Features Flame ionization detector. 3U case, 19” (48.3cm) rack mountable. Automatic calibration at user-defined intervals. Internal, multipoint sampling option. FlowGuard electronic control of fuel, air and sample. Electronic back-pressure regulator with sample bypass system ensures fast re- sponse. Discrete, multilevel concentration & fault alarms. Analog output ranges programmable from 1ppm – 100% full scale. Programmable relays for concentration alarms, events and diagnostics. Automatic FID ignition, with automatic shut-off of sample, fuel and combustion air. Remote operation via RS-232 and Ether- net. Graphical display and easy to use menu system. Optional 9000 Keeper software to moni- tor and control the instrument remotely. Analyzer The Baseline® Series 9000 is the candidate of choice whenever accurate, reliable total hydrocarbon analysis is required. The Series 9000 analyzer provides unparalleled flexibility and offers continuous, fully automated gas analysis over a broad range of concentrations. The 9000 has a dynamic range from less than 10 ppb on the low range version and up to 100% as methane on the high range version. The analyzer has a generous complement of analog, digital and logic output capabilities. The analyzer is based on a flame ionization detector (FID) with FlowGuard electronic control that delivers a small portion of the sample gas to the detector flame. During the combustion process, organic or hydrocarbon-based gases in the sample are ionized then they are detected by the instrument and reported as a concentration. The Series 9000 can be configured with internal components for single or multipoint analysis of non-condensing gas samples. The automatic calibration feature enhances the long-term analytical stability of the instrument. These features place the instrument well ahead of the competition in performance, automation, and configurability. Applications The Series 9000 is designed to continuously monitor the total hydrocarbon content of non-condensing gases in a variety of applications, such as: Beverage grade CO 2 analysis Scrubber & oxidizer efficiency Carbon bed breakthrough detection Contaminant analysis in pure/ultrapure inert gases Well logging Industrial hygiene & safety monitoring Fenceline (perimeter) monitoring around industrial sites LEL monitoring www.Baseline-MOCON.com Series 9000 Total Hydrocarbon Analyzer

Upload: others

Post on 16-Sep-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Total Hydrocarbon Analyzer3.imimg.com/.../MY-1224407/ambient-hydrocarbon-analyzer.pdfThe analyzer is based on a flame ionization detector (FID) and a reduction catalyst that is switched

FeaturesFlame ionization detector.•3U case, 19” (48.3cm) rack mountable.•Automatic calibration at user-defined •intervals.Internal, multipoint sampling option.•FlowGuard• electronic control of fuel, air and sample.Electronic back-pressure regulator with •sample bypass system ensures fast re-sponse.Discrete, multilevel concentration & fault •alarms.Analog output ranges programmable •from 1ppm – 100% full scale.Programmable relays for concentration •alarms, events and diagnostics.Automatic FID ignition, with automatic •shut-off of sample, fuel and combustion air.Remote operation via RS-232 and Ether-•net.Graphical display and easy to use menu •system.Optional 9000 Keeper software to moni-•tor and control the instrument remotely.

AnalyzerThe Baseline® Series 9000 is the candidate of choice whenever accurate, reliable total hydrocarbon analysis is required. The Series 9000 analyzer provides unparalleled flexibility and offers continuous, fully automated gas analysis over a broad range of concentrations.

The 9000 has a dynamic range from less than 10 ppb on the low range version and up to 100% as methane on the high range version. The analyzer has a generous complement of analog, digital and logic output capabilities.

The analyzer is based on a flame ionization detector (FID) with FlowGuard electronic control that delivers a small portion of the sample gas to the detector flame. During the combustion process, organic or hydrocarbon-based gases in the sample are ionized then they are detected by the instrument and reported as a concentration.

The Series 9000 can be configured with internal components for single or multipoint analysis of non-condensing gas samples. The automatic calibration feature enhances the long-term analytical stability of the instrument. These features place the instrument well ahead of the competition in performance, automation, and configurability.

ApplicationsThe Series 9000 is designed to continuously monitor the total hydrocarbon content of non-condensing gases in a variety of applications, such as:

Beverage grade CO• 2 analysisScrubber & oxidizer efficiency•Carbon bed breakthrough detection•Contaminant analysis in pure/ultrapure inert gases•Well logging•Industrial hygiene & safety monitoring•Fenceline (perimeter) monitoring around industrial sites•LEL monitoring• www.Baseline-MOCON.com

Series 9000Total Hydrocarbon Analyzer

Page 2: Total Hydrocarbon Analyzer3.imimg.com/.../MY-1224407/ambient-hydrocarbon-analyzer.pdfThe analyzer is based on a flame ionization detector (FID) and a reduction catalyst that is switched

Series 9000Total Hydrocarbon Analyzer

AnalyticalDetector: Flame ionization detector (FID).

Ranges: User definable based upon calibration within: MDQ 0.01ppm, Full-scale 1 - 200ppm (Methane) MDQ 0.1ppm, Full-scale 1 - 2,000ppm (Methane)MDQ 0.3ppm, Full-scale 1 -20,000ppm (Methane)MDQ 0.003%, Full-scale 0.01 - 100% (Methane)

Repeatability: +/- 1% Full-scale response.

Drift: +/- 1% of full-scale over 24 hours.

Response Time: < 5 seconds to 90% of final reading.

Alarms: Multilevel concentration and fault alarms that result in audible and visual alarms. Alarms may also be mapped to relays to control external equipment.

Sampling: Internal single or multipoint modules, for pre-filtered (<0.1 microns) non-condensing samples, with or without sample pump.

Calibration: Programmable automatic or manual calibration.Support Gas: Hydrogen 35 cc/min, Air 175 cc/min (typical). Hydrocarbon content must be less than 1 ppm. Fuel blend options available, consult Baseline®.

PhysicalDimensions: 19.00” (48.3cm) W x 14.25” (36.2cm) D x 5.25” (13.3cm) H.Weight: < 20 lb (9.07kg).

Configuration: Bench-top or rack-mount 19” (48.3cm) panel.Connections: 1/4” (6.35mm) tube fitting connectors.

Operating Temperature: 32 - 104 °F (0 - 40 °C).

Operating Humidity: 0 – 95% (non-condensing).

ElectricalPower: 90 – 230 VAC, 50/60Hz, 3A.

Display: 3.4” (8.6cm) x 4.5” (11.4cm) graphical display.

Relay Outputs: 5 programmable form A relays rated to 3A @ 230V AC (optional additional 9 relay output board).

Analog Outputs: 1 programmable 0-20mA or 4-20mA isolated output (optional additional 3 analog output board).

Digital Outputs: RS-232, Ethernet.

P.O. Box 649, 19661 Highway 36 • Lyons, CO 80540P:1.800.321.4665•1.303.823.6661F: 303.823.5151 E: [email protected]

D011.4

Page 3: Total Hydrocarbon Analyzer3.imimg.com/.../MY-1224407/ambient-hydrocarbon-analyzer.pdfThe analyzer is based on a flame ionization detector (FID) and a reduction catalyst that is switched

FeaturesFlame ionization detector.•3U case, 19” (48.3cm) rack mountable.•Hydrocarbon detection from sub-ppm to •1000ppm (Methane).Automatic calibration at user-defined •intervals.Internal, multipoint sampling option.•FlowGuard• electronic control of fuel, air and sample.Electronic back-pressure regulator with •sample bypass system ensures fast re-sponse.Discrete, multilevel concentration & fault •alarms.Programmable analog output ranges.•Programmable relays for concentration •alarms, events and diagnostics.Automatic FID ignition, with automatic •shut-off of sample, fuel and combustion air.Remote operation via RS-232 and Ether-•net.Graphical display and easy to use menu •system.Optional 9000 Keeper software to moni-•tor and control the instrument remotely.

AnalyzerThe Baseline® Series 9000 NMHC is a specialized member of the extraordinary Series 9000 family of gas analyzers. The Series 9000 is the candidate of choice whenever accurate, reliable total, Methane, and NonMethane analysis is required. The Series 9000 analyzer provides unparalleled flexibility and offers continuous, fully automated gas analysis over a broad range of concentrations.

The Series 9000 NMHC is designed to measure the total, methane, and non-methane hydrocarbon content of gas samples. The analyzer has a generous complement of analog, digital and logic output capabilities.

The analyzer is based on a flame ionization detector (FID) and an oxidation catalyst that is switched in and out of a portion of the sample stream. The catalyst oxidizes all hydrocarbons except methane, for a methane measurement. The methane value is then automatically subtracted from the total hydrocarbon concentration to determine the non-methane hydrocarbon reading.

The Series 9000 NMHC can be configured with internal components for single or multipoint analysis of non-condensing gas samples. The automatic calibration feature enhances the long-term analytical stability of the instrument. These features place the instrument well ahead of the competition in performance, automation, and configurability.

ApplicationsThe Series 9000 NMHC is designed to continuously monitor the total hydrocarbon content of non-condensing gases in a variety of applications, such as:

Ambient Air Monitoring Networks•Clean Rooms•Air & Oxygen Purity in Air Separation Plants•

www.Baseline-MOCON.com

Series 9000 NMHCNonMethane Hydrocarbon Analyzer

Page 4: Total Hydrocarbon Analyzer3.imimg.com/.../MY-1224407/ambient-hydrocarbon-analyzer.pdfThe analyzer is based on a flame ionization detector (FID) and a reduction catalyst that is switched

Series 9000 NMHC

AnalyticalDetector: Flame ionization detector (FID).

Oxidizer: Oxidizes NonMethane gases.

Range: User definable based upon calibration within:MDQ 0.06ppm, Full-scale 1 - 200ppm (Air Balance)MDQ 0.1ppm, Full-scale 1 - 1000ppm (Air Balance)MDQ 0.03ppm, Full-scale 1 - 200ppm (O2 Balance)

Repeatability: +/- 1% Full-scale response.

Drift: +/- 1% of Full-Scale over 24 hours.

Response Time: T90 < 30 sec.

Alarms: Multilevel concentration and fault alarms that result in audible and visual alarms. Alarms may also be mapped to relays to control external equipment.

Sampling: Internal single or multipoint modules for pre-filtered (<0.1 microns), non-condensing samples with or without sample pump.

Calibration: Programmable automatic or manual calibration.Support Gas: Hydrogen 35 cc/min (H2He Blend 100cc/min), Air 175 cc/min (typical). Hydrocarbon content must be less than 1 ppm. Fuel blend options available, consult Baseline®.

PhysicalDimensions: 19.00” (48.3cm) W x 14.25” (36.2cm) D x 5.25” (13.3cm) H.Weight: < 20 lb (9.07kg).

Configuration: Bench-top or rack-mount 19” (48.3cm)panel.Connections: 1/4” (6.35mm) tube fitting connectors.

Operating Temperature: 32 - 104 °F (0 - 40 °C).

Operating Humidity: 0 – 95% (non-condensing).

ElectricalPower: 90 – 230 VAC, 50/60Hz, 3A.

Display: 3.4” (8.6cm) x 4.5” (11.4cm) graphical display.

Relay Outputs: 5 programmable form A relays rated to 3A @ 230V AC (optional additional 9 relay output board).

Analog Outputs: 1 programmable 0-20mA or 4-20mA isolated output (optional additional 3 analog output board).

Digital Outputs: RS-232, Ethernet.

P.O. Box 649, 19661 Highway 36 • Lyons, CO 80540P:1.800.321.4665•1.303.823.6661F: 303.823.5151 E: [email protected]

NonMethane Hydrocarbon Analyzer

D018.5

Page 5: Total Hydrocarbon Analyzer3.imimg.com/.../MY-1224407/ambient-hydrocarbon-analyzer.pdfThe analyzer is based on a flame ionization detector (FID) and a reduction catalyst that is switched

FeaturesFlame ionization detector.•3U case, 19” (48.3cm) rack mountable.•Hydrocarbon detection from sub-ppm to •100 ppm levels.Automatic calibration at user-defined •intervals.Internal, multipoint sampling option.•FlowGuard• electronic control of fuel, air and sample.Electronic back-pressure regulator with •sample bypass system ensures fast re-sponse.Discrete, multilevel concentration & fault •alarms.Analog output ranges programmable •from 1ppm – 100% full scale.Programmable relays for concentration •alarms, events and diagnostics.Automatic FID ignition, with automatic •shut-off of sample, fuel and combustion air.Remote operation via RS-232 and Ether-•net.Graphical display and easy to use menu •system.Optional 9000 Keeper software to moni-•tor and control the instrument remotely.

AnalyzerThe Baseline® Series 9000 TCA is a specialized member of the extraordinary Series 9000 family of gas analyzers. The Series 9000 is the candidate of choice whenever accurate, reliable total hydrocarbon and total carbon analysis is required. The Series 9000 analyzer provides unparalleled flexibility and offers continuous, fully automated gas analysis over a broad range of concentrations.

The Series 9000 TCA has a dynamic range from less than 0.1 ppm to 100 ppm, it’s designed to measure total hydrocarbon content and provide a combined carbon monoxide (CO) and carbon dioxide (CO2) reading. The analyzer has a generous complement of analog, digital and logic output capabilities.

The analyzer is based on a flame ionization detector (FID) and a reduction catalyst that is switched in and out of a portion of the sample stream. Total hydrocarbons are measured first, and then the catalyst switches in. The catalyst converts the CO & CO2 to methane and leaves the hydrocarbons unchanged for a total carbon measurement. The total hydrocarbon value is then automatically subtracted from the total carbon concentration to determine the combined CO/CO2 reading

The Series 9000 TCA can be configured with internal components for single or multipoint analysis of non-condensing gas samples. The automatic calibration feature enhances the long-term analytical stability of the instrument.

ApplicationsThe Series 9000 TCA is designed to continuously monitor total hydrocarbons and trace levels of CO/CO2 in order to provide a total carbon reading.

This specialized analyzer is configured for the detection of hydrocarbon and CO/CO2 impurities in pure and ultrapure inert gases such as argon, helium, hydrogen, and nitrogen. Samples must be free of oxygen.

www.Baseline-MOCON.com

Series 9000 TCATotal Carbon Analyzer

Page 6: Total Hydrocarbon Analyzer3.imimg.com/.../MY-1224407/ambient-hydrocarbon-analyzer.pdfThe analyzer is based on a flame ionization detector (FID) and a reduction catalyst that is switched

Series 9000 TCA

AnalyticalDetector: Flame ionization detector (FID).

Methanizer: Converts low levels of CO and CO2 to methane.Range: User definable based upon calibration within: 0 - 100 ppm (Methane).MDQ: 0.1 ppm (100 ppb).

Repeatability: +/- 1% Full-scale response.

Drift: Zero: +/- 1% Full-scale over 24 hours

Response Time: T90 < 30 sec.

Alarms: Multilevel concentration and fault alarms that result in audible and visual alarms. Alarms may also be mapped to relays to control external equipment.

Sampling: Internal single or multipoint modules for pre-filtered (<0.1 microns), non-condensing samples, with or without sample pump.

Calibration: Programmable automatic or manual calibration.Support Gas: Hydrogen 40 cc/min, Air 200 cc/min. (typical). Hydrocarbon content must be less than 1 ppm. Fuel blend options available, consult Baseline®

PhysicalDimensions: 19.00” (48.3cm) W x 14.25” (36.2cm) D x 5.25” (13.3cm) H.Weight: < 20 lb (9.07kg).

Configuration: Bench-top or rack-mount 19” (48.3cm) panel.Connections: 1/4” (6.35mm) tube fitting connectors.

Operating Temperature: 32 - 104 °F (0 - 40 °C).

Operating Humidity: 0 – 95% (non-condensing).

ElectricalPower: 90 – 230 VAC, 50/60Hz, 3A.

Display: 3.4” (8.6cm) x 4.5” (11.4cm) graphical display.

Relay Outputs: 5 programmable form A relays rated to 3A @ 230V AC (optional additional 9 relay output board).

Analog Outputs: 1 programmable 0-20mA or 4-20mA isolated output (optional additional 3 analog output board).

Digital Outputs: RS-232, Ethernet.

P.O. Box 649, 19661 Highway 36 • Lyons, CO 80540P:1.800.321.4665•1.303.823.6661F: 303.823.5151 E: [email protected]

Total Carbon Analyzer

D019.4

Page 7: Total Hydrocarbon Analyzer3.imimg.com/.../MY-1224407/ambient-hydrocarbon-analyzer.pdfThe analyzer is based on a flame ionization detector (FID) and a reduction catalyst that is switched

FeaturesFlame ionization detector.•3U case, 19” (48.3cm) rack mountable.•Hydrocarbon detection from sub-ppm to •50% levels (Methane).Automatic calibration at user-defined •intervals.FlowGuard• electronic control of fuel, air and sample.Electronic back-pressure regulator with •sample bypass system ensures fast re-sponse.Discrete, multilevel concentration & fault •alarms.Programmable relays for concentration •alarms, events and diagnostics.Automatic FID ignition, with automatic •shut-off of sample, fuel and combustion air.Remote operation via RS-232 and Ether-•net.Graphical display and easy to use menu •system.Optional 9000 Keeper software to moni-•tor and control the instrument remotely.

AnalyzerThe Baseline® Series 9000 H is a high-temperature member of the extraordinary Series 9000 family of gas analyzers. The Series 9000 is the candidate of choice whenever accurate, reliable hydrocarbon analysis is required.

The Series 9000 H has a dynamic range from 0.1 ppm to 10% (Propane) or 0.3 ppm to 50% (Methane). The analyzer has a generous complement of analog, digital and logic output capabilities. These features place the instrument well ahead of the competition in performance, automation, and configurability.

The analyzer is based on a flame ionization detector (FID) with FlowGuard electronic control that delivers a small portion of the sample gas to the detector flame. During the combustion process, organic or hydrocarbon-based gases in the sample are ionized then they can be detected by the instrument and reported as a concentration.

The Series 9000 H is configured for single point analysis of samples heated up to 376 °F (191°C). The automatic calibration feature enhances the long-term analytical stability of the instrument.

ApplicationsThe Series 9000 H is designed to continuously monitor the total hydrocarbon content of gases and maintain the temperature of heated sample above their dew point. This extremely versitile instrument can be configured to support a variety of applications, such as:

Compliance monitoring for EPA Methods 25A & 503•Continuous Emission Monitoring (CEM) of source •hydrocarbonsScrubber & oxidizer efficiency•Carbon bed breakthrough detection•LEL monitoring•Vehicle emissions•Chemical process blending• www.Baseline-MOCON.com

Series 9000 HHeated Hydrocarbon Analyzer

Page 8: Total Hydrocarbon Analyzer3.imimg.com/.../MY-1224407/ambient-hydrocarbon-analyzer.pdfThe analyzer is based on a flame ionization detector (FID) and a reduction catalyst that is switched

Series 9000 H

AnalyticalDetector: Flame ionization detector (FID).

Sample Temperature: Up to 376 °F (191 °C).

Ranges: User definable based upon calibration within:MDQ 0.3ppm, Full-scale 1 to 2,000ppm (Methane)MDQ 0.6ppm, Full-scale 1 – 20,000ppm (Methane)MDQ 0.003%, Full-scale 0.01 – 50% (Methane)

Repeatability: +/- 1% Full-scale response.

Drift: +/- 1% of Full-Scale over 24 hours.

Response Time: T90 < 5 sec.

Alarms: Multilevel concentration and fault alarms that result in audible and visual alarms. Alarms may also be mapped to relays to control external equipment.

Sampling: Internal single module for pre-filtered (<0.1 microns), non-condensing samples, with or without sample pump.

Calibration: Programmable automatic or manual calibration.Support Gas: Hydrogen 40 cc/min, Air 200 cc/min. (typical). Hydrocarbon content must be less than 1 ppm. Fuel blend options available, consult Baseline®

PhysicalDimensions: 19.00” (48.3cm) W x 14.25” (36.2cm) D x 5.25” (13.3cm) H.Weight: < 25 lb (11.3kg).

Configuration: Bench-top or rack-mount 19” (48.3cm) panel.Connections: 1/4” (6.35mm) tube fitting connectors.

Operating Temperature: 32 - 104 °F (0 - 40 °C).

Operating Humidity: 0 – 95% (non-condensing).

ElectricalPower: 90 – 230 VAC, 50/60Hz, 3A.

Display: 3.4” (8.6cm) x 4.5” (11.4cm) graphical display.

Relay Outputs: 5 programmable form A relays rated to 3A @ 230V AC.Analog Outputs: 1 programmable 0-20mA or 4-20mA isolated output.Digital Outputs: RS-232, Ethernet.

P.O. Box 649, 19661 Highway 36 • Lyons, CO 80540P:1.800.321.4665•1.303.823.6661F: 303.823.5151 E: [email protected]

Heated Hydrocarbon Analyzer

D025.3

Page 9: Total Hydrocarbon Analyzer3.imimg.com/.../MY-1224407/ambient-hydrocarbon-analyzer.pdfThe analyzer is based on a flame ionization detector (FID) and a reduction catalyst that is switched

Series 9000 Analyzer Versions

www.Baseline-MOCON.com P.O. Box 649, 19661 Highway 36 • Lyons, CO 80540

P: 1.800.321.4665 • 1.303.823.6661 F: 303.823.5151 E: [email protected] D031.1

Model 9000 THA 9000 NMHC 9000 H 9000 TCA

Description Fully Automated Hydrocarbon Analyzer

Methane / Non-Methane Hydrocarbons

Fully Automated Heated Hydrocarbon Analyzer

Total Carbon Analyzer (THC+CO+CO2 in inert

gases, except air & O2)Detector FID FID with Oxidizer Catalyst FID FID with Methanizer Catalyst

Sample Points 1 and 4 point 1 and 4 point 1 point 1 and 4 pointSampling Type 4 point is Sequential 4 point is Sequential Continuous 4 point is Sequential

Outputs

Standard: 5 relays, 1 analog, RS-232, Ethernet

Optional: Additional 9 relays & 3 analogs

Standard: 5 relays, 1 analog, RS-232, Ethernet

Optional: Additional 9 relays & 3 analogs

Standard: 5 relays, 1 analog, RS-232, Ethernet

Standard: 5 relays, 1 analog, RS-232, Ethernet

Optional: Additional 9 relays & 3 analogs

Internal Wetted Parts Stainless Steel Stainless Steel Stainless Steel Stainless SteelSample Temperature Ambient Ambient Up to 191°C (376°F) Ambient

Repeatability Drift +/- 1% full scale+/- 1% over 24 hours

+/- 1% full scale+/- 1% over 24 hours

+/- 1% full scale+/- 1% over 24 hours

+/- 1% full scale+/- 1% over 24 hours

Gas Flow Control Electronic Electronic Electronic Electronic

Ranges & MDQ(as Methane)

User selectable within:0.01 to 100ppm0.1 to 2000ppm0.3 to 20000ppm0.003% to 100%

User selectable within:0.06 to 200ppm (Air balance)0.1 to 1000ppm (Air balance)0.03 to 200ppm (O2 balance)

User selectable within:0.3 to 2000ppm0.6 to 20000ppm0.003% to 50%

User selectable within:0.1 to 100ppm

Calibration Automatic or Manual Automatic or Manual Automatic or Manual Automatic or Manual

Calibration Plumbing Optional: Internal Zero/Span Valves

Optional: Internal Zero/Span Valves

Optional: Internal Zero/Span Valves

Optional: Internal Zero/Span Valves