate for the interval time,

then begin delivery of the first dose.

Page 19: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

OPTIONS Total Infused, Standby Function

Vista™ basic Instructions for Use 18/25950787 Rev 6/02

TOTAL INFUSEDThe Total Infused information screen displays the total amount of fluid in ml thathas been pumped since the last time the Total Infused was cleared.

To Clear the Total Infused1. Press OPTS until display shows Tot Infused=.2. Press .3. Press END.

STANDBY This function allows the pump to be put on standby while retaining all set infusiondata. This function is only available in Continuous Mode.

1. Press OPTS until display shows STANDBY2. Press ON.3. Enter time in hours and minutes the pump is to stand by.4. Press C to clear the current data and re-enter new data. Press OK to accept new data.4. Press OK.

The pump will count down to zero and then alarm Pump Stopped!

To Exit Standby Function1. Press EXIT to return to Continuous Mode.

Note: If countdown has already started, press END to return to ContinuousMode.

Page 20: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

OPTIONS Occlusion Pressure, Drop Sensor, Battery

Vista™ basic Instructions for Use 19/25950787 Rev 6/02

OCCLUSION PRESSUREThe pressure at which a downstream occlusion alarm will be triggered may bevaried. The higher the pressure limit, the less sensitive the pump is to changesin fluid resistance.

1. Press OPTS until display shows OCCLUSION PRESSURE.2. Press (--) to change pressure to Low. Press (+) to change pressure to High.3. Press END to return to Continuous Mode.

NOTE: Low corresponds to approximately 8 PSI and high to approximately 17 PSI.High pressure limits may increase the severity of an infiltration without alarming.

DROP SENSORUsing a drop sensor eliminates the need for pre-setting a VTBD in ContinuousMode and senses upstream occlusions and empty solution containers.

NOTE: Turning the Drop Sensor feature off forces volumes to be entered and preventssensing of upstream occlusions.

To Turn On the Drop Sensor1. Press OPTS until display shows DROP SENSOR.2. Press ON.3. Press END to return to Continuous Mode.

To Turn Off the Drop Sensor1. Press OPTS until display shows DROP SENSOR.2. Press OFF.3. Press END to return to Continuous Mode.

BATTERY CAPACITYThis is an indicator of the remaining battery time in hours and minutes. Thisscreen also indicates the amount of time the pump has been in operation.

1. Press OPTS until display shows Battery Left =.2. Press END to return to Continuous Mode

Page 21: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

OPTIONS Data Lock, Contrast Selection

Vista™ basic Instructions for Use 20/25950787 Rev 6/02

DATA LOCKThis features prevents unauthorized use by locking out all keys except ,

,and .

1. Press OPTS until display shows DATA LOCK.2. Press ON.3. Press END to return to Continuous Mode.

To Turn Off Data Lock1. Press OPTS until display shows DATA LOCK.2. Press the decimal key.3. Press OFF.4. Press END to return to Continuous Mode.

CONTRAST SELECTIONThis feature allows the contrast of the screen to be changed for best possibleviewing.

1. Press OPTS until display shows CONTRAST SELECTION.2. Press (+) to increase the contrast (make the text darker). Press (--) to

decrease the contrast (make the text lighter). 3. Press END to return to Continuous Mode.

Page 22: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

OPTIONS Adjust Clock, Loudness

Vista™ basic Instructions for Use 21/25950787 Rev 6/02

ADJUST CLOCKThis feature allows the time and date to be changed.

1. Press OPTS until display shows ADJUST CLOCK.2. Press DAT to enter the date. (Press C to clear the current data and re- enter new data.)3. Enter the date using two digits for month, day and year.

Example: November 5, 2001 is entered as 11 05 014. Press OK5. Press TIME to enter the time of day.

Note: The time is set using 24 hour Military Time. Example: 6:30pm is entered as 18:30

6. Press OK7. Press END to return to Continuous Mode.

LOUDNESSThis feature allows the volume of the alarm tone to be changed.

1. Press OPTS until display shows LOUDNESS.2. Press (+) to increase the alarm volume. Press (--) to decrease the volume.3. Press END to return to Continuous Mode.

Page 23: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

OPTIONS Alarms

Vista™ basic Instructions for Use 22/25950787 Rev 6/02

In alarm situations, press to silence the alarm. Correct the problem. Resumethe infusion by pressing .

Alarm Possible Cause ActionAir-In-Line! Air bubble in air-in-line sensor Remove air from system

Check DropSensor!Free Flow!Not Enough Drops!Too Many Drops!

Solution container emptyRoller clamp closedSet loaded incorrectlyKinks, closed clamps, orblockageDroplets on wall of drip chamberDrop sensor out of position

Replace solution containerOpen roller clampRe-load setCheck for blockage

Shake drip chamberReposition drop sensor

Door Open! Door opened during infusion Close door

DwnstreamOcclusion!

Downstream tubing kinkedFilter blockedClosed clampIV positional or infiltratedCatheter or vein too smallOcclusion limit too low

Correct kinksCorrect blocked filterOpen clampsCheck IV siteIncrease occlusion limit

Invalid data! Entered data is out of range Press ; re-enter data

Keep Vein OpenRate: InfusionComplete

Programmed infusion iscomplete

Stop infusion

Low Battery! 30 minutes until battery is fullydischarged

Silence alarm and connectpump to AC power

Plug In Now! Battery is discharged Connect pump to AC power

Pump Stopped! No key panel activity for 2minutes or Standby time expired

Silence alarm; begin infusion

Unit Malfunction! Internal error Press until display reads:“Do not press any key untildisplay is off;” wait until pumpturns off, turn device on. Ifalarm repeats, return forservice.

Page 24: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

AC, 12 V, BATTERY, & CC INTERFACE

Vista™ basic Instructions for Use 23/25950787 Rev 6/02

AC and 12 V♦ Check AC voltage as per

rating label.♦ Insert/screw on power lead

(12 V lead in ambulance car)on back.

Battery Operation♦ Switching to internal battery is

automatic in the event ofpower failure or if 12 V or ACis not connected.

♦ Charge battery on initial use,following a battery alarm, orafter 2 months of non-use.

♦ Charging time: approx. 16hours. Longer charging hasno detrimental effect.

♦ Full battery provides operationfor 3.0 hours at highestdelivery rate.

♦ Battery capacity is approx.50% after 2 years.

♦ The service life of the batterycan be extended bycompletely discharging it fromtime to time and then fullyrecharging it on AC power.

CC InterfaceInterface descriptions available fromB. Braun. Connection to interfaceinput uses MFC plug.

Documentation♦ All operating data of the

Vista™ basic can be called upand logged via externalcomputer

Page 25: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

MAINTENANCE, STORAGE, & SERVICE

Vista™ basic Instructions for Use 24/25950787 Rev 6/02

MaintenanceClean the pump with a soft, lint-freecloth or swab dampened with soapand water, a general non stainingchemical disinfectant, isopropylalcohol (15%), derivatives ofphenylphenol (5%) or ammoniumchloride (10%).

As a general recommendation forcleaning and disinfecting thisdevice, use a lint-free clothdampened (not soaked) with a 0.5%bleach (Sodium Hypochlorite)solution (10% solution with water).The solution is most effective whenprepared weekly and allowed toremain on the device forapproximately 10 minutes prior towiping off.♦ Do not use acetone solutions

containing glutaraldehyde orabrasive cleansers on thepump. If necessary, Betadineor Iodine solutions may beused, but they will stain thepump.

♦ Do not use abrasive cleaningtools or pressurized sprayingdevices as these may damagethe device.

♦ Do not sterilize the pump usingethylene oxide (EtO) gas.

♦ Do not steam autoclave orimmerse the pump in any fluidsor cleaning solutions.

♦ Always turn off the pump anddisconnect it from the electricaloutlet prior to cleaning.

StorageStore the pump away from excessiveheat, cold, or humidity. Keep devicesplugged into electrical power duringstorage.

ServiceIf the pump fails to respond to theoperating or troubleshooting procedureslisted in this manual and the causecannot be determined, discontinue useand forward it to an authorized B. BraunService Center.

Should it be necessary to return a pumpfor repair, contact Technical Support atB. Braun Customer Service at 800-627-PUMP. A Returned MaterialsAuthorization number will be provided.Carefully pack the pump (preferably inthe original packing) and ship it prepaidto:

B. Braun Medical Inc.Attn: Service Manager1601 Wallace DriveSuite 150Carrollton, TX 75006 USA

B. Braun cannot assume responsibilityfor loss or damage to returned pumpswhile they are in transit.

Product complaints may be sent to theQuality Manager at the above address.With each complaint, please includepump serial number and full descriptionof difficulty, including all settings, typesof fluids, times, and alarm messages.Return set used if possible. ContactTechnical Support for an RMA numberprior to return.

Page 26: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

SPECIFICATIONS

Vista™ basic Instructions for Use 25/25950787 Rev 6/02

Mechanism Linear peristaltic volumetric infusion pumpDimensions WxHxD 5.5 in (14cm) x 9.4 in (24cm) x 7.8 in (20cm) Weight 6.8 lbs. (3.1 kg)Power Requirements 100-120 VAC, 50-60 Hz, 13 WattsLeakage Current Max 30µABattery 7.2V, 1.2 Ah NiCd (rechargeable)Battery Operating Time ≈3.5 hours at 125 mL/hrRate Range 0.1 to 800.0 mL/hr in 0.1 mL increments with Vista™

tubingVolume to be InfusedRange

0.1 to 9999.9 mL

Occlusion Pressure Variable Settings Low: 8 PSI High: 17 PSIKVO Rate 1 mL/hr or selected rate, whichever is lowerFlow Rate Accuracy ± 5% measurement time 8hr, temp. 71.6° FAlarms Air-in-Line, Low Battery, Plug In Now, Bag Empty,

Downstream Occlusion, Door Open, Pump Stopped,KVO, Unit Malfunction

Air-in-Line Alarm Air bubble size 0.3mL* or 1.5mL/hr***Can be set from 0.01 to 0.3 mL in serviceprogram**Can be set from 0.5 to 3.5 mL/hr in serviceprogram

Memory User option to retain last Continuous and PiggybackMode data upon power up. Permanent dataretention in Program and Ramp/Taper Modes afterpump is turned off

Tubing B. Braun Vista™ administration setsModes/Features Continuous and Piggyback Modes, Program Mode

for variable infusions, Ramp/Taper Mode for TPNinfusions and Delay A Mode for postponing infusionstart and finish

OperatingTemperatures

Operating: 64.4° F to 95° F, 10.15 PSI to 15.37PSINon-Operating: -13° F to 131° F, 10.15 PSI to15.37 PSI

Page 27: Vista basic - InfuSystem. Braun Vista Basic.pdf · WARNINGS & CAUTIONS Vista™ basic Instructions for Use 3/25 950787 Rev 6/02 The following should be observed to avoid patient/user

Vista™ basic Instructions for Use950787 Rev 6/02

©2001 B. Braun Medical Inc.

B. Braun Medical Inc.824 Twelfth Ave.Bethlehem, PA 18018