unit 9 atomic absorption

Upload: krishnakkumar-krish

Post on 07-Apr-2018

221 views

Category:

Documents


3 download

TRANSCRIPT

  • 8/3/2019 Unit 9 Atomic Absorption

    1/26

    5

    Atomic Absorption

    SpectrophotometryUNIT 9 ATOMIC ABSORPTION

    SPECTROPHOTOMETRY

    Structure

    9.1 IntroductionObjectives

    9.2 Principle of Atomic Absorption SpectrophotometryConcentration Dependence of AbsorptionQuantitative Methodology

    9.3 Instrumentation for Atomic Absorption SpectrophotometryRadiation SourcesAtomisers

    Monochromators

    Detectors

    Readout Devices

    9.4 Graphite Furnace Atomic Absorption SpectrophotometryElectrothermal Atomisers

    Handling Background Absorption in GFAASAdvantages and Disadvantages of GFAAS

    9.5 Atomic Absorption SpectrophotometersSingle Beam Atomic Absorption Spectrophotometer

    Double Beam Atomic Absorption Spectrophotometer

    9.6 Interferences in Atomic Absorption SpectrophotometrySpectral Interferences

    Chemical InterferencesPhysical Interferences

    9.7 Sample Handling in Atomic Absorption SpectrophotometryPreparation of the Sample

    Use of Organic Solvents

    Microwave DigestionSample Introduction Methods

    9.8 Applications of Atomic Absorption Spectrophotometry9.9 Summary9.10 Terminal Questions9.11 Answers9.1 INTRODUCTIONYou have learnt in Block 3 that in atomic spectrometry, the elements present in a

    sample are converted into gaseous atoms by a process called atomisation and their

    interaction with the radiation is measured. In Units 7 and 8 of the third block youhave learnt about flame photometry and atomic fluorescence spectrometry. In flamephotometry we measure the emission of radiation by thermally excited atoms whereas

    in atomic fluorescence spectrometry we monitor the fluorescence emission from the

    radiationally excited atoms. In this unit you would learn about atomic absorptionspectrophotometry (AAS) that concerns the absorption of radiation by the atomised

    analyte element in the ground state. The atomisation is achieved by the thermal

    energy of the flame or electrothermally in an electrical furnace. The wavelength(s) ofthe radiation absorbed and the extent of the absorption form the basis of thequalitative and quantitative determinations respectively.

    The atomic absorption methods using flame are rapid and precise and are applicable

    to about 67 elements. Electrothermal methods of analysis on the other hand are slower

    and less precise; however, these are more sensitive and need much smaller samples.

    As the absorption of resonance radiation is highly selective and also very sensitive,the technique of AAS has became a powerful method of analysis, which is used for

  • 8/3/2019 Unit 9 Atomic Absorption

    2/26

    6

    Atomic SpectroscopicMethods-II

    trace elemental determinations in most analytical laboratories for a wide variety of

    applications.

    We begin the unit with an understanding of the origin of atomic absorption spectrum

    and learn about the principle behind atomic absorption spectrophotometry being used

    as an important analytical technique. Then we will take up the instrumentationrequired for the measurement of atomic absorption spectrum. An account of the

    possible interferences in atomic absorption spectrophotometry will be followed by

    sample handling procedures, like preparing and loading the sample for the spectral

    measurements. The qualitative and quantitative applications of atomic absorptionspectrophotometry will be followed by the merits and demerits of the method. In the

    next unit you would learn about atomic emission spectrometry and its applications indiverse areas.

    Objectives

    After studying this unit, you will be able to:

    explain the principle of atomic absorption spectrophotometry, outline the quantitative methodology of the atomic absorption

    spectrophotometry,

    draw a schematic diagram illustrating different components of a flame atomicabsorption spectrophotometer,

    justify the usage of line radiation sources in atomic absorptionspectrophotometry,

    describe the functioning of different types of nebulisers used in atomicabsorption spectrophotometry,

    compare the flame and flameless atomisation of the analyte in terms ofsensitivity and detection limits,

    outline the importance of sample handling in atomic absorptionspectrophotometry,

    discuss the interferences observed in atomic absorption spectrophotometricdeterminations, and

    state the merits and limitations of the atomic absorption spectrophotometrictechnique.

    9.2 PRINCIPLE OF ATOMIC ABSORPTIONSPECTROPHOTOMETRY

    The concept of atomic absorption spectrometry (AAS) was proposed by two groups in

    1955, A. Walsh of Australia and another one of C T J Alkamade and J M W Milatz

    from The Netherlands. You have learnt that in atomic spectroscopy, the analyte must

    be present in the atomic vapour state. In atomic absorption spectrophotometry theatomisation is performed by aspirating the sample solution into a flame where the

    analyte element is converted into gaseous phase atoms. Alternatively, the sample isfed into a graphite furnace where the atomisation is achieved electrothermally atrelatively lower temperature, below 3000 K. As the temperature of atomisation is low;

    most of the atoms remain in the ground state which can absorb characteristic radiation

    from the radiation source made from the analyte element. The atomic vapourscontaining free atoms of an element in the ground state are illuminated by a radiation

    source emitting the characteristic radiation of the analyte. You would recall fromUnit 8 that in halogen cathode lamp the cathode is made of the element that needs to

    be determined and gives radiations characteristic of the element.

    It may be noted that as

    only ground state atomsare involved in this

    process. Therefore, the

    ionisation occurring due

    to high temperature of theflame needs to be kept toa minimum.

  • 8/3/2019 Unit 9 Atomic Absorption

    3/26

    7

    Atomic Absorption

    SpectrophotometryThe radiation is absorbed by the analyte vapours and its intensity decreases. This is

    similar to spectrophotometry which you have learnt earlier in Unit 2; molecules beingreplaced by atoms and the lamp changed to a line source. The degree of absorption is

    a quantitative measure of the concentration of ground state atoms in the vapours. The

    analysis is done by comparing the observed absorption with the one obtained bysuitable standard samples of the analyte under similar experimental conditions, i.e., a

    calibration curve method is generally employed.

    9.2.1 Concentration Dependence of AbsorptionYou have learnt earlier that according to Boltzmann distribution law, the population

    of the ground state i.e., the number of species in the ground state is highest and it

    keeps on decreasing as we go to higher energy levels. It can be shown that for mostelements at moderate temperatures prevailing in a flame, nearly all atoms are in

    ground state leaving only a few atoms in excited state (Refer Example 1, Unit 7, page8, Block 3). The absorption follows Lambert-Beers law so that the concentration of

    an analyte element in the vapours in the flame may be determined.

    You would recall from Unit 2 that according to Lambert-Beers law, the extent of

    radiation absorbed by the absorbing species is a function of the path length and theconcentration of the absorbing species.

    Mathematically, bcP

    P=0log

    where,Po = radiant power of incident light,

    P = radiant power of transmitted light,b = thickness of the absorbing medium,

    = absorption coefficient, and

    c = concentration of absorbing analyte atoms.

    The term, log Po/P is called absorbance and is represented as A. Therefore we can

    write it as follows.

    bcAP

    P

    ==

    0

    log

    Thus, absorbance of the sample is directly proportional to the concentration of theanalyte. Therefore, a calibration plot of concentration of analyte element versus

    absorbance is drawn from the standard solutions and the concentration of element in

    unknown solution is read directly from the graph. However, such a linear relationship

    between the absorption and the concentration can be observed only if all radiationpassing through the sample is absorbed to the same extent by the analyte atoms.However, the experimental concentration versus intensity calibration curve is

    observed to be deviating from the linearity as a result of the presence of nonabsorbedradiation and other interferences. Therefore, suitable measures need to be taken so as

    to minimise the interferences and obtain the linearity in the calibration curves. We

    would discuss about these interferences in Section 9.6. Let us learn about the

    methodology used in quantitative determinations using atomic absorptionspectrophotometry.

    9.2.2 Quantitative MethodologyLike many other analytical methods, AAS is also not an absolute method of analysis.

    The routine analytical methodology for quantitative determinations using AAS is

    based on calibration method. Besides this, the internal standard method and standard

    addition methods are also employed. You have learnt about these methods in Unit 7in the context of flame photometry. These are briefly recalled here.

    Typical absorbance must

    be in the range 0.1 to 0.3or else precision is poorer

    at the extremes due to

    instrumental noise.

    In AAS the absorption of

    resonant radiation byground state atoms of the

    analyte is used as the

    analytical signal.

  • 8/3/2019 Unit 9 Atomic Absorption

    4/26

    8

    Atomic SpectroscopicMethods-II Calibration plot method

    In this method, a calibration plot is drawn by aspirating standard solutions of knownconcentration into the flame and measuring absorbance for each solution. The

    concentration of the unknown solution is then determined from the calibration plot.

    Despite the fact that Beers law is followed in AAS, in practice the departures fromlinearity are encountered as shown in Fig. 9.1. The nonlinearity is due to the

    transmission of unabsorbed light from the radiation source. In addition, a number of

    uncontrolled variables in atomisation and absorbance measurements may also affect

    the measurements. Therefore, we need to find the concentration range in which theLambert-Beers law holds i.e., we get a straight line.

    Fig. 9.1: Typical calibration plot between the absorbance and the concentration of

    analyte element

    In practice, however, a single calibration does not serve the purpose. We need to take3-4 standards of different concentration and a bank to obtain a suitable calibrationplot. As you can see in Fig. 9.1, a single standard calibration plot does not hold good.

    Further, if the analyte concentration happens to be outside the limits of the standards

    used for calibration then the analyte sample should be suitably diluted orconcentrated.

    As the test solution is often a complex whose all constituents are not known, it

    becomes almost impossible to prepare standard solutions having a similarcomposition to the analyte sample to obtain a calibration plot. In such cases we have

    to use internal standard method and the standard addition method.

    Internal standard method

    In this method, a fixed amount of an internal standard which is chemically similar tothe analyte being determined and absorbs at similar wavelength, is added to the

    standard solution and the test sample. The intensity ratio of the analyte and internalstandard is plotted as a function of the analyte concentration in the standard solution.For example, while determining Na or K in blood serum Li is used as internal

    standard. A typical plot obtained in an atomic absorption spectrophotometric

    determination using internal standard method is given in Fig. 9.2.

  • 8/3/2019 Unit 9 Atomic Absorption

    5/26

    9

    Atomic Absorption

    Spectrophotometry

    Fig. 9.2: A typical calibration plot for internal standard method in AAS

    If the aspiration fluctuates then each signal is affected to the same extent and the ratio

    at a given analyte concentration remains constant.

    Standard addition method

    As you have learnt in Unit 7, this method is especially applicable when the signal is

    altered by the sample matrix. In this method a known amount of the standard solutionof known increasing concentrations of the analyte is added to a number of aliquots of

    the sample solution. The resulting solutions are diluted to the same final volume and

    their absorbances are measured. A graph is drawn between the absorbance and theadded concentrations of the analyte. It is then extrapolated to the concentration axis toobtain the concentration of sample solution. If the plot is nonlinear then extrapolation

    is not possible. It is essential to perform blank correction in such a case. The

    calibration plot obtained by using standard addition method is shown in Fig. 9.3.

    Fig. 9.3: Calibration plot for standard addition method indicating Cx as the

    concentration for unknown sample

    AAS is a promising analytical method that is extensively employed for quantitative

    determinations of different elements in wide range of samples. A major disadvantage

    of the AAS measurements is that only a single element can be determined at a time asa separate radiation source is required for each element. However, nowadays modern

    instruments are equipped to undertake multielement determinations. Let us learn

    about the instrumental aspects of atomic absorption spectrometry.

    SAQ 1

    What is the importance of calibration plot in atomic absorption spectrophotometry?

    ..

    ..

    ..

  • 8/3/2019 Unit 9 Atomic Absorption

    6/26

    10

    Atomic SpectroscopicMethods-II 9.3 INSTRUMENTATION FOR ATOMIC ABSORPTION

    SPECTROPHOTOMETRY

    You have learnt above that in AAS the absorption of resonant radiation by groundstate atoms of the analyte is used as the analytical signal. Accordingly, a sourcedelivering the characteristic resonant radiation of the analyte is required along with an

    atom reservoir into which the analyte is introduced and atomised. The basic

    requirements of atomic absorption spectrophotometer instrument are similar to any

    regular spectrophotometer; the sample holder cell being replaced by a flame or someother atomiser. A typical atomic absorption spectrophotometer consists of the

    following components.

    Radiation source Atom reservoir Monochromator Detector Readout deviceA block diagram showing the basic components of an atomic absorptionspectrophotometer is given in Fig. 9.4.

    Fig. 9.4: Schematic diagram of an atomic absorption spectrophotometer showing its

    basic componentsIn a typical flame atomic absorption spectrophotometric determination, the radiationfrom a hollow cathode lamp is made to fall on the sample of the analyte aspirated into

    the flame, where a part of it is absorbed. The transmitted radiation is then dispersed

    by a monochromator and sent to the detector. The detector output is suitablyprocessed and is displayed by appropriate readout device. These single channel

    instruments can perform measurements at a single wavelength only in one channel.

    Nowadays, dual-channel instruments are also available that permit simultaneousmeasurements at two different wavelengths. These contain two independentmonochromators for the purpose. Thus, these can be used for the simultaneous

    determination of two elements; one can be the analyte to be determined and the other

    may be a reference element. You would learn about different types of AASinstruments in Section 9.5.

    Let us learn about various components of an atomic absorption spectrophotometer.

    9.3.1 Radiation SourcesAll commercially available atomic absorption spectrophotometers use a radiation

    source that emits the characteristic spectrum of the element to be determined. Theessential requirement of the radiation source is that it gives a constant and intense

    output. Generally two types of sources are in use: line sources and continuum

    sources. Initially continuum sources were used and the primary radiation required

  • 8/3/2019 Unit 9 Atomic Absorption

    7/26

    11

    Atomic Absorption

    Spectrophotometrywas isolated with a high resolution monochromator. However, these had low radiant

    densities and did not provide sufficiently high sensitivity. Nowadays, the hollow-cathode lamp (HCL), belonging to the first type, has been most widely used. The

    electrodeless discharge lamps(EDL) - another line sourcesare also frequently

    employed for the purpose and in fact are superior for the elements such as As, Se andTe with low melting points. You have learnt about HCL and EDL in the context of

    atomic fluorescence spectrometry in Unit 8.

    When the bandwidth of the primary radiation is low with respect to the profile of the

    analyte absorption, a given amount of analyte would absorb more radiation.Therefore, the radiation sources having low widths of the emitted analyte lines arepreferred. Accordingly, the radiation sources are designed so as to operate at muchlower temperatures and pressures as compared to that of the flame and furnaces used

    for atomisation. As a consequence, the emitted lines are much sharper than the

    absorption lines to be measured. In such a set up, sufficient accurate measurements of

    the peak absorption can be made without using elaborate optics. It is referred to assource resolution. Fig. 9.5 illustrates source resolution achieved by using radiationsources emitting sharper lines.

    Fig. 9.5: A schematic diagram illustrating the source resolution achieved by usingradiation sources emitting sharper lines

    9.3.2 AtomisersThe purpose of atomiser is to provide a representative portion of the analyte in theoptical path and convert it into free neutral ground state atoms. In atomic absorptionspectrophotometry, the flames and furnaces that generate a temperature in the range

    of 1500 to 3000 C are the most common methods of atomisation. Two common types

    of atomisers used for generating atomic species in the vapour phase are flame

    atomisers and electrothermal atomisers. Let us learn about flame atomisers. You willlearn about electrothermal alone use in flameless atomic absorption spectrum.

    Flame atomiser

    You have learnt about flame atomisers in Unit 7. In a typical flame atomisationprocess, the analyte solutions are generally nebulised with the help of a nebuliser (see

    Sec. 7, Unit 7) into a spray chamber. The aerosol so produced along with a mixture ofa burning gas and an oxidant is directed into a suitable burner. As already described inUnit 7, Section 7.5, flame temperature depends on fuel-oxidant ratio and the requisite

    temperature for analysis can be obtained by varying the fuel-oxidant ratio. The fuel-

    oxidant combinations commonly used in AAS, the corresponding combustion

    reactions and the flame temperatures are given in Table 9.1.

    The availability of narrow

    band and tunable laser

    sources have opened upnewer areas of application

    of the technique.

  • 8/3/2019 Unit 9 Atomic Absorption

    8/26

    12

    Atomic SpectroscopicMethods-II Table 9.1: The fuel-oxidant combinations commonly used in AAS, the

    corresponding chemical reaction and the corresponding flame

    temperatures

    Fuel-oxidant Combustion reaction* Flame

    temperature

    (K)

    C3H8 air C3H8+5O2+20N2 3CO2+4H2O+20N2 2267

    H2 air 2H2+O2+4N2 2H2O+4N2 2380

    C2H2 air C2H2+O2+4N2 2CO+H2+4N2 2540

    H2 O2 2H2+O2 2H2O 3080

    C3H8 O2 C3H8+5O2 3CO2+4H2O 3094

    C2H2 N2O C2H2+5N2O 2CO2+H2O+5N2 3150

    C2H2 O2 C2H2+2O2 2CO2+H2 3342

    *N2 is included in air just to show its stoichiometry.

    While analysing liquid samples, flame is considered to be superior in terms of

    performance characteristics and reproducible behaviour though sampling efficiency

    and sensitivity of other methods are better. This is because large amount of the

    sample flows down the drain and the residence time of individual atoms in the pathlength of flame is of the order of ~0.1ms. The region of maximum absorption is

    restricted to specific areas of the flame.

    Concentration of atoms may vary widely if the flame is moved relative to the light

    path either vertically or laterally from the resonance line source. The position of

    observation in the flame and the fuel-oxidant ratio must be suitably optimised foreach element in AAS. The fuel-oxidant ratio and observation heights are so chosen asto provide the maximum number of free atoms while minimising interferences from

    emission, ionisation or compound formation.

    Burners

    Two major types of nebuliser burners used in AAS are premix nebuliser-burner

    system and total consumption burner. You have learnt about these in the context of

    flame photometry in Unit 7. You would recall that in premix type burner, liquid issprayed into a mixing chamber where the droplets are mixed with the combustion gas

    and are sent to the burner. Fig. 9.6 gives a schematic drawing of such a burner used inAAS.

    On the other hand, in the total consumption burner, the nebuliser and burner are

    combined. This is also called turbulent flow burner (Fig. 7.1). Several factors are

    involved in the choice of a burner. Generally speaking, a premix burner is preferredfor atomic absorption work, except when a high burning-velocity flame must be used.Turbulent flow burners are widely used for atomic emission measurements about

    which you would learn in the next unit.

    The flame atomisation method discussed above is used more often than the others. As

    many as 67 elements can be determined by employing simple, easy to use,

  • 8/3/2019 Unit 9 Atomic Absorption

    9/26

    13

    Atomic Absorption

    Spectrophotometryequipments based on flame atomisation. It is facilitated by the usage of sharp line

    light sources that result in the ease and reliability in selection of resonance lines.More so, the method is rapid and gives a precision of as high as 0.1 % in some

    cases.

    Fig. 9.6: Schematic diagram of premix nebuliser burner system used in AAS

    Flame atomisation, however, has following disadvantages.

    i) Only about 10% of the nebulised sample reaches the flame and it is then furtherdiluted by the fuel and oxidant gases so that test material has very smallconcentration in the flame.

    ii) A minimum sample volume of 0.5-1.0 mL is needed to give a reliablemeasurement.

    iii) Viscous samples such as blood, serum, oils etc require dilution with a solvent.In order to avoid such problems, nonflame methods involving electrical heating havebeen developed for atomisation about which you would learn in the next section.

    9.3.3 MonochromatorsYou know that the monochromators are the devices that can selectively provideradiation of a desired wavelength out of the range of wavelengths emitted by the

    source or emitted by analyte sample. In AAS, the monochromators select a given

    emission line and isolate it from other lines due to molecular band emissions and allnon absorbed lines. Some of these lines originate from the filler gas in the hollow

    cathode lamp while some others are the spectral emissions of various samplecomponents during atomisation. Most commercial AAS instruments use diffractiongratings as monochromators.

    9.3.4 DetectorsAs the wavelengths of resonance lines fall in UV region, the most commonly useddetector in atomic absorption spectrophotometry is photomultiplier (PM) tube whose

    output is fed to a readout system. The radiation received by the detector may originate

    The flames are not idealatomisers as for a number

    of elements theatomisation is not

    quantitative. The

    sensitivity is loweredconsiderably due to this

    and by dilution of theanalyte atom population

    with gases in the flame.

  • 8/3/2019 Unit 9 Atomic Absorption

    10/26

    14

    Atomic SpectroscopicMethods-II

    not only from the selected resonance lines but also from the emission within the

    flame. Therefore, in addition to absorption signal intensityIA, the detector mayreceive signal intensity of (IA + S) where S is the intensity of emitted radiation from

    flame. Actually one requires only the signal intensity due to absorption, it is therefore,

    important to eliminate effects due to flame emission. This is achieved by modulatingthe emission from the emission line source using a mechanical chopper device.

    9.3.5 Readout DevicesThe readout systems include meters, chart recorders and digital display meters. Thesedays, however, microprocessor controlled systems are commercially available where

    everything can be done by touch of a button. Modern instruments provide a fast

    display of the experimental conditions, absorbance data, statistical values andcalibration curves, etc.

    SAQ 2

    Name the line sources employed in atomic absorption spectrophotometry.

    ..

    ..

    ..

    ..

    ..

    ..

    9.4 GRAPHITE FURNACE ATOMIC ABSORPTIONSPECTROPHOTOMETRY

    As mentioned earlier, the flame atomisation method suffers from some drawbacks. Inorder to overcome these problems some flameless methods of AAS have been

    developed. Two types of flameless atomisers are generally used. These are graphite

    tube or LVov furnace and the carbon rod or filament. AAS using graphite furnaceis called GraphiteFurnace Atomic Absorption Spectrophotometry (GFAAS)

    which is highly sensitive (100 to 1000 times as compared to flame AAS) and requiresa very small sample size. It has a further advantage of not requiring any samplepreparation. So much so that solid samples do not require sample dissolution. Let us

    learn about the electrothermal or flameless atomic absorption spectrophotometry.

    The basic principle of flameless AAS is similar to flame AAS. The analyte isconverted into vaporised atoms in ground state that are subjected to the characteristicresonance radiation emitted by a line source. The absorption of radiation and its

    extent form the basis of analytical applications. As regards the instrumental aspects,the two techniques are similar to a good extent, the difference being in the atomiser

    and the atom reservoir. Rest of the components of the instrument are the same,

    however, a faster electronics is required to process the rapidly obtained transientsignal in GFAAS. Let us learn about the electrodeless or electrothermal atomisers.GFAAS is also termed as electrothermal AAS because of the electrothermal atomisers

    used.

    9.4.1 Electrothermal AtomisersThe use of furnaces as atomisers for quantitative AAS goes back to the work of

    LVov, which led to the breakthrough of atomic absorption spectrophotometry

    towards very low absolute detection limits. The essence of LVov method is to

  • 8/3/2019 Unit 9 Atomic Absorption

    11/26

    15

    Atomic Absorption

    Spectrophotometrycompletely vaporise a small amount of analyte sample in a graphite tube furnace and

    obtain a good concentration of the analyte species in the vapour phase which can besuitably determined. In electrothermal AAS, graphite or metallic tube or cup furnaces

    undergo resistive heating to attain temperatures required for complete atomisation of

    the analyte. For volatile elements this can be accomplished at temperatures of 1000 Kwhereas for more refractory elements the temperatures should be up to 3000 K. Let us

    learn about the working of graphite furnace as an atomiser.

    Graphite furnace

    A graphite furnace consists of a hollow graphite cylinder having a length of about5 cm and diameter of about 1 cm. The tube is surrounded by a metal jacket through

    which water is circulated and remains separated from the tube by a gas space where

    an inert gas such as argon or nitrogen is circulated as schematically shown in Fig. 9.7.A small amount of analyte sample solution (1-100 L) is introduced in the sample

    cell or holder by inserting the tip of micropipette through a port in the outer jacket,and into the gas inlet orifice in the centre of the graphite tube. Alternatively the

    powered analyte sample (about 10-500 g) is introduced directly into the graphitetube.

    Fig. 9.7: Schematic diagram of the cross section of an electrically heated graphite

    furnaceThe nature and design of cuvettes or sample holder is of great importance in GFAAS.

    Different types of cuvettes are commercially available. The standard cuvette madefrom electrographite is suitable for the determination of volatile elements such as Pband Cd. Extended lifetime cuvettes can sustain faster heating rate and have longer

    lifetime and are especially useful in the determination of refractory elements.

    The graphite tube is heated by the passage of an electric current to a temperaturecapable of evapourating the solvent from the solution. The current is then increased in

    such a way that initially the sample is ashed and then ultimately it is vaporised

    producing metal atoms. In other words, a heating cycle as shown in Fig. 9.8(a) isfollowed. For reproducibility, the temperatures and the time of the drying, ashing and

    atomisation process are carefully selected depending on the metal to be analyzed.

    The radiation from the line source is passed through the central hollow graphite tubecontaining the vaporised analyte. The absorption signal produced by this method is a

    transient one and lasts for a few seconds Fig. 9.8(b). The signal will be obtained onlywhen the analyte is atomised. You may note the correspondence between theatomization step and the analyte signal. This can be recorded on a suitable chart

    recorder. This is in contrast with the flame atomisation technique wherein a steady

    absorption signal is obtained. Each graphite tube can be used for 100-200 analysesdepending upon the nature of material.

  • 8/3/2019 Unit 9 Atomic Absorption

    12/26

    16

    Atomic SpectroscopicMethods-II

    (a) (b)

    Fig. 9.8: (a) A schematic heating cycle profile for graphite furnace (b) the transient

    absorption signal obtained by GAAFS

    The sensitivity of GFAAS is much higher as compared to the flame AAS and the

    detection limits are lower by 2-3 orders of magnitude than that in flame AAS. This isso because in the furnace a much higher concentration of atomic vapour can be

    maintained as compared with flames. Furthermore, in this method, the dilution of theanalyte by the solvent is avoided as the solvent is evapourated before the atomisation

    step.

    9.4.2 Handling Background Absorption in GFAASHigh background absorption is a problem area in furnaces. This may be solved by

    diluting the sample or selecting another resonance wavelength line. The use of matrix

    modifier is a commonly acceptable method used to reduce background effects. In thismethod a reagent is added to the sample that may modify the matrix behaviour and

    thereby tackle the problem of background. Sometimes the added matrix may modify

    the analyte also. Following are the reasons for adding matrix modifier.

    It stabilises the analyte during the ashing stage. It converts the interfering matrix into a volatile compound that may be removed

    during ashing.

    It helps to obtain isothermal conditions in the graphite tube by delaying theanalyte atomisation.

    9.4.3 Advantages and Disadvantages of GFAASThe most significant advantages of this flameless vaporisation method are:

    It eliminates the possibility of the interaction of the sample with differentcomponents of the flame, thereby eliminating anomalous results.

    The longer residence time for the analyte in the path of incident radiation leadsto a greater sensitivity.

    As a higher proportion of the analyte sample is converted into vapours, thesensitivity is further enhanced.

    It provides an ability to deal with very small sample sizes. This becomes quiteimportant in the context of clinical samples.

    However, it has the following disadvantages too.

    The background absorption effects are more serious. Analyte may be lost during ashing especially for the volatile compounds. The sample may not be completely atomised and it may produce memory

    effect within the furnace.

    Memory effect refers tothe contributions from the

    remains of the previousdeterminations.

    While using GFAAS,

    great care is requiredduring sample preparation

    because of contaminationarising out from

    glassware and volumetricpipettes.

  • 8/3/2019 Unit 9 Atomic Absorption

    13/26

    17

    Atomic Absorption

    Spectrophotometry Precision is poorer than in flame AAS. However, furnace auto samplers have

    enhanced the precision of furnace AAS.

    Problems due to interferences and high background may become serious.Before proceeding further try to answer the following SAQ.

    SAQ 3

    What do you understand by heating cycle in the context of graphite furnace?

    ..

    ..

    ..

    .

    9.5 ATOMIC ABSORPTION SPECTROPHOTOMETERHaving learnt about the essential components of an atomic absorptionspectrophotometer let us now learn about the different types of atomic absorption

    spectrophotometers. You would recall the spectrophotometers used in UV-VISspectrophotometry. It is pertinent to mention here that the atomic and molecular

    absorption spectrometries are quite similar in principle and instrumentation; theradiation source and the sample holder being different. Let us learn about the two

    types of atomic absorption spectrophotometers given below.

    Single beam atomic absorption spectrophotometer Double beam atomic absorption spectrophotometer9.5.1 Single Beam Atomic Absorption SpectrophotometerA simplified sketch of a single beam flame atomic absorption spectrophotometer is

    shown in Fig. 9.9. It consists of hollow cathode lamp (HCL), a radiation source, flame

    as an atomisation device, a monochromator, a photomultiplier detector and a

    recording system. The HCL radiation is chopped to eliminate the background signalarising from the radiation emitted by the sample itself, focused on the atomic vapour

    produced by the atomiser and then directed to the monochromator where the atomicline of interest is isolated.

    Fig. 9.9: Schematic diagram of a single beam flame atomic absorption

    spectrophotometer

  • 8/3/2019 Unit 9 Atomic Absorption

    14/26

    18

    Atomic SpectroscopicMethods-II

    The electronic amplifier is synchronised with the chopper so that the signal

    component generated by emission from the sample is not detected. The attenuation ofthe source radiation by the analyte atomic vapour is detected by the photomultiplier.

    A blank is aspirated into the flame and the transmittance is adjusted to 100%.

    9.5.2 Double Beam Atomic Absorption SpectrophotometerIn this instrument the beam from the HCL is split by a mirrored chopper, one halfpassing through the atomiser and the other half around it, as schematically shown in

    Fig. 9.10. The two beams are then recombined by a half-silvered mirror and passedthrough the monochromator. The ratio between the reference and sample signal is

    then amplified and fed to the readout display and recorder. These instruments correct

    the fluctuations in the intensity of radiation coming from the radiation source and forchanges in the sensitivity of the detector. It must be noted that reference beam in

    double beam instruments does not pass through the flame and thus corrects for theloss of radiant power due to absorption or scattering by the flame itself.

    Fig. 9.10: Schematic diagram of a double beam atomic absorption spectrophotometer

    Changes in peak shape or position can be indicative of interference problems. Mostinstruments are equipped with a device for background correction. These days

    simultaneous multielemental atomic absorption spectrophotometers with line source

    for 2 to 10 elements have become available. The most successful design of asimultaneous multielement AAS is based on continuum source and a multichannel

    direct reading spectrophotometer.

    Modern atomic absorption spectrophotometers generally have the following features:

    These have a lamp turret capable of holding at least four hollow cathode lampsemitting the absorption lines for different elements. These have an independent

    current stabilised supply for each element.

    The sample area is capable of incorporating an autosampler which can workwith both flame and furnace atomisers. Improved analytical precision isobtained when an autosampler is used in conjunction with a furnace atomiser.

    The monochromator is capable of high resolution typically 0.04 nm, a featuremore desirable if the AAS is adapted for flame emission work though good

    resolution is also desirable for many elements in AAS.

    The photomultipliers are able to function over a wide wavelength range of180 800 nm.

    The instruments are equipped with a background correction facility.

  • 8/3/2019 Unit 9 Atomic Absorption

    15/26

    19

    Atomic Absorption

    Spectrophotometry The instruments have an integral video screen facility for the ease of operation.

    A modern software package includes help facilities, full graphical data

    presentation, complete data storage and flexible data generation. The softwaremust optimise flame and spectrometer parameters including furnacetemperature.

    SAQ 4

    In what way is a double beam atomic absorption spectrophotometer better than a

    single beam spectrophotometer?

    ..

    ..

    ..

    ..

    9.6 INTERFERENCES IN ATOMIC ABSORPTIONSPECTROPHOTOMETRY

    All spectrometric methods have associated interferences that need to be addressed to

    so as to put the technique to analytical use. We may define an interference to be achemical, physical or spectral effect that may cause the analyte signal to be altered inintensity. Accordingly, there are three types of interferences in AAS. These arespectral, chemical and physical interferences. Let us learn about these in the contexts

    of flame AAS and GFAAS.

    9.6.1 Spectral InterferencesThese refer to the presence of another atomic absorption line or a molecular

    absorption band close to the spectral line of the analyte element being monitored. Itqualifies to be an interference if it is not resolved by the monochromator. Most

    probable spectral interferences are the ones of the molecular emissions from oxides ofother elements in the sample. In case of AAS, such interferences occur if a dc

    instrument is used and can be eliminated by employing an ac instrument. Similarly apositive interference may occur if an element or molecule is capable of absorbing

    radiation from a continuous source. This may be minimised but not eliminated

    altogether by using a line source.

    Another source of spectral interference is the light scattering or absorption by solid

    particles, unvaporised solvent droplets or molecular species in the flame. This

    problem is significant at wavelengths less than 300 nm when solutions of high saltcontent are aspirated. This arises because of incomplete desolvation and is calledbackground absorption or blank. This can be corrected by measuring the absorbance

    of a line close to the absorption line of the analyte element but not absorbed by theelement itself. The measurements should be made at two other lines from the hollowcathode lamp or a nearby line from a second hollow cathode lamp. Analyte test

    element should always be aspirated to check that it does not absorb the backgroundcorrection line. This technique requires two separate measurements on the sample.

    A yet another source of spectral interference is the background emission from theflame. This may be corrected by modulation of the output of the radiation source andthe ac detection system. Several background correction schemes have been developed

    and incorporated in spectrophotometers such as the deuterium background correction,

    Zeeman correction system and the Smith-Hieftie system. These are not discussedhere, you can obtain information about these from the reference texts listed at the end

    of the block.

    Absorption due to

    molecular species andscattering are more

    problematic withelectrothermalatomization.

  • 8/3/2019 Unit 9 Atomic Absorption

    16/26

    20

    Atomic SpectroscopicMethods-II 9.6.2 Chemical Interferences

    These include interferences due to ionisation, formation of low volatility compounds,

    dissociation, etc. During atomisation in the flame, several reactions occur resulting in

    the formation of analyte compounds which decrease atomic population in the cell.

    Most important chemical interference is due to anions and form atom of compoundsof low volatility from the analyte element. For example, the refractory elements suchas Ti, W, U, V, Mo, Zr and elements like B, Al, and Fe may combine with O and OH

    species in the flame producing thermally stable oxides and hydroxides. Similarly,

    absorbance due to Ca is decreased in presence of phosphate because of the formationof calcium phosphate having low volatility.

    Such interferences can be avoided by increasing the flame temperature whence these

    interfering compounds are decomposed. In some cases chemical interferences may beeliminated by using a releasing agent that react with interfering species and avoids its

    reaction with the analyte element. For example, in the determination of Ca, Sr and Lacan be used as releasing agents to minimise phosphate interference as these would

    react preferentially with the phosphate.

    9.6.3 Physical InterferencesThese are independent of the analyte type and have nearly the same effect on

    emission, absorption and fluorescence with a given type of atomiser. These may bedue to variations in the gas flow rates, changes in the solution viscosity affecting itsrate of aspiration into the flame which may finally change the atomic concentration in

    the flame. Viscosity of standards and samples may be different if the sample contains

    organic solvent or a high concentration of the salt which is not the case with thestandard. Errors due to changes in viscosity may be avoided by matching the matrix

    and by performing frequent calibrations. Some instruments offer the capability ofusing internal standards that can partially compensate for changes in physical

    parameters including flame temperature.

    SAQ 5

    a) How does phosphate interfere in the quantitative determination of calcium byatomic absorption spectrophotometry?

    ..

    ..

    ..

    ..

    b) How is this interference handled?..

    ..

    ..

    ..

    9.7 SAMPLE HANDLING IN ATOMIC ABSORPTIONSPECTROPHOTOMETRY

    In principle, the sample in solid, liquid or in the gas phase can be analysed by flameAAS. However, in most cases, sample analysed by AAS is in the solution form.Therefore, the solid sample is first dissolved and converted into a solution. Solids

  • 8/3/2019 Unit 9 Atomic Absorption

    17/26

    21

    Atomic Absorption

    Spectrophotometrycould be analysed directly also by using an electrothermal furnace. The gaseous

    samples, on the other hand, generally are pretreated by scrubbing before the resultantsolution. Alternatively, the gases may be adsorbed on a solid surface and then leached

    into solution with suitable reagents. Let us learn how the dissolution of the solid

    sample, an important step in AAS, is carried out.

    9.7.1 Preparation of the SampleThe choice of proper reagents and techniques of decomposition and dissolution of the

    sample is a critical step for the success of AAS determination. This is often done byacid digestion, which produces a clear solution without loss of any of the elements to

    be determined. It is therefore, essential that all the reagents and solvents used in wet

    decomposition should be of highest purity as any impurity may raise the blank value.Common acids used for dissolution are HCl, HNO3, aqua regia (HCl : HNO3 :: 3:1) or

    perchloric acid (HClO4) which dissolve most of the inorganic materials. For thedecomposition of silicate materials, however, HF must be used. A combination of

    nitric acid and perchloric acid is especially useful for the complete destruction of fatsand proteins in biological samples.

    In a typical dissolution step, a suspension of the sample in acid is heated by flame or ahot plate until complete dissolution i.e., when the entire solid has disappeared and atransparent solution is obtained. The decomposition temperature is the boiling point

    of acid. However, such a wet decomposition in open vessels may give rise tosystematic errors due to volatilisation losses and contamination caused by thereagents and container material, and loss of elements caused by adsorption on the

    vessel surface. More so, sometimes the dissolution may not be complete and it may

    also cause errors. In order to avoid such errors, wet decomposition methods in closedsystems have been developed. These have the following advantages.

    There are no volatilisation losses. These have a shorter reaction time and improved decomposition due to high

    temperatures.

    The blank values are low. These do not have contamination from external sources.If the concentration of the elements to be determined is too high, then the solution

    must be diluted quantitatively before commencing the absorbance measurements.

    Conversely, if the concentration of the metal in the test solution is too low, aconcentration procedure such as solvent extraction or ion-exchange must be followed.

    While analysing halogens and some other elements like S, Se, P, B, Hg, As and Sb, it

    is advantageous to use combustion in an oxygen flask. The combustion is carried outin a sealed container and the reaction products are absorbed in a suitable solvent.

    Metals and alloys can usually be dissolved in acids, whereas dissolution of glass

    requires alkaline or acid fusion. Generally speaking, the final solution of the analyte

    should not contain acid concentration more than about 1Mor else aspiration ofcorrosive solution may damage the burner.

    9.7.2 Use of Organic SolventsIn the early development stages of AAS, it was observed that analyte solutionscontaining organic solvents of low molar mass e.g. alcohols, ethers, ketones and

    esters enhanced absorption peaks. This has been attributed to the increased rate of

    aspiration, nebulisation efficiency, formation of finer droplets, and more efficientevaporation or combustion of the solvent. In favourable cases, up to three fold

    increase in sensitivity could be obtained by adding a miscible organic solvent such as

    Sample preparation is a

    crucial step in the AASdetermination.

  • 8/3/2019 Unit 9 Atomic Absorption

    18/26

    22

    Atomic SpectroscopicMethods-II

    acetone to the solution. However, this makes the sample solution more dilute which

    more or less defeats the purpose of achieving enhanced sensitivity.

    Therefore, to obtain increased sensitivity, the technique of solvent extraction is

    usually employed. The metal extracted into the organic phase is directly aspirated intothe flame. This method has following advantages.

    The analyte element is separated from the bulk matrix of the sample therebyeliminating chemical interferences.

    Complexation of metal with organic solvent increases atomisation efficiencycausing upto ten fold signal enhancement.

    The analyte element may be extracted into a smaller volume of organic solventwith 10

    to 100 fold gain in concentration. Methylisobutyl ketone (MIBK) is an

    ideal solvent which is easily aspirated into the flame.

    While using an organic solvent, flame should be adjusted before aspirating the solventwhich must be burned along with the fuel. If the flame is too rich in fuel, the solventwill not be burnt resulting in smoky flame. Thus fuel-oxidant ratio may be adjusted

    while using organic solvent to offset the presence of organic solvent. Solvent should

    be aspirated between samples because the hot lean flame will heat up the burner.However, the lean mixture results in lower flame temperature, thus increasing the

    possibility of chemical interferences. Therefore, suitable safety procedure must befollowed while using organic solvents.

    9.7.3 Microwave DigestionIn another method of preparing sample for AAS determination, microwave radiations

    are employed. A microwave digestion system (MDS) offers more rapid and efficient

    decomposition of complex matrices of geological and biological samples. Theconcept of microwave ovens for the decomposition of inorganic and organic samples

    was first proposed during mid 1970s. This method has an advantage overconventional methods as it takes less time because of rapid heating ability ofmicrowaves.

    In contrast to conventional flame/hot plate heating method based on conduction,microwave energy is directly transferred to all the molecules of solution almost

    simultaneously without heating the vessel and thus boiling temperature is reachedvery quickly due to increased pressure in the vessel. In addition, small amounts of

    reagents are used and evaporative losses are avoided, thus reducing interferences byreagent contamination. As MDS can be easily automated it greatly reduces the

    operator time to prepare samples for analysis. These days multi-vessel MDS with 4, 6

    or 8 vessels are commercially available where more number of samples can besimultaneously dissolved.

    Microwave digestion vessels are constructed from low-loss materials that aretransparent to microwave radiation. Teflon is an ideal material for many of the acids

    including HF commonly used for dissolution. Not only it is transparent to microwaves

    but it has low melting point of ~300 oC which is of course lower than boiling point ofH2SO4 and H3PO4. For these acids, quartz and borosilicate glass vessels are used.A typical closed vessel microwave digestion system consisting of teflon body, capand a safety relief valve is shown in Fig. 9.11. The maximum recommended

    temperature obtained with this device is 250 C. When overpressurisation occurs,safety valve gets distorted similar to home pressure cookers and the excess pressure is

    released.

    Microwave region ofelectromagnetic spectrum

    corresponds to lower

    energy (or longer

    wavelength) than IRradiation. Most

    commonly used

    frequency is 2450 MHz asset by international

    convention for use forindustrial, scientific and

    medical purposes.

  • 8/3/2019 Unit 9 Atomic Absorption

    19/26

    23

    Atomic Absorption

    Spectrophotometry

    Fig. 9.11: A schematic diagram of closed vessel microwave digestion system

    Commercially available MDS incorporates corrosion protection for the interior with

    variables such as sample mass (0.1-2g), digestion acids (HCl, HF, HNO3 and H3BO4),

    power setting and heating time (1-20min), etc.

    9.7.4 Sample Introduction MethodsThe sample introduction into the flame is an important step in flame AASmeasurements as its accuracy, precision and detection limits depend on how the

    analyte sample is introduced. The aim of a sample introduction system is to transfer a

    reproducible and representative portion of a sample into an atomiser. It depends onthe physical and chemical state (i.e. solid, liquid or gas) of the analyte and the sample

    matrix such as soil, water, blood, plant leaves, etc. For solution and gaseous samples,

    the introduction step is quite simple but for solid, it poses a major problem.

    You have learnt in Unit 7 that a simple method of sample introduction into the flame

    is nebulisation. It is a process of thermal vaporisation and dissociation of aerosol

    particles at high temperatures thus producing small particle size with high residencetime. Some nebulisation methods are given below.

    pneumatic nebulisation ultrasonic nebulisation electrothermal vaporisation hydride generationYou have learnt about pneumatic nebulisation and the nebulisers in Unit 7; let us

    learn about the other three types of nebulisers.

    Ultrasonic nebulisation: In this case the sample is pumped on to the surface of apiezoelectric crystal that vibrates at a frequency of 20 kHz to a few MHz (Fig. 9.12).

    The waves so produced are very efficient in turning the sample into a fine aerosol,

    which is carried by a stream of argon, first through a heated tube and then to arefrigerated tube to condense out the solvent. Such nebulisers produce more dense

    and homogeneous aerosols because of desolvation and there is no cooling effect. Thisimproves detection limit by a factor of 10 to 20 as compared to pneumatic nebulisers.

  • 8/3/2019 Unit 9 Atomic Absorption

    20/26

    24

    Atomic SpectroscopicMethods-II

    Fig. 9.12: Schematic diagram of an ultrasonic nebuliser

    Electrothermal vaporisation: An electrothermal vaporiser (ETV) is an evaporator

    located in a closed chamber through which an inert gas such as argon flows to carrythe vaporised sample into the atomiser. The sample can be vaporised from an ETV on

    a conductor such as carbon rod or a tube furnace or a heated metal filament

    commonly used in AAS. A schematic diagram of a Lvov platform used aselectrothermal vaporiser is given in Fig. 9.13. In contrast to the nebuliserarrangements, ETV system produces a discrete signal rather a continuous one so that

    signal from the atomised sample increases to a maximum and then decreases to zero

    as the sample is swept through the atomiser. Each preheated electrode is introducedinto the aperture of a preheated cuvette and heated by means of a separate power

    supply. It enables 1-200 L samples to yield approximately an order of improvement

    in detection limit in the range 0.1-500 pg. The major problems of electrothermalatomisation system are interferences due to sample matrix.

    Fig. 9.13: Schematic diagram of Lvov platforman electrothermal vaporiser in a

    graphite furnace

    Hydride generation technique: This provides a method for introducing samples

    containing As, Sb, Sn, Se, Bi and Pb into an atomiser as their representative hydrides.This enhances detection limits by a factor of 10 to 100. As some of these elements are

    highly toxic and occur in the environment at very low levels, their determination at

    low concentrations is extremely important. The volatile hydrides of these elementsmay be generated by the reaction of an acidified aqueous solution of the sample to

    1% aqueous solution of NaBH4 in a glass vessel. A representative reaction of As (III)

    with NaBH4 to form arsine (AsH3) is as follows.

    3NaBH4+3HCl+4H3AsO3 3H3BO3+4AsH3+3H2O+3NaCl

    The volatile hydrides such as AsH3, BiH3, SbH3, H2Se etc. are swept out of thesolution into the atomisation chamber by an inert gas carrier. The chamber is usually

    a silica tube heated in a tube furnace or in a flame where hydride gets decomposedleading to the formation of analyte element whose concentration is then determined

  • 8/3/2019 Unit 9 Atomic Absorption

    21/26

    25

    Atomic Absorption

    Spectrophotometryfrom the atomic absorption signal. The signal is a peak similar to that obtained with

    electrothermal atomisation. Schematic diagram of basic system used for the hydridegeneration and atomisation is shown in Fig. 9.14.

    Fig. 9.14: Schematic diagram of a basic system for hydride generation technique

    Commercial hydride generation system use either electrically heated or flame heated

    quartz tube for atomisation. Its main advantage is enhanced sensitivity and freedomfrom matrix interferences as the element is separated from all other accompanying

    elements. In Table 9.2 are given the comparison of detection limits by hydride

    generation and graphitic furnace AAS.

    Table 9.2: Comparison of detection limits of for selected elements by hydride

    generation AAS with graphite furnace AAS

    Limit of Detection (g/L)Element

    Hydride generation Graphite furnace

    As 0.01 0.3

    Sb 0.02 0.2

    Bi 0.02 0.2

    Se 0.01 1.0

    Sn 0.04 0.2

    SAQ 6

    What is the principle of ultrasonic nebuliser?

    ..

    ..

    ..

    ..

    9.8 APPLICATIONS OF ATOMIC ABSORPTIONSPECTROPHOTOMETRY

    Atomic absorption spectrophotometry (AAS) is now a routinely and widely employedtechnique for trace and ultratrace analysis of complex matrices of geological,

  • 8/3/2019 Unit 9 Atomic Absorption

    22/26

    26

    Atomic SpectroscopicMethods-II

    biological, environmental, industrial, glass, cement, marine sediment, pharmaceutical,

    engine oil or any other kind of samples. It has been employed for the determination ofmore than 60 elements at trace and ultratrace levels. It is frequently used for the cases

    where the sample size is small e.g. in case of metalloproteins.

    Accuracy in AAS method is generally limited by random errors and noise to about

    0.5 5%. Spectral and chemical interferences may however cause systematic errors.

    Precision of AAS measurements is typically 0.3 1% at absorbance larger than 0.1 or

    0.2 for flame atomisation and 1 5% with electrothermal atomisation. The detection

    limits and sensitivities provide a means of comparing characteristics of AAS for agiven element. The detection limits of the AAS method lie in the range of ppb; theGFAAS giving better detection limits as compared to the flame AAS. The detectionlimits of the two methods along with the resonance lines for some commonly

    determined elements is compiled in Table 9.3.

    The data in Table 9.3 gives only representative detection limits which may vary withthe analyte matrix, nebulisation conditions, flame temperature, sample path length,positioning of burner and other factors including interferences. You would learn in

    details about the applications of atomic absorption spectrometry in Unit 11 along withthat of atomic emission spectrometry.

    Table 9.3: The resonance lines and approximate detection limits of some selected

    elements by flame AAS and GFAAS

    Detection Limit (ppb)Element Resonance

    line (nm)

    Air-acetylene

    flame

    Graphite furnace

    Ag

    Ba

    328.1

    553.6

    0.9

    8

    0.005

    0.1

    Ca 422.7 2 0.3

    Cd 228.8 5 0.1

    Cr 357.9 5 0.5

    Cu 324.5 4 0.1

    Fe

    Hg

    248.3

    253.7

    4

    200

    3

    1

    K 766.5 4 1

    Mg

    Mn

    Na

    285.2

    279.5

    589.0

    3

    1

    0.2

    0.05

    0.1

    0.05

    Ni

    Sb

    Ti

    232.0

    217.6

    364.3

    5

    30

    .0.2

    1

    0.2

    0.05

    Zn 213.9 1 0.006

    As AAS is a sensitivetechnique, all possiblesources of contamination

    such as due to storage

    containers, impurities inreagents, and solvents and

    incomplete removal of

    earlier sample from the

    nebuliser system shouldbe avoided.

  • 8/3/2019 Unit 9 Atomic Absorption

    23/26

    27

    Atomic Absorption

    SpectrophotometryMerits and Limitations of Atomic Absorption Spectrophotometry

    Some of the merits of atomic absorption spectrometry are as given below.

    The equipment is easy to use It is a robust technique The techniques has a small turn around time; of the order of few seconds Moderate cost of analysis per sample Low detection limitsSome limitations of AA spectrophotometry are given below.

    Requirement of furnace for the analysis of refractory elements Use of flammable gases Non-automated analytical procedureLet us summarise what have we learnt in this unit.

    9.9 SUMMARYAtomic absorption spectrophotometry (AAS) concerns the absorption of radiation by

    the atomised analyte element in the ground state. The atomisation is achieved by thethermal energy of the flame or electrothermally in an electrical furnace. Thewavelength(s) of the radiation absorbed and the extent of the absorption form the

    basis of the qualitative and quantitative determinations respectively. As atomic

    absorption spectrophotometry is not an absolute method of analysis, the routineanalytical methodology for quantitative determinations using AAS is based on

    calibration method. Besides this the internal standard method and standard additionmethods are also employed.

    A typical atomic absorption spectrophotometer consists of a source delivering the

    characteristic resonant radiation of the analyte, an atom reservoir into which the

    analyte is introduced and atomised, a monochromator, a detector and a readoutdevice. In a typical flame atomic absorption spectrophotometric determination, the

    radiation from a hollow cathode lamp (or electrodeless discharge lamp) is made to fallon the sample of the analyte aspirated into the flame (or in the cuvette of a Lvovgraphite furnace), where a part of it is absorbed. The transmitted radiation is then

    dispersed by a monochromator and sent to the detector. The detector output is suitably

    processed and is displayed by appropriate readout device. Like, UV-VISspectrophotometers the atomic absorption spectrophotometers are also of two typesviz., single beam atomic absorption spectrophotometers and double beam atomic

    absorption spectrophotometers

    GFAAS is a much more sensitive as compared to flame AAS and requires a very

    small sample size. More so, it does not require any sample preparation; even solid

    samples can be analysed without dissolution. However, the background absorptioneffects are quite serious. These are generally sorted out by diluting the sample or

    selecting another resonance wavelength line. In matrix modifier method a reagent is

    added to the sample that may modify the matrix behaviour and thereby tackle the

    problem of background. Sometimes the added matrix may modify the analyte also.

    Three types of interferences viz., spectral, chemical and physical interferences areencountered in AAS. These need to be suitably addressed to so as to put the techniqueto analytical use. Sample preparation is a crucial step in the AAS determination.

    Though in principle, the sample in solid, liquid or in the gas phase can be analysed by

    flame AAS but in practice the sample is taken in the solution form. The solution of

  • 8/3/2019 Unit 9 Atomic Absorption

    24/26

    28

    Atomic SpectroscopicMethods-II

    the solids is generally prepared by wet dissolution method using a suitable acid. The

    presence of organic solvents of low molar mass e.g. alcohols, ethers, ketones andesters are found to enhance absorption peaks and hence increase sensitivity. A

    microwave digestion system (MDS) offers more rapid and efficient decomposition of

    complex matrices of geological and biological samples. It greatly reduces the operatortime to prepare samples for analysis. More so, it can be easily automated also.

    The accuracy, precision and detection limits of flame AAS depend on how the analyte

    sample is introduced into the atomiser. We need to transfer a reproducible and

    representative portion of a sample into an atomiser which depends on the physical and

    chemical state of the analyte and the sample matrix. The sample introduction isachieved with the help of a nebuliser. The commonly used nebulisation methods arepneumatic nebulisation, ultrasonic nebulisation, electrothermal vaporisation andhydride generation.

    Atomic absorption spectrophotometry (AAS) is now a routinely and widely employed

    technique for trace and ultratrace analysis of complex matrices of geological,biological, environmental, industrial, glass, cement, marine sediment, pharmaceutical,engine oil or any other kind of samples. The atomic absorption methods using flame

    are rapid and precise and are applicable to about 67 elements. Electrothermal methodsof analysis on the other hand are slower and less precise; however, these are more

    sensitive and need much smaller samples.

    9.10 TERMINAL QUESTIONS1. Why a sharp line source is required in atomic absorption spectrophotometry?2. Why atomic absorption spectrophotometry is not an ideal method for the

    determination of alkali metals? Which atomic spectrometric method would yourecommend for these elements?

    3. How does the hydride generation method of sample introduction improve thesensitivity of some elements?

    4. What do you understand by matrix modifier? What is its importance?5. Why do furnace atomisers provide enhanced sensitivity over flame atomisers in

    AAS measurements?

    6. Explain the role of organic solvents in atomisation.7. In what way is the signal obtained in GFAAS different from that obtained in

    flame AAS?

    9.11 ANSWERSSelf Assessment Questions

    1. In principle the absorption of radiation in AAS is directly proportional to theconcentration i.e., the Lambert-Beers law holds. However, a number ofexternal factors like the background emission, the flame radiation and other

    types of interferences cause deviations from the law. In such a case a

    dependable determination can be obtained only with the help of a calibrationplot.

    2. The atomic absorption spectrophotometers generally use line sources. Twocommonly used line sources are hollow-cathode lamp (HCL) and electrodelessdischarge lamps (EDL).

  • 8/3/2019 Unit 9 Atomic Absorption

    25/26

    29

    Atomic Absorption

    Spectrophotometry3. In graphite furnace the electrothermal heating is done in three stages. In the first

    stage the temperature is adequate for evapourating the solvent from the samplesolution. In the next stage the sample is ashed and then finally it is vaporised

    producing metal atoms. This three stage heating is called heating cycle.

    4. Double beam atomic absorption spectrophotometers are better than a singlebeam spectrophotometer because these correct the fluctuations in the intensity

    of radiation coming from the radiation source and for changes in the sensitivity

    of the detector.

    5. a) Phosphate interferes in the quantitative determination of calcium as itforms calcium phosphate having low volatility which decreases atomicpopulation in the cell.

    b) The interference of phosphate in the determination of Ca can be managedby using a releasing agent like Sr or La. These act by reactingpreferentially with the phosphate.

    6. The ultrasonic nebuliser uses a piezoelectric crystal which vibrates at extremelyhigh frequencies and the waves so produced turn the sample into a fine aerosol.

    The aerosol is then carried by a stream of argon, first through a heated tube and

    then to a refrigerated tube to condense out the solvent.

    Terminal Questions

    1. When the bandwidth of the primary radiation is low with respect to the profileof the analyte absorption, a given amount of analyte would absorb more

    radiation. Therefore, the radiation sources having low widths of the emittedanalyte lines are preferred. The continuous radiation sources on the other hand

    have low radiant densities and do not provide sufficiently high sensitivity.

    2. Alkali metals have low ionisation energies and even at the low temperaturesobtained in the flame a sufficient amount of the sample may be in the excited or

    ionised state. This means that the concentration of the analyte atoms in theground state will not be a true representation of the analyte concentration. As

    the AAS method depends on the radiation absorption by the analyte in ground

    state it is not ideal for such determinations. Flame photometry would be themethod of choice for such determinations.

    3. The hydride generation method of sample enhances the detection limits by afactor of 10 to 100 by converting the analyte element into a volatile hydride.Some of the elements for which this method can be used are As, Sb, Sn, Se, Bi

    and Pb.

    4. The matrix modifier is a commonly acceptable method used to reducebackground effects in GFAAS. In this method, a reagent is added to the sample

    that acts by modifying the sample matrix and thereby reduces the problem of

    background. In some cases the modifier may modify the analyte also.

    5. The sensitivity of electrothermal AAS is much higher as compared to the flameAAS because in the furnace a much higher concentration of atomic vapour can

    be maintained as compared with flames. Furthermore, in this method, thedilution of the analyte by the solvent is avoided as the solvent is evaporated

    before the atomisation step.

    6. The organic solvents of low molar mass like alcohols, ethers, and ketonesenhance the absorption peaks by increasing rate of aspiration, nebulisation

    efficiency, formation of finer droplets, and more efficient evapouration or

  • 8/3/2019 Unit 9 Atomic Absorption

    26/26

    Atomic SpectroscopicMethods-II

    combustion of the solvent. However, their presence makes the sample solution

    more dilute and the advantage is lost. Therefore, the analyte is extracted into asuitable solvent and the organic phase is directly aspirated into the flame. This

    method increases the atomisation efficiency and eliminates a number of

    chemical interferences.

    7. A transient signal that lasts for a few seconds is produced in GFAAS whereas inflame AAS a steady absorption signal is obtained.