Electrode Maintenance

PeriFlux 6000 | tcpO2 made intelligent

Page 2: Tcpo2 electrode maintenance

Disclaimer

The information contained in this document is intended to provide general

information only. It is not intended to be, nor does it constitute, medical advice.

Under no circumstances is the information contained in this document to be

interpreted as a recommendation for a particular treatment for specific

individuals. In all cases it is recommended that clinicians perform their own

interpretations of data in conjunction with the clinical assessment of their patient.

Due to Perimed’s commitment to continuous improvement of our products, all

specifications are subject to change without notice.

All information and content in this document is protected by copyright. All rights

are reserved. Users are prohibited from modifying, copying, distributing,

transmitting, displaying, publishing, selling, licensing, creating derivative works,

or using any information available in or through the document for commercial or

public purposes. All responsibility for any liability, loss or risk, personal or

otherwise, which is incurred as a consequence, directly or indirectly, of the use

and application of any of the material in this document is specifically disclaimed.

Page 3: Tcpo2 electrode maintenance

• Calibration

• tcpO2 Site Selection

• Electrode Maintenance

• Interpreting Curves and Results

Page 4: Tcpo2 electrode maintenance

Remembraning

• Electrodes build up a layer of silver oxide during use that

must be removed

• Remembrane electrodes weekly or when they come into

contact with any sort of infectious disease (e.g. MRSA)

• Calibrate twice after remembraning

• Remembraning is performed with the supplied kit

Page 5: Tcpo2 electrode maintenance

Remembraning Procedure

Use O-ring removal tool Remove both O-rings Remove membrane

Clean electrode face Clean circular groove Two drops of electrolyte

solution – NO bubbles!

Page 6: Tcpo2 electrode maintenance

Remembraning Procedure

Insert electrode into

cartridge and compress New membrane and

two O-rings loaded

Page 7: Tcpo2 electrode maintenance

Electrode Care

• Electrodes can be cleaned with the following water-based disinfectants:

– Hi-Tor Plus (Ecolab Inc.)

– MadaCide-FD (MADA Inc.)

– Control III (Maril Products Inc.)

• Never immerse plug in liquid

• Avoid exposing lead to alcohol-based products (e.g. hand lotion, disinfectant)

• When electrodes are not in use, store with cap on and two drops of contact liquid