using statistics in validation -- small molecule … · using statistics in validation -- small...

78
USING STATISTICS IN VALIDATION -- SMALL MOLECULE CASE STUDY Jane Weitzel [email protected]

Upload: dodang

Post on 04-May-2018

227 views

Category:

Documents


0 download

TRANSCRIPT

USING STATISTICS IN

VALIDATION -- SMALL

MOLECULE CASE STUDY

Jane Weitzel

[email protected]

IVT’S LAB WEEK SERIES

Statistics in Validation14-16 JUNE 2016 • PHILADELPHIA

[email protected] 2

Jane Weitzel Biosketch

Jane Weitzel has been working in analytical chemistry for

over 35 years for mining and pharmaceutical companies with

the last 5 years at the director/associate director level. She

is currently a consultant, auditor, and trainer. Jane has

applied Quality Systems and statistical techniques, including

the estimation and use of measurement uncertainty, in a

wide variety of technical and scientific businesses. She has

obtained the American Society for Quality Certification for

both Quality Engineer and Quality Manager.

In 2014 she was pointed to the Chinese National Drug

Reference Standards Committee and attended their

inaugural meeting in Beijing

For the 2015 – 2020 cycle, Jane is a member of the USP

Statistics Expert Committee and Expert Panel on Method

Validation and Verification.

[email protected] 3

Disclaimer

This presentation reflects the speaker’s

perspective on this topic and does not

necessarily represent the views of USP or

any other organization.

[email protected] 4

Based on Approach

http://www.friesenpress.com/book

store/title/119734000004601536

[email protected]

Example Based on Paper

Volume XX Issue 4

[email protected] 6

Authors of the Paper

Jane Weitzel

Robert Forbes

Principal Scientist of Roche Diabetes Care

Formerly with Eli Lilly

Ron Snee

Founder and president of Snee Associates

[email protected] 7

TWO BASIC STATISTICAL

CONCEPTS

Central Limit Theorem

Confidence Interval of the Standard Deviation (SD)

[email protected] 8

Central Limit Theorem

The distribution of an average is normal.

The sampling distribution of sample means

tends to be a normal distribution.

[email protected] 9

Standard Error of the Mean

[email protected] 10

ns /

S and SEM

[email protected] 11

Confidence Interval of the Standard

Deviation (SD) Each time a standard deviation is estimated,

a different value may be obtained

The confidence interval for the standard

deviation can be calculated

A range within which you expect to find the SD

with a certain level of confidence

Width of the range depends upon number of

measurements used to calculate the SD and the

confidence level you choose

Let’s look at an example

[email protected] 12

TMU and MU variability

Replicate Experiment 1 Experiment 2 Experiment 3 Experiment 4 Experiment 5 Experiment 6

1 98.81 100.68 100.51 100.53 98.74 99.15

2 99.21 100.72 100.44 99.81 99.51 100.86

3 100.41 99.17 99.58 100.47 100.38 100.04

4 99.16 99.85 99.12 99.44 100.98 101.27

5 100.90 99.46 100.84 99.49 100.43 99.28

6 99.96 100.85 99.68 99.95 100.42 99.16

7 98.93 99.34 100.65 99.94 98.85 100.18

Standard

Deviation (s) 0.81 0.72 0.65 0.43 0.87 0.85

Minimum s 0.43

Maximum s 0.87

[email protected]

Upper & Lower Factors

Chi Square distribution

[email protected] 14

Degrees

of

freedom

= 0.05 = 0.01 = 0.001

BU BL BU BL BU BL

1 17.79 .358 86.31 .297 844.4 .248

2 4.86 .458 10.70 .388 33.3 .329

3 3.18 .518 5.45 .445 11.6 .382

4 2.57 .519 3.89 .486 6.94 .422

5 2.25 .590 3.18 .518 5.08 .453

10 1.69 .678 2.06 .612 2.69 .549

15 1.51 .724 1.76 .663 2.14 .603

20 1.42 .754 1.61 .697 1.89 .640

Example

N=6, DF=5

SD = 0.21

Confidence Level 95% (α=0.05)

We are 95% confident that if we make another 6

measurements, the SD will be within this range

UL = 0.21 * 2.25 = 0.12

LL = 0.21 * 0.590 = 0.47

[email protected] 15

Examples with Chart

[email protected] 16

Trial Trial Std Dev Low CI High CI

1 1.64 0.97 3.70

2 2.07 1.22 4.66

3 1.03 0.61 2.31

4 1.72 1.01 3.86

5 1.58 0.93 3.55

6 2.11 1.25 4.75

7 1.57 0.92 3.52

8 1.86 1.10 4.18

9 1.53 0.91 3.45

10 2.37 1.40 5.34

0.00

1.00

2.00

3.00

4.00

5.00

6.00

1 2 3 4 5 6 7 8 9 10

Trial Std Dev

Low CI

High CI

LIFECYCLE APPROACH TO

VALIDATION

[email protected] 17

Lifecycle Approach to Validation

Current regulatory guidance around process

validation is moving away from a simple

once-and-done validation, and toward a

continuous process verification approach, al

recent FDA method validation draft guidance

discusses lifecycle approach. FDA, Guidance for Industry Analytical Procedures and Methods

Validation for Drugs and Biologics (Rockville, MD, Draft--February 2014),

online,

www.fda.gov/downloads/drugs/guidancecomplianceregulatoryinformatio

n/guidances/ucm386366.pdf

[email protected] 18

Gage R&R

The use of traditional Gage R&R statistical

analysis and control sample charting, which

are commonly used in manufacturing process

validation and improvement, will also be

illustrated in the development of an analytical

procedure.

Method is analytical procedure (AP)

[email protected] 19

Intended Use can be Linked to

Clinical Requirement - MU

[email protected]

MU

Decision Makers

The lifecycle

approach involves the

decision makers in

establishing the

intended use of the

reportable result.

[email protected] 21

Analytical Target Profile

Analytical Target Profile (ATP) [1], analogous

to the Quality Target Product Profile (QTPP)

in the ICH

ICH, Q8(R2), Pharmaceutical Development – Part

II: Pharmaceutical Development - Annex, Step 4

version (August 2009).

The ATP

defines the requirements for the result of a test method

based on the suitability for use of that result.

[email protected] 22

CQA

The test ensures that the DS is of adequate

potency, so that when used in the DP, the

patient will receive the appropriate dose of

the active ingredient.

In this way, potency is a critical quality

attribute (CQA) of the DS that is linked to a

CQA for the DP.

[email protected] 23

Example Parameters

An example of the potency test for a drug

substance (DS) with specification limits of

98.0% to 102.0%.

Assume the limits are on the as-is basis, and

no correction for water or solvents is

necessary

There are other CQA’s but this example will

focus on potency only.

[email protected] 24

Focus of This Example

Focus on the

precision and

uncertainty performance characteristics of the

potency test method

For Procedure Performance Qualification and

Verification demonstrate the

gage R&R tool, and

experimentally designed intermediate precision

studies

[email protected] 25

Acceptance Criterion for Precision

Set using a Decision Rule (DR)

DR uses acceptable probability of making a

wrong decision

Null Hypothesis

The potency of the sample is within specification

Two types of wrong decisions False Positive or Type I Error or Producer’s Risk or α

False negative or Type II Error or Manufacturer’s Risk or

β

[email protected] 26

False Positive or Type I Error or

Producer’s Risk or α A batch that has

acceptable potency

may test outside the

specification limits

False positive because

it is a positive test for

unacceptable potency

True value is 99.0%

[email protected] 27

Lower Limit 98

Nominal Concentration (Central Value) 99

Upper Limit 102

Measurement Uncertainty 1

% Below Lower Limit Total % Outside Limits % Above Upper Limit

15.87% 16.00% 0.13%

95 97 99 101 103 105

Concentration

LL

UL

False negative or Type II Error or

Manufacturer’s Risk or β A batch that does not

have acceptable

potency tests inside

the specification limits

False negative

because it is a

negative test for

unacceptable potency

True value is 97.0%

[email protected] 28

Lower Limit 98

Nominal Concentration (Central Value) 97

Upper Limit 102

Measurement Uncertainty 1

% Below Lower Limit Total % Outside Limits % Above Upper Limit

84.13% 84.13% 0.00%

95 97 99 101 103 105

Concentration

LL

UL

ATP uses

ATP uses the probability of each of these

errors to define the maximum allowed

measurement variation (uncertainty) for the

reported result.

This translates into an acceptance criterion

for the qualification of the analytical

procedure.

[email protected] 29

Decision Rule

The product will be considered compliant if

the measurement result is within the

acceptance zone.

94 96 98 100 102 104 106

Simple Decision Rule

Upper LimitLower Limit

Specification Zone

Rejection Zone Rejection ZoneAcceptance Zone

[email protected] 30

Type II error rate is negligible

Good Quality DS

Probability for either error

is caused by both

process variability and

measurement uncertainty.

Assume Type II error rate

is negligible because

Drug Substance is high

quality, few impurities at

very low concentrations.

Not Likely

Lower Limit 98

Nominal Concentration (Central Value) 97

Upper Limit 102

Measurement Uncertainty 1

% Below Lower Limit Total % Outside Limits % Above Upper Limit

84.13% 84.13% 0.00%

95 97 99 101 103 105

Concentration

LL

UL

[email protected] 31

Probability of Type I Error

Measurement Uncertainty

Probability of Type I error

comes from

measurement variation.

Measurement uncertainty

(MU)

Hence, the probability of

Type I error determines

the probability that is

acceptable.

Likely True Value In-Spec

[email protected] 32

Lower Limit 98

Nominal Concentration (Central Value) 99

Upper Limit 102

Measurement Uncertainty 1

% Below Lower Limit Total % Outside Limits % Above Upper Limit

15.87% 16.00% 0.13%

95 97 99 101 103 105

Concentration

LL

UL

Specification is Set

The specification is set by regulators, so it

cannot be adjusted.

The only parameter that can be adjusted to

meet the probability of the decision rule is the

Measurement Uncertainty.

[email protected] 33

Measurement Uncertainty

MU Definition from the VIM:

non-negative parameter characterizing the

dispersion of the quantity values being

attributed to a measurand (the quantity being

measured), based on the information used

JCGM, International vocabulary of metrology –

Basic and general concepts and associated terms

(VIM), 3rd ed., JCGM 200:2012(E/F), online,

www.bipm.org/utils/common/documents/jcgm/JC

GM_200_2012.pdf

[email protected] 34

Intermediate Precision

Test will be carried out in a single laboratory

Intermediate Precision is suitable parameter

for precision to estimate the analytical

procedure capability

Must challenge likely and appropriate analytical

procedure performance characteristics that will

impact the precision

[email protected] 35

GAGE R&R

[email protected] 36

Terminology

repeatability will be used to describe short-

term variability (i.e., repeat preparations

within a single run)

intermediate precision to describe long-term

variability within a single lab (i.e.,

preparations analyzed over different

runs/setups)

Gage R and R

Repeatability and Reproducibility

[email protected] 37

Gage R&R

Terminology associated with the use of the

Gage R&R tool does not make the distinction

between within-lab and inter-lab precision

estimates,

S in this example, the term “reproducibility”

means the same thing as “intermediate

precision.”

[email protected] 38

MU Distribution

Assume the MU distribution is normal

basis for this assumption is discussed in

Annex G of the GUM.

Even if the distribution is not normal, the same

approach will be used.

ISO/IEC, Uncertainty of measurement – Part 3:

Guide to the expression of uncertainty in

measurement (GUM:1995), 98-3:2008.

[email protected] 39

What does this look like?

[email protected] 40

Lower Limit 98

Nominal Concentration (Central Value) 100

Upper Limit 102

Measurement Uncertainty 1

% Below Lower Limit Total % Outside Limits % Above Upper Limit

2.28% 4.55% 2.28%

95 97 99 101 103 105

Concentration

LL

UL

Remember,

production

variability is

assumed to be

negligible.

MU is 1.0%, Probability is 5%

Add 5% to DR

The batch of DS will be

considered compliant if

the result is within 98.0 to

102.0%, with a 5%

probability of being

wrong.

This probability is

“theoretical” or ideal.

Simple Decision Rule

[email protected] 41

94 96 98 100 102 104 106

Simple Decision Rule

Upper LimitLower Limit

Specification Zone

Rejection Zone Rejection ZoneAcceptance Zone

MU at 1.0%

Any analytical procedure that produces

results with an MU of 1.0% or less is fit for its

intended use.

The Target Measurement Uncertainty (TMU)

is 1.0%.

TMU is defined in the VIM as the “measurement

uncertainty specified as an upper limit and

decided on the basis of the intended use of the

measurement results” VIM,http://www.bipm.org/en/publications/guides/vim.html

[email protected] 42

Analytical Target Profile

ATP can now be set.

The laboratory qualifies the analytical

procedure for a range larger than 98.0 to

102.0%.

There is only one known impurity, Impurity A,

and its concentration is <0.1%.

[email protected] 43

ATP

The analytical procedure must be able to

quantify the drug substance in the presence

of impurity A, and potential degradation

products, over a range of 80% to 120% of the

nominal concentration with an accuracy and

uncertainty so that the reportable result falls

within ±2% of the true value with at least a

95% probability.

[email protected] 44

Performance Characteristics for the

Analytical Procedure Use the ATP to define performance

characteristics for the analytical procedure

Accuracy, precision, specificity, linearity and

range

As long as their impact on the accuracy and

precision is such that the target measurement

uncertainty can be met, they are acceptable.

[email protected] 45

Advantage of this approach – you can define “Good Enough”

Acceptance Criteria

The target uncertainty is NMT 1.0%. The

combination of bias and uncertainty need to

be such that the probability of not meeting the

specification is NMT 5%.

In order to meet the ATP, the acceptance

criterion for the study is that the overall

variability, the combined uncertainty, must be

NMT 1.0% as a standard uncertainty or

standard deviation.

[email protected] 46

VALIDATION STUDY

Intermediate Precision

[email protected] 47

HPLC Technique is Used

Designed experiments

Analytical procedure is developed and

qualified

Example of intermediate precision study:

[email protected] 48

Analyst Setup Instrument Column Prepa

ration

Potency

(%)

AVG(%)

A 1 HPLC-A A 1 99.58 99.64

2 99.70

A 2 HPLC-B B 1 99.50 99.58

2 99.65

B 3 HPLC-C C 1 99.43 99.28

2 99.13

B 4 HPLC-D D 1 99.69 99.53

2 99.36

C 5 HPLC-E E 1 99.61 99.64

2 99.66

C 6 HPLC-F F 1 99.60 99.62

2 99.64

D 7 HPLC-G G 1 99.55 99.53

2 99.51

D 8 HPLC-H H 1 99.79 99.79

2 99.79

SD 0.17 0.15

80%

UCL

0.23 0.20

[email protected] 49

SD < TMU

The overall standard deviation (SD) obtained

in this study was 0.15%, with an 80% UCL of

0.20%, which met the acceptance criteria of

NMT 1.0% SD

The analytical procedure is fit for its intended

use.

[email protected] 50

INSTALL AP IN QC

LABORATORY

[email protected] 51

Install AP in QC Laboratory

Write procedure performance qualification

protocol

From the perspective of a lifecycle approach,

a key element of the transfer activity is to

show that the precision of the method as

performed in the QCL is sufficient to meet the

required uncertainty as described in the ATP.

Precision is NMT 1.0%RSD

[email protected] 52

Use Gage R&R

Could repeat the intermediate precision

experiment

An alternative is to conduct a Gage R&R

study

Remember in Gage Repeatability a&

Reproducibility, the latter is the same as

intermediate precision

[email protected] 53

Structure of R&R

In a Gage R&R Study

5-10 samples are evaluated by

2-4 analysts (reproducibility) using

2-4 repeat tests (repeatability)

sometimes involving 2-4 test instruments

(reproducibility).

For example if 3 analysts measure each of 10

samples of a product in duplicate the study will

produce 3x10x2=60 test results.

[email protected] 54

Experimental Design

Samples from

five API batches were tested on

four setups (run) by

two analysts using

each of two instruments

each instrument used a unique column for total of two

columns

New mobile phase was prepared for each setup and each

analyst prepared a fresh set of standards for each run.

Samples were prepared and analyzed in

duplicate on each run, producing a total of 5x2x2x2 = 40

Potency(%) test results.

[email protected] 55

[email protected] 56

API Batch Analyst

HPLC/ Column Test

Potency (%)

API Batch Analyst

HPLC Column Test

Potency (%)

1 A 1 1 100.0 3 A 2 1 100.4

1 A 1 2 100.1 3 A 2 2 99.6

1 B 2 1 99.4 3 B 1 1 100.6

1 B 2 2 100.3 3 B 1 2 100.8

1 A 2 1 100.4 4 A 1 1 99.5

1 A 2 2 100.8 4 A 1 2 99.3

1 B 1 1 100.8 4 B 2 1 100.8

1 B 1 2 100.3 4 B 2 2 101.5

2 A 1 1 99.8 4 A 2 1 100.4

2 A 1 2 99.5 4 A 2 2 100.2

2 B 2 1 101.3 4 B 1 1 100.2

2 B 2 2 101.1 4 B 1 2 100.3

2 A 2 1 100.9 5 A 1 1 99.8

2 A 2 2 100.5 5 A 1 2 99.6

2 B 1 1 100.6 5 B 2 1 101.3

2 B 1 2 100.4 5 B 2 2 101.0

3 A 1 1 99.2 5 A 2 1 100.2

3 A 1 2 98.9 5 A 2 2 99.9

3 B 2 1 101.3 5 B 1 1 100.7

3 B 2 2 101.6 5 B 1 2 100.2

[email protected] 57

API Batch Analyst

HPLC/ Column Test

Potency (%)

API Batch Analyst

HPLC Column Test

Potency (%)

1 A 1 1 100.0 3 A 2 1 100.4

1 A 1 2 100.1 3 A 2 2 99.6

1 B 2 1 99.4 3 B 1 1 100.6

1 B 2 2 100.3 3 B 1 2 100.8

1 A 2 1 100.4 4 A 1 1 99.5

1 A 2 2 100.8 4 A 1 2 99.3

1 B 1 1 100.8 4 B 2 1 100.8

1 B 1 2 100.3 4 B 2 2 101.5

2 A 1 1 99.8 4 A 2 1 100.4

2 A 1 2 99.5 4 A 2 2 100.2

2 B 2 1 101.3 4 B 1 1 100.2

2 B 2 2 101.1 4 B 1 2 100.3

2 A 2 1 100.9 5 A 1 1 99.8

2 A 2 2 100.5 5 A 1 2 99.6

2 B 1 1 100.6 5 B 2 1 101.3

2 B 1 2 100.4 5 B 2 2 101.0

3 A 1 1 99.2 5 A 2 1 100.2

3 A 1 2 98.9 5 A 2 2 99.9

3 B 2 1 101.3 5 B 1 1 100.7

3 B 2 2 101.6 5 B 1 2 100.2

Data Organized by Run/Setup (1& 2)

[email protected] 58

Run/Set-up Analyst HPLC/ColumnLot/Sample

API Batch

Replicate

TestPotency(%)

1 A 1 1 1 100

1 A 1 1 2 100.1

1 A 1 2 1 99.8

1 A 1 2 2 99.5

1 A 1 3 1 99.2

1 A 1 3 2 98.9

1 A 1 4 1 99.5

1 A 1 4 2 99.3

1 A 1 5 1 99.8

1 A 1 5 2 99.6

2 B 2 1 1 99.4

2 B 2 1 2 100.3

2 B 2 2 1 101.3

2 B 2 2 2 101.1

2 B 2 3 1 101.3

2 B 2 3 2 101.6

2 B 2 4 1 100.8

2 B 2 4 2 101.5

2 B 2 5 1 101.3

2 B 2 5 2 101

Data Organized by Run/Setup (3&4)

[email protected] 59

Run/Set-up Analyst HPLC/ColumnLot/Sample

API Batch

Replicate

TestPotency(%)

3 A 2 1 1 100.4

3 A 2 1 2 100.8

3 A 2 2 1 100.9

3 A 2 2 2 100.5

3 A 2 3 1 100.4

3 A 2 3 2 99.6

3 A 2 4 1 100.4

3 A 2 4 2 100.2

3 A 2 5 1 100.2

3 A 2 5 2 99.9

4 B 1 1 1 100.8

4 B 1 1 2 100.3

4 B 1 2 1 100.6

4 B 1 2 2 100.4

4 B 1 3 1 100.6

4 B 1 3 2 100.8

4 B 1 4 1 100.2

4 B 1 4 2 100.3

4 B 1 5 1 100.7

4 B 1 5 2 100.2

Variance components analysis

Used Minitab

Estimates repeatability (within set-up

variability)

Estimates reproducibility (setup to setup

variability)

multiple analysts, instruments and columns

were used to obtain more realistic variability

estimates, and could not be separately

evaluated for statistically significant effects

[email protected] 60

Gage R &R Variance Components

results of the variance components analysis

for the potency measurements are

summarized

[email protected] 61

Source of Variation Variance % of Total Std Dev

Repeatability 0.086750 16.0 0.294534

Reproducibility 0.456167 84.0 0.675401

Total Gage R&R 0.542917 100.0 0.736829

Variability Chart showing the results of the

Gage R&R study

[email protected] 62

Intermediate Precision

Reportable value is average of the two

potencies analyzed in a single setup

The standard deviation is 0.71%

With an 80%upper confidence limit of 1.16%

Acceptance criterion is NMT 1.0%

The standard deviation passes but the upper

confidence limit does not

[email protected] 63

Fail Confidence Limit Criterion

If the verification protocol utilized an

acceptance criterion of NMT 1.0% for the

80% UCL of the SD, these results would not

pass the criterion

[email protected] 64

Pool results for all the batches

Make an assumption there are no significant

differences between the batches

Support this assumption with detailed analytical

results from the supplier of the drug substance

The standard deviation becomes 0.63% with

an 80% Upper Confidence Limit of 1.03%

Rounds to 1.0% that meets the acceptance

criterion

[email protected] 65

Impact of Replication

Another way to reduce the standard deviation

is to use replication

Equation to calculate intermediate precision

SD

[email protected] 66

0.71%

√((0.675401)2/1 + (0.294534)2/(1*2)) =

0.71%

[email protected] 67

Source of Variation Variance % of Total Std Dev

Repeatability 0.086750 16.0 0.294534

Reproducibility 0.456167 84.0 0.675401

Total Gage R&R 0.542917 100.0 0.736829

Impact of Replication

Increasing number of replicates within a

run/setup has little impact

Since between setup standard deviation is

larger component of variance, replicate

between setup/runs.

[email protected] 68

Number of Setups Number of Within

Setup Replicates Method Standard Deviation

(Intermediate Precision) 1 1 0.74%

1 2 0.71%

2 1 0.52%

2 2 0.50%

Benefit of Using R&R Study

You understand the variance components

Don’t waste effort by preparing replicates at

the repeatability level (within setup/run).

Waste resources

Have a false sense of assurance that precision is

improved

[email protected] 69

Compare Validation to Transfer

Estimate of the standard deviation obtained

in the QCL (0.63%, 80% UCL = 1.0%,

pooling the batches) was larger than the

standard deviation estimate obtained during

method validation (0.15%, 80% UCL =

0.20%)

Both met acceptance criterion

May be many reasons for this difference

Analysts performing validation may have been

more experienced

[email protected] 70

PRODUCTION & ANALYTICAL

VARIABILITY

[email protected] 71

EXCEL and Variance

[email protected] 72

Enter Analytical Uncertainty and Production Standard Deviaion

Lower Limit 98

Nominal Concentration (Central Value) 100 Analytical Production Pooled

Upper Limit 102 Uncertainty Std Dev Std Dev

0.5 1 1.118034

Pooled Std Dev (MU with Procction Std Dev) 1.118034

% Below Lower Limit Total % Outside Limits % Above Upper Limit

3.68% 7.36% 3.68% Impact of Combined Analytical & Production Variability

Enter values for various scenarios in the table

Analytical Production Pooled % Below % Above Total %

Uncertainty Std Dev Std Dev Limit Limit Outside

Limits

85 90 95 100 105 110 115

Concentration

LL

UL

Equation

Combine standard deviations as their

variances

sc=√(sA2 + sP

2)

Sc combined standard deviation

sA analytical measurement uncertainty

sP production standard deviation

Pooled

Std Dev

=SQRT(Y5^2+Z5^2)

[email protected] 73

Calculate Production Standard

Deviation Rearrange equation to calculate production

standard deviation

√(sc2 - sA

2) = sP

[email protected] 74

Various Measurement Capability

Recommendations The Automotive Industry Action Group

(AIAG) recommends the use of 6s in gage

R&R studies.

The variation that is due to the measuring

system, either as a percent of study variation

or as a percent of tolerance, is less than

10%. According to AIAG guidelines, this is

acceptable.

[email protected] 75

For Pharmaceutical

Such recommendations may or may not work

for pharmaceutical

With knowledge and understanding of

statistics and use of experimental designs

such as Gage R&R, we can determine the

capability of our measurement systems.

[email protected] 76

Conclusion

We can use the lifecycle approach to

analytical procedures to determine

Decision rules

TMU

ATP

We can use designs such as Gage R&R to

study our analytical procedures and

demonstrate they are fit for their intended

purpose.

[email protected] 77

SPREADSHEET

CI of SD

[email protected] 78