Page 1: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

DICOM: Definitions and Testing

Using DICOM for Advanced Technology Clinical Trials

Walter R. Bosch, D.Sc.

Page 2: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

Bosch – AAPM 2004 2

NCI Advanced Technology QA Consortium

• Members• Image-guided Therapy Center (ITC)

• J. Purdy, W. Bosch, J. Matthews, Wm. Straube, J. Michalski• Resource Center for Emerging Technologies (RCET)

• J. Palta, V. Frouhar, J. Dempsey• Radiological Physics Center (RPC)

• G. Ibbott, D. Followill• Radiation Therapy Oncology Group (RTOG)

• E. Martin, R. Lustig• Quality Assurance Review Center (QARC)

• T.J. FitzGerald, M. Urie, K. Ulin

• Support• NIH U24 grant CA81647, “Advanced Technology QA Center”

Page 3: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

Bosch – AAPM 2004 3

Advanced Technology Clinical Trials• Collect digital treatment planning

data to link with clinical outcomes • Evaluate response statistics• Develop dose-response models

• Treatment planning based on volumetric model of patient anatomy and 3-D dose distributions

• Maintain temporal pattern of dose delivery

Page 4: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

Bosch – AAPM 2004 4

Data for Advanced-Technology Clinical Trials in Radiotherapy

• Primary data (patient model, dosimetry)• Patient Volumetric Image – CT Image• Structures: GTV, CTV, PTV, OAR – RT

Structure Set• 3-D Dose Distribution (including

fractionation information) – RT Dose

• Secondary data (QA of primary data)• Treatment Plan Specification – RT Plan• Dose-Volume Histograms – RT Dose• Digital Simulator and Portal Images – RT

Image

Page 5: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

Bosch – AAPM 2004 5

Special Requirements for Attributes• The ATC DICOM File Set Reader Conformance

Statement* is a detailed list of the attributes required for each DICOM object• Attributes whose values must be specified for clinical trials

submissions (including DICOM Type 2 or 3 attributes)• Special requirements for attributes values

• Reference Guide to the ATC DICOM File Set Reader Conformance Statement* summarizes the conformance statement and highlights special requirements.

*Available online at the ATC web site: http://atc.wustl.edu

Page 6: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

Bosch – AAPM 2004 6

ATC Requirements: DICOM RT Structure Set

• Structure contours must be planar and defined in transverse planes whose axial positions coincide with those of images.

• For external beam protocols, an external patient contour (“skin”) is required on all axial positions.

• Structures must reflect the extent of volumes contoured by the physician, e.g., target volumes must not interfere with overlapping, contoured structures.

• Structures are to be submitted with protocol-defined structure names (see http://atc.wustl.edu ).

Page 7: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

Bosch – AAPM 2004 7

ATC Requirements: DICOM RT Dose• Dose distributions are to be represented as a single, multi-

frame object (rather than multiple, planar doses).• A separate dose array is required for each fraction group

(set of beams or brachytherapy sources which are delivered together, in constant proportion).

• Doses must be specified as absolute, physical dose (in Gy).• Dose array must represent all meaningful dose (outside the

array, dose is assumed to be zero).• A DVH is to be submitted for total dose plan only. (DVHs

may be submitted graphically.)

Page 8: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

Bosch – AAPM 2004 8

ATC Requirements: DICOM RT Plan• RT Plan Geometry (300A,000C) must be PATIENT• External Beam 3DCRT – beam energy, aperture, beam

modifiers are used for setup and beam weight QA.• External Beam IMRT – beam information is not currently

used for QA.• Permanent Implant Brachytherapy

• Seed model and strength at implant• Seed locations in patient space

• HDR Brachytherapy• Source geometry and dwell positions in patient space• Source strength• Dwell times

Page 9: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

Bosch – AAPM 2004 9

ATC Requirements: DICOM RT Image

• Requirements are protocol-specific• RT Image objects can be used for

• DRRs• Digitized simulator images/films• Digitized port images/films

• RT Image is preferred to Secondary Capture or Computed Radiography when relationship to beam is known

Page 10: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

Bosch – AAPM 2004 10

ATC Compliant Treatment Planning Systems

• ATC compliant treatment planning systems are listed on the ATC website http://atc.wustl.edu.

• Represents• Complete, protocol-

compliant submissions

• Clinical users • Released software

8/8/038/11/03

10/18/03

3/15/04

Page 11: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

Bosch – AAPM 2004 11

Progress over the past 12 months*

• Four treatment planning systems have ATC compliant DICOM export capabilities• Eight additional TPS are “works in progress”

• 40 institutions have been credentialed using DICOM submissions • Some for multiple protocols• Some with multiple TPS

• 194+ DICOM data submission to ITC• 123 “dry run” submissions• 50 protocol case data sets (RTOG protocols)• 21 phantom data sets (evaluated by RPC)

* as of July 2004

Page 12: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

Bosch – AAPM 2004 12

ATC/AAPM/DICOM WG7 Demonstration Phase 1: DICOM Export (2004)1. Two test data sets (DICOM CT Image Series and RT

Structure Set object, planning instructions) were distributed to TPS manufacturers through DICOM WG7:a. ATC04PR02 – 3DCRT Prostateb. ATC04HN02 – IMRT Head/Neck

2. Instructions given at ATC DICOM Workshop and via emaila. Import images, structuresb. Create plansc. Export CT images, RT Structure Set; RT Plan, RT Dose, RT Imaged. Submit to ITC by June 25, 2004

3. Display and compare submitted data at AAPM• ATC web-based review tools• Compare submitted data to vendors’ conformance statements

Page 13: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

Bosch – AAPM 2004 13

Phase 1 Test Data Sets• Two test data sets were created as a starting point for

treatment planning:1. 3DCRT Prostate, based on RTOG 0126 – ATC04PR022. IMRT Nasopharynx, based on RTOG 0225 – ATC04HN02

• Data were de-identified and attempts were made to facilitate import into treatment planning systems:• Private attributes were removed from DICOM objects.• Structure Sets contour Z positions coincided with those of CT

slices.• However, one of the data sets had irregularly spaced CT slices.

• John Matthews at ITC received, imported, and evaluated data sets and gave feedback to participants.

Page 14: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

Bosch – AAPM 2004 14

Case 1 (ATC04PR02) –3DCRT Prostate

• Based on RTOG 0126• 97 CT slices• 10 Structures:

• PTV2 (high-dose)• PTV1 (low-dose)• Prostate/SV [CTV1]• CTV2• Bladder• Rectum• RT Femur• Lt Femur• Penile bulb• Skin

Page 15: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

Bosch – AAPM 2004 15

Case 1 (ATC04PR02) 3DCRT Treatment Plan Summary

• Prescription (ICRU Ref Dose)PTV1: 57.97 GyPTV2: 82.28 Gy (44 fractions)

• 2 fraction groups (sub-plans) with 6 beams each:• RPO, Rlat, RAO, LAO, Llat, LPO

50%35%25%15%

60 Gy65 Gy70 Gy75 GyRectum

65 Gy70 Gy75 Gy80 GyBladder

Maximum Dose to Percent of Volume

Page 16: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

Bosch – AAPM 2004 16

DICOM Objects Submitted to ATC• Eight manufacturers submitted

DICOM data sets.• Data from nine TPS included

• CT Image• RT Structure Set• RT Plan• RT Dose (3D dose array).

• Partial data sets were submitted from three other systems.

• See ATC Booth #1634 for a display of data sets submitted by demonstration participants.

Page 17: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

Bosch – AAPM 2004 17

ATC Review Tools

Remote Review Tool (RRT)Image-guided Therapy QA Center

Computational Environment for Radiotherapy Research (CERR)James Alaly, Emiliano Spezi, Joseph Deasy

Page 18: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

Bosch – AAPM 2004 18

ATC/AAPM/DICOM WG7 Demonstration Phase 2: DICOM Import (2005)1. Use phase 1 data to create a library of

manufacturers’ DICOM objects for off-line testing:a. From each TPS: CT Image, RT Structure Set, RT Plan,

RT Dose, RT Imageb. ATC web-based review tools provide displays of what

data should look like.2. Demonstrate exchange using …

a. Manufacturer-to-manufacturer network exchange,b. Central Test Node, orc. Media exchange.

Page 19: Using DICOM for Advanced Technology Clinical TrialsUsing DICOM for Advanced Technology Clinical Trials Walter R. Bosch, D.Sc. Bosch – AAPM 2004 2 ... overlapping, contoured structures

Bosch – AAPM 2004 19

Summary• In the past 12 months, DICOM has become a viable format

for the submission of treatment planning data in advanced-technology clinical trials.

• The ATC DICOM Conformance Statement provides a uniform “profile” for use of RT objects in clinical trials.

• The Image-Guided Therapy QA Center (an ATC member) provides assistance to vendors in their development of DICOM export capabilities through data set evaluation and web-based review tools.

• The ATC/AAPM/DICOM WG7 Demonstration appears to have accelerated the process of vendor implementation of DICOM export capabilities for advanced-technology clinical trials.

• Data collected this year should prove valuable for a demonstration of import capabilities in 2005.