Table -Top Autoclavesmodels 1730, 2340, 2540, 3140, 3850, 3870 M & MK
1730MK Valueklave
Cat. No. MAN205-0309001EN Rev. H Tuttnauer Europe b.v., Paardeweide 36, 4824 EH, Breda, P.O. Box 7191, 4800 GD Breda, Netherlands. +31/76-5423510, Fax: +31/76-5423540
1
TABLE OF CONTENTS
PARAGRAPH PAGE NO.
1 INTRODUCTION ........................................................................................3
2 PERIODICAL TESTS .................................................................................3
3 SYMBOL DESCRIPTION ...........................................................................3
4 SAFETY INSTRUCTIONS..........................................................................4
5 WATER QUALITY ......................................................................................5
5.1 Water for Generating Steam .............................................................5
5.2 Reverse Osmosis................................................................................5
6 MAINTAINING AND REPLACING PARTS ...........................................10
6.1 Safety Tests after Repair .................................................................10
6.2 Dismantling the Outer Cover of the Autoclave...............................11
6.3 Cleaning and Replacing Air Trap Jet .............................................12
6.4 Replacing the Safety Valve .............................................................13
6.5 Replacing the circuit breaker..........................................................14
6.6 Temperature Safety Thermostat......................................................15
6.7 Raising the Working the Temperature of the Safety Thermostat...15
6.8 Cut-Off Thermostat .........................................................................16
6.9 Replacing Heating Elements...........................................................17
6.10 Replacing Multi-Purpose Valve ......................................................18
6.11 Unclogging the multi-Purpose Valve or Chamber .........................20
6.12 Pressure Door Lock System ............................................................20
6.13 Replacing the Door Bellows............................................................21
6.14 Replacing the thermostat B10.........................................................22
6.15 Replacement of the Door Cover ......................................................23
6.16 Replacing the Locking Device ........................................................25
6.17 Replacing the Door Switch (models 2540, 3150, 3850, 3870) ........26
6.18 Replacing the Drain Valve ..............................................................27
7 TROUBLESHOOTING .............................................................................28
8 LIST OF SPARE PARTS...........................................................................36
9 PRESSURE VS TEMPERATURE FOR SATURATED STEAM .............40
2
TABLE OF CONTENT (Cont.)
DRAWINGS PAGE NO.
Front View Model 1730 M, MK-Valueklave................................................................6
Front View Model 2340/2540 M, MK ..........................................................................7
Front View Model 3850/3870 M, MK ..........................................................................8
Rear View ......................................................................................................................9
General View of Vessel, Door and Accessories .........................................................32
Autoclave Cover ..........................................................................................................33
Door Tightening Bolt – Assembly ..............................................................................34
Multi-Purpose Valve Assembly...................................................................................35
Drawing of Electrical System of Table Autoclave Models 1730M, MK ...................44
Drawing of Electrical System of Table Autoclave Models 2340/2540 M, MK .........45
Drawing of Electric System of Table Autoclave Model 3140 M ...............................46
Drawing of Electric System of Table Autoclave Models 3850 M..............................47
Drawing of Electrical System of Table Autoclave Models 3870 M...........................48
Piping Diagram Table Top Autoclave Models: M and MK ......................................49
3
1 INTRODUCTIONThis manual, together with the operator’s manual, forms the complete edition of the Operation and Maintenance instructions. This manual is intended for the use of the technician. It is forbidden for unqualified and unauthorized personnel to service the autoclave in accordance with the instructions in this manual. Any unauthorized service may result in the invalidation of the manufacturer’s guarantee.
The qualified technician shall be an authorized electrician with the right qualifications in electronics and shall be familiar with the local technical/electrical regulations.
2 PERIODICAL TESTS
PERIOD TEST
1 months Test the safety valve by operating it.
6 monthsRemove the autoclave’s cover, tighten the heaters’ screws and electrical connections and valves.Check the continuity of the grounding connections.Perform validation of the autoclave.Check the precise operation of the earth leakage relay.Check that the autoclave is leveled.Check the safety elements; safety valve, door locking bellows and door locking mechanisms.Run the autoclave and verify that it operates as specified.Check the water reservoir, piping, plastic parts and electric wires.Check and tighten the piping joints to avoid leakage.Check and tighten all screw connections, heaters and valves and instrumentation.
Year
Calibrate the temperature and pressure once a year or in reference to local rules or regulations (refer to the section on Calibration).
5 years Observe the closing device for excessive wearSafety tests (pressure vessel, efficiency, electrical) shall be performed in accordance with local rules or regulations, by an authorized inspector.
Only an authorized technician shall perform the 6-months and yearly tests!
3 SYMBOL DESCRIPTION
Caution! Consult accompanying documents
Caution! Hot Surface.
Caution! Hot steam.
Ground
4
4 SAFETY INSTRUCTIONS
The autoclave has unique characteristics. Please read and understand the operation instructions before first operation of the autoclave. The following issues may require instructions guidance provided by the manufacturer: how to operate the autoclave, the door safety mechanism, the dangers involved in circumventing safety means, how to ensure that the door is closed, and how to select a correct sterilization program.Autoclave maintenance is crucial for the correct and efficient function of the device. We enclose a log booklet that includes maintenance recommendations, with every device.1. Make sure that you know where the main power switch is.2. Never use the autoclave to sterilize corrosive products, such as: acids,
bases and phenols, volatile compounds or solutions such ethanol, methanol or chloroform nor radioactive substances.
3. All autoclave users must receive training in proper usage from an experienced employee. Every new employee must undergo a training period under an experienced employee.
4. A written procedure must be established for autoclave operation, including: daily safety tests, seal inspection and door hinge inspection, smooth action of the closing mechanism, chamber cleaning, prevention of clogging and preservation from corrosion, what is permitted and what is prohibited for sterilization and choosing a sterilization program.
5. Before use, check inside the autoclave chamber to ensure that no items have been left from the previous cycle.
6. Load trays in such a way as to allow steam to move freely among all items.
7. Do not attempt to sterilize liquids since this autoclave is not intended to sterilize liquids.
8. When sterilizing plastic materials, make sure that the item can withstand sterilization temperature. Plastic that melts in the chamber is liable to cause a great deal of damage.
9. On closing the device door, make sure it is properly locked before activating.
10. Verify once again that you have chosen the appropriate sterilization program.
11. Before withdrawing trays, wear heat resistant gloves.12. Before opening the door, verify that there is no pressure in the chamber
(chamber pressure gauge is located on the autoclave's front panel).13. Open the door slowly to allow steam to escape and wait 5 minutes
before you remove the load. 14. Once a month, ensure that the safety valves are functioning, and once
annually a certified tester must conduct pressure chamber safety tests.15. Once annually, or more frequently, effective tests must be performed,
i.e., calibration and validation.16. Examine the condition of assemblies on a regular basis. Make sure there
are no leaks, breaks, blockages, whistles or strange noises.17. It is required to conduct maintenance operations as instructed.18. Immediately notify the person in charge of any deviation or risk for the
proper function of the device.
5
5 WATER QUALITY 5.1 Water for Generating Steam
The distilled or mineral – free water supplied to the sterlizer shall be according to the table below:
Physical Characteristics and Maximum acceptable contaminants levels in steam for sterlizers
(According to EN 13060:2004).
Element Condensate – allowable content
Silicium oxide. SiO2 ≤0.1 mg/kg
Iron ≤0.1 mg/kg
Cadmium ≤0.005 mg/kg
Lead ≤ 0.05 mg/kgRest of metals except iron, cadmium, lead
≤0.1 mg/kg
Chloride (Cl) ≤0.1 mg/kg
Phosphate (P2O5) ≤0.1 mg/kg
Conductivity (at 20°C) ≤3 μs/cm
pH value (degree of acidity) 5 to 7
Appearance Colourless clean without sedimentHardness (Σ Ions of alkaline earth)
≤0.02 mmol/l
Compliance with the above data should be tested in accordance with acknowledged analytical methods, by an authorized laboratory.
Attention: We recommend testing the water quality once a month. The use of water that does not comply with the table above may have severe impact on the working life of the sterilizer and can invalidate the manufacturer’s guarantee.
5.2 Reverse OsmosisA Reverse Osmosis system may be used to improve the quality of the water used to generate steam in the autoclave chamber. The use of mineral free will contribute to better performance and longer life of the autoclave.
6
FRONT VIEW MODEL 1730 M, MK-Valueklave
No. description No. description
1. Water reservoir cover 10. Dry indicator light
2. Water reservoir 11. Heat indicator light
3. Safety valve 12. Multipurpose valve
4. Air trap jet 13. Front legs
5. Timer 14. Reservoir water drain valve
6. Main power switch 15. Door Closing Device
7. Pressure gauge 16. Door Micro-switch
8. Thermostat (B10) knob 17. Door cover
9. Power indicator light 18. Autoclave cover
14
13
11
12
8
9
10
5
6
7
18
17
1 2 3 4
16
15
7
FRONT VIEW MODEL 2340/2540 M, MK
No. description No. description
1. Water reservoir cover 10. Heat indicator light
2. Water reservoir 11. Thermostat (B10) knob
3. Safety valve 12. Multipurpose valve
4. Air trap jet 13. Front legs
5. Pressure gauge 14. Rear legs
6. Main power switch 16. Door Closing Device
7. Timer 18. Door cover
8. Power indicator light 19. Autoclave cover
9. Dry indicator light
16
14
13
11
12
8
9
10
5
6
7
19
18
1 2 3 4
17
15
8
FRONT VIEW MODEL 3850/3870 M, MK
No. description No. description
1. Water reservoir cover 10. Heat indicator light
2. Water reservoir 11. Thermostat (B10) knob
3. Safety valve 12. Multipurpose valve
4. Air trap jet 13. Front legs
5. Pressure gauge 14. Reservoir water drain valve
6. Main power switch 15. Door Closing Device
7. Timer 16. Door Micro-switch
8. Power indicator light 17. Door cover
9. Dry indicator light 18. Autoclave cover
14
13
11
12
8
9
10
5
6
7
18
17
1 2 3 4
16
15
9
REAR VIEW
10
6 MAINTAINING AND REPLACING PARTS6.1 Safety Tests after Repair
ATTENTION!After every repair or dismantling the enclosure, the autoclave should pass two safety electrical test by the Service Engineer. The following shall be performed:
Warning!When re-installing the enclosure, connect the earthing to the cover. On installing the rear cover, connect the earthing before accomplishing the installation of the rear cover.
6.1.1 Enclosure Leakage Current Test.Every autoclave should pass this test as follows:
1. Connect the electrical cord to the autoclave.2. Turn on the main switch and the circuit breaker.3. Short-circuit the L and N pins on the cord's plug. 4. Connect the Short-circuit pins to the L pole on the Megger.5. Connect the earth pins to the earth pole on the Megger.6. Impose an electrical potential of 500-1000V on the tested
autoclave. The insulation resistance should be at least 2 MΩ. The test is successful if there was no leakage.
6.1.2 Protective Earth Impedance Test1. Connect the grounding pin of the power cord plug to one
pole of an Ohmmeter.2. Connect any other metallic part (preferable – the metallic
part of the locking screw) to the second pole of the Ohmmeter.
3. The resistance should not exceed 0.3 Ω.
After performing these tests, the Service Engineer should complete and sign the Work Order.
11
6.2 Dismantling the Outer Cover of the Autoclave.
Caution:Allow the instrument to cool before removing the outer covers.
Warning: Before starting disconnect the instrument from the power source and make sure there is no pressure in the autoclave.
Then proceed as follows:
1. Remove the screws holding the rear cover (1).
2. Remove the screws holding the cover to the base (2).
3. Pull the cover upwards.
12
6.3 Cleaning and Replacing Air Trap Jet (Located in the water reservoir)
The elimination of air pockets from the sterilization chamber during heating and sterilization phases is achieved by means of the air trap jet.This device consists of a small orifice that is obtrusive and opened by a small wire moving forth and back. The air pockets and small steam quantities are pushed up by the steam pressure and evacuated through this orifice.
Caution:
Before starting, ensure that the electric cord is disconnected and that there is no pressure in the autoclave.
1. Remove the water reservoir cover.
2. Clean the hole of the jet by manipulating the air trap wire back and forth (A).
3. In case it is necessary to replace the air trap jet, allow the instrument to cool and the pressure to drop to 0 before removing the jet.
It is important to clean the hole of the air trap, as described at point 2 before starting operation of the autoclave, for the first time.
13
6.4 Replacing the Safety Valve
CautionBefore starting, be sure that the electric cord is disconnected and that there is no pressure in the autoclave. Note: These instructions are valid for both, CE-marked and ASME type safety valves.
1. Remove the water reservoir cover.
2. Unscrew the safety valve and remove it from the safety valve base.
3. Replace it with a new safety valve (ensure the safety valve is an original one!)
4. Test all autoclave.
ASME approved Type
CE marked Type
14
6.5 Replacing the circuit breaker
Caution!Before starting, disconnect the instrument from the power source.
1. Remove the autoclave cover (see para. 6.2 “Dismantling the Outer Covers of the Autoclave”).
2. Disconnect the wires from the circuit breaker.
3. Remove the four screws connecting the circuit breaker to the panel (1).
4. Replace the circuit breaker with a new one.
5. Reconnect the electrical wires.
6. Reassemble the cover.
7. Turn on the autoclave and verify it operates correctly.
8. Move the circuit breaker’s lever to the “tripped” position and verify that the autoclave turns off.
Make sure that the correct circuit breaker is installed as marked in the table below!
1730 2340 2540 3140 3850 3870
M
Model
Voltage M MKMK-V
M MK M MK standard special
M M
1ph, 120V, 50/60 Hz
15 A 15 A 15 A — 15 A — — — — —
1ph, 230V, 50/60 Hz
10 A 10 A 10 A 15 A 10 A 15 A 10 A 15 A 15 A 15 A
Circuit breaker lever
15
6.6 Temperature Safety Thermostat(Located on the rear side of the heaters)
The autoclave is supplied with a temperature thermostat that maintains the temperature during the dry stage, by connecting and disconnecting the electric power.
This device automatically disconnects the heating elements in case of a rise in temperature. The electric power is automatically reconnected when the chamber cools down.
To replace this safety thermostat, remove the rear cover, unscrew the thermostat and replace it.
6.7 Raising the Working the Temperature of the Safety Thermostat
Caution:
Only authorized technician should do this operation!
1. Unscrew the rear cover of the autoclave.
2. With a screwdriver, turn the central screw slightly clockwise to raise the temperature.
3. Replace the rear cover.
Safety Thermostat
16
6.8 Cut-Off Thermostat This thermostat cuts out power to the autoclave, in the event that all other safety systems do not function. For example: if the operator forgets to fill the chamber with water, and starts the sterilization cycle, the chamber will heat up and activate the cut-out thermostat. In order to restart the operation, press the RESET button. If the autoclave is operated according to the instructions, and the thermostat cuts off, a replace the thermostat.
The thermostat has been calibrated by the manufacturer of the autoclave. Do not attempt to re-calibrate it
Cut-Off Thermostat
Reset Button
17
6.9 Replacing Heating Elements
Caution:Before starting, ensure that the electric cord is disconnected from the power source and there is no pressure in the autoclave chamber.
1. Dismantle the autoclave cover (see para. 6.2 “Dismantling the Outer Covers of the Autoclave”).
2. Release the two terminal wires from the heating element.
3. Remove the heater tightening bolts (1).
4. Replace the damaged heating element with a new one and reconnect the two terminal wires.
4.1 On the 3140M special model, the front heating element (3) (adjacent to the chamber's door) is with grooves (4). Assemble it with the grooves frontward.
5. Re-assemble the autoclave cover.
6. Test all the autoclave cycles.
1
2
4
3
2
1
3140 Special Model
Standard Model
18
6.10 Replacing Multi-Purpose Valve
Caution:
Before starting, make sure that the electric cord is disconnected from the power source and there is no pressure in the autoclave chamber.
1. Dismantle the autoclave cover (see para. 6.2 “Dismantling the Outer Covers of the Autoclave”).
2. Drain the water from the water reservoir.
3. Pull the valve knob out.
4. Unscrew the 3 nuts that tighten the copper tubes to the multi-purpose valve.
5. Unscrew the nut holding the valve to its base.
6. Pull out the valve.
7. Replace it with a new one. Make sure that the valve is tightened to the valve base.
8. Reconnect the three tube nuts.
9. Return the valve knob to its place.
10. Turn the valve knob to FILL position.
11. Pour water into the reservoir.
12. Check the copper tube connections for leakage.
13. Replace the cover and tighten it to the base.
Occasionally, it is necessary to take off the screws that are on the shaft of the valve in order to take the valve out of its base.
For Position of Micro-Switches see next page
After installing the new valve, the screws should be replaced according to drawing "Multi-Purpose Valve Assembly ".
19
Position of Micro-Switches and their Operation Situation
M.Sw Stage -0- Fill Ste. Exh. + Dry
M.Sw.1 Tight Tight Loose Loose
M.Sw.2 Loose Loose Loose Tight
M.Sw.3 Loose Loose Loose Tight
Notes:
1. Microswitches MSw1 - STER. and MSw. 2- DRY are actuated by the multi-purpose valve.
2. Microswitch MSw3 - Door Sw. is actuated by the door and is in pos. ON when the door is closed.
20
6.11 Unclogging the multi-Purpose Valve or Chamber
VERY IMPORTANT!When sterilizing cotton wool or pads, it is essential to wrap them in paper or cotton bags in order to prevent the multi-purpose valve and the autoclave openings from becoming clogged with remnants of the material.1. Pour distilled water into the chamber, according to quantities
mentioned in the table below:
1730 2340/2540 3140 3850 387010-12
oz300-350
ml12-15
oz350-450
ml14-16
oz420-480
ml20-23
oz600-690
ml24-27
oz720-810
ml.
2. Close the door.
3. Turn the multi-purpose valve to STE. position.
4. Turn the timer knob to 20 min.
5. Turn the thermostat (B10) knob to either 250°F or 274°F° (121°Cto 134°C).
6. Turn the main switch to START position.
7. After the timer has reached “0” turn the multi-purpose valve (clockwise) to the FILL position.
In most cases, the pressure pushes the obstructing substance out, and the steam exhausts into the water reservoir.
8. When the pressure gauge reaches 0, turn the multi-purpose valve to the OFF position, and the main switch to STOP.
9. Open the door.
10. Replace the water in the water reservoir.
The autoclave is ready for the next cycle.
11. If this procedure does not clear up the clogging, replace the multi-purpose valve.
6.12 Pressure Door Lock System
This safety device prevents the door from opening when the chamber is pressurized. The system is based on the built up pressure in the chamber that pushes the Silicon-rubber bellows and the pin into the grove of the tightening bolt. This prevents the operator from opening the door. When the steam is released, this bellow returns to its original position, drawing the pin with it thus releasing the tightening bolt.
Should there be no pressure in the chamber, and the door cannot be opened, the following procedure should be observed.
1. Turn the handle of the multi-purpose valve to EXH. & DRY.
2. The steam exhaust valve pipe is open and air inserts the chamber. In this stage the door can be opened.
21
6.13 Replacing the Door Bellows (Located in the Door Bridge)Caution:Before starting, be sure that there is no pressure in the autoclave chamber.
1. Open the door.
2. Unscrew and remove the tightening screw (5).
3. Gently pull out the door safety device locking pin (13).
4. It is possible that the washers (7, 10) will be stuck - if so, push them out by introducing pressurized air through the steam inlet hole.
5. Reconnect the door device locking pin (13) into a new silicone bellows (12).
6. Put the silicone bellows (12) and pin (13) into the bellows housing (14) and replace the washers (7, 10).
7. Reconnect all the above into the door bridge.
8. Re-screw and tighten the tightening screw (5).
9. Test all autoclave cycles.
LOK240-0026 LOK240-0025 LOK240-0023 GAS080-0020 LOK240-0024 GAS080-0006
22
6.14 Replacing the thermostat B10Caution:Before starting, be sure that the electric cord is disconnected from the power source and that there is no pressure in the autoclave chamber.
1. Remove the autoclave cover (see para. 6.2 “Dismantling the Outer Covers of the Autoclave”).
2. Unscrew the nut (1) connecting the pressure pipe (the pipe connecting the thermostat to the chamber).
3. Remove the isolating cover.
4. Remove the thermostat knob.
5. Unscrew the 2 screws connecting the thermostat to the panel (located under the thermostat knob).
6. Unscrew the nuts connecting the wires to the thermostat (2).
7. Remove the thermostat and replace it with a new one.
8. Reconnect the electrical wires.
9. Reassemble the thermostat to the panel.
10. Reassemble the knob and the pressure pipe.
11. Re-assemble the isolating and the autoclave cover.
12. Test and calibrate the pressure switch.
13. Test all the autoclave cycles and verify it operates correctly.
1
2
23
6.15 Replacement of the Door CoverCaution:Before starting, be sure that the electric cord is disconnected from the power source and that there is no pressure in the autoclave chamber.
1. Unscrew the four screws attaching the door cover and remove the door cover. Since the screw pressing the door microswitch includes two washers and a spring, be aware not to lose them.
2. Reassemble the new cover.3. When assembling covers on models 3850/3870 place a washer (7)
between the door cover (6) and the door flange (5) – on all 4 flanges (in the drawing below the washer is hidden).
4. To assemble the screw that is pressing the door switch, insert screw (1) until dimension A is approximately 15 mm.4.1 Please note that on model 3140 two washers (3) are placed
between the spring (4) and the door.5. Perform final adjustment of the screw as follows:
5.1 While the autoclave is disconnected from electricity turn on the circuit breaker.
5.2 Connect the electrical plug to a multi-meter.5.3 Press the microswitch and verify that the microswitch
functions as required.5.4 Close the door and verify that the microswitch operates.5.5 If the microswitch does not operate unscrew the screw one
turn counter- clockwise and check per para. 4.4. Repeat until microswitch operates.
5.6 Connect the autoclave to electricity.5.7 Close the door until the microswitch indicates that the door is
closed. Operate the autoclave and verify that there is no steam or pressure leak.
5.8 If there is steam leak, stop the autoclave’s operation, reduce steam pressure, open the door and turn the screw one turn clockwise and check per para. 5.7. Repeat until leakage ceases.
6. After assembling the door cover stick the model label (see spare parts list) in the sunken surface.
24
No. Description Model Cat. No. No. Description Model Cat. No.
2340, 2540 BOL191-0032 3 washer All models ELE036-0009
4 spring All models SPR177-00121 screw 1730, 3140 3850, 3870
BOL191-00911730 POL065-0001
1730 BOL191-0140 2340, 2540 POL065-0053
2340, 2540 BOL191-0033 3140 COV314-0001
3140 BOL191-0115
6 door cover
3850, 3870 POL065-00442 screw
3850, 3870 BOL191-0082 7washer(hidden)
3850, 3870 NUT193-0270
A
7
6
2
15 3 4 3
25
6.16 Replacing the Locking DeviceCaution:Before starting, verify that there is no pressure in the autoclave chamber.
1. Remove the security ring (9) using a special tool.
2. Remove pin (6).
3. Remove locking device. Take care not to lose the Teflon disk (10).
4. Reassemble the new locking device.
5. Insert the pin (6).
6. Reassemble the security ring (9).
CLOSING DEVICE
No. Description No. Description
1 Bushing 6 Door locking device pin
2Door tightening bolt assembly
7 Bakelite handle
3 Locking screw housing 8 Closing bridge “c” clip
4 Locking base 9 Cotter pin
5 Locking housing axis 10 Teflon disk
26
6.17 Replacing the Door Switch (models 2540, 3150, 3850, 3870)Caution!Before starting, disconnect the instrument from the power source and ensure that there is no pressure in the autoclave.Allow the autoclave to cool before removing outer covers.
1. Remove the autoclave cover (see para. 6.2 “Dismantling the Outer Covers of the Autoclave”).
2. Disconnect the wires (1), (2) from the door switch (3).
3. Remove the microswitch and replace it with a new one.
4. Reconnect the wires the microswitch. Verify that the wire is placed on the isolating cover (4) and does not touch the chamber.
5. Reassemble the door cover.
6. Test the connection with an ohmmeter. In “open” position the ohmmeter shows disconnection and in “close” position the ohmmeter shows connection.
3
2
1
4
27
6.18 Replacing the Drain ValveCaution!Before starting, disconnect the instrument from the power source and ensure that there is no pressure in the autoclave.Allow the autoclave to cool before removing outer covers.
1. Remove the autoclave cover (see para. 6.2 “Dismantling the Outer Covers of the Autoclave”).
2. Disconnect the drainpipe from the valve, using a 9/16” wrench.
3. Remove the nut (3) and the “ring for drain valve” (2).
4. Remove the drain valve (1) from the panel.
5. Install a new valve according to the drawing below.
6. Verify that there is no leakage.
CMT240-0020 CMT240-0003 VLV170-0066
Item Cat No.
1 GAS082-0020
2 GAS082-0021
28
Cor
rect
ion
s
1.1
Fol
low
inst
ruct
ions
in p
ara.
6.1
0.
1.2
Fol
low
inst
ruct
ions
in p
ara.
6.1
0.
2.1
Che
ck a
nd r
epla
ce it
if n
eces
sary
.
2.2
Rep
lace
“P
OW
ER
” bu
lb.
2.3
Che
ck a
nd r
epai
r th
e li
ne f
rom
the
entr
y un
til t
he m
ain
swit
ch.
2.4
Rep
lace
the
“Pow
er”
ligh
t.
2.5
Che
ck o
ut th
e un
it f
or a
n in
tern
al
elec
tric
al p
robl
em.
Pos
sibl
e ca
use
ch
eck-
up
and
test
s
1.1
Mul
ti-p
urpo
se v
alve
of
cham
ber
is c
logg
ed.
1.2
The
pip
e is
clo
gged
.
2.1
Mai
n S
wit
ch is
def
ecti
ve.
2.2
Set
tim
er f
or 1
5 m
inut
es. T
urn
the
mul
ti-
purp
ose
valv
e to
“D
RY
” po
siti
on.
If “
HE
AT
” &
“D
RY
” li
ghts
are
on
then
“P
OW
ER
” bu
lb is
bur
nt.
2.3
If a
lso
“HE
AT
” &
“D
RY
” li
ghts
are
not
on
then
the
elec
tric
al li
ne is
fau
lty.
2.4
Tur
n th
e m
ulti
-pur
pose
val
ve to
the
“Exh
/Dry
” po
siti
on a
nd s
et th
e ti
mer
to 1
5 m
inut
es.
If th
e “D
ry”
ligh
t is
on th
en th
e “P
ower
” li
ght
is b
urne
d ou
t. W
hen
fini
shed
turn
the
tim
er
back
to 0
min
utes
2.5
The
“D
ry”
ligh
t doe
s no
t com
e on
.
7T
RO
UB
LE
SH
OO
TIN
G
Sym
ptom
1.M
ulti
-pur
pose
val
ve is
inF
ILL
pos
itio
n.
Wat
er d
oes
not e
nter
into
th
e ch
ambe
r.
2.W
hile
mai
n sw
itch
is in
S
TA
RT
pos
itio
n, p
ower
is
supp
lied
, C.B
. in
ON
po
siti
on, P
OW
ER
in
dica
tor
ligh
t doe
s no
t li
ght u
p.
29
Cor
rect
ion
s
3.1.
1R
epla
ce f
ault
y bu
lb.
3.2.
1A
djus
t the
ther
mos
tat(
B10
).
3.2.
2R
epla
ce th
e fa
ulty
Mic
rosw
itch
.
3.2.
3F
ix th
e br
idge
.
3.3.
1F
ixor
rep
lace
the
faul
tyth
erm
osta
t(B
10).
3.3.
2F
ix o
r re
plac
e th
e fa
ulty
ther
mos
tat.
3.3.
3F
ix o
r re
plac
e th
e fa
ulty
tim
er.
3.3.
4F
ix o
r re
plac
e th
e fa
ulty
mic
rosw
itch
.
3.3.
5F
ixor
rep
lace
the
faul
tyth
erm
osta
t(B
10).
3.3.
6F
ix o
r re
plac
e th
e fa
ulty
doo
r sw
itch
.
4.1
Cal
ibra
te th
e th
erm
osta
t (B
10).
4.2
Pul
l saf
ety
valv
e ri
ng f
or 2
sec
onds
, the
n re
leas
e. I
f le
akag
e co
ntin
ues,
rep
lace
it.
4.3
Rep
lace
the
air
trap
jet.
4.4
Tig
hten
doo
r lo
ckin
g bo
lt. I
f le
akag
e pe
rsis
ts, r
epla
ce d
oor
seal
.
4.5
Loc
ate
leak
age
and
repa
ir f
ault
y pi
ping
co
nnec
tion
.
Pos
sibl
e ca
use
ch
eck-
up
and
test
s
3.1.
1B
urnt
bul
b.
3.2.
1T
herm
osta
t (B
10)
set t
o to
o lo
w te
mpe
ratu
re
3.2.
2M
icro
swit
ch N
o.2
is f
ault
y.
3.2.
3B
ridg
e N
o. 1
0 is
fau
lty.
3.3.
1T
herm
osta
t (B
10)
is f
ault
y.
3.3.
2S
afet
y th
erm
osta
t fau
lty.
3.3.
3T
imer
is f
ault
y.
3.3.
4M
icro
swit
ch N
o.3
is f
ault
y.
3.3.
5T
herm
osta
t (B
10)
is f
ault
y.
3.3.
6D
oor
swit
ch is
fau
lty.
4.1
The
rmos
tat (
B10
) is
not
cal
ibra
ted.
4.2
Ste
am e
scap
es f
rom
saf
ety
valv
e.
4.3
Air
trap
jet h
ole
leak
s ex
cess
ivel
y.
4.4
Ste
am e
scap
es f
rom
the
door
sea
l.
4.5
Ste
am e
scap
es f
rom
pip
ing
conn
ecti
ons.
Sym
ptom
3.M
ulti
-pur
pose
val
ve is
in
ST
E. p
osit
ion,
mai
n sw
itch
in O
N p
osit
ion.
T
imer
and
ther
mos
tat
(B10
) ar
e in
any
wor
king
po
siti
on a
nd th
e do
or is
cl
osed
tigh
tly.
3.1
“HE
AT
” li
ght i
s no
t on
but
the
auto
clav
e op
erat
es.
3.2
“HE
AT
” li
ght i
s no
t on
and
hea
ting
is
insu
ffic
ient
.
3.3
“HE
AT
” li
ght i
s no
t on
and
ther
e is
no
heat
ing.
4.M
ulti
-pur
pose
val
ve o
n S
TE
. pos
itio
n. P
OW
ER
an
d H
EA
T in
dica
tor
ligh
ts
are
lit.
and
wat
er le
vel i
s as
spe
cifi
ed
Tem
pera
ture
and
pre
ssur
e ar
e no
t suf
fici
ent.
30
Cor
rect
ion
s
5.1
Che
ck a
nd r
epla
ce h
eate
rs if
nec
essa
ry.
5.2
Che
ck if
cha
mbe
r ho
lds
the
corr
ect
amou
nt o
f w
ater
(se
e pa
ra. 6
.11)
.
5.3
Tig
hten
a b
it m
ore,
if le
akag
e co
ntin
ues,
re
plac
e ga
sket
.
5.4
Cal
ibra
te th
e th
erm
osta
t.
5.5
Doo
r be
llow
s is
leak
ing.
Rep
lace
the
bell
ows.
5.6
Act
ivat
e th
e sa
fety
val
ve (
see
"Ope
rati
on
& M
aint
enan
ce M
anua
l".I
f le
akin
gpe
rsis
ts r
epla
ce th
e S
afet
y va
lve.
6.1
Set
the
ther
mos
tat t
o th
e ri
ght
tem
pera
ture
.
6.2
Rep
lace
the
valv
e (s
ee p
ara.
6.1
0).
6.3
Rep
lace
the
safe
ty v
alve
(se
e pa
ra. 6
.4).
6.4
Rep
lace
air
jet (
see
para
. 6.3
)
7.1
Fix
or
repl
ace
the
ther
mos
tat (
B10
).
7.2
Rep
lace
the
Pre
ssur
e sa
fety
val
ve.
Pos
sibl
e ca
use
ch
eck-
up
and
test
s
5.1
One
or
mor
e he
ater
s ar
e bu
rnt.
5.2
Too
muc
h w
ater
in th
e ch
ambe
r.
5.3
Doo
r ga
sket
leak
age.
5.4
Saf
ety
ther
mos
tat i
s no
t set
to th
e ri
ght t
emp
(sto
ps h
eati
ng in
the
“inc
reas
e pr
essu
re”
stag
e).
5.5
Ste
am is
leak
ing
at th
e cl
osin
g de
vice
5.6
Saf
ety
Val
ve is
leak
ing
6.1
The
ther
mos
tat i
s no
t set
to th
e ri
ght
tem
pera
ture
.
6.2
Mul
ti-p
urpo
se v
alve
is le
akin
g. W
ater
ret
urns
to
wat
er r
eser
voir
.
6.3
Saf
ety
valv
e is
leak
ing.
6.4
Air
trap
jet l
eaks
exc
essi
vely
.
7.1
Fau
lty
ther
mos
tat (
B10
) do
es n
ot s
top
heat
ing
whe
n re
achi
ng r
equi
red
tem
p.
7.2
Pre
ssur
e sa
fety
val
ve is
fau
lty.
Sym
ptom
5.P
ress
ure
buil
ds u
p ve
ry
slow
ly.
6.T
empe
ratu
re s
afet
y de
vice
is
act
ivat
ed d
urin
g th
e st
e.
cycl
es d
ue to
ove
rhea
ting
an
d w
ater
am
ount
is
suff
icie
nt.
7.A
utoc
lave
in S
TE
. po
siti
on. P
ress
ure
safe
ty
valv
e is
act
ivat
ed a
nd
heat
ing
cont
inue
s.
31
Cor
rect
ion
s
8.1
Rep
lace
the
“Dry
” li
ght.
9.1
Sli
ghtl
y tu
rn h
andl
e in
clo
sing
dir
ecti
on
(clo
ckw
ise)
, the
n at
tem
pt to
ope
n.
9.2
If p
robl
em p
ersi
sts,
ref
er to
“P
ress
ure
Doo
r L
ock
Sys
tem
”. A
fter
ope
ning
the
door
, rep
lace
the
bell
ows.
10.1
Doo
r be
llow
s is
leak
ing.
Rep
lace
the
bell
ows.
10.2
Act
ivat
e th
e sa
fety
val
ve (
see
"Ope
rati
on
& M
aint
enan
ce M
anua
l". I
f le
akin
g pe
rsis
ts r
epla
ce th
e S
afet
y va
lve.
11.1
Rep
lace
the
mul
ti-p
urpo
se v
alve
.
12.1
Rep
lace
the
mul
ti-p
urpo
se v
alve
.
13.1
Rep
lace
the
tim
er.
14.1
Rep
lace
the
tim
er.
Pos
sibl
e ca
use
ch
eck-
up
and
test
s
8.1
The
“D
ry”
ligh
t is
burn
ed o
ut.
9.1
Doo
r pi
n se
t in
groo
ve.
9.2
Doo
r lo
ckin
g sy
stem
stu
ck o
r be
llow
s da
mag
ed.
10.1
Ste
am is
leak
ing
at th
e cl
osin
g de
vice
.
10.2
Saf
ety
Val
ve is
leak
ing.
11.1
The
inte
rnal
spr
ing
in th
e m
ulti
-pur
pose
val
ve
has
brok
en.
12.1
Poo
r m
aint
enan
ce w
ill r
esul
t in
the
mul
ti-
purp
ose
valv
e bi
ndin
g.
13.1
Inte
rnal
gea
ring
has
wor
n do
wn.
14.1
The
ham
mer
on
the
tim
er b
ell h
as b
roke
n of
f.
Sym
ptom
8.D
ry in
dica
tor
ligh
t doe
s no
t lig
ht u
p at
the
begi
nnin
g of
the
dry
cycl
e.
The
Pow
er li
ght i
s on
and
th
e un
it d
oes
heat
up.
9.D
oor
hand
le c
anno
t be
turn
ed c
ount
er c
lock
wis
e fo
r op
enin
g.
10.
Item
s in
the
cham
ber
are
burn
ing
or m
elti
ng.
11.
Mul
ti-p
urpo
se v
alve
turn
s ba
ckw
ards
.
12.
Mul
ti-p
urpo
se v
alve
doe
s no
t tur
n.
13.
Tim
er d
oes
not t
ime
dow
n.
14.
Tim
er b
ell d
oes
not r
ing.
32
GENERAL VIEW OF VESSEL, DOOR AND ACCESSORIES
Door
Closing device
Chamber
33
AUTOCLAVE COVER
34
DOOR TIGHTENING BOLT – ASSEMBLY
Cat. No.No. Description
1730, 2340, 2540 3140, 3850, 3870
1 Bushing LOK240-0003 LOK387-0003
2 Door tightening bolt assembly LOK240-0034 LOK387-0007
3 Locking screw housing LOK240-0005 LOK387-0006
4 Locking base LOK240-0030 LOK387-0012
5 Locking housing axis LOK240-0014 LOK387-0014
6 Door locking device pin LOK240-0019 LOK387-0016
7 Bakelite handle HAN071-0003 HAN071-0006
8 Closing bridge “c” clip NUT193-0339 NUT193-0300
9 Cotter pin LOK692-0039 LOK692-0039
10 Okolon disc LOK240-0017 LOK387-0017
Bushing (1) + Locking screw housing (3) + Closing bridge “c” clip (8)
LOK240-0002 LOK387-0002
Tightening bolt , Door w/o locking solenoid, assembly
LOK240-0001 LOK387-0032
35
MULTI-PURPOSE VALVE ASSEMBLY
Model 1730
Models 2340/2540/3140/3850/3870
36
8 LIST OF SPARE PARTS Cat. No.
Description1730 2340 2540 3140 3850 3870
Thermostat, Cut-Off, TY95-H, Campini
THE005-0014 THE005-0014 THE005-0014 THE005-0014 THE005-0014 THE005-0014
Thermostat, Safety, 180C, TY95/AC, Campini
THE005-0003 THE005-0003 THE005-0003 THE005-0003 THE005-0003 THE005-0003
Heating Element, 120V, 350W 1730 M/E
HEA009-0001 —— —— —— —— ——
Heating Element, 120V, 350W 2340 M/E
—— HEA009-0002 —— —— —— ——
Heating Element, 120V, 350W 2540 M/E
—— —— HEA009-0017 —— —— ——
Heating Element, 230V, 350W, 1730 M/E
HEA009-0004 —— —— —— —— ——
Heating Element, 230V, 350W, 2340 M/E
—— HEA009-0005 —— —— —— ——
Heating Element, 230V, 350W, 2540 M/E
—— —— HEA009-0006 —— —— ——
Heating Element, 120V, 450W, 1730 MK/EK
HEA010-0007 —— —— —— —— ——
Heating Element, 230V, 450W, 1730 MK/EK
HEA010-0008 —— —— —— —— ——
Heating Element, 230V, 550W, 2340 MK/EK
—— HEA010-0003 —— —— —— ——
Heating Element, 230V, 550W, 2540 MK/EK
—— —— HEA010-0004 —— —— ——
Heating Element 230V 600W 3140 M/E
—— —— —— HEA009-0014 —— ——
Heating Element 230V 800W 3140 M/E, w/o groove
—— —— —— HEA009-0015 —— ——
Sp
ecia
l Mod
el
Heating Element 230V 800W 3140 M/E, with groove
—— —— —— HEA009-0016 —— ——
Heating Element, 230V, 600W, 3850 M/E
—— —— —— —— HEA009-0007 ——
Heating Element, 230V, 500W, 3870 M/E
—— —— —— —— —— HEA009-0008
Heating Element, 240V, 350W, 1730 M/E
HEA009-0009 —— —— —— —— ——
Heating Element, 240V, 350W, 2340 M/E
—— HEA009-0010 —— —— —— ——
Heating Element, 240V, 350W, 2540 M/E
—— —— HEA009-0011 —— —— ——
Heating Element, 240V, 450W, 1730 MK/EK
HEA010-0010 —— —— —— —— ——
Heating Element, 240V, 550W, 2340 MK/EK
—— HEA010-0005 —— —— —— ——
Heating Element, 240V, 550W, 2540 MK/EK
—— —— HEA010-0006 —— —— ——
37
Cat. No.Description
1730 2340 2540 3140 3850 3870Heating Element, 240V, 800W, 3850 M/E
—— —— —— —— HEA009-0012 ——
Heating Element, 240V, 600W, 3870 M/E
—— —— —— —— —— HEA009-0013
Circuit Breaker, 1PH, 10A, Carlingswitch
ELE035-0069 ELE035-0069 ELE035-0069 ELE035-0069 — —
Circuit Breaker, Rail, 1PH, 15A, Carlingswitch
ELE035-0021 ELE035-0021 ELE035-0021 ELE035-0021 ELE035-0021 ELE035-0021
Circuit Breaker, 1-PH, 25A, Carlingswitch
— — — ELE035-0060 — —
Timer, Mechanical, 0-60 min, Faucigny
ELE033-0001 ELE033-0001 ELE033-0001 ELE033-0001 ELE033-0001 ELE033-0001
Switch, Rocker, 16A ELE035-0012 ELE035-0012 ELE035-0012 ELE035-0012 ELE035-0012 ELE035-0012
Microswitch, E13-00M, 15A, 125/250VAC, 3/4HP, Cheery
ELE036-0001 ELE036-0001 ELE036-0001 ELE036-0001 ELE036-0001 ELE036-0001
Microswitch, E11-00-H, Cheery
ELE036-0002 ELE036-0002 ELE036-0002 ELE036-0002 ELE036-0002 ELE036-0002
Lamp, Orange, 110V, 8mm
ELE038-0003 ELE038-0003 ELE038-0003 ELE038-0003 ELE038-0003 ELE038-0003
Lamp, Orange, 230V, 8mm
ELE038-0006 ELE038-0006 ELE038-0006 ELE038-0006 ELE038-0006 ELE038-0006
Lamp, Green, 110V, 8mm ELE038-0002 ELE038-0002 ELE038-0002 ELE038-0002 ELE038-0002 ELE038-0002
Lamp, Green, 230V, 8mm ELE038-0005 ELE038-0005 ELE038-0005 ELE038-0005 ELE038-0005 ELE038-0005
Gauge, Pressure, Steam, 0-60 psi, Red Pointer
— GAU029-0005 GAU029-0005 GAU029-0005 GAU029-0005 GAU029-0005
Gauge, Pressure, Steam, 0-60 psi, 1.5"
GAU029-0008 — — — — —
Handle, Door, Bakelite for TTA (522)
HAN071-0003 HAN071-0003 HAN071-0003 HAN071-0006 HAN071-0006 HAN071-0006
Knob, Timer HAN071-0011 HAN071-0011 HAN071-0011 HAN071-0011 HAN071-0011 HAN071-0011
Knob, Thermostat (B10) HAN071-0012 HAN071-0012 HAN071-0012 HAN071-0012 HAN071-0012 HAN071-0012
Cover, Door POL065-0001 POL065-0053 POL065-0053 COV314-0001 POL065-0044 POL065-0044
M — LAB048-0289 LAB048-0295 — LAB048-0250 LAB048-0249Label, Door
MK — LAB048-0290 LAB048-0296 — — —
Dipstick, Reservoir, Water, Superp.
POL067-0005 POL067-0005 POL067-0005 POL067-0005 POL067-0005 POL067-0005
Cover, Reservoir, Water, Superp.
POL067-0004 POL067-0004 POL067-0004 POL067-0004 POL067-0004 POL067-0004
Bellows, Door Lock GAS080-0020 GAS080-0020 GAS080-0020 GAS080-0020 GAS080-0020 GAS080-0020
Gasket, Door, Blue, 1730
GAS080-0197 — — — — —
Gasket, Door, Red, 2340 — GAS080-0196 — — — —
Gasket, Door, Black, 2540
— — GAS080-0184 — — —
38
Cat. No.Description
1730 2340 2540 3140 3850 3870Gasket, Door, Gray, 31XX
— — — GAS080-0195 — —
Gasket, Door, Green, 38XX
— — — — GAS080-0198 GAS080-0198
Disc, Silicone, Door Bellows
GAS080-0006 GAS080-0006 GAS080-0006 GAS080-0006 GAS080-0006 GAS080-0006
Gasket, Silicone, Water Reservoir
GAS080-0007 GAS080-0007 GAS080-0007 GAS080-0007 GAS080-0007 GAS080-0007
Cable, Plug+Socket 230V 10A, EUR
WIR040-0003 WIR040-0003 WIR040-0003 WIR040-0003 WIR040-0003 WIR040-0003
Cable, Plug+Socket 110V 15A, USA
WIR040-0004 WIR040-0004 WIR040-0004 WIR040-0004 WIR040-0004 WIR040-0004
Cable, Plug+Socket 220V 15A, USA
WIR040-0005 WIR040-0005 WIR040-0005 WIR040-0005 WIR040-0005 WIR040-0005
Multi-purpose valve assy. complete with harness
CMT173-0033 CMT240-0028 CMT240-0028 CMT314-0006 CMT385-0005 CMT387-0027
Multi-purpose valve with M.Sw.
CMT173-0026 CMT240-0046 CMT240-0046 CMT240-0046 CMT240-0046 CMT240-0046
Valve, Multipurpose, Assembly+Base
CMT173-0027 CMT240-0016 CMT240-0016 CMT240-0016 CMT240-0016 CMT240-0016
Valve, Multi-Purpose VLV170-0065 VLV170-0065 VLV170-0065 VLV170-0065 VLV170-0065 VLV170-0065
Harness, Electrical, Valve, Multipurpose
ELE032-0020 ELE032-0001 ELE032-0001 ELC314-0005 ELC385-0006 ELC387-0015
CE marked 1/4 x 2.8 Bar
SVL029-0028 SVL029-0028 SVL029-0028 SVL029-0028 SVL029-0028 SVL029-0028Safety valve ASME
1/4"-40 psiSVL029-0004 SVL029-0004 SVL029-0004 SVL029-0004 SVL029-0004 SVL029-0004
Air Jet, Red MK CMT100-0003 CMT100-0003 CMT100-0003
Air Jet, Black M CMT100-0006 CMT100-0006 CMT100-0006CMT100-0003 CMT100-0006 CMT100-0006
Socket for electric cord, 15A
WIR040-0016 WIR040-0016 WIR040-0016 — — —
Socket for electric cord, 10A
WIR040-0003 WIR040-0003 WIR040-0003 — — —
Leg, Front, TTA WHE070-0012 WHE070-0012 WHE070-0012 — — —
Leg, Front, Long, TTA — — — WHE070-0013 WHE070-0013 WHE070-0013
Leg, Rubber, Plug Type, 25x1/4
WHE070-0016 WHE070-0016 WHE070-0016 WHE070-0016 WHE070-0016 WHE070-0016
Reservoir, Water, Assembly
CMT173-0025 CMT240-0025 CMT240-0025 CMT240-0025 CMT387-0024 CMT387-0024
Cover, Outer COV173-0002 COV240-0002 COV240-0002 COV314-0002 COV385-0002 COV387-0002
Rear cover RCV173-0002 RCV240-0002 RCV240-0002 RCV314-0001 RCV387-0005 RCV387-0005
Drain valve VLV170-0066 VLV170-0066 VLV170-0066 VLV170-0066 VLV170-0066 VLV170-0066
Brass spacer for drain valve
CMT240-0003 CMT240-0003 CMT240-0003 CMT240-0003 CMT240-0003 CMT240-0003
Nut for drain valve CMT240-0020 CMT240-0020 CMT240-0020 CMT240-0020 CMT240-0020 CMT240-0020
O-Ring (drain valve) 10 x 2.5
GAS082-0020 GAS082-0020 GAS082-0020 GAS082-0020 GAS082-0020 GAS082-0020
39
Cat. No.Description
1730 2340 2540 3140 3850 3870O-Ring (drain valve) 6 x 2
GAS082-0021 GAS082-0021 GAS082-0021 GAS082-0021 GAS082-0021 GAS082-0021
Autoclave vessel ASM173-0001 ASM234-0001 ASM254-0001 CHM314-0000 ASM385-0001 ASM387-0001
Door Assembly DOR173-1000 DOR234-1000 DOR254-0000 DOR314-0000 DOR387-0001 DOR387-0001
Cooling coil PIP254-0041 PIP254-0041 PIP254-0041 PIP254-0041 PIP387-0068 PIP387-0068
Bushing for selector valve
CMT240-0019 CMT240-0019 CMT240-0019 CMT240-0019 CMT240-0019 CMT240-0019
Microswitch D48X ELE036-0012 ELE036-0012 ELE036-0012 ELE036-0012 ELE036-0012 ELE036-0012
Tightening Nut, Door Locking Bellows
LOK240-0026 LOK240-0026 LOK240-0026 LOK240-0026 LOK240-0026 LOK240-0026
Safety membrane housing
LOK240-0025 LOK240-0025 LOK240-0025 LOK240-0025 LOK240-0025 LOK240-0025
Bellows pin L0K240-0023 L0K240-0023 L0K240-0023 L0K240-0023 L0K240-0023 L0K240-0023
Bushing (1) + Locking screw housing (3) + Closing bridge “c” clip (8)
LOK240-0002 LOK240-0002 LOK240-0002 LOK387-0002 LOK387-0002 LOK387-0002
Door tightening bolt –assembly
LOK240-0001 LOK240-0001 LOK240-0001 LOK387-0032 LOK387-0032 LOK387-0032
Inner bushing for bellow CMT067-0002 CMT067-0002 CMT067-0002 CMT067-0002 CMT067-0002 CMT067-0002
Thermostat, B10, Robert Show
THE005-0002 THE005-0002 THE005-0002 THE005-0002 THE005-0002 THE005-0002
Upper CPN064-0023 CPN064-0023 CPN064-0023Control panel
LowerCPN064-0025 CPN064-0022 CPN064-0022
CPN064-0024 CPN064-0024 CPN064-0024
40
9 PRESSURE VS TEMPERATURE FOR SATURATED STEAM°CkPaBar°Fpsigpsia°CkPaBar°FInHgpsia
104.3117.91.18219.72.417.145.8100.10114.52.951.5104.4118.61.18219.92.517.254.1150.15129.34.442.2104.5118.61.19220.12.517.260.1200.20140.25.902.9104.6119.31.19220.32.617.365.0250.25149.17.393.6104.7120.01.20220.52.717.468.9300.30156.48.864.4104.8120.01.20220.62.717.472.7350.35162.910.345.1104.9120.41.20220.82.817.575.9400.40168.611.815.8105.0120.71.21221.02.817.578.8450.45173.813.306.5105.1121.31.21221.22.917.681.3500.50178.414.767.3105.2122.01.22221.43.017.7105.3122.01.22221.53.017.7°CkPaBar°Fpsigpsia105.4122.71.23221.73.117.8100.0101.31.01212.00.014.7105.5122.71.23221.93.117.8100.1101.71.02212.20.114.8105.6123.41.23222.13.217.9100.2102.11.02212.40.114.8105.7124.11.24222.33.318.0100.3102.41.02212.50.214.9105.8124.11.24222.43.318.0100.4102.81.03212.70.214.9105.9124.71.24222.63.418.1100.5103.21.03212.90.315.0106.0125.11.25222.83.518.2100.6103.61.04213.10.315.0106.1125.51.26223.03.518.2100.7104.01.04213.30.415.1106.2126.01.26223.23.618.3100.8104.31.04213.40.415.1106.3126.21.26223.33.618.3100.9104.71.05213.60.515.2106.4126.81.27223.53.718.4101.0105.11.05213.80.515.2106.5127.21.27223.73.818.5101.1105.41.05214.00.615.3106.6127.71.28223.93.818.5101.2105.81.06214.20.715.4106.7128.11.28224.13.918.6101.3106.21.06214.30.715.4106.8128.51.29224.23.918.6101.4106.61.07214.50.815.5106.9129.01.29224.44.018.7101.5106.91.07214.70.815.5107.0129.61.29224.64.118.8101.6107.31.07214.90.915.6107.1129.91.30224.84.218.9101.7107.71.08215.10.915.6107.2130.41.30225.04.218.9101.8108.11.08215.21.015.7107.3130.81.31225.14.319.0101.9108.41.08215.41.015.7107.4131.31.31225.34.319.0102.0108.81.09215.61.115.8107.5131.71.32225.54.419.1102.1109.21.09215.81.115.8107.6132.21.32225.74.519.2102.2109.61.10216.01.215.9107.7132.61.33225.94.619.3102.4110.01.10216.31.316.0107.8133.11.33226.04.619.3102.5110.71.11216.51.416.1107.9133.51.34226.24.719.4102.6111.11.11216.71.416.1108.0134.01.34226.44.719.4102.7111.51.12216.91.516.2108.1134.41.34226.64.819.5102.8111.91.12217.01.516.2108.2134.91.35226.84.919.6102.9112.31.12217.21.616.3108.3135.31.35226.94.919.6103.0112.71.13217.41.716.4108.4135.81.36227.15.019.7103.1113.11.13217.61.716.4108.5136.21.36227.35.119.8103.2113.51.14217.81.816.5108.6136.71.37227.55.119.8103.3114.01.14217.91.816.5108.7137.11.37227.75.219.9103.4114.31.14218.11.916.6108.8137.61.38227.85.219.9103.5114.71.15218.31.916.6108.9138.11.38228.05.320.0103.6115.11.15218.52.016.7109.0138.51.39228.25.420.1103.7115.61.16218.72.116.8109.1139.01.39228.45.520.2103.8116.01.16218.82.116.8109.2139.51.39228.65.620.3103.9116.31.16219.02.216.9109.3140.01.40228.75.620.3104.0116.71.17219.22.216.9109.4140.51.40228.95.720.4104.1117.11.17219.42.317.0109.5140.91.41229.15.720.4104.2117.51.18219.62.417.1
41
°CkPaBar°Fpsigpsia°CkPaBar°Fpsigpsia115.1169.71.70239.29.924.6109.6141.41.41229.35.820.5115.2170.21.70239.410.024.7109.7142.01.42229.55.920.6115.3170.81.71239.510.024.7109.8142.41.42229.65.920.6115.4171.31.71239.710.124.8109.9142.91.43229.86.020.7115.5171.81.72239.910.224.9110.0143.31.43230.06.120.8115.6172.41.72240.110.325.0110.1143.91.44230.26.220.9115.7173.11.73240.310.425.1110.2144.31.44230.46.321.0115.8173.61.74240.410.525.2110.3144.81.45230.56.321.0115.9174.11.74240.610.625.3110.4145.31.45230.76.421.1116.0174.71.75240.810.625.3110.5145.81.46230.96.421.1116.1175.31.75241.010.725.4110.6146.21.46231.16.521.2116.2175.91.76241.210.825.5110.7146.71.47231.36.621.3116.3176.41.76241.310.925.6110.8147.21.47231.46.621.3116.4177.01.77241.511.025.7110.9147.71.48231.66.721.4116.5177.61.78241.711.125.8111.0148.21.48231.86.821.5116.6178.21.78241.911.225.9111.1148.61.49232.06.921.6116.7178.71.79242.111.225.9111.2149.61.49232.27.021.7116.8179.31.79242.211.326.0111.3149.61.50232.37.021.7116.9180.01.80242.411.426.1111.4150.31.50232.57.121.8117.0180.51.80242.611.526.2111.5151.01.51232.77.221.9117.1181.11.81242.811.626.3111.6151.01.51232.97.221.9117.2181.61.82243.011.726.4111.7151.71.52233.17.322.0117.3182.21.82243.111.726.4111.8152.21.52233.27.422.1117.4182.81.83243.311.826.5111.9152.71.53233.47.422.1117.5183.41.83243.511.926.6112.0153.21.53233.67.522.2117.6184.01.84243.712.026.7112.1153.81.54233.87.622.3117.7184.51.85243.912.126.8112.2154.31.54234.07.722.4117.8185.11.85244.012.126.8112.3154.81.55234.17.722.4117.9185.71.86244.212.226.9112.4155.31.55234.37.822.5118.0186.31.86244.412.327.0112.5155.81.56234.57.922.6118.1186.91.87244.612.427.1112.6156.31.56234.78.022.7118.2187.51.88244.812.527.2112.7156.81.57234.98.122.8118.3188.21.88244.912.627.3112.8157.31.57235.08.122.8118.4188.81.89245.112.727.4112.9157.91.58235.28.222.9118.5189.41.89245.312.827.5113.0158.41.58235.48.323.0118.6190.01.90245.512.927.6113.1158.91.59235.68.423.1118.7190.61.91245.713.027.7113.2159.41.59235.88.423.1118.8191.21.91245.813.027.7113.3159.91.60235.98.523.2118.9191.81.92246.013.127.8113.4160.41.60236.18.623.3119.0192.41.92246.213.227.9113.5160.01.61236.38.723.4119.1193.01.93246.413.328.0113.6161.51.62236.58.723.4119.2193.71.94246.613.428.1113.7162.11.62236.78.823.5119.3194.31.94246.713.528.2113.8162.61.63236.88.923.6119.4194.91.95246.913.628.3113.9163.11.63237.09.023.7119.5195.51.95247.113.728.4114.0163.71.64237.29.023.7119.6196.11.96247.313.828.5114.1164.21.64237.49.123.8119.7196.71.97247.513.928.6114.2164.81.65237.69.223.9119.8197.31.97247.613.928.6114.3165.31.65237.79.324.0119.9197.91.98247.814.028.7114.4165.91.66237.99.424.1120.0198.51.99248.014.128.8114.5166.41.66238.19.424.1120.1199.21.99248.214.228.9114.6167.01.67238.39.524.2120.2199.82.00248.414.329.0114.7167.51.67238.59.624.3120.3200.52.00248.514.429.1114.8168.01.68238.69.724.4120.4201.12.01248.714.529.2114.9168.61.69238.89.724.4120.5201.82.02248.914.629.3115.0169.11.69239.09.824.5
42
°CkPaBar°Fpsigpsia°CkPaBar°Fpsigpsia125.9238.72.39258.619.934.6120.6202.42.02249.114.729.4126.0239.42.39258.820.034.7120.7203.12.03249.314.829.5126.1240.22.40259.020.134.8120.8203.72.04249.414.829.5126.2240.92.41259.220.234.9120.9204.42.04249.614.929.6126.3241.62.42259.320.335.0121.0205.02.05249.815.029.7126.4242.32.42259.520.435.1121.1205.72.06250.015.329.8126.5243.12.43259.720.635.3121.2206.32.06250.215.429.9126.6243.82.44259.920.735.4121.3207.02.07250.315.530.0126.7244.52.45260.120.835.5121.4207.62.08250.515.630.1126.8245.32.45260.220.935.6121.5208.32.08250.715.630.3126.9246.02.46260.421.035.7121.6208.92.09250.915.830.5127.0246.82.47260.621.135.8121.7209.62.10251.115.830.5127.1247.62.48260.821.235.9121.8210.22.10251.215.930.6127.2248.32.48261.021.336.0121.9210.82.11251.416.030.7127.3249.12.49261.121.436.1122.0211.52.11251.616.130.8127.4249.92.50261.321.536.2122.1212.12.12251.816.331.0127.5250.62.51261.521.836.5122.2212.82.13252.016.331.0127.6251.42.51261.721.836.5122.3213.52.13252.116.431.1127.7252.22.52261.921.936.6122.4214.22.14252.316.531.2127.8252.92.53262.022.036.7122.5214.82.15252.516.631.3127.9253.72.54262.222.136.8122.6215.22.16252.716.731.4128.0254.52.54262.422.236.9122.7216.22.16252.916.831.5128.1255.22.55262.622.337.0122.8216.92.17253.016.931.6128.2256.02.56262.822.437.1122.9217.62.18253.217.031.7128.3256.82.57262.922.537.2123.0218.32.18253.417.131.8128.4257.52.58263.122.737.4123.1218.92.19253.617.131.8128.5258.32.58263.322.837.5123.2219.62.20253.817.231.9128.6259.12.59263.522.937.6123.3220.32.20253.917.332.0128.7259.82.60263.723.037.7123.4221.02.21254.117.432.1128.8260.62.61263.823.137.8123.5221.72.22254.317.532.2128.9261.42.61264.023.237.9123.6222.42.22254.517.632.3129.0262.22.62264.223.338.0123.7223.12.23254.717.732.4129.1263.02.63264.423.438.1123.8223.72.24254.817.832.5129.2263.82.64264.623.638.3123.9224.42.24255.017.932.6129.3264.62.65264.723.738.4124.0225.12.25255.217.932.6129.4265.42.65264.923.838.5124.1225.82.26255.418.032.7129.5266.22.66265.123.938.6124.2226.52.26255.618.132.8129.6267.02.67265.324.038.7124.3227.22.27255.718.232.9129.7267.82.68265.524.138.8124.4227.92.28255.918.333.0129.8268.62.69265.624.339.0124.5228.62.29256.118.433.1129.9269.42.69265.824.439.1124.6229.32.29256.318.633.3130.0270.32.70266.024.539.2124.7230.02.30256.518.733.4130.1271.12.71266.224.639.3124.8230.72.31256.618.833.5130.2271.92.72266.424.739.4124.9231.52.31256.818.933.6130.3272.72.73266.524.839.5125.0232.22.32257.019.033.7130.4273.52.73266.725.039.7125.1232.92.33257.219.133.8130.5274.32.74266.925.139.8125.2233.62.34257.419.233.9130.6275.12.75267.125.239.9125.3234.42.34257.519.334.0130.7275.92.76267.325.340.0125.4235.12.35257.719.434.1130.8276.72.77267.425.440.1125.5235.82.36257.919.534.2130.9277.52.78267.625.640.3125.6236.52.37258.119.634.3131.0278.32.78267.825.740.4125.7237.32.37258.319.734.4131.1279.12.79268.025.840.5125.8238.02.38258.419.834.5
43
°CkPaBar°Fpsigpsia°CkPaBar°Fpsigpsia135.2315.03.15275.431.245.7131.2280.02.80268.225.940.6135.3315.93.16275.531.345.8131.3280.92.81268.326.040.7135.4316.83.17275.731.545.9131.4281.72.82268.526.240.9135.5317.73.18275.931.646.1131.5282.62.83268.726.341.0135.6318.63.19276.131.746.2131.6283.42.83268.926.441.1135.7319.53.20276.231.946.3131.7284.32.84269.126.541.2135.8320.53.20276.432.046.5131.8285.12.85269.226.741.4135.9321.43.21276.632.146.6131.9286.02.86269.426.841.5136.0322.43.22276.832.346.8132.0286.82.87269.626.941.6136.1323.33.23277.032.446.9132.1287.72.88269.827.041.7136.2324.33.24277.232.647.0132.2288.52.89270.027.141.8136.3325.23.25277.332.747.2132.3289.42.89270.127.342.0136.4326.23.26277.532.847.3132.4290.22.90270.327.442.1136.5327.13.27277.733.047.4132.5291.12.91270.527.542.2136.6328.13.28277.933.147.6132.6291.92.92270.727.642.3136.7329.03.29278.133.247.7132.7292.82.93270.927.842.5136.8330.03.30278.233.347.9132.8293.62.94271.027.942.6136.9330.93.31278.433.348.0132.9294.52.94271.228.042.7137.0331.93.32278.633.448.1133.0295.42.95271.428.142.8137.1332.83.33278.833.648.3133.1296.22.96271.628.343.0137.2333.83.34279.033.748.4133.2297.12.97271.828.443.1137.3334.73.35279.133.848.5133.3297.92.98271.928.543.2137.4335.63.36279.334.048.7133.4298.82.99272.128.643.3137.5336.63.37279.534.148.8133.5299.73.00272.328.843.5137.6337.53.38279.734.349.0133.6300.63.01272.528.943.6137.7338.53.38279.934.449.1133.7301.53.01272.729.043.7137.8339.43.39280.034.549.2133.8302.43.02272.829.243.9137.9340.43.40280.234.749.4133.9303.33.03273.029.344.0138.0341.43.41280.434.849.5134.0304.23.04273.229.444.1138.1342.43.42280.635.049.7134.1305.13.05273.429.544.2138.2343.43.43280.835.149.8134.2306.03.06273.629.744.4138.3344.43.44280.935.249.9134.3306.93.07273.729.844.5138.4345.43.45281.135.450.1134.4307.83.08273.929.944.6138.5346.43.46281.335.550.2134.5308.73.09274.130.144.8138.6347.43.47281.535.750.4134.6309.6310274.330.244.9138.7348.43.48281.735.950.6134.7310.53.10274.530.345.0138.8349.43.49281.836.050.7134.8311.43.11274.630.545.2138.9350.43.50282.036.150.8134.9312.33.12274.830.645.3139.0351.43.51282.236.351.0135.0313.23.13275.030.745.4139.1352.43.52282.436.451.1135.1314.13.14275.231.145.6
Legend:psia ─ absolute pressure in psiPsig ─ gauge pressure in psikPa ─ absolute pressure in kilo-PascalInHg ─ pressure (vacuum) in inch-Mercury
44
DRAWING OF ELECTRICAL SYSTEM OF TABLE AUTOCLAVE MODELS 1730M, MK
45
DRAWING OF ELECTRICAL SYSTEM OF TABLE AUTOCLAVE MODELS 2340/2540 M, MK
46
DRAWING OF ELECTRIC SYSTEM OF TABLE AUTOCLAVE MODEL 3140 M
47
DRAWING OF ELECTRIC SYSTEM OF TABLE AUTOCLAVE MODELS 3850 M
48
DRAWING OF ELECTRICAL SYSTEM OF TABLE AUTOCLAVE MODELS 3870 M
49
PIPING DIAGRAM TABLE TOP AUTOCLAVE MODELS: M AND MK