the risks of promiscuous mva of tof-sims data: the mva …

40
The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA Quickie Daniel J. Graham NESAC/BIO University of Washington Department of Bioengineering SIMS Europe Sept 19-21, 2010 Münster Germany NESAC/BIO

Upload: others

Post on 31-Oct-2021

8 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA Quickie

Daniel J. GrahamNESAC/BIO

University of WashingtonDepartment of Bioengineering

SIMS Europe Sept 19-21, 2010Münster Germany

NESAC/BIO

Page 2: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

ToF-SIMS Is Complicated• Spectra contain hundreds of peaks• Peak intensities can be interrelated• Matrix effects can cause non-linear changes in peak

intensities– Due to sample composition– Due to the presence of oxides– Due to the presence of salts

• Peak intensities may or may not correlate with surface composition

• Peak intensities may vary due to differential sputtering

• Often heavy fragmentation and lack of molecular ion signals (assuming you know what signal to expect)

?0

2 0 0 0 0

4 0 0 0 0

6 0 0 0 0

8 0 0 0 0

1 0 0 0 0 0

1 2 0 0 0 0

1 4 0 0 0 0

1 6 0 0 0 0

1 8 0 0 0 0

0 2 0 4 0 6 0 8 0 1 0 0 1 2 0 1 4 0 1 6 0 1 8 0 2 0 0

m / z

Inte

nsity

(Cou

nts)

Page 3: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

ToF-SIMS Is Complicated• No one surface analysis method can provide a complete

surface characterization alone• ToF-SIMS in particular benefits from data from other

methods• The more complicated the surface chemistry, the more

important this becomes

XPS

IR

AFMSFG

NEXAFS

SIMS

Now it's startingto make sense

Page 4: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

Proper ToF-SIMS Analysis Requires:• Good experimental plans (controls)• Proper sample preparation• Careful data collection• Consistent data calibration• Sound understanding of the fundamentals of

mass spectral analysis• Knowledge of how to properly use the available

tools to help with the analysis

Page 5: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

MVA “To The Rescue...”• MVA is becoming increasingly popular for ToF-

SIMS spectra and images• MVA can aid in reducing the complexity of a

data set with regards to the magnitude of the data one needs to focus on (what is changing, where is it different, what peaks are changing)

• MVA cannot:– Remove all effects of contaminants or matrix effects– Fix a poorly designed experiment– Interpret your data– Find things that are not there

Page 6: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

Avoid the MVA Quickie• MVA should not be treated like a black box• MVA should not be something you do as a last

minute decision, it should be part of the experimental design from the beginning

• If someone says, “Lets see what we get when we use SIMS and MVA on my samples”...RUN...Or at least stop to think if it really makes sense

MVADATA Magical ResultsX

Page 7: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

Plan• What is the

question you want to answer?

• What samples do you need to answer that question?

• How many samples/ replicates do you need?

Remember MVA will find the main differences between any samples

If you input garbage in

You will get garbage out!!!

MVA

Page 8: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

When Should MVA be Used?• MVA should be used to help answer questions

– Are surfaces A and B different?– How does treatment X change the surface chemistry?– How is fragmentation pattern affected by ____?– Can TOF-SIMS data distinguish Protein A from

Protein B?– What regions in my image are different and why?

• The question should be part of the experimental design and not an afterthought

Page 9: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

9

Structures of the polyurethane units

Control sample

Poly(caprolactone) diol (PCL)

PUU 817, LDI/PCL/Hyd =8/1/7

L-lysine diisocyanate (LDI)

PPUU 8161, LDI/PCL/Hyd/(Gly/Hyp/Pro/Hyd) =8/1/6/1

Structure of the Poly(peptide-urethaneurea)

The Hard segment (HS)The Soft segment (SS)

The Oligo-Polypeptide Segment (OPS)

PCL n

n

*Slide and data from Gilad Zorn Ph.D., NESAC/BIO University of Washington

Page 10: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

10

PC1 NH4 18.036

CH2N 28.019

CH4N 30.035

N2H4 32.038

N2H5 33.048

C5H10N 84.086

C3H3 39.023

C3H5 41.037

C3H3O 55.017

C5H9 69.073

C6H9O 97.068

C6H11O2 115.089

PCL (SS) **LDI-Hyd (HS) *

Characteristic peaksPrincipal Component Analysis (PCA) of ToF-SIMS data

* Comparable to ToF-SIMS studies of poly(L-lysine) * * Similar to published ToF-SIMS data of PCL.

PCL

PPUU 8161 333

27.022732.0378

33.04855.0168

86.0628

70.0705

68.0516

-0.7

-0.6

-0.5

-0.4

-0.3

-0.2

-0.1

0

0.1

0.2

0.3

0 50 100 150 200

m /z

340A

PUU 817

C4H6N 68.052 Proline

C4H8N 70.071 Proline

C4H8NO 86.063 Hydroxy-proline

PC2

OPS peaks:

PPUU 8161

PCL

*Slide and data from Gilad Zorn Ph.D., NESAC/BIO University of Washington

Page 11: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

Steps to MVA

• Plan Experiment and controls• Collect data

– (What samples, how many replicates?)• Calibrate spectra

– Calibration should be consistent• Select peaks (Which?)• Normalize the data (How?)• Pre-process the data (How?)• Interpret the results (What are you looking at?)

Page 12: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

Experimental Design/Data Collection• Not all systems are well defined, but

your experimental design can be:– Think about what you want to learn from

SIMS– Simplify the number of variables you are

dealing with per experiment– Plan appropriate controls– Run enough replicates to determine

reproducibility• Homogeneous => 3 to 5 spots on 2

samples• Inhomogeneous => 5 to 7 spots on 3 to 5

samples

Page 13: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

Proteins adsorbed onto Mica: PCA

-0.08

-0.06

-0.04

-0.02

0

0.02

0.04

0.06

-0.12 -0.1 -0.08 -0.06 -0.04 -0.02 0 0.02 0.04 0.06 0.08 0.1Score on PC 1 (51%)

Scor

e on

PC

2 (2

1%)

Cytochrome c

Myoglobin

Hemoglobin

HumanFibrinogen

BovineFibrinogen

Collagen

Fibronectin

Immunoglobulin G

γ-Globulins

α2-Macroglobulin

Lysozyme

Papain

Albumin

TransferrinLactoferrin

Kininogen

443 spectra,16 different proteins Modified fromWagner & Castner, Langmuir 17 (2001) 4649.

Page 14: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

Proteins adsorbed onto Mica: PCA

-0.08

-0.06

-0.04

-0.02

0

0.02

0.04

0.06

-0.12 -0.1 -0.08 -0.06 -0.04 -0.02 0 0.02 0.04 0.06 0.08 0.1Score on PC 1 (51%)

Scor

e on

PC

2 (2

1%)

Cytochrome c

Myoglobin

Hemoglobin

HumanFibrinogen

BovineFibrinogen

Collagen

Fibronectin

Immunoglobulin G

γ-Globulins

α2-Macroglobulin

Lysozyme

Papain

Albumin

TransferrinLactoferrin

Kininogen

443 spectra,16 different proteins Wagner & Castner, Langmuir 17 (2001) 4649.

Page 15: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

Synthesis of an IPN of P(AAm-co-EG/AA)

OO

On[ ]

HN

HNO

O

OH2N

HN

HNO

O

CH3COCH3

CH3OH

hv

hv

OO

OO

OOSi

Si

HO

OCQ

CQ

PEG(NH2)2

H2Nn

[ ] NH2O

EDCSulfo-NHSMES/NaCl BufferPH 7.0

Substrate

IPN

PEG spacer

Page 16: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

PCA of ATC-AAm

samplesATCATCATCATCATCATC

ATCATC

ATC

AAmAAmAAmAAmAAmAAmAAmAAmAAm

-0.15

-0.1

-0.05

0

0.05

0.1

0 5 10 15 20

Sample

PC1

scor

es (9

3.2%

)

-1-0.8-0.6-0.4-0.20

0.20.4

0 50 100 150 200 250 300 350 400 450 500m/z

PC1

load

ings

(93.

2 %

)

Al

Si,Si(isotope)

SiOHC3H8N

C3H6NOCH2NO

C3H3ONH4

OOOOO

O Si Si

PCA ofATC and AAm

samples

-PCA clearly separates the two sample types

-low scatter in the data

Page 17: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

AAmAAmAAmAAmAAmAAmAAmAAmAAm

AAcAAcAAc

AAcAAcAAcAAcAAcAAc

-0.04

-0.02

0

0.02

0.04

0.06

0.08

0 5 10 15 20Sample

PC1

scor

es (9

0.8%

)

-0.5-0.3-0.10.10.30.50.70.9

0 50 100 150 200 250 300 350 400 450 500m/z

PC1

load

ings

(90.

8%)

C3H5 C4H7

C4H9

C5H9

CH3O

C2H5O

C3H7O

C3H5O2

CH2NO

C3H6NO

PCA ofAAm and AAc

samples

-PCA separates the two sample types

-scatter seen in the AAc datasuggesting less uniform chemistry spot to spot.

Page 18: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

AAcAAcAAc

AAcAAcAAcAAcAAcAAc

NH2NH2NH2

NH2NH2NH2NH2NH2NH2

-0.08-0.06-0.04-0.02

00.020.040.060.080.1

0 5 10 15 20Sample

PC1

scor

es (9

1.1%

)

-0.6-0.4-0.20

0.20.40.60.81

0 50 100 150 200 250 300 350 400 450 500m/z

PC1

load

ings

(91.

1%)

C3H5C4H7

C5H7,C5H9

Na

CH3OC2H5O

C3H5O2

C4H7O2

PCA ofAAc and NH2

samples

-PCA separates the two sample types (some overlap with 1 NH2 sample)

-1 sample from each set is seen to have noticeably different scores (reason unknown)

Page 19: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

Conclusions• PCA has great potential to aid in spectral

interpretation and analysis– can aid in determining sample differences– requires well thought out experiments– cannot do analysis for you

• Plan your experiments with a central question and minimize the number of variables– This can greatly simplify the interpretation– Can maximize what you get out of your data

Page 20: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

• mvsa.nb.uw.edu– Tutorials– References– Links– Software– Practice Data Sets

Website is online nowIf you have information,tutorials, links, softwareyou would like to contributeplease contact me at:[email protected]

Page 21: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA Quickie

Daniel J. GrahamNESAC/BIO

University of WashingtonDepartment of Bioengineering

SIMS Europe Sept 19-21, 2010Münster Germany

NESAC/BIO

The goal of this presentation is to drive home the fact that MVA is a tool that can aid in the analysis of ToF-SIMS data, but it cannot do the analysis for you. Proper use of MVA requires good planning, sound experimental design and careful review and interpretation of the results.

Page 22: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

ToF-SIMS Is Complicated• Spectra contain hundreds of peaks• Peak intensities can be interrelated• Matrix effects can cause non-linear changes in peak

intensities– Due to sample composition– Due to the presence of oxides– Due to the presence of salts

• Peak intensities may or may not correlate with surface composition

• Peak intensities may vary due to differential sputtering

• Often heavy fragmentation and lack of molecular ion signals (assuming you know what signal to expect)

?0

2 0 0 0 0

4 0 0 0 0

6 0 0 0 0

8 0 0 0 0

1 0 0 0 0 0

1 2 0 0 0 0

1 4 0 0 0 0

1 6 0 0 0 0

1 8 0 0 0 0

0 2 0 4 0 6 0 8 0 1 0 0 1 2 0 1 4 0 1 6 0 1 8 0 2 0 0

m / z

Inte

nsity

(Cou

nts)

There is no way around it. ToF-SIMS data is complicated. There can be multiple factors that can change the relative intensities of peaks within a spectrum or image. These changes can be due to instrumentation, sample preparation, sample composition and more.

Page 23: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

ToF-SIMS Is Complicated• No one surface analysis method can provide a complete

surface characterization alone• ToF-SIMS in particular benefits from data from other

methods• The more complicated the surface chemistry, the more

important this becomes

XPS

IR

AFMSFG

NEXAFS

SIMS

Now it's startingto make sense

As with many surface analytical methods, ToF-SIMS should not be used alone. Complimentary information from other methods can help elucidate and clarify the interpretation of the ToF-SIMS data.

Page 24: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

Proper ToF-SIMS Analysis Requires:• Good experimental plans (controls)• Proper sample preparation• Careful data collection• Consistent data calibration• Sound understanding of the fundamentals of

mass spectral analysis• Knowledge of how to properly use the available

tools to help with the analysis

One should always plan carefully when doing any experiment. This is particularly important when using ToF-SIMS. Controlling extraneous variables and sources of potential variation within a data can be critical to the success of an experiment.

Page 25: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

MVA “To The Rescue...”• MVA is becoming increasingly popular for ToF-

SIMS spectra and images• MVA can aid in reducing the complexity of a

data set with regards to the magnitude of the data one needs to focus on (what is changing, where is it different, what peaks are changing)

• MVA cannot:– Remove all effects of contaminants or matrix effects– Fix a poorly designed experiment– Interpret your data– Find things that are not there

Due to the complexities of ToF-SIMS data, researchers a turning to MVA to aid in sorting through their data. MVA can aid in reducing the complexity of a data set and help highlight what is changing, however MVA cannot interpret your data for you.

Page 26: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

Avoid the MVA Quickie• MVA should not be treated like a black box• MVA should not be something you do as a last

minute decision, it should be part of the experimental design from the beginning

• If someone says, “Lets see what we get when we use SIMS and MVA on my samples”...RUN...Or at least stop to think if it really makes sense

MVADATA Magical ResultsXMVA should not be treated as a black box. It cannot help make sense of a poorly designed experiment.

Page 27: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

Plan• What is the

question you want to answer?

• What samples do you need to answer that question?

• How many samples/ replicates do you need?

Remember MVA will find the main differences between any samples

If you input garbage in

You will get garbage out!!!

MVA

Garbage in = Garbage out

Plan before you start any experiment.

Page 28: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

When Should MVA be Used?• MVA should be used to help answer questions

– Are surfaces A and B different?– How does treatment X change the surface chemistry?– How is fragmentation pattern affected by ____?– Can TOF-SIMS data distinguish Protein A from

Protein B?– What regions in my image are different and why?

• The question should be part of the experimental design and not an afterthought

MVA should be part of the experimental design, not an afterthought of “wow this is complex, maybe we should use MVA.”

Design your experiments around a specific question/hypothesis to be answered. Think about how MVA can help and which method would be best suited to the experimental design.

Page 29: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

9

Structures of the polyurethane units

Control sample

Poly(caprolactone) diol (PCL)

PUU 817, LDI/PCL/Hyd =8/1/7

L-lysine diisocyanate (LDI)

PPUU 8161, LDI/PCL/Hyd/(Gly/Hyp/Pro/Hyd) =8/1/6/1

Structure of the Poly(peptide-urethaneurea)

The Hard segment (HS)The Soft segment (SS)

The Oligo-Polypeptide Segment (OPS)

PCL n

n

*Slide and data from Gilad Zorn Ph.D., NESAC/BIO University of Washington

In this example from Zorn .et .al. the authors compared a set of polyurethane samples. Both polymers had a PCL softsegment. One of the polymers included a polypeptide segment within the hard segment.

The authors wanted to verify the presence of the peptide within the polymer and determine which peaks were characteristic of each material. For this they used PCA.

Page 30: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

10

PC1 NH4 18.036

CH2N 28.019

CH4N 30.035

N2H4 32.038

N2H5 33.048

C5H10N 84.086

C3H3 39.023

C3H5 41.037

C3H3O 55.017

C5H9 69.073

C6H9O 97.068

C6H11O2 115.089

PCL (SS) **LDI-Hyd (HS) *

Characteristic peaksPrincipal Component Analysis (PCA) of ToF-SIMS data

* Comparable to ToF-SIMS studies of poly(L-lysine) * * Similar to published ToF-SIMS data of PCL.

PCL

PPUU 8161 333

27.022732.0378

33.04855.0168

86.0628

70.0705

68.0516

-0.7

-0.6

-0.5

-0.4

-0.3

-0.2

-0.1

0

0.1

0.2

0.3

0 50 100 150 200

m/z

340A

PUU 817

C4H6N 68.052 Proline

C4H8N 70.071 Proline

C4H8NO 86.063 Hydroxy-proline

PC2

OPS peaks:

PPUU 8161

PCL

*Slide and data from Gilad Zorn Ph.D., NESAC/BIO University of Washington

It was found that the PCA PC1 scores clearly separated the PCL soft segment from the two different hard segment materials. PC2 separated the two different polymers. It was seen that peaks characteristic of the peptide segment were found to have high loadings corresponding with the peptide containing polymer (negative scores and loadings on PC2).

Page 31: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

Steps to MVA

• Plan Experiment and controls• Collect data

– (What samples, how many replicates?)• Calibrate spectra

– Calibration should be consistent• Select peaks (Which?)• Normalize the data (How?)• Pre-process the data (How?)• Interpret the results (What are you looking at?)

This graphic illustrates the “Steps” required to properly carrying out MVA of ToF-SIMS data. This presentation will not cover all of them. These steps are covered in more detail in the “PCA Step by Step” tutorial on the NESAC/BIO MVSA website (http://mvsa.nb.uw.edu).

Page 32: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

Experimental Design/Data Collection• Not all systems are well defined, but

your experimental design can be:– Think about what you want to learn from

SIMS– Simplify the number of variables you are

dealing with per experiment– Plan appropriate controls– Run enough replicates to determine

reproducibility• Homogeneous => 3 to 5 spots on 2

samples• Inhomogeneous => 5 to 7 spots on 3 to 5

samples

It is important to think about what one wants to learn from ToF-SIMS data and also to plan experiments where a minimal number of variables change within a sample set. By reducing the number of variables that are changing, one can more easily relate changes seen in the spectra to the variable of interest. Ideally only one variable should change within a given set of samples.

It is also important to include the proper number of replicates in order to get statistical significance to the MVA results. This slide provides a general guideline, but ultimately the number of data points to collect will depend on the sample type. In general, more homogenous samples require fewer data points, and less homogeneous samples require more data points.

Page 33: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

Proteins adsorbed onto Mica: PCA

-0.08

-0.06

-0.04

-0.02

0

0.02

0.04

0.06

-0.12 -0.1 -0.08 -0.06 -0.04 -0.02 0 0.02 0.04 0.06 0.08 0.1Score on PC 1 (51%)

Scor

e on

PC

2 (2

1%)

Cytochrome c

Myoglobin

Hemoglobin

HumanFibrinogen

BovineFibrinogen

Collagen

Fibronectin

Immunoglobulin G

γ-Globulins

α2-Macroglobulin

Lysozyme

Papain

Albumin

TransferrinLactoferrin

Kininogen

443 spectra,16 different proteins Modified fromWagner & Castner, Langmuir 17 (2001) 4649.

This slide was adapted from the work of Matt Wagner and Dave Castner. In this slide I have deleted most of the data points from the protein data set generated by Matt. With this set of data points one could look at the data and conclude that all of the proteins are clearly separated and that the scatter in the data is minimal.

However if you add back in all the data points, the story changes....

Page 34: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

Proteins adsorbed onto Mica: PCA

-0.08

-0.06

-0.04

-0.02

0

0.02

0.04

0.06

-0.12 -0.1 -0.08 -0.06 -0.04 -0.02 0 0.02 0.04 0.06 0.08 0.1Score on PC 1 (51%)

Scor

e on

PC

2 (2

1%)

Cytochrome c

Myoglobin

Hemoglobin

HumanFibrinogen

BovineFibrinogen

Collagen

Fibronectin

Immunoglobulin G

γ-Globulins

α2-Macroglobulin

Lysozyme

Papain

Albumin

TransferrinLactoferrin

Kininogen

443 spectra,16 different proteins Wagner & Castner, Langmuir 17 (2001) 4649.

This slide shows the Wagner data with all of the data points. Most all of the proteins are still clearly separated from each other, however there is significant scatter in the data of some protein. Collecting a proper number of data points is critical in understanding the true variance within a data set.

Page 35: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

Synthesis of an IPN of P(AAm-co-EG/AA)

OO

On[ ]

HN

HNO

O

OH2N

HN

HNO

O

CH3COCH3

CH3OH

hv

hv

OO

OO

OOSi

Si

HO

OCQ

CQ

PEG(NH2)2

H2Nn

[ ] NH2O

EDCSulfo-NHSMES/NaCl BufferPH 7.0

Substrate

IPN

PEG spacer

This example is taken from data collected from samples prepared from Kevin Healy's group at Berkely. The sample consists of an interpenetrating polymer network (IPN) created by sequentially attaching various compounds to the substrate.

Due to the similarity in the chemistry of the various compounds when looking at all of the data together, much of the data overlaps in the PCA scores. However, when the samples are compared step by step throughout the IPN creation process one can separate out the various chemistries and find peaks that are characteristic of each compound.

Page 36: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

PCA of ATC-AAm

samplesATCATCATCATCATCATC

ATCATC

ATC

AAmAAmAAmAAmAAmAAmAAmAAmAAm

-0.15

-0.1

-0.05

0

0.05

0.1

0 5 10 15 20

SamplePC

1 sc

ores

(93.

2%)

-1-0.8-0.6-0.4-0.20

0.20.4

0 50 100 150 200 250 300 350 400 450 500m/z

PC1

load

ings

(93.

2 %

)

Al

Si,Si(isotope)

SiOHC3H8N

C3H6NOCH2NO

C3H3ONH4

OOOOO

O Si Si

PCA ofATC and AAm

samples

-PCA clearly separates the two sample types

-low scatter in the data

Here the base silanized substrate is compared to the surface after addition of the AAm polymer. PCA separates the two sample surfaces and highlights the characteristic peaks for each of the compounds on the surface.

Page 37: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

AAmAAmAAmAAmAAmAAmAAmAAmAAm

AAcAAcAAc

AAcAAcAAcAAcAAcAAc

-0.04

-0.02

0

0.02

0.04

0.06

0.08

0 5 10 15 20Sample

PC1

scor

es (9

0.8%

)

-0.5-0.3-0.10.10.30.50.70.9

0 50 100 150 200 250 300 350 400 450 500m/z

PC1

load

ings

(90.

8%)

C3H5 C4H7

C4H9

C5H9

CH3O

C2H5O

C3H7O

C3H5O2

CH2NO

C3H6NO

PCA ofAAm and AAc

samples

-PCA separates the two sample types

-scatter seen in the AAc datasuggesting less uniform chemistry spot to spot.

Here the AAm polymer is compared to the AAc polymer. The two samples types are separated on the scores plot, though it is noted that there is more scatter within the AAc samples.

Page 38: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

AAcAAcAAc

AAcAAcAAcAAcAAcAAc

NH2NH2NH2

NH2NH2NH2NH2NH2NH2

-0.08-0.06-0.04-0.02

00.020.040.060.080.1

0 5 10 15 20Sample

PC1

scor

es (9

1.1%

)

-0.6-0.4-0.20

0.20.40.60.81

0 50 100 150 200 250 300 350 400 450 500m/z

PC1

load

ings

(91.

1%)

C3H5C4H7

C5H7,C5H9

Na

CH3OC2H5O

C3H5O2

C4H7O2

PCA ofAAc and NH2

samples

-PCA separates the two sample types (some overlap with 1 NH2 sample)

-1 sample from each set is seen to have noticeably different scores (reason unknown)

Comparing the AAc samples to the NH2 samples shows some overlap on PC1 and scatter in both samples types. However, overall the two samples types are separated on PC1. Since both polymers contain similar chemical structures, the loadings appear to be showing that the NH2 samples have a higher relative intensity of the oxygen containing fragments than the AAc samples. This is verified in the raw data (not shown).

Page 39: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

Conclusions• PCA has great potential to aid in spectral

interpretation and analysis– can aid in determining sample differences– requires well thought out experiments– cannot do analysis for you

• Plan your experiments with a central question and minimize the number of variables– This can greatly simplify the interpretation– Can maximize what you get out of your data

MVA encompasses a powerful set of methods that can be valuable for the ToF-SIMS analyst. However, MVA is a tool and not a replacement of common sense and good experimental planning.

The ToF-SIMS analyst still needs to know “standard” data analysis methodologies and also needs to understand why they are using a given MVA method and the assumptions that go along with each method.

Page 40: The Risks of Promiscuous MVA of ToF-SIMS Data: The MVA …

• mvsa.nb.uw.edu– Tutorials– References– Links– Software– Practice Data Sets

Website is online nowIf you have information,tutorials, links, softwareyou would like to contributeplease contact me at:[email protected]

The NESAC/BIO MVSA website is a community resource providing information about the application of MVA to surface analytical data.

The website provides:Tutorials, references, links and software to aid in the application of MVA to ToF-SIMS and other surface analytical data.

If you have materials that could be useful for the community, please consider contributing them to the website.