troubleshooting problems with lateral flow assays
TRANSCRIPT
Troubleshooting Problems with Lateral Flow Assays
Michael A. Mansfield
OEM Diagnostics Group
EMD Millipore
Bedford, MA
2
M
A
T
E
R
I
A
L
S
P
R
O
C
E
S
S
E
S
Complexity of a Lateral Flow Test
BIOLOGICAL CHEMICAL
PHYSICAL ENGINEERING
TROUBLESHOOTING | 25 JUNE 2015
3
The Design Goal
Develop processes that result in a foolproof product.
TROUBLESHOOTING | 25 JUNE 2015
4
Complaints
A defect is reported.
Product is returned.
Defective product is analyzed.
The failure mode is identified.
Corrective action is implemented.
FAST!
Often not so fast!
What is a typical sequence of events?
TROUBLESHOOTING | 25 JUNE 2015
5
Why does the description of the failure not suggest a root cause?
The reported failure is an endpoint observation, often from a QC
technician or end user.
– Low sensitivity
– False negative
– False positive
– High background
– Irregular or slow flow
There are many possible causes for “typical” failures.
TROUBLESHOOTING | 25 JUNE 2015
6
What are the possible causes of “low sensitivity?”
Membrane flow too fast
Inactive capture reagent
Insufficient capture reagent
Insufficient conjugate in test
Inactive conjugate
Inactive analyte
pH or salts incorrect
System flowing too fast
Not enough conjugate reaching the
test line
Not enough sample reaching the test
line
Each of these possibilities
has material, design, and
process elements.
TROUBLESHOOTING | 25 JUNE 2015
7
Where do you start troubleshooting?
Confirm the reported failure.
– Is the failure described adequately and correctly?
– What is the frequency of the failure?
– Can the failure be reproduced?
– Does the confirmation process reveal any other problems?
Are other problems observed that are consistent with the
described failure?
TROUBLESHOOTING | 25 JUNE 2015
8
How do you go about distinguishing the possible root causes?
Basic Rule: Check the obvious and simple
things first.
TROUBLESHOOTING | 25 JUNE 2015
9
#1 - Investigate the basics.
Materials history
Process variability
Unusual observations during production
This is an obvious place to start and may provide a clue where to
look first.
TROUBLESHOOTING | 25 JUNE 2015
10
#2 - Start with the finished device.
Take the device apart before starting lengthy experimental
investigations.
Inspect the strips.
– Consistency of assembly
– Contamination
– Physical damage
A housing and tapes can hide a lot of problems.
TROUBLESHOOTING | 25 JUNE 2015
11
#3 - Substitute components that are known to work.
If a material is thought to be the problem, substitute a lot from a
previously released batch of devices.
If a chemical or biological component is suspect, substitute a lot
from a previously released batch of devices.
This assumes:
– There are retains of the key components.
– There are no stability issues.
TROUBLESHOOTING | 25 JUNE 2015
12
Absorbent pad
#4a - Employ the half-stick assay to speed the investigation.
Easier to troubleshoot problems with the
membrane as well as biological and
chemical components
Eliminates variation from conjugate
pads, sample pads, housings, and
associated assembly processes
Many variables can be checked quickly.Membrane
T
C
TROUBLESHOOTING | 25 JUNE 2015
13
#4b - Increase the complexity as needed.
Add sample padAdd conjugate pad
CONJUGATE PAD
SAMPLE PAD
TROUBLESHOOTING | 25 JUNE 2015
14
What corrective actions are needed?
Consider the nature of the manufacturing operation.
Automated Processes
- Machinery
- Engineering controls
- Process variables
- Material consistency
- Operator training
Manual Processes
- Large number of employees
- Consistency of training
- Consistency of execution
day to day
person to person
TROUBLESHOOTING | 25 JUNE 2015
15
What corrective actions are needed?
Vendors
– Improve consistency
– Additional (relevant) specifications
Manufacturing and QC
– Institute relevant incoming QC
– Refine processes
– Enhanced training
– Are the specifications too stringent?
End user
– Improve operating instructions
TROUBLESHOOTING | 25 JUNE 2015
16
Troubleshooting Examples
TROUBLESHOOTING | 14 MAY 2013
17
EXAMPLE #1 - The problem
The strips run inconsistently, and sensitivity
is variable. Strips made with membrane
from the same roll don’t work the same. In
some cases, the flow is so slow that the strip
never finishes running. All of our chemistries
are optimized.
TROUBLESHOOTING | 25 JUNE 2015
18
EXAMPLE #1 - What did the analysis show?
Absorbent Pad
Membrane
Conjugate Pad
Sample Pad
Absorbent Pad
Membrane
Conjugate Pad
Sample Pad
Absorbent Pad
Membrane
Conjugate Pad
Sample Pad
Top View of
Master Cards
TROUBLESHOOTING | 25 JUNE 2015
19
EXAMPLE #1 - What did the analysis show?
Absorbent Pad
Membrane
Conjugate Pad
Sample Pad
Angled View toward Conjugate Pad
TROUBLESHOOTING | 25 JUNE 2015
20
EXAMPLE #1 - What did the analysis show?
The conjugate pad was not consistently overlapping or contacting
the membrane.
The cards were being assembled by hand.
In final devices, variations in overlaps were obscured by the
housing.
Corrective Actions:
Increased training of assembly personnel
Higher level of visual inspection during and after assembly
process
Further investigation revealed:
TROUBLESHOOTING | 25 JUNE 2015
21
Example #2 - Misalignment of Pad Materials
Strips developed in R&D
– Excellent sensitivity & specificity
– Highly reproducible
Transferred into manufacturing
– Automation equipment designed in house
Major problems encountered
– Manufacturing using the same lots of materials
– Strip performance very inconsistent
– Product unable to meet QC specifications
TROUBLESHOOTING | 25 JUNE 2015
22
EXAMPLE #2 - Misalignment of Pad Materials
TROUBLESHOOTING | 25 JUNE 2015
23
EXAMPLE #2 - Materials Crushed During Assembly
TROUBLESHOOTING | 25 JUNE 2015
24
EXAMPLE #2 - Inconsistent Contact
TROUBLESHOOTING | 25 JUNE 2015
25
EXAMPLE #2 - What happened?
A mechanical engineer was told to design and build an assembly
machine.
He wasn’t told about the criticality of alignment and the
compressibility of the materials.
When these requirements were explained, the engineer added
guides to maintain alignment and reduced the compression to
prevent crushing.
TROUBLESHOOTING | 25 JUNE 2015
26
The Challenges of Cutting
The cutting stroke has to be strong enough to cut through the
thickest material – the backing card – cleanly and consistently.
The cutting stroke has to be gentle enough that there is no
damage to the most fragile materials – the membrane and glass
fiber pad.
The cutting stroke must not disturb the position of the materials.
TROUBLESHOOTING | 25 JUNE 2015
27
Damage from Cutting
Nicked Edges
Stray Fibers
TROUBLESHOOTING | 25 JUNE 2015
28
Example #3 – Poor Test Strip Performance
Flow of the sample was not uniform.
Hydrophobic spots appeared on the membrane when strips were
run.
Variations in sensitivity and specificity – Some membrane lots
were OK while others were not.
“We need a membrane that performs more consistently.”
TROUBLESHOOTING | 25 JUNE 2015
29
Example #3 – Poor Test Strip Performance
Strips were removed from the cassette to observe the entire flow
stream.
TEST
LINE
FLOW
AIR LOCK
TIME
TROUBLESHOOTING | 25 JUNE 2015
30
Damage from Cutting
NORMAL DAMAGED
TOP LAMINATE
MEMBRANE
POLYESTER
ADHESIVE CARD
EDGE VIEW
DAMAGED(cross section)
channel
Example #3 – Poor Test Strip Performance
TROUBLESHOOTING | 25 JUNE 2015
31
Example #3 – Corrective Actions
The cutting machine needed to be re-evaluated for:
– Frequency of cleaning
– Frequency of routine maintenance
The adhesive materials needed to be reassessed.
Then, they could rationally look at other problems.
TROUBLESHOOTING | 25 JUNE 2015
32
EXAMPLE #4 – Delamination
Delamination is separation of the top laminate from the strip.
TOP LAMINATE
MEMBRANE
POLYESTER FILM
ADHESIVE CARD
Adhesive not adhering
to membrane
Membrane not adhering
to film
???
TROUBLESHOOTING | 25 JUNE 2015
33
EXAMPLE #4 – Delamination
The manufacturer received a new lot of membrane.
Some slits from the new lot showed varying levels of delamination
of the cover tape from the membrane.
The artifact was difficult to control by working with the cutting
fixture.
TROUBLESHOOTING | 25 JUNE 2015
34
EXAMPLE #4 – Delamination
ABSORBENT PADCONJUGATE PAD MEMBRANE
INT
ER
ME
DIA
TE
DE
LA
MIN
AT
ED
Side View of Strips
TROUBLESHOOTING | 25 JUNE 2015
35
Using a properly maintained cutter, very thin
strips can be produced.
Delamination – Not Always Width Dependent
3 mm
1 mm
TROUBLESHOOTING | 25 JUNE 2015
36
Assembly into Cassettes
The cassette must be compatible with the dimensions
of the strip.
If the cassette is merely a holder for the test strip, it
must not interfere with the function and flow of the test
strip.
TROUBLESHOOTING | 25 JUNE 2015
37
EXAMPLE #5 – Assembly Problems
Indentations Cracks Leakage
Finished devices were not meeting specification for sensitivity and
specificity.
Cassettes were taken apart after running strips.
TROUBLESHOOTING | 25 JUNE 2015
38
EXAMPLE #6 – Abnormal Sample Flow
TROUBLESHOOTING | 25 JUNE 2015
39
EXAMPLE #6 – Abnormal Sample Flow
Damage to the membrane from placement into the housing was
having a large negative impact on sample flow.
Changing the physical design of the strip and/or housing was
required to produce a uniform flow front along the entire length of
the strip.
TROUBLESHOOTING | 25 JUNE 2015
40
Example #7 – False positives
2 min
10 min
T C
TROUBLESHOOTING | 25 JUNE 2015
41
Example #7 – False positivesSide View of Conjugate Pad Region
Initial
After 12 min
TROUBLESHOOTING | 25 JUNE 2015
42
Example #7 – False positives
Summary of test strip artifacts
– Colloidal gold is the wrong color.
– False positive signal develops as soon as the gold particles
pass the test line.
– Colloidal gold flows in streaks.
– Release of colloidal gold is very slow.
Approaches to fixing the artifacts
– First, redevelop the colloidal gold conjugates.
– Then, address the other issues with test strip performance.
TROUBLESHOOTING | 25 JUNE 2015
43
Conclusions
Effective manufacture of lateral flow test strips is a combination of
biological, chemical, physical, and engineering issues.
All aspects of manufacture must be considered when attempting to
minimize variation in the final product.
Problems that are seen on the membrane can be caused by
design and assembly issues affecting other parts of the test.
TROUBLESHOOTING | 25 JUNE 2015
44
To learn more about our products and services, visit
http://www.emdmillipore.com/diagnostics
TROUBLESHOOTING | 25 JUNE 2015