the effect of parallel combined steady and oscillatory shear … · 2007. 4. 3. · g. vlastos et...

13
Rheol Acta 36:160-172 (1997) © Steinkopff Verlag 1997 Georgios Vlastos Dietmar Lerche Brigitte Koch Odette Samba Manfred Pohl The effect of parallel combined steady and oscillatory shear flows on blood and polymer solutions Received: 25 July 1996 Accepted: 22 November 1996 Paper in part presented at the Symposium on Rheology and Computational Fluid Me- chanics dedicated to the memory of Prof. A.C. Papanastasiou, University of Cyprus, Nicosia, July 4-5, 1996 G. Vlastos • O. Samba. M. Pohl Institute for Medical Physics and Biophysics University Hospital Charit6 Humboldt University Ziegelstrage 5/9 10117 Berlin, Germany E-mail: vlastos @rz.charite.hu-berlin.de Prof. Dr. D. Lerche (~) . B. Koch L. U. M. GmbH Rudower Chaussee 5 (IGZ) 12489 Berlin, Germany E-mail: lerche @lum.fta-berlin.de Abstract Human blood at physio- logical volume concentration exhi- bits non-Newtonian and thixotropic properties. The blood flow in the microcirculation is pulsatile, in- itiated from the heart pulse and can be considered as superposition of two partial flows: a) a steady shear, and b) an oscillatory shear. Until now steady and viscoelastic behav- ior were separately investigated. Here we present the response to the combination of steady and oscilla- tory shear for human blood, a high molecular weight aqueous polymer solution (polyacrylamide AP 273E) and an aqueous xanthan gum solu- tion. The polyacrylamide and xanthan solutions are fluids that model the theological properties of human blood. In general, parameters describing blood viscoelasticity be- came less pronounced as superim- posed steady shear increased, espe- cially at low shear region and by elasticity, associated with reduction in RBC aggregation. The response of polymer solutions to superposi- tion shows qualitative similarities with blood by elasticity, but their quantitative response differed from that of blood. By viscosity another behavior was observed. The super- position effect on viscous compo- nent was described by a modified Carreau equation and for the elastic component by an exponential equa- tion. Key words Viscoelasticity - blood - aggregation - model fluid - polyacrylamide - xanthan Introduction The blood flow near the heart (aorta and large arteries) is pulsatile due to the periodic action of the heart. Even in the microcirculation it was possible to detect signifi- cant pulsatile flow at arterioles, capillaries and venules (Gaethgens, 1970; Intaglietta et al., 1971). It is well known, that human blood at physiological volume con- centration displays non-Newtonian and thixotropic be- havior (Chien, 1967a, b; Cokelet, 1987). The blood vis- coelasticity has also been explored (Lessner et al., 1971; Thurston, 1972; Chien et al., 1975). Thus an ex- amination of the pulsatile nature of blood flow must combine an oscillatory flow at constant frequency simi- lar to that of heart pulse and a steady forward flow, in respect to blood viscoelasticity. The steady blood flow which has been often regarded as main blood flow at vessels, and oscillatory flow were up to now separately investigated. We studied the combination of steady and oscillatory shear for human blood at standard hemato- crit and for two model fluids (an aqueous polyacryla- mide AP 273E and an aqueous xanthan gum solution). These model fluids are blood-like at certain concentra- tions and are used as fluids modeling blood behavior for heart valve testing (Lerche et al., 1993 a; Vlastos et al., 1994). Some measurements and visualization techniques for hydrodynamic performance tests are not appropriate using blood, that is not always obtainable in required quantifies. Therefore, fluids modeling blood rheological behavior have to be used for tests of artificial organs or simulating flow disturbances in respect to atherosclero-

Upload: others

Post on 10-Feb-2021

0 views

Category:

Documents


0 download

TRANSCRIPT

  • Rheol Acta 36:160-172 (1997) © Steinkopff Verlag 1997

    Georgios Vlastos Dietmar Lerche Brigitte Koch Odette Samba Manfred Pohl

    The effect of parallel combined steady and oscillatory shear flows on blood and polymer solutions

    Received: 25 July 1996 Accepted: 22 November 1996

    Paper in part presented at the Symposium on Rheology and Computational Fluid Me- chanics dedicated to the memory of Prof. A.C. Papanastasiou, University of Cyprus, Nicosia, July 4-5, 1996

    G. Vlastos • O. Samba. M. Pohl Institute for Medical Physics and Biophysics University Hospital Charit6 Humboldt University Ziegelstrage 5/9 10117 Berlin, Germany E-mail: vlastos @ rz.charite.hu-berlin.de

    Prof. Dr. D. Lerche ( ~ ) . B. Koch L. U. M. GmbH Rudower Chaussee 5 (IGZ) 12489 Berlin, Germany E-mail: lerche @ lum.fta-berlin.de

    Abstract Human blood at physio- logical volume concentration exhi- bits non-Newtonian and thixotropic properties. The blood flow in the microcirculation is pulsatile, in- itiated from the heart pulse and can be considered as superposition of two partial flows: a) a steady shear, and b) an oscillatory shear. Until now steady and viscoelastic behav- ior were separately investigated. Here we present the response to the combination of steady and oscilla- tory shear for human blood, a high molecular weight aqueous polymer solution (polyacrylamide AP 273E) and an aqueous xanthan gum solu- tion. The polyacrylamide and xanthan solutions are fluids that model the theological properties of human blood. In general, parameters

    describing blood viscoelasticity be- came less pronounced as superim- posed steady shear increased, espe- cially at low shear region and by elasticity, associated with reduction in RBC aggregation. The response of polymer solutions to superposi- tion shows qualitative similarities with blood by elasticity, but their quantitative response differed from that of blood. By viscosity another behavior was observed. The super- position effect on viscous compo- nent was described by a modified Carreau equation and for the elastic component by an exponential equa- tion.

    Key words Viscoelasticity - blood - aggregation - model fluid - polyacrylamide - xanthan

    Introduction

    The blood flow near the heart (aorta and large arteries) is pulsatile due to the periodic action of the heart. Even in the microcirculation it was possible to detect signifi- cant pulsatile flow at arterioles, capillaries and venules (Gaethgens, 1970; Intaglietta et al., 1971). It is well known, that human blood at physiological volume con- centration displays non-Newtonian and thixotropic be- havior (Chien, 1967a, b; Cokelet, 1987). The blood vis- coelasticity has also been explored (Lessner et al., 1971; Thurston, 1972; Chien et al., 1975). Thus an ex- amination of the pulsatile nature of blood flow must combine an oscillatory flow at constant frequency simi- lar to that of heart pulse and a steady forward flow, in respect to blood viscoelasticity. The steady blood flow

    which has been often regarded as main blood flow at vessels, and oscillatory flow were up to now separately investigated. We studied the combination of steady and oscillatory shear for human blood at standard hemato- crit and for two model fluids (an aqueous polyacryla- mide AP 273E and an aqueous xanthan gum solution). These model fluids are blood-like at certain concentra- tions and are used as fluids modeling blood behavior for heart valve testing (Lerche et al., 1993 a; Vlastos et al., 1994).

    Some measurements and visualization techniques for hydrodynamic performance tests are not appropriate using blood, that is not always obtainable in required quantifies. Therefore, fluids modeling blood rheological behavior have to be used for tests of artificial organs or simulating flow disturbances in respect to atherosclero-

  • G. Vlastos et al. 161 The effect of parallel combined steady and oscillatory shear flows

    sis (Ku and Liepsch, 1986; Brookshier and Tarbell, 1993). In investigations at cardiovascular assist devices testing artificial heart valves under pulsatile flow condi- tions, Pohl et al. (1996) measured distinct hydrody- namic parameters by using Newtonian and non-Newto- nian model fluids. Therefore, it is beneficial to study the response of model fluids, under conditions existing in cardiovascular apparatus, i.e. unsteady flow condi- tions. The separate response of model fluids to steady shear flow and/or dynamic conditions can answer the question of whether a fluid simulates blood satisfactori- ly. Additionally to this, the response to combined steady and oscillatory flow accounts for changes in the structure of model fluids under conditions where blood also is exposed. Therefore, the investigation of parallel combined flows is an additional important procedure, for an improved rheological comparison between blood and model fluids.

    Several experimental studies on superposition were undertaken for different polymer solutions like polystyr- enes (Osaki et al., 1965), polyacrylamides (Jones and Waiters, 1971; Tanner and Williams, 1971) polyisobuty- lenes (Malkin et al., 1975; De Cleyn and Mewis, 1981) and propylenes (Booij, 1966a). Theoretical work on superposition of steady on small amplitude oscillatory shear has been performed, based on Oldroyd's theory (Booij, 1966b; Barnes et al., 1971), and on Lodge's network theory (Tanner and Simmons, 1967). Most of the above-mentioned work concerned a parallel super- position of steady shear on small amplitude oscillatory shear, measured by the Weissenberg Rheogoniometer. Measurements of orthogonal superposition for colloidal dispersions (carbon black in mineral oil) have been car- ried out from Mewis and Schoukens (1978). They con- cluded that for certain dispersions, superposition mea- surements provide more detailed information about the colloidal structure than viscosity or elasticity measure- ments. Studies of large amplitude superposition for polymers are rare in the literature. Isayev and Wong (1988) investigated the parallel large amplitude superpo- sition of various polymer systems experimentally, and compared their data with theoretical results based on the Leonov model.

    Investigations on superposition of human blood have not been extensively performed. Measurements have been reported only in small rigid tubes at 2 Hz (Thur- ston, 1975), in Couette geometry at 0.5 Hz (Lerche et al., 1993a), and in the erythro-aggregometer device at 0.5 Hz (Riha and Stolz, 1996). Hence, we explored the effect of parallel combined steady and oscillatory shear with increasing shear strain amplitude at constant fre- quency (f=0.5Hz), for the above-mentioned fluids. Furthermore, blood was subjected to a combined flow at low shear amplitudes by varying the oscillation fre- quency. The appropriate concentrations of model solu- tions matching blood behavior were obtained using the

    same measuring protocols as for blood. The response of blood-like model fluids to pulsatile character of blood flow, i.e., to superposition, should be regarded as an ad- ditional criterion for the simulation of rheological prop- erties of blood.

    Experimental

    Materials

    Human blood from healthy donors was drawn by veni- puncture and anticoagulated with heparin. The hemato- crit was adjusted to 0.45 by autologous plasma. The samples were stored at room temperature and measured within 5 h. The polyacrylamide (Nordfloc AP 273E, Nordmann, Rasmann GmbH & Co., approximate mo- lecular weight 15x106), and the polysaccharide xanthan gum (Xanthomonas campestris, Fluka, approximate mo- lecular weight 2×106 ) were used to prepare model fluids. Both fluids had a wide molecular weight distri- bution. Stock solutions of 1000 ppm were prepared dis- persing the powders in distilled water, under continuous gentle stirring for 4 h. Solutions of different concentra- tions were subsequently prepared from the stock solu- tions by dilution with distilled water. The solutions pre- pared were stored at T=4°C and measured the next day. Fresh solutions were always used. The time span between preparation, storing and measurement was con- stant. The steady shear viscosity of the samples was regularly tested and compared with values from past measurements of the same fluids. No shear degradation was detected, in agreement with other studies concern- ing the above-mentioned solutions (Argumedo et al., 1978; Rochefort and Middleman, 1987).

    Experimental conditions

    Steady shear viscosity and viscoelasticity were deter- mined using the computer controlled Contraves LS 40 Low Shear Couette viscometer. All measurements were carried out with the DIN 412 Couette measuring system (Rc=6.5 ram, Rb=6 mm and L=18 ram, where Rc and Rb represent the radii of measuring cup and measuring bob and L the length of the bob), at a temperature T=(25_+0.1)°C. The test substance is located in the gap between the rotating measuring cup and the stationary measuring bob. The friction between the substance and the measuring bob exerts a torque on measuring bob. The bob is coupled to a deflection system in the mea- suring head (type A). The deflection system with a compensator and a mirror is suspended from a load thread. A stator, on the other hand, is permanently fixed to the measuring head. The deflection transmitted to the

  • 162 Rheologica Acta, Vol. 36, No. 2 (1997) © Steinkopff Verlag 1997

    Table 1 Experimental conditions and statistical analysis of deviations in percent from oil standard viscosities (column 4) after n= 11 calibra- tion measurements, by equilibrium viscosity and flow curve calibra- tion experiment. T=25 °C

    Calibration Standard Measuring Deviation in % Shear rate oil viscosity method (mean_+s.d.) range (s -l)

    (mPa s)

    2A 1.425 Equilibrium 2.2+0.7 30-100 Flow curve 2.5_+1.2 1-100

    20B 24.02 Equilibrium -1.4_+0.6 5-15 Flow curve 1 .7_+1 .2 0.06-31.62

    100B 102.0 Equilibrium -0.1_+1.1 0.5-1.5 Flow curve 0 . 7 _ + 1 . 4 0.012-10.0

    system by the bob is scanned optically and compen- sated. The measuring head measures the torque exerted.

    The steady shear viscosity was obtained starting from a shear rate of 100 s 1 and decreasing in 25 steps to 0.01 s -~. The viscoelasticity was determined with measur- ing protocols incrementing the shear strain amplitude 70 from 0.05 to 13.11 at a fixed frequency f=0.5 Hz (pre- shearing time 20 s at 100 s -1, 10 s at 10 s -~ and 20 s at rest, automatic sensitivity range, measuring time 3.5 min). The viscoelasticity protocols for blood in Fig. 1 consisted of two measuring phases (sensitivity range 4), the first phase with increasing shear strain am- plitude and the second phase beginning after 5 s, with an addition of a superimposed steady shear rate ~sup = 2 s l on oscillatory shear. The frequency dependence was mea- sured by decreasing the oscillation frequency f from 5 Hz to 0.03 Hz at a constant small shear strain amplitude ~o=0.2 (the _~reshearing time was 20 s at 100 s -1 and 20 s at 30 s - , automatic sensitivity range, measuring time 5.2 min). A steady shear rate ~)sup (1 s -t -5 s -1) was then superimposed in same direction on oscillatory shear. Virgin samples were used in each measurement. Data analysis was performed by means of the software package SWR 40 (2.30).

    The viscometer was calibrated using three calibration oils (Physikalisch-Technische Bundesanstalt, Braun- schweig) with known standard viscosities, in order to cover most of viscosity range of blood. The viscosity of three oils was determined using two experimental pro- cedures, a) the equilibrium viscosity experiment where, by increasing the shear rate, the corresponding shear stress was obtained. The time required for the oils achieving a state of equilibrium was set at 2%/min (shear stress change in percent per minute) as equili- brium criterion, and b) the flow curve experiment where shear stress was measured by ramp decreasing of shear rate.

    Experimental conditions and deviations in percent from oil standard viscosities after n = l l calibration measurements (over a period of a year) are shown in

    Table 1, for both calibration experiments. The calibra- tion for the dynamic measurements was tested by mea- suring the viscoelasticity of the above mentioned three calibration oils. The oils were Newtonian, so they ex- hibited no elasticity. The deviations of the viscous com- ponent t/' (mean values over the whole shear amplitude range) in percent from the standard viscosities of the oils (cf. Table 1, column 1), were 1.2% for the 2A oil, -0.3% for the 20B oil, and 4.5% for the 100B oil.

    Results

    Under normal physiological conditions, blood is a con- centrated suspension of deformable, near colloidal-size cells (erythrocytes, leukocytes and platelets) in a com- plex aqueous solution (plasma). The rheology of blood is determined from the red blood cells (RBC's) which are dominant on cell population because of their number and volume concentration (Cokelet, 1987). The presence of larger molecules in plasma (e.g., fibrinogen) at low shear rates (i _0,45), network formation leads to measurable yield shear stress (Kiesewetter et al., 1982).

    Human blood exhibits shear thinning behavior at a constant RBC volume concentration of 0.45, as shown in Fig. 1. Chien (1970) explained the shear dependence of blood viscosity as result of shear dependence of the effective cell volume. He correlated the viscometric data with the microrheological behavior of erythrocytes and rigid particles. The effective cell volume represents the actual volume of suspended particles and an addi- tional volume of external fluid (plasma), hydrodynami- cally immobilized. At higher shear rates (?)>50 s -1) the aggregates are being dispersed, and now the individual erythrocytes are oriented and deformed in shear field. Thurston (1989) explained the behavior of both steady shear viscosity and viscoelasticity on shear rate, by pro- gressive disaggregation of aggregates with increasing shear rate, followed by cell alignment and formation of compacted cell layers. The process is accompanied by release of trapped plasma on which cell layers slide. Wells and Schmid-Schtnbein (1969) proposed a mecha- nism allowing the RBC's to participate in flow like a fluid drop. They observed red cell deformation under flow and a tank-treadlike motion of erythrocyte mem- brane around the cell contents. The shear stress is trans- mitted across the membrane, forcing the interior of cell to flow.

    Hence, at low shear conditions blood behaves like a suspension of interactive rigid particles and at high shear conditions as an emulsion. Additionally, deforma- tion takes place at constant cell surface area, because of large surface area to volume ratio (Lerche et al.,

  • G. Vlastos et al. 163 The effect of parallel combined steady and oscillatory shear flows

    Fig. 1 Steady shear rate depen- dence of steady shear viscosity for n = 3 2 healthy donors (mean_+std. dev.) and oscillatory shear rate dependence of visco- elasticity of human blood for n = 7 healthy donors (mean_std. dev.). The open symbols repre- sent data with a superimposed steady shear rate ~sup=2 s 1 on simple oscillatory shear (full symbols) for the same donor group (n=7). The drawings adja- cent to the curves display the ef- fects of RBC aggregation, disag- gregation and deformation at ap- propriate shear rates. Hct. = 0.45, f= 0.5 Hz, T = 25 °C

    ° J l

    r, l j

    100

    10

    • r/' osciiiatory shear r/'oscillatory shear

    --4>-- r/' oscillatory + steady shear q" osciilatory + steady shear

    R B C D I S A G G R E G A T I O N

    A N D

    R B C D E F O R M A T I O N

    R B C A G G R E G A T I O N

    i i i 1 [

    0.01

    @

    @ @

    i [ i r l l i l l i I r I I I H I r i i i i i i i L i i i i l r l l l i ~ t

    0.1 1 10 100 1000

    Steady and oscillatory shear rate, [s "l]

    1993b). The physical condition of red blood cells in plasma is shown by drawings adjacent to the curves, at appropriate shear rates in Fig. 1. For pathological blood, e.g., in cardiovascular diseases, an elevated shear resis- tance of aggregates has been observed, correlated to pathologically induced aggregation. According to this, the limiting stress values for dispersion of aggregates are displaced to higher values, i.e. to the right side of Fig. 1 (Lerche et al., 1991).

    While steady flow measurements mirror only energy losses, viscoelasticity provides information on both en- ergy loss and energy storage following structural changes. In Fig. 1 the viscous component of complex viscosity 17' is lower than the steady shear viscosity at low shear rates. At shear rates above 8 s -I the values begin to converge. A similar tendency is observed for the viscous component of complex viscosity at a con- stant superimposed steady shear rate ~?sup=2 s -1. The common values at higher shear rate.s, imply that energy dissipation in oscillatory and steady shear flow are equal. At low shear rates, shear strain increases continu- ously with time by steady shear. This leads to cell-to- cell contacts, producing large viscous energy dissipa- tion. By oscillatory shear at low shear rates, if the sinu- soidal strain value is below unity, the resultant energy dissipation would be lower than in steady shear (Thurston, 1994).

    The elastic component of complex viscosity ~" starts to depend on oscillatory shear rate for ~o> 1 s- as the viscous one. For ~o> 1 s -1 begins disaggregation of ery- throcytes, lowering both viscous and elastic component. At shear rates 7;'o > 30 s 1 the disaggregation is relatively complete and the only mechanism for storing elastic en- ergy is the deformation of individual red blood cells. The sharp decrease of ~/" at higher shear rates reflects

    the loss of ability by storing elastic energy in shear de- formation process. The values of viscous component are higher than the values of elastic component, for whole shear rate range.

    The effect of a superimposed steady shear rate ~sup=2 s 1 on simple oscillatory flow for the same healthy donor group, is a slight decrease of viscous ~/' and a remarkable decrease of elastic component 17" of complex viscosity. The values of q" decrease approxi- mately 10 times the values without superposition at low shear region. Simultaneously, there is a displacement of the oscillatory shear rate where the nonlinear region be- gins, to higher shear rates especially pronounced for q ". The q' curves without and with superposition coincide with the steady shear viscosity curve for ~o-> 8 s -1. The elastic component after superposition tends to converge with the same component without superposition for ~_>30 s -1.

    For a quantitative evaluation of the superposition ef- fect, five different steady shear rates )':sup (1 s -1 to 5 s -l) were superimposed on oscillatory shear (Fig. 2). The decrease of both t/' and q" with incrasing superimposed shear is obvious, especially at low shear region. The de- crease of r/" is much more pronounced than that of r/'. At higher shear region, the r/' curves with superposition converge with the curve with simple oscillation for 7o_>3 and there is a displacement of the critical strain at which the non-linear region starts, to higher amplitudes. The elastic component takes the values of t/" without superposition also at high shear region, but at higher amplitudes (?o >4).

    The steady shear stress should act additively on os- cillatory stress, and causes displacement of critical strain amplitude to behave nonlinearly, which becomes larger with increasing )sup. At higher shear region, the

  • Rheologica Acta, Vol. 36, No. 2 (1997) 0 Steinkopff Verlaa 1997

    -

    Shear strain amplitude

    Fig. 2 Viscous q' and elastic q" component of complex viscosity for human blood from a representative healthy donor (hct.=0.45) vs. shear strain amplitude, at different superimposed steady shear rates j,,, in sC1 (open symbols). The full symbols represent data without superimposed steady shear (y,,, =0). f = 0.5 Hz, T=25 "C

    larger strains weaken the effect of superimposed shear, i.e., all curves with superposition tend to converge to the curves with simple oscillation (Fig. 2). At low shear region steady shear predominates the oscillatory shear. A low shear strain amplitude, e.g. yo=0.21 corresponds to an oscillatory shear rate at 0.5 Hz, jo=o yo= (2710 y0=0.66 s-I (maximal shear rate during one oscil- lation cycle). This shear rate value is smaller than the superimposed ones (I s-' l ys,,55 s-I), i.e. the low shear behavior will be controlled from steady shear. At higher shear region, for yo =2.92 the corresponding os- cillatory shear rate is yo= 9.17 s-' which is larger than the range of superimposed steady shear rates, thus oscil- latory shear dominates at this region.

    In Fig. 3 the blood viscoelasticity from the same healthy donor as in Fig. 2 was measured in dependence on oscillation frequency, without and with application of five superimposed steady shear rates of magnitude equal to those in Fig. 2. The oscillation amplitude was kept constant and low enough (yo=0.2) in order to obtain the linear viscoelastic properties of blood, but in view of instrument limitations (resonance frequencies) measure-

    Frequency, !Hz]

    Fig. 3 Frequency dependence of viscous component of complex vis- cosity q' at small amplitude for a representative blood sample, at five different superimposed steady shear rates j,,, (open symbols). The full symbols display values of q' without superposition of steady shear (j,,,=O). The crosses represent the frequency dependence of elastic component q" at a superimposed steady shear rate j,,,= 1 s-'. Hct.=0.45, T=25 "C

    ments at very low amplitudes could not be attained. Blood exhibits a frequency dependence over the fre- quency range investigated, implying a broad spectrum of relaxation times. The viscous component q ' of com- plex viscosity decreases with increasing superimposed steady shear at low frequency region. At higher frequen- cies the q' values after superposition tend to coincide with these without steady superposition. The elastic com- ponent q" of complex viscosity without superposition (data not shown) displays a similar frequency depen- dence as q l , but lower values. At very low frequencies blood elasticity tends to diminish for f ~ 0 . 0 7 Hz, since for f -t 0 the flow becomes steady, and y ' values approach the zero steady shear viscosity value 70.

    The frequency-dependent elasticity after superposi- tion of steady shear could be measured only for js,, = 1 s-l. Above j,,, > 1 s-', considering the sharp de- crease of the elastic component with increasing super- imposed steady shear, the values of q" were out of measurement range. The elasticity decreases markedly with superposition at low frequencies and tends faster to zero values with decreasing frequency. The sharp de- crease of q" shows similarities with the character of G ' curves displaying negative values at low frequencies, obtained for polystyrene and polyethylene oxide solu- tions in Acroclor-1248 after steady superposition (Lau- fer et al., 1975).

    The addition of superimposed shear on small oscilla- tory shear modifies the distribution of relaxation times, as has been investigated for the orthogonal superposi- tion of steady and oscillatory flow for polymer solu-

  • G. Vlastos et al. 165 The effect of parallel combined steady and oscillatory shear flows

    Fig. 4 Shear strain dependence of viscous r/' and elastic compo- nent J/" of complex viscosity for 100 an aqueous polyacrylamide ~ (AP 273E) and an aqueous xanthan solution (representative data) at different concentrations 10 in ppm. The thick solid lines re- . ~ present the viscous t I ' (top) and ~" elastic component r/" (bottom) of complex viscosity for the 99% percentile of human blood from a healthy control group (n = 7, mean+_2 std. dev.) at hematocrit 0.45. f=0.5 Hz, T=25 °C

    POLYACRYLAMIDE AP 273E

    1 . . . . . . . . I . . . . . . . . I . . . . . . . . i

    0.01 0.1 1 10

    I00

    10

    1 0,01

    1000 750 5O0

    250

    125 # , , # @ @ ~ $ ~

    XANTHAN

    0.1 1 10

    1000 100 1 10oo ~ _

    10

    0.1 . . . . . . . . I . . . . . . . . t . . . . . . . . J 0,1

    0.01 0.1 1 t0 0 .01 0.1 1 10

    S h e a r s t r a i n a m p l i t u d e S h e a r s t r a i n a m p l i t u d e

    tions (Osaki et al., 1965; Tanner and Williams, 1971). The effect of superimposed shear flow on viscoelastic- ity in Fig. 3, can be associated with the cut of long time end of the relaxation spectrum. The missing longer relaxation times then lead to decrease of viscosity t/'. Mewis and Schoukens (1978) calculated the equili- brium relaxation spectra after orthogonal superposition for colloidal dispersions, and compared them with theo- retical predictions based on a chain-like structural mod- el. They used the relaxation spectrum to calculate chain dimensions during flow.

    In order to obtain model fluids that mimic the theo- logical behavior of blood, the appropriate concentration being blood-like for the fluids under investigation must be determined. The AP 273E is high molecular weight polyacrylamide. Polyacrylamides are anionic and drag- reducing when diluted in water (Vlassopoulos and Schowalter, 1994). Xanthan gum is high molecular weight extracellular polysaccharide. It is highly pseudo- plastic and its conformation in solution is rodlike with some flexibility (Whitcomb and Macosko, 1978). We measured the shear strain amplitude dependence of an aqueous polyacrylamide AP273E and an aqueous xanthan solution at different concentrations and com- pared them with normal values of blood from 7 healthy donors (99% percentile at hematocrit 0.45, frequency 0.5 Hz) in respect to viscous and elastic component of complex viscosity.

    The polyacrylamide solution at concentrations from 50 to 750 ppm in Fig. 4 displays viscoelasticity, with

    the peculiarity that elastic component disposes higher values than viscous component at all concentrations and whole shear range, in contrast to xanthan. The visco- elasticity values increase with increasing polymer con- centration. The concentration of 125 ppm comes closer to blood viscoelasticity values (hct.=0.45) for r/', but t/" displays higher values than blood.

    The xanthan gum solution (Fig. 4) shows a nonlinear shear strain dependence for 7o > 1 and at a concentration of 500 ppm approaches better the "normal" interval of blood. The elastic component of xanthan solution shows a sharper decrease with increasing strain ampli- tude as the polyacrylamide solution, especially for 1000 ppm concentration. Lim et al. (1984) found the strain amplitude at which transition from linear to non- linear viscoelastic behavior occurs, depending on xanthan concentration. They investigated concentrated xanthan solutions (smaller concentration 1000ppm), being nonlinear for 500% strain at an angular frequency of 1.0 rad/s. In contrast, the shear strain at which visco- elasticity starts to decrease in Fig. 4 is nearly the same at all concentrations for both solutions (70~ 1). That means, the relaxation processes dominating at 0.5 Hz are not changed by dilution in this concentration range, observed also for similar measurements at 2 Hz from Thurston and Pope (1981).

    In Fig. 5 is shown the response of polyacrylamide solution to application of different superimposed steady shear rates in dependence on shear strain amplitude. The experimental conditions were identical with those

  • 166 Rheologica Acta, Vol. 36, No. 2 (1997) © Steinkopff Vertag 1997

    Fig. 5 Viscous ~I' and elastic */" component of complex viscosity for an aqueous polyacrylamide (AP 273E) and an aqueous -g xanthan solution vs. shear strain amplitude, at different rates of superimposed steady shear 9~p (representative data). The sym- bols m'e the same as for blood in "m Fig. 2. f=0.5 Hz, T=25 °C 10

    20

    10

    p•p 273E, 125 ppm !~

    I I t r l l l I I r J ~ l l l l , I I I I I l l [

    0. l 1 10 --Q POLYACRYLAMIDE AP 273E, 125 ppm

    0.1 1 10

    Shear strain amplitude

    30

    20

    10

    I0

    1

    0.1

    XANTHAN 500 ppm

    i i t l ~ l l i i i i J l l l [ i i i l l l l t [

    0.1 1 10

    XANTHAN 500 ppm

    0.1 1 10

    Shear strain amplitude

    for blood in Fig. 2. With increasing superimposed steady shear rate ?)~uv from 1 s -I to 5 s -1 a slight de- crease of viscous *7' and a pronounced decrease of elas- tic *7" component of complex viscosity is observed. There is also a displacement of strain amplitude at which transition to nonlinear viscoelastic behavior oc- curs, to higher shear amplitudes especially for .7". The .7' curves of superimposed shear tend to coincide with the curve of simple oscillatory shear for Yo >--8. The elas- tic component of superimposed shear tends to converge with the same component of simple oscillatory shear for 7o_> 10, particularly for the viscous component.

    The xanthan solution at blood-like concentration of 500ppm was subjected to different superimposed steady shear rates and the shear strain dependence was measured (Fig. 5). The behavior of xanthan solution does not differ that much from that of polyacrylamide solutions. The decrease of *7' and especially ~/" with in- creasing superimposed shear, at low shear region, mark the behavior of xanthan solution as well. There is a dis- placement of the beginning of nonlinear region, to high- er amplitudes especially for .7". The .7' curves of super- imposed shear tend to coincide with the curve of simple oscillatory shear for 7o>8. The elastic component of superimposed shear tends to converge with the same component of simple oscillatory shear for yo>__9, but it seems that this tendency is not further maintained for 7o_> 10. In regard to superposition of steady shear rate on oscillatory shear, both solutions exhibit a similar

    qualitative behavior with blood, i.e. the decrease of vis- cous and elastic component of complex viscosity with increasing superimposed shear.

    In order to demonstrate the influence of superim- posed steady shear rate ))sup on viscoelastic parameters *7' and '7" at constant frequency, we replotted in Fig. 6 the data from Fig. 2 for human blood at 0.5 Hz for three selected shear strain amplitudes 7o. The three shear strain amplitudes (0.21, 0.93 and 2.92), corre- spond to lower, transition and higher shear regions, re- spectively.

    The viscous component of blood decreases gradually with incrasing ))sup. At higher shear strains, e.g. 7o=2.92, there is only a slight dependence on superim- posed shear rate and *7' is almost independent of super- imposed shear. In contrast, the dependence of elastic component *7" on superimposed steady shear rate is very strong and decreases exponentially with increasing superimposed shear rate for ))sup_

  • G. Vlastos et al. 167 The effect of parallel combined steady and oscillatory shear flows

    16 ~ • Y° = 0"21 ' ~ • /o=0.93

    10[ ~ • •

    g L_ I I I I I I

    0 1 2 3 4 5

    12 • Yo =0.21

    • Yo = 0.93 • Y0 =2.92

    \ \ \

    a 4 ~

    0 ---'-- --~2~_::Z=-| I I I I [ I

    0 1 2 3 4 5

    S u p e r i m p o s e d s t e a d y s h e a r r a t e , I s d ]

    Fig. 6 Dependence of viscous ~/' and elastic component 1/" of com- plex viscosity on superimposed steady shear rate at three selected shear strain amplitudes Yo, for healthy human blood (representative data). The solid lines represent a data fit by the modified Carreau

    , . , . ~ . 2 - p equation q 0 ,~ , )=q (Ysup=0)[l+(XYsup)] . The dashed lines repre- sent data fit by the equation t/"(gsup)=~'0)sup=0)e -bk~p, Hct.=0.45, f=0.5 Hz, T=25°C

    20 [ • )'o=0.4

    V " ro =3,o .

    2a- ]2

    8 [ I _ _ L _ [ I I

    0 1 2 3 4 5 30

    • • Yo = 0 . 4 \ \ • 7o=3.0 \

    \ A /o=9 .0 ~" 20 \

    wv... • N-,, X N N N

    10 ~ \

    0 L t ! t t p t

    0 1 2 3 4 5

    S u p e r i m p o s e d s t e a d y s h e a r r a t e , Is -1]

    Fig. 7 Dependence of viscous q ' and elastic component r/" of com- plex viscosity on superimposed steady shear rate )?sup (representative data) at 3 selected shear strain amplitudes 2o, for an aqueous poly- acrylamide solution (AP 273 E, concentration 125 ppm). The solid lines represent data fit by the modified Carreau Eq. (1). The dashed lines represent data fit by the exponential Eq. (2). f=0.5 Hz, T=25 °C

    )~sup. The curve for Yo = 0.4 exhibits a slight decrease up to ~)s~p = 1 s -~ and above 1 s -~ the decrease becomes marked. At higher shear strains, e.g. 7o=9.0, there is only a slight dependence on superimposed shear rate and r/' is almost independent of superimposed shear. The elastic compo- nent r/" decreases exponentially with increasing superim- posed shear rate for whole shear rate region and for all amplitudes. For Ysup > 5 s -t t/" reaches extremely low val- ues for all shear strain amplitudes. At higher shear amplL tudes, e.g. 7o=9,0, q" is nearly independent of superim- posed shear.

    The results on superimposed steady shear depen- dence for the aqueous xanthan solution at a 500 ppm concentration, are shown in Fig. 8 (data from Fig. 5) at 0.5 Hz, for three shear strain amplitudes 7o (0.4, 3.0 and 9.0). The viscous component of xanthan solution decreases also with increasing ~)~p. The curve for 70=0.4 is independent of superimposed shear up to ))s- =2-1 and above 2 s -t the decrease becomes marked, following power law behavior. The elastic component r/" decreases also exponentially with increasing super-

    imposed shear rate for whole shear rate region and for all amplitudes. For ~sup_>5 s -1 /I" reaches very low val- ues for all Yo. At higher shear amplitudes, e.g. 7o=9.0, both q' and q" are independent of superimposed shear.

    The dependence of i/' on superimposed steady shear rate ?)sup looks similar to steady shear rate dependence of apparent viscosity for these solutions (Vlastos et al., 1992; Lerche et at., 1993a). In order to compare blood and polymer solutions quantitatively, the Carreau equa- tion A (Carreau, 1972) describing the steady shear rate dependence of viscosity, was modified. The steady shear viscosity r/ of Carreau equation was replaced by the viscous component of complex viscosity as a func- tion of superimposed steady shear rate I/' (gsup) in mPa s, the zero shear viscosity r/o by the viscous com- ponent at simple oscillatory shear ~/' (gsup=0) in mPa s, and the steady shear rate 9 of Carreau equation by the superimposed steady shear rate 9sup in s -I. Then, the modified Carreau equation would take the form

    f / (gsup) : r/(~)sup : 0)[1 -~- (}@sup)2] - p (1)

  • 168 Rheologica Acta, Vol. 36, No. 2 (1997) © Steinkopff Verlag 1997

    28

    24

    20

    16

    • ; , o = 0 . 4

    • 70 = 3 . 0

    ~ • • Y0 = 9 .0

    • A - , , L - •

    12 I I I I I I 0 1 2 3 4 5

    16 • yo = 0 .4

    • J'o = 3 .0

    • Y0 = 9 , 0

    \ \

    12 \

    ~., ...

    8

    4 . . . . I ~ . . . .

    0 I [ I I I I

    0 1 2 3 4 5

    Superimposed steady shear rate, [s -1]

    Fig. 8 Dependence of viscous //' and elastic component r/" of com- plex viscosity on superimposed steady shear rate )~p (representative data) at three selected shear strain amplitudes ~o, for an aqueous xanthan solution (concentration 500 ppm). The solid lines represent data fit by the modified Carreau Eq. (1). The dashed lines represent data fit by the exponential Eq. (2). f=0.5 Hz, T=25 °C

    where 2 is a constant in s and p a dimensionless param- eter representing the slope of power law region. The re- sults after fitting by Eq. (1) are shown as solid lines at the top of Figs. 6, 7 and 8 and the fit values outlines Table 2.

    The viscous component of complex viscosity at sim- ple oscillatory shear ~/' (gsup=0) for xanthan and polya- crylamide solutions, is higher than blood as the result of selected polymer concentration, at low shear region. The parameter 2 which is characteristic for every fluid, is larger for blood than for the polymer solutions, and reaches its smallest value for the xanthan solution. The parameter p shows an inverse behavior, i.e. it is larger for the xanthan solution and becomes the smallest value for blood. The data for q" were fitted by the exponen- tial equation

    //tt()sup) = /~/t(~)sLtp = 0)e b)sup (2)

    where J/"(9~up) represents the elastic component of complex viscosity after superposition in mPa s, ,/" (9~up=0) the value of elastic component 17" at simple

    Table 2 Values of equation parameters of modified Can-eau equation / / ' ( ) ) s u p ) = / / ' ( ) ) s u p = 0 ) [ l + ( ) ~ s u p ) 2 ] -p, for blood, an aqueous polyacryla- mide (AP 273E), and an aqueous xanthan solution at 70=0.4, after superposition of steady on oscillatory shear. The blood parameters refer to 70=0.21 amplitude, f=0.5 Hz, T=25°C

    //' (~)~p = 0) 2 p (mPa s) (s)

    Blood hct. = 0.45 15.9 1.0 0.2 Polyacrylamide 125 ppm 18.0 0.3 0.3 Xanthan 500 ppm 24.3 0.03 10.5

    Table 3 Parameters of exponential equation //" (Tsup) =//" (?)sup =0) e m~up for blood, an aqueous polyacrylamide (AP 273E), and an aqu- eous xanthan solution at 7o=0.4, after superposition of steady on os- cillatory shear. The blood parameters refer to 7o=0.21 amplitude. f=0.5 Hz, T=25 °C

    J/" ())sup = 0) b (mPa s) (s)

    Blood hct. = 0.45 11.4 1. l Polyacrylamide 125 ppm 25.7 0.5 Xanthan 500 ppm 14.3 0.3

    oscillation (~sup=0) in mPas, b_>0 is a fit parameter and ~sup represents the superimposed steady shear rate in s - . The dashed lines in Figs. 6, 7 and 8 (bottom) display the fit results by Eq. (2). Table 3 describes the values of equation parameters for 7/".

    The elastic component at simple oscillatory shear is larger for the polyacrylamide solution following from that of xanthan solution. Blood shows the smaller value at low shear region. The parameter b, representing the intensity of exponential dependence, is higher for blood than for the polymer solutions. The polyacrylamide shows a sharper exponential dependence than xanthan.

    With the above-mentioned equations the viscoelastic parameters */' (~sup) and q"())sup) after superposition of steady on oscillatory shear could be calculated, by repla- cing the measured viscoelastic parameters at simple oscil- latory shear conditions q' (9~up=0) and */" (9~up=0), in Eqs. (1) and (2), at a constant frequency and shear strain amplitude. Although Carreau equation is not appropriate describing blood (Easthope and Brooks, 1980), its use was made for comparison to polymer solutions.

    For the frequency dependence of t/' in Fig. 3, if the amplitude of oscillatory shear is small, then the effect of superimposed steady shear on r/' in terms of stresses should be additive resulting from two stress contribu- tions, a) from a steady shear stress due to steady shear flow, and b) from an oscillatory shear stress due to os- cillatory motion, as has been already theoretically stud- ied (Pipkin and Owen, 1967). From parallel superposi- tion experiments of glass bead suspensions (Masi et al., 1984), the above-mentioned principle of stress contribu- tion was specified as a sum hypothesis stating that, if

  • G. Vlastos et al. 169 The effect of parallel combined steady and oscillatory shear flows

    100

    m

    ~.~ 10

    "a-

    • .

    1 r i r ~ l r l r I r E ~ l r ] _ _ ~ i , ~ i r l l ]

    0.l 1 10

    = L p + ;0 , [s-'l

    Fig. 9 Viscous component 0' of complex viscosity for a representa- tive blood sample, alter replotting the data from Fig. 3 vs. the total maximal shear r a t e ) t = ) ) s u p q - ) ) o where ))sup represents the superim- posed steady shear rate, and ?)o=e)•0 the oscillatory shear rate. The symbols indicate measurements at different superimposed steady shear rates same as those in Fig. 3. Hct.=0.45, T=25 °C

    r/' values (frequency-dependent) at different superim- posed steady shear rates were replotted versus a total shear rate, defined as the sum of steady and oscillatory shear rate (maximal during one oscillation cycle), then the data should assemble in a master curve. Conse- quently, we replotted in Fig. 9 the t/' data from Fig. 3 versus a maximal total shear rate ~t defined as ~)t=~)sup-t-~)0 where ~sup represents the superimposed steady shear rate and ?)0=o~ ?~o the maximal oscillatory shear rate during one oscillation cycle.

    The data points tend to gather in a single curve. However, agreement with the above-mentioned hypoth- esis is not precise, due to very low viscosity values for the curves under high superimposed shear measured at the limit of instrument sensitivity, and the constant small amplitude value being on the upper limit of linear viscoelastic range. The replotting of data from the other type of oscillatory experiment in Fig. 2 according to the above proposed sum criterion (data not shown) failed to yield in a single curve, indicating that at this type of flow the partial stresses cannot be added for blood.

    Discussion

    The investigation of pulsatility of blood flow can eluci- date some phenomena in the microcirculation, as an ad- ditional factor determining the in vivo rheological state of circulating blood. The fact that human blood is non- Newtonian and under unsteady flow conditions is a vis- coelastic fluid, was often disregarded in microrheologi-

    cal studies (Cokelet, 1987). The blood flow can be con- sidered as superposition of steady on oscillatory flow at a constant frequency initiated from the heart beat. The in vitro rheological measurements of such a flow pro- vide quantitative considerations on possible changes of blood structure caused by changes in type of flow. The most significant effect of steady superposition on oscil- latory flow for blood is the decrease of elasticity r/" at low shear deformations. At this region the elasticity of blood is characterized mainly by the elasticity of large RBC aggregates. The decrease of elasticity after super- position indicates that the aggregation process of RBC's is reduced, as has been confirmed in optical aggrego- metric studies (Riha and Stolz, 1996).

    There exist two modes managing the parallel super- position of steady on oscillatory shear, in order to ex- amine their interaction, a) the oscillatory shear ampli- tude is fixed and the steady shear is varied, or b) the steady is constant and the oscillatory shear amplitude is varied. We applied the last mode in our experiments. From other investigations dealing with blood superposi- tion in rigid tubes (Thurston, 1975), a small decrease of r/' and a great reduction of r/" is manifested, when the combination of partial flows results in mode (b), which is in agreement with our findings. In contrast, it was observed that in mode (a) changes at the added oscilla- tory amplitude do not modify the steady component. Hence, the question of whether combination of partial flows results in a linear way, should depend from the fact on which flow the other one has been superim- posed on, and from their quantitative relationship, i.e., which type of flow dominates at the actual shear range. Nevertheless, the complex reality of stress distribution in the microcirculation is perplexing when we attempt to determine some, at least qualitatively, outlined inter- actions between partial flows resulting in the pulsatile character of blood flow. Moreover, this point demands further investigation.

    Another phenomenon by superposition, mentioned elsewhere (Booij, 1968), is the existence of large oscilla- tory normal stress measured for non-Newtonian liquids (ethylene-propylene), when they are exposed to both small oscillatory and steady shear. The amplitude of nor- mal stress is proportional to that of oscillatory shear. The real part of complex normal stress modulus, as function of frequency shows a maximum and the imaginary part is positive at low frequencies and negative at high fre- quencies. However, both components of normal stress de- pend on value of superimposed shear and increase with increasing superimposed steady shear rate. If we take into account that the existence of normal forces in blood is controversial or they are too small to be measured (Copley and King, 1975), the application of superim- posed steady shear could produce also in blood large forces in perpendicular direction to flow, causing interac- tions with the arterial wall, and leading to high concentra-

  • 170 Rheologica Acta, Vol. 36, No. 2 (1997) © Steinkopff Verlag 1997

    don of local stresses. This has been mentioned as a pos- sible important localizing factor in atherosclerosis (Oka, 1981; Liepsch, 1990). Mann and Tarbell (1990) found that aqueous xanthan solutions exhibit normal stresses about one-third that of aqueous polyacrylamide solu- tions, measured in an atherogenic artery model. Further investigations combined with in vivo and visual studies have to be carded out.

    Caution is necessary in interpreting in vitro rheologi- cal data for conditions existing in the microcirculation (Schmid-Sch6nbein, 1981; Usami, 1982), because at ar- teries, veins, and capillaries various effects occur link- ing to each other such as axial migration, very high shear stresses, and alterations in vessel geometry. How- ever, in vitro viscometric data constructs a basis con- taining many influencing factors (large shear rate de- pendence, hematocrit variability, etc.), on which phe- nomena of microcirculation can be better evaluated.

    The rheological behavior of human blood in respect to viscoelasticity can be imitated by model fluids of appro- priate concentration (see Fig. 4). They exhibit similar re- sponse to superposition for both components of complex viscosity. In regard to blood, they exhibit similar qualita- tive dependence by elasticity (see Figs. 6, 7 and 8) but another type of behavior for the viscous component.

    The quantitative agreement with blood can be achieved selecting appropriate molecular weight, con- centration and solvent. In respect to steady viscosity, polymers with broad molecular weight distribution (MWD) show onset of shear thinning at lower shear rates than polymers with narrow MWD (Graessley, 1974). The fluids used in this study possess this feature, so they can better match the steady dependence of blood (Vlastos et al., 1992). A proper molecular charac- terization of model fluids is necessary in order to achieve a more accurate comparison with blood. In respect to the above, it would be useful for the future to compare other material properties of blood and model fluids, like relaxation times or intrinsic viscosities. Vlas- sopoulos and Schowalter (1993) examined dilute aqu- eous polyacrylamide solutions in steady streaming flow. They found that in order to obtain the same Maxwell relaxation times of the fluids, the frequency must be in- creased at the same concentration, or the concentration must be increased at a constant frequency.

    Viscoelasticity determined at a higher fixed fre- quency of oscillation displaces the onset to nonlinear re-

    gion at higher shear strain amplitudes for blood (Thur- ston, 1972). For polymer solutions, however, employ- ment of higher frequency seems not to change the onset value to nonlinear region very prominently (Thurston and Pope, 198!). Since the onset to nonlinear region occurs at lower shear strains by blood than by polymer solutions, measurements at higher frequencies could re- duce the discrepancy in respect to departure of non- linear region.

    The increase of ionic strength of solutions using electrolytes (MgC12) as solvents, has been already tested (Vlastos et al., 1992). The addition of salt de- creased steady shear viscosity, but increased the critical shear rate to reach limiting low shear conditions (New- tonian behavior). This deviates the behavior of model fluids even more from that of blood.

    Blood and model fluids feature a different structure. The form of erythrocytes in blood is unlike the random- coil polymer macromolecules. The orientation of flow is becoming more significant for blood as for the fluids investigated. Due to their biconcave form, erythrocytes are more sensible to the direction of applied stresses, whereas for spherical polymer molecules orientation is less substantial. Additionally, the absence of particle ag- gregation is a pronounced distinction between blood and model fluids investigated. Some investigators devel- oped blood-like model fluids simulating the aggregation of red blood cells occurring in blood. Fukada et al. (1989) used polystyrene microspheres to mimic red blood cells, and dextran to imitate plasma proteins. Although polystyrene particles are not deformable like red blood cells, an appropriate choice of particle dimen- sions and concentrations combined with suitable sol- vents could produce promising model fluids.

    The distinct viscoelastic behavior in respect to super- position for blood and model fluids and correspond- ingly to pulsatile flow, establishes new claims for the development and characterization of blood-like model fluids. It is essential that model fluids should imitate not only steady shear and viscoelastic behavior of hu- man blood, but they should also display analogous be- havior in respect to superposition of steady on oscilla- tory shear.

    Conclusions

    We investigated the influence of parallel superimposed steady shear on oscillatory shear at constant frequency, for human blood and two blood-like model fluids (aqu- eous polyacrylamide AP 273E and xanthan gum solu- tions) in terms of viscoelastic parameters 71' and ,/". The appropriate concentrations of model fluids were de- rived from concentration dependence measurements, comparing the shear strain amplitude dependence of model fluids with that of human blood at standard he- matocrit. The polyacrylamide solution AP 273E dis- played viscoelastic behavior close to that of blood at a concentration of 125 ppm and the xanthan solution at 500 ppm, at low shear region. At the transition and higher shear region, all model fluids at the above con- centrations dispose higher viscous and elastic compo- nent than blood.

    The effect of different superimposed steady shear rates on oscillatory shear for blood is the decrease of

  • G. Vlastos et al. 171 The effect of parallel combined steady and oscillatory shear flows

    viscous and elastic component of complex viscosity. This effect is more pronounced at lower shear region, where as higher the superimposed shear rate, as lower are the components t/' and q". The most prominent al- teration occurred by elasticity, whereas the change by the viscous component is relatively small. At higher shear region, there was no noticeable variation of visco- elastic parameters. The superposition of steady shear holds the aggregation tendency of red blood cells at low shear region, and delays the start of disaggregation. The dependence of r/" on superimposed steady shear rate is exponential.

    For blood, the superposition of steady shear on small amplitude oscillatory shear modified the relaxation times, leading to decrease of viscous component of complex viscosity, with increasing steady shear. The elastic component could be detected only for one super- imposed steady shear, in consideration of marked de- crease at low frequency region, tending to zero values with decreasing frequency.

    The response of model fluids to superposition ap- pears qualitatively similar to blood for elasticity, but shows different quantitative dependence. For the vis- cous component, the superimposed shear dependence of all model fluids investigated differs from that of blood and resembles that of steady shear viscosity. The depen- dencies of r/' and t/" on superimposed steady shear rate were described by a modified Carreau equation for r/', and an exponential equation for 17". Further investiga- tions have to be performed to elucidate the phenomena of RBC aggregation and deformation, like dextran-in- duced aggregation experiments, time-dependent studies and normal stress measurements, in respect to interac- tions between steady and oscillatory shear.

    Acknowledgment The authors thank Prof. Lutz-Gfinther Fleischer from Technical University of Berlin for the helpful discussions, Mrs. Margrit Neumann for the tireless supply of literature, and Mrs. Chris- tine Koch for the technical assistance. The authors thank also Prof. Georgios Georgiou from the University of Cyprus and the Hellenic Society of Rheology, for the excellent organization of the Symposium dedicated to the memory of Prof. A.C. Papanastasiou.

    References

    Argumedo A, Tung TT, Chang KI (1978) Rheological property measurements of drag-reducing polyacrylamide solutions. J Rheol 22:449-470

    Barnes HA, Townsend R Waiters K (197[) On pulsatile flow of non-Newtonian fluids. Rheol Acta 10:517-527

    Booji HC (1966a) Influence of superimposed steady shear flow on the dynamic proper- ties of non-Newtonian fluids. I. Measure- merits of non-Newtonian solutions. Rheol Acta 5:215-221

    Booji HC (1966b) Influence of superimposed steady shear flow on the dynamic proper- ties of non-Newtonian fluids. II. Theoreti- cal approach based on the Oldroyd theory. Rheol Acta 5:222-227

    Booji HC (1968) Influence of superimposed steady shear flow on the dynamic proper- ties of non-Newtonian fluids. IlL Mea- surements on oscillatory normal stress components. Rheol Acta 7:202-207

    Brookshier KA, Tarbell JM (1993) Evalua- tion of a transparent blood analog fluid: Aqueous xanthan gum/glycerin. Biorheol- ogy 30:107-116

    Carreau PJ (1972) Rheological equations from molecular network theories. Trans Soc Rheol 16:99-127

    Chien S (1970) Shear dependence of effec- tive cell volume as determinant of blood viscosity. Science 168:977-979

    Chien S, Usami S, Dellenback RJ, Gregersen MI (1967a) Blood viscosity: Influence o~C erythrocyte deformation. Science 157: 827-829

    Chien S, Usami S, Dellenback R J, Gregersen MI, Nanninga LB, Guest MM (1967b) Blood viscosity: Influence of erythrocyte aggregation. Science 157:829-831

    Chien S, King RG, Skalak R, Usami S, Cop- ley AL (1975) Viscoelastic properties of human blood and red cell suspensions. Biorheology 12:341-346

    Cokelet GR (1987) The rheology and tube flow of blood. In: Skalak R, Chien S (eds) Handbook of Bioengineering. McGraw-Hill, pp 14.1-14.17

    Copley AL, King RG (1975) On the visco- elasticity of anticoagulated wheol human blood in steady shear as tested by rheogo- niometric measurements of normal forces. Biorheology 12:5-10

    De Cleyn G, Mewis J (1981) A constitutive equation for polymer liquids: Application to shear flow. J Non-Newtonian Fluid Mech 9:91-105

    Easthope PL, Brooks DE (1980) A compari- son of theological constitutive functions for whole human blood. Biorheology 17:235-247

    Fukada E, Seaman GVE Liepsch D, Lee M, Friis-Baastad L (1989) Blood modeling using polystyrene microspheres. Biorheol- ogy 26:401-413

    Gaethgens P (1970) Pulsatile pressure and flow in the mesenteric vascular bed of the cat. Pfltigers Arch 316:140-151

    Graessley WW (1974) The entanglement concept in polymer theology. Adv Polym Sci 16:1-179

    Intaglietta M, Richardson DR, Tompkins WR (1971) Blood pressure, flow, and elastic properties in microvessels of cat omen- turn. Am J Physiol 221:922-928

    Isayev AI, Wong CM (1988) Parallel super- position of small- and large-amplitude os- cillations upon steady shear flow of poly- mer fluids. J Polym Sci Pol Phys 26:2303-2327

    Jones TER, Waiters K (1971) The behaviour of materials under combined steady and oscillatory shear. J Phys A Gen Phys 4:85-100

    Kiesewetter H, Radtke H, Jung E Schmid- SchOnbein H, Wortberg G (1982) Deter- mination of yield point: Methods and re- view. Biorheology 19:363-374

    Ku DN, Liepsch D (1986) The effects of non-Newtonian viscoelasticity and wall elasticity at a 90 ° bifurcation. Biorheol- ogy 23:359-370

    Laufer Z, Jalink HL, Staverman AJ (1975) Dynamic properties of some polymer so- lutions subjected to a steady shear super- imposed on an oscillatory shear flow. I. Experimental results. Rheol Acta 14:641- 649

    Lerche D, Bfiumler H, Kucera W, Meier W, Paulitschke M (1991) Flow properties of blood and hemorheological methods of quantification. In: Sch~tt W, Klinkmann H, Lamprecht I, Wilson T (eds) Physical characterization of biological cells. Vertag Gesundheit GmbH, Berlin, pp 189-214

  • 172 Rheologica Acta, Vol. 36, No. 2 (1997) © Steinkopff Verlag 1997

    Lerche D, Vlastos G, Koch B, Pohl M, Af- feld K (1993 a) Viscoelastic behaviour of human blood and polyacrylamide model fluids for heart valve testing. J Phys lII France 3:1283-1289

    Lerche D, Koch B, Vlastos G (1993b) Flow behaviour of blood. Rheology 3:105-112

    Lessner A, Zahavi J, Sitberberg A, Frei EH, Dreyfus F (1971) The viscoelastic proper- ties of human blood. In: Hartert H, Cop- ley AL (eds) Theoretical and Clinical He- morheology. Springer, Berlin, pp 194-205

    Liepsch D (ed) Biofluid Mechanics. Blood flow in large vessels. Springer Verlag

    Lira T, Uhl JT, Prud'homme RK (1984) Rheology of self-associating concentrated xanthan solutions. J Rheol 28:367-379

    Malkin AY, Isayev AI, Vinogradov GV (1975) Power estimate of the effective- ness of vibrational thinning of polymer and filled systems. Polym Mech USSR 11:439-444

    Mann DE, Tarbell JM (1990) Flow of non- Newtonian blood analogs fluids in rigid curved and straight artery models. Biorheology 27:711-733

    Masi R Nicodemo L, Nicolais L, Taglialatela G (1984) Viscosity of glass bead suspen- sions in parallel superimposed flows. Rheol Acta 23:324-326

    Mewis J, Schoukens G (1978) Mechanical spectroscopy of colloidal dispersions. Faraday Discuss 65:58-64

    Oka S (1981) Cardiovascular hemorheology. Cambridge University Press

    Osaki K, Tamura M, Kurata M, Kotaka T (1965) Complex modulus of concentrated polymer solutions in steady shear. J Phys Chem 69:4183-4191

    Pipkin AC, Owen DR (1967) Nearly visco- metric flows. Phys Fluids 10:836-843

    Pohl M, Wendt MO, Wemer S, Koch B, Lerche D (1996) In vitro testing of artifi- cial heart valves: Comparison between Newtonian and non-Newtonian fluids. Artif Organs 20:37-46

    Riha P, Stolz JF (1996) Flow oscillations as a natural factor of reduction of the effect of RBC aggregation on blood flow. Clin Hemorheol 16:4348

    Rochefort WE, Middleman S (1987) Rheol- ogy of xanthan gum: Salt, temperature, and strain effects in oscillatory and steady shear experiments. J Rheol 31:337-369

    Schmid-Sch6nbein H (1981) Blood rheology and physiology of microcirculation. La Ricerca Clin Lab 11:13-33

    Tanner RI, Simmons JM (1967) Combined simple and sinusoidal shearing in elastic liquids. Chem Eng Sci 22:1803-1815

    Tanner RI, Williams GW (1971) On the orthogonal superposition of simple shear- ing and small-strain oscillatory motions. Rheol Acta 10:528-538

    Thurston GB (1972) Viscoelasticity of human blood. Biophys J 12:1205-1217

    Thurston GB (1975) Elastic effects in pulsa- tile blood flow. Microvasc Res 9:145-157

    Thurston GB (1994) Non-Newtonian viscosi- ty of human blood: Flow-induced changes in microstructure. Biorheology 31:179-192

    Thurston GB, Pope GA (1981) Shear rate de- pendence of the viscoelasticity of poly- mer solutions. II. Xanthan gum. J Non- Newtonian Fluid Mech 9:69-78

    Usami S (1982) Physiological significance of blood rheology. Biorheology 19:29-46

    Vlassopoulos D, Schowalter WR (1993) Characterization of the non-Newtonian flow behavior of drag-reducing fluids. J Non-Newtonian Fluid Mech 49:205-250

    Vlassopoulos D, Schowalter WR (1994) Steady viscometric properties and charac- terization of dilute drag-reducing polymer solutions. J Rheol 38:1427-1446

    Vlastos G, Koch B, Lerche D, Pohl M (1992) Polymer solutions as model fluids for the non-Newtonian behaviour of blood. In: Moldenaers P, Keunings R (eds) Theoretical and Applied Rheology, Vol. 2. Elsevier, Amsterdam, pp 747-749

    Vlastos G, Lerche D, Koch B (1994) Visco- elastic behaviour of human blood and blood-like model fluids with emphasis on low oscillatory shear rates. In: Gallegos C (ed) Progress and Trends in Rheology IV. Steinkopff Verlag, Darmstadt, pp 32-34

    Wells R, Schmid-Sch6nbein H (1969) Red cell deformability and fluidity of concen- trated cell suspensions. J Appl Physiol 27:213-217

    Whitcomb PJ, Macosko CW (1978) Rheol- ogy of xanthan gum. J Rheol 22:493-505