    SAN JOS STATE UNIVERSITY

    The Undersigned Project Committee Approves the Masters Project Titled

    MECHANICAL PROPERTIES OF CARBON HYBRID BRAIDED STRUCTURE FORLOWER LIMB PROSTHESIS

    byBhavana Shekhar

    Salma RiaziShirin Rahmanian

    APPROVED FOR THE DEPARTMENT OF GENERAL ENGINEERING

    Dr. Arthur Diaz, Department of Chemical & Materials Engineering Date

    Dr. Richard Chung, Department of Chemical & Materials Engineering Date

    Dr. Leonard Wesley, MSE Director, General Engineering Date

    iii

  • 8/2/2019 Team 10 Report Final-Version

    4/71

    ABSTRACT

    MECHANICAL PROPERTIES OF CARBON HYBRID BRAIDED STRUCTURE FORLOWER LIMB PROSTHESIS

    by Bhavana ShekharSalma Riazi

    Shirin Rahmanian

    The aim of this project was to test and evaluate different types of materials for lower limb

    prosthesis in order to choose the most appropriate material in terms of performance and

    cost. Breakey Prosthetics Inc. has provided the materials chosen for this project, which

    include Carbon-Carbon, Spectra-Carbon, and Spectra-Nylon fiber composites. The tests

    conducted on these materials consisted of Instron tensile test, hardness test, and SEM

    failure analysis. The tensile and hardness results indicate that Carbon-Carbon fibers have

    the highest tensile strength and hardness. The SEM results showed that Carbon-Carbon

    has poor bonding to the resin, while the Spectra-nylon has the best bonding. Spectra-

    Carbon had an average bonding. Economic analysis was also conducted to determine the

    viability of this project and justify its completion.

    iv

  • 8/2/2019 Team 10 Report Final-Version

    5/71

    Acknowledgements

    We would like to express our gratitude to Dr. Richard Chung, Professor, Chemical and

    Materials Engineering, SJSU, for giving us an opportunity to work with him. This projectwould not have been possible without his support.

    We would like to thank Mike Gidding and Chris Pimental of Breakey Prosthetics forsponsoring our project.

    We would like to thank Dr. Arthur Diaz, Professor, Chemical and Materials Engineering,SJSU, for his guidance.

    We would like to thank Dr. Leonard Wesley, Associate Professor, ComputerEngineering, and Dr. Micheal Jennings, Department Chair, Chemical and Materials

    Engineering, for their advice and support.

    We are thankful to Jaron Nimori, SEM lab; Neil Peters, Materials lab; and Craig,Machine shop, SJSU for their assistance.

    Lastly, we are grateful to our family and friends for their support throughout the project.

    v

  • 8/2/2019 Team 10 Report Final-Version

    6/71

    Table of Contents

    1.0 Introduction ..................................................................................................................1

    2.0 Literature Review ........................................................................................................32.1 Introduction to Literature Review ..............................................................................32.2 Background ................................................................................................................32.3 Pre-amputation and Post-amputation Procedures ......................................................42.4 Components of a Below Knee Prosthesis ..................................................................72.5 Manufacturing Process of Prosthetic Limbs ............................................................102.6 Materials ..................................................................................................................12

    2.6.1Historical Development of Composites for Orthopedics ...................................172.6.2 Properties of Spectra .........................................................................................18

    2.7 Experimental Methods .............................................................................................19

    3.0 Materials and Methods ..............................................................................................203.1 Tensile Testing .........................................................................................................203.2 Hardness Test ...........................................................................................................273.3 SEM Analysis ..........................................................................................................303.4 Discussion of Results ...............................................................................................35

    4.0 Economic Justification...............................................................................................37

    4.1 Executive Summary .................................................................................................374.2 Problem Statement ...................................................................................................384.3 Solution and Value Proposition ...............................................................................384.4 Market size ...............................................................................................................394.5 Competitors ..............................................................................................................404.6 Customers ................................................................................................................414.7 Cost ..........................................................................................................................42

    4.7.1 Fixed Cost .........................................................................................................434.7.2 Variable Cost ....................................................................................................45

    4.8 Price Point ................................................................................................................464.9 SWOT Assessment ..................................................................................................464.10 Investment Capital Requirements ..........................................................................47

    4.10.1 Norden-Rayleigh Financial Profile .................................................................504.11 Personnel ................................................................................................................534.12 Business and Revenue Model ................................................................................554.13Strategic Alliances and Partners .............................................................................56

    4.14 Exit Strategy...........................................................................................................56

    5.0 Project Schedule .........................................................................................................58

    6.0 Conclusion ..................................................................................................................60

    7.0 References.....................................................................................................................61

    vi

  • 8/2/2019 Team 10 Report Final-Version

    7/71

    List of Figures

    Figure 1: Components of transtibial prosthesis

    Figure 2: Sample specifications

    Figure 3: Stress-Strain curve of Carbon-Carbon sample 1

    Figure 4: Stress-Strain curve of Carbon-Carbon sample 2

    Figure 5: Stress-Strain curve of Carbon-Carbon sample 3

    Figure 6: Stress-Strain curve of Spectra-Carbon sample 1

    Figure 7: Stress-Strain curve of Spectra-Carbon sample 2

    Figure 8: Stress-Strain curve of Spectra-Carbon sample 3

    Figure 9: Stress-Strain curve of Spectra-nylon sample 1

    Figure 10: Stress-Strain curve of Spectra-nylon sample 2

    Figure 11: Stress-Strain curve of Spectra-nylon sample 3

    Figure 12: Box plot for Hardness test Data

    Figure 13: SEM images of Carbon samples. (a) Crack area at 50x magnification (b) at

    400x magnification (c) at 6000x magnification

    Figure 14: SEM images of Spectra-Carbon samples. (a) Crack area at 6000x

    magnification (b) at 2400x magnification (c) and (d) at 10000x magnification

    Figure 15: SEM images of Spectra-Nylon samples at 3000x magnification (a) top surface

    (b) fiber pull-out

    Figure 16: Percent concentration of limb prosthesis companies in the U.S.

    Figure 17: Profit and Loss chart

    Figure 18: Break-even chart

    vii

  • 8/2/2019 Team 10 Report Final-Version

    8/71

    viii

    Figure 19: Cumulative Distribution Function for Norden Rayleigh

    Figure 20: Probability Density Function curve for Norden-Rayleigh

    Figure 21: ROI Chart

  • 8/2/2019 Team 10 Report Final-Version

    9/71

    List of Tables

    Table 1: Mechanical properties of some materials used for prosthesis fabrication

    Table 2: Comparison of Mechanical Properties of different fibers

    Table 3: Mechanical Properties of Carbon-Carbon

    Table 4: Mechanical Properties of Spectra-Carbon

    Table 5: Mechanical Properties of Spectra-Nylon

    Table 6: Hardness test results for Spectra-Nylon, Spectra-Carbon and Carbon composites

    Table 7: Major Competitors

    Table 8: Fixed Cost Break Down

    Table 9: The Variable Cost Break Down

    Table 10: Price Point Analysis

    Table 11: SWOT analysis

    Table 12: Expected Profit and Loss

    Table 13: Funding Break Down

    Table 14: Cost Drivers

    Table 15: Probability density function and Cumulative distribution function for Norden-

    Rayleigh

    Table 16: Return on Investment for Prosthetics Labs

  • 8/2/2019 Team 10 Report Final-Version

    10/71

    1.0INTRODUCTIONThere are about 1.7 million people in the United States alone who have lost at least

    one limb. Need to be fitted with a prosthetic limb to enable them to carry out normal

    daily activities. Currently, only one type of fiber (for e.g. Carbon fiber) is used in the

    fabrication of below knee prosthesis for all different kinds of patients. Athletes for

    instance would require stronger sockets than would an average person. Hence different

    socket materials can be chosen based on usage.

    Although the first prosthetic leg was a wood stump, we have come a long way since

    then. Some of the materials used for the fabrication of sockets include: Carbon fibers,

    Kevlar fibers, glass fibers, and thermoplastic polymers. We have tested three different

    composite materials that are used to make the socket of below-knee prosthesis. The

    materials tested are braided carbon fiber, spectra-carbon, and spectra-nylon composites

    which were provided by Breakey Prosthetics Inc., San Jose, CA. To characterize and

    compare the mechanical properties of these materials, tensile test, hardness test, and

    failure analysis were used. The samples were then cut into dog-bone shapes in the SJSU

    machine shop to conduct the tests. All of the tests were conducted in SJSU Materials

    Engineering laboratories. For the tensile testing, three samples of each

    material were tested using the Instron Machine. The Rockwell hardness machine wasused to measure the hardness of the samples. Twenty test values for each composite

    were obtained. Finally, for failure analysis, a Scanning Electron Microscope was used to

    view the images of the failed samples to discover the cause of failure in each composite.

    1

  • 8/2/2019 Team 10 Report Final-Version

    11/71

    To further justify the completion of this project, it is important to economically

    evaluate it. The work done in this project can be viewed as a service provided by a small

    start-up company offered to limb prosthetic companies to test their materials and also

    give them consultations. Prosthetic Labs, Inc., is a Limited Liability Company (LLC)

    which provides testing and consulting services specific to prosthetic limb manufacturers.

    For tax purposes, Prosthetic Labs files as a sole proprietorship. Prosthetic Labs Inc. will

    initially consist of about 9 employees, including technicians to the test, expert analysts,

    marketing consultant, and the CEO. What sets this company apart from other material

    testing companies is that this company will be the only company that offer tests and

    consultation exclusive to limb prosthetic companies and meet tailored needs. The

    economical analysis conducted has shown that the market size is large enough for starting

    such service and considerable profit can be made in a short period with a relatively low

    budget.

    2

  • 8/2/2019 Team 10 Report Final-Version

    12/71

    2.0 LITERATURE REVIEW

    2.1 Introduction to Literature Review

    Prosthetics allow people with limb amputation to resume normal daily activities.

    Developments in limb prosthesis have significantly improved over the last few years that

    even in patients being able to participate in extreme sports. For instance, in the 2000

    Paralympic Games, Sydney, a new world record was set for the 100m sprint for below-

    knee amputees at 11.09 seconds (Gutfleisch, 2003). The 2008 record for fit and healthy

    athletes was 9.69 seconds. This astounding result was due to the superior performance of

    lower limb prosthesis along with the determination and talent of the athlete. This chapter

    includes a literature review of materials used in lower limb prosthesis.

    2.2 Background

    The word prosthesis comes from the Greek word prostithenai which means to

    add to. A prosthetic device is an artificial device which mimics the function of a missing

    body part. The use of Peg legs, carved out of wood, as early as 3000-1800 B.C. has

    been documented in Indian literature (Cochrane, Orsi, & P., 2001). James Potts

    constructed The Anglesea leg in 1800. The socket and shank were made of wood; the

    knee joint was made of steel, and it had an articulated foot. The shank was attached to a

    leather thigh corset with the help of metal hinges and side bars (Bannister, 1978). Even in

    the 1940s sockets for below knee prosthesis was made out of either blocks of wood or

    leather (Verrall, 1940). The basic design of the prosthetic limb remained unchanged till

    1950.

    3

  • 8/2/2019 Team 10 Report Final-Version

    13/71

    Research on improving the design of the lower limb prosthesis was going on at the

    University of California, Berkeley; and in 1958, they made their design public. It was

    called Patellar Tendon Bearing (PTB) prosthesis. The socket was made of plastic and

    lined with rubber and soft leather. The hollow wooden shank was reinforced with a

    plastic laminate, and the articulated foot was replaced with a rubber solid ankle cushion

    heel (SACH) foot. Leather straps were used above the patella to provide suspension.

    Though it had a huge advantage of being much lighter than the older version and fit

    properly to the residual limb; the drawback limited knee flexion, skin abrasions, and

    dermatitis. In 1964, the PTB prosthesis was further modified such that the sides of the

    socket extended beyond the femoral condyles. The liner was built-in with one or two

    wedges to provide suspension. This design was called the PTS (patellar tendon

    supracondylar) prosthesis (Bannister, 1978). The Botta prosthesis was an adaptation of

    the PTS prosthesis. The socket was light in weight and was fabricated with polyester or

    other synthetic resin strengthened with carbon fiber (Marquardt & J., 1984).

    2.3 Pre-Amputation and Post Amputation Procedures

    Some amputations of the lower limb, which are due to some diseases, can be

    predicted. In this preamputation phase, before the surgery, the patients will attend certain

    meetings to learn about how this surgery will affect their lives and how to deal with it.

    They will also get emotional support by meeting with an amputee in the same situation as

    they are. Providing information to individuals might also prevent some traumatic and

    nontraumatic amputations. For example, an amputation care facility program held by the

    4

  • 8/2/2019 Team 10 Report Final-Version

    14/71

    Department of Veterans Affairs, has decreased the number of non-traumatic amputations

    by forty percent each year (Pasquina, Bryant, et al. 2006).

    Different surgical techniques must be used for each amputee. Using a general

    technique will not work for all individuals. In order to attach the remaining muscles two

    different techniques are used which are called myodesis, and myoplasty. In myoplasty the

    opposite muscles of the amputated limb are sewn to each other, whereas in myodesis the

    muscles are attached to the periosteum of the amputated bone (Pasquina, Bryant, et al.

    2006).

    After the surgery is done, there are certain issues to have in mind and precautions to

    take. The amputated part has to be taken care of properly, in order to be able to attach the

    prosthetic to it without encountering any problems in the future. The cut limb has to

    undergo a skin desensitization program which involves massaging the severed section of

    the limb. This is done to heal the scar and not let it adhere to the bone underneath. Also,

    edema should be prevented by using a residual limb stump. The cut end of the limb is

    then introduced to increasing amounts of pressure so that it adapts to the forces that it will

    undergo after the prosthetic is placed and utilized (Kelly, Spires and Restrepo, 2007).

    A study done by Moore, Hall et al., has proved the advantages of immediate

    postoperative techniques which has been done about 30 days after the amputation, in

    comparison to a later post operative action. The advantages include a quicker

    rehabilitation, lower number in mortality and healing rate. Therefore, the time from the

    5

  • 8/2/2019 Team 10 Report Final-Version

    15/71

    amputation to the fitting of the prosthesis is critical and can affect the procedure directly

    (Moore, Hall, et al. 1972).

    For a prosthetic surgery to be successful and make it as comfortable as possible to

    the recipient of the prosthetic limb many factors must be considered. Some of these

    include proper diagnosis, to begin with, the patients weight and other physical

    characteristics, and also their activity level (Kelly, Spires and Restrepo, 2007).

    It should also be noted that for children extra consideration must be taken because

    they are constantly growing. This is why adjustments to the prosthesis must be made

    more regularly than those of adults (Pasquina, Bryant, et al. 2006). On average the life

    expectancy of the prosthetic devices is approximately three to five years depending on

    the patients physical activity, and the materials with which the prosthetics are made of.

    Also, the location where the socket of the prosthesis and the limb are in contact with each

    other is of great importance. The socket must not cause any damage, irritation, or allergic

    reactions to the skin, since the skin at the severed end of the limb is very sensitive. The

    most important issue which has to be considered is the amputated tissues response to the

    applied load, after the prosthesis has been placed. Depending on the individual and their

    body type the tissue will get more susceptible to pain after the amputation (Kelly, Spires

    and Restrepo, 2007). Other than pain, some temperature changes might be observed in

    the area which load is more. When the tissue undergoes an amount of load, due to

    decrease in blood circulation, the temperature of the tissue will drop. After unloading the

    tissue however, the temperature will rise (Mak, Zhang & Boone, 2001). The skin in the

    contacting region of the prosthesis and the tissue might get abraded as a result of rubbing

    6

  • 8/2/2019 Team 10 Report Final-Version

    16/71

    against the device. This might lead to some skin problems, such as skin thickening and

    blisters (Mak, Zhang & Boone, 2001).

    2.4 Components of a Below Knee Prosthesis

    Socket

    Socket Adapter

    Tube Clamp AdapterPylon

    Endoskeletal finish

    Ankle

    Foot

    Figure 1. Components of transtibial prosthesisSource: Below Knee (Transtibial) Prosthesis. (n.d.). Retrieved March 12, 2009, from ProstheticConcepts Web site: http://www.prostheticconcepts.ca/belowknee.pdf

    The components of a transtibial prosthesis are shown in Figure 1. The main components

    of the prosthesis include:

    1. The Socket connects the amputated limb to the prosthesis. It transfers the bodyweight to the prosthesis and may contain liners that act like padding. It protects the

    amputated limb and enables the user to stand, walk, and move freely. Mostly in

    transtibial amputation, patellar-tendon bearing socket (PTB) is used (Carroll, 2006,

    7

  • 8/2/2019 Team 10 Report Final-Version

    17/71

    Kelly et al. 2007). Before this socket was introduced, some open-ended plug-fit

    sockets were used which had some disadvantages such as skin irritation and chronic

    choke syndromes (Kelly, Spires and Restrepo, 2007).

    However, the socket fit and comfort depends on the individual and has to be chosen

    by them and the help of the prosthetist. Some common sockets are: PTB, Total-

    surface bearing (TSB), PTB-supracondylar, and joint and corset (Kelly, Spires and

    Restrepo, 2007). All these sockets have their own pros and cons due to their

    mechanism of load bearing, pressure transmission to the amputated area, and comfort.

    As technology advances, recent and more applicable materials are being introduced

    for this purpose. Sockets are now being manufactured with computer-aided designs

    and are custom fitted for the patients. Some carbon graphite sockets that are now

    being used are light weight and last longer. The linings of these sockets are also more

    flexible and comfortable. By being custom made, the socket manufacturing will not

    only be cheaper, it will also save time and accelerate the delivery time to patients

    (Carroll, 2006).

    2. The Anklejoins the foot and prosthesis. This part is made of some joints which allowaxial rotation. This part should also be capable of energy storing and absorption

    (Kelly, Spires and Restrepo, 2007). The ankle is made from different material

    depending on its use and the socket its connecting to. There are ankle joints made of

    wood with a laminated outer shell. Some other ankle joints are made out of metal,

    plastic or carbon fiber. There are different types of prosthetic ankles with different

    8

  • 8/2/2019 Team 10 Report Final-Version

    18/71

    movement axes. A study done by Rita, Mason, et al. has shown that these ankles

    show different results in shock absorption. Although it has to be taken into

    consideration that stride length, age, weight and some other factors are involved as

    well (Witra, Mason, et al. 1991).

    It is common for this part to be available together with the socket; however, it is also

    available separately for use in sports and heavy physical activity (Kelly, Spires and

    Restrepo, 2007).

    3. The Foot acts as a base support. Not only this part must be able to bear all the weightacted on it, it must also transfer the body weight to ground, act as a shock absorber,

    and replace lost muscle function and biomechanics of the foot ( Kelly, Spires and

    Restrepo, 2007). It must also provide cosmetic appearance and should fit in normal

    shoes.The recent artificial feet are now being made to reproduce a healthy foots

    function. They have energy absorption mechanism in multiple planes, as well as

    being able to absorb vertical and torsional shocks which are acted upon them ( Kelly,

    Spires and Restrepo, 2007).

    4. Suspension systems are used to hold the prosthesis on the body (Below KneeProsthesis). The socket must have some suspension mechanism in order to not to fall

    off. Different mechanisms are used for this purpose such as suction, harness, belt, and

    gel suspension. A combination of them is also available (Moore, Hall, & Lim, 1972).

    By using a gel elastomer, the gel suspension system provides cushioning to the

    residual limb. It also provides an acceptable cosmesis to the amputee. Using a thin gel

    can improve the sense of propioception in the patients, allowing them to function

    9

  • 8/2/2019 Team 10 Report Final-Version

    19/71

    better in their activities (Carroll, 2006). Another common suspension system is using

    sleeves. Sleeves have their own drawbacks, since the material can get punctured

    easily; it has to be replaced quite often. They can also have ventilation problems as

    well as getting hot in warm and humid regions (Carroll, 2006). This problem

    however, has recently been solved by putting a valve on the sleeve to provide air

    transfer.

    The transtibial prosthesis is also comprised of some other components which are

    mainly used for connecting the major parts to each other. They consist of a socket

    adaptor which connects the socket to the rest of the prosthesis and aligns the

    prosthesis; a tube clamp adaptor, which is used to connect the pipe to the socket; a

    pylon, which is used to transfer the body weight and should be adjusted to achieve

    proper height of the prosthetic; and finally an endoskeletal finish which envelopes the

    whole prosthesis and hence protects the internal parts from dust, dirt, and moisture

    (Osbourne, 2009).

    2.5 Manufacturing Process of Prosthetic Limbs

    After amputation, a medical doctor has to prescribe prosthesis to the patient. This

    device cannot be bought in stores and is not mass produced. Following the prescription,

    the patient must consult with the prosthetist and physical therapist in order to choose the

    best prosthesis for his use. Some parts of the prosthesis are manufactured in factories, and

    some other parts like the socket, can be custom made for each patient.

    10

  • 8/2/2019 Team 10 Report Final-Version

    20/71

    The procedure starts with the prosthetist studying the amputees residual limb, taking

    an impression of it and measuring some body segments of the patient. The impression

    and the measurements are used to make a plaster cast of the stump, which is then used to

    make the stump itself. Afterwards, a clear sheet of heated thermoplastic is placed onto the

    mold and put into a vacuum chamber. When the air is taken out of the chamber the

    thermoplastic sheet smoothly presses against the mold taking its shape without having

    any air bubbles trapped between the sheet and the mold. The product of this process is the

    test socket. The reason why it is clear is so that it will be easier for the prosthetist to

    check the fit of the prosthesis.

    The penultimate process involves the prosthetist to check and confirm that the test

    socket that was just constructed properly fits the patient. After the patient puts it on he or

    she walks while the prosthetist analyzes the gait. Adjustments are made both in response

    to the appearance of the gait and the comfort of the patient. Only after this is done and

    both patient and prosthetist are satisfied do they move on to the final procedure. Finally,

    the permanent socket is formed usually using polypropylene, again, using vacuum

    forming as the production or shaping process. Over time, if any anatomical changes occur

    to the appendage changes to the prosthesis are made accordingly.

    The manufacture of prosthetic limbs involves a wide variety of materials and

    methods. Many are made using different types of plastics which are formed using

    vacuum-forming, extruding, or injection molding. Those that are made of metallic

    11

  • 8/2/2019 Team 10 Report Final-Version

    21/71

    components usually use titanium and aluminum parts which are die cast and then finished

    by drilling, sanding, polishing, and other finalizing processes.

    After the different parts and components that make up the prosthesis are

    manufactured the assembly is done by the prosthetist technician using various tools such

    as wrenches, screwdrivers and other such hand or power tools. The patient is fitted with

    the assembled by the prosthetist. (Stacey, Blachford, & Cengage, 2002)

    2.6 Materials

    Metal, leather, and wood were used as prosthetic materials before mid 19th century.

    Lower limb prostheses made before 1984 were rigid and were manufactured from metal,

    leather, or plastic laminate with a foam toe filler (Lange, 1992). Although these devices

    did not restore function lost by amputation, they maintained rollover in the toe section.

    Since these devices were rigid, a lot of effort was put in to make partially rigid devices

    which integrated foam toe fillers with clear and flesh-tone plastics that were flexible.

    Cosmesis was an issue although the function and fit was satisfactory.

    During 1984-1986, thermoplastics were used to fabricate custom prosthetic sockets.

    Flexible Surlyn below-knee (BK) sockets which were supported within semi-rigid

    polypropylene socket retainers were made at the University of Virginia Medical Center

    (Schuch, 1991). Polyethylene was used to make flexible sockets, and polypropylene was

    used for making socket retainers. The advantages of prosthetic sockets made of

    thermoplastics were flexibility, light weight, quick and simple fabrication (Schuch,

    12

  • 8/2/2019 Team 10 Report Final-Version

    22/71

    1991). The drawback of this material was that it was not durable, i.e., it would split or

    tear. Also, thermoplastics shrink and this would loosen the fit.

    Shrinkage of the material was found to be greatest, if it was stretched at the time of

    fabrication. Hence, it must be carefully draped and vacuum-formed rapidly to optimize

    the results (Rothschild, Fox, Michael, Rothschild, & Playfair, 1991).

    Next, laminated silicone sockets were attached to hollow foot shells with silicone

    sealant (Lange, 1992). To improve cosmesis, another layer of lamination was put above

    the socket and the foot shell to join them permanently and a zipper was used at the back

    of the socket (Lange, 1992).

    Langes silicone partial foot prosthesis had two laminations and incorporated the

    foot shell between the two. The result was that the foot shell was permanently bound to

    the socket thereby producing an elastic, resistive toe lever. The hollow toe in the foot

    shell was filled with room temperature vulcanized (RTV) foam to attain more natural

    ankle motion (Lange, 1992). This design also had a zipper closure behind the socket.

    Instead of adding pigments to the silicone, a nylon stockinette matching the skin tone of

    the patient was coated on the foot shell to provide better cosmesis (Lange, 1992).

    Fiber reinforced plastics are laminated composites made by applying resin to one or

    more layers of fibers. The material properties of the resin and the fiber, the extent of

    bonding between the two and the resulting structural architecture determines the

    strength of the laminate (Phillips & Craelius, 2005). The tensile and flexural strengths of

    the fiber reinforced plastics were found to be high along the fiber axis. From Table 1, it

    13

  • 8/2/2019 Team 10 Report Final-Version

    23/71

    can be noted that the Youngs modulus of nylon fiber is slightly lesser than that of the

    SPT (SPT Technology, Inc., Minneapolis) resin.

    Although the nylon fibers are supposed to increase the strength since they are the

    reinforcing materials, they do not do so because of their low modulus of elasticity when

    compared to that of the resin (Phillips & Craelius, 2005). Hence it is important to choose

    a fiber whose modulus of elasticity is greater than that of the resin else the material

    properties will be completely controlled by the resin.

    Table 1Mechanical properties of some materials used for prosthesis fabrication

    Source: Phillips, S. L., & Craelius, W. (2005). Material Properties of Selected ProstheticLaminates.Journal of Prosthetics & Orthotics , 27-32.

    While glass and carbon fibers have high strengths, the main drawback is that they

    are brittle; during post-fabrication modification, instead of stretching, they break (Phillips

    & Craelius, 2005).

    A strut was designed by Madden using Kevlar-49, carbon, and S-2 glass fibers.

    Carbon fiber was used to make the inner core, Kevlar-49 was used to fabricate the outer

    layer, and S-2 glass was used in the middle layer. The S-2 glass acts as a transition layer

    for the stress changes which is an intrinsic to the strut design. Carbon fibers and Kevlar-

    14

  • 8/2/2019 Team 10 Report Final-Version

    24/71

    49 are comparatively weak under tensile and compressive loads respectively, S-2 glass

    acts as a balance between the two, facilitating stress transfer between the layers. This

    design has been used successfully for fabricating rigid frames for flexible socket

    systems in prosthetics (Madden, 1991).

    As discussed previously the initial material made of wood and metals for artificial

    legs had major drawbacks, since they were limited by their weight, and had poor

    durability to corrosion and moisture induced swelling (Ramakrishna et al., 2001). These

    limitations resulted in restricting the user to slow and non-strenuous activities due to poor

    elastic response during stance (Ramakrishna et al., 2001).

    Due to these limitations, polymer composites were introduced for material of choice

    for limb systems, because of their lightweight, corrosion resistance, fatigue resistance,

    aesthetics, and ease of fabrication (Ramakrishna et al., 2001). Polymer composites can

    be either thermosetting or thermoplastics composites that are reinforced with glass,

    carbon, or Kevlar fibers (Ramakrishna et al., 2001).

    Thermoplastic polymers have the advantage of having strong intermolecular bonds

    that result in good biocompatibility and resistance to moisture damage (Evans, &

    Gregson, 1998). Polyetheretherketone (PEEK), polyaryletherketone (PAEK),

    polyethylene, and polysulfone have been widely used for orthopedic use (Evans, &

    Gregson, 1998). While these materials have excellent biocompatibility and good

    durability in the physiological environment, they are difficult in fabrication of long fiber-

    reinforced composites, and thus difficult for prosthetics with sufficient strength for

    highly loaded applications (Evans, & Gregson, 1998).

    15

  • 8/2/2019 Team 10 Report Final-Version

    25/71

    Thermosetting polymers such as epoxy resins have not been so common in

    orthopedics because of variable biocompatibility and durability characteristics (Evans, &

    Gregson, 1998). However studies have shown that properly processed and selected epoxy

    resins can have excellent biocompatibility, and these materials can have much better

    processing characteristics than the thermoplastics, allowing the fabrication of more

    sophisticated composite structures (Evans, & Gregson, 1998).

    During the past 10 years the most notable reinforcing materials for orthopedic use

    have been carbon fibers (Evans, & Gregson, 1998). Especially for lower limb prosthesis

    carbon composite lay-ups are very popular (Strike & Hillery, 2000). These composites

    are chosen or their flexibility and energy storage and release properties (Strike & Hillery,

    2000). The fibers can be fabricated in different ways such as being braided, woven,

    knitted, or laminated. According to a lower limb design by Strike & Hillery, the

    lamination would allow them to have specific tensile strength and stiffness by changing

    the resin and controlling the angles between successive layers (Strike & Hillery, 2000).

    There are also other reinforcing materials that have been used for prosthetic use.

    One of the most recent ones are Aramid fibers such as Kevlar, which have excellent

    tensile properties (Evans, & Gregson, 1998). However, these fibers have poor

    compressive strength and stiffness and thus have limitation during bending loads (Evans,

    & Gregson, 1998).

    Morever, according to Ramakrishna et al., the sockets currently in the market, are

    made using a combination of knitted or braided carbon or glass fiber fabrics and water-

    curable (water-activated) resins (Ramakrishna et al., 2001). The braided fabric

    16

  • 8/2/2019 Team 10 Report Final-Version

    26/71

    reinforced sockets have the advantage of being stiff and strong, whereas the knitted fabric

    reinforced sockets have the advantage of being flexible and more conformable to the

    patient's stump (Ramakrishna et al., 2001).

    2.6.1 Historical Development of Composites for Orthopedics

    While composites have been used throughout history, the emergence of composite

    materials as we know today is fairly recent, becoming most popular during the early

    1950s due to extensive research done in aerospace industry. Development of fiberglass

    in the 1930s has led to using these materials as reinforcement for polymers, and thus

    improving composite technology (Erwin, 1947). As it has previously referred to, polymer

    composites can be reinforced with glass, carbon, or Kevlar fibers (Ramakrishna et al.,

    2001).

    Major breakthrough in the composite development was the use of reinforcing fibers.

    In 1961, the first carbon fiber was produced by Shindo et al. (Shindo, 1969). Carbon

    fibers have the advantage of having a low density, high elastic modulus, and high tensile

    strength. Also, due to their high specific strength and good fatigue resistance (when used

    as a reinforcing polymer) makes them a good candidate for orthopedic use (Pigott &

    Harris, 1980). The major drawbacks of carbon reinforced thermosetting composites are

    their brittleness yielding low fracture toughness and poor impact resistance.

    The next major emergence of reinforcing fibers in composite development was the

    production of Kevlar by DuPont in 1971. Thermoset resins that are reinforced by Kevlar

    fibers display high fracture toughness and good resistance to tensile loading (Pigott &

    17

  • 8/2/2019 Team 10 Report Final-Version

    27/71

    Harris, 1980). However they have the disadvantage of being weak under compression and

    they are also hard to fabricate because they are difficult to shape, cut, and saturate with

    resins (Berry, 1987)

    2.6.2 Properties of Spectra

    Spectra fiber is a polyethylene fiber that is produced from a gel-spinning process

    by Honeywell, Inc. (Honeywell). Spectra Fibers are available in three different series:

    Spectra 900, Spectra 1000, and Spectra 2000. Spectra fibers have tensile strength that is

    higher than aramid fibers at temperatures below ~ 1000

    C and above this temperature their

    tensile strength will decrease (Lewin, Sello, & Preston, 1996). Spectra fibers can

    withstand twisting without losing their strength. These fibers also have very good

    abrasion resistance, can creep well under static load, and have a good impact resistance,

    when compared to aramids. The mechanical properties of Spectra fibers are shown in

    Table 2 and are compared to other composite materials (Lewin, Sello, & Preston, 1996).

    Table 2Comparison of Mechanical Properties of different fibers

    Source: Lewin, M., Sello, S. B., & Preston, J. (1996).Handbook of fiber science andtechnology (Vol. III). New York: Marcel Dekker, Inc.

    18

  • 8/2/2019 Team 10 Report Final-Version

    28/71

    2.7 Experimental Methods

    Samples of braided carbon fiber, spectra-nylon, and spectra-carbon have been

    provided by Breakey Prosthetics. At least three tensile (dog bone shape) samples of these

    fiber composites will be made for mechanical testing. The dimension of the samples will

    be as shown in Figure 2. The samples will be tested for their tensile strength, hardness

    and bending. Fractured samples will be analyzed using a Scanning Electron Microscope.

    The results will be compared to the values required by ISO standards to see if they meet

    the minimum requirements.

    Figure 2. Sample specificationsSource: (2008). Quality and testing. In B. A. Strong, Fundamentals of CompositesManufacturing: Materials, Methods and Application (p. 277). SME.

    19

  • 8/2/2019 Team 10 Report Final-Version

    29/71

    3.0 MATERIALS &METHODS

    3.1 Tensile Testing

    Tensile test is a mechanical test in which a machine is used to deform a specimen

    under gradually increasing tension load. Tension test can be used to plot a stress-strain

    curve and several mechanical properties can be obtained from this curve. Some of the

    most important mechanical properties that can be obtained from the stress-strain curve

    include yield strength, modulus of elasticity (Youngs modulus), ultimate tensile strength,

    ductility, and toughness.

    The yield strength measure the stress level at which a material starts to plastically

    deform. This means that up to yield strength, the material can return to its original length

    after deformation. Beyond yield strength, the material will undergo plastic deformation,

    in which the deformation is permanent. During elastic deformation (before reaching

    yield strength), stress and strain are proportional to each other. This relationship is known

    as Hooks law and the constant of proportionality is called Youngs modulus, or modulus

    of elasticity. The modulus can be obtained by measuring the slope of linear portion of a

    stress-strain curve. The greater the modulus, the more brittle the material is, meaning it

    will go under less strain before yielding.

    Another important mechanical property is the tensile strength (UTS), which is the

    maximum strength in the stress-strain curve. UTS corresponds to the strain that a material

    20

  • 8/2/2019 Team 10 Report Final-Version

    30/71

    can sustain under tension. If such stress is applied and maintained, the material will start

    to non-uniformly deform and fracture will occur.

    Ductility and toughness are another mechanical properties that can be obtained

    from the stress-strain curve. Ductility is a measure of the amount of plastic deformation

    at fracture. It can be expressed as percent elongation or percent of reduction in area the

    fracture. Toughness is a measure of the ability of the material to sustain energy up to the

    fracture point. Toughness can be measured by calculating the entire area under the stress-

    strain curve.

    Tensile test was conducted on the three different types of composite samples

    under study, which were carbon-carbon, Spectra-carbon, and Spectra-nylon. The tensile

    done test was done using the Instron machine on dog-bone shaped sample. For each

    material several tensile tests were conducted and the three most consist results were

    selected for each type of material for analysis. The strain rate used for all the samples

    was 8.0 mm/min.

    The stress-strain curves of the samples are shown in Figures 3-11. Figures 3, 4

    and 5 show the stress-strain curve of three different carbon-carbon samples. Figures 6, 7

    and 8 show the stress-strain curve of three different spectra-carbon samples. Similarly,

    Figures 9, 10 and 11 show the stress-strain curve of three different spectra-nylon samples.

    21

  • 8/2/2019 Team 10 Report Final-Version

    31/71

    Figure 3. Stress-Strain curve of Carbon-Carbon sample 1.

    Figure 4. Stress-Strain curve of Carbon-Carbon sample 2.

    22

  • 8/2/2019 Team 10 Report Final-Version

    32/71

    Figure 5. Stress-Strain curve of Carbon-Carbon sample 3.

    Figure 6. Stress-Strain curve of Spectra-Carbon sample 1.

    23

  • 8/2/2019 Team 10 Report Final-Version

    33/71

    Figure 7. Stress-Strain curve of Spectra-Carbon sample 2.

    Figure 8. Stress-Strain curve of Spectra-Carbon sample 3.

    24

  • 8/2/2019 Team 10 Report Final-Version

    34/71

    Figure 9. Stress-Strain curve of Spectra-nylon sample 1.

    Figure 10. Stress-Strain curve of Spectra-nylon sample 2.

    Figure 11. Stress-Strain curve of Spectra-nylon sample 3.

    25

  • 8/2/2019 Team 10 Report Final-Version

    35/71

    From above curves the important mechanical properties were calculated,

    including yield strength, yield strain, tensile strength (UTS), ductility, strain at failure, E

    modulus, and toughness. Tables 3, 4, and 5 show the results of these measurements along

    with mean and standard deviation of each property for carbon-carbon, Spectra-carbon,

    and Spectra-nylon respectively.

    Table 3Tensile Properties of Carbon-Carbon composites

    Table 4Tensile Properties of Spectra-Carbon Composites

    26

  • 8/2/2019 Team 10 Report Final-Version

    36/71

    Table 5Tensile Properties of Spectra-Nylon Composites

    3.2 Hardness Test

    Hardness is the ability of a material to resist plastic deformation. The Rockwell

    hardness tester uses an indenter to press against the surface of the material under study.

    There is a minor load (10 grams) which constantly presses on the indenter; a major load

    will be applied to the material gradually until equilibrium has been reached. Then the

    major load will be removed and some of the penetration it caused will recover. At this

    point the residual penetration is measured which is the hardness. The three studiedsamples, which were Spectra-Nylon, Spectra-Carbon, and Carbon composites were tested

    for their hardness. The hardness test was done using a Rockwell machine on an M scale

    with minor load being 10 kgf and the major load 100 kgf.

    As Table 6 shows, the hardness of spectra-nylon and spectra carbon are very

    similar, being tenths different, whereas Carbons hardness is twice as much as the other

    two. This result shows that carbon can withstand plastic deformation much more than the

    other two samples. Therefore in case of hardness and plastic deformation, carbon would

    be the most suitable for lower limb prostheses application. Evidently lower limb

    27

  • 8/2/2019 Team 10 Report Final-Version

    37/71

    prostheses are constantly under load during the time they are being used, therefore it is

    crucial that they do not fail and deform plastically. From the Box-Whisker plot, we can

    see that the hardness value of 75.6 in Carbon sample is an outlier.

    Table 6Hardness test results for Spectra-Nylon, Spectra-Carbon and Carbon composites

    28

  • 8/2/2019 Team 10 Report Final-Version

    38/71

    Figure 12. Box plot for Hardness test Data

    29

  • 8/2/2019 Team 10 Report Final-Version

    39/71

    3.3 SEM Analysis

    A Scanning Electron Microscopic analysis on the tensile test samples were done

    in the SEM lab at San Jose State University. The samples were placed in a low vacuum,

    slightly humid environment inside the SEM. The fractured area was at the (0, 0) co-

    ordinates of the mount so as to focus on the crack area. Figure 13 shows the SEM

    pictures of the carbon fiber samples at different magnifications. Figure 13(c) shows that

    the fiber surface at the crack is smooth which indicates that the bonding between the resin

    and fibers is poor. We can conclude that the resin broke first which was followed by fiber

    pullout.

    Fiber pullout

    (a)

    (b)

    30

  • 8/2/2019 Team 10 Report Final-Version

    40/71

    (c)

    Figure 13. SEM images of Carbon-Carbon samples. (a) Crack area at 50x magnification(b) at 400x magnification (c) at 6000x magnification

    Figure 14 are SEM images of Spectra-Carbon samples. Figure 14(a) indicates that

    the failure mechanism was de-lamination. It can be seen in Figures 14(b) and 14(c) that

    the resin is still attached to the fiber. Figure 14(d) shows the breakage of resin. Theseimages indicate that although the bonding between resin and fibers in the Spectra-Carbon

    samples were poor, it was better than in Carbon fibers. Since Spectra-Carbon is a hybrid

    of polyethylene (spectra) and carbon fibers, it is difficult to tell which of these fibers

    caused this failure mode. An Energy Dispersive X-ray analysis could be done to

    differentiate between the fibers by identifying their composition.

    31

  • 8/2/2019 Team 10 Report Final-Version

    41/71

    (a)

    Fiber

    Matrix

    (b)

    32

  • 8/2/2019 Team 10 Report Final-Version

    42/71

    (c)

    Resinbreakage

    (d)Figure 14. SEM images of Spectra-Carbon samples. (a) Crack area at 6000xmagnification (b) at 2400x magnification (c) and (d) at 10000x magnification

    Figure 15 shows SEM images related to Spectra-Nylon samples. Figure 15(a)

    shows the resin crack and 15(b) shows the fiber breakage. The fiber has good adhesion

    33

  • 8/2/2019 Team 10 Report Final-Version

    43/71

    with the resin but they are not strong enough. As such, the fiber was pulled out with resin

    still attached to it.

    Cracks in the

    (a)

    34

  • 8/2/2019 Team 10 Report Final-Version

    44/71

    Resin

    Fiber

    (b)Figure 15. SEM images of Spectra-Nylon samples at 3000x magnification (a) top surface

    (b) fiber pull-out

    3.4 Discussion of Results

    The hardness and tensile tests were done in accordance to ASTM D 3039 standard. The

    results of the tensile test show that the tensile strength of carbon samples to be in the

    range of 83-89 MPa spectra-carbon - 27 to 36 MPa, and spectra-nylon to be in the range

    of 28 to 30 MPa. According to the literature review, the tensile strength of Carbon

    fiber/Epoxy resin composite was 76.8 MPa, while that of spectralon, which is a hybrid of

    spectra and nylon, was 25MPa. The findings of this project are close to those in the

    literature review. The ultimate tensile strength (UTS) of cortical bone is between 80 and

    115 MPa (Mechanical Properties of Bone). The mean UTS of carbon sample is 92 MPa,

    which is in the range of that of the cortical bone. From stress-strain curves of spectra-

    carbon, and spectra-nylon, it can be seen that the UTS is very close to the point of

    35

  • 8/2/2019 Team 10 Report Final-Version

    45/71

    fracture. This indicates that the samples are very brittle. The fact that Spectra containing

    composites have no yielding and the Carbon-Carbon does have plastic deformation can

    be due to the bonding strength of resin. That is the Carbon-Carbon sample continues to

    plastically deform after yield strength because the resin is no longer bound to the carbon

    fiber while the Carbon fibers by themselves have not failed completely. On the other

    hand, Spectra-Carbon and Spectra-Nylon had good bonding with resins, therefore making

    them brittle and failing without plastic deformation.

    36

  • 8/2/2019 Team 10 Report Final-Version

    46/71

    4.0ECONOMIC JUSTIFICATION

    4.1 Executive Summary

    Limb prosthetics need to have certain mechanical properties to perform well and

    meet different requirements and regulations. Prosthetic companies need to test their

    material to optimize the performance and cost. Prosthetic Labs offers material testing and

    consulting services to prosthetic limb manufacturers in order to evaluate their materials.

    What sets Prosthetics Labs apart from other material testing companies is that the

    services are exclusive to limb prosthetic companies. Hence, the customers can get

    specialized and tailored tastings from experts in their field to choose the best and

    cheapest material.

    The prosthetic market was about 1.45 billion in 2006, with an estimated growth

    rate of 3.9 %, which makes the testing market potentially lucrative. With more than 500

    limb prosthetic companies in the U.S, the current market for material testing of these

    companies is estimated to be $3,500,000. The goal of Prosthetics Lab is to gain 30% of

    the market share amounting to $1,050,000. The main competitors in this market are

    testing companies that offer services to a wide variety of industries, therefore being

    exclusive to limb prosthetics, Prosthetic Labs has a major competitive advantage.

    However, the major weakness is the limited number of potential customers.

    The average price point of the different services is $7000, which was calculated

    from number of test and the costs incurred. The company will start in year 2010 with an

    estimated budget required is $415,816. The company will break even in the second

    37

  • 8/2/2019 Team 10 Report Final-Version

    47/71

    quarter of 2012. The required capital will be funded from bank loans, family and friends

    contribution, and venture capitals. These investments will be returned starting from 2013.

    Prosthetics Labs will offer its service by using web cataloguing. The revenue will

    be generated through customer referrals and advertisement in biomedical journals and

    magazines. Also, advertisement space will be sold in the companys website to material

    manufactures for prosthetics.

    4.2 Problem Statement

    In 2007, there were nearly 185,000 amputees in the United States. Due to the

    large number of amputees, the prosthetic limb market is quite large. In order for the

    prostheses to perform well and not fail, their material must be tested for its mechanical

    properties and failure. Therefore Prosthetics Labs is offering its services to test these

    materials for this purpose. The customers of Prosthetics Labs are companies which are

    manufacturers of prosthetic limbs. These manufacturing companies also need

    consultation in order to reduce their material cost. Also, FDA requires the prostheses

    limbs to possess certain qualities for safety and efficacy.

    4.3 Solution and Value Proposition

    Prosthetic Labs is the only company which offers testing and consulting services

    exclusive to prosthetic limb market. Working with only prosthetic limbs, makes

    Prosthetic Labs an expert in this field. The company offers material selection and tailored

    tests for the customers. Other companies offering testing are not aimed at prostheses.

    38

  • 8/2/2019 Team 10 Report Final-Version

    48/71

    They offer a wide range of device testing in the biomedical field, which makes them not

    as specialized as Prosthetic Labs in prostheses area.

    Consulting services offered by this company will result in choosing the best and

    cheapest material which has the required properties for its use. The prostheses

    manufacturing companies often use unnecessary amount of expensive materials in order

    to improve the performance of the prosthesis. Also, there are different prostheses for

    various uses; prostheses for athletes must be much stronger and much more flexible than

    prostheses for the elderly. Therefore a consulting service is required to understand these

    differences and make changes to the material amount to save cost and improve

    performance.

    4.4 Market Size

    According to a study by Frost & Sullivan, the U.S. market for prosthetics was

    about $1.45 billion in 2006, and this value is estimated to reach $1.85 billion by 2013

    (Prosthetics Market Growing, 2007). Based on Frost & Sullivan study the average annual

    growth rate will be 3.9 % which equates to annual revenue of $56.5 million. This means

    that the estimated revenue of $1,135,000 will be 2.38 % of the total market growth. This

    is important because the growth rate of Prosthetics Labs completely depends on the

    prostheses market. Hence, there is a great potential in the prosthetics industry which

    creates a good opportunity for Prosthetics Lab Company.

    39

  • 8/2/2019 Team 10 Report Final-Version

    49/71

    To calculate the market size for limb prosthesis testing services, the number of

    limb prostheses manufacturing companies needs to be found. . There are about 500 limb

    prosthesis companies in the U.S. Some of the companies in the United States that

    manufacture limb prostheses are: Ossur, Becker Orthopedic Appliance Company, Hanger

    Orthopedic Group Inc (Prosthetics and Orthotics, 2009).

    The entire market size of material testing and consulting for prosthesis companies

    can be estimated by multiplying the number of all potential customers by the average

    service charge of $7000. Therefore, the estimated current market size is $3,500,000. The

    goal of Prosthetics Lab is to attract 150 customers or more of these possible 500

    companies by year 2013. This means that Prosthetics Lab will gain 30% of the prosthetic

    limb testing market share which amounts to $1,050,000.

    4.5 Competitors

    Table 7 shows the major competitors for Prosthetic Labs including their annual

    revenue, geographic location and the industries they serve. As it can be seen in Table 7,

    most of these competitors serve wide variety of industries and located throughout the

    country and worldwide. The companies that only serve Medical Device have much

    smaller revenues. Furthermore, there are no competitors that are exclusive to prosthesis

    companies, thus most of the competitors will be large corporate companies.

    40

  • 8/2/2019 Team 10 Report Final-Version

    50/71

    Table 7Major Competitors

    4.6 Customers

    The customers for Prosthetics labs are the companies that manufacture limb

    prosthesis. The principal manufacturing or R&D engineers of these companies will be the

    targets for sale.

    There are currently about 500 limb prosthesis companies located in the United

    States. The population distribution of these companies is show in Figure 16. The states

    with highest percentage of the potential customers are New York, Ohio, Florida,

    Pennsylvania, and California, each having 6-7.5 % of the prosthetic companies. Since

    Florida has one of the highest potential customer bases, and corporate income is not

    taxable in Florida for a single-member LLC, Prosthetic Labs will be located in

    41

  • 8/2/2019 Team 10 Report Final-Version

    51/71

    Florida.

    Figure 16. Percent concentration of limb prosthesis companies in the U.S.

    4.7 Cost

    The cost for starting the company and providing the services for the years 2010-

    2011 can be broken down into two different categories. These costs include fixed cost

    and variable cost which will be discussed in detail in this section.

    42

  • 8/2/2019 Team 10 Report Final-Version

    52/71

    4.7.1 Fixed Cost

    Fixed cost is the cost needed to provide the service that does not change with the

    amount of service. The fixed cost break down of Prosthetics Labs Inc. is shown in Table

    8 for two years. The fixed cost of the company includes:

    Startup Cost, which is the initial cost needed to set up the office including thebusiness license fees, lease deposit, rent, furniture, computers, printers, fax machine,

    and etc.

    Office rent Stationary, telephone and internet services, website, insurance Equipment rental, which includes Instron tensile test machine, SEM machine,

    Rockwell hardness test machine, and three point bending test.

    Salaries, which include salaries of seven permanent employees. The three of the

    employees including the CEO, the senior consultant and the senior analyst have been

    assumed to work for free for the first three months. They will start getting paid from

    the first quarter, although with no benefits until the end of the first year. From the

    other four employees, two of them are testing technicians. The first year only one

    testing technician will be hired; however from the beginning of the second year

    another testing technician will be added. SEM analyst is another employee who will

    be working part-time for the first year, and full time in the second year. The last

    employee would be the janitor, working 10 hours a week. Benefits have also been

    included in the salaries for the full time employees in the cost break down.

    43

  • 8/2/2019 Team 10 Report Final-Version

    53/71

    Since Prosthetics Labs Inc. only provides research and consulting services, there are no

    manufacturing and material costs. There will also be no patent and intellectual property

    costs.

    Table 8Fixed Cost Break Down

    44

  • 8/2/2019 Team 10 Report Final-Version

    54/71

    4.7.2 Variable Cost

    The variable costs of Prosthetics Labs include the costs which are for hiring

    contract professionals and advertising. The two contract employees are one marketing

    analyst and another scientific analyst. Due to the low number of customers in the first

    quarters, the scientific analyst is only needed a few hours a week. However, as the

    company progresses and attracts more customers, the scientific analyst will be hired for

    longer hours per week.

    As oppose to the scientific analyst, the marketing analyst will be needed more in

    the initial quarters. As the company grows and settles, the need for marketing as well as

    advertising will decrease. Other variable costs are incurred by packaging, shipping,

    equipment calibration and maintenance. The variable cost break down is shown in Table

    9.

    Table 9The Variable Cost Break Down

    45

  • 8/2/2019 Team 10 Report Final-Version

    55/71

  • 8/2/2019 Team 10 Report Final-Version

    56/71

    Table 11SWOT analysis

    4.10 Investment Capital Requirements

    In order to calculate the budget needed to start the company and make a profit, the

    total income and losses of company need to be calculated. The income is the revenue

    that the company will be generating based on the number of customers. The expected

    number of customers has been calculated based on quantitative forecasting.

    The profit/loss then is calculated by subtracting the total cost (the sum of fixed andvariable costs) from the total income. Table 12 and Figure 17 show the expected profit

    and loss (P&L) of the company for years 2010 to 2012.

    47

  • 8/2/2019 Team 10 Report Final-Version

    57/71

    Table 12Expected Profit and Loss

    Figure 17. Profit and Loss Chart

    48

  • 8/2/2019 Team 10 Report Final-Version

    58/71

    The time at which a company starts to make a profit is called the break-even point.

    Figure 18 shows the break-even graph. As can be seen from this figure, Prosthetics Lab

    will be break-even from losses and start making profit of $36,450 in the second quarter

    of 2012. The total capital required will be the sum of all the losses before break-even.

    Therefore, the expected budget for Prosthetics will be $415,816.

    Figure 18. Break-even chart

    To fund the budget three different sources will be used which are bank loans,

    venture capital and money borrowed from family and friends. The break down of the

    funding is shown in Table 13.

    Table 13

    Funding Break Down

    49

  • 8/2/2019 Team 10 Report Final-Version

    59/71

    4.10.1 Norden-Rayleigh Financial Profile

    The Norden Rayleigh cost profile indicates the total budget spent until breakeven.

    Table 15 shows the calculations of cumulative distribution function and probability

    density for the curve.

    Cumulative distribution function is given by (Wesley, 2009):

    Where, t-time to breakeven

    V(t) total amount spent

    d- estimated budget

    a-cost drivers

    Cost drivers are those that cost money to a company. Assuming the cost drivers to

    be as in Table 14, a will be equal to 0.55.

    Table 14Cost Drivers

    Number of cost drivers, N=6

    Hence, a=Total/N=0.008

    50

  • 8/2/2019 Team 10 Report Final-Version

    60/71

    The probability density function is given by (Wesley, 2009):

    From Figure 19, it can be seen that maximum budget is spent at the end of three months.

    Table 15Probability density function and Cumulative distribution function for Norden-Rayleigh

    Figure 19. Cumulative Distribution Function for Norden Rayleigh

    51

  • 8/2/2019 Team 10 Report Final-Version

    61/71

    Figure 20. Probability Density Function curve for Norden-Rayleigh

    4.11 Personnel

    As a start-up company, Prosthetics Lab will only need a few employees to start

    the business. There are three founders of the company who will occupy the top positions.

    All three of the founders will share the same set of technical and scientific skills. The

    skills include having at least an M.S. degree in materials or biomedical engineering. They

    must have the knowledge of material science, especially in composites and biomaterials

    and experience in mechanical testing of material and analysis. They must also have a

    good understanding of the prosthetics industry.

    The first of these three positions is the CEO. The CEO will be responsible for

    management of the employees and financial resources of the company. She will also be

    52

  • 8/2/2019 Team 10 Report Final-Version

    62/71

    responsible for legal matters and hiring. The second position would be the senior analyst

    whose responsibility is to analyze the data gathered from the material testing.

    Subsequently the third top position is the senior consultant. She will be in charge of

    consulting the costumers based on the analysis of the results and coming up with the most

    effective material for the intended purpose.

    There will also be three other permanent full time positions in the Prosthetic Labs.

    The positions are two material testing technicians and one SEM technician. All three

    positions must have 2 years or more experience relevant to the tasks. The material testing

    technicians must have a high school diploma or higher. They must be trained to perform

    the tests offered by the company including Instron tensile test, bending, and Rockwell

    hardness test. The testing technicians will also be responsible for cutting the samples in

    order to fit in the test machines. They must also be computer literate and have the

    knowledge of operating specific software used for the tests. The SEM technician must

    have at least a B.S degree in material science or material engineering. He must be

    experienced to operate the SEM machine and have basic knowledge of failure analysis.

    There will also be another permanent janitorial position which will only be ten hours a

    week.

    The rest of the employees are contract employees which are the marketing analyst

    and contract scientific analyst. Since the company is a start-up the marketing analyst will

    be extensively needed in the beginning phases. The marketing analyst must have five

    years or more experience working with start-ups in the biomedical field, preferably

    53

  • 8/2/2019 Team 10 Report Final-Version

    63/71

    prosthetics and have at least an M.B.A degree. The scientific analyst must have a Ph.D.

    degree in material or biomedical engineering. The skills required for this position is to

    give professional advice to the consultant and the scientific analyst. Due to low number

    of customers in the starting period the contract scientific analyst will only be hired for

    two to six hours a week depending on work load. As the company develops, the need for

    the scientific analyst will increase and he will be hired for longer hours per week.

    4.12 Business and Revenue Model

    Any business needs a strategy to sell its product and also generate revenue, which

    is called business and revenue model. In addition to on location order processing,

    Prosthetics Lab will use web cataloguing to enable its customers to choose the required

    testing services for their product. Web cataloguing has the advantage of being easier for

    those customers who find it hard to commute to the office location. The customers will

    also have all the information about the services and the amount of samples needed online.

    The samples can be shipped to the testing location, after which they will be tested and

    analyzed. The consultation services can either be offered by email, fax or in person as

    desired by the customer.

    The strategy that Prosthetics Lab plans to make revenue is by advertising and

    customer referrals. Advertising will be done in several ways. One way is to advertise

    Prosthetics Lab in journals related to materials and medical devices. Prosthetics Labs will

    also be advertised on the internet, on different prosthesis manufacturers websites.

    54

  • 8/2/2019 Team 10 Report Final-Version

    64/71

    Another way is to sell advertising space of the companys website to materials suppliers

    for medical devices.

    4.13 Strategic Alliances and Partners

    Prosthetic Labs will be a sole operator and will have no strategic alliances or

    partners.

    4.14 Exit Strategy

    Return on Investment is calculated by dividing the profit by the total cost of

    investment. It signifies how profitable an investment is at each time period. Table 16 and

    Figure 21 show the ROI for Prosthetic Labs.

    Table 16Return on Investment for Prosthetics Labs

    55

  • 8/2/2019 Team 10 Report Final-Version

    65/71

    ROI

    -100.00

    -80.00

    -60.00

    -40.00

    -20.00

    0.00

    20.00

    40.00

    Q1,

    201

    0

    Q2,

    201

    0

    Q3,

    201

    0

    Q4,

    201

    0

    Q1,

    201

    1

    Q2,

    201

    1

    Q3,

    201

    1

    Q4,

    201

    1

    Q1,

    201

    2

    Q2,20

    12

    Q3,20

    12

    Q4,

    201

    2

    Time

    %ROI

    ROI

    Figure 21. ROI Chart

    The ROI is negative in the first nine quarters which shows that the company is not

    making any profit. The rate of increase in the ROI is maximal between the first and third

    quarter of 2012. It is by the second quarter of year 2012 that Prosthetics Labs will be able

    to return the capital investments. Assuming a ten fold return for the venture capital,

    Prosthetics Labs will repay about 35% of the profit starting from year 2013 for the next

    20 years. The bank loans will be repaid based on their interest rate.

    56

  • 8/2/2019 Team 10 Report Final-Version

    66/71

    5.0 PROJECT SCHEDULE

    The schedule for the entire project has been split in to two charts: one for ENGR 281 andthe other for ENGR 298

    Gantt Chart for ENGR 281

    57

  • 8/2/2019 Team 10 Report Final-Version

    67/71

    58

    Gantt Chart for ENGR 281

  • 8/2/2019 Team 10 Report Final-Version

    68/71

    6.0 CONCLUSION AND FUTURE DIRECTION

    In this project the mechanical properties of Carbon-Carbon, Spectra-Carbon, and

    Spectra-Nylon composites has been tested and assessed for the specific application of

    lower limb prosthesis. According to the results the Carbon-Carbon composite sample had

    the highest tensile strength of 92 11 and hardness of 93.5 5.8. The SEM results show

    that the binding of carbon fibers and resin for Carbon-Carbon composites is poor

    compared to the other samples. This result is consistent with the results from stress-strain

    diagrams from which the brittleness of spectra-carbon is seen, indicating strong binding.

    While it is assumed that Spectra-carbon can be a cost-effective alternative to

    carbon composites, Carbon-Carbon composite is still the best material of choice for this

    application because of superior mechanical properties. Future research could be done on

    carbon fiber and resin binding in order to make stronger binding between the two.

    Furthermore, in future more types of samples and tests could be used for testing and

    analyzing for this purpose.

    59

  • 8/2/2019 Team 10 Report Final-Version

    69/71

    References

    (2008). Quality and testing. In B. A. Strong, Fundamentals of Composites

    Manufacturing: Materials, Methods and Application (p. 277). SME.

    Bannister, P. (1978). The Prosthetic Problems of Below Knee Amputees with FlexionContractures. The Australian Journal of Physiotherapy , 21-25.

    Below Knee (Transtibial) Prosthesis. (n.d.). Retrieved March 12, 2009, from ProstheticConcepts Web site: http://www.prostheticconcepts.ca/belowknee.pdf

    Berry, D.A. (1987), Composite materials for orthotics and prosthetics. Orthot Prosthet,40,3543.

    Carroll, K. (2006). Lower Extremity Socket Design and Suspension. Physical Medicationand Rehabilitation Clinics of North America, 17, 31-48.

    Cochrane, H., Orsi, K., & P., R. (2001). Lower limb amputation. Prosthetics andOrthotics International , 21-28.

    Erwin, J.R. (1947): Fiberglass plastics.Industrial Medicine, 16, 439-41

    Evans, S. L., & Gregson, P. J. (1998). Composite technology in load-bearing orthopaedicimplants. Biomaterials, 19(15), 1329-1342

    Fabrication Materials and Supplies. (n.d.). Retrieved May 12, 2009, from ForseeOrthopedic Products: http://www.4corthopedic.com/products/fab/epox.htm

    Gutfleisch, O. (2003). Peg legs and bionic limbs:the development of lower extremityprosthetics.Interdisciplinary Science Reviews .

    Honeywell. (n.d.).Honeywell Advanced Fibers and Composites. Retrieved may 10, 2009,from Spectra Fiber : http://www51.honeywell.com/sm/afc/products-details/fiber.html

    Kelly, B. (2009). Lower Limb Prosthetics,

    http://emedicine.medscape.com/article/317358-overview, Extracted 13 March 2009.Kelly, B. M., Spires, M. C., & Restrepo, J., A. (2007). Orthotic and Prosthetic

    Prescriptions for Today and Tomorrow. Physical Medicine and RehabilitationClinics of North America, 18, 185-858.

    60

  • 8/2/2019 Team 10 Report Final-Version

    70/71

    Lange, L. R. (1992). The Lange Silicone Partial Foot Prosthesis.Journal of Prostheticsand Orthotics , 56-61.

    Lewin, M., Sello, S. B., & Preston, J. (1996).Handbook of fiber science and technology

    (Vol. III). New York: Marcel Dekker, Inc.

    Madden, M. (1991). Fabricating the Spiral and Hemi-Spiral Orthosis Using a CompositeLay-Up Technique.Journal of Prosthetics & Orthosis , 94-99.

    Mak, A., F., T., Zhang, M., & Boone, D., A. (2001). State-of-the-art research in Lower-limb prosthetic biomechanics socket interface: A review. Journal of RehabilitationResearch and Development, 38(2), 161-174.

    Marquardt, E., & J., C. (1984). Amputations and Prostheses for the Lower Limb.International Orthopaedics , 139-146.

    Mechanical Properties of Bone. (n.d.). Retrieved November 10, 2009, from ICB Dent Web site:http://www.feppd.org/ICB-Dent/campus/biomechanics_in_dentistry/ldv_data/mech/basic_bone.htm

    Moore, W.,S., Hall, A., D., & Lim, R., C. (1972). Below the knee amputation forischemic gangrene: Comparative results of conventional operation and immediatepostoperative fitting technique. The American Journal of Surgery, 124(2), 127-134.

    Osbourne, P., E., (2009). Below Knee ( transtibial) Prosthesis. Prosthetic Concepts,http://www.prostheticconcepts.ca/belowknee.pdf, extracted on 12 March 2009.

    Pasquina, P., F., Bryant, P., R., Huang, M., E., Roberts, T., L., Nelson, V., S., & Flood,K., M. (2006). Limb deficiency and prosthetic management: focused review.Advances in Amputee Care.American Academy of Physical Rehabilitation, 87, 34-43.

    Phillips, S. L., & Craelius, W. (2005). Material Properties of Selected ProstheticLaminates.

    Journal of Prosthetics & Orthotics, 27-32.

    Piggot, M.R., & Harris B. (1980). Compression strength of carbon, glass and Kevlar-49fiber reinforced polyester resins.Journal of material Science, 15, 2523-2538

    Ramakrishna, S., Mayer, J., Wintermantel, E., & Leong, K. W. (2001). Biomedicalapplications of polymer-composite materials: A review. Composites Science andTechnology, 61(9), 1189-1224

    61

    http://www.prostheticconcepts.ca/belowknee.pdfhttp://www.prostheticconcepts.ca/belowknee.pdf
  • 8/2/2019 Team 10 Report Final-Version

    71/71

    Rothschild, V. R., Fox, J. R., Michael, J. W., Rothschild, R. J., & Playfair, G. (1991).Clinical Experience with Total Thermoplastic Lower Limb Prostheses.Journal ofProsthetics and Orthotics , 51-54.

    Schuch, M. C. (1991). Thermoplastic Applications in Lower Extremity Prosthetics.Journal of Prosthetics and Orthotics , 1-8.

    Shindo, A., Nakanishi, Y., & Soma, I. (1969). Carbon fibers from cellulose fibers.ApplPolym Symp, 9 , 271284

    Stacey, E., Blachford, L., Cengage, G., (2002). eNotes.com.2006. Retrieved 17 April,2009, from http://www.enotes.com/how-products-encyclopedia/artificial-limb

    Strike, S., & Hillery, M. (2000). The design and testing of a composite lower limbprosthesis. Proceedings of the Institution of Mechanical Engineers.Part H, Journal

    of Engineering in Medicine, 214(6), 603-614

    Verrall, J. P. (1940). Amputation Stumps and Artificial Limbs.British Medical Journal ,62-64.

    Wan, Y. Z., Chen, G. C., Huang, Y., Li, Q. Y., Zhou, F. G., Xin, J. Y., et al. (2005).Characterization of three-dimensional braided carbon/Kevlar hybrid composites fororthopedic usage. Materials Science and Engineering A, 398(1-2), 227-232

    Wesley, L. P. (2009).Lecture 2: Economic Justification Section for Your Report andSoftware & Hardware Project Cost Estimation [PowerPoint slides]. Retrieved from

    http://www.engr.sjsu.edu/lwesley/teaching/ENGR295B/lecture_slides/Lec2.pdf

    Witra, R., W., Mason, R., Calvo, K., & Golbranson, F., L. (1991). Effect on gait usingvarious prosthetic ankle-foot devices.Journal of Rehabilitation Research andDevelopment, 28(2), 13-24.