Page 1: UV - VIS SW

UV/VIS SOFTWAREGamal A. Hamid

Page 2: UV - VIS SW

To everyone who has helped us with support, new books, hard / soft ware And over the internet Special thanks

for THERMO

Thanks

Page 3: UV - VIS SW

3 Gamal A. Hamid

Contents Introduction

Software

Menus

Toolbar

Data Store

Status Bar

Reporting

Applications

Page 4: UV - VIS SW

Introduction

Page 5: UV - VIS SW

5 Gamal A. Hamid

UV/VIS An analytical for measure the concentration of

samples or samples components according to

Beer’s lambert Law depending on the ratio

between incident and transmitted light

through the cuvette.

Absorbance is expressed as

A = εbc where ε is a constant called absorptivity

Absorbance is directly proportional to the path

length, b, and the concentration, c, of the

absorbing molecule.

Page 6: UV - VIS SW

6 Gamal A. Hamid

UV/VIS Region The UV is the electromagnetic

spectrum

shorter than those of ordinary, visible

violet light but longer than those of x-

rays.

The human eye is only sensitive

between approximately 380 and 780

nm and within this area we perceive

the colors of the rainbow from violet

through to red.

Page 7: UV - VIS SW

7 Gamal A. Hamid

Spectrometer

A spectrophotometer is an instrument for

measuring the transmittance or

absorbance of a sample as a function of

the wavelength of electromagnetic

radiation.

Single beam instrument.

Double beam instrument.

Page 8: UV - VIS SW

8 Gamal A. Hamid

Instrument Components

1. Light Source

2. Sample Chamber

3. Optical System

4. Detector

Page 9: UV - VIS SW

Software

Page 10: UV - VIS SW

10 Gamal A. Hamid

Main window

Page 11: UV - VIS SW

Menus

11

Page 12: UV - VIS SW

12 Gamal A. Hamid

Menus

File

View

Tools

Application

Command

Quant

Data Store

Table

Window

The Main Menu Bar gives you access to all of the

operations needed to obtain, process, store and

retrieve data.

Each pull-down menu gives you access to one or

more sub-menus.

The most frequently-used operations may also be

accessed directly via icons on the toolbar.

Page 13: UV - VIS SW

13 Gamal A. Hamid

File menu

The File Menu is used for all operations connected with creating, opening, saving and printing files.

The File Menu is used for all operations

connected with creating, opening, saving

and printing files.

Page 14: UV - VIS SW

14 Gamal A. Hamid

View menu

Customize

This opens a tabbed dialog which

gives you control over which

toolbar icons are displayed and

how they are grouped.

You also have some choices about

the appearance of the toolbar

buttons.

Page 15: UV - VIS SW

15 Gamal A. Hamid

Tools menu

The Tools Menu enables you to configure

the software to suit your preferences.

Options

This item opens a tabbed dialog through

which you can control the following:

File Load/Save Behavior

System Settings

Batch Information Printout

Environment Information

Page 16: UV - VIS SW

16 Gamal A. Hamid

Application menu

The Applications Toolbar enables you to

move between one application and

another with a single mouse click,

instead of accessing the applications via

the Application Menu.

Page 17: UV - VIS SW

17 Gamal A. Hamid

Command menu

The available options in the Command

Menu will depend on which instrument

you are using and on the current context.

Page 18: UV - VIS SW

18 Gamal A. Hamid

Baseline

100%T Baseline

Causes the instrument to do a Baseline

Scan over the Wavelength Range and Data

Interval specified in the current method.

This is used to compensate for introduced

differences between the sample and

reference beams other than the presence

of the sample (e.g., differences in

absorption between the Reference position

or sample and the Sample position or

sample).

Page 19: UV - VIS SW

19 Gamal A. Hamid

Math menu

The Math Menu contains all the commands

needed to process your data post-run.

If you are going to repeat the manipulation

on a number of batches of results, the

window can be left open and other batches

dragged to it.

The Math Manipulation window will

remain on top on the work area all the

time it is open.

Page 20: UV - VIS SW

20 Gamal A. Hamid

Quant menu

No curve fit

Select all curve fits

Select Linear fit

Select Linear through zero fit

Select Quadratic fit

Select Quadratic fit through zero

Page 21: UV - VIS SW

21 Gamal A. Hamid

Data Store menu

The Data Store menu gives you access to

commands that operate on data in the

Data Store.

Not all commands will be applicable in all

circumstances.

Commands not available in the current

context will be grayed out.

Page 22: UV - VIS SW

22 Gamal A. Hamid

Table- Graph- Report menus

When the Results Table has the

focus Table is added to the Main

Menu Bar.

When the Graph window has the

focus Graph is added to the Main

Menu Bar.

When a Report window exists and

has the focus Report is added to

the Main Menu Bar.

Page 23: UV - VIS SW

23 Gamal A. Hamid

Window menu

The options in the Window Menu give

you control over the way in which

windows are displayed on the program

work area and enable you to transfer

the focus from one window to another.

Clicking on your selected window will

transfer the focus to that window and

re-open it if that window has previously

been closed or minimized.

Page 24: UV - VIS SW

Toolbars

24

Page 25: UV - VIS SW

25 Gamal A. Hamid

The toolbars

The toolbars contain icons for the most

commonly used operations, enabling them

to be accessed with a single mouse click

rather than via the pull-down menus on

the Main Menu Bar.

Page 26: UV - VIS SW

26 Gamal A. Hamid

The toolbars groups

File operation group

Application group

Analysis group

Window management group

Window group

Graph group

Page 27: UV - VIS SW

27 Gamal A. Hamid

File operation group

Create a new method in the current application

Load a method for the current application from disk

Save the Sample Method to disk using its current name

Load a result file for the current application from disk

Save the file currently selected in the Data Store to disk using its current name

Page 28: UV - VIS SW

28 Gamal A. Hamid

Application group

Starts the Scan Application (shows the appearance of the button for the active application)

Starts the Fixed Application

Starts the Quant Application

Starts the Rate Application (available only if VISIONlife is installed)

Starts the MCA (Multicomponent Analysis) Application

Starts the DNA Melting Application (available only if VISIONlife and suitable accessories are

installed)

Page 29: UV - VIS SW

29 Gamal A. Hamid

Analysis group

Starts a run in the current application using the current Application and Accessory Method

parameters.

Toggle to pause start/stop data collection.

Start a Zero measurement or a baseline scan as applicable using the current Application

Method parameters.

Stop the current run

Access this Help system

Page 30: UV - VIS SW

30 Gamal A. Hamid

Window management group

Remember the Current Window Layout

Restore the remembered window layout

Arrange the windows as non-overlapping tiles

Cascade the Windows

Load a Window arrangement from disk. (Window arrangements can be saved via the

Save/Window Layout command in the File Menu)

Page 31: UV - VIS SW

31 Gamal A. Hamid

Window group

Restores the Method window

Restores the Graph window

Restores the Results window

Restores the Calibration window

Restores the Kinetics window

Restores the Data Store

Restores the Report Composer window

Page 32: UV - VIS SW

32 Gamal A. Hamid

Graph group

Cancels Tracking

Sets tracking

Clears the graph, without clearing the axes (Use the Graph Menu to clear axes)

Steps through grid displays (dots, lines, no grid)

Sets up Axes

Toggles display of the legend

Toggles display of selected labels

Toggles Graph background between white and gray

Toggles Graph Line Type between Solid and Points

Page 33: UV - VIS SW

Data Store

33

Page 34: UV - VIS SW

34 Gamal A. Hamid

Data Store

The Data Store lists all of the batches of

results that are currently held in memory.

When a batch of results has been run it will

appear in the Data Store.

When a results file is loaded from disk it

will be placed in the Data Store.

A batch is represented by

A sample is represented by

A cycle is represented by

Page 35: UV - VIS SW

Status Bar

35

Page 36: UV - VIS SW

36 Gamal A. Hamid

Status Bar

The following will be displayed on the status bar when appropriate:

The effect of a left mouse-click at the current cursor position (e.g., reminders of the action of toolbar

icons)

Icons for currently active accessories (e.g., sipper, Cell Changer, temperature control, stirrer)

The status of active accessories (e.g., current cell, cell temperature, equilibration status)

The presence of any Performance Verification accessory

Lamp status

The current absorbance reading

The current wavelength

The status of the instrument

Zero or user Baseline status

Page 37: UV - VIS SW

Reporting

37

Page 38: UV - VIS SW

38 Gamal A. Hamid

Report The Report Menu only appears on the Main Menu Bar when the Report Composer

window has the focus.

Click on the item you require to select it, Move the cursor into the Report window.

Page 39: UV - VIS SW

Applications

39

Page 40: UV - VIS SW

40 Gamal A. Hamid

Applications

Scan

Fixed

Quant

Rate

MCA

DNA

Page 41: UV - VIS SW

41

1.Scan

Spectrum recording.

The spectrophotometer measures the amount

of light that passes through the sample over a

continuous range of wavelengths and plots a

graph of the resulting spectrum.

The starting and finishing wavelengths of the

scan are both set .

Select the appropriate Baseline Correction you

wish to perform.

Maximum wavelength range is determined by

the working range of the spectrophotometer.

Page 42: UV - VIS SW

42

Scan Parameters

Page 43: UV - VIS SW

43

Definitions No. Item Definition

1 Data Mode The Data Mode enables you to select what will be used as the vertical axis when the spectrum is first plotted.

2 Bandwidth The Effective Spectral Slit width of the monochromator.

3 Scan speed A measure of how fast the monochromator changes the wavelength of the light falling on the sample.

4 Data interval The Data Interval is the difference in wavelength between successive measurement points.

5 Lamp changeControl an instrument with a deuterium lamp installed to cover the UV wavelengths and a tungsten lamp to cover the visible range.It will not be available if the instrument has a xenon lamp installed,

6 Cycle Single measurement on a sample

7 Baseline correction The standard User Baseline Correction selection will be the 100%T .

Page 44: UV - VIS SW

44 Gamal A. Hamid

Display Parameters

Off No results will appear in the Results Table

Manual The Results Table will contain the values of the Y-axis variable at up to 100 specified wavelengths.

Peak PickThe Results Table will contain the wavelengths and Y-axis variable values of up to 100 peaks, valleys or zero-crossings.

Peak Area

The Results Table will report the area of a selected peak.

Peak Height

The Results Table will report the height of a selected peak.

 

( Scan and MCA) methods – display Parameters – results table, enables you to choose from a number of ways of reporting and at run-time

Page 45: UV - VIS SW

45

2. Fixed

Recording spectrometer readings at

different dedicated wavelengths and

performing simple calculations

In the Fixed application the instrument

measures the amount of light passing

through the sample at one or more

wavelengths.

Set the wavelengths at which the

instrument will make measurements.

Page 46: UV - VIS SW

46

Fixed Parameters

Page 47: UV - VIS SW

47

3. Quantitative

Quantifying sample concentrations via

calibration, based on Lambert- Beers law

The Quantification application is used to obtain

quantitative concentration data on a single

component from absorbance measurements.

A valid calibration must be present before

samples can be measured.

Enter the number of standards that will be used.

Once a calibration has been performed it can be

saved as part of the method and will be available

for use next time the method is loaded.

Page 48: UV - VIS SW

48

Quant. Parameters

Page 49: UV - VIS SW

49

4. Rate

The Rate application is used to measure the change

in concentration of a reactant or product with time.

The absorbance of the sample is measured

repeatedly at a fixed wavelength over a period of

time appropriate to the reaction being studied. The

software then offers a variety of methods to

calculate the rate of reaction from the variation of

absorbance with time.

Recording time dependent readings at a fixed

wavelength and is dedicated for measuring rate

curves.

Page 50: UV - VIS SW

50

Rate Parameters

Page 51: UV - VIS SW

51

5. MCA

It provides the ability to perform

simultaneous quantitative analysis on

samples for up to 20 components

Standards and components are scanned,

and each data point of the scan

contributes to the calculation.

A series of Standards, in which the

concentrations of the components are

known, is prepared and scanned.

There must be at least as many standards

as there are components to be analyzed.

Page 52: UV - VIS SW

52

MCA Parameters

Page 53: UV - VIS SW

53

6. DNA

During a DNA experiment, the sample is

heated and Absorbance v Temperature

data are collected at a suitable

wavelength.

Tm, the "melting point" of the DNA, may

be calculated at run-time, or post-run via

the Math menu.

Page 54: UV - VIS SW

54

DNA Parameters

Page 55: UV - VIS SW

Accessories Software

55

Page 56: UV - VIS SW

56

Rapid Mixing Accessory

The Rapid Mixing accessory provides a very low

dead time of 8 ms, giving you access to reactions

that occur up to 1000 times faster than those

visible to manual mixing experiments.

The Rapid Mixing accessory provides a very low

dead time of 8 ms, giving you access to reactions

that occur up to 1000 times faster than those

visible to manual mixing experiments.

The Anaerobic accessory consists of a manifold,

which is mounted over the base of the drive

syringes and is purged with a steady stream of

inert gas.

Page 57: UV - VIS SW

57

Stirrer Parameters

Off The stirrer is off

Continuous The cell is stirred

continuously

Off Before Measurement The stirrer is normally

on but is switch off prior to and during

measurement.

A programmable settling time elapses before the

measurement is made.

Page 58: UV - VIS SW

58

Air-Cooled Peltier

The Air-cooled Peltier Accessory allows the cell to be

thermostatted during measurements. It comprises a

dedicated cell holder and an electronic control device, which

is connected to the PC.

The Air-cooled Peltier accessory delivers reliable temperature

control .

Wider temperature range from 0 to 110 °C

Multiple cell efficiency for higher sample throughput.

Designed for biologically relevant assays that require

temperature control at 25°, 37°, 40° and 50 °C, the Air-cooled

Peltier accessory is ideal for the life science laboratory.

Page 59: UV - VIS SW

59

Temp. Control Parameters

Method off not selected

Tolerance Target Temperature ±

Tolerance

Start on using block or probe

The Equilibration Time permitted

range and start a measurement.

The Equilibrate Using sample or

sample and reference.

Page 60: UV - VIS SW

60

Cell Changer

Using a cell changer, you load the system

with a batch of samples before starting

the measurement.

During the measurement the cell changer

moves the required sample automatically

into the sample beam of the

spectrophotometer.

This type of measurement procedure is

advisable for time consuming analyses,

e.g. rate measurements.

Page 61: UV - VIS SW

61

Cell Changer Parameters

Method off not selected

Cell changer type. When an instrument

with a cell changer installed is connected,

this field will automatically show the

correct type.

Each cell can contain a Sample (green) or a

Reference (red), or not be used.

Page 62: UV - VIS SW

62

Sipper

In sipper operation the sipper pump

transports the sample to the flow cell

located in the sample beam. Then the

measurement starts automatically.

Therefore, there is no need to change and

clean the cells.

This operation mode is recommended for

high sample throughput in routine

analyses.

Page 63: UV - VIS SW

63

Sipper Parameters

Method off not selected

Sipper Mode. Choose between Sip Only, Sip and

Run, Continuous.

Sample Direction. When Waste is selected the

sample is pumped to waste after it has been

measured. Return sample is returned to the

sample container .

Sample Delay.

The Sample time between the end of pumping

the air gap and the start of sample measurement.

The purpose of the Sample Delay is to allow time

for the sample to settle down in the flowcell.

Page 64: UV - VIS SW

Thanks