ven$lator**hksccm.org/files/presentations/ventilator_final_version_180215.pdfgas flows along a...

64
Ven$lator

Upload: others

Post on 25-Jun-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Ven$lator    

Page 2: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

•  Posi$ve  pressure  •  Nega$ve  pressure  

Page 3: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

•  The  first  iron  lung  was  used  on  october  12  at    children  hospital  boston  

•  used  in  a  child  unconscious  from  respiratory  failure  

Page 4: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted
Page 5: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Iron  lung  ward  filled  with  Polio  pa$ents,    Rancho  Los  Amigos  Hospital,  ca.  1953    

Page 6: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Sessions    

Page 7: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted
Page 8: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted
Page 9: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

•  Mechanical  ven$lators    used  increasingly  in    

•  Anesthesia  and  intensive  care  •  To  treat  polio  pa$ents  and    •  The  increasing  use  of  muscle  relaxants  during  anesthesia  

Page 10: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Indica$on  for  intuba$on  

•  Protect  airway  •  Maintain  airway  •  Mechanical  ven$la$on  •  Bronchial  toile$ng  

Page 11: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Indica$on  for  mechanical  ven$la$on  

•  Ven$la$on  Failure    •  Oxygena$on  Failure    

Page 12: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

. gas flows along a pressure gradient between the upper airway and the alveoli

Flow is either volume targeted and pressure variable, or pressure limited and volume variable.

The pattern of flow may be either sinusoidal (which is normal), decelerating or constant. Flow is controlled by an array of sensors and microprocessors.

expiration is passive

Mechanical  ven$lator  

Page 13: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Control  

•  Either  Volume  Controlled  (volume  limited,  volume  targeted)  and  Pressure  Variable  or  

•  Pressure  Controlled  (pressure  limited,  pressure  targeted)  and  Volume  Variable  or    

•  Dual  Controlled  (volume  targeted  (guaranteed)  pressure  limited)  

Page 14: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Cycling    

•  Time  cycled  -­‐  such  as  in  pressure  controlled  ven$la$on  

•  Flow  cycled  -­‐  such  as  in  pressure  support  

•  Volume  cycled  -­‐  the  ven$lator  cycles  to  expira$on  once  a  set  $dal  volume  has  been  delivered:  this  occurs  in  volume  controlled  ven$la$on  

Page 15: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Triggering    

•  what  causes  the  ven$lator  to  cycle  to  inspira$on?  •  Ven$lators  may  be  $me  triggered,    •  pressure  triggered  or    •  flow  triggered.  •   Time:  the  ven$lator  cycles  at  a  set  frequency  as  determined  by  the  controlled  rate.  

•  Pressure:  the  ven$lator  senses  the  pa$ent's  inspiratory  effort  by  way  of  a  decrease  in  the  baseline  pressure.  

Page 16: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Flow  trigger  

• modern  ven$lators  deliver  a  constant  flow  around  the  circuit  throughout  the  respiratory  cycle  (flow-­‐by).    

•  A  deflec$on  in  this  flow  by  pa$ent  inspira$on,  is  monitored  by  the  ven$lator  and  it  delivers  a  breath  

•  This  mechanism  requires  less  work  by  the  pa$ent  than  pressure  triggering.  

•         

Page 17: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Breath  are  either  

•  Mandatory  (controlled)  -­‐  which  is  determined  by  the  respiratory  rate.  

•  Assisted  -­‐    (as  in  assist  control,  synchronized  intermi^ent  mandatory  ven$la$on,  pressure  support)  

•  Spontaneous-­‐      (no  addi$onal  assistance  in  inspira$on,  as  in  CPAP)  

Page 18: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Flow  pa^ern  

•  Sinusoidal  =  this  is  the  flow  pa^ern  seen  in  spontaneous  breathing  and  CPAP  

•  Decelera$ng  =  the  flow  pa^ern  seen  in  pressure  targeted  ven$la$on  

•  inspira$on  slows  down  as  alveolar  pressure  increases  (there  is  a  high  ini$al  flow).  

Page 19: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Flow  pa^ern  

•  Constant    -­‐  flow  con$nues  at  a  constant  rate  un$l  the  set  $dal  volume  is  delivered  

•  Accelera$ng    -­‐  flow  increases  progressively  as  the  breath  is  delivered.  This  should  not  be  used  in  clinical  prac$ce.  

Page 20: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Various  modes  of  mechanical  ven$la$on    

Page 21: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Control  Modes  

– every breath is fully supported by the ventilator –  in classic control modes, patients were unable

to breathe except at the controlled set rate –  in newer control modes, machines may act in

assist-control, with a minimum set rate and all triggered breaths above that rate also fully supported.

Page 22: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

CMV  

Page 23: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Assist-­‐control  

Ingento  EP  &  Drazen  J:  Mechanical  Ven$lators,  in  Hall  JB,  Scmidt  GA,  &  Wood  LDH(eds.):  Principles  of  Cri.cal  Care  

Page 24: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted
Page 25: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

SIMV  

Page 26: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

SIMV  

Ingento  EP  &  Drazen  J:  Mechanical  Ven$lators,  in  Hall  JB,  Scmidt  GA,  &  Wood  LDH(eds.):  Principles  of  Cri.cal  Care  

Page 27: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Volume  Ven$la$on  Tidal  volume,  is  not  affected  by  the    

rapidly  changing  pulmonary  mechanics    

Compliance  ↓      

 Pressure  Ven1la1on:  Volume  Ven1la1on:    Decreased  Tidal  Volume  Increased  Pressure  

Volume  

Pressure  Pressure  

Volume  

Page 28: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted
Page 29: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Volume  targeted  

Page 30: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Pressure  Control  vs.  Pressure  Support  

Constant  insp.  pressure  Decelera$ng,  variable  

inspiratory  flow  rate    Time  cycled:  (A)  •  Pressure  Control  Flow  cycled:  (B)  •  Pressure  Support  

Pressure  

Flow  

A   B  

Time  Cycled  

       Flow  Cycled  

Page 31: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Pressure  targeted  

Page 32: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted
Page 33: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted
Page 34: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Posi$ve  Airway  Pressure  Can  Be  Either  Pressure  or  Flow  Controlled—But  Not  Both  Simultaneously    

Dependent  Variable  

Dependent  Variable  Set  Variable  

Set  Variable  

Page 35: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Rise  $me  

Page 36: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

How  to  set  Ti  in  a  spontaneous  breathing  pa$ent  on  a  pressure  support  mode  ?  

Flow  

Pressure  

Tinsp. PIP

Peak Flow

25%

Pressure  Control   Pressure  Support  

“Flow termination criteria”

Page 37: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted
Page 38: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted
Page 39: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Termina$on  Sensi$vity  =  Cycle-­‐off  Criteria  

Flow  

Peak  Flow  (100%)  

TS  5%  

Tinsp.  (eff.)  Set  (max)  Tinsp.  

Leak  

Time  

Page 40: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Combina$on  “Dual  Control”  Modes    

Combina$on  or  “dual  control”  modes  combine  features  of  pressure  and  volume  targe$ng  to  accomplish  ven$latory  objec$ves  which  might  remain  unmet  by  either  used  independently.  

 Combina$on  modes  are  pressure  targeted  

Par$al  support  is  generally  provided  by    pressure  support  Full  support  is  provided  by  Pressure  Control  

Page 41: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Combina$on  “Dual  Control”  Modes    

Volume Assured Pressure Support (Pressure Augmentation)

Volume Support (Variable Pressure Support) Pressure Regulated Volume Control

(Variable Pressure Control, or Autoflow) Airway Pressure Release (Bi-Level, Bi-PAP)

Page 42: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

PRVC  (Pressure  regulated  volume  control)  

A control mode, which delivers a set tidal volume with each breath at the lowest possible peak pressure.

Delivers the breath with a decelerating flow

pattern that is thought to be less injurious to the lung…… “the guided hand”.

Page 43: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted
Page 44: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

PRVC  Automa$cally  Adjusts  To  Compliance  Changes  

Page 45: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Servo  PRVC  

Page 46: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

PRCV:    Advantages  

Decelera$ng  inspiratory  flow  pa^ern  Pressure  automa$cally  adjusted  for  changes  in  compliance  and  resistance  within  a  set  range  Tidal  volume  guaranteed  Limits  volutrauma  Prevents  hypoven$la$on  

 

Page 47: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

PRVC:  Disadvantages  Pressure  delivered  is  dependent  on  $dal  volume  achieved  on  

last  breath  Intermi^ent  pa$ent  effort  ⇒  variable  $dal  volumes  

Pres

sure

Fl

ow

Volu

me

Set tidal volume

 ©  Charles  Gomersall  2003  

Page 48: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

Pres

sure

Fl

ow

Volu

me

Set tidal volume

PRVC:  Disadvantages  Pressure  delivered  is  dependent  on  $dal  volume  achieved  on  

last  breath  Intermi^ent  pa$ent  effort  ⇒  variable  $dal  volumes  

 ©  Charles  Gomersall  2003  

Page 49: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

APRV  

•  BiVent  -­‐  (Servo-­‐i  ven$lator  by  Maquet)  •  BiLevel  -­‐  (Puritan  Benne^  840  ven$lator  by  Covidien)  

•  DuoPAP  -­‐  (  C-­‐1  ven$lator  by  Hamilton)  

Page 50: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

APRV  

Page 51: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted
Page 52: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted
Page 53: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

•  An  applica$on  of  CPAP  •  Con$nuous  posi$ve  airway  pressure  (CPAP)  with  an  intermi^ent  release  phase.  

•  Pa$ent  cycles  between  two  levels  of  CPAP—higher  one  P  High(P1),  the  lower  P  Low(P2)  

•  The  pa$ent  can  breath  spontaneously  at  either  level  •  Maintains  an  op$mal  FRC  maintain  adequate  lung  volume  and  alveolar  recruitment.  

•  Occasional  pressure  releases  augments  CO2removal    

Page 54: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

•  P  High–the  upper  CPAP  level.  Analogous  to  MAP  (mean  airway  pressure)  and  thus  affects  oxygena$on  

•  P  Low/PEEPis  the  lower  pressure  seong.  •  T  High-­‐is  the  $me  spent  at  P  High  •  T  Low-­‐is  the  release  $me  allowing  CO2elimina$on  

Page 55: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted
Page 56: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

•  Applica$on  of  ‘P  high’  and  ‘T  high’:(80-­‐95%)  of  the  cycle  $me—”open  lung”  

•  Mean  airway  pressure  (MAP)  is  increased  •  Insures  almost  constant  lung  recruitment  

– Decreases  repe$$ve  infla$on/defla$onDecreases  risk  of  VILI  

– Decreases  need  for  recruitment  maneuvers  

Page 57: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

•  Minute  ven$la$on/CO2  removal  depend  upon:Lung  compliance  

•  Airway  resistance  •  Magnitude  and  dura$on  of  pressure  release  •  Pa$ent’s  spontaneous  breathing  efforts  

Page 58: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted
Page 59: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted
Page 60: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted
Page 61: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

•  Pressure  seongs  P  High–desired  mean  airway  pressure  +  3  cmH2O  (20  to  30  cm)  

•  P  Low–usually  set  at  0  cm  H2O  (  0  to  5  cm)  •  Time  SeongsT  High–usually  set  at  5.0  seconds  (4.5-­‐6.0  sec)  

•  T  Low–usually  set  at  0.6  seconds  (0.5-­‐.8  sec)  •  FiO2  

Page 62: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

 1)  Volume  support  monitors  minute  

ven$la$on  and  $dal  volume  ,  changing  the  level  of  pressure  support  to  achieve  a  volume  target.    

2)  Volume  assured  pressure  support  allows  the  pa$ent  to  breathe  with  pressure  support,  supplemen$ng  the  breath  with  constant  flow  when  needed  to  achieve  the  targeted  $dal  volume  within  an  allocated  $me.      

3)  Propor$onal  assist  varies  pressure  output  in  direct  rela$on  to  pa$ent  effort.  

Several  modes  allow  for  variability  in  pa$ent  efforts  while  achieving  a  targeted  goal.    

Page 63: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

•  Pa$ent  status  

Page 64: Ven$lator**hksccm.org/files/Presentations/ventilator_final_version_180215.pdfgas flows along a pressure gradient between the upper airway and the alveoli Flow is either volume targeted

•  Alarm