Page 1: Validation of HID EVOlutionâ„¢ System: Automation and Integration

Jacki BenfieldApplied Markets Division, Applied Biosystems

Validation of HID EVOlution™System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

Page 2: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

2

Disclosure

This presentation will include commercial products from:• Applied Biosystems• Tecan• Other laboratory product suppliers

Page 3: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

3

Outline of Presentation

• Scope• Developmental Validation Studies:

• Quantifiler® Human and Y• AmpFℓSTR® STR kits• Complete System Integration

• Summary

Page 4: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

4

HID EVOlution™ System: Scope

• Develop validated automated solution to streamline the human identification (HID) workflow

• HID EVOlution™ System• Tecan Freedom EVO® 150

• 4 probe disposable tip configuration• AB custom carriers

• HID EVOlution™ System Software• Designed specifically for AB chemistry and instrumentation• Generates instrument setup files and reports• Seamless transfer of sample information and data

Page 5: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

5

Automated Protocols• Quantifiler® Quantification

• Human and Y • Duo (in development)

• 5 dye AmpFℓSTR® PCR Amplification Kits• Identifiler®

• MiniFiler™

• Yfiler®

• 4 dye AmpFℓSTR® PCR Amplification Kits• SGM Plus®

• Profiler Plus®

• COfiler®

• SEfiler™

Page 6: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

6

HID EVOlution™ Development and Optimization

• Robot movement • Disposable tip size • Liquid class optimization • Liquid level detection• Mixing protocols• Aspiration/Dispensing mode• Reagent and sample volume requirements• Dilution schemes

Page 7: Validation of HID EVOlutionâ„¢ System: Automation and Integration

Quantifiler® Human and Y Human Male Studies

Page 8: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

8

Quantifiler® Kit Validation Studies

• Standard Curve Study• Precision and Accuracy

• Quantification Study• Precision

• 0.025, 0.05, 0.1, 1.0, 5.0 and 25.0 ng/µL of DNA in 12 replicates• Compared to manual operations

• Reproducibility• Precision study - run over three days

Page 9: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

9

Quantifiler® Standard Curve Precision and Accuracy Study

Kit Parameter Typical Range Observed Range

Slope -2.9 to -3.3 -3.052 to -3.512

R2 >0.98 0.991 to 0.9999

Slope -3.0 to -3.6 -3.217 to -3.386

R2 >0.98 0.984 to 0.991

Quantifiler® Human (N=75)

Quantifiler® Y (N=40)

Page 10: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

10

Quantifiler® Human: Manual vs HID EVOlution™ System

DNA

0.01

0.1

1

10

100

0.025ng/ul 0.05ng/ul 0.1ng/ul 1ng/ul 5ng/ul 25ng/ul

Ave

rage

DN

A Q

uant

ity (n

g/ul

)

ManualPlate1Plate2Plate3Tube1Tube2Tube3

N =12

Page 11: Validation of HID EVOlutionâ„¢ System: Automation and Integration

Sample Normalization and PCR Setup for AmpFℓSTR® Kits

Page 12: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

12

DNA Normalization & AmpFℓSTR® Kit Automation Validation

• Accuracy, Precision and Reproducibility - 0.025, 0.1, 0.5, 1.0, 2.0, 5.0,10.0 and 50.0 ng/µL of DNA normalized to target input amount

• Microcentrifuge tube and plate configuration • Each configuration prepared in triplicate• Compared with manual operation• Metrics:

• Genotype concordance• Peak Height• Peak Balance

Page 13: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

13

Identifiler®: Precision and Reproducibility Study

Green: Plate

Yellow: Tube

N =171

Page 14: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

14

Identifiler®: Manual vs HID EVOlution™ System

Green: Manual

Yellow: Tecan

N =171

0.025 0.1 0.5 2.01.0 5.0 10.0 50.025.00.05 PC

Page 15: Validation of HID EVOlutionâ„¢ System: Automation and Integration

Complete System Integration Study

Page 16: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

16

HID EVOlution™ System Validation

Report

SDS Real-time set-up file SDS Real-time

sample results file

CE Instrument set-up File

Report

Page 17: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

17

• Automated protocols tested:• Genomic DNA at concentrations: 0.025, 0.1, 0.5, 1, 2, 5, 10,

and 50 ng/µL• 10 replicates per concentration• Quantifiler® Human on a 7500 Real-time PCR System• 4 and 5 dye AmpFℓSTR® Kits • 3130xl Genetic Analyzer

• Position ID (Barcode)• Cross Contamination• Reports

Complete Integrated System Studies

Page 18: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

18

Integrated System Validation – Quantification Results

DNA Quantification

0.001

0.01

0.1

1

10

100

CMBi0646

CFBi06

47CMBi06

40CMBi06

38CFB

i0631

CMBi0625

CMBi0628

BMBi0697

CFBi06

21BMBi06

91BFB

i0681

BFBi06

85BFB

i0686

BMBi0671

BMBi0672

BMBi0662

BMBi0669

BMBi0668

CMBi0650

BMBi0655

DN

A Q

uant

ity (n

g/ul

)

Quantifiler® Human quantification results used for normalization

50 510 0.5 0.12 1 0.025

ng/µL

direct1 step2 step

N=10

Page 19: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

19

Integrated System Average STR Peak Height Results

AmpFℓSTR® Identifiler® PCR Amplification Kit -10 replicates per DNA concentration

0.025 0.1 0.5 2.01.0 5.0 10.0 50.0 Pos

direct 1 step 2 step

ng/µL N=176

Page 20: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

20

Integrated System Heterozygote Peak Height Ratio

N =10

AmpFℓSTR® Identifiler® PCR Amplification Kit

Page 21: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

21

Position ID Confirmation Study

• Barcode scanning of sample tubes and tray confirmation• 80 different DNA samples at varying concentrations• The STR genotype for each sample was compared with

the previously generated known genotype

Results• Complete and accurate barcode record• 100% genotype concordance

Page 22: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

22

Cross Contamination Study• Samples containing 100ng/uL and TE Buffer only• Two plate configurations• Two runs per configuration

No contamination was observed

Stripe Checkerboard

A

BC

D

E

F

G

H

01 02 03 04 05 06 07 08 09 10 11 12

100 ng/uL

TE BufferABCDEFGH

01 02 03 04 05 06 07 08 09 10 11 12100 ng/uLTE Buffer

Page 23: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

23

Quantifiler® Setup Report

Page 24: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

24

Quantifiler® PCR Setup Report

Page 25: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

25

Normalization/STR PCR Sample Information Report

Page 26: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

26

Automation Development and Validation Summary

• Automated protocols and system software have been developed, optimized and validated for the Quantifiler®

Quantification Kits and the AmpFℓSTR® Amplification Kits• Validation studies demonstrated:

• Robust, reproducible and reliable liquid handling • Clean liquid handling capabilities• Accurate data integration and management

Page 27: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

27

Advantages of a Turn Key Automated System

• Optimized and validated protocols enable:• Quick implementation• Improved overall first pass success

• Customized software allows for more effective forensic sample processing

• Complete instrument and data integration minimizes:• Assay set-up time• Transcriptional errors • Data review time

For more information visit: evo.appliedbiosystems.com

Page 28: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

28

Co-authors and AcknowledgementsR&D

Ada Wong Jason LiuJai ShewaleMike CasselRixun FangRob LagaceSonny YoungVivian T. NguyenManohar Furtado

Technical MarketingJacki Benfield

Product ManagementHeidi Kijenski

TecanDirk AbelnLynda Treat-ClemonsMichael DahmsWendy LauberTony Mamone

Page 29: Validation of HID EVOlutionâ„¢ System: Automation and Integration

© 2008 Applied Biosystems02/23/2008

Validation of HID EVOlution™ System: Automation and Integration of Quantification and STR Analysis for High Throughput Sample Processing

29

Legal information

For Research, Forensic or Paternity Use Only. Not for use in diagnostic procedures.

Applera, Applied Biosystems, AB (Design), AmpFLSTR, COfiler, Identifiler, Profiler Plus, SGM Plus, SEfiler, Quantifiler, TaqMan, VIC, and Y-filer are registered trademarks and FAM, MiniFiler, and NED are trademarks of Applera Corporation or its subsidiaries in the US and/or certain other countries.

AmpliTaq Gold is a registered trademark of Roche Molecular Systems, Inc.

EVO® is a trademark of Tecan Corporation.

All other trademarks are the sole property of their respective owners.

Page 30: Validation of HID EVOlutionâ„¢ System: Automation and Integration

HID EVOlution™ System

Thank you