Correlation and Best-Case BP Measurement Errors BP Error with 1-Point Subject-Specific Calibration Big Picture: To measure BP surrogate(s) from BCG and calibrate to BP! Our mathematical modeling 1 revealed that the primary mechanism for the genesis of BCG is aortic blood pressure (BP) gradients (Fig. 2). Ultra-Convenient Cardiovascular Monitoring with Heartbeat-Induced Body Motions Jin-Oh Hahn (Maryland ), Omer Inan (Georgia Tech ), Ramakrishna Mukkamala (Michigan State ) Ballistocardiogram (BCG) is heartbeat-induced body motion (Fig. 1). BCG is closely associated with cardiac function. Further, BCG is amenable to ultra-convenient measurement. Hence, BCG is anticipated to drastically advance cardiovascular (CV) health monitoring. However, its impact on CV health monitoring has been limited due to limited physical understanding. BCG: Heartbeat-Induced Body Motion This material is based on work supported by the U.S. National Institutes of Health under Grant 1U01EB018818-01. Any opinions, findings, and conclusions or recommendations expressed in this material are those of the authors and do not necessarily reflect the views of the National Institutes of Health. BCG: Physiological Mechanism Ultra-Convenient Blood Pressure Tracking with BCG 1) A simple mathematical modeling analysis reveals that the primary mechanism responsible for the genesis of BCG is blood pressure (BP) gradients in the ascending and descending aorta (Fig. 2(a)). 2) Based on the revealed mechanism, BCG waveform may have the following implications for clinically significant CV parameters (Fig. 2(b)-(d)): (a) I wave is related to the onset of ascending aortic BP (b) I-J interval is related to aortic pulse transit time (c) J amplitude is related to aortic pulse pressure (d) J-K amplitude is related to distal pulse pressure 3) The mathematical (a) model could predict waveforms similar to BCG from experimental aortic BP waveforms obtained from humans (Fig. 2(e)). Fig. 2: Mechanism and utility of BCG for cardiovascular monitoring. (a) Simple mathematical model to reveal mechanism of BCG (b)-(d) Relationship between BCG and blood pressures in the aorta (e) Predicted BCG from experimental aortic BPs in humans Fig. 1: Ballistocardiogram (BCG): Each heartbeat induced head-to-foot directional body motion. 1 : Kim et al., “Ballistocardiogram: Mechanism and Potential for Unobtrusive Cardiovascular Health Monitoring,” Scientific Reports, Vol. 6, Article Number 31297, 2016. Trend of BP and BP Surrogates in Response to BP-Perturbing Interventions R: Rest | MA: Mental Arithmetic | CP: Cold Pressor | PE: Post Exercise R: Rest | MA: Mental Arithmetic | CP: Cold Pressor | PE: Post Exercise [mmHg] BCG PTT BCG I-J Interval PAT RMSE 9.2+/-1.0 11.3+/-1.2 14.1+/-1.4 Mean AE 7.7 9.6 11.1

Upload: others

Post on 06-Aug-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Ultra-Convenient Cardiovascular Monitoring with Heartbeat ... · 1) A simple mathematical modeling analysis reveals that the primary mechanism responsible for the genesis of BCG is

Correlation and Best-Case BP Measurement Errors

BP Error with 1-Point Subject-Specific Calibration

Big Picture: To measure BP surrogate(s) from BCG and calibrate to BP!

Our mathematical modeling1 revealed that the primary mechanism for the genesis ofBCG is aortic blood pressure (BP) gradients (Fig. 2).

Ultra-Convenient Cardiovascular Monitoring with Heartbeat-Induced Body Motions Jin-Oh Hahn (Maryland ), Omer Inan (Georgia Tech ), Ramakrishna Mukkamala (Michigan State )

Ballistocardiogram (BCG) is heartbeat-induced body motion (Fig. 1).

BCG is closely associated with cardiac function. Further, BCG is amenableto ultra-convenient measurement. Hence, BCG is anticipated to drasticallyadvance cardiovascular (CV) health monitoring. However, its impact on CVhealth monitoring has been limited due to limited physical understanding.

BCG: Heartbeat-Induced Body Motion

This material is based on work supported by the U.S. National Institutes of Health under Grant 1U01EB018818-01. Any opinions, findings, and conclusions or recommendationsexpressed in this material are those of the authors and do not necessarily reflect the views of the National Institutes of Health.

BCG: Physiological Mechanism

Ultra-Convenient Blood Pressure Tracking with BCG

1) A simple mathematical modeling analysis reveals thatthe primary mechanism responsible for the genesis ofBCG is blood pressure (BP) gradients in the ascendingand descending aorta (Fig. 2(a)).

2) Based on the revealed mechanism, BCG waveform mayhave the following implications for clinically significantCV parameters (Fig. 2(b)-(d)):(a) I wave is related to the onset of ascending aortic BP(b) I-J interval is related to aortic pulse transit time(c) J amplitude is related to aortic pulse pressure(d) J-K amplitude is related to distal pulse pressure

3) The mathematical (a) model could predict waveformssimilar to BCG from experimental aortic BP waveformsobtained from humans (Fig. 2(e)).

Fig. 2: Mechanism and utility of BCG for cardiovascular monitoring.(a) Simple mathematical model to reveal mechanism of BCG(b)-(d) Relationship between BCG and blood pressures in the aorta(e) Predicted BCG from experimental aortic BPs in humans

Fig. 1: Ballistocardiogram (BCG): Each heartbeat induced head-to-foot directional body motion.

1: Kim et al., “Ballistocardiogram: Mechanism and Potential for UnobtrusiveCardiovascular Health Monitoring,” Scientific Reports, Vol. 6, Article Number31297, 2016.

Trend of BP and BP Surrogates in Response to BP-Perturbing Interventions

R: Rest | MA: Mental Arithmetic | CP: Cold Pressor | PE: Post Exercise

R: Rest | MA: Mental Arithmetic | CP: Cold Pressor | PE: Post Exercise

[mmHg] BCG PTT BCG I-J Interval PATRMSE 9.2+/-1.0 11.3+/-1.2 14.1+/-1.4

Mean AE 7.7 9.6 11.1