Page 1: Use of Drager Pump & Tubes. Preparation Pump Leak Test  Insert unopened tube into the socket  Squeeze pump completely and release  Pump is adequately

Use of Drager Pump & Tubes

Page 2: Use of Drager Pump & Tubes. Preparation Pump Leak Test  Insert unopened tube into the socket  Squeeze pump completely and release  Pump is adequately

Preparation

Page 3: Use of Drager Pump & Tubes. Preparation Pump Leak Test  Insert unopened tube into the socket  Squeeze pump completely and release  Pump is adequately

Pump Leak Test Insert unopened tube into the socketSqueeze pump completely and releasePump is adequately leak-proof if the

end-of-stroke volume indicator has not appeared after 15 minutes

Remove tube from socketPress button to reset counter to zero

Page 4: Use of Drager Pump & Tubes. Preparation Pump Leak Test  Insert unopened tube into the socket  Squeeze pump completely and release  Pump is adequately

Preparing the TubeSelect tube for the measurement

required.Follow instructions for use (tube). Insert tip in opening on the lower pump

plate; keep it pressed against the ceramic edge while turning it, so that the glass is scored.

Page 5: Use of Drager Pump & Tubes. Preparation Pump Leak Test  Insert unopened tube into the socket  Squeeze pump completely and release  Pump is adequately

Pull tube over at an angle until the tip breaks off.

Discard broken tip.Open other end of tube in the same

way. Insert tube into pump, the arrow must

point towards the pump.

Page 6: Use of Drager Pump & Tubes. Preparation Pump Leak Test  Insert unopened tube into the socket  Squeeze pump completely and release  Pump is adequately

Making a MeasurementCheck number of strokes required for the

tube.Put forefinger and ringfinger in the middle

of the lower pump plate and squeeze completely.

Release pump.Evaluate the result.Remove used tube from socket.Flush pump.

Page 7: Use of Drager Pump & Tubes. Preparation Pump Leak Test  Insert unopened tube into the socket  Squeeze pump completely and release  Pump is adequately

Carbon Dioxide TubesDetermination of CO2 in airMeasuring range – 100 to 3000ppmNumber of strokes – 10Time of measurement – approx. 4 minsColour change – white to violet

Page 8: Use of Drager Pump & Tubes. Preparation Pump Leak Test  Insert unopened tube into the socket  Squeeze pump completely and release  Pump is adequately

Carbon Dioxide

(Simple Asphyxiants)

Page 9: Use of Drager Pump & Tubes. Preparation Pump Leak Test  Insert unopened tube into the socket  Squeeze pump completely and release  Pump is adequately

Acute EffectsRapid loss of consciousnessComaAnoxic brain injury

Page 10: Use of Drager Pump & Tubes. Preparation Pump Leak Test  Insert unopened tube into the socket  Squeeze pump completely and release  Pump is adequately

Chronic EffectsResidual anoxic injury

Page 11: Use of Drager Pump & Tubes. Preparation Pump Leak Test  Insert unopened tube into the socket  Squeeze pump completely and release  Pump is adequately

Exposure LimitsAtmospheric oxygen should > 18% at

normal barometric pressure

Page 12: Use of Drager Pump & Tubes. Preparation Pump Leak Test  Insert unopened tube into the socket  Squeeze pump completely and release  Pump is adequately

General ConsiderationsAsphyxiants gases are those inhalants

that act predominantly by oxygen deprivation.

This may occur by simply displacing oxygen from inspired air.

Page 13: Use of Drager Pump & Tubes. Preparation Pump Leak Test  Insert unopened tube into the socket  Squeeze pump completely and release  Pump is adequately

Occupational & Environmental ExposureOne of confined space exposure

(storage tanks or mines)Product of combustion and fermentation

processes.Natural volcanic lake release