Electrode Maintenance
PeriFlux 6000 | tcpO2 made intelligent
![Page 2: Tcpo2 electrode maintenance](https://reader038.vdocuments.us/reader038/viewer/2022100801/58ec9b3b1a28abb4778b4611/html5/thumbnails/2.jpg)
Disclaimer
The information contained in this document is intended to provide general
information only. It is not intended to be, nor does it constitute, medical advice.
Under no circumstances is the information contained in this document to be
interpreted as a recommendation for a particular treatment for specific
individuals. In all cases it is recommended that clinicians perform their own
interpretations of data in conjunction with the clinical assessment of their patient.
Due to Perimed’s commitment to continuous improvement of our products, all
specifications are subject to change without notice.
All information and content in this document is protected by copyright. All rights
are reserved. Users are prohibited from modifying, copying, distributing,
transmitting, displaying, publishing, selling, licensing, creating derivative works,
or using any information available in or through the document for commercial or
public purposes. All responsibility for any liability, loss or risk, personal or
otherwise, which is incurred as a consequence, directly or indirectly, of the use
and application of any of the material in this document is specifically disclaimed.
![Page 3: Tcpo2 electrode maintenance](https://reader038.vdocuments.us/reader038/viewer/2022100801/58ec9b3b1a28abb4778b4611/html5/thumbnails/3.jpg)
• Calibration
• tcpO2 Site Selection
• Electrode Maintenance
• Interpreting Curves and Results
![Page 4: Tcpo2 electrode maintenance](https://reader038.vdocuments.us/reader038/viewer/2022100801/58ec9b3b1a28abb4778b4611/html5/thumbnails/4.jpg)
Remembraning
• Electrodes build up a layer of silver oxide during use that
must be removed
• Remembrane electrodes weekly or when they come into
contact with any sort of infectious disease (e.g. MRSA)
• Calibrate twice after remembraning
• Remembraning is performed with the supplied kit
![Page 5: Tcpo2 electrode maintenance](https://reader038.vdocuments.us/reader038/viewer/2022100801/58ec9b3b1a28abb4778b4611/html5/thumbnails/5.jpg)
Remembraning Procedure
Use O-ring removal tool Remove both O-rings Remove membrane
Clean electrode face Clean circular groove Two drops of electrolyte
solution – NO bubbles!
![Page 6: Tcpo2 electrode maintenance](https://reader038.vdocuments.us/reader038/viewer/2022100801/58ec9b3b1a28abb4778b4611/html5/thumbnails/6.jpg)
Remembraning Procedure
Insert electrode into
cartridge and compress New membrane and
two O-rings loaded
![Page 7: Tcpo2 electrode maintenance](https://reader038.vdocuments.us/reader038/viewer/2022100801/58ec9b3b1a28abb4778b4611/html5/thumbnails/7.jpg)
Electrode Care
• Electrodes can be cleaned with the following water-based disinfectants:
– Hi-Tor Plus (Ecolab Inc.)
– MadaCide-FD (MADA Inc.)
– Control III (Maril Products Inc.)
• Never immerse plug in liquid
• Avoid exposing lead to alcohol-based products (e.g. hand lotion, disinfectant)
• When electrodes are not in use, store with cap on and two drops of contact liquid