tis 93025 syscal reference standard device itica c 11353 reed hartman hwy, suite 110 cincinnati, oh...

1
Cooper-Atkins Corporation | Accuracy to the Highest Degree 11353 Reed Hartman Hwy, Suite 110 | Cincinnati, OH 45241-2929 U.S.A. 888-533-6900 | 513-793-5366 | Fax: 513-793-4895 | www.cooper-atkins.com ENTERPRISE WIRELESS MONITORING 67-1489 V1017 93025 SysCal Reference Standard Device OVERVIEW SysCal was developed to provide a highly accurate, referenceable, comparative device to assist customers in spot checking the accuracy of the TempTrak transmitters and solid simulator probes in situ. Some customers are interested in utilizing SysCal to perform their own calibrations* of Cooper-Atkins TempTrak transmitters and solid simulator probes in situ. * It is the sole responsibility of the party performing calibrations with SysCal to collect, document and maintain all necessary information to achieve calibration standards and regulatory requirements. Due to legal restrictions and to ensure the accuracy, completeness and reliability of the data collected during customer performed calibrations, Cooper-Atkins is unable to assist in the process of collection, maintenance, validation, and documentation of all calibration requirements. This includes any certificate generation and data integration into TempTrak databases. Cooper-Atkins is not liable, does not guarantee and is not responsible for the accuracy of any calibration performed by customers. SysCal™ Reference Standard Device #93025 • Modified Solid Simulator • On-site testing / spot checking • Simple and easy to use FEATURES BENEFITS Provides a way to insert the probe to allow temperature stabilization • Allows for probe validation • Training times reduced SPECIFICATIONS System Components: (1) 35025 thermocouple instrument and (2) 50219-K probes System Range: -25°F to 180°F (-32°C to 82°C) System Accuracy: ±0.5°F (±0.3°C) when probe/instrument are calibrated together and ambient temperature is between 68°F and 86°F (20°C and 30°C) Ambient Temperature Range: Operating: 32°F to 122°F (0°C to 50°C) Ambient Accuracy: Operating: At ambient temperatures between 32°F to 68°F (0°C to 20°C) and 86°F to 122°F (30°C to 50°C), add ±0.1° error per °F or °C deviation from 68°F and 86°F (20°C and 30°C) Resolution: 0.1°/1° user selectable Display: 0.5” (13 mm) LCD Battery Type: (2) 1.5V (AAA) batteries (ANSI/NEDA-24F, IEC-R03) Battery Life: 1800 hours typical Compliance/Regulatory: Calibrated in an accredited lab, CE certified, NSF listed, WEEE marked, RoHS compliant, NIST traceable The reading from the TempTrak transmitter is compared to the reference standard recorded by the SysCal thermocouple. If that reading is within ±2.0°F (±1.1°C) the transmitter and solid simulator probe are considered to be calibrated within the stated accuracy. (1) 35025 Thermocouple instrument (2) 50219-K probes (1) Certificate of calibration (1) Calibration log sheet (1) Hard carrying case (48) Calibration labels (48) Blank labels

Upload: others

Post on 23-Mar-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: TIS 93025 SysCal Reference Standard Device ITICA C 11353 Reed Hartman Hwy, Suite 110 Cincinnati, OH 45241-2929 .S.A. 888-533-6900 513-93-5366 Fa: 513-93-4895  TIS ISS ITI 6 …

Cooper-Atkins Corporation | Accuracy to the Highest Degree11353 Reed Hartman Hwy, Suite 110 | Cincinnati, OH 45241-2929 U.S.A. 888-533-6900 | 513-793-5366 | Fax: 513-793-4895 | www.cooper-atkins.com

ENTERPRISEWIRELESS

MONITORING

67-1489V1017

93025 SysCal™ Reference Standard Device

OVERVIEWSysCal was developed to provide a highly accurate, referenceable, comparative device to assist customers in spot checking the accuracy of the TempTrak transmitters and solid simulator probes in situ. Some customers are interested in utilizing SysCal to perform their own calibrations* of Cooper-Atkins TempTrak transmitters and solid simulator probes in situ.

* It is the sole responsibility of the party performing calibrations with SysCal to collect, document and maintain all necessary information to achieve calibration standards and regulatory requirements. Due to legal restrictions and to ensure the accuracy, completeness and reliability of the data collected during customer performed calibrations, Cooper-Atkins is unable to assist in the process of collection, maintenance, validation, and documentation of all calibration requirements. This includes any certificate generation and data integration into TempTrak databases. Cooper-Atkins is not liable, does not guarantee and is not responsible for the accuracy of any calibration performed by customers.

SysCal™ Reference Standard Device#93025

• Modified Solid Simulator • On-site testing / spot checking• Simple and easy to use

FEATURES BENEFITS• Provides a way to insert the probe to allow temperature stabilization • Allows for probe validation• Training times reduced

SPECIFICATIONS• System Components: (1) 35025 thermocouple instrument and

(2) 50219-K probes• System Range: -25°F to 180°F (-32°C to 82°C)• System Accuracy: ±0.5°F (±0.3°C) when probe/instrument are

calibrated together and ambient temperature is between 68°F and 86°F (20°C and 30°C)

• Ambient Temperature Range: Operating: 32°F to 122°F (0°C to 50°C)• Ambient Accuracy: Operating: At ambient temperatures between 32°F

to 68°F (0°C to 20°C) and 86°F to 122°F (30°C to 50°C), add ±0.1° error per °F or °C deviation from 68°F and 86°F (20°C and 30°C)

• Resolution: 0.1°/1° user selectable• Display: 0.5” (13 mm) LCD• Battery Type: (2) 1.5V (AAA) batteries (ANSI/NEDA-24F, IEC-R03)• Battery Life: 1800 hours typical• Compliance/Regulatory: Calibrated in an accredited lab, CE certified,

NSF listed, WEEE marked, RoHS compliant, NIST traceable

The reading from the TempTrak transmitter is compared to the reference standard recorded by the SysCal thermocouple. If that reading is within ±2.0°F (±1.1°C) the transmitter and solid simulator probe are considered to be calibrated within the stated accuracy.

(1) 35025 Thermocouple instrument (2) 50219-K probes (1) Certificate of calibration (1) Calibration log sheet (1) Hard carrying case (48) Calibration labels(48) Blank labels