type ms02 · to measure the free chlorine, the sensor cube type ms02 marked cl 2 and a sensor cube...
TRANSCRIPT
Operating Instructions
TypeMS02
Chlorine sensor cubeChlorine dioxide sensor cube
We reserve the right to make technical changes without notice.Technische Änderungen vorbehalten.Sous réserve de modifications techniques.
© Bürkert SAS, 2014-2016
Operating Instructions 1606/01_EU-ML 00566497 Original EN_FR
1
Contents
1 AbouttheoperAtinginstructions...........................................................................................................................4
1.1 symbolsused..........................................................................................................................................................................4
1.2 Definitionofthewordproduct........................................................................................................................................4
1.3 Definitionofthewordsystem.........................................................................................................................................5
1.4 Definitionofthewordbüs................................................................................................................................................5
1.5 Definitionofthetermchlorine-containingcompound......................................................................................5
2 intenDeDuse....................................................................................................................................................................................5
3 bAsicsAfetyinformAtion....................................................................................................................................................6
4 generAlinformAtion................................................................................................................................................................7
4.1 contact........................................................................................................................................................................................7
4.2 Warrantyconditions..............................................................................................................................................................7
4.3 informationsontheinternet............................................................................................................................................7
5 Description.......................................................................................................................................................................................8
6 technicAlDAtA................................................................................................................................................................................9
6.1 conditionsofuse...................................................................................................................................................................9
6.2 conformitytostandardsanddirectives....................................................................................................................9
6.3 materialstheproductismadeof..................................................................................................................................9
6.4 fluiddata....................................................................................................................................................................................9
6.5 measurementdata.............................................................................................................................................................10
6.6 electricaldata.......................................................................................................................................................................10
6.7 communication....................................................................................................................................................................10
7 instAllAtion...................................................................................................................................................................................11
7.1 safetyinstructions.............................................................................................................................................................11
7.2 mountingtheproductonthebackplane................................................................................................................11
8 commissioning............................................................................................................................................................................12
8.1 safetyinstructions.............................................................................................................................................................12
8.2 commissioningtheproduct.........................................................................................................................................12
Type MS02
English
2
Contents
9 ADjustmentAnDoperAtion..............................................................................................................................................13
9.1 safetyinstructions.............................................................................................................................................................13
9.2 howtoadjusttheproduct.............................................................................................................................................13
9.3 generalinformationonthedisplaysoftwaretypeme21.........................................................................................................................................................13
9.4 Deviceviewoftheproduct............................................................................................................................................14
9.5 "function"viewoftheproduct.....................................................................................................................................15
9.6 Detailedviewsofthebüsfunction...........................................................................................................................16
9.7 parameterviewofthesensor......................................................................................................................................18
9.7.1 Changingtheunitofameasuredphysicalquantity...................................................................18
9.7.2 Measuringthefreechlorineconcentrationwithachlorinesensor,Cl2.................................19
9.7.3 DeactivatingthepHcompensationandmeasuringtheactivechlorineconcentration......19
9.7.4 Reducingthemeasurementvariationsoftheconcentration....................................................19
9.7.5 Monitoringthechlorine-containingcompoundconcentrationinthewater..........................20
9.7.6 Monitoringthewatertemperature..................................................................................................20
9.7.7 Resettingcertainparameters..........................................................................................................22
9.7.8 Resettingtheprocessvaluemonitoringparameters..................................................................22
9.8 Diagnosticviewofthesensor.....................................................................................................................................22
9.8.1 Readingthemeasuredvalueofthechlorine-containingcompoundconcentra-tioninthewatersample....................................................................................................................22
9.8.2 Readingthevalueofthecurrentmeasuredbythechlorine-containingcompoundcell...22
9.8.3 Readingtheadjustmentvalueofthezero-pointofthechlorine-containingcom-poundcell.............................................................................................................................................23
9.8.4 Readingtheconcentrationoffsetvalue.........................................................................................23
9.8.5 Readingthesensorslopevalue......................................................................................................23
9.8.6 Readingthemeasuredvalueofthewatersampletemperature..............................................23
9.8.7 Readingthetemperatureoffsetvalue............................................................................................23
9.8.8 Readingthemeasuredvalueofthetemperatureoftheelectronicmeasurementboard...23
9.8.9 Readingthepolarisingvoltagesetpointvalueinthechlorine-containingcompoundcell24
9.8.10 Readingthepolarisingvoltagevaluemeasuredinthechlorine-containingcom-poundcell.............................................................................................................................................24
9.8.11 Readingtheoperatingtimeoftheproduct..................................................................................24
9.8.12 Readingthecalibrationlimitsofthesensorslopevalue...........................................................24
9.9 maintenanceviewofthesensor.................................................................................................................................25
9.9.1 Manuallysettingthezeropointoftheproduct............................................................................25
9.9.2 Calibratingandautomaticallyadjustingthezeropointoftheproduct...................................25
9.9.3 Manuallysettingthesensorslope..................................................................................................26
English
3
9.9.4 1-pointcalibrationandautomaticadjustmentofthesensorslope,off-line..........................26
9.9.5 1-pointcalibrationandautomaticadjustmentofthesensorslope,withtheprocesswater......................................................................................................................................27
9.9.6 Settingthetemperatureoffsetvalueoftheprocesswater......................................................29
9.9.7 Simulatingprocessvaluestocheckthattheproductisworkingproperly,stop-pingthesimulation.............................................................................................................................30
9.9.8 Readingthedateofthelastcalibration........................................................................................30
9.9.9 Readingthedateofthenextduecalibration...............................................................................30
9.9.10 Settingthetimeintervalbetweentwocalibrations.....................................................................31
10 mAintenAnceAnDtroubleshooting.......................................................................................................................32
10.1 safetyinstructions.............................................................................................................................................................32
10.2 cleaningoftheproduct...................................................................................................................................................32
10.3 regularmaintenanceoperations...............................................................................................................................32
10.4 troubleshoutingifnomessageisdisplayed.......................................................................................................32
10.5 troubleshootingifthestatusleDoftheproductisredororange........................................................33
10.5.1 Message"Toohighchlorinevalue"or"Toohighchlorinedioxidevalue"..............................33
10.5.2 Message"Toolowchlorinevalue"or"Toolowchlorinedioxidevalue".................................34
10.5.3 Message"Toohightemperaturevalue".........................................................................................34
10.5.4 Message"Toolowtemperaturevalue"..........................................................................................35
11 spArepArtsAnDAccessories......................................................................................................................................36
12 pAckAging,trAnsport..........................................................................................................................................................36
13 storAge..............................................................................................................................................................................................36
13.1 tostoretheproduct..........................................................................................................................................................36
13.2 Afterstorage..........................................................................................................................................................................36
14 DisposAloftheproDuct..................................................................................................................................................37
English
Type MS02
4
AbouttheOperatingInstructions
1 AbouT The operATing inSTrucTionSTheOperatingInstructionsdescribetheentirelifecycleoftheproduct.PleasekeeptheOperatingInstructionsinasafeplace,accessibletoallusersandanynewowners.
theoperatinginstructionscontainimportantsafetyinformation.
Failuretocomplywiththeseinstructionscanleadtohazardoussituations.
▶ TheOperatingInstructionsmustbereadandunderstood.
1.1 Symbols used
danger
Warnsagainstanimminentdanger.
▶ Failuretoobservethiswarningcanresultindeathorinseriousinjury.
Warning
Warnsagainstapotentiallydangeroussituation.
▶ Failuretoobservethiswarningcanresultinseriousinjuryorevendeath.
attention
Warnsagainstapossiblerisk.
▶ Failuretoobservethiswarningcanresultinsubstantialorminorinjuries.
note
Warnsagainstmaterialdamage.
• Failuretoobservethiswarningmayresultindamagetotheproductorsystem.
Indicatesadditionalinformation,adviceorimportantrecommendations.
ReferstoinformationcontainedintheOperatingInstructionsorinotherdocuments.
▶ Indicatesaninstructiontobecarriedouttoavoidadanger,awarningorapossiblerisk.
→ Indicatesaproceduretobecarriedout.
Indicatestheresultofaspecificinstruction.
1.2 Definition of the word product
Theword"product"usedintheseOperatingInstructionsalwaysreferstothechlorinesensorcube(Cl2)typeMS02orthechlorinedioxidesensorcube(ClO2)typeMS02.
English
Type MS02
5
AbouttheOperatingInstructions
1.3 Definition of the word system
Theword"system"usedwithintheseOperatingInstructionsalwaysreferstotheOnlineAnalysisSystemtype8905.
1.4 Definition of the word büS
Theword"büS"usedintheseOperatingInstructionsreferstothecommunicationbusdeveloppedbyBürkertandbasedontheCANopenprotocol.
1.5 Definition of the term chlorine-containing compound
Dependingonyourproduct,theterm"chlorine-containingcompound"usedinthismanualrefers:
• eithertoCl2(activechlorine,ifpHcompensationisinactive,orfreechlorine,ifpHcompensationisactive),
• ortoClO2(chlorinedioxide).
2 inTenDeD uSe
useofthisproductthatdoesnotcomplywiththeinstructionscouldpresentriskstopeople,nearbyinstallationsandtheenvironment.
▶ ProductsmarkedCl2areintendedexclusivelyformeasuringtheconcentrationoftheCl2(activechlorineorfreechlorine)inthewaterinan8905system.
▶ Tomeasurethefreechlorine,thesensorcubetypeMS02markedCl2andasensorcubetypeMS01forpHcompensationmustbeusedincombinationonthesamefluidcircuitandthesamebüS.
▶ ProductsmarkedClO2areintendedexclusivelyformeasuringtheconcentrationoftheClO2(chlorinedioxide)inthewaterinan8905system.
▶ Thisproductmustbeprotectedagainstelectromagneticinterference,ultravioletraysand,wheninstalledoutdoors,theeffectsofclimaticconditions.
▶ UsetheproductonlyincombinationforeigndevicesorforeigncomponentsrecommendedorapprovedbyBürkert.
▶ ThisproductmustbeusedincompliancewiththecharacteristicsandcommissioninganduseconditionsspecifiedinthecontractualdocumentsandintheOperatingInstructions.
▶ Requirementsforthesafeandproperoperationoftheproductarepropertransport,storageandinstallation,aswellascarefuloperationandmaintenance.
▶ Onlyusetheproductasintended.
English
Type MS02
6
Basicsafetyinformation
3 bASic SAfeTy inforMATionThissafetyinformationdoesnottakeintoaccount:
• anycontingenciesoroccurrencesthatmayariseduringassembly,useandmaintenanceoftheproduct.
• thelocalsafetyregulationsthattheoperatormustensurethestaffinchargeofinstallationandmaintenanceobserve.
Variousdangeroussituations.
Toavoidinjurytakecare:
▶ topreventanyunintentionalpowersupplyswitch-on.
▶ tocarryouttheinstallationandmaintenanceworkbyqualifiedandskilledstaffwiththeappropriatetools.
▶ tousetheproductonlyifinperfectworkingorderandincompliancewiththeinstructionsprovidedintheseOperatingInstructions.
▶ toobservethegeneraltechnicalrulesduringtheplanninganduseoftheproduct.
▶ nottousethisproductinexplosiveatmospheres.
▶ nottousethisproductinanenvironmentincompatiblewiththematerialsfromwhichitismade.
▶ nottomakeanymodificationstotheproduct.
note
elements/componentssensitivetoelectrostaticdischarges
• Thisproductcontainselectroniccomponentssensitivetoelectrostaticdischarges.Theymaybedamagediftheyaretouchedbyanelectrostaticallychargedpersonorobject.Intheworstcasescenario,thesecompo-nentsareinstantlydestroyedorgooutoforderassoonastheyareactivated.
• Tominimiseorevenavoidalldamageduetoanelectrostaticdischarge,takealltheprecautionsdescribedintheEN61340-5-1norm.
• Alsoensurethatyoudonottouchanyoftheliveelectricalcomponents.
English
Type MS02
7
Generalinformation
4 generAl inforMATion
4.1 contact
Tocontactthemanufactureroftheproductusefollowingaddress:
BürkertSAS
RueduGiessen
BP21
F-67220TRIEMBACH-AU-VAL
TheaddressesofourinternationalbranchescanbefoundontheInternetat:www.burkert.com
4.2 Warranty conditions
TheconditiongoverningthelegalwarrantyistheconforminguseoftheproductinobservanceoftheoperatingconditionsspecifiedintheseOperatingInstructions.
4.3 informations on the internet
YoucanfindtheOperatingInstructionsandtechnicaldatasheetsregardingthetypeMS02at:www.burkert.com
English
Type MS02
8
Description
5 DeScripTionTheproductisusedinasystemtype8905.
Theelectricalandfluidconnectionsaremadeviatheconnectionpanelofthesystemtype8905.
Leverto:
- lock/unlocktheproduct.
- carryoutmaintenanceoperations.
Unlockedposition
Lockedposition
Maintenanceposition
Push-buttonwithmea-suredprocessvariable
Fluidconnections
Electricalinterface
Pin
Adaptationpins
SensorcubestatusLED
Slotforthememorycard
Fig. 1: Description of the product
English
Type MS02
9
Technicaldata
6 TechnicAl DATA
6.1 conditions of use
Ambienttemperature 0...+40°C
Airhumidity <90%,withoutcondensation
Protectionratingacc.toEN60529
• IP65,whenpluggedinthebackplane
• IP20,asstandaloneproduct
Max.heightabovesealevel 2000m
6.2 conformity to standards and directives
TheproductconformstotheCEdirectivesofthesystemtype8905,onlywhentheproducttypeMS02ispluggedinthesystemtype8905.
6.3 Materials the product is made of
part material
Housing PPE+PS
Seal EPDM
Lever PC
6.4 fluid data
• Typeoffluid
• pHvalue
• Conductivityvalue
• Water,withoutparticles:drinkingwater,industrialwater
• pH5...9
• >50µS/cmMinimalflowrate >6l/h
Watersamplepressure PN3bar
Watersampletemperature +3...+40°C
English
Type MS02
10
Technicaldata
6.5 Measurement data
Chlorinemeasurement(Cl2)
• Measuringrange
• Sensitivity
• Sensorresolution
• Measurementdeviation("measurementbias",asdefinedinthestandardJCGM200:2012)
• Linearity
• Repeatability
• Responsetime(t90)
• 0.01...5ppm
• -11nA/ppm
• 0.01ppm
• ±0.03ppmor±5%ofthemeasuredvalue,whicheverisgreater
• ±0.02ppmofthemeasuredvalue
• ±0.02ppmofthemeasuredvalue
• 30sChlorinedioxidemeasurement(ClO2)
• Measuringrange
• Sensitivity
• Sensorresolution
• Measurementdeviation("measurementbias",asdefinedinthestandardJCGM200:2012)
• Linearity
• Repeatability
• Responsetime(t90)
• 0.005...5ppm
• –4nA/ppm
• 0.001ppm
• ±0.05ppmor±3%ofthemeasuredvalue,whicheverisgreater
• ±0.01ppmor±3%ofthemeasuredvalue,whicheverisgreater
• ±0.01ppmor±3%ofthemeasuredvalue,whicheverisgreater
• 30sTemperaturemeasurement
• Measuringrange
• Measurementsensor
• 0...+50°C
• Pt1000ClassB,nocontactwiththewatersampleMaintenanceintervalofthechlorinecell 12months,nominal,dependingonthewaterquality
6.6 electrical data
Operatingvoltage 24VDCthroughthebackplaneofthesystemtype8905
Powerconsumption 0.8VA
6.7 communication
Internalcommunication throughbüS
ProductstatusLED accordingtoNAMURNE107
English
Type MS02
11
Installation
7 inSTAllATion
7.1 Safety instructions
note
riskofdamagetotheproductduetonon-conforminginstallation.
• Theelectricalandfluidicinstallationscanonlybecarriedoutbyqualifiedandskilledstaffwiththeappropriatetools.
• Respecttheinstallationinstructionsforthesystem.
note
riskofdamagetotheproductduetothepowersupply
• Shutdownandisolatetheelectricalpowersourcebeforecarryingoutworkonthesystem.
note
riskofdamagetotheproductduetotheenvironment
• Protecttheproductagainstelectromagneticinterference,ultravioletraysand,wheninstalledoutdoors,theeffectsoftheclimaticconditions.
7.2 Mounting the product on the backplane
TheproductispluggedinthebackplaneofthesystemType8905.
4
3
3
1. Drythesurfaceofthebackplanethatwillbeincontactwiththeproduct.
2. Drythesurfaceoftheproduct.
3. Whilethepush-buttonispushed,turnthebayonetlevertotheright,on
theunlockedposition, .Donotpushthelevertothemaintenanceposition.
4. Insertthetwoadaptationpinsintheirholesandthenplugtheproductinthebackplane.
5. Whilethepush-buttonispushed,turnthebayonetlevertotheleft,on
thelockedposition, .
Fig. 2: Mounting a product on the backplane of the system
English
Type MS02
12
Commissioning
8 coMMiSSioning
8.1 Safety instructions
note
riskofdamagetotheproductduetonon-conformingcommissioning.
• Beforecommissioning,makesurethatthestaffinchargehavereadandfullyunderstoodthecontentsoftheseOperatingInstructions.
• Inparticular,observethesafetyrecommendationsandintendeduse.
• Theproduct/theinstallationmustonlybecommissionedbysuitablytrainedstaff.
note
riskofdamagetotheproductduetotheenvironment.
▶ Thisproductmustbeprotectedagainstelectromagneticinterference,ultravioletraysand,wheninstalledoutdoors,theeffectsofclimaticconditions.
8.2 commissioning the product
Aftertheproducthasbeeninstalledonthebackplane:
→ checkthefluidicandelectricalinstallationsofthesystem.
→ checkthetightnessofthesystem.
→ operatetheproductfor12hourswiththecirculatingprocesswater.
→ afterthe12hourshaveelapsed,comparethemeasuredvaluewiththeactualvalueoftheprocesswater.
→ ifnecessary,calibrateandadjusttheproduct.
English
Type MS02
13
Adjustmentandoperation
9 ADjuSTMenT AnD operATion
9.1 Safety instructions
note
riskofdamagetotheproductduetonon-conformingadjustment.
• TheoperatorsinchargeofadjustmentmusthavereadandunderstoodthecontentsoftheseOperatingInstructions.
• TheoperatorsinchargeofadjustmentmusthavereadandunderstoodthecontentsoftheOperatingInstruc-tionsofthedisplaysoftwaretypeME21and/orthecontentsoftheOperatingInstructionsoftheBürkertCommunicatorsoftwaretype8920and/ortheOperatingInstructionsofthecontrollermoduletypeME25.
• Inparticular,observethesafetyrecommendationsandintendeduse.
• Theproduct/installationmustonlybeadjustedbysuitablytrainedstaff.
note
riskofdamagetotheproductduetonon-conformingoperation.
• TheoperatorsinchargeofoperationmusthavereadandunderstoodthecontentsoftheseOperatingInstructions.
• Inparticular,observethesafetyrecommendationsandintendeduse.
• Theproduct/installationmustonlybeoperatedbysuitablytrainedstaff.
9.2 how to adjust the product
Theadjustmentoftheproductcanbemade:
• eitherwiththedisplayofthesystemtype8905.ThedisplayofthesystemismanagedbythesoftwaretypeME21.Seechap.9.3.
• orwithaPCandtheBürkertCommunicatorsoftwaretype8920.Togetgeneralinformationaboutthesoftwaretype8920,refertotheOperatingInstructionsofthetype8920.
9.3 general information on the display software type Me21
TheseOperatingInstructionsexplaintheadjustmentsthatarespecifictothechlorinesensorcubetypeMS02.
→ TogetgeneralinformationaboutthedisplaysoftwaretypeME21,refertotheOperatingInstructionsofthetypeME21thatisontheCDdeliveredwiththesystemandavailableontheinternetatwww.burkert.com.
TheOperatingInstructionsofthedisplaysoftwaretypeME21:
• givegeneralinformationonthesoftware,suchasthedescriptionoftheuserinterface,thestructureofthemenus,thedescriptionofthepossibleviews("Device"viewforexample),thedescriptionofthenavigationbuttons.
• explainhowtomakethegeneraladjustmentssuchas:thedisplaylanguage,thelocatingoftheproduct...
• explainhowtoconfigureandcustomizethe"Desktop"viewswithvaluesorgraphs.
• givegeneralinformationontheerrormessagesandtheoperatingofthesystemstatuslight.
English
Type MS02
14
Adjustmentandoperation
9.4 Device view of the product
The"Device"viewshowssomeofthemeasurementdatarelatedtotheproduct.
→ Ifthedisplayshowsa"Desktop"view,tap toaccessthe"Device"view.
→ Todisplaythe"Device"viewoftheproduct,selecttheproductinthelistofdevicesontheleftofthedisplay.
Controller
Springwater
Device Spring water
SensorChlorine 0.00g/l Cellcurrent 0.00mATemperature 307,74K
Deviceview
HMIU
12.08.201416:55
Cl2
ListofdevicesthatareconnectedtobüS
Thefollowingdatacanbereadfromthe"Device"viewoftheproduct:
• thedevicesthatareconnectedonbüSwiththeir"Description".TheDescriptioncanbemodifiedintheFunctionviewofbüS.
• themeasuredvalueofthechlorine-containingcom-poundconcentrationinthewatersample.
• thevalueofthecurrentmeasuredbythechlorine-containingcompoundcell.
• themeasuredvalueofthetemperatureofthewatersample.
Table 1: "Device" view of the product (example of the chlorine sensor cube Cl2 with the description "Spring water")
→ Todisplaythe"Function"viewoftheproduct,tap .
English
Type MS02
15
Adjustmentandoperation
9.5 "function" view of the product
The"Function"viewshowsthefunctionsavailableforaproductand,foreachfunction,themaindatarelatedtoeachfunction.
Todisplaythe"Function"viewoftheproduct:
1. selecttheproductinthelistofdevices,ontheleftofthedisplay,
2. tap .
büS
function büS
Uniquename 5648310001 Location TreatmentUnitNr1
Description Springwater
ChlorineSensorSpringwater 12.08.201416:55
CL2 Sensor
"function"viewofbüs
→ Todisplaythe"Function"viewofbüStheproductisconnectedto,select"büS"inthelistofdevicesontheleftofthedisplay.
→ Tap toaccessthe"Parameter"view.Seechap.9.6.
Sensor
function Sensor
Chlorine 0.00g/l Chlorinecurrent -0.00mA
Slope -10.00nA/ppm
Temperature 24.10°COffset 0.00°C
ChlorineSensorSpringwater 12.08.201416:55
büS
Cl2
"function"viewofthesensor
→ Todisplaythe"Function"viewofthesensor,select"Sensor"inthelistofdevicesontheleftofthedisplay.
→ Tap toaccessthe"Parameter"view.Seechap.9.7.
Table 2: "Function" views of the product; Function view of büS and Function view of the sensor (example of the chlorine sensor cube Cl2)
English
Type MS02
16
Adjustmentandoperation
9.6 Detailed views of the büS function
parameter büS
Basesettings Uniquedevicename 5648310001
TreatmentUnitNr1
Springwater
13
Location
Description
Address
ChlorinesensorSpringwater 04.04.201412:21"parameter"detailedviewofbüs
• uniquedevicenameisusedbytheparticipantsconnectedonbüS(includingpossibledisplays).Iftheuniquedevicenameischanged,thepartici-pantslosethelinkwiththeproduct.OnlyauserwithInstalleraccessrightscanmodifytheuniquedevicename.Onlymodifytheuniquedevicenameif2productsconnectedonbüShavethesamename(forexample2Cl2sensorcubes).
→ Inthiscase,chooseauniquedevicenamethatexplicitlyidentifiestheproduct,becauseiftheuniquedevicenameischanged,allthebüSpara-metersmustbechanged.
• locationisusedtoentertheproductlocation.
• AddressoftheCANopennode.
TheDescriptionparameterisusedtoidentifytheproductinthesystem.Thedescriptionisdisplayed:
- inthelistofdevices,nexttothesymbolcorres-pondingtotheproduct,
- intheheaderofeachdetailedviewoftheproduct,nexttothedevicename,
- andinthecustomviews.
Table 3: "Parameter" view of the "büS" function
English
Type MS02
17
Adjustmentandoperation
Diagnosis büS
büSinformation Address 126
500kbit/s
Turbiditysensor
564834
Baudrate
Devicename
Ident.number
ChlorinesensorSpringwater 04.04.201412:21
Deviceinformation
1.34hOperatingduration
Mode büSmode
A.07.06.00büSversion
Diagnosis büS
565544
A.03.00.01
A.00.00.00
Softwareidentnumber
Softwareversion
Hardwareversion
ChlorinesensorSpringwater 04.04.201412:21
Devicedriver
1020Serialnumber
A.02.00.00Driverversion
"Diagnosis"detailedviewofbüs
Readonlyfields,whatevertheuserlevel.
InformationaboutbüS:
• AddressoftheCANopennode.
• baudrateforinformationtransmittedviabüS.
• modeofoperationofthecommunicationbus.
• büsversion
Informationabouttheproduct:
• Devicename
• operatingdurationindicatesthenumberofope-ratinghoursoftheproduct
• ident.number:productordercode.
• softwareidentnumber:ordercodeforthesoftwareinstalledontheproduct.
• softwareversion:versionofthesoftwareinstalledontheproduct.
• hardwareversion:hardwareversionoftheproduct.
• serialnumber:serialnumberoftheproduct.
• Driverversion:versionofthedriverinstalledontheproduct.
Table 4: "Diagnosis" view of the "büS" function
Maintenance büS
Device Reset
ChlorinesensorSpringwater 12:21:16"maintenance"detailedviewofbüs
ThisfunctionisavailablefromtheAdvanceduserlevel.
Usedtorestarttheproduct.
Table 5: "Maintenance" view of the "büS" function
English
Type MS02
18
Adjustmentandoperation
9.7 parameter view of the sensor
→ Inthe"Function"view,tap toaccessthe"Parameter"view.
parameter Sensor
Temperature Unit °C
µAUnit
Configure2)
Device2)
ChlorineSensorSpringwater 12:21:16
Chlorine1)
Cellcurrent
Off2)Status2)
Unit g/l
Value2)
Notconfigured2)
Notconfigured2)
pHcompensation2)
parameter Sensor
Warnings
Errors
Messagestriggers
ChlorineSensorSpringwater 12:21:16
Parameter
Messagestriggers
Restorefactoryvalues
Filter Responsetime 12s
1)Orchlorinedioxide,forachlorinedioxidesensorcube(ClO2)2)Onlychlorinesensorcube(Cl2)
• Theseparameters,withtheexceptionoftheunitparameters,canbeeditedwithauserlevelofatleast:Advanceduser.
• Theunitparameterscanbeeditedbyanyuserlevel.
The"Parameter"viewofthesensormakesitpossible:
• tochangetheunitsofameasuredphysicalquantity.Seechap.9.7.1.
• tomeasuretheconcentrationoffreechlorinewithachlorinesensorcubeCl2.Seechap.9.7.2.
• deactivatethepHcompensationandmeasuretheactivechlorineconcentration.Seechap.9.7.3.
• dampenthemeasurementvariationsoftheconcen-tration.Seechap.9.7.4.
• monitorthechlorine-containingcompoundconcen-trationinthewater.Seechap.9.7.5.
• monitorthewatertemperature.Seechap.9.7.6.
• resetcertainparameters.Seechap.9.7.7andchap.9.7.8.
Table 6: "Parameter" view of the sensor
9.7.1 changing the unit of a measured physical quantity
Thefollowingmeasuredquantityunitscanbechanged:
• thechlorine-containingcompoundconcentration,
• thecurrentmeasuredbythechlorine-containingcompoundcell,
• thetemperature.
Ifyouchangeaunit,theassociatedvaluesareautomaticallyrecalculated(e.g.,thethresholdvaluesformonitoringtheprocessvalues).
procedureforchangingtheunitbyameasuredphysicalquantity:
1. Inthe"Parameter"viewofthesensor,selectthephysicalquantity.
2. Selectaunit.
3. Confirm.
Theunitsareset.
English
Type MS02
19
Adjustmentandoperation
9.7.2 Measuring the free chlorine concentration with a chlorine sensor, cl2
Whenitleavesthefactory,thechlorinesensorCl2issettomeasuretheactivechlorineconcentration.
→ Tomeasurethefreechlorineconcentration,activatethepHcompensation.
ToactivatethepHcompensation,apHsensorcubetypeMS01mustbeinstalledonthesamefluidcircuitandconnectedtothesamebüS.
Procedure:
1. Tapphcompensation configure inthe"Parameter"viewofthesensor.
2. SelectthepHsensorcubeusedforthepHcompensation.
3. Confirm.
4. Tapphcompensation status off inthe"Parameter"viewofthesensor.
5. Tapoff:onisdisplayed.
6. Confirm.
ThepHcompensationisactivatedandtheproductmeasuresthefreechlorineconcentration.
9.7.3 Deactivating the ph compensation and measuring the active chlorine concentration
WhenthepHcompensationisactivated,theproductmeasuresthefreechlorine.
→ Tomeasuretheactivechlorineconcentration,thepHcompensationmustbedeactivated.
Procedure:
1. Tapphcompensation status on inthe"Parameter"viewofthesensor.
2. Tapon:offisdisplayed.
3. Confirm.
ThepHcompensationisdeactivatedandtheproductmeasurestheactivechlorineconcentration.
9.7.4 reducing the measurement variations of the concentration
Toreducethemeasurementvariationsoftheconcentration,adjustthedigitalfilter.
→ DefinearesponsetimeT10-90forthesignalmeasured.Thelongertheresponsetime,themorethemeas-urementvariationsarereduced.
Procedure:
1. Tapfilter responsetime 30s inthe"Parameter"viewofthesensor.
2. Enteravalueinseconds.
3. Confirm.
Themeasurementvariationsarereduced.
English
Type MS02
20
Adjustmentandoperation
9.7.5 Monitoring the chlorine-containing compound concentration in the water
Thechlorine-containingcompoundconcentrationcanbemonitoredtodetectanyproblemsintheprocess.
2minimumvaluesand2maximumvaluescanbeentered:
• 1maximumwarningvalueand1minimumwarningvalue.Ifoneofthesevaluesisreached,awarningmessageistriggeredandthesystemstatuslightisorange.
• 1maximumerrorvalueand1minimumerrorvalue.Ifoneofthesevaluesisreached,anerrormessageistrig-geredandthesystemstatuslightisred.
Whenoneofthesevaluesisreachedandcorrespondingactivationflagisselected,anerrororwarningmessage
istriggeredinthe"MessageList":tap togotothe"MessageList".Seechap.10.5TroubleshootingifthestatusLEDoftheproductisredororangeformoredetailsontroubleshooting.
Activationprocedureforthewarningmessagetrigger:
1. Tapmessagestriggers Warnings inthe"Parameter"viewofthesensor.
2. TapWarningtriggers lowconcentration andentertheminimumvalueforthechlorine-containingcompoundcontentforwhichawarningmessageistriggered.
3. Confirm.
4. TapWarningtriggers highconcentration andenterthemaximumvalueforthechlorine-containingcompoundcontentforwhichawarningmessageistriggered.
5. Confirm.
6. TapWarningtriggers Activationflag andselectlowconcentrationand/orhighconcentration.
7. Confirm.
Activationprocedurefortheerrormessagetrigger:
1. Tapmessagestriggers errors inthe"Parameter"viewofthesensor.
2. Taperrortriggers lowconcentration andentertheminimumvalueforthechlorine-containingcompoundcontentforwhichanerrormessageistriggered.
3. Confirm.
4. Taperrortriggers highconcentration andenterthemaximumvalueforthechlorine-containingcompoundcontentforwhichanerrormessageistriggered.
5. Confirm.
6. Taperrortriggers Activationflag andselectlowconcentrationand/orhighconcentration.
7. Confirm.
Monitoringofthemaximumchlorine-containingcompoundconcentrationandtheminimumchlorine-containingcompoundconcentrationisactive.
9.7.6 Monitoring the water temperature
Thewatertemperaturecanbemonitoredtodetectanyproblemsintheprocess.
2minimumvaluesand2maximumvaluescanbeentered:
English
Type MS02
21
Adjustmentandoperation
• 1maximumwarningvalueand1minimumwarningvalue.Ifoneofthesevaluesisreached,awarningmessageistriggeredandthesystemstatuslightisorange.
• 1maximumerrorvalueand1minimumerrorvalue.Ifoneofthesevaluesisreached,anerrormessageistrig-geredandthesystemstatuslightisred.
Whenoneofthesevaluesisreachedandcorrespondingactivationflagisselected,anerrororwarningmessage
istriggeredinthe"MessageList":tap togotothe"MessageList".Seechap.10.5TroubleshootingifthestatusLEDoftheproductisredororangeformoredetailsontroubleshooting.
Activationprocedureforthewarningmessagetrigger:
1. Tapmessagestriggers Warnings inthe"Parameter"viewofthesensor.
2. TapWarningtriggers lowtemperature andentertheminimumtemperaturevalueforwhichawarningmessageistriggered.
3. Confirm.
4. TapWarningtriggers hightemperature andenterthemaximumtemperaturevalueforwhichawarningmessageistriggered.
5. Confirm.
6. TapWarningtriggers Activationflag andselectlowtemperatureand/orhightemperature.
7. Confirm.
Activationprocedurefortheerrormessagetrigger:
1. Tapmessagestriggers errors inthe"Parameter"viewofthesensor.
2. Taperrortriggers lowtemperature andentertheminimumtemperaturevalueforwhichanerrormessageistriggered.
3. Confirm.
4. Taperrortriggers hightemperature andenterthemaximumtemperaturevalueforwhichanerrormessageistriggered.
5. Confirm.
6. Taperrortriggers Activationflag andselectlowtemperatureand/orhightemperature.
7. Confirm.
Monitoringofthemaximumwatertemperatureandtheminimumwatertemperatureisactive.
English
Type MS02
22
Adjustmentandoperation
9.7.7 resetting certain parameters
Thisfunctionisusedtoresetthefollowingparameters:
• thetemperatureoffsetvalue(chap.9.9.6),
• themeasurementfiltervalue(chap.9.7.4),
• thepHcompensationparameters(chap.9.7.2).
procedure:
1. Taprestorefactoryvalues parameter inthe"Parameter"viewofthesensor.
2. Confirm.
Theparametersarereset.
9.7.8 resetting the process value monitoring parameters
Thisfunctionisusedtoresettheparametersformonitoringtheprocessvalues(chap.9.7.5andchap.9.7.6).
procedure:
1. Taprestorefactoryvalues messagestriggers inthe"Parameter"viewofthesensor.
2. Confirm.
Theparametersformonitoringtheprocessvaluesarereset.
9.8 Diagnostic view of the sensor
9.8.1 reading the measured value of the chlorine-containing compound concentration in the water sample
→ Inthe"Parameter"viewofthesensor,tap togotothe"Diagnostic"viewofthesensor.
→ Inthe"Diagnostic"viewofthesensor,takeareadingofthevalueinthefieldmonitoredvalueschlorineorinthefieldmonitoredvalueschlorinedioxide.
9.8.2 reading the value of the current measured by the chlorine-containing compound cell
→ Inthe"Parameter"viewofthesensor,tap togotothe"Diagnostic"viewofthesensor.
→ Inthe"Diagnostic"viewofthesensor,takeareadingofthevalueinthefieldmonitoredvaluescellcurrent.
English
Type MS02
23
Adjustmentandoperation
9.8.3 reading the adjustment value of the zero-point of the chlorine-containing compound cell
→ Inthe"Parameter"viewofthesensor,tap togotothe"Diagnostic"viewofthesensor.
→ Inthe"Diagnostic"viewofthesensor,takeareadingofthevalueinthefieldmonitoredvaluesZeroadjustment.
9.8.4 reading the concentration offset value
→ Inthe"Parameter"viewofthesensor,tap togotothe"Diagnostic"viewofthesensor.
→ Inthe"Diagnostic"viewofthesensor,takeareadingofthevalueinthefieldmonitoredvaluesconcen-trationoffset.
9.8.5 reading the sensor slope value
Theslopevalueofthechlorine-containingcompoundhasbeeneithercalculatedduringthecalibrationorenteredinthe"Maintenance"viewofthesensor.
→ Inthe"Parameter"viewofthesensor,tap togotothe"Diagnostic"viewofthesensor.
→ Inthe"Diagnostic"viewofthesensor,takeareadingofthevalueinthefieldmonitoredvaluesslope.
9.8.6 reading the measured value of the water sample temperature
→ Inthe"Parameter"viewofthesensor,tap togotothe"Diagnostic"viewofthesensor.
→ Inthe"Diagnostic"viewofthesensor,takeareadingofthevalueinthefieldmonitoredvaluestemperature.
9.8.7 reading the temperature offset value
Thetemperatureoffsetvaluehasbeenenteredinthe"Maintenance"viewofthesensor.
→ Inthe"Parameter"viewofthesensor,tap togotothe"Diagnostic"viewofthesensor.
→ Inthe"Diagnostic"viewofthesensor,takeareadingofthevalueinthefieldmonitoredvaluestemper-atureoffset.
9.8.8 reading the measured value of the temperature of the electronic measurement board
→ Inthe"Parameter"viewofthesensor,tap togotothe"Diagnostic"viewofthesensor.
→ Inthe"Diagnostic"viewofthesensor,takeareadingofthevalueinthefieldmonitoredvaluespcbtemperature.
English
Type MS02
24
Adjustmentandoperation
9.8.9 reading the polarising voltage set point value in the chlorine-containing compound cell
→ Inthe"Parameter"viewofthesensor,tap togotothe"Diagnostic"viewofthesensor.
→ Inthe"Diagnostic"viewofthesensor,takeareadingofthevalueinthefieldmonitoredvaluesrequiredpolarization.
9.8.10 reading the polarising voltage value measured in the chlorine-containing compound cell
→ Inthe"Parameter"viewofthesensor,tap togotothe"Diagnostic"viewofthesensor.
→ Inthe"Diagnostic"viewofthesensor,takeareadingofthevalueinthefieldmonitoredvaluesmeasuredpolarization.
9.8.11 reading the operating time of the product
Thisfunctionshowstheoperatingtimeoftheproductsinceinitialcommissioning.
→ Inthe"Parameter"viewofthesensor,tap togotothe"Diagnostic"viewofthesensor.
→ Inthe"Diagnostic"viewofthesensor,takeareadingofthevalueinthefieldmonitoredvaluescellworkingtime.
9.8.12 reading the calibration limits of the sensor slope value
Thisfunctionshowstheoperatingtimeoftheproductsinceinitialcommissioning.
→ Inthe"Parameter"viewofthesensor,tap togotothe"Diagnostic"viewofthesensor.
→ Inthe"Diagnostic"viewofthesensor,takeareadingofthevaluesinthefields:
- slopelimitsWarninglimitlow
- slopelimitsWarninglimithigh
- slopelimitserrorlimitlow
- slopelimitserrorlimithigh
English
Type MS02
25
Adjustmentandoperation
9.9 Maintenance view of the sensor
9.9.1 Manually setting the zero point of the product
→ Tocalibrateandautomaticallyadjustthezeropointoftheproduct,seechap.9.9.2.
Todothesettingsinthe"Maintenance"viewofthesensor,youmusthaveuserlevelinstaller.
procedureformanualsettingofthezeropointoftheproduct:
1. Inthe"Parameter"viewofthesensor,tap togotothe"Maintenance"viewofthesensor.
2. Inthe"Maintenance"viewofthesensor,tapchlorineZeroadjustment orchlorinedioxideZeroadjustment .
3. Enterthevalue.
4. Confirm.
Thezerocurrentpointofthesensorisadjusted.
9.9.2 calibrating and automatically adjusting the zero point of the product
Todothesettingsinthe"Maintenance"viewofthesensor,youmusthaveuserlevelinstaller.
→ Tomanuallysetthezeropointoftheproduct,seechap.9.9.1.
procedureforcalibratingandautomaticallyadjustingthezeropointoftheproduct:
1. Inthe"Parameter"viewofthesensor,tap togotothe"Maintenance"viewofthesensor.
2. Inthe"Maintenance"viewofthesensor,tapchlorinecalibrationZeroadjustment orchlorinedioxidecalibrationZeroadjustment .
3. Step1/4:preparethereferencesolution.
4. Letthereferencesolutionflowthroughthesensor.
5.
6. Step2/4:waituntilthevaluesdisplayedarestable.
7. Whenthevaluesdisplayedarestable.
8. Step3/4:
→ toaborttheadjustmentofthezeropointoftheproduct,tap .
Thezeropointoftheproductisnotadjusted.
English
Type MS02
26
Adjustmentandoperation
→ toautomaticallyadjustthezeropointoftheproduct,tap .
Thezeropointoftheproductisadjusted.
9.9.3 Manually setting the sensor slope
Todothesettingsinthe"Maintenance"viewofthesensor,youmusthaveuserlevelinstaller.
Youcansetthesensorcubeslopewithoutmakingacalibration.
Tomakea1-pointcalibrationand,ifnecessary,automaticallyadjustthesensorslope,seechap.9.9.4andchap.9.9.5.
procedureformanuallysettingthesensorslopevalue:
1. inthe"Parameter"viewofthesensortogotothe"Maintenance"viewofthesensor.
2. Inthe"Maintenance"viewofthesensor,tapchlorineslope orchlorinedioxideslope .
3. Enterthevalue.
4. Confirm.
Thesensorslopeisset.
9.9.4 1-point calibration and automatic adjustment of the sensor slope, off-line
Toperformacalibration,youmusthaveuserlevelinstaller.
danger
Dangerduetothenatureofthereferencesolution.
▶ Respecttheregulationsonaccidentpreventionandsafetyrelatingtotheuseofaggressivefluids.
Thisprocedureisusedforthe1-pointcalibrationofthesensor,off-line,toautomaticallyadjustthesensorslopevalue.
→ Tomakea1-pointcalibrationwiththeprocesswater,followonlytheinstructionsinchap.9.9.5.
1. Connectaperistalticpumpinthewatersampleinletcircuitofthesystem,asshowninFig.3.Ifthefluidicconnectionsarenotdonecorrectly,theproductandthesystemcanbedamagedbecausethepressureintheproductandinthesystemistoohigh.
2. Makesurethedirectionoftheflowiscorrect.
3. Makesurethefluidicinstallationforthecalibrationistight.
4. Letcleanwaterflowthroughthesystemtorinsetheproduct.
5. Prepareasolution2to3timesmoreconcentratedinchlorine-containingcompoundthantheconcentrationinthewatersampletobeanalysed.
6. Letthepreparedsolutionflowthroughthesensorcubefor4hours.
English
Type MS02
27
Adjustmentandoperation
7. EitherdeterminethefreechlorineconcentrationofthepreparedsolutionbycolorimetrywiththereagentDPD-1.Iftheproductmeasurestheactivechlorine,calculatetheactivechlorineconcentrationusingthefollowingformula:[activechlorine]=[freechlorine]/(1+10pH–pKa),wherepKa=7.53at25°C.
8. OrdeterminethechlorinedioxideconcentrationofthepreparedsolutionbycolorimetrywiththereagentDPD-1andmultiplythatvalueby1.9.
9. Followinstructions3to11inchap.9.9.51-pointcalibrationandautomaticadjustmentofthesensorslope,withtheprocesswater.
+-
OUT IN
Peristalticpump
Referencesolution
Powersupplyofthesystem
Fig. 3: Installation for calibration of the product off-line
9.9.5 1-point calibration and automatic adjustment of the sensor slope, with the process water
Toperformacalibration,youmusthaveuserlevelinstaller.
danger
Dangerduetothenatureofthereferencesolution.
▶ Respecttheregulationsonaccidentpreventionandsafetyrelatingtotheuseofaggressivefluids.
Thisprocedureisusedforthe1-pointcalibrationofthesensorwiththeprocesswatertoautomaticallyadjustthesensorslopevalue.
→ Tomakea1-pointcalibrationoff-line,followonlytheinstructionsinchap.9.9.4.
→ Beforecalibratingthesensorwiththeprocesswater,runthefullcommissioningprocedure.
English
Type MS02
28
Adjustmentandoperation
1. EitherdeterminethefreechlorineconcentrationoftheprocesswaterbycolorimetrywiththereagentDPD-1.Iftheproductmeasurestheactivechlorine,calculatetheactivechlorineconcentrationusingthefollowingformula:[activechlorine]=[freechlorine]/(1+10pH–pKa),wherepKa=7.53at25°C.
2. OrdeterminethechlorinedioxideconcentrationoftheprocesswaterbycolorimetrywiththereagentDPD-1andmultiplythatvalueby1.9.
3. Inthe"Maintenance"viewofthesensor,tapchlorinecalibration1point orchlorinedioxidecalibration1point .
4. Step1/5:waituntilthemeasurementisstable.
5. Tap .
6. Step2/5:toadjustthesensorcube,tapinputreferencesolutionvalue andenterthechlorine-containingcompoundconcentrationofthesolutionpreviouslydetermined.
7. Confirm.
8. Tap .
9. Step3/5:whenthemeasurementofthechlorine-containingcompoundconcentrationisstable,tap .
10. Step4/5:
• iftheadjustmenthassucceeded,thecalculatedslopevalueisdisplayedandthedateofthelastcalibrationis
updated(seechap.9.9.8).Tap togotothestep5/5.
• ifanerrormessageisdisplayed,refertoTable7.
• ifawarningmessageisdisplayed,refertoTable8.
11. Step5/5:confirmorcanceltheadjustment.
Displayedmessage Computedcalibrationvalueoutoferrorlimitssymboldisplayedontheiconoftheproduct,inthelistofdevicespossiblecause Thesensorcalibrationhasfailedbecausethecalculatedslopevalueis
outoftheerrorrange.Whattodo? 1. Comparethecalculatedslopevaluewiththecalibrationlimitsthat
canbereadinthe"Diagnosis"viewofthesensor.
2. Replacetheproductbyanewone.
Table 7: Error message at the end of the 1 point calibration of the product
English
Type MS02
29
Adjustmentandoperation
Displayedmessage Computedcalibrationvalueoutofwarninglimitssymboldisplayedontheiconoftheproduct,inthelistofdevicespossiblecause Thecalculatedslopevalueisoutofthewarningrangebecause:Whattodo? • eitherawrongsolutionhasbeen
usedforthecalibration.• ortheproductisaging.
solution 1. Makesurethereferencesolutionusedisthecorrectone.
2. Useacorrectreferencesolutionanddothecompletecalibrationprocedureagain.
→ Youcanchoosetoeitherval-idateorcanceltheadjustment.
• Ifyouvalidatetheadjustment,thenewcalculatedslopevalueisusedtodeterminethechlorineconcentration,andthelastcali-brationdateisupdated.
• Ifyoucanceltheadjustment,thecurrentslopevalueisusedtodeterminethechlorineconcen-tration,andthelastcalibrationdateisnotupdated.
→ Plantoreplacetheproduct.
Table 8: Warning message at the end of the 1 point calibration of the product
9.9.6 Setting the temperature offset value of the process water
Todothesettingsinthe"Maintenance"viewofthesensor,youmusthaveuserlevelinstaller.
Thetemperatureoftheprocesswatermeasuredbythesensorcanbecorrectedwithanoffsetvalue.
1. inthe"Parameter"viewofthesensortogototheMaintenanceviewofthesensor.
2. Inthe"Maintenance"viewofthesensor,taptemperaturecalibrationoffset .
3. Enterthevalueintheunitsdisplayed.
4. Confirm.
5. Tochangethetemperatureunits,gotothe"Parameter"viewofthesensorandtaptemperatureunit .
6. Selectthetemperatureunits:thetemperatureoffsetvalueisautomaticallyconverted.
7. Confirm.
Theoffsetvalueoftheprocesswatertemperatureismodified.
English
Type MS02
30
Adjustmentandoperation
9.9.7 Simulating process values to check that the product is working properly, stopping the simulation
Tosimulateprocessvalues,youmusthaveuserlevelinstaller.
Thisfunctioncanbeused,forexample,toensurethatthewarningand/orerrorlimitshavebeenenteredcorrectly.
1. inthe"Parameter"viewofthesensortogototheMaintenanceviewofthesensor.
2. Inthe"Maintenance"viewofthesensor,tapsimulationstatus .
3. Tapoff:onisdisplayed.
4. Confirm.
5. Tosimulateachlorine-containingcompoundconcentration,tapsimulationconcentration andenterthechlorine-containingcompoundconcentrationvaluetobesimulated.
6. Confirm.
7. Tosimulateatemperaturevalue,tapsimulationtemperature andenterthetemperaturevaluetobesimulated.
8. Confirm.
Thevaluesenteredareemitted.
9. Tostopthesimulationofprocessvalues,tapsimulationstatus .
10. Tapon:offisdisplayed.
11. Confirm.
Thesimulationisstopped.
9.9.8 reading the date of the last calibration
Toreadthedateofthelastcalibrationthathassucceeded.
→ inthe"Parameter"viewofthesensortogototheMaintenanceviewofthesensor.
→ Inthe"Maintenance"viewofthesensor,readthedateinthefieldcalibrationschedulelast.
9.9.9 reading the date of the next due calibration
Toreadthedateofthenextcalibrationthatmustdone.
→ inthe"Parameter"viewofthesensortogototheMaintenanceviewofthesensor.
→ Inthe"Maintenance"viewofthesensor,readthedateinthefieldcalibrationschedulenext.
English
Type MS02
31
Adjustmentandoperation
9.9.10 Setting the time interval between two calibrations
Todothesettingsinthe"Maintenance"viewofthesensor,youmusthaveuserlevelinstaller.
Tosetthetimeinterval,indays,betweentwocalibrations:
1. inthe"Parameter"viewofthesensortogototheMaintenanceviewofthesensor.
2. Inthe"Maintenance"viewofthesensor,tapcalibrationscheduleintervalindays .
3. Tapthenumberofdaysbetweentwocalibrations.Werecommendtocalibratetheproducteverymonth.
4. Confirm.
Thetimeintervalbetween2calibrationsisset.Whentheduecalibrationdateisreached,awarningmessageisdisplayedintheMessagesList.
English
Type MS02
32
Maintenanceandtroubleshooting
10 MAinTenAnce AnD TroubleShooTing
10.1 Safety instructions
Warning
riskofinjuryduetonon-conformingmaintenance.
▶ Maintenancemustonlybecarriedoutbyqualifiedandskilledstaffwiththeappropriatetools.
note
riskofdamagetotheproductduetothepowersupply
• Shutdownandisolatetheelectricalpowersourcebeforecarryingoutworkonthesystem.
10.2 cleaning of the product
danger
riskofinjuryduetothenatureofthedetergent.
▶ Respecttheregulationsonaccidentpreventionandsafetyrelatingtotheuseofaggressivefluids.
Theproductcanbecleanedwithaclothdampenedwithwater(max.40°C)orwithanaciddetergent(withmax.5%ofhydrochloricacid).
10.3 regular maintenance operations
→ Onceamonth,checkthemeasurements.
→ Calibratetheproductand,ifnecessary,adjusttheproductaccordingly.Seechap.9.9.3Manuallysettingthesensorslopeorchap.9.9.51-pointcalibrationandautomaticadjustmentofthesensorslope,withtheprocesswater.
10.4 Troubleshouting if no message is displayed
colouroftheproductstatusleD OFF
possiblecause Theproduct/thesystemisnotenergized.
Whattodo? 1. Checkthewiring.
2. Makesurethevoltagesupplyis24VDC.
3. Checkthatthepowersupplysourceisworkingproperly.
English
Type MS02
33
Maintenanceandtroubleshooting
10.5 Troubleshooting if the status leD of the product is red or orange
Ifanerrororawarningmessagehasbeengeneratedbythesystem:
• thestatuslightofthesystemisredororange,
• thestatusLEDoftheproductisredororange,
• thesymbol or appearsontheUniqueDeviceNameoftheproduct,inthelistofdevices,
• thesymbol appearsinthetopleftcornerofthedisplay.
→ Tap toaccessthe"MessagesList".
10.5.1 Message "Too high chlorine value" or "Too high chlorine dioxide value"
Displayedmessage Toohighchlorinevalue
or
Toohighchlorinedioxidevaluesymboldisplayedontheiconoftheproduct,inthelistofdevicespossiblecause Thevalueofthechlorine-containingcompoundconcentrationinthe
watersampleexceedsthemaximumerrorvaluedefinedbytheuser.
Seechap.9.7.5Monitoringthechlorine-containingcompoundconcen-trationinthewater.
Whattodo? → Checktheprocess.
Displayedmessage Toohighchlorinevalue
or
Toohighchlorinedioxidevaluesymboldisplayedontheiconoftheproduct,inthelistofdevicespossiblecause Thevalueofthechlorine-containingcompoundconcentrationinthe
watersampleexceedsthemaximumwarningvaluedefinedbytheuser.
Seechap.9.7.5Monitoringthechlorine-containingcompoundconcen-trationinthewater.
Whattodo? → Checktheprocess.
English
Type MS02
34
Maintenanceandtroubleshooting
10.5.2 Message "Too low chlorine value" or "Too low chlorine dioxide value"
Displayedmessage Toolowchlorinevalue
or
Toolowchlorinedioxidevaluesymboldisplayedontheiconoftheproduct,inthelistofdevicespossiblecause Thevalueofthechlorine-containingcompoundconcentrationinthe
watersampleexceedstheminimumerrorvaluedefinedbytheuser.
Seechap.9.7.5Monitoringthechlorine-containingcompoundconcen-trationinthewater.
Whattodo? → Checktheprocess.
Displayedmessage Toolowchlorinevalue
or
Toolowchlorinedioxidevaluesymboldisplayedontheiconoftheproduct,inthelistofdevicespossiblecause Thevalueofthechlorine-containingcompoundconcentrationinthe
watersampleexceedstheminimumwarningvaluedefinedbytheuser.
Seechap.9.7.5Monitoringthechlorine-containingcompoundconcen-trationinthewater.
Whattodo? → Checktheprocess.
10.5.3 Message "Too high temperature value"
Displayedmessage Toohightemperaturevaluesymboldisplayedontheiconoftheproduct,inthelistofdevicespossiblecause Thetemperaturevalueofthewatersampleexceedsthemaximumerror
valuedefinedbytheuser.
Seechap.9.7.6Surveillerlatempératuredel'eau.Whattodo? → Checktheprocess.
Displayedmessage Toohightemperaturevaluesymboldisplayedontheiconoftheproduct,inthelistofdevicespossiblecause Thetemperaturevalueofthewatersampleexceedsthemaximum
warningvaluedefinedbytheuser.
Seechap.9.7.6Surveillerlatempératuredel'eau.Whattodo? → Checktheprocess.
English
Type MS02
35
Maintenanceandtroubleshooting
10.5.4 Message "Too low temperature value"
Displayedmessage Toolowtemperaturevaluesymboldisplayedontheiconoftheproduct,inthelistofdevicespossiblecause Thetemperaturevalueofthewatersampleexceedstheminimumerror
valuedefinedbytheuser.
Seechap.9.7.6Surveillerlatempératuredel'eau.Whattodo? → Checktheprocess.
Displayedmessage Toolowtemperaturevaluesymboldisplayedontheiconoftheproduct,inthelistofdevicespossiblecause Thetemperaturevalueofthewatersampleexceedstheminimum
warningvaluedefinedbytheuser.
Seechap.9.7.6Surveillerlatempératuredel'eau.Whattodo? → Checktheprocess.
English
Type MS02
36
Sparepartsandaccessories
11 SpAre pArTS AnD AcceSSorieS
caution
riskofinjuryand/ordamagecausedbytheuseofunsuitableparts.
Incorrectaccessoriesandunsuitablereplacementpartsmaycauseinjuriesanddamagetheproductandthesurroundingarea.
▶ UseonlyoriginalaccessoriesandoriginalreplacementpartsfromBürkert.
12 pAckAging, TrAnSporT
note
Damageduetotransport
Transportmaydamageaninsufficientlyprotectedproduct.
• Transporttheproductinshock-resistantpackagingandawayfromhumidityanddirt.
• Donotexposetheproducttotemperaturesthatmayexceedtheadmissiblestoragetemperaturerange.
• Protecttheelectricalinterfacesusingprotectiveplugs.
13 STorAge
note
poorstoragecandamagetheproduct.
▶ Obeythestorageinstructionsinthechapter13.1.
▶ Afteranystorageperiod,obeytheinstructionsinchap.13.2.
13.1 To store the product
1. Rinsetheproductwithtapwater.
2. Purgetheproductwithairatamax.pressureof2bar.
3. Storetheproductatroomtemperature(about23°C±10°C).
4. Storetheproductinadryplaceawayfromdust.
13.2 After storage
Beforecommisioningaproductthathasbeenstored:
1. Plugtheproductinthesystem.
2. Letthewatersampleflowthroughtheproductforatleast12hours.
3. Calibratetheproduct.Seechap.9.7.
English
Type MS02
37
Sparepartsandaccessories
14 DiSpoSAl of The proDucT→ Disposeoftheproductanditspackaginginanenvironmentally-friendlyway.
noteDamagetotheenvironmentcausedbyproductscontaminatedbyfluids.
• Keeptotheexistingprovisionsonthesubjectofwastedisposalandenvironmentalprotection.
English
Type MS02
38
English
Type MS02