LT1057X1 • 8/6/99 Serial Number Type 47900 & 48000 Furnaces OPERATION MANUAL AND PARTS LIST SERIES 1057 Model Numbers F47910 220-240 F47910-33 220-240 F47914 100 F47915 120 F47918 208 F47920 220-240 F47920-33 220-240 F47924 100 F47925 120 F47928 208 F47920-80 220-240 F47920-33-80 220-240 F47924-80 100 F47925-80 120 F47928-80 208 F47950 220-240 F47950-33 220-240 F47954 100 F47955 120 F47958 208 F48010 220-240 F48010-33 220-240 F48014 100 F48015 120 F48018 208 F48020 220-240 F48020-33 220-240 F48024 100 F48025 120 F48028 208 F48020-80 220-240 F48020-33-80 220-240 F48024-80 100 F48025-80 120 F48028-80 208 F48050 220-240 F48050-33 220-240 F48055 120 F48058 208 BARNSTEAD THERMOLYNE

Upload: others

Post on 20-Oct-2021

6 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Type 47900 & 48000 Furnaces - NIST

1LT1057X1 • 8/6/99 Serial Number

Type 47900 & 48000Furnaces

OPERATION MANUALAND PARTS LIST

SERIES 1057

Model Numbers

F47910 220-240F47910-33 220-240F47914 100F47915 120F47918 208F47920 220-240F47920-33 220-240F47924 100F47925 120F47928 208F47920-80 220-240F47920-33-80 220-240F47924-80 100F47925-80 120F47928-80 208F47950 220-240F47950-33 220-240F47954 100F47955 120F47958 208

F48010 220-240F48010-33 220-240F48014 100F48015 120F48018 208F48020 220-240F48020-33 220-240F48024 100F48025 120F48028 208F48020-80 220-240F48020-33-80 220-240F48024-80 100F48025-80 120F48028-80 208F48050 220-240F48050-33 220-240F48055 120F48058 208

BARNSTEAD THERMOLYNE

Page 2: Type 47900 & 48000 Furnaces - NIST

2

Thank you for buying this Thermolyne furnace. We believe that you will find it to be the bestfurnace of its kind available. Please note that this manual contains important operating andsafety information. You must carefully read and understand the contents of this manual priorto the use of this furnace.

Important Information

Page 3: Type 47900 & 48000 Furnaces - NIST

3

Model Information

MODEL NO. VOLTAGE CONTROL DISPLAY

F47910220-240 BASIC AUTOMATIC °CF47910-33 220-240 BASIC AUTOMATIC °C

F47914 100 BASIC AUTOMATIC °F47915 120 BASIC AUTOMATIC °CF47918 208 BASIC AUTOMATIC °CF47920 220-240 AUTOMATIC °F & °C

F47920-33 220-240 AUTOMATIC/EUROPEAN °F & °CF47924 100 AUTOMATIC °F & °CF47925 120 AUTOMATIC °F & °CF47928 208 AUTOMATIC °F & °C

F47920-80 220-240 PROGRAMMABLE (4 RAMP & 4 DWELL) °FF47920-33-80 220-240 PROGRAMMABLE/EUROPEAN (4 RAMP & 4 DWELL) °F

F47924-80 100 PROGRAMMABLE (4 RAMP & 4 DWELL) °FF47925-80 120 PROGRAMMABLE (4 RAMP & 4 DWELL) °FF47928-80 208 PROGRAMMABLE (4 RAMP & 4 DWELL) °F

F47950 220-240 PROGRAMMABLE W/COMMUNICATIONS (8 RAMP & 8 DWELL) °F & °CF47950-33 220-240 PROGRAMMABLE/EUROPEAN W/COMMUNICATIONS °F & °C

(8 RAMP & 8 DWELL)F47954 100 PROGRAMMABLE W/COMMUNICATIONS (8 RAMP & 8 DWELL) °F & °CF47955 120 PROGRAMMABLE W/COMMUNICATIONS (8 RAMP & 8 DWELL) °F & °CF47958 208 PROGRAMMABLE W/COMMUNICATIONS (8 RAMP & 8 DWELL) °F & °CF48010 220-240 BASIC AUTOMATIC °C

F48010-33 220-240 BASIC AUTOMATIC °CF48014 100 BASIC AUTOMATIC °CF48015 120 BASIC AUTOMATIC °CF48018 208 BASIC AUTOMATIC °CF48020 220-240 AUTOMATIC °F & °C

F48020-33 220-240 AUTOMATIC/EUROPEAN °F & °CF48024 100 AUTOMATIC °F & °CF48025 120 AUTOMATIC °F & °CF48028 208 AUTOMATIC °F & °C

F48020-80 220-240 PROGAMMABLE ( 4 RAM & 4 DWELL) °FF48020-33-80 220-240 PROGRAMMABLE/EUROPEAN (4 RAMP & 4 DWELL) °F

F48024-80 100 PROGRAMMABLE ( 4 RAMP & 4 DWELL) °FF48025-80 120 PROGRAMMABLE ( 4 RAMP & 4 DWELL) °FF48028-80 208 PROGRAMMABLE (4 RAMP & 4 DWELL) °F

F48050 220-240 PROGRAMMABLE W/COMMUNICATIONS (8 RAMP & 8 DWELL) °F & °CF48050-33 220-240 PROGRAMMABLE/EUROPEAN W/COMMUNICATIONS °F & °C

(8 RAMP & 8 DWELL)F48054 100 PROGRAMMABLE W/COMMUNICATIONS (8 RAMP & 8 DWELL) °F & °CF48055 120 PROGRAMMABLE 2/COMMUNICATIONS (8 RAMP & 8 DWELL) °F & °CF48058 208 PROGRAMMABLE W/COMMUNICATIONS (8 RAMP & 8 DWELL) °F & °C

Page 4: Type 47900 & 48000 Furnaces - NIST

4

Table of Contents

Important Information ........................................................................................................................................... 2Model Information ................................................................................................................................................ 3Safety Information ................................................................................................................................................ 5Introduction .......................................................................................................................................................... 7 General Usage ................................................................................................................................................... 7General Specifications ......................................................................................................................................... 8 Environmental Conditions .................................................................................................................................. 8Declaration of Conformity ..................................................................................................................................... 9Installation .......................................................................................................................................................... 10

Unpacking ...................................................................................................................................................... 10Site Selection ................................................................................................................................................. 10Electrical Connections ................................................................................................................................... 10

Furnace Operation, All Models ........................................................................................................................... 11Basic Automatic Control ..................................................................................................................................... 12Single Setpoint Temperature Control (Automatic) ............................................................................................. 154 Ramp & 4 Dwell Programmable Models ......................................................................................................... 188 Ramp & 8 Dwell Multi-Programmable Models ................................................................................................ 26Furnace Loading ................................................................................................................................................ 37Preventative Maintenance.................................................................................................................................. 38Problem Solving Tips ......................................................................................................................................... 39Maintenance and Servicing ................................................................................................................................ 41Replacement Parts List ...................................................................................................................................... 48Wiring Diagrams................................................................................................................................................. 49Material Safety Data Sheet ................................................................................................................................ 53Ordering Procedures .......................................................................................................................................... 55One Year Limited Warranty ................................................................................................................................ 56

Page 5: Type 47900 & 48000 Furnaces - NIST

5

Your Thermolyne Type 47900 or 48000 Furnacehas been designed with function, reliability, andsafety in mind. It is your responsibility to install it inconformance with local electrical codes. For safeoperation, please pay attention to the alert signalsthroughout the manual.

WarningsTo avoid electrical shock, this furnace must:1. Use a properly grounded electrical outlet

of correct voltage and current handlingcapacity.

2. Be disconnected from the power supplyprior to maintenance and servicing.

3. Have the door switch operating properly.

To avoid personal injury:

1. Do not use in the presence of flammableor combustible materials; fire or explosionmay result. This device containscomponents which may ignite suchmaterial.

2. Caution: Hot Surface - Avoid Contact. Toavoid burns, do not touch the exterior orinterior surfaces of this furnace during useor for a period of time after use.

3. Refer servicing to qualified personnel.

Safety Information

Hot SurfaceHot surfaces alert you to a

possibility of personal injury if you come in contact with a surface during use or for a period of time after use.

WarningWarnings alert you to a

possibility of personal injury.

CautionCautions alert you to apossibility of damage to theequipment.

NoteNotes alert you ro pertinentfacts and conditions.

Alert Signals

Page 6: Type 47900 & 48000 Furnaces - NIST

6

WARNINGThis warning is presented for compliance with California Proposition 65 and other regulatoryagencies and only applies to the insulation in this product. This product contains refractoryceramic, refractory ceramic fiber or fiberglass insulation, which can produce respirable dustor fibers during disassembly. Dust or fibers can cause irritation and can aggravate pre-existing respiratory diseases. Refractory ceramic and refractory ceramic fibers (afterreaching 1000°C) contain crystalline silica, which can cause lung damage (silicosis). TheInternational Agency for Research on Cancer (IARC) has classified refractory ceramic fiberand fiberglass as possibly carcinogenic (Group 2B), and crystalline silica as carcinogenic tohumans (Group 1).

The insulating materials can be located in the door, the hearth collar, in the chamber of theproduct or under the hot plate top. Tests performed by the manufacturer indicate that thereis no risk of exposure to dust or respirable fibers resulting from operation of this productunder normal conditions. However, there may be a risk of exposure to respirable dust orfibers when repairing or maintaining the insulating materials, or when otherwise disturbingthem in a manner which causes release of dust or fibers. By using proper handlingprocedures and protective equipment you can work safely with these insulating materialsand minimize any exposure. Refer to the appropriate Material Safety Data Sheets (MSDS)for information regarding proper handling and recommended protective equipment. Foradditional MSDS copies, or additional information concerning the handling of refractoryceramic products, please contact the Customer Service Department atBarnstead|Thermolyne Corporation at 1-800-553-0039.

WARNINGREFER SERVICING TO QUALIFIED PERSONNEL.

1-800-446-60601-800-553-0039

PLEASE NOTE THE FOLLOWING WARNINGS:

Page 7: Type 47900 & 48000 Furnaces - NIST

7

Intended UseThe Type 47900 and 48000 furnaces are generalpurpose laboratory and heat treating furnaces. Foroptimum element life, Thermolyne recommendsobserving these temperature ranges: from 100°C(212°F) to 1093°C (2000°F) for continuous use, orfrom 1093°C (2000°F) to 1200°C (2192°F) forintermittent use. Continuous use is operating thefurnace for more than 3 hours and intermittent useis operating the furnace for less than 3 hours.

All furnaces consist of: 1) a vented heatingchamber; 2) a temperature controller; and 3) a doorinterlock relay for operator safety.

General usageDo not use this product for anything other thanintended usage.

Principles of Operation

Furnace:The furnace chamber is heated by two open coilelectric resistance heaters and is insulated with aceramic fiber insulation. The control is locatedunder the furnace chamber and is well insulatedfrom the heat generated in the furnace chamber. Adoor safety switch removes power to the heatingelements whenever the furnace door is opened.The temperature is controlled by one of four typesof controllers.

Introduction

CautionDo not exceed limitations forcontinuous or intermittentoperating temperatureshown in the GeneralSpecifications. Exceedingthese limits will result inseverely reduced heatingelement life.

Page 8: Type 47900 & 48000 Furnaces - NIST

8

General Specifications

F479 ModelsDimensions

Chamber — Width: 12.7 cm. Height: 10.2 cm. Depth: 15.2 cm.Overall — Width: 28.6 cm. Height: 45.7 cm. Depth: 39.4 cm.

Weight — 16.8 kg.Electrical Ratings — Volts: 220-240. Amps: 4.2. Watts: 1000. Frequency: 50/60. Phase: 1Temperature

Operating Range (continuous): 1093°C; (intermittent): 1200°C.Environmental Conditions

Operating: 17°C - 27°C; 20% - 80% relative humidity, non-condensing. InstallationCategory II (over-voltage) in accordance with IEC 664. Pollution Degree 2 inaccordance with IEC 664. Altitude limit: 2,000 meters.

Storage: -25°C - 65°C; 20% - 80% relative humidity.

F480 ModelsDimensions

Chamber — Width: 17.8 cm. Height: 12.7 cm. Depth: 25.4 cm.Overall — Width: 33.7 cm. Height: 48.3 cm. Depth: 49.5 cm.

Weight — 25.4 kg.Electrical Ratings — Volts: 220-240. Amps: 7.5. Watts: 1800. Frequency: 50/60. Phase: 1Temperature

Operating Range (continuous): 1093°C; (intermittent): 1200°C.Environmental Conditions

Operating: 17°C - 27°C; 20% - 80% relative humidity, non-condensing. InstallationCategory II (over-voltage) in accordance with IEC 664. Pollution Degree 2 inaccordance with IEC 664. Altitude limit: 2,000 meters.

Storage: -25°C - 65°C; 20% - 80% relative humidity.

Environmental ConditionsOperating: 17°C - 27°C; 20% to 80% relative humidity, non-condensing. InstallationCategory II (over-voltage) in accordance with IEC 664. Pollution Degree 2 in accordancewith IEC 664. Altitude limit: 2,000 meters.Storage: -25°C to 65°C; 20% to 80% relative humidity.

Page 9: Type 47900 & 48000 Furnaces - NIST

9

Declaration of ConformityBarnstead|Thermolyne hereby declares under its sole responsibility that this productconforms with the technical requirements of the following standards:

EMC: EN50081-1 Generic Emission Standard; EN 50082-1 Generic ImmunityStandard.

Safety: IEC 1010-1-92 Safety requirements for electrical equipment for measurement,control and laboratory use; Part I: General RequirementsIEC 1010-2-010 Part II Particular requirements for laboratory equipment for theheating of materials

per the provisions of the Electromagnetic Compatability Directive 89/336/EEC, as amendedby 92/31/EEC and 93/68/EEC, and per the provisions of the Low Voltage Directive 73/23/EEC, as amended by 93/68/EEC.

The authorized representative located within the European Community is:Electrothermal Engineering Ltd.419 Sutton RoadSouthend On SeaEssex SS2 5PHUnited Kingdom

Copies of the Declaration of Conformity are available upon request

Page 10: Type 47900 & 48000 Furnaces - NIST

10

UnpackingVisually check for any physical damage to theshipping container. Inspect the equipment surfacesthat are adjacent to any damaged area. Open thefurnace door and remove the packing material frominside the furnace chamber. Vacuum the chamberprior to use to remove the insulation dust due toshipment. (The Type 48000 furnace is supplied withone ceramic shelf.)

Retain the original packaging material ifreshipment is foreseen or required.

Site SelectionInstall furnace on a sturdy surface and allowadequate space for ventilation.

Electrical ConnectionsThe electrical specifications are located on the

specification plate on the back of the furnace.Consult Barnstead/Thermolyne if your electricalservice is different than those listed on thespecification plate. Be sure the front power switch isin the OFF position before connecting the furnaceto your electrical supply.

Installation

CautionBe sure ambient temperaturedoes not exceed 104°F(40°C). Ambients above thislevel may result in damageto the controller.

Allow at least six inches ofspace between the furnaceand any combustiblesurface. This permits theheat from the furnace caseto escape so as not to createa possible fire hazard.

WarningTo avoid electrical shock,this furnace must always usea properly groundedelectrical outlet of correctvoltage and current handlingcapacity.

Page 11: Type 47900 & 48000 Furnaces - NIST

11

Furnace Operation, All Models

Power SwitchBoth the ON/OFF power switch and the digitaldisplay will illuminate when power is turned ON.The furnace will begin to heat to its controller'scurrent setpoint. (See the instructions for your typeof controller for information on checking and settingthe setpoint.)

Cycle LightThe amber cycle light will illuminate whenever thepower is being applied to the heating elements. Thecycle light will turn on and off as the furnacereaches the setpoint.

Door Safety SwitchThe door safety switch removes power from theheating elements when the door is opened. Openand close the door a few times; note that the amberCYCLE light will be out when door is open. Thischeck must be done when the furnace is heating orwhen the cycle light is illuminated. If this conditionis not true, consult the Troubleshooting sectionbefore proceeding.

To avoid personal injury, donot use in the presence offlammable or combustiblechemicals — fire orexplosion may result. Thisdevice contains componentswhich may ignite suchmaterials.

Caution: Hot Surface —Avoid Contact. To avoidburns, this furnace must notbe touched on the exterior orinterior surfaces during useor for a period of time afteruse.

Warning

NoteThe furnace will begin toheat to its controller's currentsetpoint each time you turnthe power switch "ON."

WarningTo avoid electrical shock,this furnace must have thedoor switch operatingproperly.

Page 12: Type 47900 & 48000 Furnaces - NIST

12

This furnace controller is a single set pointcontroller. This controller provides a single digitaldisplay to indicate either chamber temperature orsetpoint temperature. Also, this controllerincorporates sensor break protection, but does nothave over-temperature protection.

Digital ReadoutThe digital readout continuously displays chambertemperature unless the Push To Set Temperaturebutton is depressed. When the Push To SetTemperature button is depressed, the setpointtemperature is displayed.

Push To Set Temperature ButtonWhen this button is depressed, the digital displaywill indicate setpoint temperature. When the buttonis released, the actual chamber temperature isdisplayed.

Basic Automatic Control

Page 13: Type 47900 & 48000 Furnaces - NIST

13

Temperature Adjustment KnobTurning this knob clockwise will increase the setpointtemperature. Turning the knob counterclockwise willdecrease the setpoint temperature.

Operation of Basic AutomaticControlThis controller provides accurate control at one singletemperature setting. To set the temperature, simply:

a) Turn the Power switch ON.

b) While depressing the Push To SetTemperature button, turn the temperatureknob clockwise to increase the setpoint orcounterclockwise to decrease the setpoint tothe desired setpoint temperature asindicated on digital display.

c) Release the Push To Set Temperaturebutton.

BASIC AUTOMATIC CONTROL

Figure 1: Basic Automatic Controller

ON/OFFPower Switch

Cycle Light

Push To Set Temperature Button

Temperature SetpointAdjustment Knob

Page 14: Type 47900 & 48000 Furnaces - NIST

14

BASIC AUTOMATIC CONTROL

The digital display will now indicate the actualchamber temperature. The furnace will heat to thenew setpoint temperature. The CYCLE light willremain on until the furnace temperature is within1°C of the setpoint temperature; the cycle light willthen turn on and off as the controller maintains thesetpoint temperature.

Sensor Break ProtectionThis controller provides sensor break protection inthe event the thermocouple opens. If an openthermocouple condition occurs, the digital displaywill indicate 5 degrees or less and the power to theheating element will be shut off (CYCLE light willextinguish).

NoteIf at any time theTemperature Adjustmentknob is turned in eitherdirection, the setpoint willchange, even if the PushTo Set Temperature buttonis not depressed. To viewthe current setpointtemperature, depress thePush to Set Temperaturebutton.

Page 15: Type 47900 & 48000 Furnaces - NIST

15

Single Setpoint TemperatureControl (Automatic)

Note

Figure 2: Single Setpoint Controller

This furnace controller consists of amicroprocessor-based three-mode (Proportional,Integral, Derivative), single setpoint temperaturecontroller. The controller features programmableover–temperature protection, thermocouple breakindication and self-tuning capability.

Digital Readout:The Digital Readout continuously displays chamber(upper display) and setpoint (lower display)temperatures unless the “SCROLL” button isdepressed.

Initial Startup:When the power switch is turned on the controllerwill perform a self-test to make sure controller isoperating properly. (If all four 1’s do not light up orfails to go to “8888” contact Thermolyne.)

This temperature control is asingle setpoint device. Byusing the "UP" or "DOWN"buttons, a specifictemperature can be chosen.The control will cause thefurnace chamber to heat tothe chosen temperature andhold it at this temperatureuntil you turn off the powerswitch or select anothertemperature.

Furnace Chamber Temperature

Temperature Setpoint

Alarm Indicator

Cycle Light

Secret Key

Down Button Up Button

Scroll Button

Page 16: Type 47900 & 48000 Furnaces - NIST

16

Temperature Indication:This controller will display all temperatures in °C.Contact Thermolyne if control needs to be changedto °F.

Adjustment of ParametersThis controller's buttons are incorporated into itsfront panel. To illuminate the “DOWN” button,“SCROLL” button, and “UP” button, touch anywhereon front panel.

Setpoint Temperature:Push the “UP” button or the “DOWN” button tomodify the temperature setpoint (shown on thelower digital display). Once set, the setpoint willremain unchanged until you reset it. This allows youto set the controller for a consistent procedure, thensimply turn the furnace on and off with the powerswitch.

High Alarm (Over-TemperatureProtection OTP):Push the “SCROLL” button until “AL.SP” (AlarmSetpoint) appears.

Depress either the “UP” or “DOWN” button toselect the OTP value you desire. Thermolynerecommends that you set the value either at themaximum operating temperature of the furnace(1210°C) or a value of 20 degrees above yourworking temperature if you desire to provideprotection for your workload.

SINGLE SETPOINT TEMPERATURE CONTROL (AUTOMATIC)

NoteWhen performing thesesteps, remember that if morethan eight to ten secondselapse before the buttonsare used again, the displayscreen will automaticallyswitch back to displayingchamber temperature. If thishappens, light up the frontpanel again and step througheach parameter until youreach the point at which theinterruption occurred. Theparameter values youchecked earlier, however,will not be lost or altered.Holding down on the“SCROLL” button allowslonger viewing time.

NoteOvertemperatureprotection: The controller isfitted with a mechanical relaywhich is de-energized in thealarm mode. This relay,when de-energized, removespower from the heatingelements. If the primarycontrol circuit fails, the OTPwill control the furnacetemperature at the presetvalue you have entered. Itdoes not shut off the furnace,but will maintain the chambertemperature at that value.

Page 17: Type 47900 & 48000 Furnaces - NIST

17

Tuning:This control incorporates a self tuning feature whichdetermines the optimum control parameters for thebest temperature accuracy with your load andsetpoint. Use this feature the first time you use yourfurnace and each time you change either yoursetpoint or the type of load you are heating.Barnstead/Thermolyne highly recommendsusing this feature to provide the besttemperature accuracy the controller can attain.To use the Tuning feature:

First, adjust the setpoint to your desired value.Then, push the “SCROLL” button until “tunE”

appears. To start the tuning function push the “UP”button.

When the tuning process is started, the lowerdisplay will flash “tunE” along with the furnacetemperature setpoint. During tuning thetemperature setpoint cannot be changed. Tochange temperature setpoint, “tunE” must be turned“OFF.” To stop the tuning function push the“DOWN” button.

The next time you use your furnace, push the“SCROLL” button until “tunE” appears. Insure that“tunE” is “OFF.”

SINGLE SETPOINT TEMPERATURE CONTROL (AUTOMATIC)

NoteIf the power to furnace is“turned off” or interruptedwhile in “tuning,” uponreturning power to furnace,the controller display willindicate “LinE FAIL” becausesampled data could bequestionable. To restarttuning, refer to “Tuning”procedure.

If the controller cannotmaintain temperaturesetpoint, “tunE FAIL” willappear on display. Firstcorrect the problemresponsible for notmaintaining temperaturesetpoint, then restart “tuning”procedure.

Page 18: Type 47900 & 48000 Furnaces - NIST

18

This controller consists of a microprocessor basedthree-mode (Proportional, Integral, Derivative),programmable temperature controller with "Fuzzy"logic capability, programmable over-temperatureprotection and appropriate output switching devicesto control the furnace. The digital readoutcontinuously displays chamber (upper display) andsetpoint (lower display) temperatures unless the“SEL” button is depressed.

The programmable control can be used as asingle setpoint control or as a programmablecontrol.

Single Setpoint OperationTo use as a single setpoint control simply push“UP” or “DOWN” button to choose a specifictemperature.

1. The setpoint temperature presently set inthe control will be read out on the lowerdisplay.

4 Ramp & 4 DwellProgrammable Models

Figure 3: Programmable Control (4 ramp & 4 dwell)

Output Indicator

Enter ButtonUp Button

Furnace ChamberTemperature

Non-functional

Alarm Indicator

Non-functional

Setpoint Temperature

Down Button"SEL" Button

To avoid electrical shock,this furnace must have thedoor switch operatingproperly.

Warning

For optimum control of asingle setpoint operation, setthe "CTrL" parameter to"FUZY" to take advantage ofthe controller's fuzzy logiccapabilities.

Note

Page 19: Type 47900 & 48000 Furnaces - NIST

19

2. To change this setpoint, depress the “UP”or “DOWN” button until the desiredsetpoint value is displayed, then releasethe button.

3. At this point the furnace will begin to heatif the new setpoint temperature you havechosen is higher than the present chambertemperature.

Programmed Operation

Control ParametersYou can gain access to the control parameters by

holding the “SEL” button depressed for about 3seconds. "LoCt" will be displayed in the upperdisplay.

Depressing “SEL” once again shows the nextparameter and its current value on the display. Theparameter value can either be modified with thebuttons or left unmodified. Pressing the enter button(>) will allow you to modify the current value usingthe "Up" and "Down" buttons. Press the enterbutton again to set the new value as the currentvalue. (See "Programming the Controller" forinformation on programming.)

The control parameters in order are:

LoCt - Parameter lock level. Determines thenumber of parameters displayed when you press"SEL". It should be set to "0002." If it does not,set to "0002."

Nod - Mode. By pushing “UP” or “DOWN” buttons,can be set to "Auto" or "NAn" (Manual). Set to"Auto". In Manual mode, you must enter values tocontrol the furnace chamber temperature.

NoteWhen examining theparameters, if you depressand release either the “SEL,”“UP,” or “DOWN” buttonsand more than 1 minuteelapses before the buttonsare used again, the displayscreen will automaticallyswitch to displaying chambertemperature. If this happens,you will have to step througheach parameter until youreach the point at which theinterruption occurred. Theparameter values youchecked earlier, however,will not be lost or altered.

4 RAMP & 4 DWELL PROGRAMMABLE MODELS

NoteThe furnace will begin toheat to the current setpointtemperature when you turnthe power switch "ON." Toexamine the controlparameters without heatingthe furnace, depress andhold the “DOWN” button untilthe setpoint (lower) displayreads "20".

Page 20: Type 47900 & 48000 Furnaces - NIST

20

At - Autotuning feature. Will automatically setoptimum PID values for your process. ForTuning, set to "on." For normal operation, set to"oFF."

StAt- Status. Displays current status of programand cannot be changed.

Tine - Time. Displays the time remaining in thecurrent program.

ProG - Program options. Can be set to one ofthree options: "oFF", "rUn" or "hold". "oFF" willstop any running program and reset the programto the beginning. "rUn" will start a program fromthe beginning or resume a program from "hold"from the point it was held. "hold" will interrupt arunning program without resetting the program tothe beginning.

Sü-1 - Setpoint 1. The temperature at which thefirst ramp is aiming and at which the first dwellwill be held.

tN1r - Ramp time 1. The time the controller willtake to change the chamber temperature from itspresent temperature to the value of Setpoint 1.This time may be limited by the physicalcapabilities of the furnace.

tN1S - Dwell time 1. The length of time thecontroller will hold the furnace at the Setpoint 1temperature.

Sü-2- Setpoint 2.

tN2r - Ramp time 2.

tN2S - Dwell time 2.

Sü-3 - Setpoint 3.

4 RAMP & 4 DWELL PROGRAMMABLE MODELS

Page 21: Type 47900 & 48000 Furnaces - NIST

21

tN3r - Ramp time 3.

tN3S - Dwell time 3.

Sü-4 - Setpoint 4.

tN4r - Ramp time 4.

tN4S - Dwell time 4.

P-on - Power-on start. Push the "Up" or "Down"Buttons to set to "Yes" or "No." When set to"Yes," your furnace will immediately beginrunning the set program when you power on yourfurnace. For most applications, set to "No."

The next three parameters - Proportional band,Integral and Derivative - are the three controlparameters of a P.I.D. control system. The valuesfor these parameters are set during the tuningprocess. You should not change them; doing somay cause your controller to operate incorrectly.

P - Proportional band

i - Integral

d - Derivative

AL1T - Alarm type. This parameter determineshow the controller responds to any alarm input. Itshould display "0031." If it does not, set it to"0031."

AL11 - Furnace overtemperature alarm. Thisparameter determines the absolute upper limit onthe furnace chamber temperature. Werecommend a setting of "1200" since 1200°C isthe maximum temperature your furnace canobtain without damage to furnace components.

NoteOvertemperature protection:The controller is fitted with amechanical relay which is de-energized in the alarm mode.This relay, when de-energized,removes power from theheating elements. If theprimary control circuit fails, theOTP will control the furnacetemperature at the value youhave entered for AL11. It doesnot shut off the furnace, but willmaintain the chambertemperature at that value.

4 RAMP & 4 DWELL PROGRAMMABLE MODELS

Page 22: Type 47900 & 48000 Furnaces - NIST

22

You may wish to set the parameter lower if youneed to ensure a crucial process is protectedfrom high temperature.

AL12 - High deviation alarm. This parameterdetermines how far the furnace chambertemperature can rise above your setpoint. Werecommend a setting of 20; since a P.I.D. controlprocess involves some degree of temperatureovershoot, a setting of less than 20 may interferewith the function of your controller.

LooP - Thermocouple break alarm detectiontime. For proper operation of your furnace, thisparameter should be set to "0.00". If it is not, setit to "0.00".

CTrL - Control type. Push the "Up" or "Down"buttons to set to "Pid" or "FUZY". Programmedoperation requires P.I.D. control. Single setpointoperation will achieve better control using Fuzzylogic control.

Programming ControllerThis controller allows you a maximum of 4 ramp

and 4 dwell segment combinations, thus enabling 4different setpoint levels to be achieved. Each rampis programmed by specifying the endingtemperature level and the required time to ramp tothat temperature. The controller then automaticallycalculates the ramp rate required to attain theending temperature level based on the desiredramp time. Dwell segments then can be attached toeach temperature level to hold that temperature fora specified amount of time.To run a program, first determine your ramp times,ending temperature levels and dwell times. It ishelpful to graph your program out for ease ofloading the program into the controller.

Typical Profile for Programmable Models(Other Profiles May Be Formed)

CautionRemember that wheneverthe power switch is turned“ON,” the furnace will beginto heat at the setpointtemperature that waspreviously set. This value willremain unchanged for up toa year without power beingapplied to the control.

4 RAMP & 4 DWELL PROGRAMMABLE MODELS

Page 23: Type 47900 & 48000 Furnaces - NIST

23

If you desire to skip a rampor dwell segment, follow thisprocedure:

• for a dwell segment, enter asetting of “0” minutes.

• for a ramp segment, enter asetting of "00.00"hour.minutes. This will causecontroller to skip to nextsegment as fast as thefurnace can heat or cooldown.

Note

Make sure "ProG" (program options) is set to"oFF" (to stop the program) when entering programvalues.

Push “SEL” button until “Sü-1” is displayed. Pushthe Enter button (>) to select the current value.Push “UP” or “DOWN” button and set Setpoint 1in °C. Push the Enter button to enter the setvalue as the current value.

Push “SEL” button until “TN1r” is displayed. Pushthe Enter button (>) to select the current value.Push “UP” or “DOWN” button and set Ramp time1 in hours and minutes in the format "X.XX"(hour.minutes). This is the time Ramp 1 will taketo reach Setpoint 1. Push the Enter button toenter the set value as the current value.

Push “SEL” button until “TN1S” is displayed. Pushthe Enter button (>) to select the current value.Push “UP” or “DOWN” button and set Dwell time1 in hours and minutes in the format "X.XX"(hour.minutes). Push the Enter button to enterthe set value as the current value.

Set “Sü-2” “TN2r” “TN2S” “Sü-3” “TN3r” “TN3S”“Sü-4” “TN4r” and “TN4S” in the same manner as“Sü-1” “TN1r” “TN1S”.

Program ExecutionPush “SEL” button until “ProG” is displayed.Pushing “UP” or “DOWN” button, select "run."When “run” is selected, the program will start fromthe actual furnace temperature at that point in time.

SetpointWhile the program is in “run” or “Hold,” the setpointshown on the bottom display is the current workingsetpoint.

4 RAMP & 4 DWELL PROGRAMMABLE MODELS

Page 24: Type 47900 & 48000 Furnaces - NIST

24

Program EndWhen the program ends, the controller will hold thefurnace at the value of Setpoint 4 until you stop theprogram.

Program StopTo stop program, push “SEL” button until “ProG” isdisplayed. Push “UP” or “DOWN” button until “oFF”is displayed. This will terminate program.

Self Tuning FeatureThis programmable control has an automatic tuningfeature which installs optimum tuning parameters togive the best temperature accuracy with your loadand setpoint or program. No manual loading oftuning parameters is needed. Barnstead/Thermolyne highly recommends using thisfeature to provide the best temperatureaccuracy the controller can attain. Use thisfeature the first time you use your furnace and eachtime you change your setpoint or program or thetype of load you are heating. To use the Tuningfeature:

1. Select the temperature at which you intendto operate. If you will be running aprogram, enter the value of Setpoint 1 as asingle setpoint.

2. Load your furnace with a characteristicsample of the load you will be heating.

3. Push “SEL” button until “AT” is displayed,then push “UP” or “DOWN” button to turn“AT” “on.”

4 RAMP & 4 DWELL PROGRAMMABLE MODELS

The temperature control inthese models is aprogrammable “and”automatic single setpointdevice. When the programhas ended, the controller willmaintain the chambertemperature at a value equalto the last programmed level(Setpoint 4) until the programis cancelled. It will notautomatically cool to ambientunless Setpoint 4 is set atambient. When a program iscancelled, the controller willmaintain the chambertemperature at a value equalto the single setpoint modesetpoint.

Note

NoteFor processes which cannottolerate the overshootnaturally generated duringthe tuning process, select"Lo" rather than "on" for the"AT" setting. The "Lo" settingwill tune your furnace at atemperature below yoursetpoint, then apply theresult to your setpoint. Thisapproach prevents thenecessary tuning overshootfrom exceeding your actualsetpoint.

Page 25: Type 47900 & 48000 Furnaces - NIST

25

During the operation, “TunE” flashes in the lowerdisplay. Do not make any adjustments to thecontroller parameters during this period. The selftuning is finished when “TunE” no longer flashes inthe lower display.

Self tuning will calculate values for:Proportional band - PIntegral - iDerivative - d

• Self tuning cannot be initiated whilerunning a program.

• A power failure will cause the ATparameter to revert back to “OFF”. (Resettune parameter to “ON” using the “UP”button).

• In the case of alarm conditions duringtuning, those conditions will flashalternately with "TunE."

The next time you use your furnace, push “SEL”button until “AT” is displayed, then push “UP” or“DOWN” button to turn tune “oFF.”

4 RAMP & 4 DWELL PROGRAMMABLE MODELS

Page 26: Type 47900 & 48000 Furnaces - NIST

26

This furnace controller consists of amicroprocessor-based three-mode (Proportional,Integral, Derivative), programmable control withovertemperature protection and appropriate outputswitching devices to control the furnace. The upperdigital display indicates the measured chambertemperature and the lower display indicates themain setpoint temperature unless the scroll buttonis depressed. (See Figure 4.)

The programmable control can be used as asingle setpoint control or as a programmablecontrol.

Single Set Point OperationTo use as a single set point control, simply push upor down buttons to choose a specific temperature.

1. The setpoint temperature presently set inthe control will be read out on the lowerdisplay.

8 Ramp & 8 Dwell Multi-Programmable Models

Figure 4: Multi-Programmable Control (8 ramp & 8 dwell)

NoteTo change from °C indicationto °F indication, contactBarnstead/Thermolyne.

Alarm Active Temperature Setpoint

Screw Retaining Instrument in Sleeve

Furnace ChamberTemperature(At Startup, DisplaysSoftware VersionNumber)DigitalCommunications

Condition of Program,Ramp or Dwell and ifProgram is Held, RampDwell Segment CurrentlyRunningUp Button

Run/Hold Button

Down Button

ParameterSelected onSecondary Display:SP = SetpointOP = Output PowerTime = TimeRemaining inProgram

Adaptive or SelfTune Selected (AT)

Scroll Button

Nonfunctional

Page 27: Type 47900 & 48000 Furnaces - NIST

27

2. To change this set point, depress the “UP”or “DOWN” push button until the desiredsetpoint value is displayed, then releasethe button.

3. At this point the furnace will begin to heatif the new set point temperature you havechosen is higher than the present chambertemperature.

4. The upper display indicates actual furnacetemperature.

The two center push buttons are inactive and notused.

Programmed Operation

Control ParametersControl parameters are viewed and changed

using the “SCROLL” button. If the scroll button isdepressed and released, the lower display willindicate output power (OP) or setpoint (SP). This isreferred to as the “short scroll.” Continued singlestep action of scroll button will cause lower displayto alternate between setpoint (SP) and outputpower (OP).

To enter the main scroll list (list of all controllerparameters that are accessed through frontkeyboard), the scroll button should be helddepressed. Pnr1 (program number 1) will appear.To progress through the parameter list, release thescroll button, then use single step depression toadvance through the list. You can rapidly progressthrough the parameter list by holding the scrollbutton depressed. For all the parameters, pushingthe up or down button will give current setting.

8 RAMP & 8 DWELL MULTI-PROGRAMMABLE MODELS

NoteThe furnace will begin toheat to the current setpointtemperature when you turnthe power switch "ON." Toexamine the controlparameters without heatingthe furnace, depress andhold the “DOWN” button untilthe setpoint (lower) displayreads "20".

Do not exceed limitations forcontinuous or intermittentoperating temperatureshown in the GeneralSpecifications. Exceedingthese limits will result inseverely reduced heatingelement life.

Caution

Page 28: Type 47900 & 48000 Furnaces - NIST

28

The temperature control parameter values inorder are:

Pnr 1 - Program number. By pushing the up ordown button, you can select a program from 1 to4. Each program will contain independently thefollowing parameters:

PR1 - Program Ramp Rate. The rate of heatincrease or decrease in °C/minutes.

PL1 - Program Level. The temperature to whichthe furnace needs to attain.

PD1 - Program Dwell 1. Amount of time in minutesto hold PL1 program level temperature entered.

The remaining Program Ramp Rates PR2 - PR8,Program Levels PL2 - PL8, and Program DwellsPD2 - PD8 follow with the same structure as PR1,PL1, and PD1.

Cnt - Continued. Allows linking of programs. Cntmay be selected as “y” (yes) or “n” (no) bypushing the up or down button.

HB - Holdback. Automatically places the controllerinto “Hold” if the measured value deviates morethan a specified amount from controller setpoint.When measured value re-enters the holdbackband, the timing for the segment resumes.(Parameter is expressed in °C and only functionswhen running a program).

PLC - Program Loop Count. The number of timesa program will be repeated.

SP1 - Setpoint One. Indicates current setpoint.

SP2 - Setpoint two. Not configured into control andnonfunctional.

8 RAMP & 8 DWELL MULTI-PROGRAMMABLE MODELS

NoteWhen examining theparameters, if you depressand release either the “PAR,”“UP,” or “DOWN” buttonsand more than 6 secondselapse before the buttonsare used again, the displayscreen will automaticallyswitch to displaying chambertemperature. If this happens,you will have to step througheach parameter until youreach the point at which theinterruption occurred. Theparameter values youchecked earlier, however,will not be lost or altered.Holding down on “PAR”allows longer viewing time.

NoteOnce the desired parameterhas been selected,depressing either the raise orlower button will cause theparameter to be replacedwith the new value. At thispoint, the “top dot” of theleast significant digit of thesecondary display will flashon and off. Any further use ofup or down buttons willchange parameter value. Inall cases, the value shownon the display is the currentworking value of thatparameter.

Page 29: Type 47900 & 48000 Furnaces - NIST

29

ST - Self Tune. Automatically loads PID values oninitial start up. (See Furnace Operation forfunction of Self Tune).

AT - Adaptive Tune. Analyzes and inputs optimumPID values when temperature has reachedsetpoint. (See Furnace Operation for function ofAdaptive Tune).

SAT - Self Adaptive Tune. When engaged, startscontroller off in self-tune mode then automaticallyswitches to Adaptive Tune (AT). (See FurnaceOperation for function of Self Adaptive Tune).

ATR - Adaptive Tune Band setting. Determinesthe operational band width of the adaptive tuningfunction. Self-tuning automatically determinesthis setting.

AL1 - Alarm 1. A full scale alarm which protectsload and furnace when temperature exceedspreset value. Furnace will control temperature atthe preset temperature value; it will not shut offfurnace.The next three parameters – Proportional (PB)

Integral (+i) and Derivative (+d) – are for highaccuracy control. Self-Adaptive Tuning will changethese parameters to their optimum values. Youshould not change them; doing so may cause yourcontroller to operate incorrectly.

The next two parameters – cutback low (cbl) andcutback high (cbh) – are to aid the control inpreventing temperature overshoots andtemperature undershoots. The point from setpointwhere the power starts “cutting back” is defined asthe cutback value. These values are alsoautomatically adjusted by the Self Tune andAdaptive Tune features. These values cannot bechanged by the user; the controller automaticallyinstalls optimum cutback values when in Self andAdaptive Tuning.

8 RAMP & 8 DWELL MULTI-PROGRAMMABLE MODELS

Page 30: Type 47900 & 48000 Furnaces - NIST

30

HL - Output Power limits the average maximumpower that is applied to the heating elements.Normal setting is 100%. If you plan to use thefurnace below 260°C (500°F) the output powermay be reduced. This will significantly shortenthe time it takes for stabilization. It will alsoreduce drastic temperature overshoots. ContactBarnstead/Thermolyne Customer Service foradvice on the proper value to use. Rememberthat this parameter does not reduce the voltageto the elements. It reduces the average power tothe elements by cycling power on and off.

HC - Cycle Time is the rate at which power issupplied to power control switch.

Sbr - the percent of power that is supplied to thecontrol output terminals if an openthermocouple condition exists. Push up ordown button. 0.0 will be displayed. Thisparameter cannot be changed; if 0.0 is notdisplayed, contact Barnstead/Thermolyne. Theupper display will flash “OR” if an openthermocouple condition exists.

Programming ControllerThe multi-programmable controller in these unitsprovides up to 4 separate programs of 8 ramps and8 dwells each. This controller also allows you to linkprograms together, which allows you to achieve 64total segments (4 programs X 16 segments). Thesefunctions are controlled by two of the controller’sprogramming parameters, “Pnr” and “Cnt.”

To run a program, first determine your ramp rate,dwell times and program levels. It is helpful tograph your program out for ease of loading programinto controller.

A maximum of 8 ramp and 8 dwell segmentcombinations are available per program, thusenabling eight different setpoint levels to be

8 RAMP & 8 DWELL MULTI-PROGRAMMABLE MODELS

Page 31: Type 47900 & 48000 Furnaces - NIST

31

achieved. Each ramp is programmed by specifyingthe program level (PL) and the required ramp rate(PR). The controller then automatically calculatesthe time that is required to attain the program level(PL) based on the desired ramp rate (PR). Dwellsegments (soak) then can be attached to eachprogram level (PL) to hold that temperature for aspecified time.

To select program numberPush scroll button until “Pnr 1” is displayed. Pushthe up or down button to select a program numberfrom 1 to 4.

To link programs togetherPush scroll button until “Cnt n” is displayed. Pressand release the up and down buttons to switchbetween “Cnt y” (continue yes) and “Cnt n”(continue no). The effect of selecting “Cnt y” is tocontinue the program to the next program number.For example, if in program #3 you select “Cnt y,”when program #3 is complete, program #4 will runautomatically. Setting “Cnt y” in program #4 willinitiate the start of program #1 upon the completionof program #4. Each program will complete theselected number of loops before continuing (seeLoop Count). If you do not want to link programs,set Cnt to “Cnt n” (continue no).

Program Entry1. With the controller not operating, indicated

by the bottom right hand side of thedisplay extinguished, depress scroll buttonuntil PR1 is displayed. Push the up ordown button to scroll to the desired value,which is degrees per minute.

Scrolling down below zero will give three otheroptions for the ramp:

8 RAMP & 8 DWELL MULTI-PROGRAMMABLE MODELS

Page 32: Type 47900 & 48000 Furnaces - NIST

32

NONE-which will force the program to skipto the next segment;

END-which will cause the program to stopor restart if loops remaining is not zero;

STEP-which will cause the program tostep to the next program level.

All other ramps in the program are set in a similarfashion by selecting “PR” followed by the relevantramp number.

2. The level to which the first ramp is aimingis entered by scrolling through the mainscroll list until “PL1” is displayed. Bypressing either the up or down button thepresent value of this level is indicated indisplay units. Using the up or down buttonwill scroll the present value to the newvalue required. All other levels in theprogram are set in a similar fashion byselecting “PL” followed by the relevantnumber.

3. To set the dwell time for the first level,scroll through the main scroll list until“Pd1” is displayed. Pressing the up ordown button will reveal the current value oftime in minutes. Using the up or downbutton will scroll the present value to thenew value required. Scrolling this valuedown will allow a setting of “END.” Asetting of “END” will terminate theprogram, or force it to restart if loopsremaining are not zero at the beginning ofthat dwell.

All other dwells in the program can be set in asimilar fashion by selecting “PD” followed by therelevant dwell number.

8 RAMP & 8 DWELL MULTI-PROGRAMMABLE MODELS

Page 33: Type 47900 & 48000 Furnaces - NIST

33

4. Scrolling through the main scroll until theparameter “PLC” is displayed and thendepressing the up or down button willreveal the present setting of the loopcount. This is the number of times that theentered program will be repeated before acontinuous setpoint at the last level of theprogram is achieved. By pushing the up ordown button the number of loops can beset at any value from 1 to 999.

HoldbackThis controller features a “holdback” (Hb) functionto ensure programmed parameter values areadhered to. Holdback is set in display units(degrees) and represents the allowable excursionof measured value away from the current setpoint,either above or below, before the program is forcedinto hold (clock stops).

The program will remain in hold until themeasured value comes within the holdback band(clock starts). This feature is active the whole timethat the program is running. (Holdback functionsonly while running a program).

Scroll through the main scroll list until “HB” isdisplayed. Push the up or down button to reveal thecurrent value of holdback. The up or down buttoncan now be depressed to scroll to the requiredvalue. Select desired holdback setting – a setting of20° is recommended. Holdback is set in displayunits and represents the allowable excursion ofmeasured value away from the current setpoint,either above or below, before the program is forcedinto hold. The program will remain in hold until themeasured value comes within holdback limits. Thisfeature is active the whole time that the program isrunning. When hold is forced onto the program byholdback, the “HOLD” legend is not illuminated, buteither the “RAMP” or “DWELL” legend will flash.

8 RAMP & 8 DWELL MULTI-PROGRAMMABLE MODELS

Page 34: Type 47900 & 48000 Furnaces - NIST

34

When the program is in “holdback,” it effectivelylengthens the time of the program – if the holdbackband “Hb” is set too low, the program will neverescape the holdback band, thus the program willnever be completed. If you do not want to use theholdback function, set “Hb” to an extremely largevalue.

Program ExecutionOnce the program has been entered it can be runby depressing the “RUN/HOLD” push button on thefront.

With the run initiated, the program will commenceand the legend on the display will indicate if a rampor dwell is being performed. While a program isrunning, the short scroll will contain a thirdparameter “TIME.” Push the scroll button once; timeremaining for the current segment, either ramp ordwell, will be indicated. If the loop counter has beenset to any value other than one, then the aboveprocedure will be repeated for each loop.At the end of the complete program, an “E” willappear on the display.

Parameter Change While RunningThe previous parameters can be inspected but notchanged while a program is running. If it isnecessary to alter a parameter while a program isrunning, the program must be placed into the holdcondition. To put program into hold, push the “RUN/HOLD” button once. After modification of theparameter, returning the program to the run statewill cause the program to continue with thechanged value(s) installed. Push “RUN/HOLD”button again to restart program.

8 RAMP & 8 DWELL MULTI-PROGRAMMABLE MODELS

NoteBe sure to select theprogram number beforepressing “RUN/HOLD”.

The temperature control inthese models is aprogrammable “and”automatic single setpointdevice. When the programhas ended, the controller willmaintain the chambertemperature at a value equalto the last programmed level(PL) until the program iscancelled. It will notautomatically cool to ambientunless last programmedlevel (PL) is set at ambient.When a program iscancelled, the controller willmaintain the chambertemperature at a value equalto the main setpoint (SP1 orSP). To cancel a program,depress and release the“UP” and “DOWN” pushbuttons simultaneously. Besure single set point mode isset to 20 degrees asdescribed earlier in thismanual.

Note

Page 35: Type 47900 & 48000 Furnaces - NIST

35

Loop CountIf the loop count is set to values other than one,then the number of loops remaining in a runningprogram can be displayed. To determine which loopis being performed depress scroll button until "LR"is displayed and by pushing either the up or downbutton the remaining number of loops, excludingthe one being executed, is displayed.

Program HoldA running program can be forced into hold at anystage by depressing the “RUN/HOLD” push buttonon the front. When a running program is forced intohold, the “HOLD” legend will appear on the displaytogether with the segment type and will be flashing.Pushing “RUN/HOLD” button again will return theprogram to a run situation and extinguish the“HOLD” legend.

Program ResetA running, held or finished program can be reset bydepressing the up and down push buttons together.

When the reset has been enabled, the parts ofthe display associated with programming will beextinguished and the controller will operate as asingle setpoint control.

SAT Self Adaptive TuningThis programmable control has automatic tuningfeatures which install optimum tuning parameters togive the best temperature accuracy with your loadand setpoint. No manual loading of tuningparameters is needed. Thermolyne highlyrecommends using these features to providethe best temperature accuracy the controller

8 RAMP & 8 DWELL MULTI-PROGRAMMABLE MODELS

Page 36: Type 47900 & 48000 Furnaces - NIST

36

can attain. Use this feature the first time you useyour furnace and each time you change yoursetpoint or program or the type of load you areheating.

The following paragraphs instruct you on how toinitiate the SAT Self and Adaptive Tuning feature.This feature starts the controller in the Self Tunemode then automatically switches over to theAdaptive Tuning Feature.

Self Tuning is a one-time function which permitsyou to retune the instrument control parameters tosuit new process conditions. Adaptive tuning takesover when the self tune is completed andcontinuously reevaluates the tuned parameters.Adaptive tuning will then automatically install newvalues if a better response could have beenattained. To use the tuning features:

1. Depress scroll button until SAT isdisplayed. Depress the up and downbuttons simultaneously to start self tuning.The A-T indicator is then illuminated(upper right hand corner) and the lowerdisplay indicates the setpoint at which theself-tune sequence will occur. The “SP”indicator will flash for 1 minute, duringwhich time the setpoint may be changed ifit is required to retune at a new setpointeither above or below the process valueindicated on the upper display (usetemperature setpoint that your applicationrequires). At the end of the minute, the“SP” indicator will stop flashing, indicatingthat the setpoint can no longer bechanged. The A-T indicator will startflashing and continue to flash until the selftune has completed. Once the self tune iscompleted, adaptive tune takes over andthe A-T indicator will remain illuminated.

To stop tuning, function scroll until SAT isdisplayed and simultaneously push up and downbuttons.

8 RAMP & 8 DWELL MULTI-PROGRAMMABLE MODELS

Page 37: Type 47900 & 48000 Furnaces - NIST

37

For best results, use only the center two-thirds ofthe furnace chamber.

If you are heating a number of small parts,spread them throughout the center two-thirds of thefurnace chamber.

Keep objects away from the thermocouple.Use insulated tongs and mittens when loading

and unloading the furnace.Always wear safety glasses.Never come into contact with the heating

elements. Hitting the elements with tongs or layingthe load against them will cause the elements toburn out prematurely.

Furnace Loading

Do not overload your furnacechamber. If the load is to beheated uniformly, it shouldnot occupy more than two-thirds of the furnacechamber. Maintain at least a3/4 inch gap between theload and elements. Failure toobserve this caution couldresult in damage to furnacecomponents.

Caution

Heating Elements

Thermocouple Sample Port

Thermocouple

Page 38: Type 47900 & 48000 Furnaces - NIST

38

This unit is equipped with a venting system on thetop of the furnace. This is for the removal of fumesfrom the chamber of the unit. Contamination is amajor cause of element failure, therefore, removeall fume forming material before heating. (e.g. cleancutting oil from tool steel).

Housekeeping is vital to your electric furnace –KEEP IT CLEAN. Run your furnace up to 871°C(1600°F) empty occasionally to burn off thecontamination that may exist on the insulation andelements. Maintain 871°C (1600°F) for at least 4hours to insure complete ashing of foreignmaterials.

Element life is reduced somewhat by repeatedheating and cooling. If the furnace is to be usedagain within a few hours, it is best to keep it at theoperating temperature or at a reduced level such as260°C (500°F).

Thermolyne highly recommends that you replacethe thermocouple periodically (once every sixmonths) to ensure temperature accuracy.

General Cleaning InstructionsWipe exterior surfaces with lightly dampened clothcontaining mild soap solution.

Preventative Maintenance

Page 39: Type 47900 & 48000 Furnaces - NIST

39

The Problem Solving Tips section is intended to aid in defining and correcting possibleservice problems. When using the chart, select the problem category that resembles themalfunction. Then proceed to the possible causes category and take necessary correctiveaction.

Problem Solving Tips

PROBLEM POSSIBLE CAUSES CORRECTIVE ACTION

The cycle light does not illuminate. The furnace is not connected topower supply.

Incorrect power source.

ON and OFF power switchdefective.

Defective cycle light.

Door switch defective.

Reconnect furnace to powersupply.

Connect to correct power source.

Replace power switch.

Replace cycle light.

Re-align or replace furnace doorswitch.

The furnace does not heat. No power.

Defective electrical hookup.

Thermocouple has oxidized andopened the circuit.

Controller malfunction.

Heating elements in 208V, 230V or240V furnaces are burned out.

Door switch malfunction.

Defective safety relay.

Defective solid state relay.

Check power source and fuses orbreakers.

Repair electrical hookup.

Replace thermocouple.

Replace controller.

Replace defective elements.

Re-align or replace door safetyswitch.

Replace safety relay.

Replace output relay.

Door switch does not cut power toheating elements.

Door switch is not functioning.

Safety relay malfunction.

Re-align or replace door safetyswitch.

Replace safety relay.

Page 40: Type 47900 & 48000 Furnaces - NIST

40

PROBLEM SOLVING TIPS

PROBLEM POSSIBLE CAUSES CORRECTIVE ACTION

Controller over-temperature alarmdoes not cut power to furnacechamber.

Alarm output device malfunction.

Mechanical relay malfunction.

Element shorted to ground.

Replace control unit.

Replace mechanical relay.

Replace faulty element.

Slow heatup. Low line voltage.

Heavy load in chamber.

Wrong heating element.

One heating element is burned outin 100V & 120V furnace.

Install line of sufficient size andproper voltage. (Isolate furnace fromother electrical loads.)

Lighten load in chamber to allowheat to circulate.

Install proper element.

Replace burned out element.

Repeated element burnout. Overheating furnace.

Heating harmful materials.

Contamination present fromprevious burnout.

Keep furnace under maximumtemperature. Closer supervision ofcontrol setting.

Enclose material in container. Cleanup spills in and on chamber.Ventilate chamber by leaving topvent slightly open when heatingknown harmful reagents.

Replace insulation material.

Inaccurate temperature readout. Oxidized or contaminatedthermocouple.

Poor thermocouple connection.

Improper loading procedures.

Poor ventilation of base.

Thermocouple connectionsreversed.

Replace thermocouple.

Tighten connections.

Use proper loading procedures.

Clear area around furnace base.

Reconnect thermocouple correctly.

Page 41: Type 47900 & 48000 Furnaces - NIST

41

To Replace a Heating Element:1. Disconnect the furnace from the power

supply.

2. Remove the back terminal cover of thefurnace. (Note placement and connectionof wires.)

3. Loosen the nuts on the terminal points ofthe four heating element lead wires andremove the element wires from theterminals.

4. Loosen the thermocouple hold-down clipand carefully remove the thermocouplefrom the rear of the furnace chamber.

5. Remove both the back of the furnace andthe blanket insulation.

6. Grasp the element lead wires and pull outboth elements and the back chamberinsulation.

7. Replace the defective element.

Maintenance and Servicing

WarningTo avoid electrical shock,this furnace must always bedisconnected from the powersupply prior to maintenanceand service.

Perform only maintenancedescribed in this manual.Contact an authorized dealeror our factory for parts andassistance.

Refer servicing to qualifiedpersonnel.

Page 42: Type 47900 & 48000 Furnaces - NIST

42

8. Reverse the disassembly procedure,making sure you thread all element leadwires through the insulating porcelainbushings on the back of the furnace andcut off any excess element lead wire aftersecuring the leads to the terminal points.

9. Reconnect the furnace to the powersupply.

10. Test the operation of the furnace.

To Replace a Chromel/AlumelThermocouple (All Furnaces):

1. Disconnect the furnace from the powersupply.

2. Remove both back covers. (Noteplacement and connection of wires.)

3. Remove the clip holding the thermocouplein place (1 screw), and remove the twoscrews on the thermocouple terminals.

MAINTENANCE AND SERVICING

Page 43: Type 47900 & 48000 Furnaces - NIST

43

4. Remove the thermocouple (Note: first pullthe thermocouple straight out of the hole inthe chamber to avoid damage to theinsulation.)

5. Guide the looped ends of the newthermocouple through the plastic bushingswith the red (-) lead to the right as youface the back of the furnace.

6. Form a 90° bend towards the furnace, withthe red (-) lead to the right. Insert thethermocouple straight through the hole inthe chamber.

7. Secure the thermocouple with clip andscrew. Connect the looped ends of thethermocouple to the terminals with + to +

(yellow) and - to - (red). (A polarity test ofthe thermocouple and lead wire is easilymade with the use of a magnet. OnChromel/Alumel thermocouples and leadwire, the non-magnetic wire is positive (

+)

and the magnetic wire is negative (-).)

8. Replace both back covers.

MAINTENANCE AND SERVICING

Thermocouple Leads+_

Page 44: Type 47900 & 48000 Furnaces - NIST

44

9. Reconnect to the power supply.

10. Test the operation of the furnace.

To Replace Door Switch (Micro-Switch):1. Disconnect the furnace from the power

supply.

2. Remove the two top screws on the frontdial and the two bottom screws and 1 lockwasher on the back cover.

3. Slide the control section forward. (Note: donot pull excessively on the internal wires.)

4. Disconnect the wires from the door switch.(Note the connection placement of thewires to the micro-switch.)

5. Remove the two screws, lock washers andnuts from the micro-switch.

6. Insert a new micro-switch and secure withthe screws, lock washers and nutsremoved in step 5.

MAINTENANCE AND SERVICING

Page 45: Type 47900 & 48000 Furnaces - NIST

45

MAINTENANCE AND SERVICING

7. Reconnect the wires identified or markedin step 4 to the new door switch.

8. To realign the door switch, see thefollowing section: To Realign DoorSwitch.

9. Slide the control section back and replacethe screws and lock washer described instep 2.

10. Reconnect to the power supply.

11. Test the operation of the door switch. (Seenext section, step 10.)

To Realign Door Switch (Micro-Switch):

1. Disconnect the furnace from the powersupply.

2. Remove the two top screws on the frontdial and the two bottom screws and lockwasher on the back cover.

Door Switch

Page 46: Type 47900 & 48000 Furnaces - NIST

46

3. Slide the control section forward. (Note: donot pull excessively on the internal wires.)

4. With the door closed, loosen the screwson the micro-switch and slide the switchdownward, so that the screws are at thebottom of the slots in the mountingbracket.

5. Finger tighten both screws. While holdingdown the rear of the micro-switch housing,gently push up on the front of the switchuntil you hear a click.

6. Open and close the door; the switchshould click when the door is openedapproximately 3" and 1" to 2” before thedoor is closed. Slide the front of the switchup to increase the distance, down todecrease the distance.

7. Tighten the two screws to secure themicro-switch. Check the operation of theswitch as described in step 6 aftertightening the screws.

MAINTENANCE AND SERVICING

Page 47: Type 47900 & 48000 Furnaces - NIST

47

8. Slide the control section back and replacethe screws and lock washer described instep 2.

9. Reconnect to power supply.

10. To test the operation of the door switch:turn the power switch on, set the control toa setting high enough to keep the controlfrom cycling, open and close the door; thecycle light should turn off when the door isopened approximately 3" and turn back on1” to 2” before the door is closed.

To Replace the Automatic orProgrammable Controller:

The controller plugs into a panel mounting sleevewhich should be left permanently installed in thefurnace housing. To remove the controller, graspthe top and bottom of the controller front, squeezetogether and pull out. Do not attempt to dismantlethis unit further, but replace it either with aThermolyne loaner or a new unit.

NoteTo gain access to controlsection components, simplyremove the two top screwson the front dial and the twobottom screws (with one lockwasher) on the back cover.Slide control section forward.Do not pull excessively onthe internal wires.

MAINTENANCE AND SERVICING

Page 48: Type 47900 & 48000 Furnaces - NIST

48

Replacement Parts List

*Basic automatic control models include:F47910, F47910-26, F47910-33, F47914, F47915, F47918F48010, F48010-26, F48010-33, F48014, F48015, F48018

.ONTRAP NOITPIRCSED NODESUSLEDOM

94X17NC sledom.otua,eludomlortnoC

,33-02974F,62-02974F,02974F,52974F,42974F

33-02084F,02084F,82974F,42084F,62-02084F

82084F,52084F

151XNC elbammargorp,eludomlortnoC)llewd4&pmar4(sledom

,08-33-02974F,08-62-02974F,08-02974F,08-52974F,08-42974F,08-02084F,08-82974F

,08-42084F,08-33-02084F,08-62-02084F08-82084F,08-52084F

75X17NCelbammargorp-tilum,eludomlortnoC

8/pmar8(noitponoitacinummoc/w)llewd

33-05974F,62-05974F,05974F,55974F,45974F,05084F,85974F

,45084F,33-05084F,62-05084F85084F,55084F

A1X974LE ).d'qer2(edis,tnemelegnitaeH sledom00974,V042-022,021

A2X974LE ).d'qer2(edis,tnemelegnitaeH sledom00974,V802&001

A1X084LE ).d'qer2(edis,tnemelegnitaeH sledom00084,V042-022,021

A2X084LE ).d'qer2(edis,tnemelegnitaeH sledom00084,V802&001

A1X653WS hctiwsytefasrooD sledomllA

A1X753CT Kepyt,elpuocomrehT sledomllA

A1X057CP )lortnoccitamotuacisab(draobCP 51084F,51974F

A2X057CP )lortnoccitamotuacisab(draobCP )33-(&)62-(,01084F,)33-(&)62-(,01974F

A3X057CP )lortnoccitamotuacisab(draobCP 41084F,41974F

A4X057CP )lortnoccitamotuacisab(draobCP 81084F,81974F

67XLP )rebma(thgilelcyC cisabtpecxesledomV021&001llA*sledomlortnoccitamotua

28XLP )rebma(thgilelcyC cisabtpecxesledomV042-022,802llA*sledomcitamotua

65XYR yaleRlacinahceM cisabtpecxesledomV021&001llA*sledomlortnoccitamotua

75XYR yaleRlacinahceM cisabtpecxesledomV042-022,802llA*sledomlortnoccitamotua

43XYR yaleRetatSdiloS lortnoccitamotuacisabtpecxesledomllA*sledom

45XYR yaleRetatSdiloS *sledomlortnoccitamotuacisabllA

86XWS hctiwsrewopFFO/NO lortnoccitamotuacisabtpecxesledomllA*sledom

431XWS hctiwsrewopFFO/NO 51084F,41084F,51974F,41974F

441XWS hctiwsrewopFFO/NO ,81974F,)33-(&)62-(,01974F81084F,)33-(&)62-(,01084F

A1X547WS hctiws.pmeTteSothsuP *sledomlortnoccitamotuacisabllA

A1X547SR tatsoehr.tsujdAtniopteS.pmeT *sledomlortnoccitamotuacisabllA

SWX163

Page 49: Type 47900 & 48000 Furnaces - NIST

49

Wiring Diagrams

Basic Automatic Control ModelsF47910, F47910-33, F47910-33, F47914, F47915, F47918F48010, F48010-33, F48010-33, F48014, F48015, F48018

DE

SC

RIP

TIO

NR

EF.

NO

.

CN

1C

ON

TRO

L

DIA

GR

AM

CO

MP

ON

EN

T LI

ST

DS

1P

ILO

T LI

GH

T

HE

ATI

NG

ELE

ME

NT

H1

RE

SIS

TOR

, PO

TR

1R

Y1

RE

LAY

, SO

LID

STA

TES

WIT

CH

, PO

WE

RS

1S

2S

WIT

CH

, DO

OR

S3

SW

ITC

H, D

OO

R

TB1

TER

MIN

AL

BLO

CK

TC1

THE

RM

OC

OU

PLE

J6

S4

R1

CN

1

GN

D

J3J1H

IA

CJ2

J5LOW

AC

J8J9-

+

J7

SS

R

YE

L.+ J4 - RE

D

TB1

YE

LLO

W

RE

D

TC1

S1

12

- 43

+

RY

1

DS

1

S2

HE

ATI

NG

ELE

ME

NT

H2

H1

H2

T2T3

100/

120V

T2T3

H1

H2

208/

240V

T4T1 T1

T4

L2L1

GN

D

GN

D

F479

10-3

3O

NLY

FL1

FILT

ER

S3

S4

SW

ITC

H, S

ET

PO

INT

F1FU

SE

F2FU

SE

FL1

GN

D

L2L1

F1F2

MO

DE

L N

O. A

ND

OU

R P

AR

T N

O. (

s)R

EF.

NO

.D

ES

CR

IPTI

ON

F480

10F4

8010

-33

F480

14F4

8015

F480

18C

N1

CO

NTR

OL

PC

750X

2AP

C75

0X2A

PC

750X

3AP

C75

0X1A

PC

750X

4AD

S1

PIL

OT

LIG

HT

PLX

82P

LX82

PLX

76P

LX76

PLX

82F1

FUS

EFZ

X30

F2FU

SE

FZX

30FL

1FI

LTE

RC

AX

98H

1H

EA

TIN

G E

LEM

EN

TE

L480

X1A

EL4

80X

1AE

L480

X2A

EL4

80X

1AE

L480

X2A

H2

HE

ATI

NG

ELE

ME

NT

EL4

80X

1AE

L480

X1A

EL4

80X

2AE

L480

X1A

EL4

80X

2AR

1R

ES

ISTO

R, P

OT

RS

X31

4R

SX

314

RS

X31

4R

SX

314

RS

X31

4R

Y1

RE

LAY

, SO

LID

STA

TER

YX

54R

YX

54R

YX

54R

YX

54R

YX

54S

1S

WIT

CH

, PO

WE

RS

WX

144

SW

X14

4S

WX

143

SW

X14

3S

WX

144

S2

SW

ITC

H, D

OO

RS

WX

163

SW

X16

3S

WX

163

SW

X16

3S

WX

163

S3

SW

ITC

H, D

OO

RS

WX

163

SW

X16

3S

WX

163

SW

X16

3S

WX

163

S4

SW

ITC

H,S

ET

PO

INT

SW

X63

SW

X63

SW

X63

SW

X63

SW

X63

TB1

TER

MIN

AL

BLO

CK

TRX

136

TRX

136

TRX

136

TRX

136

TRX

136

TC1

THE

RM

OC

OU

PLE

TC35

7X1A

TC35

7X1A

TC35

7X1A

TC35

7X1A

TC35

7X1A

Page 50: Type 47900 & 48000 Furnaces - NIST

50

WIRING DIAGRAMS

LINE

LOAD

TC-

TC+

GN

D

GN

D

TC1

T4T3

T1T2

H1

2 4 53C

N1

RY

1

CY

CLE

RY

24 3

1 2

+

WIR

ING

DIA

GR

AM

NO

TES

:D

AS

HE

D L

INE

S "

"

DE

NO

TE 1

00/1

20 V

.W

IRE

CO

NN

EC

TIO

NS

.

1

7 9 1086

1211

CN

1D

S1

H1

H2

RY

2R

Y1

S1

S2

SO

LID

STA

TE R

ELA

Y

HE

ATI

NG

ELE

ME

NT

RE

LAY

, DP

ST,

N.O

.

SW

ITC

HD

OO

R S

WIT

CH

HE

ATI

NG

ELE

ME

NT

PIL

OT

LIG

HT,

NE

ON

CO

NTR

OL

DIA

GR

AM

CO

MP

ON

EN

T LI

ST

RE

F.N

O.

DE

SC

RIP

TIO

N

H2

FL1

FILT

ER

, EM

I

LIN

E

L1 L2

GN

D F479

20-3

3 O

NLY

DO

OR

SW

ITC

HS

3

F1FU

SE

F2FU

SE

DS

1

G

LIN

EL2L1 G

GN

D

S3

S2

L1L2

S1

FL1

F1 F2

THE

RM

OC

OU

PLE

TC1

Automatic ModelsF47920, F47920-33, F47924, F47925, F47928,F48020, F48020-33, F48024, F48025, F48028

MO

DE

L N

O. A

ND

OU

R P

AR

T N

OS

(s)

RE

F. N

O.

DE

SC

RIP

TIO

NF4

8020

F480

20-3

3F4

8024

F480

25F4

8028

CN

1C

ON

TRO

LC

NX

146

CN

X14

6C

NX

146

CN

X14

6C

NX

146

DS

1P

ILO

T LI

GH

T, N

EO

NP

LX82

PLX

82P

LX76

PLX

76P

LX82

F1FU

SE

FZX

30F2

FUS

EFZ

X30

FL1

FILT

ER

, EM

IC

AX

94H

1H

EA

TIN

G E

LEM

EN

TE

L480

X1A

EL4

80X

1AE

L480

X2A

EL4

80X

1AE

L480

X2A

H2

HE

ATI

NG

ELE

ME

NT

EL4

80X

1AE

L480

X1A

EL4

80X

2AE

L480

X1A

EL4

80X

2AR

Y1

RE

LAY

, DP

ST,

N.O

.R

YX

57R

YX

57R

YX

56R

YX

56R

YX

57R

Y2

SO

LID

STA

TE R

ELA

YR

YX

34R

YX

34R

YX

34R

YX

34R

YX

34S

1S

WIT

CH

, PO

WE

RS

WX

144

SW

X14

4S

WX

143

SW

X14

3S

WX

144

S2

SW

ITC

H, D

OO

RS

WX

163

SW

X16

3S

WX

163

SW

X16

3S

WX

163

S3

SW

ITC

H, D

OO

RS

WX

163

SW

X16

3S

WX

163

SW

X16

3S

WX

163

TC1

THE

RM

OC

OU

PLE

TC35

7X1A

TC35

7X1A

TC35

7X1A

TC35

7X1A

TC35

7X1A

Page 51: Type 47900 & 48000 Furnaces - NIST

51

WIRING DIAGRAMS

Programmable Models (4 Ramp & 4 Dwell):F479020-80, F47920-33-80, F47924-80, F47925-80, F47928-80F48020-80, F48020-33-80, F48024-80, F48025-80, F48028-80

LINE

LOAD

TC-

TC+

GN

D

GN

D

TC1

T4T3

T1T2

H1

2 4 53C

N1

RY

1

CY

CLE

RY

24 3

1 2

+

WIR

ING

DIA

GR

AM

NO

TES

:D

AS

HE

D L

INE

S "

"

DE

NO

TE 1

00/1

20 V

.W

IRE

CO

NN

EC

TIO

NS

.

1

7 9 108611

12

CN

1D

S1

H1

H2

RY

2R

Y1

S1

S2

SO

LID

STA

TE R

ELA

Y

HE

ATI

NG

ELE

ME

NT

RE

LAY

, DP

ST,

N.O

.

SW

ITC

HD

OO

R S

WIT

CH

HE

ATI

NG

ELE

ME

NT

PIL

OT

LIG

HT,

NE

ON

CO

NTR

OL

DIA

GR

AM

CO

MP

ON

EN

T LI

ST

RE

F.N

O.

DE

SC

RIP

TIO

N

H2

FL1

FILT

ER

, EM

I

LIN

E

L1 L2

GN

D F479

20-3

3-80

ON

LY

THE

RM

OC

OU

PLE

TC1

F1FU

SE

F2FU

SE

DS

1

G

LIN

EL2L1 G

GN

D

S3

S2

L1L2

S1

F1 F2FL

1

S3

DO

OR

SW

ITC

H

MO

DE

L N

O. A

ND

OU

T P

AR

T N

O. (

s)R

EF.

NO

.D

ES

CR

IPTI

ON

F480

20-8

0F4

8020

-33-

80F4

8024

-80

F480

25-8

0F4

8028

-80

CN

1C

ON

TRO

LC

NX

151

CN

X15

1C

NX

151

CN

X15

1C

NX

151

DS

1P

ILO

T LI

GH

T, N

EO

NP

LX82

PLX

82P

LX76

PLX

76P

LX82

F1FU

SE

FZX

30F2

FUS

EFZ

X30

FL1

FILT

ER

, EM

IC

AX

94H

1H

EA

TIN

G E

LEM

EN

TE

L480

X1A

EL4

80X

1AE

L480

X2A

EL4

80X

1AE

L480

X2A

H2

HE

ATI

NG

ELE

ME

NT

EL4

80X

1AE

L480

X1A

EL4

80X

2AE

L480

X1A

EL4

80X

2AR

Y1

RE

LAY

, DP

ST,

N.O

.R

YX

57R

YX

57R

YX

56R

YX

56R

YX

57R

Y2

SO

LID

STA

TE R

ELA

YR

YX

34R

YX

34R

YX

34R

YX

34R

YX

34S

1S

WIT

CH

, PO

WE

RS

WX

144

SW

X14

4S

WX

143

SW

X14

3S

WX

144

S2

SW

ITC

H, D

OO

RS

WX

163

SW

X16

3S

WX

163

SW

X16

3S

WX

163

S3

SW

ITC

H, D

OO

RS

WX

163

SW

X16

3S

WX

163

SW

X16

3S

WX

163

TC1

THE

RM

OC

OU

PLE

TC35

7X1A

TC35

7X1A

TC35

7X1A

TC35

7X1A

TC35

7X1A

Page 52: Type 47900 & 48000 Furnaces - NIST

52

WIRING DIAGRAMS

Programmable Models (8 Ramp & 8 Dwell):F47950, F47950-33, F47954, F47955, F47958F48050, F48050-33, F48054, F48055, F48058

LINE

LOAD

TC-

TC+

GN

D

GN

D

TC1

T4T3

T1T2

H1

5 63

CN

1

RY

1

CY

CLE

RY

24 3

1 2

+

WIR

ING

DIA

GR

AM

NO

TES

:D

AS

HE

D L

INE

S "

"

DE

NO

TE 1

00/1

20 V

.W

IRE

CO

NN

EC

TIO

NS

.

1

33 2631

1825

CN

1D

S1

H1

H2

RY

2R

Y1

S1

S2

SO

LID

STA

TE R

ELA

Y

HE

ATI

NG

ELE

ME

NT

RE

LAY

, DP

ST,

N.O

.

SW

ITC

HD

OO

R S

WIT

CH

HE

ATI

NG

ELE

ME

NT

PIL

OT

LIG

HT,

NE

ON

CO

NTR

OL

DIA

GR

AM

CO

MP

ON

EN

T LI

ST

RE

F.N

O.

DE

SC

RIP

TIO

N

H2

FL1

FILT

ER

, EM

I

LIN

E

L1 L2

GN

D F479

50-3

3 O

NLY

DO

OR

SW

ITC

HS

3

F1FU

SE

F2FU

SE

DS

1

G

LIN

EL2L1 G

GN

D

S3

S2

L1L2

S1

GN

D

FL1

F1 F2THE

RM

OC

OU

PLE

TC1

MO

DE

L N

O. A

ND

OU

R P

AR

T N

O. (

s)R

EF.

NO

.D

ES

CR

IPTI

ON

F480

50F4

8050

-33

F480

55F4

8058

CN

1C

ON

TRO

LC

N71

X57

CN

71X

157

CN

71X

157

CN

71X

157

DS

1P

ILO

T LI

GH

T, N

EO

NP

LX82

PLX

82P

LX76

PLX

82F1

FUS

EFZ

X30

F2FU

SE

FZX

30FL

1FI

LTE

R, E

MI

CA

X94

H1

HE

ATI

NG

ELE

ME

NT

EL4

80X

1AE

L480

X1A

EL4

80X

2AE

L480

X2A

H2

HE

ATI

NG

ELE

ME

NT

EL4

80X

1AE

L480

X1A

EL4

80X

2AE

L480

X2A

RY

1R

ELA

Y, D

PS

T, N

.O.

RY

X57

RY

X57

RY

X56

RY

X57

RY

2S

OLI

D S

TATE

RE

LAY

RY

X34

RY

X34

RY

X34

RY

X34

S1

SW

ITC

H, P

OW

ER

SW

X14

4S

WX

144

SW

X14

3S

WX

144

S2

SW

ITC

H, D

OO

RS

WX

163

SW

X16

3S

WX

163

SW

X16

3S

3S

WIT

CH

, DO

OR

SW

X16

3S

WX

163

SW

X16

3S

WX

163

TC1

THE

RM

OC

OU

PLE

TC35

7X1A

TC35

7X1A

TC35

7X1A

TC35

7X1A

Page 53: Type 47900 & 48000 Furnaces - NIST

53

Material Safety Data Sheet

Thermal CeramicsDate Revised: 7/2/91

PRODUCT IDENTIFICATIONTrade Name(s): CERAFIBERGeneric Name: REFRACTORY CERAMIC FIBER INSULATION CAS #: 65997-17-3Chemical Name: ALUMINA SILICA Formula: MIXTUREManufacturer: Thermal Ceramics Telephone: (404) 796-4200Address: P.0. BOX 923, 2102 Old Savannah RoadCity: Augusta State: Georgia Zip: 30903

PRODUCT INGREDIENTS

PHYSICAL DATAAppearance and Odor: WHITE FIBER-NO ODOR.Boiling Point: NA Evaporation Rate (NA = 1): NAVapor Pressure: NA Specific Gravity (water = 1 ): 2.6Water Solubility (%): NIL Melting Point: >3000°FVapor Density (Air= l): NA % Volatile by Volume: 0

FIRE AND EXPLOSION DATA

HEALTH HAZARDSSummary/Risks

Summary: EXPOSURE TO DUST FROM THIS PRODUCT SHOULD BE MINIMIZED. ANIMAL INHALATION AND ARTIFICIALIMPLANTATION STUDIES HAVE REPORTED THE DEVELOPMENT OF TUMORS. BASED ON PRELIMINARY RESULTS, ANOTICE OF SUBSTANTIAL RISK HAS BEEN FILED WITH THE EPA ACCORDING TO SECTION 8(e) OF THE TOXICSUBSTANCES CONTROL ACT. BASED ON ANIMAL STUDIES, IARC HAS CLASSIFIED RCF AS POSSIBLY CARCINOGENICFOR HUMANS (2B). DATA FROM HUMAN EPIDEMIOLOGICAL STUDIES IS INSUFFICIENT. THIS SUBSTANCE OR MIXTUREHAS NOT BEEN CLASSIFIED A CARCINOGEN BY NTP OR OSHA.

Medical conditions which may be aggravated: AS WITH ANY DUST, PRE-EXISTING UPPER RESPIRATORY AND LUNG DISEASESMAY BE AGGRAVATED.

Target Organ(s): LUNGS, SKIN AND EYES .Acute Health Effects: PRODUCT IS A MECHANICAL IRRITANT TO SKIN, EYES AND UPPER RESPIRATORY SYSTEM.Chronic Health Effects: EXCESSIVE EXPOSURE TO RCF DUSTS AND AFTER SERVICE FIBERS MAY CAUSE LUNG DAMAGE

(FIBROSIS). IARC STATES THERE IS SUFFICIENT EVIDENCE IN ANIMALS AND LIMITED EVIDENCE IN HUMANS TOCLASSIFY CRYSTALLINE SILICA AS A PROBABLE CARCINOGEN (2A) AND RCF AS A POSSIBLE CARCINOGEN (2B).

Primary Entry Route(s): INHALATION, SKIN AND EYE CONTACT.

Signs/Symptoms of OverexposureInhalation: IRRITATION OR SORENESS IN THROAT & NOSE. IN EXTREME EXPOSURES SOME CONGESTION MAY OCCUR.Skin Contact: TEMPORARY IRRITATION OR RASH.Skin Absorption: NAIngestion: NOT HAZARDOUS WHEN INGESTED. MAY CAUSE TEMPORARY IRRITATION TO GI TRACT.Eyes: TEMPORARY IRRITATION OR INFLAMMATION

Page 54: Type 47900 & 48000 Furnaces - NIST

54

First Aid/Emergency ProceduresInhalation: REMOVE TO FRESH AIR. DRINK WATER TO CLEAR THROAT AND BLOW NOSE TO EVACUATE FIBERS.Skin Contact: WASH AFFECTED AREAS GENTLY WITH SOAP AND WARM WATER.Skin Absorption: NAIngestion: NAEyes: FLUSH EYES WITH COPIOUS QUANTITIES OF WATER. IF IRRITATION PERSISTS CONSULT A PHYSICIAN.

REACTIVITY DATAMATERIAL IS STABLE. HAZARDOUS POLYMERIZATION CANNOT OCCUR.Chemical Incompatibilities: HYDROFLUORIC ACIDConditions to Avoid: NONE IN DESIGNED USE.Hazardous Decomposition Products: NONE

SPILL OR LEAK PROCEDURESProcedures for Spill/Leak: VACUUM CLEAN DUST WITH EQUIPMENT FITTED WITH HEPA FILTER. IF SWEEPING IS NECESSARY

USE A DUST SUPPRESSANT.Waste Management: WASTES ARE NOT HAZARDOUS AS DEFINED BY RCRA (40 CFR PART 261 ) . COMPLY WITH FEDERAL,

STATE & LOCAL REGULATIONS. METHOD OF DISPOSAL - LANDFILL. RQ - N/A.

SPECIAL PROTECTION INFORMATIONGoggles: GOGGLES OR SAFETY GLASSES WITH SIDE SHIELDS ARE RECOMMENDED.Gloves: GLOVES ARE RECOMMENDED.Respirator: <1 F/CC, USE 3M 9900; <10 F/CC, USE MSA COMFO II WITH H FILTER; <50 F/CC, USE MSA ULTR”At”WIN H FILTER;

OR EQUIVALENTS. SEE SECTION IX-OTHER.Ventilation: USE SUFFICIENT NATURAL OR MECHANICAL VENTILATION TO KEEP DUST LEVEL TO BELOW PEL/TLV/WEG

(WORKPLACE EXPOSURE GUIDELINE) USE DUST COLLECTION WHEN TEARING OUT.Other: WEAR LOOSE FITTING, LONG SLEEVED CLOTHING. WASH EXPOSED AREAS WITH SOAP & WARM WATER AFTER

HANDLING. WASH WORK CLOTHES SEPARATELY FROM OTHER CLOTHING; RINSE WASHER THOROUGHLY.Special Considerations for repair/maintenance of contaminated equipment: CRISTOBALITE RESPIRATOR: <10X PEL, USE 3M

9900; <100X PEL, USE MSA ULTR”At”WIN H FILTER; OR EQUIV. SEE SEC IX-OTHER.

SPECIAL PRECAUTIONS*** ALWAYS SEGREGATE MATERIALS BY MAJOR HAZARD CLASS ***

THIS PRODUCT CONTAINS A CHEMICAL KNOWN TO THE STATE OF CALIFORNIA TO CAUSE CANCER.Storage Segregation Hazard Classes: IRRITANTSpecial Handling/Storage: KEEP MATERIAL DRY.Special Workplace Engineering Controls: ADEQUATE VENTILATION TO KEEP DUST LEVEL TO BELOW PEL/TLV/WEG (WORKPLACE

EXPOSURE GUIDELINE).Other: ADDITIONAL INFORMATION ON THE HEALTH AND SAFETY ASPECTS OF REFRACTORY CERAMIC FIBERS IS AVAILABLE.

† Copyright© 1980, National Fire Protection Assoc., Quincy, MA 02269. This reprinted material is not the complete and official position ofthe NFPA on the referenced subject, which is represented only by the standard in its entirety. MSSM/226-21:26/00030. As of the date ofpreparation of this document, the foregoing information is believed to be accurate and is provided in good faith to comply with applicablefederal and state law(s). However, no warranty or representation with respect to such information is intended or given.

MSDS/MSD3 FORM REV. 7/2/91

MATERIAL SAFETY DATA SHEET

Page 55: Type 47900 & 48000 Furnaces - NIST

55

Ordering Procedures

Please refer to the Specification Plate for the complete model number, serial number, andseries number when requesting service, replacement parts or in any correspondenceconcerning this unit.

All parts listed herein may be ordered from the Barnstead|Thermolyne dealer from whomyou purchased this unit or can be obtained promptly from the factory. When service orreplacement parts are needed we ask that you check first with your dealer. If the dealercannot handle your request, then contact our Customer Service Department at 319-556-2241 or 800-553-0039.

Prior to returning any materials to Barnstead|Thermolyne Corp., please contact ourCustomer Service Department for a “Return Goods Authorization” number (RGA). Materialreturned without a RGA number will be refused.

Page 56: Type 47900 & 48000 Furnaces - NIST

56

Barnstead|Thermolyne Corporation warrants that if a product manufactured by Barnstead|Thermolyne andsold by it within the continental United States or Canada proves to be defective in material or construction, itwill provide you, without charge, for a period of ninety (90) days, the labor, and a period of one (1) year, theparts, necessary to remedy any such defect. Outside the continental United States and Canada, the warrantyprovides, for one (1) year, the parts necessary to remedy any such defect. The warranty period shallcommence either six (6) months following the date the product is sold by Barnstead|Thermolyne or on thedate it is purchased by the original retail consumer, whichever date occurs first.

All warranty inspections and repairs must be performed by and parts obtained from an authorizedBarnstead|Thermolyne dealer or Barnstead|Thermolyne (at its own discretion). Heating elements, however,because of their susceptibility to overheating and contamination, must be returned to our factory, and if, uponinspection, it is concluded that failure is not due to excessive high temperature or contamination, warrantyreplacement will be provided by Barnstead|Thermolyne. The name of the authorized Barnstead|Thermolynedealer nearest you may be obtained by calling 1-800-446-6060 (319-556-2241) or writing to:

Barnstead|ThermolyneP.O. Box 797

2555 Kerper BoulevardDubuque, IA 52004-0797

USAFAX: (319) 589-0516

E-MAIL ADDRESS: [email protected]