ultrasonic doppler modes · 2018. 11. 16. · microsoft powerpoint - cachard_doppler_shkodra.ppt...

44
Ultrasonic Ultrasonic Doppler Modes Doppler Modes Piero Tortoli Piero Tortoli Information Engineering Department Information Engineering Department Università degli Studi di Firenze Christian Cachard Christian Cachard CREATIS CREATIS Université Claude Bernard Lyon 1

Upload: others

Post on 01-Feb-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

  • Ultrasonic Ultrasonic

    Doppler ModesDoppler Modes

    Piero TortoliPiero Tortoli

    Information Engineering DepartmentInformation Engineering Department

    Università degli Studi di Firenze

    Christian CachardChristian Cachard

    CREATISCREATIS

    Université Claude Bernard Lyon 1

  • OutlineOutline

    • Doppler effect

    • CW/PW Doppler systems building-blocks

    • Pulsed Wave (PW) mode:

    PRF, sample volume, spectral broadening, mean frequency estimation…

    Advanced Doppler systems and methods:

    �Single-gate (TCD, Duplex)

    �Multi-gate

    �Flow-imaging

    �Power, Harmonic & Tissue Doppler imaging

    • Doppler artefacts (aliasing, blooming…)

  • Doppler effectDoppler effect

    Change in the observed frequency of a wave, due to motion

    Fixed Tx and Rx

    Rx approaching

    Tx

    Rx receding from

    Tx

    fr = ft fr > ft fr < ft

  • A moving reflector / scatterer returns echoes with:

    • higher frequency if it is approaching the source/receiver

    • lower frequency if it is moving away from the source/receiver

    If the transmitter and the receiver are still but a reflector or scatterer is moving, the Doppler effect

    takes place in the same way as in A) or B)

    Doppler effectDoppler effect

  • Doppler effectDoppler effect

  • Doppler effectDoppler effect

    f0 (or ft ) transmitted frequency

    ϑ angle between directions of sound propagation and of target path

    c sound wave velocity

    f0 = 5 MHz

    c= 1500 m/s

    ϑ= 60°

    v = 30 cm/s

    fd ≅ 1 kHz

    vc

    ffd ×= ϑcos2

    0

    Doppler frequency: difference between Tx and Rx frequencies:

    0d Rf f f= −

  • Doppler equationDoppler equation

    The Doppler frequency shift is proportional to:• the target velocity, v• the transmitted frequency, f0• cos ϑ (angle between directions of sound propagation and of target path)

    � it decreases as ϑ� 90°� in particular is 0, when ϑ = 90° (probe

    perpandicular to the vessel axis)

    vc

    ffd ×= ϑcos2

    0

  • Doppler equationDoppler equation

    The frequency ratio is equal to the velocity ratio

    vc

    ffd ×= ϑcos2

    0

    Simple case: ϑ = 60°; cos ϑ = 1/2

    0

    df v

    f c=

  • Vessel Speed range (cm/s) Doppler frequency range (Hz)

    Carotid artery 100 - 150 2000 - 3000

    Ascending aorta 20 - 290 400 - 5800

    Descending aorta 25 - 250 500 -5000

    Abdominal aorta 50 - 60 1000 - 1200

    Femoral artery 100 - 120 2000 – 2400

    Inferior cava vein 15 - 40 300 - 800

    Arterioles 0.5 - 1 10 - 20

    Capillaries 0.02 - 0.17 0,4 - 3,4

    Blood velocitiesBlood velocities

    02 cosd

    ff v

    cϑ= × f0 = 3 MHz

    ϑ = 60°; cos ϑ = 1/2

  • Doppler frequency detection:Doppler frequency detection:

    Audio outputAudio output

    All Doppler frequencies fall in the audio range

    �The sound produced by loudspeakers

    provides immediate (but qualitative and

    operator-dependent) information on blood

    movement

    Jugular vein Common carotid artery

  • Spectral analysis of the Doppler signal allows distinct velocity contributions to be discriminated

    In Doppler spectrograms, subsequent spectra are grey-scale coded and displayed in adjacent vertical lines

    Fre

    quen

    cy (

    kH

    z)

    Time (s)

    Doppler frequency detection: Doppler frequency detection:

    Spectral AnalysisSpectral Analysis

  • Doppler instrumentationDoppler instrumentation

    � 2 - 6 MHz

    Abdominal ultrasound, obstetrical and gynaecological

    exam, echocardiography, transcranial Doppler;

    � 7.5 - 14 MHz

    Small parts, vascular Doppler;

    � 10 - 20 MHz

    Ophthalmology, special vascular exam;

    � 20 - 50 MHz

    Intra-Vascular UltraSound (IVUS), ultrasound

    biomicroscopy (ophthalmology, dermatology);

  • Ultrasound Doppler equipmentUltrasound Doppler equipmentHandheld systems

    (fetal monitoring,

    PAOD…)

    Portable systems (TCD, bedside echo-

    cardiography…)

    Advanced systems (assessment of stenosis, hemodynamics, heart valve function, TDI…)

    Integrated Ultrasound equipment

  • Doppler systems building blocks: Doppler systems building blocks:

    Continuous Continuous WaveWave (CW) systems(CW) systems

    Transmitter

    Flow

    Distincttransmitting and receiving transducers

    US energy is

    continuously transmitted into the body

    Receiver

    Processing

    Audiooutput

    Display

  • 1515

    CW Doppler (Continuous Wave Doppler)

    ( )

    ++=+= t

    DtDtPtdtpt

    rs ωωω

    0cos

    00cos

    0)()(

    p(t): signal from non moving tissue,

    transmitted frequency

    d(t): signal from moving tissue,

    transmitted frequency + Doppler frequency

    f0= 3 MHZ

    f0++fD = 3,001 MHZ

  • CW systems: benefits and drawbacksCW systems: benefits and drawbacks

    TXRX

    • Large investigated volume

    � easy transducer positioning

    Flow

    � no possibility of selecting the region for investigation

    � no discrimination between

    different flow contributions

    � No aliasing (no ambiguity)

    � strong “clutter”

    Echoes backscattered from the region where TX and RX

    beams overlap, are

    integrated in the receiver

    Used only in cardiologyUsed only in cardiology

    2nd Flow

  • 1717

    • The velocity of wall vessel is in the range 5 à 10 mm/s

    • The Doppler frequency induced by the wall motion is in the range 10 to 30 Hz

    • High pass filter: wall filter

    Wall motionWall motion

  • 1818

    Frequency spectrum of the Doppler signal

  • Doppler systems building blocks: Doppler systems building blocks:

    Pulsed wave (PW) systemsPulsed wave (PW) systems

    Transmitter

    Flow

    ·A single transduceracts as transmitter and receiver

    Bursts of US energy are transmitted into the body at rate PRF

    GateReceiver

    Processing

    Display

    Audiooutput

  • • Selection of the R.O.I. �

    � possible discrimination

    between different flow

    contributions

    � possible difficulties in

    transducer positioning

    � Risk of aliasingFlow

    Only the echoes

    backscattered from the selected sample volume are

    gated in the receiver

    PW systems: benefits and drawbacksPW systems: benefits and drawbacks

    Used in most advanced instrumentsUsed in most advanced instruments

  • Doppler equationDoppler equation

    vc

    ffd ×= ϑcos2

    0

    A moving scatterer returns echoes with:

    • fr > f0 ; fd = (fr – fo) > 0 if it is approaching the source/receiver • fr < f0 ; fd = (fr – fo) > 0 if it is moving away from the source/receiver

  • Doppler receiver architectureDoppler receiver architecture

    × LPF HPF

    LPF HPF

    90°

    L.O.

    AFromtransducer

    Q

    I

    Complexdemodulator Sample

    & Hold

    Band-pass filters

    Gate

    Low-noiseamplifier

    •The In-Phase & Quadrature channels are needed to distinguish positive from negative Doppler shifts

    •In recent systems, the echo-signal is sampled at rf (digital processing)

    ×

    The wall filter suppresses

    the signal from tissue

  • PW receiver: Gate (Sample and Hold)PW receiver: Gate (Sample and Hold)

    Range gate

    TX burst

    Received echoes

    1st wall 2nd wall

    vesselUS transducerskin

    •Bursts are transmitted at PRF rate: for each TX burst, one

    sample of the Doppler signal is obtained (time sampling).

    • The electronic gate selects the information backscattered

    only from the region of interest (sample volume).

  • PW systems: Sample Volume (SV) PW systems: Sample Volume (SV)

    SV depth

    SV length

    Transducer

    Blood/tissue volume contributing to the Doppler signal

    • The depth and the length can be set by the operator

    • The width depends on the transducer features/settings

    � Better resolution

    � Worst S/NSmall SV

    SV width

  • Pulse Repetition Frequency Pulse Repetition Frequency

    (PRF)(PRF)

    The rate (1/PRI) at which the bursts are transmitted is called PRF

    • The delay of each echo is proportional to the depth of

    the target• The PRF should be low enough that a new burst is not

    transmitted before the last echo from the max depth (Dmax) has not come back (PRFmax = c / (2xDmax)

    For each Pulse Repetition

    Interval (PRI), one burst of few cycles at frequency fo is transmitted.

    1 / fo

  • Application of spectral analysis:Application of spectral analysis:

    Stenosis detection in the carotid arteryStenosis detection in the carotid artery

    a. Healthy

    subject

    b. Patient with proximal stenosis

    (turbulence)

    (Courtesy of Johan Thijssen)

  • The flow is interrogated by a range of angles around the

    nominal Doppler angle, ϑ

    beam axis

    flow direction

    ϑ +∅∅∅∅

    ∅∅∅∅

    The corresponding Doppler spectrum

    extends over a range around the nominal Doppler frequency

    ϑ ----∅∅∅∅

    Geometric spectral broadeningGeometric spectral broadening

  • Maximum and mean frequencyMaximum and mean frequency

    Mean frequency curveMean frequency curve

  • Mean frequency estimation:Mean frequency estimation:

    CrossCross--correlationcorrelationτ

    SN(t) : Nth Doppler echo

    SN+1(t): [N+1]th Doppler echo

    τ can be estimated as the value maximizing:

    dt ) (tS (t)S 1NN τ++∫

    When such echoes are stored in a digital memory, τcorresponds to the shift needed to make them overlap

  • Quantification of Doppler Quantification of Doppler

    waveformswaveforms

    Pulsatility index (1-10):

    PI = (A-B)/mean

    = (peak systole-peak diastole)mean

    Resistance index (0-1):

    RI = (A-C)/A

    = (peak systole-end diastole) peak systole

    (Courtesy of Johan Thijssen)

  • Peripheral vascular Peripheral vascular ““damping factordamping factor””

    Damping factor:

    DF = PI2 / PI1

    (Courtesy of Johan Thijssen)

  • EchoEcho--Doppler (Duplex) PW systemsDoppler (Duplex) PW systems

    • A (M-mode) scan line

    can be superimposed

    to a B-mode image

    • Over the scan line, a

    specific sample volume

    can be selected

    • The Doppler signal

    produced from the

    gated sample volume

    is analysed

  • PW-Mode

    Ideal for stenosis assessment

  • EchoEcho--Doppler (Duplex) examplesDoppler (Duplex) examples

    Healthy Common Carotid Artery Stenotic Internal Carotid Artery

  • Flow imaging modeFlow imaging mode

    • Real-time 2D velocity maps are obtained

    – By firing several pulses for each scan line

    – By estimating the mean frequency (velocity) detected at each depth

    – By color-coding consecutive pixels according to the detected mean frequencies

    – By scanning a 2-D region

    • Frame rate limitations

    • Poor sensitivity

    • Unsuitable for low velocities Multigate processing applied to multiple scan lines

  • Power Doppler modePower Doppler mode

    • For each scan line, the power of Doppler echoes is detected and integrated (persistance)

    ☺High S/N

    ☺ Ideal for small vessels

    � All movements (including “slow” blood and tissue movements) are detected

    � Qualitative (velocity magnitude is ignored)

    Same hardware as for

    Doppler imaging: software differences

    Un

    idirectio

    na

    l P

    DB

    idirection

    al P

    D

  • Doppler modesDoppler modes

    withwith

    contrast agentscontrast agents

  • Echo enhancement Echo enhancement

    Injection of US contrast agents (microbubbles in a shell) generates strong backscattering (echo enhancement)

    Useful for small, deep, hardly accessible

    vessels (eg:TCD analysis)

  • Harmonic Doppler modeHarmonic Doppler mode

    • Non-linear behaviour of US contrast agents yields 2nd harmonic (2×ft ) components much stronger than those generated by tissue

    Need for wideband transducers – Suitable for perfusion assessment

  • Harmonic Doppler modesHarmonic Doppler modes• Detection of 2nd harmonic echoes allows to reverse the roles of blood & tissue in US images• More sophisticated TX-RX strategies (eg: pulse inversion) allow further increments of blood/tissue ratio to be obtained

    ConventionalUS imaging

    HarmonicDoppler imaging

    Pulse inversion

    Doppler imaging

  • Tissue Doppler ImagingTissue Doppler Imaging

    (TDI)(TDI)

  • Tissue Doppler imaging (TDI)Tissue Doppler imaging (TDI)

    Signal

    from

    tissue Signal from blood

    Conventional

    wall filter

    • The wall filter partially suppresses

    the echo-signal from tissue

    • In TDI, the blood signal is suppressed!

    Frequency/velocity

    Power spectral density

    Signal

    from

    tissue Signal from blood

    Conventional

    wall filter

    Power spectral density

  • Tissue Doppler images of left ventricle Tissue Doppler images of left ventricle

    Diastole Systole

    High signal, low velocity image

    (Courtesy of Johan Thijssen)

  • Doppler angle ambiguitiesDoppler angle ambiguities

    • Frequency/color changes due to a change in the angle of insonation

    • If the angle is not known, the frequency cannot be

    converted to velocity

    The detected frequency depends on the Doppler angle

    ϑcos2 0fc

    vfd =