Tungsten Cable Failure Analysis
Brad James Ph.D. P.E.
![Page 2: Tungsten Cable Failure Analysis Brad James Ph.D. P.E](https://reader034.vdocuments.us/reader034/viewer/2022051412/55195678550346b5698b460c/html5/thumbnails/2.jpg)
Tungsten Cable Failure Analysis
Background• Medical device company uses tungsten cables as
part of a surgical apparatus• Non-sterilized cables withstand hundreds of
durability test cycles (consists of cyclic and static loading)
• Sterilized cables fail after a few durability test cycles
• Tensile testing shows no real difference between sterilized and non-sterilized cables
![Page 3: Tungsten Cable Failure Analysis Brad James Ph.D. P.E](https://reader034.vdocuments.us/reader034/viewer/2022051412/55195678550346b5698b460c/html5/thumbnails/3.jpg)
Tungsten Cable Failure Analysis
![Page 4: Tungsten Cable Failure Analysis Brad James Ph.D. P.E](https://reader034.vdocuments.us/reader034/viewer/2022051412/55195678550346b5698b460c/html5/thumbnails/4.jpg)
Tungsten Cable Failure Analysis
Sterilized Non-Sterilized
![Page 5: Tungsten Cable Failure Analysis Brad James Ph.D. P.E](https://reader034.vdocuments.us/reader034/viewer/2022051412/55195678550346b5698b460c/html5/thumbnails/5.jpg)
Tungsten Cable Failure Analysis
Sterilized Cable After Durability Testing
![Page 6: Tungsten Cable Failure Analysis Brad James Ph.D. P.E](https://reader034.vdocuments.us/reader034/viewer/2022051412/55195678550346b5698b460c/html5/thumbnails/6.jpg)
Tungsten Cable Failure Analysis
Sterilized Cable After Durability Testing
![Page 7: Tungsten Cable Failure Analysis Brad James Ph.D. P.E](https://reader034.vdocuments.us/reader034/viewer/2022051412/55195678550346b5698b460c/html5/thumbnails/7.jpg)
Tungsten Cable Failure Analysis
Cables After Tensile Testing:Sterilized Non-Sterilized
![Page 8: Tungsten Cable Failure Analysis Brad James Ph.D. P.E](https://reader034.vdocuments.us/reader034/viewer/2022051412/55195678550346b5698b460c/html5/thumbnails/8.jpg)
Tungsten Cable Failure Analysis
Hydrogen content tests (Leco Interstitial)
Non-sterilized cable: ~20 ppm hydrogenSterilized cable: ~ 60 ppm hydrogen
![Page 9: Tungsten Cable Failure Analysis Brad James Ph.D. P.E](https://reader034.vdocuments.us/reader034/viewer/2022051412/55195678550346b5698b460c/html5/thumbnails/9.jpg)
Tungsten Cable Failure Analysis
Conclusions• Classic case of hydrogen embrittlement! (tungsten is
known to be susceptible)• Corrosion during sterilization is the source of hydrogen• Embrittlement is stress/time dependent• Tensile testing not long enough to allow diffusion of
hydrogen to high stress locations• Static loads during durability allowed hydrogen to
diffuse to higher stress locations – resulting in fracture