uniformity testing of gamma -...

16
Uniformity testing of gamma cameras at high count-rates NICK BATES UNIVERSITY OF SURREY ROYAL SURREY COUNTY HOSPITAL NATIONAL PHYSICAL LABORATORY UK

Upload: others

Post on 30-Dec-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Uniformity testing of gamma - MRTDosimetrymrtdosimetry-empir.eu/.../05/N-Bates-Optima640Uniformity.pdfy = 0.0002x2 + 0.004x + 8.6525 R² = 0.9936 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5

Uniformity testing of gamma cameras at high count-ratesNICK BATESUNIVERSITY OF SURREYROYAL SURREY COUNTY HOSPITALNATIONAL PHYSICAL LABORATORYUK

Page 2: Uniformity testing of gamma - MRTDosimetrymrtdosimetry-empir.eu/.../05/N-Bates-Optima640Uniformity.pdfy = 0.0002x2 + 0.004x + 8.6525 R² = 0.9936 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5

Background• Imaging patients after therapeutic

administration of radiopharmaceutical causes significant dead-time effects in gamma cameras

• Gamma cameras behave as paralysabledetectors so the count-rate can saturate at high activities

• To deal with this, manufacturers have implemented high count-rate modes. Some switch in automatically, others operate separately from “normal mode”

• GE SPECT/CT cameras have a “fast mode” setting that must be set-up for each radionuclide

Input Count Rate

Ou

tpu

t C

ou

nt R

ate

Gamma camera

Page 3: Uniformity testing of gamma - MRTDosimetrymrtdosimetry-empir.eu/.../05/N-Bates-Optima640Uniformity.pdfy = 0.0002x2 + 0.004x + 8.6525 R² = 0.9936 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5

Background• Prior to carrying out patient imaging, appropriate

Quality Assurance should be completed.• Flood-field uniformity is a standard test of overall

camera performance• Response of the detector to a uniform flux of radiation• Ideally equal image intensity over entire field of view• Non-uniformity indicates detector drift / inaccurate

corrections

• The stability of uniformity with time and activity in “fast mode” is unknown…

• As part of the set-up for the SELIMETRY study, RSCH has investigated the dependence of “fast mode” uniformity on source activity and the stability over time

Page 4: Uniformity testing of gamma - MRTDosimetrymrtdosimetry-empir.eu/.../05/N-Bates-Optima640Uniformity.pdfy = 0.0002x2 + 0.004x + 8.6525 R² = 0.9936 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5

Initial Set-up and Testing• GE engineers enabled “fast mode” after preliminary testing revealed

paralysis of detector in “normal mode” at ~1.5GBq source activity (~25kcps)

• A 60M count uniformity correction map was acquired using a 500MBq point source of I-131 positioned >5xUFOV distance from the uncollimated detector head.• This activity was chosen as it gave a count-rate of ~70kcps • For Tc-99m fast mode, GE recommend acquiring calibrations at 45-

100kcps, so we assumed the same would apply for I-131• Uniformity, energy peak position and FWHM was assessed by

acquiring 20M count flood-field with sources ranging from 50MBq (8kcps) to 700MBq (94kcps).

Page 5: Uniformity testing of gamma - MRTDosimetrymrtdosimetry-empir.eu/.../05/N-Bates-Optima640Uniformity.pdfy = 0.0002x2 + 0.004x + 8.6525 R² = 0.9936 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5

Initial Uniformity Results

y = -0.1675x + 13.844R² = 0.994

0.00

2.00

4.00

6.00

8.00

10.00

12.00

14.00

16.00

0 20 40 60 80 100

CF

OV

In

teg

ral

Un

ifo

rmit

y (

%)

Count rate (kcps)

y = 9E-05x2 + 0.0086x + 8.5638R² = 0.9988

8.4

8.6

8.8

9.0

9.2

9.4

9.6

9.8

10.0

10.2

10.4

360.9

361.0

361.1

361.2

361.3

361.4

361.5

361.6

361.7

361.8

0 20 40 60 80 100

Peak F

WH

M (

%)

Peak P

osit

ion

(keV

)Count Rate (kcps)

Page 6: Uniformity testing of gamma - MRTDosimetrymrtdosimetry-empir.eu/.../05/N-Bates-Optima640Uniformity.pdfy = 0.0002x2 + 0.004x + 8.6525 R² = 0.9936 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5

Initial Uniformity Results• Uniformity degrades linearly either side of the count-rate used for calibrations.• Uniformity becomes clinically unacceptable (>3%) within 10kcps of calibration

count-rate• Peak position does not change significantly with count-rate• Peak FWHM increased with count rate (8.6% at 8kcps to 10.2% at 94kcps)

Page 7: Uniformity testing of gamma - MRTDosimetrymrtdosimetry-empir.eu/.../05/N-Bates-Optima640Uniformity.pdfy = 0.0002x2 + 0.004x + 8.6525 R² = 0.9936 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5

Repeat testing• Suggestion from other hospitals (Oxford) to acquire uniformity map at lower

count-rate• ~140MBq source used ~20kcps• 60M count uniformity correction map as before• Prepared 7 sources (10, 20, 50, 100, 140, 400, 700MBq)

• Acquired individually and in combination to study dependence of uniformity on count-rate in more detail

Page 8: Uniformity testing of gamma - MRTDosimetrymrtdosimetry-empir.eu/.../05/N-Bates-Optima640Uniformity.pdfy = 0.0002x2 + 0.004x + 8.6525 R² = 0.9936 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5

Repeat Testing Results: Uniformity

0

5

10

15

20

25

30

35

40

0 20 40 60 80 100 120 140

Un

ifo

rmit

y (%

)

Count Rate (kcps)

CFOV Int

CFOV Diff

UFOV Int

UFOV Diff

Page 9: Uniformity testing of gamma - MRTDosimetrymrtdosimetry-empir.eu/.../05/N-Bates-Optima640Uniformity.pdfy = 0.0002x2 + 0.004x + 8.6525 R² = 0.9936 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5

Repeat Testing Results: Uniformity

y = -0.0001x3 + 0.014x2 - 0.405x + 6.1714R² = 0.9797

0

2

4

6

8

10

12

0 10 20 30 40 50 60 70

Un

ifo

rmit

y (%

)

Count Rate (kcps)

CFOV Int

Polyn. (CFOV Int)

Page 10: Uniformity testing of gamma - MRTDosimetrymrtdosimetry-empir.eu/.../05/N-Bates-Optima640Uniformity.pdfy = 0.0002x2 + 0.004x + 8.6525 R² = 0.9936 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5

Repeat Testing Results: Peaking

360

360.5

361

361.5

362

362.5

363

363.5

0 50 100 150

Ph

oto

pe

ak p

osi

tio

n (

keV

)

Count Rate (kcps)

y = 0.0002x2 + 0.004x + 8.6525R² = 0.9936

8

8.5

9

9.5

10

10.5

11

11.5

12

12.5

0 50 100 150

FWH

M (

%)

Count Rate (kcps)

Page 11: Uniformity testing of gamma - MRTDosimetrymrtdosimetry-empir.eu/.../05/N-Bates-Optima640Uniformity.pdfy = 0.0002x2 + 0.004x + 8.6525 R² = 0.9936 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5

Observable Changes

Centres of PMTs have lower counts at higher count-rates

Increased count rate

20kcps (140MBq) 100kcps (890MBq)

Page 12: Uniformity testing of gamma - MRTDosimetrymrtdosimetry-empir.eu/.../05/N-Bates-Optima640Uniformity.pdfy = 0.0002x2 + 0.004x + 8.6525 R² = 0.9936 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5

Observable Changes

• Event pile-up towards centre of PMTs at very high rates

• Streaking between centres of PMTs – more prominent horizontally

Page 13: Uniformity testing of gamma - MRTDosimetrymrtdosimetry-empir.eu/.../05/N-Bates-Optima640Uniformity.pdfy = 0.0002x2 + 0.004x + 8.6525 R² = 0.9936 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5

Observable Changes20kcps (140MBq)

Decreased count rate

1.5kcps (10MBq)

Centres of PMTs have higher counts at lower count-rates

Page 14: Uniformity testing of gamma - MRTDosimetrymrtdosimetry-empir.eu/.../05/N-Bates-Optima640Uniformity.pdfy = 0.0002x2 + 0.004x + 8.6525 R² = 0.9936 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5

Possible causes• Events outside energy window contribute to

dead-time• Count-rate dependent spatial linearity

distortions?• Amplifier saturation?• GE factory response:

• Significant scatter downward shift in energy peak…?

• Tests have been conducted with increased event threshold; no effect

• Suggesting adding copper behind source to reduce scatter

• What about patient scatter?!

Page 15: Uniformity testing of gamma - MRTDosimetrymrtdosimetry-empir.eu/.../05/N-Bates-Optima640Uniformity.pdfy = 0.0002x2 + 0.004x + 8.6525 R² = 0.9936 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5

Stability with time

2

2.2

2.4

2.6

2.8

3

3.2

3.4

0 5 10 15 20 25 30CF

OV

In

teg

ral

Un

ifo

rmit

y (

%)

Days since calibration

Measured Predicted based on count-rate

• Possibly random variation versus predicted values?

• No signficant trend in uniformity seen over 1 month time-scale

• Longer-term stability still to be tested

• 1 month stipulation for SELIMETRY study seems sensible

Page 16: Uniformity testing of gamma - MRTDosimetrymrtdosimetry-empir.eu/.../05/N-Bates-Optima640Uniformity.pdfy = 0.0002x2 + 0.004x + 8.6525 R² = 0.9936 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5

Conclusion• Image uniformity degrades at count-rates above and below that used

for calibrations• Unusual artefacts in images at high count rates which pile up towards

centres of PMT, with streaks between the tubes• Multiple uniformity calibrations or list-mode data processing may be

required• Uniformity at calibrated count-rate is stable over ~1 month timescale• Experiments may need to be repeated with Cu behind source to reduce

scatter • Similar experiments are planned for “normal” mode, and with other

radionuclides• Has anyone else studied this?

• Thoughts from the audience would be greatly appreciated