    Specifications

    True North

    Each True North DGA Sample is carefully prepared from a bulk standard. A Certificate of Concentration isdelivered with each sample as a guarantee of accuracy and traceability. The Certificate of Concentrationcontains the actual concentrations of the different gases found in the sample. Once the sample is analyzed,the reported results should correspond to the values found on the Certificate of Concentration. If theresults are outside the laboratory accepted standard deviation, procedures and calibrations are not beingfollowed and should be verified.

    Certificate of Concentration

    5110 de Courtrai Avenue, Montreal, Quebec, Canada H3W 1A7Tel.: +514.739.1967 Fax: +514.739.0434Web: www.morganschaffer.com

    MORGAN

    SCHAFFER

    Transformers - The Inside View

    Syringe S/N: Lot No.: Date Filled: Prepared by: QC Approved:

    COMPONENT*

    H2 (Hydrogen)

    O2 (Oxygen)

    N2 (Nitrogen)

    CH4 (Methane)CO (Carbon Monoxide)

    CONCENTRATION (PPM)

    88 3%

    11163 3%

    40368 3%

    96 3%89 3%

    COMPONENT*

    CO2 (Carbon Dioxide)

    C2H4 (Ethylene)

    C2H6 (Ethane)

    C2H2 (Acetylene)

    CONCENTRATION (PPM)

    123 5%

    90 3%

    92 4%

    84 4%

    (*) Milliliters of gas at 273K and 760 torr per cubic meter of oil.Concentrations are determined by the method of preparation and are verified by laboratory analysis.

    THER MS SEVICES

    HIGH SENSITIVITYDGA

    MOISTURE CONTENT

    PCB ANALYSIS

    STANDARD OIL TESTS

    FURAN ANALYSIS

    INHIBITOR CONTENT

    APPLICATION

    Calibration, verification andtrouble shooting of laboratory DGA(Dissolved Gas Analysis) systems.

    PREPARATION METHOD

    NIST traceable gas mixture dissolvedin degassed oil in accordance withprinciples outlined in ASTM MethodD3612-01, Annex A2 (2001) and IECpublication 60567, clause 6.2(1992).

    DESCRIPTION

    30cc glass, gas tight syringe containing9 gases dissolved in dielectric fluid.

    Syringe has a 3 way valve with LuerLock fitting. A unique identificationnumber for each True North samplecan be found on the Certificate ofConcentration.

    DISSOLVED GAS QUANTITIES

    Stated in each sample Certificateof Concentration.Typical values: 100 ppm for faultgases (H2, CH4, CO, CO2, C2H4, C2H6,C2H2). Balance gas industrial air.

    OIL TYPE

    New Voltesso 35 dielectricinsulating fluid.

    ENVIRONMENT

    Storage: 15C to 30C; sample mustbe kept in its box.Use: 20C to 25C

    STABILITYBulk storage prior to shipping hasa shelf life of at least 12 months.Shelf life of sample: 1 month fromdate of delivery.

    PHYSICAL10 cm (4) high10 cm (4) wide30 cm (12) deep0.5 kg (1 lb)

    WARRANTY AND SUPPORT

    Morgan Schaffers True North DGAOil Standard is backed by a fullreturn policy on unused samples.Technical support is provided freeof charge to all True North users.True North and other MorganSchaffer products and services areavailable through Morgan Schaffersworldwide network of authorizeddistributors and agents.

    NOTE: Continuing research and improvementsmay result in specification or appearancechanges at any time.

    5110 de Courtrai AvenueMontreal, QuebecCanada H3W 1A7