vdw-dental.com VDW.GOLD ® RECIPROC ® Tips for Successful Length Determination General Tips for Your Success Always wear gloves; use of rubber dam is recommended Always put protective silicon cover over contra-angle for simultaneous length control Always insert lip clip into the cheek opposite the tooth being treated Always dry access cavity with air blower or cotton ball Separate length determination: choose suitable hand file size (matching the root canal diameter) First determine working length using a hand file, then control length simultaneously via the contra-angle during preparation 2. Separate Length Determination Clamp hand instrument into file clip and hang lip clip into patient's cheek pouch 3. Simultaneous Length Control Insert rotary or reciprocating file into contra-angle and hang lip clip into patient's cheek pouch 4. Option: APICAL-AUTO STOP can be activated via the menu MENU APEX STOP >ON The working length is set at the 3rd green LED 3 blue LEDs: Median coronal section 3 green LEDs/ 1 orange LED: Apical section (apical constriction up to apical foramen) 1 red LED: Apical foramen surpassed 1. Activating the Apex Locator Touch micromotor with lip clip Display will show: APEX LOCATOR ON

Upload: others

Post on 28-Sep-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: VDW.GOLDRECIPROC – Tips for Successful Length Determination · • Separate length determination: choose suitable hand file size (matching the root canal diameter) • First determine

vdw-dental.com

VDW.GOLD®RECIPROC® – Tips for Successful Length Determination

General Tips for Your Success

• Always wear gloves; use of rubber dam is recommended

• Always put protective silicon cover over contra-angle for simultaneous length control

• Always insert lip clip into the cheek opposite the tooth being treated

• Always dry access cavity with air blower or cotton ball

• Separate length determination: choose suitable hand file size (matching the root canal diameter)

• First determine working length using a hand file, then control length simultaneously via the contra-angle during preparation

2. Separate Length Determination

Clamp hand instrument into file clip and hang lip clip into patient's cheek pouch

3. Simultaneous Length Control

Insert rotary or reciprocating file into contra-angle and hang lip clip into patient's cheek pouch

4. Option: APICAL-AUTO STOP can be activated via the menu

MENUAPEX STOP >ON

The working length is set at the 3rd green LED

3 blue LEDs:

Median coronal section

3 green LEDs/ 1 orange LED:

Apical section(apical constriction up to apical foramen)

1 red LED: Apical foramen surpassed

1. Activating the Apex Locator

Touch micromotor with lip clip Display will show: APEX LOCATOR ON

Page 2: VDW.GOLDRECIPROC – Tips for Successful Length Determination · • Separate length determination: choose suitable hand file size (matching the root canal diameter) • First determine

vdw-dental.com

VDW.GOLD®RECIPROC® – Additional Recommendations for...

VW

0003

51 R

ev2/

07.0

8.20

18

➀ ➃

➁ ➄

1. Lubricating the 6:1 Contra-Angle (With the help of compressed air)

DO NOT spray display directly!

✔➀

Aldehyde-free solution

2. Surface Disinfection

Charging time approx. 3 h

Consider the following points before charging:• Battery symbol already changes from

red to green after approx. 30 min.• At this point, the battery has a capacity

of only 20 %.• Do not stop charging; charge battery

for 3 full h.

After battery is fully recharged, a continuous use of 2 h is possible.

Symbol flashing red: Charge

Symbol flashing green: DO NOT charge

3. Charging the Battery

AIR 5 sec. MON TUE WED THU FRI SAT SUN

OIL 1 sec. MON TUE WED THU FRI SAT SUN

MON TUE WED THU FRI SAT SUN1 sec.OIL