UL and the UL logo are trademarks of UL LLC © 2013
Integrating Human Factors nto the R&D Process
Presented at
Presented by
Michael Wiklund, PE, CHFP General Manager – Human Factors Engineering, UL-Wiklund
November 5, 2014
INTRODUCTION
My background
3
Michael Wiklund, MS, PE
General Manager – Human Factors Engineering, UL
Certified Human Factors Professional
Professor, Tufts University
Author/Editor: Usability Testing of Medical Devices
Handbook of Human Factors in Medical Device Design
Designing Usability into Medical Products
Usability in Practice
Member:AAMI Human Factors Engineering Committee
IEC Human Factors Engineering Committee
Human factors specialists
Seek to optimize the quality of interaction between people and products, making them:
• Safe
• Effective
• Efficient
• Satisfying
4
Participant did not notice drug injection pen was expired.
Credit: http://marylandepinephrineresourceschoolnurses.yolasite.com/resources/Epi%20pen%20demo.jpg
5
Photo source: http://media.propertycasualty360.com/propertycasualty360/article/2013/01/08/ConctructionDefect_324-resize-380x300.jpg
To err is human
Photo source: http://www.justjared.com/photo-gallery/2586425/katie-holmes-suri-ice-cream-17/fullsize/
Source: http://www.langleyflyingschool.com/Photos/Safety%20Management%20System/car_err.JPG
To err when using a medical device can be harmful or deadly.
10
11
1999: Deaths resulting from use errors involving medical devices = 10% of at least 45,000/year in the USA
Participant did not notice drug injection pen was expired.
Credit: http://marylandepinephrineresourceschoolnurses.yolasite.com/resources/Epi%20pen%20demo.jpg
12
Participant did not notice drug injection pen was expired.
ROOT CAUSES
•Inadequate training •European format•Small text
Credit: http://marylandepinephrineresourceschoolnurses.yolasite.com/resources/Epi%20pen%20demo.jpg
13
Participant programmed IV pump to deliver antibiotic fluid (“secondary” infusion) at “primary” rate intended for hydration fluid.
Credit: http://humanfactors.ca/wp-content/uploads/2011/06/Lab-CNursingWard1_6114383052-960x475.jpg
14
Participant programmed IV pump to deliver antibiotic fluid (“secondary” infusion) at “primary” rate intended for hydration fluid.
Credit: http://humanfactors.ca/wp-content/uploads/2011/06/Lab-CNursingWard1_6114383052-960x475.jpg
POSSIBLE ROOT CAUSES
•Use of abbreviations•No confirmation•No prompt about IV bag height•Closed clamp•No dose limits
15
Participant did not breathe out completely before taking puff from inhaler.
Credit: http://www.visualphotos.com/photo/2x3376353/man_using_inhaler_42-17590020.jpg
16
Participant did not breathe out completely before taking puff from inhaler.
Credit: http://www.visualphotos.com/photo/2x3376353/man_using_inhaler_42-17590020.jpg
POSSIBLE ROOT CAUSES
•Step not described during training•Step not illustrated in instructions
17
Participant drew blood sample from fingertip still wet with disinfectant.
Credit: http://1.bp.blogspot.com/-pfZ60W9YpLo/ThYVUI5Ad-I/AAAAAAAAAME/6bpsNlSsaVI/s1600/Diabetes_3.jpg
18
Participant drew blood sample from fingertip still wet with disinfectant.
Credit: http://1.bp.blogspot.com/-pfZ60W9YpLo/ThYVUI5Ad-I/AAAAAAAAAME/6bpsNlSsaVI/s1600/Diabetes_3.jpg
POSSIBLE ROOT CAUSES
•Not taught to air dry finger during training•Instructions suggest using disinfectant, but don’t direct users to air dry finger
19
Participant did not detect LVAD’s low battery alarm.
Credit: http://www.umcvc.org/sites/default/files/styles/large/public/JeromeWilson-DrToddKoelling.jpg?itok=xsgl8W_y
20
Participant did not detect LVAD’s battery low alarm.
Credit: http://www.umcvc.org/sites/default/files/styles/large/public/JeromeWilson-DrToddKoelling.jpg?itok=xsgl8W_y
POSSIBLE ROOT CAUSES
•Pizoelectric buzzer emits 3500 Hz signal; above threshold for older individuals with presbycussis•Parallel visual alarm is inconspicuous
21
THE HUMAN FACTORS ENGINEERING IMPERATIVE
Imperative to apply HFE
• Today, FDA and other regulatory agencies expect medical devices to reflect good HFE.
• Applying HFE is a cornerstone of overall risk management.
25
Timeline
1993: ANSI/AAMI HE48:1993 (design guidelines)
1995: Joint AAMI and FDA conference on HF
1996: Quality System Regulation (QSR); indirect requirements for HFE added
1997: End of “grace period” to incorporate HFE in medical device design process
1999: IOM report on medical error
2001: ANSI/AAMI HE74:2001 (HFE process standard)
2006: IEC 60601-1-6 collateral standard (AAMI HE74 is informative annex)
2007: ISO/IEC 62366:2007 (AAMI HE74 included as informative annex)
2008: EU adopts ISO/IEC 62366:2007 as basis for CE mark
2008: FDA’s HFE team moves into Office of Device Evaluation
2009: ANSI/AAMI HE75:2009 (HFE methods and design guidelines)
2011: FDA publishes draft HFE guidelines
2012: Adoption of 3rd edition of IEC 60601 by Europe and Canada
2013: Adoption of 3rd edition of IEC 60601 by USA
Quality System Regulation (Effective Date: June 1, 1997)
Subpart C -- Design Controls, § 820.30 Design controls. The need for human factors techniques or data in the design process is implicit in paragraphs c, f, and g of Section 820.3.
( c ) Design input: "Each manufacturer shall establish and maintain procedures to
ensure that the design requirements relating to a device are appropriate and
address the intended use of the device, including the needs of the users and
patient.”
( f ) Design verification: "Each manufacturer shall establish and maintain
procedures for verifying the design input. Design verification shall confirm that
the design output meets the design input requirements.”
( g ) Design validation:Design validation shall ensure that devices conform to
defined user needs and intended uses, and shall include testing of production
units under actual or simulated use conditions."
Key documents
28
IEC 62366FDA’s Draft Guidance
Basic expectations
• Implement an HFE (UE) program
• Define intended users, use environments, potential hazards, potential use errors, and use-related risks
• Mitigate use-related risks
• Validate user interface designs, proving risk control measures work
• Document HFE activities and outcomes
29
HFE Design History File (one example, partial, simplified)
• HFE program plan
• User profiles and use environment descriptions
• Function and task analysis
• Use-related hazard analysis
• Known problems analysis
• Use-related risk analysis (e.g., Use-FMEA); iterative
• User interface requirements
• Application specification, usability specification, and other documents (per IEC 62366)
• User interface design products (reflecting application of HFE principles)
• Formative evaluation plans and reports
• User interface verification report
• Summative usability test plan and report (to validate user interface)
• HFE Report
Standards apply to diverse medical devices
SAMPLE HFE TECHNIQUES AND END-PRODUCTS
Sample user profile: registered nurses
Nurse often multi-task, placing relatively heavy demand on their short-term memoryNurse often multi-task, placing relatively heavy demand on their short-term memory
Use environment description (sample)
Patient care areas might be brightly or dimly lit.Patient care areas might be brightly or dimly lit.
• “I’d like to be able to read the screen without a magnifying glass.”
• “The bG result should stand out from everything else. Make it big!”
• “You should be able to confirm the time and date so that you know the readings are being logged properly.”
• “The information labels don’t have to be as big.”
• “It shouldn’t have too few or too many controls. Too many controls would be intimidating.”
• “I want it to make a noise when I press a button so that I know it got my input.”
• “I want to know that the device is alive – awake – at all the times.”
• “Don’t let the screen time out too quickly.”
User needs – glucose meter
• Text (capital letters and numbers) shall be ≥14 point (5 mm).
• Blood glucose readout shall be ≥ 60 point (21 mm).
• Main screen shall include the time and date.
• Labels shall be visually subordinate to primary onscreen content.
• There shall be no more than 4 primary hardware controls used to navigate among screens.
• Device shall provide audible feedback in response to all button presses (user option).
• At least one visual element shall be constantly dynamic to indicate the display has not failed.
• Power-down screen after ≥ 2 minutes of inactivity.
Derived user requirements – glucose meter
Design concepts – glucose meter
Task analysis
Deconstruct an overall user interactions with a device into its myriad components: perceptions (P), cognitive processes (C), and actions (A). Yields: PCA analysis.
Sample flow diagram showing decisions and events
Source: http://www.pre-diabetes.com/wp-content/uploads/2012/02/9176220_s.jpg
Use errors derived from task analysis
• User inserts wrong test strip
• User inserts test strip in wrong orientation
• User inserts test strip in wrong port
• User damages test strip during handling
• User applies blood to wrong part of test strip
• User applies too much blood to test strip
• User applies too little blood to test strip
• User does not select re-test when prompted
• User does not remove used strip from meter
• User misreads the blood glucose readout
• User mistakes units of measure as mmol/L versus mg/dL
•
Sample risk mitigations
Advisory messageAudible feedbackClear instructionsColor codingConfirmation messageEmergency power cutoffFamiliar symbolLack of parting linesLarge labelInterlockNeedle guardNon-glare displayOrientation cue
Quick reference cardResistance forceSetting limitsShape codingSize codingSwitch coverTactile feedbackTextured gripTrainingWarning labelWarning lightWider pushbutton spacingCable/tube strain relief
Life of a use error
Life of a use error (continued)
Select a sample size (continued)
• Formative tests typically involve ≤12 participants, yielding excellent insights while preserving resources for additional tests.
• Summative tests typically involve either 15 participants per distinct user group, or 25 participants if the user population is relatively homogeneous (i.e., there is 1 distinct user group).
Operating room simulator
Hospital meeting room
Usability testing laboratory (at UL-Wiklund)
Sample use error report (abbreviated)
Did not attach needle securely
Task
5. Deliver 20 units of insulin.
Task priority
4 out of 16
Risk identifier(s)
Use-FMEA: Items 4.3, 16.1, 22.3
Potential harms
1 –Insulin underdose causes hyperglycemia. 2 – Needlestick injury causes pain and infection. 3 – Needle contamination causes local or systemic infection.
Occurrences
5 test participants committed this use error one or more times during 5 repeated injection trials. The use error occurred 8 times out of the 150 opportunities to err, yielding a use error rate of 5.3%.
Description
3 participants pressed the needle on to the injector but did not twist it to lock it in place. 2 participants initially twisted the needle to lock it in place, but then unlocked it when they removed the needle cap by means of a twisting rather than pulling motion.
Participant reported root causes
3 participants said they forgot to twist the needle to lock it in place. 2 participants speculated that they must have initially gripped the needle’s hub rather than the cap when removing the cap.
Root cause analysis
There is no visual feedback to distinguished a needle that is locked in place from one that is not. The needle cap gripping surface is adjacent to the needle gripping surface, making it vulnerable to unintended twisting during cap removal.
THE HFE REPORT
Excerpted from: Applying Human Factors and Usability Engineering to Optimize Medical Device Design
FDA’s guidance calls for an HFE/UE report
An HFE Report is like a court case.
•Opening statement about the manufacturer’s HFE efforts
•Presentation of HFE evidence and testimony (i.e., results of summative usability test)
•Closing argument that device is safe and effective
Analogy
HFE Report – Sections 1, 2, and 3
HFE Report – Sections 4, 5, and 6
HFE Report – Section 7
Device is adequately safe and effective for the intended users, its intended uses, and use environments.
Claim
BENEFITS AND COSTS
HFE benefits and costs
Benefits
•Faster time to market (regulatory approval)•Simpler user documentation and training•Increased sales related to better first impressions, ease of learning, and initial satisfaction•Reduced risk of recalls, products liability claims•Lower customer support demand•Extended marketability
Costs
•Possibly $250K-$500K for a moderately complex device
Return on investment (ROI)
•Could be the range of 5:1 to 10:1•ROI factors in the cost of lost sales due to a device’s comparatively poor usability, a reject submission to regulators, a recall, and a lawsuit.
Promise of a high ROI
HFE PRACTICE MODELS
Who takes the lead?
Initially:
•CEO (when inspired)
•Regulatory (when facing FDA pushback)
•Quality (when procedure driven)
Long-term:
•R&D (when properly integrated)
Briefly:
•Marketing (when seeking claims)
60
Contracting for HFE services (outsourcing)
• HFE service broker (in-house)
• Engage consultants as needed
• Establish strategic relationship with a preferred HFE vendors
• Establish strategic relationship with multiple HFE vendors
61
In-house capability (core competency)
• Establish a central HFE department that consults to project teams
• Embed HFE specialists within project teams
62
CONCLUSION
Experience suggests…
• Today, HFE is a mandate rather than an option.
• HFE is cost-effective
• Existing staff can perform a substantial amount of HFE work, but specialists are needed at times
• HFE, if implemented in a timely manner, is not a paperwork exercise. It leads to better devices.
• HFE offers commercial advantages.
UL and the UL logo are trademarks of UL LLC © 2013
Questions?
Contact information
UL – Wiklund R&D300 Baker Avenue, Suite 200
Concord, Massachusetts 01742 USA
Telephone: 001 (978) 371–2700
URL: www.ul.com/ul-hfe
Management team:
Michael WiklundGeneral Manager – Human Factors Engineering, E: [email protected]
Jonathan KendlerDesign Director – Human Factors Engineering, E: [email protected]
Allison StrochlicResearch Director – Human Factors Engineering, E: [email protected]