using dicom for advanced technology clinical trialsusing dicom for advanced technology clinical...
TRANSCRIPT
![Page 1: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/1.jpg)
DICOM: Definitions and Testing
Using DICOM for Advanced Technology Clinical Trials
Walter R. Bosch, D.Sc.
![Page 2: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/2.jpg)
Bosch – AAPM 2004 2
NCI Advanced Technology QA Consortium
• Members• Image-guided Therapy Center (ITC)
• J. Purdy, W. Bosch, J. Matthews, Wm. Straube, J. Michalski• Resource Center for Emerging Technologies (RCET)
• J. Palta, V. Frouhar, J. Dempsey• Radiological Physics Center (RPC)
• G. Ibbott, D. Followill• Radiation Therapy Oncology Group (RTOG)
• E. Martin, R. Lustig• Quality Assurance Review Center (QARC)
• T.J. FitzGerald, M. Urie, K. Ulin
• Support• NIH U24 grant CA81647, “Advanced Technology QA Center”
![Page 3: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/3.jpg)
Bosch – AAPM 2004 3
Advanced Technology Clinical Trials• Collect digital treatment planning
data to link with clinical outcomes • Evaluate response statistics• Develop dose-response models
• Treatment planning based on volumetric model of patient anatomy and 3-D dose distributions
• Maintain temporal pattern of dose delivery
![Page 4: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/4.jpg)
Bosch – AAPM 2004 4
Data for Advanced-Technology Clinical Trials in Radiotherapy
• Primary data (patient model, dosimetry)• Patient Volumetric Image – CT Image• Structures: GTV, CTV, PTV, OAR – RT
Structure Set• 3-D Dose Distribution (including
fractionation information) – RT Dose
• Secondary data (QA of primary data)• Treatment Plan Specification – RT Plan• Dose-Volume Histograms – RT Dose• Digital Simulator and Portal Images – RT
Image
![Page 5: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/5.jpg)
Bosch – AAPM 2004 5
Special Requirements for Attributes• The ATC DICOM File Set Reader Conformance
Statement* is a detailed list of the attributes required for each DICOM object• Attributes whose values must be specified for clinical trials
submissions (including DICOM Type 2 or 3 attributes)• Special requirements for attributes values
• Reference Guide to the ATC DICOM File Set Reader Conformance Statement* summarizes the conformance statement and highlights special requirements.
*Available online at the ATC web site: http://atc.wustl.edu
![Page 6: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/6.jpg)
Bosch – AAPM 2004 6
ATC Requirements: DICOM RT Structure Set
• Structure contours must be planar and defined in transverse planes whose axial positions coincide with those of images.
• For external beam protocols, an external patient contour (“skin”) is required on all axial positions.
• Structures must reflect the extent of volumes contoured by the physician, e.g., target volumes must not interfere with overlapping, contoured structures.
• Structures are to be submitted with protocol-defined structure names (see http://atc.wustl.edu ).
![Page 7: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/7.jpg)
Bosch – AAPM 2004 7
ATC Requirements: DICOM RT Dose• Dose distributions are to be represented as a single, multi-
frame object (rather than multiple, planar doses).• A separate dose array is required for each fraction group
(set of beams or brachytherapy sources which are delivered together, in constant proportion).
• Doses must be specified as absolute, physical dose (in Gy).• Dose array must represent all meaningful dose (outside the
array, dose is assumed to be zero).• A DVH is to be submitted for total dose plan only. (DVHs
may be submitted graphically.)
![Page 8: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/8.jpg)
Bosch – AAPM 2004 8
ATC Requirements: DICOM RT Plan• RT Plan Geometry (300A,000C) must be PATIENT• External Beam 3DCRT – beam energy, aperture, beam
modifiers are used for setup and beam weight QA.• External Beam IMRT – beam information is not currently
used for QA.• Permanent Implant Brachytherapy
• Seed model and strength at implant• Seed locations in patient space
• HDR Brachytherapy• Source geometry and dwell positions in patient space• Source strength• Dwell times
![Page 9: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/9.jpg)
Bosch – AAPM 2004 9
ATC Requirements: DICOM RT Image
• Requirements are protocol-specific• RT Image objects can be used for
• DRRs• Digitized simulator images/films• Digitized port images/films
• RT Image is preferred to Secondary Capture or Computed Radiography when relationship to beam is known
![Page 10: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/10.jpg)
Bosch – AAPM 2004 10
ATC Compliant Treatment Planning Systems
• ATC compliant treatment planning systems are listed on the ATC website http://atc.wustl.edu.
• Represents• Complete, protocol-
compliant submissions
• Clinical users • Released software
8/8/038/11/03
10/18/03
3/15/04
![Page 11: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/11.jpg)
Bosch – AAPM 2004 11
Progress over the past 12 months*
• Four treatment planning systems have ATC compliant DICOM export capabilities• Eight additional TPS are “works in progress”
• 40 institutions have been credentialed using DICOM submissions • Some for multiple protocols• Some with multiple TPS
• 194+ DICOM data submission to ITC• 123 “dry run” submissions• 50 protocol case data sets (RTOG protocols)• 21 phantom data sets (evaluated by RPC)
* as of July 2004
![Page 12: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/12.jpg)
Bosch – AAPM 2004 12
ATC/AAPM/DICOM WG7 Demonstration Phase 1: DICOM Export (2004)1. Two test data sets (DICOM CT Image Series and RT
Structure Set object, planning instructions) were distributed to TPS manufacturers through DICOM WG7:a. ATC04PR02 – 3DCRT Prostateb. ATC04HN02 – IMRT Head/Neck
2. Instructions given at ATC DICOM Workshop and via emaila. Import images, structuresb. Create plansc. Export CT images, RT Structure Set; RT Plan, RT Dose, RT Imaged. Submit to ITC by June 25, 2004
3. Display and compare submitted data at AAPM• ATC web-based review tools• Compare submitted data to vendors’ conformance statements
![Page 13: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/13.jpg)
Bosch – AAPM 2004 13
Phase 1 Test Data Sets• Two test data sets were created as a starting point for
treatment planning:1. 3DCRT Prostate, based on RTOG 0126 – ATC04PR022. IMRT Nasopharynx, based on RTOG 0225 – ATC04HN02
• Data were de-identified and attempts were made to facilitate import into treatment planning systems:• Private attributes were removed from DICOM objects.• Structure Sets contour Z positions coincided with those of CT
slices.• However, one of the data sets had irregularly spaced CT slices.
• John Matthews at ITC received, imported, and evaluated data sets and gave feedback to participants.
![Page 14: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/14.jpg)
Bosch – AAPM 2004 14
Case 1 (ATC04PR02) –3DCRT Prostate
• Based on RTOG 0126• 97 CT slices• 10 Structures:
• PTV2 (high-dose)• PTV1 (low-dose)• Prostate/SV [CTV1]• CTV2• Bladder• Rectum• RT Femur• Lt Femur• Penile bulb• Skin
![Page 15: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/15.jpg)
Bosch – AAPM 2004 15
Case 1 (ATC04PR02) 3DCRT Treatment Plan Summary
• Prescription (ICRU Ref Dose)PTV1: 57.97 GyPTV2: 82.28 Gy (44 fractions)
• 2 fraction groups (sub-plans) with 6 beams each:• RPO, Rlat, RAO, LAO, Llat, LPO
50%35%25%15%
60 Gy65 Gy70 Gy75 GyRectum
65 Gy70 Gy75 Gy80 GyBladder
Maximum Dose to Percent of Volume
![Page 16: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/16.jpg)
Bosch – AAPM 2004 16
DICOM Objects Submitted to ATC• Eight manufacturers submitted
DICOM data sets.• Data from nine TPS included
• CT Image• RT Structure Set• RT Plan• RT Dose (3D dose array).
• Partial data sets were submitted from three other systems.
• See ATC Booth #1634 for a display of data sets submitted by demonstration participants.
![Page 17: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/17.jpg)
Bosch – AAPM 2004 17
ATC Review Tools
Remote Review Tool (RRT)Image-guided Therapy QA Center
Computational Environment for Radiotherapy Research (CERR)James Alaly, Emiliano Spezi, Joseph Deasy
![Page 18: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/18.jpg)
Bosch – AAPM 2004 18
ATC/AAPM/DICOM WG7 Demonstration Phase 2: DICOM Import (2005)1. Use phase 1 data to create a library of
manufacturers’ DICOM objects for off-line testing:a. From each TPS: CT Image, RT Structure Set, RT Plan,
RT Dose, RT Imageb. ATC web-based review tools provide displays of what
data should look like.2. Demonstrate exchange using …
a. Manufacturer-to-manufacturer network exchange,b. Central Test Node, orc. Media exchange.
![Page 19: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures](https://reader034.vdocuments.us/reader034/viewer/2022042016/5e74e191228bf4677d7f7b95/html5/thumbnails/19.jpg)
Bosch – AAPM 2004 19
Summary• In the past 12 months, DICOM has become a viable format
for the submission of treatment planning data in advanced-technology clinical trials.
• The ATC DICOM Conformance Statement provides a uniform “profile” for use of RT objects in clinical trials.
• The Image-Guided Therapy QA Center (an ATC member) provides assistance to vendors in their development of DICOM export capabilities through data set evaluation and web-based review tools.
• The ATC/AAPM/DICOM WG7 Demonstration appears to have accelerated the process of vendor implementation of DICOM export capabilities for advanced-technology clinical trials.
• Data collected this year should prove valuable for a demonstration of import capabilities in 2005.