user guide - landinst.nl · user guide issue 2: march 2010 publication no. aeqa 198.283 ... if you...

24
Aluminium Extrusion & Quench Thermometer System - Actuator Copyright © 2003 - 2010 Land Instruments International User Guide Issue 2: March 2010 Publication N o . AEQA 198.283 Language: English

Upload: others

Post on 23-Oct-2020

5 views

Category:

Documents


0 download

TRANSCRIPT

  • Aluminium Extrusion & Quench Thermometer System - Actuator

    Copyright © 2003 - 2010 Land Instruments International

    User Guide

    Issue 2: March 2010 Publication No. AEQA 198.283

    Language: English

  • Design and Manufacturing Standards Signs and Symbols Used on Equipment and Documentation

    Caution, risk of electric shock. This item or material can be recycled.

    Caution, attention to possibility of risk of damage to the product, process or surroundings. Refer to instruction manual.

    Caution, hot surface.

    This item or material can be recycled.

    This item or material must be disposed of in accordance with the Waste Electrical and Electronic Equipment directive as applied by local regulations.

    Protective Conductor Terminal.

    Observe precautions for handling electrostatic discharge sensitive devices.

    IMPORTANT INFORMATION - PLEASE READ

    Important Health and Safety InformationEquipment Operation Use of this instrument in a manner not specified by Land Instruments International may be hazardous.

    Electrical Power Supply Before working on the electrical connections all of the electrical power lines to the equipment must be isolated. All the electrical cables and signal cables must be connected exactly as indicated in these operating instructions. If in doubt contact Land Instruments International.

    Face and Eye ProtectionSuitable face and eye protection must be worn when working on hot vessels and ducts! Special safety measures must be taken when working on a high-pressure duct.

    Protective ClothingProtective clothing must always be worn when working in the vicinity of hot vessels or ducts.

    StorageThe instrument should be stored in its packaging, in a dry sheltered area.

    UnpackingCheck all packages for external signs of damage. Check the contents against the packing note.

    Return of Damaged Goods

    IMPORTANT If any item has been damaged in transit, this should be reported to the carrier and to the supplier immediately. Damage caused in transit is the responsibility of the carrier not the supplier.

    DO NOT RETURN a damaged instrument to the sender as the carrier will not then consider a claim. Save the packing with the damaged article for inspection by the carrier.

    Return of Goods for RepairIf you need to return goods for repair please contact our Customer Service Department. They will be able to advise you on the correct returns procedure.

    Any item returned to Land Instruments International should be adequately packaged to prevent damage during transit.

    You must include a written report of the problem together with your own name and contact information, address, telephone number, email address etc.

    Return of Goods for Repair Form is available for download from our websites.

    Lifting InstructionsWhere items are too heavy to be lifted manually, use suitably rated lifting equipment. Refer to the Technical Specification for weights. All lifting should be done as stated in local regulations.

    Office Locations

    UK - Dronfield Tel: +44 (0) 1246 417691 E-Mail: [email protected]: www.landinst.com

    USA - Pittsburgh AMETEK Land, Inc. Tel: +1 412 826 4444E-Mail: [email protected]: www.ametek-land.com

    For further details on all LAND/ Ametek offices, distributors and representatives, please visit our websites.

    The Quality Management System of Land Instruments International is approved to BS EN ISO 9001 for the design, manufacture and on-site servicing of combustion, environmental monitoring and non-contact temperature measuring instrumentation.

    Approvals apply in the USA

    This instrument complies with current European directives relating to Electromagnetic Compatibility 89/336/EEC and Low Voltage Directive 73/23/EEC.

    Operation of radio transmitters, telephones or other electrical/electronic devices in close proximity to the equipment while the enclosure doors of the instrument or its peripherals are open, may cause interference and possible failure where the radiated emissions exceed the EMC directive.

    The protection provided by both CE and IP classifications to this product may be invalidated if alterations or additions are made to the structural, electrical, mechanical or pneumatic parts of this system. Such changes may also invalidate the standard terms of warranty.

    DimensionsUnless otherwise stated, all measurements are given in millimetres and inches.

    Copyright

    This manual is provided as an aid to owners of Land Instruments International’s products and contains information proprietary to Land Instruments International. This manual may not, in whole or part, be copied, or reproduced without the expressed written consent of Land Instruments International Ltd.

    Copyright © 2007 - 2010 Land Instruments International.

  • PRODUCT WARRANTYThank you for purchasing your new product from Land Instruments International. This Land manufacturer’s ‘back-to-base’ warranty covers product malfunctions arising from defects in design or manufacture. The warranty period commences on the instrument despatch date from the Land Instruments International Ltd. factory in Dronfield, UK.

    36 MONTHS WARRANTYBuilding upon the reputation for reliability and longevity that System 4 and UNO thermometers have earned, Land are delighted to be able to provide our customers with an industry-leading 36 month warranty for the following products:-

    • System 4 thermometers, processors, accessories and mountings and special instruments based on System 4.

    • UNO thermometers, accessories and mountings and special instruments based on UNO.

    • Application-dedicated processors based on LANDMARK® Graphic.

    • ABTS/S and ABTS/U.

    • FTS.

    • VDT/S and VDT/U.

    • DTT.

    • FLT5/A.

    This 36 month warranty is provided as standard for all orders for the products listed above received from 1st May 2002.

    We believe that our customers expect us to set the standard in terms of performance, quality, reliability and value for money. This 36 months warranty, as a part of an on-going program of continuous improvement, is just one way in which Land strive to maintain our position as the temperature measurement partner of choice.

    12 MONTHS WARRANTYAll Land Instruments International products not provided with 36 months warranty (see list above), are provided with a 12 months warranty.

  • PRODUCT WARRANTYEXCLUSIONS FROM WARRANTY

    It should be noted that costs associated with calibration checks which may be requested during the warranty period are not covered within the warranty.

    Land reserve the right to charge for service/calibration checks undertaken during the warranty period if the cause is deemed to fall outside the terms of the warranty.

    This Land manufacturer’s warranty does not cover product malfunction arising from:-

    • incorrect electrical wiring.

    • connection to electrical power sources outside the rating of the product.

    • physical shock (being dropped, etc.) and impact damage.

    • inappropriate routing, support, physical shock & strain protection, etc. of the lightguide (Fibroptic thermometers only).

    • environmental conditions exceeding the IP / NEMA rating of the product.

    • environmental conditions outside the Ambient Temperature, Humidity and Vibration rating of the product.

    • environmental contamination (solvent vapours, deposition of airborne contamination, cooling liquids of non-neutral pH, etc.).

    • overheating as a result of interruption of water/air flow through cooling jackets or of incorrect installation.

    • inappropriate modification of product (drilling holes in thermometer bodies, etc.).

    • inappropriate recalibration which results in product calibration being taken outside specification.

    • improper resealing of thermometer following parameter adjustment (UNO, FLT5/A, etc.).

    • attempted repair by a non-Land-authorised repair centre.Land Instruments International Ltd • Dronfield S18 1DJ • England • Tel: +44 (0) 1246 417691 • Fax: +44 (0) 1246 410585 Email: [email protected] • www.landinst.comAMETEK Land, Inc . • 150 Freepor t Rd. • P i t t sburgh, PA 15238 • U.S.A. • Te l : +1 (412) 826 4444 Email: [email protected] • www.ametek-land.com

    For a complete list of our international offices, please visit www.landinst.com 06/10

  • Thermometer-Actuator User Guide

  • User Guide Thermometer-Actuator

    Contents

    1 Introduction 1

    2 Specifications 2

    3 Mounting 3

    4 Adjustment of the Angular Travel Limiter 6

    5 Electrical Connections 8

    5.1 Connection to the Mains Electrical Supply 85.2 Connection to the AET Laser Alignment Circuitry 9

    6 Primary Alignment 12

    6.1 Alignment Procedure 12

    7 Cleaning and Maintenance Instructions 13

    7.1 General 13

    8 Spares and Repairs 14

    8.1 Motor Control Unit and Signal Isolator 148.2 Motor and Gearbox Assembly 14

    Appendices 15

    Appendix 1 AET to AETS Reconfiguration 15Appendix 2 AETS Laser Alignment Unit Reconfiguration 16

  • Thermometer-Actuator User Guide

  • Page 1

    User Guide Thermometer-Actuator

    1 IntroductionThe LAND Aluminium Extrusion & Quench Thermometer Actuator is a motorised, operator controlled, mounting unit intended for use at the die exit or quench exit on aluminium extrusion presses.

    It is used in conjunction with the AET/AQT integral alignment laser, to control the position of the thermometer target spot.

    It allows for rapid and precise alignment following die changes, to reposition on multicavity strands or to maximise the signal strength (emissivity indication) on narrow sections.

    The actuator unit is controlled remotely from a small handset which features individual push buttons for the motion directions. A third (smaller) handset button also allows for remote triggering of the AET/AQT laser alignment and is powered direct from the actuator control unit.

    The handset is connected to the actuator control unit via a 12m length of screened cable.

    A mechanical clutch and angular travel limiter plate ensure that no excessive movement can occur to strain pipes or cables.

    Fig. 1 - AET/AQT Actuator Assembly and Handset 283001

  • Page 2

    Thermometer-Actuator User Guide

    2 Specifications

    Mains input: (Part No. 092.678) 110V a.c.; 50/60Hz; (Part No. 092.683) 230V a.c.; 50/60Hz

    Installation Category II Single Phase supply

    Current rating: Under-load current 0.9A (including stall current) Quiescent current 50mA

    Fuse: 1.6A (T) HRC 250V a.c.

    Angular travel limit: 30° travel limit plate fitted as standard (20° alternative travel limit plate also supplied)

    Speed of rotation: Approx. 1.5°/sec (in both directions)

    Sealing IP54

    Ambient operating temperature range: 0 to 60°C

    Pollution: Degree 2

  • Page 3

    User Guide Thermometer-Actuator

    3 MountingThe AET/AQT Actuator is intended for mounting the AET thermometer onto the press face or suitable near-by part of the press superstructure. It may also be used, where applicable, for the alignment of a quench (AQT) thermometer when mounted at the exit of a quench box or nearby structure, to allow for aiming onto sections exiting the quench.

    Fig.2showsthemountingplateandfixingpointdimensions.Theplateismountedontotheselectedstructureusingaminimumof4xM8screwsfittedwith either nyloc nuts or anti-vibration washers.

    The remote handset controls the operation of the actuator and may be separated from it by up to 12 metres. The connecting cable should be routed away from moving machinery and the handset positioned at a convenient locationatwhicheitherthelaserindicatorspotontheemergingprofiles,orthe signal strength indicator (emissivity) reading, may be clearly seen.

    Before use, ensure that all cable glands are tightened to prevent the ingress of moisture, dirt or any other contamination.

    For typical die exit use, the plate is normally mounted such that the distance from the front of the thermometer to the extrusion axis is approximately 1 to 1.5 metres. The thermometer’s viewing axis should be set (see section 6 - Alignment) so that the target spot is ‘upstream’ of any air blowers or water sprays and therefore the indicated temperature is as close to that at the die exit as possible. Also refer to the AET & AQT thermometer operating instructions for further advice on alignment.

    NOTE

    When mounting or positioning the actuator unit, the suggested orientation should be as shown in Fig. 1. Alternatively, the actuator may be positioned or mounted in any orientation as dictated by services requirements or space availability.

  • Page 4

    Thermometer-Actuator User Guide

    NOTE

    COMPATIBILITY WITH EXISTING AETS/AQTS SYSTEMSAn additional ball swivel is supplied as an accessory to allow compatibility with existing AETS and AQTS systems. The thread on this ball swivel matches the mounting holes in the environmentally protective jackets supplied with AETS and AQTS thermometers.To change the ball swivel type, loosen the ball swivel thread lock nut and unscrew the instrument off the ball swivel thread. Undo, remove and retain the ball clamp nut. Replace the ball swivel, locate and tighten the ball clamp nut. Mount the instrument onto the ball swivel thread and secure using the thread locknut. Slightly loosen the ball swivel clamp nut, align the instrument with the intended target area and re-tighten the clamp nut. For alignment details, see section 6.0.

  • Page 5

    User Guide Thermometer-Actuator

    Fig. 2 - Mounting Plate Dimensions 283002

    60,3^60,1

    4 Holes Ø

    6,5 on 75 PCD

    8 Holes Ø

    9

    4 Holes Ø

    6,5

    3

    5030

    95

    165240

    120120

    120

    420

    210

    22166

    125

    10

    50110

    190

  • Page 6

    Thermometer-Actuator User Guide

    4 Adjustment of the Angular Travel LimiterFor smaller presses it is recommended that the maximum angular displacement is limited to 20° (i.e. ±10° from the mean position). The unit is supplied as standard with a 30° (±15°) limiter, however, an alternative 20° (±10°) limiter is also included.

    To decide on the angular travel required, measure the distance form the mounting spindle/ball joint to the axis of the extrusions at the target position. Call this dimension ‘x’. Determine the maximum width of the largest possible extruded section that could be produced and call this dimension ‘y’.

    Calculate the value of the angle given by utilising one of the following two formulae:

    (Thecalculatedfigureiscommontobothformulae.)

    Takethecalculatedvalueandmultiplyby2.Ifthisfigureexceeds16°thenthe standard 30° limiter is recommended.

    To change the adjustment limiter, perform the following operation:

    (Note that the following instructions may be carried out before or after mounting). Using the handset controls, adjust the rotation of the moving spindle/jacket etc. until the notch on the pivot plate aligns with the inscribed line on the limiter plate (see Fig. 3). At this point it is possible to loosen and remove the 2x M6 cap head screws which hold the limiter plate in position onto the gearbox casing, as well as the M6 x 20 screw which acts as the limiter. The 30° plate may then be substituted by the 20° plate also supplied. Replace and re-tighten the cap head screws.

    arc tan or tan-1y2*x

    y2x

  • Page 7

    User Guide Thermometer-Actuator

    Fig. 3 - View of Actuator with Ball Joint Mounting Removed 283003

    Pivot PlateM4 x 16 Clamp ScrewsM6 x 20 Limit Peg Screws

    M6 Limit Plate Fixing Screws

    Limit Plate

    Midpoint Indicators

  • Page 8

    Thermometer-Actuator User Guide

    5 Electrical Connections

    5.1 Connection to the Mains Electrical SupplyThe following instructions must only be carried out by suitably trained personnel. All local safety warnings must be adhered to at all times.

    Remove the 4 securing screws from the lid of the motor control box to allow access to the terminal block within. The mains electricity supply connections are clearly indicated in Fig. 4 below:

    CAUTION - Risk of Electric Shock

    NOTE

    Before commencing the following instructions, ensure that the actuator is isolated from all mains power.

    Fig. 4 - Open view of control unit showing user connection block 283004

    Connection Terminal Block

    Main FuseHRC F1, 1.6A

    Laser Cable Gland Mains Input Cable Gland

  • Page 9

    User Guide Thermometer-Actuator

    The mains cable should be of type 3183-Y and must be routed through the cable gland on the box wall.

    The mains cable should be provided with strain relief, to within 150mm of the cable entry point.

    An unobscured switch or breaker should be installed in the mains circuit and situated close to the actuator. It should be marked as the disconnecting device for the actuator.

    Bare the ends of the mains electrical supply wires and connect as follows:

    •Totheterminalmarked7(protectiveearth)connectthe green/yellow wire.

    •Totheterminalmarked8(neutral)connectthebluewire.

    •Totheterminalmarked9(live)connectthebrownwire.

    When the connections have been made at the terminal strip, tighten the gland nut to prevent the ingress of dirt or moisture.

    5.2 Connection of the AET Laser Alignment CircuitryThe following instructions must only be carried out by suitably trained personnel. All local safety warnings must be adhered to at all times.

    The alignment laser is triggered into operation by a simple switch closure, and remainsactiveforaperiodof50to60seconds.Thisisusuallyquitesufficientto enable a re-alignment of the thermometer target following a change of die, or, if excessive lateral movement of the extruded section occurs.

    CAUTION - Risk of Electric Shock

    NOTE

    Before commencing the following instructions, ensure that the actuator is isolated from all mains power.

  • Page 10

    Thermometer-Actuator User Guide

    Fig. 5 - View of Backcap Terminals (with laser trigger connections) 283005

    The AET Actuator handset is provided with a green push button laser trigger.Remove the blanking plugs from the thermometer jacket back cap and the smallerholeintheactuatorcontrolbox,marked“laser”,andfitsuitablesealing glands with a minimum IP rating of 68 (Land Part No. 150.698) to both. Ensure that the seal of the gland is compatible with the size of the 2-core cable to be utilised for the interconnection between the back cap and the actuator control box.

    When judging the length of the interconnection cable allow some slack to ensure free rotation of the thermometer head and thermometer jacket during operation.

    If a connection has already been made to terminal B in the back cap (signal 4: purple wire) remove and isolate the existing connection. Connect a 2-core cable from terminals 3 and B, as shown in Fig. 5, to terminals 2 and 5 in the actuator electronic control unit, as shown in Fig. 6.

    Thermometer Connections

    Laser Trigger Connections

  • Page 11

    User Guide Thermometer-Actuator

    Fig. 6 - View of AET Actuator Terminals (laser trigger) 283006

    Mains Input Connectors

    Handset Lead Wires

    Laser Trigger Wires from Thermometer

    Backcap (use any colour)

    To Fuse

    Yello

    w

    Blu

    eRe

    d

    Whi

    te

    Bla

    ck

    To Electronic Control

    Unit

    saharrisSticky Noteimage is cutting text below.

  • Page 12

    Thermometer-Actuator User Guide

    6 Primary AlignmentThis is the process by which the thermometer is aimed to the required point on the extrusion axis, with the actuator in the nominal mid position. This procedure is carried out by adjusting the ball joint mechanism.

    6.1 Alignment ProcedureUsing the buttons on the hand set, move the pivot plate of the actuator unit to the mid-travel position. This is indicated (see Fig. 3) by aligning the notch in the plate with the inscribed line on the limit plate, mounted on the gearbox casing.

    At this point, the two holes in the pivot plate will align with the tapped holes in the limit plate. Insert the 2xM4 cap head screws into these holes and tighten. This locks the pivot plate into the central position. Using either the signal strength (emissivity) indication from the thermometer (viewing an on-axis extrusion), or the laser indicator spot, move the thermometer jacket by hand, until the required alignment is attained.

    Tighten the 36mm ball joint locking nut as well as the M16 locknut to secure the threaded spindle into the protection jacket.

    Remove the 2xM4 cap head screws to allow the unit to be operated from the handset if required.

    NOTE

    It is recommended that the notes in Section 1 of the AET/AQT Operating Instructions, are fully read before proceeding with the following instructions.

  • Page 13

    User Guide Thermometer-Actuator

    7 Cleaning and Maintenance Instructions

    7.1 GeneralThe actuator unit should not require any special regular maintenance. It is, nevertheless, recommended that in order to prevent the ingress of any dirt or contamination, the external surfaces should be periodically wiped with a damp cloth. The unit should not, under any circumstances, be subjected to a high pressure water hose or immersed in water.

  • Page 14

    Thermometer-Actuator User Guide

    8 Spares and Repairs

    8.1 Motor Control Unit and Signal Isolator

    A complete Motor Electronic Control Unit is available as a spare part:

    Specification: FullyconfiguredandinterconnectedRegenerativeDrive Unit with Mounted Signal Isolator

    Part Numbers: 031.448(configuredfor110Va.c.input) 031.502(configuredfor230Va.c.input)Supplier: Land Instruments International Limited Dronfield S18 1DJ England

    8.2 Motor and Gearbox Assembly

    A replacement motor and gearbox is available as a spare part:

    Specification: MotorandGearboxAssemblyAET/AETS ActuatorPart Number: 031.417Supplier: Land Instruments International Limited Dronfield S18 1DJ England

  • Page 15

    User Guide Thermometer-Actuator

    Appendices

    Appendix 1 - AET to AETS Reconfiguration

    Thefollowingdiagramdetailstheactionsrequiredwhenreconfiguringtheactuator from AET applications to AETS applications.

    Reconfiguration Steps

    STEP 1 Fit a link between terminals 2 & 6.

    STEP 2 Amend the termination of the lead from the hand held unit -

    Move the black wire from terminal 2 to terminal 1. Move the red wire from terminal 6 to terminal 2.

    Fig.7-ActuatorReconfiguration283007

    STEP 1-(link)

    STEP 2-(black wire)STEP 2-(red wire)

    Lead from hand held unit

    NOT TO SCALE

  • Page 16

    Thermometer-Actuator User Guide

    Appendix 2 - AETS Laser Alignment Unit Reconfiguration

    Thefollowingdiagramdetailstheactionsrequiredwhenreconfiguringtheactuator to include an optional AETS Laser Alignment Unit (092.532).

    Fig.8-OptionalAETSLaserAlignmentUnitReconfiguration283008

    Reconfiguration Steps

    STEP 1 Connect the Green wire to terminal 1.

    STEP 2 Connect the Blue wire to terminal 5.

    STEP 3 Connect the Red wire to terminal 4. This link should include a Zener Diode(1Watt,6V2or7V5)fittedby the customer, to reduce the voltage from +15V to +8/+9V.

    STEP 1-(Green wire)

    STEP 3-(Red wire)STEP 2-(Blue wire)

    NOT TO SCALE

    3-Core Laser wire

    Lead from Hand Held Unit

    Product WarrantyContents1 Introduction2 Specifications3 Mounting4 Adjustment of the Angular Travel Limiter5 Electrical Connections5.1 Connection to the Mains Electrical Supply5.2 Connection of the AET Laser Alignment Circuitry

    6 Primary Alignment6.1 Alignment Procedure

    7 Cleaning and Maintenance Instructions7.1 General

    8 Spares and Repairs8.1 Motor Control Unit and Signal Isolator8.2 Motor and Gearbox Assembly

    AppendicesAppendix 1 - AET to AETS ReconfigurationAppendix 2 - AETS Laser Alignment Unit Reconfiguration