Page 1: USER' S MANUAL ED-400 - Shenyang Tianxing Testing ... S MANUAL ED-400 COATING THICKNESS GAUGE FOR NON-FERROUS METALS TX TESTING INSTRUMENTS CO., LTD. ... All characters Probe orcircuit

TX Testing Instruments Co., Ltd.

35 Changqing Street, S h e n h e District, Shenyang 110015 China Tel: + 8 6 - 2 4 - 2 4 2 3 8 6 6 8 Fax: + 8 6 - 2 4 - 2 4 2 3 0 0 0 8 http: //www.tianxing.com.cn E-mail: [email protected]

U S E R ' S M A N U A L

ED-400

COATING THICKNESS G A U G E

FOR N O N - F E R R O U S M E T A L S

TX TESTING INSTRUMENTS CO., LTD.

Page 2: USER' S MANUAL ED-400 - Shenyang Tianxing Testing ... S MANUAL ED-400 COATING THICKNESS GAUGE FOR NON-FERROUS METALS TX TESTING INSTRUMENTS CO., LTD. ... All characters Probe orcircuit

R E A D T H E U S E R ’ S M A N U A L B E F O R E U S I N G T H E G A U G E . Contents

1. A P P L I C A T I O N

2 . T E C H N I C A L P A R A M E T E R S

3 . S T A N D A R D P A C K A G E

4 . O P E R A T I O N

5. C A L I B R A T I O N

6 . A D D I T I O N A L F U N C T I O N S -

7. N O T I C E

8 . E R R O R F A C T O R S

9 . T R O U B L E S H O O T I N G

10. D A I L Y M A I N T E N A N C E • • • • •

Page 3: USER' S MANUAL ED-400 - Shenyang Tianxing Testing ... S MANUAL ED-400 COATING THICKNESS GAUGE FOR NON-FERROUS METALS TX TESTING INSTRUMENTS CO., LTD. ... All characters Probe orcircuit

E D 4 0 0 C O A T I N G T H I C K N E S S G A U G E F O R N O N - F E R R O U S M E T A L S

1. A P P L I C A T I O N

The E D - 4 0 0 coating thickness gauge is suitable for

measuring the thickness of the non-conductive coating

on non-ferrous metal. The typical application is to

measure the thickness of the anodic oxide coating of the

aluminium alloy. It can also be used to measure the

thickness of plastic foil and paper.

The E D - 4 0 0 coating thickness gauge conforms to

international standard ISO 2360-2003.

2 . T E C H N I C A L P A R A M E T E R S

Measuring range:

Accuracy:

Resolution :

Operating temperature:

Power supply:

Power consumption:

Dimonsio 门 s:

0-500 m m ;

0-50 m m: ± 1 |j m;

50-500 m m: ± 2 %;

0-50 |J m : 0.1 |J m;

50-500 m m: 1 m m;

0-500 m m: 1 m m; (Optional)

5-45°C;

9V battery

80mw

150mm x 80mm x 30mm

Weight: 260g

ED400 COATING THICKNESS GAUGEFOR N O N - F E R R O U S METALS

3. STANDARD PACKAGE

• G a u g e

• Probe

• Plastic cover

• Standard substrate (6063 aluminum alloy)

• Calibration foil with inspection report

• User,s manual

• Certificate of conformity

• Certificate of warranty

• Carrying case

Spare Parts

• Spare probe

• Standard substrate (6063 aluminum alloy)

• Calibration foil with inspection report

4. O P E R A T I O N

Press the " O N / O F F " button and wait until "0" or

"0.0" is shown on the display.

4.1 M e a s u r e m e n t

Hold the plastic part of the probe and place it vertically

onto the clean and dry workpiece surface. Hold it steadily

and wait for the measurement to appear in the L C D and a

beep is heard (unless the beeping is disabled). Lift the

probe for at least 10 mm from the workpiece and for at

least two seconds.

Page 4: USER' S MANUAL ED-400 - Shenyang Tianxing Testing ... S MANUAL ED-400 COATING THICKNESS GAUGE FOR NON-FERROUS METALS TX TESTING INSTRUMENTS CO., LTD. ... All characters Probe orcircuit

E D 4 0 0 C O A T I N G T H I C K N E S S G A U G E F O R N O N - F E R R O U S M E T A L S

4.2 Statistics

Press the “ S T A T " button after serveral

measurements, and the following data will show in turn:

M E A N - average value

M a x 一 m a x i m u m value

MIN - minimum value

S - Standard deviation

N - Number of measurements

4.3 Delete

During measurement, if an obvious error value

appears, it can be deleted by pressing the “ DE L "

button.

5. C A L I B R A T I O N

The gauge should be calibrated under the following

circumstances: it has not been used or calibrated for a

long period of time; it is found obviously inaccurate; it is

reseted; or the probe is changed.

In calibration mode, each measurement should be

m a d e in the s a m e area. If an erroneous measurement

result appears, it can be deleted by pressing the "DEL"

button. Press once to delete the last result, and press

twice to delete all measuring results in the current

calibration sequence.

5.1 Single point calibration

E D 4 0 0 C O A T I N G T H I C K N E S S G A U G E F O R N O N - F E R R O U S M E T A L S

The standard substrate or an uncoated workpiece is

used in single point calibration, and the existing zero point

will be adjusted.

5.1.1 Press the " C A L " button, then the gauge enters

calibration m o d e . The L C D will display " Z E R O " and

" 0 . 0 " .

512 Measure on the uncoated substrate for at least 10

times. The L C D will display the average measuring

result.

5.1.3 Press the "CAL" button twice to complete the

calibration. T h e L C D will display “ ” and then

“0” or “0.0”,then the gauge is back to measuring

Mode.

5.2 Dual point calibration

The calibration foil and the standard substrate or an

uncoated workpiece are used in single point calibration,

and the full measuring range will be adjusted.

5.2.1 Press the " C A L " button, then the gauge enters

calibration m o d e . T h e L C D will display " Z E R O "

and “0.0” .

5.2.2Measure on the uncoated substrate for at least 10

times. The L C D will display the average measuring

result.

5.2.3Press the " C A L " button, then the instrument will

come to the next calibration stage. The L C D will display

"STD1" and "0.0".

Page 5: USER' S MANUAL ED-400 - Shenyang Tianxing Testing ... S MANUAL ED-400 COATING THICKNESS GAUGE FOR NON-FERROUS METALS TX TESTING INSTRUMENTS CO., LTD. ... All characters Probe orcircuit

E D 4 0 0 C O A T I N G T H I C K N E S S G A U G E F O R N O N - F E R R O U S M E T A L S

5.2.4Put the foil onto the u n c o a t e d substrate then

m e a s u r e it above ten times. T h e L C D will display

the average measuring result.

5.2.5 Press the “ •“ or “•” button to adjust thedisplay

value from 50.0 to the foil thickness marked.

5.2.6 Press the " C A L " button to finish the calibration.

After “一一”,LCD will display “0” or "0.0" . Then

you can m a k e the measurement.

Press the "CAL" button to complete the calibration.

T h e L C D will display “ ” and then "0” or

"0.0",then the gauge is back to measuring m o d e .

6 . A D D I T I O N A L F U N C T I O N S

6.1 Button combinations

By pressing two buttons at the s a m e time, the

additional functions shown in Table 1 can be used. The

L C D will display “一一一” and then “0” or “0.0”,then

the corresponding function is set.

Table 1

Combination button Function description

D E L + S T A T Reset (see Section 6.2)

S T A T + " A " Activate beeping

S T A T + “ •“ Inactivate beeping

C A L + " • ” S h o w decimal (0-50 m m)

C A L + "•“ Sho w integer (0-500 m m)

ED400 C O A T I N G T H I C K N E S S G A U G E F O R N O N - F E R R O U S M E T A L S

6.2 Reset

T h e reset operation will reload the factory settings

and calibration data. It can be completed by pressing the

" D E L " and " S T A T " buttons at the s a m e time. The

L C D will display “一一一”,then “0” or “0.0” . After the

reset operation, a dual point calibration must be

completed. 6.3 L o w Battery

W h e n the battery is low, the L C D will display

" L O B A T " . W h e n the gauge run out of battery, it will

shut d o w n automatically.

6.4 P o w e r saving

If no measurement is detected within one minute, the

gauge will turn off automatically.

7. N O T I C E

7.1 Before using the gauge, it should be checked for

accuracy. Normally it can be checked by measuring on

the uncoated substrate. For further assurance, measure

with the calibration foil and check if the measuring result

matches the marked thickness.

7.2 W h e n measurements are repeated on the s a m e

workpiece, the results m a y vary. Take the average value

for better accuracy.

Page 6: USER' S MANUAL ED-400 - Shenyang Tianxing Testing ... S MANUAL ED-400 COATING THICKNESS GAUGE FOR NON-FERROUS METALS TX TESTING INSTRUMENTS CO., LTD. ... All characters Probe orcircuit

ED400 C O A T I N G T H I C K N E S S G A U G E F O R N O N - F E R R O U S M E T A L S

8 . E R R O R F A C T O R S

According to ISO 2360-2003, the following factors

will influence the accuracy of measurement.

8.1 Thickness of coating

If the coating thickness is less than 5 m m,the average

value should be taken. If the coating thickness is less

than 3 m m, it cannot be measured accurately with this

gauge.

8.2 Edge effect

When the measuring area less than 150mm2 or the

width is less than 12mm, the gauge should be re-

calibrated on the same uncoated workpiece. 8.3

Curvature

If the diameter of the workpiece is less than 50mm,

the gauge should be re-calibrated on the same uncoated

workpiece.

8.4 Surface roughness

Measure at different spots of the workpiece to reduce

the influence of the rough surface.

8.5 Conductivity of the workpiece

8.6 Thickness of the workpiece

8.7 Gap between probe and workpiece surface

8.8 Pressing force on the probe

8.9 Perpendicularity of probe

8.10 Temperature of probe

ED400 COATING THICKNESS GAUGEFOR N O N - F E R R O U S METALS

9. TROUBLESHOOTING

9.1 Failures, reasons and solutions

Table 2

Failure Reason Solution

Power saving Press ON/OFF button

No display or No battery or incorrect

battery installation Reinstall the battery

immediate shutdown Empty battery Change battery

Broken battery connector

Contact manufacturer

No response on LCD Thickness over range Check workpiece

to measurement Probe failure Change probe

All characters Probe orcircuit Change probe or

displayed board failure Contact manufacturer

Curving surface or Calibrate on uncoated

small mearing area same workpiece

Calibration wrong Re-Calibrate

Obviously wrong

Incorrect probe

handling Press "DEL" button

measuring result Calibration data

disorder Reset and calibrate

Probe failure Change probe

Probe changed

without calibration Re-Calibrate

Unable to turn off Circuit problem Contact manufacturer

Page 7: USER' S MANUAL ED-400 - Shenyang Tianxing Testing ... S MANUAL ED-400 COATING THICKNESS GAUGE FOR NON-FERROUS METALS TX TESTING INSTRUMENTS CO., LTD. ... All characters Probe orcircuit

E D 4 0 0 C O A T I N G T H I C K N E S S G A U G E F O R N O N - F E R R O U S M E T A L S

9.2 Error codes

Table 3

C o d e Error Solution

E4 Probe failure,connection problem

or circuit board failure

Connect probe again or

Go门tact manufacturer

E5 Probe failure Change probe

E6 Measuring range exceeded Re-Calibrate

E7 Faulty calibration Re-Calibrate

10. Daily maintenance

10.1 The probe must be kept dry and clean. If the

instrument w a s used on the assembly line, the workpiece

should be wiped dry so that no water or other fluid runs

into the probe. Do not add lubricant to the probe.

10.2 W h e n not in use, the gauge should be put into the

case to avoid impact and dirt.

10.3 R e m o v e the battery if the gauge will be stored for a

long period of time.