vibration sensitivity of optical microscopes in the h lth ......... in aisc dg 11dg 11 • all other...
TRANSCRIPT
Vibration sensitivity of yoptical microscopes in the
h lth ttihealthcare setting
Hal Amick, PhD, PE Michael Gendreau, INCE.Bd.Cert., Member
Colin Gordon & Associates
Paper 4aNS6 21 May 2009
OverviewOverview
• Review of current healthcare facilityReview of current healthcare facility vibration criteria
• Process of selecting criteria• Process of selecting criteria• Process of selecting design parameters• Case Study: Surgical microscope and its
criteria• Summary of criteria we use
2
Paper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
Current Healthcare Vibration CriteriaCurrent Healthcare Vibration Criteria
• Surgical SuitesSurgical Suites– 100 µm/s (4000 µin/s), as defined by ISO and
ANSI (rms one-third octave bands)ANSI (rms one third octave bands) – Misprinted as 200 µm/s (8000 µin/s) in AISC
DG 11DG 11 • All other spaces require engineering
judgment invoking criteria for other typesjudgment invoking criteria for other types of spaces
3
Paper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
Process of Selecting Criteria (1)Process of Selecting Criteria (1)
• How is vibration a problem?How is vibration a problem?– Human perception
• Startle• Distraction• Sleep interference
A• Annoyance– Affects instrument performance
• Degrades instrument performance• Degrades instrument performance• Introduces errors into data• Affects performance of person using instrument
4
Paper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
Process of Selecting Criteria (2)Vib i ff i P lVibrations affecting People
• Startle and distraction are critical conditions to avoid in surgical suites. – ISO and ANSI standards (hence ASHRAE and AISC) use factorISO and ANSI standards (hence ASHRAE and AISC) use factor
of safety (0.5) times human threshold of perception• Sleep interference is an important issue in patient
roomsrooms. – Sleep environment is basis of ISO/ANSI recommendations for
residential-nighttime limit of perception threshold. • Annoyance can be avoided in other areas by use of• Annoyance can be avoided in other areas by use of
“office” criterion from ISO, ANSI, ASHRAE, AISC.– Allows some perceptible vibration but avoids annoyance range.
5
Paper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
Process of Selecting Criteria (3)Vib i ff i IVibrations affecting Instruments
• Degrades instrument performance; may introduce errors g p yinto data. Most likely an issue with imaging (MRI and CT) and lab equipment (commonly microscopes).– Where possible, use instrument manufacturers’ criteria (MRI, CT, etc.)– Criteria for bench microscopes can be based on Amick & Stead,
ASHRAE, AISC
• Affects performance of person using instrument. Can lead to eye fatigue or worse (misreading or miscounting in lab tests; errors, nausea or annoyance with surgical microscopes)– Only criteria for surgical microscopes are from House & Randell,
referenced in AISC; discussed in later slides
6
Paper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
Benchtop Microscope Sensitivity, p p yOmnidirectional (Amick & Stead)
10000
1000 40x
100
Velo
city
, μ
m/s
1000x
400x
100x
10
rms
V
40x100x400x1000
1000x
11 10 100 1000
Frequency, Hz
1000x
7
H. Amick and M. Stead, “Vibration Sensitivity of a Laboratory Bench Microscope,” Invited Paper, presented at the First Pan-American/Iberian Meeting on Acoustics; 144th Meeting of the Acoustical Society of America, 2-6 December 2002, Cancun, Mexico
Variation of Vibration Sensitivity i h M ifi i (A i k & S d)with Magnification (Amick & Stead)
y = 1995.1e-0.0027x
R2 = 0.9877
1000
10000
de,
μm/s With Newport Table
On Lab BenchExpon. (With Newport Table)Expon (On Lab Bench)
100
or A
mpl
itud Expon. (On Lab Bench)
AVC
y = 158.61e-0.0024x
R2 = 0.965110
Vert
ical
Flo B
C
10 200 400 600 800 1000 1200
Magnification, times
V
8
H. Amick and M. Stead, “Vibration Sensitivity of a Laboratory Bench Microscope,” Invited Paper, presented at the First Pan-American/Iberian Meeting on Acoustics; 144th Meeting of the Acoustical Society of America, 2-6 December 2002, Cancun, Mexico
Process of Selecting Design PParameters
• Footfall is generally the critical designFootfall is generally the critical design parameter
• Mechanical vibrations generally less thanMechanical vibrations generally less than those due to footfall unless something is defective or unless floor is very stiff (e.g., y ( gImaging and some MRI suites)
• Consider context in selecting footfall parameters
9
pPaper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
Context-dependent Footfall P (1)Parameters (1)
• Footfall forces are a function of pace ratep• Walker pace rate is a function of path and
activity• Path issues• Path issues
– Closed path or corridor: long path, no obstructions – high walker rate (100 or 120 paces/min, we use 100)100)
– Open path or ghost corridor: long path, some obstructions – medium walker rate (80 to 85 paces/min we use 85)paces/min, we use 85)
– Patient room, lab room, or between lab benches:short path, obstructions – slow walker rate (70 to 80 paces/min we use 75)
10
paces/min, we use 75)Paper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
Context-dependent Footfall P (2)Parameters (2)
• Walker pace rate is a function of path and p pactivity
• Activity issues– Critical care: staff often in a hurry in the public
corridors – higher walker rate (consider 120 paces/min)p )
– Non-critical care: staff are less frequently in “hurried” mode in public corridor (consider 100 paces/min)Patient room: short path obstructions hard to– Patient room: short path, obstructions – hard to develop the gait associated with fast walker (70 to 80 paces/min, we use 75)
11
Paper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
Context-dependent Footfall P (3)Parameters (3)
• Perform multiple analyses usingPerform multiple analyses using appropriate walker forces applied at “soft” spots along walker path; base design onspots along walker path; base design on the condition creating maximum floor amplitudesamplitudes
• Vibrations due to walker at 75 ppm in room may be more severe than at 100room may be more severe than at 100 ppm in nearby corridor
12
Paper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
Case Study: The “established” criteria may need some rationalcriteria may need some rational
modification
Paper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
Microscopic SurgeryMicroscopic Surgery• Some types ofSome types of
surgery require microscopy (10x tomicroscopy (10x to 50x)– OphthalmicOphthalmic– Spine
Paper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the
14
of optical microscopes in the healthcare setting”, 157th
Meeting ASA, Portland, 2009
Criteria?Criteria?• ASHRAE
– “Microsurgery, eye surgery, neurosurgery”, use 25 µm/s (1000 µin/s)
AISC / H & R d ll• AISC / House & Randell– Criterion of 50,000 / M µin/s, where M is
magnification, at frequencies between 3 and 8 Hz, g , q ,relaxed at higher frequencies; use 1250 µin/s (30 um/s) at 40x.
• Amick & Stead• Amick & Stead– Criterion of 100 um/s (4000 µin/s) for benchtop
microscope of 40x to 100x
15
Paper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
Case Study ProblemCase Study Problem
• Our client: A regional medical center withOur client: A regional medical center with four operating rooms, two dedicated to microsurgery using floor-mountedmicrosurgery, using floor mounted microscopes
One Leica one Zeiss– One Leica, one Zeiss• “Occasionally” there are vibrations which
cause the image to jigglecause the image to jiggle
16
Paper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
What we foundWhat we found …• Nice, stiff, concrete structure, ,• The “typical” ambient vibration environment in
these OR’s was below 50 um/s (2000 µin/s) (OR i i i 100 / )criterion is 100 µm/s)
• Footfall below 50 um/s (2000 µin/s)St d t t h t bl ji l• Steady-state has some acceptable jiggle
• “Problem”—high-amplitude jiggle—occurs a few minutes at a time a few times a dayminutes at a time, a few times a day
• One surgeon routinely experiences nausea during the “problem”
17
g pPaper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
Routine Floor Measurements(why such a problem?)
Vertical Floor Vibration - One-Third Octave Band10000
1000
during vibration inducedvisibility interferencewithout vibration inducedvisibility interferencePerception
Problem
100
Velo
city
, µm
/s
VC-A
10
rms
V
11 1.3 1.6 2 2.5 3.2 4 5 6.3 8 10 12.5 16 20 25 31.5 40 50 63 80
18
One-Third Octave Band Center Frequency, Hz
Paper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
Compare Floor and Eyepiece(in Narrowband)
a) Floor b) Eyepiece
Over 100 arc-sec at 50mm (explains nausea)
100
1000
w/ Interference
w/o Interference
100
1000
w/ Interference
w/o Interference
10
100
ocity
, um
/s
10
100
ocity
, um
/s12 Hz
24 Hz
1
rms
Velo
1rm
s Ve
lo
0.10 10 20 30 40 50
0.10 10 20 30 40 50
19
Frequency, Hz0 10 20 30 40 50
Frequency, HzPaper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
Compare Floor and Eyepiece(in Narrowband)
Ambient1000
w/ Visual Interference1000
100
1000
Eyepiece
Floor
100
1000
Eyepiece
Floor
10ocity
, um
/s
10
loci
ty, u
m/s
1
rms
Vel
1
rms
Vel
0.10 10 20 30 40 50
0.10 10 20 30 40 50
20
Frequency, Hz Frequency, HzPaper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
The ProblemThe Problem …
• Resonance amplification is a fact of lifeResonance amplification is a fact of life– 8 to12 Hz and 18 to 21 Hz, in this
configuration• Intermittent vibration from mechanical
equipment (12.0 Hz and harmonics) only slightly exceeded VC-A
• Improve the vibration isolation on the mechanical equipment
• Was VC-A adequate?
21
Paper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
What Criterion is Adequate?10000
Eqpt OnEqpt Off
What Criterion is Adequate?• Surgery
1000
10000
um/s
SurgeryVC-A10x (House & Randell)40x (House & Randell)100x (House & Randell)
g ycriterion notadequate
• VC A and10x
10
100
Velo
city
, u • VC-A and House & Randell 40x are
d t
40x
100x
1rms
V adequate• Consider
extending H&R0.1
1 10 100Frequency, Hz
extending H&R “dip” to the right or removing relaxation
22
relaxationPaper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
Proposed Modification to House & R d llRandell
• “Dip” in H&R 10 (H & R d ll)
pcriterion did not correspond to observed 1000
10000 10x (House & Randell)
40x (House & Randell)
100x (House & Randell)
Resonances
resonances• Solution:
Slide dip to100
Velo
city
, um
/s
3 - 8 Hz
– Slide dip to higher frequency
– Treat as single-velocity criterion
N.T.S.10
rms
V
velocity criterion, V=1250 / M(where M is magnification)
11 10 100
Frequency, Hz
8 - 13 Hz
23
magnification)Paper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
Vibration Criteria for Hospitals (1)Vibration Criteria for Hospitals (1)Type of Space
Primary Factor Criterion Rationale
This is defined by ISO and ANSI The established threshold of
Surgical Suites Human
100 µm/s (4000 µin/s)
This is defined by ISO and ANSI. The established threshold of human perception is 200 µin/s (8000 µin/s). Historically, the argument was that a factor of safety of 2 against perception avoided the risk of startling the surgeon. [Misprinted as 8000 µin/s in AISC DG 11.Correct in ASHRAE]
30 µm/s AISC DG 11 based on research by House and Randell validatedSurgical Suites
40x Surgical Microscope *
30 µm/s (1250 µin/s)
AISC DG 11, based on research by House and Randell, validated by Gendreau
Surgical 100x Surgical 12.5 µm/s (500 i / )
AISC DG 11, based on research by House and Randell Surgical Suites
100x SurgicalMicroscope *
(500 µin/s)(VC-C)
Patient Rooms Human 200 um/s
(8000 µin/s)
This is not specifically defined by international standard (differing from the case for surgical suites), but is based upon the international standard for sleeping areasRooms (8000 µin/s) international standard for sleeping areas.
24
Paper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
Vibration Criteria for Hospitals (2)Vibration Criteria for Hospitals (2)Type of Space
Primary Factor Criterion Rationalep
General Labs Instrument
50 µm/s (2000 µin/s)
(VC-A)
This is a consensus standard from a wide variety of sources, including ASHRAE, AISC, IEST, and NIH for generic laboratory space with microscopes up to 400x.
General Labs Instrument 100 µm/s
(4000 µin/s)
This is a relaxed criterion for “non-critical” laboratories with microscopes of 100x or less. Used by many universities for teaching labs (i.e., labs not used for research)
Imaging12.5 µm/s Imaging systems vary widely in their sensitivity. The 500 µin/s
criterion (approximately) is required to meet the needs of a few ofImaging Labs (MRI) Instrument (500 µin/s)
(VC-C)
criterion (approximately) is required to meet the needs of a few of the available systems. By eliminating those from consideration, the criterion can be relaxed.
25
Paper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009
ReferencesReferences• AISC DG 11: Murray, Thomas M., David E. Allen, and Eric E. Ungar (1997), “Floor Vibrations
Due to Human Activity," Steel Design Guide Series 11, American Institute of Steel Construction, y, g , ,69 pp.
• Amick and Stead: H. Amick and M. Stead, “Vibration Sensitivity of a Laboratory Bench Microscope,” Invited Paper, presented at the First Pan-American/Iberian Meeting on Acoustics; 144th Meeting of the Acoustical Society of America, 2-6 December 2002, Cancun, Mexico
• Amick and Stead: H. Amick and M. Stead, “Vibration Sensitivity of a Laboratory Bench Mi ” S d & Vib ti 41 N 2 10 17 (F b 2007)Microscope,” Sound & Vibration, v. 41, No. 2, pp. 10-17 (February 2007).
• ASHRAE: ASHRAE (2003). ASHRAE Handbook: Applications, Chapter 47 (Fig. 39 & Table 40), “Sound and Vibration Control,” 50 pp.
• House and Randell: House, M. H., and Randell, R. (1987). “Some Measurements of Acceptable Levels of Vibration in Scientific, Medical and Ophthalmic Microscopes,” Proc. SPIE Conf. Vib. Con Opt and Metrology 732 pp 74 80 [ V = 1250 / M um/s where M is magnification ]Con. Opt. and Metrology 732, pp. 74-80 [ V = 1250 / M um/s, where M is magnification ]
• IEST: Institute of Environmental Sciences and Technology, “Considerations in Cleanroom Design,” RP-CC012.2, 2005.
• ISO: International Standards Organization, ISO 2631 "Mechanical vibration and shock -Evaluation of human exposure to whole-body vibration, Parts 1 and 2." Part 1 was updated 15 July 1997 and Part 2 was updated 1 April 2003July 1997 and Part 2 was updated 1 April 2003.
• NIH: National Institutes of Health, Office of Research Fachilities (Spring 2003). NIH Design Policy and Guidelines
26
Paper 4aNS6: Amick & Gendreau, “Vibration sensitivity of optical microscopes in the healthcare setting”, 157th Meeting ASA, Portland, 2009