Olof Beck Karolinska Institutet Stockholm, Sweden Using LC-MS/MS for exhaled breath drug testing and LC-HRMS for drug screening

Upload: others

Post on 10-May-2020

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

Olof Beck Karolinska Institutet Stockholm, Sweden

Using LC-MS/MS for exhaled breath drug

testing and LC-HRMS for drug screening

Page 2: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

Drug testing in exhaled breath

- A new possibility for toxicology?

2

Page 3: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

First publication 2010

Substances in breath can be detected after sampling on filter

Page 4: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

• Amphetamine • Methamphetamine • MDMA • Methadone • EDDP • THC • Buprenorphine • Cocaine • Benzoyl ecgonine • Hydromorphone • Tramadol • O-desmethyltramadol • 6-Acetylmorphine • Morphine • Codeine

• Diazepam • Oxazepam • Alprazolam • Zolpidem • Nicotine • Cotinine • Methylphenidate • Ritalinic acid • MDPV • Budesonide • Ceftazidime • Piperacillin • Meropenem • Phosholipids • PEth

Detected substances in breath

Page 5: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

Sampling device

5

Page 6: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

Extraction procedure

6

Page 7: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

Example chromatogram

0.0 0.5 1.0 1.5 2.0 2.5 3.0

50%

100%

50%

100%

50%

100%

50%

100%

Breath RT: 1.05

Benzoylecgonine

m/z 290.1/105.0

m/z 290.1/168.1

Cocaine m/z 304.2/105.1

m/z 304.2/182.2

RT: 2.60

Time (min)

Inte

nsity

In

tens

ity

Inte

nsity

In

tens

ity 56 pg/filter

20 pg/filter

Page 8: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

Substance Breath (n=2295)

Oral fluid (n=18095)

Urine (n=35142)

Amphetamine/methamphetamine 0.17 0.11 0.27

Cannabis 0.09 0.34 1.27

Cocaine/benzoylecgonine 0.09 0.03 0.05

Opiates 0.01 0.24 1.10

Benzodiazepines 0.13 0.01 1.55

∑ 0.52 0.73 3.55

Analytical findings in routine WDT application of exhaled breath drug testing as compared with oral fluid and urine during the same time period (15 months).

Page 9: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

Tramadol in breath after 50 mg oral intake

Page 10: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

LC-HRMS

For drug screening

application

10

Page 11: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

11

Using LC-HRMS for urine drug screening Presented at poster 106

• Simple design in order to obtain a system to replace immunoassay screening • Attraction of HRMS to collect universial data

• To apply HRSIM - XIC

• More general screening method with improved sensitivity and accuracy

Page 12: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

LC-MS Thermo Scientific™ Q Exactive™ Resolving power Max 140,000 @ m/z 200

(1.5 scan/sec) Scan speed Max scan rate up to 12

Hz at resolution setting of 17,500 @ m/z 200

Mass accuracy Internal: < 1 ppm RMS External: < 5 ppm RMS

Dynamic range > 5000:1

Polarity switching One full cycle in < 1 sec (one full scan positive mode and one full scan negative mode at resolution setting of 17,500)

Multiplexing capability up to 10 precursors/scan

Page 13: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

Method design Method: Postive ESI Full MS: Mass range 100-650 Da R=70000 Targeted MS/MS (Escaline): 2.6-3.0 R=35000 Total analysis time: 6 min ”Lock” mass Sample preparation: Dilution 1:5 with internal standards Number of analytes: Plant alkaloids and drugs = 20 (Validated and Analyzed since 2014.11) NPS= 122 (Validated) Total number: 142 analytes Reporting limit: Routine: 100 ng/mL Research: appr. 5 ng/mL

Page 14: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

Identification criteria

1. Mass accuracy within ± 2.5 ppm limit (using lock mass) 2. Absolute retentions time difference in the same batch compared to calibrators ≤0.03 min.

Page 15: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

Spectral resolution (R) R = M/ΔM (50%) Isotope A+3 för M3G: 465.1835 Internal standard M3G-d3: 465.1939 R = 465.1835/0.006 = ~ 80 000

465.10 465.12 465.14 465.16 465.18 465.20 465.22 465.24 465.26 465.28 465.30m/z

0

20

40

60

80

1000

20

40

60

80

100

Rela

tive

Abun

danc

e

0

20

40

60

80

100465.1835

465.1939

465.1837

465.1962

465.1874

NL: 1.75E6Karolinska_Opiates140k_16#80-87 RT: 0.76-0.82 AV: 8 SB: 2 4.01 , 4.01 T: FTMS + p ESI Full lock ms [100.00-500.00]

NL: 4.08E6karolinska_opiates_16#165-174 RT: 0.73-0.77 AV: 10 SB: 2 4.00 , 4.00 T: FTMS + p ESI Full lock ms [100.00-500.00]

NL: 2.74E6karolinska_opiates140k_16_111117133039#411-438 RT: 0.95-1.02 AV: 28 SB: 2 4.00 , 4.00 T: FTMS + p ESI Full lock ms [100.00-500.00]

A+3 of M3G D3-M3G

140,000 R for m/z 200

70,000 R for m/z 200

35,000 R for m/z 200

ΔM = 0.006

465.1835

465.1939

465 1837

ΔM = Peak width @50%

(FWHM – Full width half maximum)

Page 16: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

d:\tracefinderdata\...\data\urin_r70000 08/21/15 13:00:59neg amfRT: 0.00 - 6.00 SM: 9G

0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0Time (min)

0

20

40

60

80

100

Rela

tive

Abun

danc

e

NL: 8.34E9TIC F: FTMS + p ESI Full ms [100.00-650.00] MS urin_Amf_r70000

RT: 2.30 - 2.50

2.30 2.32 2.34 2.36 2.38 2.40 2.42 2.44 2.46 2.48Time (min)

0

20

40

60

80

100

Rela

tive

Abun

danc

e

NL: 6.59E9TIC F: FTMS + p ESI Full ms [100.00-650.00] MS Genesis urin_r70000

RT: 0.00 - 6.00

0 1 2 3 4 5 6Time (min)

0

20

40

60

80

100

Rela

tive

Abun

danc

e

2.40

1.88 2.881.78 5.235.02 5.413.60 4.07 4.671.500.51

NL: 4.25E7Base Peak m/z= 136.11139-136.11275 F: FTMS + p ESI Full ms [100.00-650.00] MS urin_Amf_r70000

RT: 2.30 - 2.50

2.30 2.35 2.40 2.45 2.50Time (min)

0

20

40

60

80

100

Rela

tive

Abun

danc

e

NL: 2.67E5Base Peak m/z= 136.11139-136.11275 F: FTMS + p ESI Full ms [100.00-650.00] MS Genesis urin_r70000

RT: 1.00 - 6.00

1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0Time (min)

0

20

40

60

80

100

Rela

tive

Abun

danc

e

NL: 8.15E7Base Peak m/z= 135.61207-136.61207 F: FTMS + p ESI Full ms [100.00-650.00] MS Genesis urin_r70000

urin_r70000 #276 RT: 2.40 AV: 1 SB: 484 0.50-4.30 , 4.60-5.00 NL: 3.07E6T: FTMS + p ESI Full lock ms [100.00-650.00]

136.00 136.05 136.10 136.15m/z

0

20

40

60

80

100

Rela

tive

Abun

danc

e

136.06183

136.02176

136.11222136.03994

RT: 1.00 - 6.00

1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0Time (min)

0

20

40

60

80

100

Rela

tive

Abun

danc

e

Page 17: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

Z:\TraceFinderData\...\Data\Linj_250 02/11/15 00:18:08

2.50 2.55 2.60 2.65 2.70 2.75 2.80 2.85 2.90 2.95 3.00Time (min)

0

50

100

Rela

tive

Abun

danc

e

Page 18: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

Resolution vs. Scanspeed (Full MS)

0

20000

40000

60000

80000

100000

120000

140000

160000

0 2 4 6 8 10 12 14

Q-Exactive

Resolution (m/z 200) Max. scan speed (Hz) 17.500 12 35.000 7 70.000 3

140.000 1.5

Scan speed (scan/s)

Res

olut

ion

Resolution Aquisition time 17500 64 ms 35000 128 ms 70000 256 ms 140000 512 ms

Page 19: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

Orbitrap_ULMVXT_201508110705_1 08/11/15 15:54:24 K_1000

RT: 2.60 - 3.60

2.6 2.7 2.8 2.9 3.0 3.1 3.2 3.3 3.4 3.5 3.6Time (min)

0

10

20

30

40

50

60

70

80

90

100

Rela

tive

Abun

danc

e

RT: 3.13AA: 823311996

NL: 3.02E8Base Peak m/z= 215.11589-215.11989 F: FTMS + p ESI Full ms [100.00-650.00] MS Genesis Orbitrap_ULMVXT_201508110705_1

RT: 2.60 - 3.60

2.6 2.7 2.8 2.9 3.0 3.1 3.2 3.3 3.4 3.5 3.6Time (min)

0

10

20

30

40

50

60

70

80

90

100

Rela

tive

Abun

danc

e

RT: 3.14MA: 288007774

NL: 2.54E8Base Peak m/z= 215.11779-215.11799 F: FTMS + p ESI Full ms [100.00-650.00] MS Orbitrap_ULMVXT_201508110705_1

Harmaline: m/z=215.11789 Extracted: ± 2 mDa m/z: 215.11589 – 215.11989

Harmaline: m/z=215.11789 Extracted: ± 0.1 mDa m/z: 215.11779 – 215.11799

EIC, Mass accuracy (R=70000) Pitfalls

Page 20: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

False negative results: Mass accuracy fail within a batch (Hardware problems)

orbitrap_ulmvxt_201501220653_1 01/22/15 14:31:31 1000

RT: 0.00 - 6.01

0.0 0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0Time (min)

0

20

40

60

80

100

0

20

40

60

80

100

Relat

ive A

bund

ance

RT: 3.74AA: 1790144980

NL: 7.80E4Base Peak m/z= 272.19817-272.20361 F: FTMS + p ESI Full ms [100.00-650.00] MS OM_Orbitrap_ULMVXT_201503251443_1

NL: 5.51E8Base Peak m/z= 272.19817-272.20361 F: FTMS + p ESI Full ms [100.00-650.00] MS ICIS orbitrap_ulmvxt_201501220653_1

Mass accuracy for dextromethorphane is outside ±10 ppm

False neg. sample

The same sample reinjected: Mass accuracy is within ±10 ppm

Pitfalls

Page 21: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

False negative results: Integration fail (Software buggs)

Psilocyn is not integrated in the sample view – false negative But integrated in the method parameter settings

Pitfalls

Page 22: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

False positive/negative results within a low concentration levels integration difficulties

Spike False positve findings (integration difficulties)

Pitfalls

Page 23: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

Selectivity and need for reporting limit

Low concentrations? – False positive? – Reporting limit adjustment?? Sample is positive for desmethyltramadol. Background peaks with respons below limit for other analytes

Page 24: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

QC, 50 ng/mL (Values during one month period – April 2015)

46,1 47,7 1,6 4,3 3,5 9,0 35 35

Page 25: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

Method comparison

25

LC-HRMS

LC-MS/MS

Neg Pos

Neg 57 2*

Pos 0 153

*O-desmethyltramadol, interference in LC-MS/MS qualifier

Page 26: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

Results from routine application

Plant alkaloids and drugs (20 compounds) Start: Nov-2014 (10 månader): Number of samples: ~ 500 samples/month = 5000 Positive rate ~ 17% Pregabaline ~ 12% Dextromethorphane ~ 3% Ketamine ~ 2% Hydromorphone Hydrocodone Mitragynine Pethidine Dihydrocodeine Psilocin Atropine

Page 27: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

Maintenance

Clean: S-lens and ion source once/1-1.5 months (after 700 – 1500 samples)

Mass calibration Every second day

Clean: sweep cone and Ion transfer tube once/1-2 week

Page 28: Using LC-MS/MS for Exhaled Breath Drug Testing and LC-HRMS ...tools.thermofisher.com/content/sfs/brochures/WS-LC... · Method design Method: Postive ESI Full MS: Mass range 100-650

Final aspects

We see a great potential of LC-HRMS

Further development and validation of system for routine application in drug testing

• Reliable and automated data evaluation • Improved summary report forms • Electronic transfer of results to LIMS