System Components for Sample Gas Sampling Data Sheet 10/23�5.10 EN March 2004
A gas analysis system usually consists of two groups of devices: • The gas analyzer measures the concentrations of the sample
gas components and converts them into an electrical signal. • Supplementary devices sample the sample gas from the
process, condition it and feed it in the gas analysis system. In most cases, the sample gas sampled from the process cannot be processed by the gas analyzer without further conditioning. Too high a dust content, temperature and dew point as well as too high or too low pressure, and not least interference components in the sample gas can impair the operational capability of the gas analyzer and falsify the measuring result. Supplementary devices such as the sampling probe, the sample gas line, the sample gas cooler and pumps and filters ensure that the sample gas inlet conditions of the connected gas analyzers are satisfied and faultless measuring results are obtained independ-ently of the process sequence and the local conditions.
Contents Page
Modular Sample Gas Probe System 2 FE2 Filter Unit 3 Annular Heating Element 4 Heating Sleeve 5 PFE2 Filter Unit 6 PFE3 Filter Unit 13 Selection Table for Probe Tube Types and Probes 15 Type K Probe Tube 16 Type 40 Probe Tube 17 Type 42 Probe Tube (heated) 18 Type 60 Probe Tube (Cement Flue Gas Probe) 19 Probe 2 20 Probe 1, Probe 11 21 Ceramic Filter 22 Process Pressure Reducers 23 Heated Sample Gas Lines 25 CGWB 13 Heated Sample Gas Line 26 TBL 01, TBL 01-E Heated Sample Gas Lines 28 Electronic Temperature Controller and Switching Relay 30
2 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004
Modular Sample Gas Probe System
Easily adaptable to different sampling conditions thanks to modular construction
Components: • FE2, PFE2, PFE3 filter units • Probe tube types K, 40, 42, 60 • Annular heating element • Heating sleeve
Probe tubes and filter unit heatable
Connections for test gas and purging
Filter with high retention rate and long service life
Easy filter replacement
Usable in emission measuring systems conforming to 13th BImSchV and 17th BImSchV (Federal Immission Protection Law)
10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 3
FE2 Filter Unit
Connection Ports Sample Gas/Test Gas
G 1/4 internal thread to DIN ISO 228/1 Compressed Air (2 Ports)
G 1/2 internal thread to DIN ISO 228/1 Probe Tube
G 3/4 internal thread to DIN ISO 228/1 Compressed Air Supply for Blowback (See PFE2 Filter Unit, page 6) Instrument Air
Free of dust, water and oil Pressure
max. 600 kPa Ordering Information
Catalog No. Housing SS, flange 1.0718 23042-5-8018413Housing SS, with coating 23042-5-8018416Housing SS, 1 Pt 100 23042-5-8018402Housing SS, 2 Pt 100 23042-5-8018404Housing SS, with coating, 1 Pt 100 23042-5-8018406Housing SS, with coating, 2 Pt 100 23042-5-8018408Housing and flange SS 23042-5-8018414Housing and flange SS, coated 23042-5-8018417Housing and flange SS, 1 Pt 100 23042-5-8018403Housing and flange SS, 2 Pt 100 23042-5-8018405Housing and flange SS, coated, 1 Pt 100 23042-5-8018407Housing and flange SS, coated, 2 Pt 100 23042-5-8018409
Use The FE2 filter unit is a component of the modular gas sampling probe system. It is used for filtering gases containing dust. Process Gas Sampling Conditions Pressure
pabs = 50 to 600 kPa Temperature
max. 200 °C (coated max. 180 °C) Flow Rate
30 to 500 l/h, relative to 100 kPa and 0 °C Pressure Drop
approx. 0.6 hPa at 100 l/h Dust Content
max. 3 g/m3 (application recommendation); max. 40 g/m3 with probe tube type 40 with blow-back device; max. 2,000 g/m3 with probe tube type 60
Filter Characteristics Filter Element
Surface filter Filter Fineness
0.5 µm Retention Rate
99.98 % Seal Integrity
10–4 hPa l/s Dead Volume
280 ml Materials of the Gas-carrying Parts Casing and Connections
Special steel Mat. No. 1.4571 Flange
Special steel Mat. No. 1.0718 (optionally special steel Mat. No. 1.4571)
Filter
Aluminum oxide surface filter Coating
Thermoset plastic, resistant up to 180 °C Installation Flange
DN 65, PN 6, Form B to DIN 2573, flange gasket A; positioning of holes: 0°, 45°, 90°, 135°, 180°, 225°, 270°, 315°, 360°
Mounting
Flange connect or bolt to the probe tube Weight
approx. 8.0 kg
4 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004
Annular Heating Element
Dimensional Drawing (dimensions in mm)
Heating Circuit
Use For heating the FE2 filter unit, which is a component of the modular gas sampling probe system. Heating is necessary if condensation is likely from moist gases and there is consequently risk of the filter becoming clogged and corroded. Data Power Supply
115 V AC or 230 V AC, 50 to 60 Hz Heater Output
200 W, 100 W or 50 W – depending on setting The heating resistors are factory-set for heat output of 100 W
Temperature of Heater (unregulated)
approx. 120 °C to 180 ° C (measured at ambient temperature of 20 °C, still air)
Ambient Temperature
0 to 55 °C (still air) Degree of Protection of Housing
IP 50 Materials
Housing: aluminum alloy, heating surface: brass Weight
approx. 1.7 kg Electrical Connections
Terminals via cable gland PG 13.5 The electrical supply conductor must be heat resistant (≥ 200 °C).
Installation Two half shells are placed round the FE2 filter unit to be heated and closed with a wing screw. Option Regulated Heating with 1 x Pt 100 Ordering Information
Catalog No. Ring heater 230 V, 50/60 Hz 23029-5-7865312Ring heater 115 V, 50/60 Hz 23029-5-8008813Ring heater with Pt 100, 230 V, 50/60 Hz 23029-5-7865316Ring heater with Pt 100, 115 V, 50/60 Hz 23029-5-7865317
230 / 115 V AC200 W 100 W 50 W
230 / 115 V AC 230 / 115 V AC
10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 5
Heating Sleeve
Regulated Heating Sleeve for FE2 Filter Unit Heater
Micanite Holding Temperature
max. 350 °C Temperature Control
max. 200 °C (Pt 100) Power Supply
230 V AC, optionally 115 V AC Power Consumption
250 VA Connection
Terminal box with PG 9 bolted connection, protection level IP 54
Supply Cable
3 x 1 mm2 with silicone sheath, length 1 m Ambient Temperature
–20 to +45 °C Relative Humidity
≤ 75 % yearly mean; infrequent slight condensation permissible
Weight
approx. 1 kg Essential Accessories Insulating Mat Temperature Controller (see page 30) Ordering Information
Catalog No. Heated sleeve, 230 V, 50/60 Hz, 300 VA 23043-5-8018410Heated sleeve, 115 V, 50/60 Hz, 300 VA 23043-5-8018411
6 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004
PFE2 Filter Unit
Use The PFE2 filter unit consists of the FE2 filter unit with a regulat-ed heating sleeve, installed in a special-steel protective case. It is used to filter dust-containing gases and is a component of the modular gas sampling probe system. When used for gases with high dust content, the PFE2 can optionally be fitted with valves for automatic cleaning of the filter element in the FE2 filter unit. Protective Case Dimensions
450 x 450 x 400 mm (H x W x D) Material
Special steel, Mat. No. 1.4301 Degree of Protection of Housing
IP 54 Ambient Temperature
–20 to +45 °C Weight
approx. 18 kg Options – Heating sleeve, regulated (Pt 100), with insulating mat – Annular heating element, unregulated – Filter fineness 0.3 µm Pneumatic Valve Diaphragm Valve
2/2-way straight-way valve Operating Pressure/Control Pressure
600 kPa (6 bar) Nominal Diameter
DN 12 Connections
G 1/2 internal thread DIN ISO 228/1 Control air: G 1/8 internal thread DIN ISO 228/1
Material
Brass Mode of Operation
Closed in the normal position Ambient Temperature
0 to +100 °C
Control Valve Solenoid Valve
3/2-way solenoid valve Nominal Diameter
DN 2.0 Pressure Range
600 kPa (6 bar) Connections
G 1/8 internal thread DIN ISO 228/1 Material
Brass Electrical Connections
Standard connector Power Supply
115 V AC or 230 V AC, 50 to 60 Hz Power Consumption
8 VA Mode of Operation
Closed in normal position Ambient Temperature
0 to +100 °C
10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 7
PFE2 Filter Unit
Ordering Information
Catalog No. 23028-0- 0 0 0 0 0 0 ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓Design without protection box 1 with protection box IP 54 (450 x 450 x 400 mm) 2 with protection box, IP 54, anti-freeze heating 3 with protection box, IP 54, ambient temp. max. 65 °C 4 with protection box IP 54 (330 x 203 x 270 mm) 5 Filter housing (equal FE 2) Stainless steel 1 Stainless steel, with coating, max. 180 °C 2 Housing and flange SS 3 Housing and flange SS, coated 4 Stainless steel with 0,3 µm filter element 5 Stainless steel, with coating, max. 180 °C, with filter element <0,3 µm (filt. fineness) 6 Housing and flange SS, with filter element <0,3 µm (filtration fineness) 7 Housing and flange SS, coated, with filter element <0,3 µm (filtration fineness) 8 Heating Without 1 with heating unit and insulation (max. 200 °C) 2 with ring heater, unregulated 3 Control / Monitoring Without 1 with 1x Pt 100 2 with 2x Pt 100 3 Back purging unit (without compressed air unit) without reverse purging 0 with pneumatic valve 1 with 3x pneumatic-, 1x air-bleed-valve(s) 2 with check valve for AC-FNT 3 Power supply 230 V, 50 ... 60 Hz 1 115 V, 50 ... 60 Hz 2 ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓
Catalog No. 23028-0- 0 0 0 0 0 0
8 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004
PFE2 Filter Unit
Probe Protective Case Dimensional Diagram (dimensions in mm)
Test
gas
con
nect
ion
Bul
khea
d fit
ting
4/6x
1 m
m
Com
pres
sed
air
Pur
ging
Bul
khea
d fit
ting
10/1
2x1
mm
Sam
ple
gas
conn
ectio
n
Con
trol
air
Bul
khea
d fit
ting
4/6x
1 m
m
Fla
nge
plat
e w
ith p
ipe
clam
p2
1/2"
RS
T 3
7-2
hot g
alva
nize
d D
IN 3
568
for
mou
ntin
g at
pro
be m
ount
ing
adap
ter
Elo
ngat
ed h
oles
for
M8
safe
ty c
olla
r bo
lts
Ele
ctric
al c
onne
ctio
ns
Cov
er s
wiv
els
upw
ards
2x P
g21
3x P
g16
1x P
g13,
51x
Pg1
1
Take
hol
es fo
r ca
ble
glan
ds o
n th
e le
ft in
to a
ccou
ntTe
rmin
al b
ox w
ith c
able
gla
nds
Use
as
requ
ired
10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 9
PFE2 Filter Unit
PFE2 Filter Unit: Pneumatics Connections for 2-Stage Back-Purging
1 Terminal box 2 Probe protective case 3 B Bulkhead fitting compressed air 4 to 6 bar 4 E Plug (test gas connection) 5 Pilot valve, filter cleaning -Y2.1 6 Diaphragm valve, filter cleaning -Y2.2 7 C Sample gas connection 8 Pilot valve, pulsed compressed air -Y1.1 9 Diaphragm valve, pulsed compressed air -Y1.2 10 Bulkhead fitting 6 mm control air 2…6 bar 11 Bulkhead fitting 12 mm compressed air 4…6 bar 12 Heated sample gas line 13 Power supply for heated sample gas line 14 FE2 filter unit 15 Solenoid valve, venting -Y4 16 Diaphragm valve, tube and filter surface cleaning -Y3.2 17 Pilot valve, tube and filter surface cleaning -Y3.1
10 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004
PFE2 Filter Unit
Probe Protective Case, PFE2 Filter Unit Terminal Diagram, Terminal Block -X1 (Standard)
23
1
3
8
10
14
2
9
15
17
16
6
5
4
13
12
11
7
22
19
21
18
20
N
PE
N
PE
L
L
PE
N
20
23
22
18
21
19
PE
N
PE
N
L
Pt 100 three wire connectionIEC 751 Cl. B
230 V 50-60 Hzor 115 V 50-60 Hz
Probe protective case Terminal box, strip -X1
1st Pt 100
2nd Pt 100
HeaterFE2 filter unit
HeaterType 42 probe tube
FE2 filter unit
FE2 filter unit
Type 42 probe tube1st Pt 100
Type 42 probe tube2nd Pt 100
(fail safe circuit)
(fail safe circuit)
HeaterType 42 probe tube
Pt 100 three wire connectionIEC 751 Cl. B
Pt 100 three wire connectionIEC 751 Cl. B
Pt 100 three wire connectionIEC 751 Cl. B
230 V 115 V50-60 Hz 50-60 Hz
10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 11
PFE2 Filter Unit
Probe Protective Case, PFE2 Filter Unit Terminal Diagram, Terminal Block -X2 (Standard)
1
3
4
6
7
2
5
8
12
11L
N
PE
L
N
PE
9
10
13
Probe protective case Terminal box, strip -X2
Heatersample gas linewattage acc. to lengthmax. 3600 W
Anti-frost heaterfor system shut downapprox. 5 °C at ambienttemperature -20 °C30 WInternal temperature approx. 100 °CUse heat-resistant cables!
230 V 50-60 Hz or115 V 50-60 Hz
Pt 100 Three wire connectionIEC 751 Cl. B
2nd Pt 100sample gas line(fail-safe circuit)
1st Pt 100sample gas line Pt 100 three wire connection
IEC 751 Cl. B
230 V 50-60 Hz or115 V 50-60 HzThe heater operates unregulated.Switch on from outside onlywhen system shut down
12 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004
PFE2 Filter Unit
Probe Protective Case, PFE2 Filter Unit Terminal Diagram, Terminal Block -X3 (for Cleaning)
Dimensional Drawing: Mounting of Probe Protective Case with PFE2 Filter Unit (dimensions in mm)
Minimum distance xmin depending on mounting angle α:
α 10° 15° 20° 25° 30° 35° xmin /mm 229 248 268 287 307 324
1
1
10
10
11
12
13
2
2
3
4
5
6
7
8
9 9
L
L
PE
PE
N
NLPEN
LPEN
LPEN
PE PE
R
R
R
P
P
P
P
B
A
A
A
R
Pilot valve -Y2.1filter cleaning
Probe protective case Terminal box, strip -X3
Extended cleaning
Probe / flange
230 V 50-60 Hz or115 V 50-60 Hz
Pilot valve -Y3.1cleaning of filter surface and tube
Pilot valve -Y1.1Pulsed compr. air
Pilot valve -Y4Venting
Internal temp. approx. 100 °CUse heat-resistant cable
Remove with equipotential bonding
α
X
10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 13
PFE3 Filter Unit
Design The PFE3 filter unit consists of a modified FE2 filter unit with a self-regulating heater. It is installed in a stainless steel protec-tive case. Process Gas Sampling Conditions Pressure
pabs = 50 to 600 kPa Temperature
max. 200 °C Flow Rate
30 to 500 l/h, relative to 100 kPa and 0 °C Pressure Drop
approx. 0.6 hPa at 100 l/h Dust Content
max. 3 g/m3 (application recommendation) Filter Characteristics Filter Element
Surface filter Filter Fineness
0.5 µm (optional 0.3 µm) Seal Integrity
10–4 hPa l/s Dead Volume
280 ml Dimensions
50/20 x 135 mm Heater Type
Self-regulating heater (400 W) Temperature
180 °C (ambient temperature 20 °C) Temperature Alarm
approx. 150 °C Power Supply
110 to 240 V AC Option
Annular heating element (see page 4) Materials of the Gas-carrying Parts Casing, Flange and Connections
Special steel Mat. No. 1.4571 Seals
Viton
Installation Flange
DN 65, PN 6, Form B to DIN 2527; positioning of holes: 45°, 135°, 225°, 315°
Mounting
Flange connect or bolt to the probe tube Connection Ports Sample Gas
Pipe fitting 6 mm Test Gas
Bulkhead fitting 6 mm (option) Probe Tube
G 3/4 internal thread to DIN ISO 228/1 Protective Case Dimensions
270 x 203 x 330 mm (H x W x D) Material
Special steel Mat. No. 1.4301 Degree of Protection of Case
IP 54 Weight
approx. 17 kg Ambient Temperature
–20 to +45 °C
14 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004
PFE3 Filter Unit
Dimensional Drawing, Gas Connections (dim. in mm) Electrical Connections
204
2x M16
270
2x M1650
3x M20
1
2
3
1
2
3
4
5
4
PE
3
PE
2
1
9
13
11
12
10
8
7
6
5
1 Test gas connection (optional) 6 mm bulkhead fitting 2 Sample gas connection 6 mm pipe fitting 3 Check valve for ACF-NT (optional)
1 Heater for filter 2 Heater for sample gas line 3 Pt 100 (optional)
4 Pt 100 sample gas line 5 Temperature monitor
Ordering Information
Catalog No. 23046-0- 0 0 0 0 0 0 0 0 ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓Design with protection box IP 54 (330x 203x 270 mm) 1 with protection box IP 54 (330x 203x 270 mm) and check valve for ACF-NT 2 Filter housing (without connections for back purging) Housing stainless steel (filtration fineness 0,5 µm) 1 Housing and flange stainless steel (filtration fineness 0,5 µm) 2 Housing stainless steel with filter element <0,3 µm (filtration fineness) 3 Housing and flange stainless steel, with filter element <0,3 µm (filtration fineness) 4 Heating With self-regulating heating and temperature alarm (110/220 V, AC) 1 with ring heater unregulated 2 Power supply 230 V AC 1 115 V AC 2 ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓
Catalog No. 23046-0- 0 0 0 0 0 0 0 0
10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 15
Selection Table for Probe Tube Types and Probes
Pro
be
tub
e ty
pe
K
PV
DF
Pro
be
tub
e ty
pe
40
SS
1.4
571
Pro
be
tub
e ty
pe
40
SS
1.4
571
coat
ed
Pro
be
tub
e ty
pe
40
SS
1.4
893
Pro
be
tub
e ty
pe
40
silic
one
carb
ide
Pro
be
tub
e ty
pe
42
SS
hea
ted
Pro
be
tub
e ty
pe
60
wat
er c
oole
d
Pro
be
2 un
heat
ed
Pro
be
2 he
ated
Pro
be
2 p
artia
lly h
eate
d
Pro
be
60 S
w
ater
coo
led
Pro
be
H
wat
er c
oole
d
< 130 °C 2)
< 180 °C 2)
180–450 °C
180–900 °C
1300 °C
Max. temperature
1400 °C
< 50 mg/m3
< 1000 mg/m3
< 3 g/m3
> 3 g/m3 1) 1) 1) 1)
40–100 g/m3 1) 1) 1) 1) 1) 1)
Dust content
2000 g/m3 1) 1) 1)
< 65 °C Water vapor dew point > 65 °C
Acidic aerosols (acid dew pt. 130 °C)
< 10 m/s Flow velocity
> 10 m/s
Technical data see ... p. 16 p. 17 p. 17 p. 17 p. 17 p. 18 p. 19 p. 20 p. 20 p. 20 3) 3)
1) Back purging required 2) with regulated heating sleeve 3) “ACK” data sheet (Document No. 10/23-8.40 EN)
Lead-Through Tube Installation
Brickwork with metal sheeting Isolated sheet-metal channel Brickwork
1 2 5 3
1 2 5 3
α
1 2 5 34
Flow direction of process gas
Minimum distance from wall xmin depending on mounting angle α:
α 10° 15° 20° 25° 30° 35° xmin /mm 133 138 143 147 151 153
1 Lead-through tube 2 Mounting flange DN 65, PN 6,
Form B to DIN 2573 3 Gasket 4 Welded-on rectangular block 5 Sampling probe flange
Dimensions in mm
16 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004
Type K Probe Tube
Use Sampling of low-dust but strongly corrosive process gases up to 130 °C in conjunction with the FE2/PFE2 filter unit. Area of application includes gas sampling in chemical industry and in power stations after the flue gas scrubber. Process Gas Sampling Conditions Pressure
pabs = max. 500 kPa (5 bar) Temperature
max. 130 °C Flow Rate
30 to 300 l/h, relative to 1 bar, 0 °C, dry gas Pressure Drop
≤ 10 hPa at 250 l/h Flow
max. 10 m/s Ambient Temperature
–20 to +80 °C 90% Time (T90)
< 2 s at 250 l/h Dust Content
max. 50 mg/m3 in conjunction with the FE2/PFE2 filter unit Mounting Flange
DN 65, PN 6, Form B, to DIN 2573 positioning of holes: 45°, 135°, 225°, 315°
Mounting Angle
any; 8° to 90° downwards from the horizontal is recommended (see dimensional drawing)
Sample Gas Outlet
G 1/4 internal thread DIN ISO 228/1 Materials of the Gas-carrying Parts The gas sampling probe K consists of PVDF, mounting flange included. It is available in several lengths. From 1,500 mm fitting length upwards the probe includes a metal support tube hermetically enclosed in plastic tubes.
Dimensional Drawing (dimensions in mm)
1 4
1
2
1 Sample gas outlet 2 Outer ∅ 20 mm, inner ∅ 16 mm for L1 ≤ 1000 mm
Outer ∅ 32 mm, inner ∅ 16 mm for L1 ≥ 1500 mm Ordering Information
Catalog No. Length 500 mm 23025-4-0800857Length 1000 mm 23025-4-0800858Length 1500 mm 23025-4-0800859Length 2000 mm 23025-4-0800860Length 2500 mm 23025-4-0800861
10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 17
Type 40 Probe Tube
Use For sampling hot dust-containing gases in conjunction with the FE2/PFE2 filter unit Process Gas Sampling Conditions Pressure pabs = 50 to 500 kPa (0,5 to 5 bar) Temperature
SS Mat. No. 1.4571 max. 450 °C SS Mat. No. 1.4571 (coated) max. 180 °C SS Mat. No. 1.4893 max. 900 °C Silicone carbide (max. length 1.000 mm) max. 1.300 °C
Flow Rate max. 300 l/h Pressure Drop approx. 10 hPa at 30 to 90 l/h Dust Content
max. 3 g/m3 (application recommendation); max. 40 g/m3 with blow-back device in conjunction with FE2/PFE2 filter unit
Installation G 3/4 external thread DIN ISO 228/1 Mounting Angle
any; 8° to 90° downwards from the horizontal is recom-mended. It is imperative to install the silicon carbide probe tube vertically.
Dimensional Drawing (dimensions in mm)
G3/4
G3/4
G3/4
L1
L1
L1
16
21,4
ø
12,6
ø16
,8 ø
15 ø
23 ø
34 ø
16
16105
17,2
øSilicon-Carbide (max. 1300 °C)
Special Steel Material No. 1.4571 (max. 450 °C)
Special Steel Material No. 1.4893 (max. 900 °C)
Ordering Information
Catalog No. 23036-0- 0 0 0 0 0 0 0 0 0 0 ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓Version Stainless steel (1.4571), up to 500 °C 1 Stainless steel (1.4571), coated, up to 180 °C 2 Silicon carbide up to 1300 °C, max. 1000 mm 4 Stainless steel (1.4893), up to 900°C 5 Stainless steel (1.4571), up to 300 °C, heated (for cement) 6 Nominal length 500 mm 1 1000 mm 2 1500 mm 3 2000 mm 4 2500 mm 5 3500 mm (only for heated version) 6 4000 mm (only for heated version) 7 4500 mm (only for heated version) 8 ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓
Catalog No. 23046-0- 0 0 0 0 0 0 0 0 0 0
18 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004
Type 42 Probe Tube (heated)
Use • Low gas temperature • High moisture content in gas • Corrosive components in the process gas Typical places of use are: • Sampling points in wet desulphurization plants
(crude gas, clean gas, part-cleaned clean gas) • Waste incineration plants • Auxiliary boilers and stack emission measurements Process Gas Sampling Conditions Pressure
pabs = 200 kPa (2 bar) Temperature
< 200 °C Flow Rate
max. 250 l/h Flow Velocity
max. 12 m/s (> 1,000 mm max. 7 m/s) Dust Content
max. 3 g/m3 (application recommendation); max. 40 g/m3 with blow-back device in conjunction with FE2/PFE2 filter unit
T90 Time
approx. 0.4 s at 250 l/h
Probe Tube Heater Heat Output
nominal length 1.000 mm: 2 x 200 W nominal length 1.500 mm: 2 x 300 W nominal length 2.000 mm: 2 x 400 W
Nominal Voltage
230 V/115 V AC, 50 to 60 Hz Control
Control temperature max. 200 °C Temperature Sensor
Pt 100 Continuous Operation
at 200 °C flue gas temperature 450 °C Ordering Information
Catalog No. Length 1000 mm 23041-4-8018392Length 1500 mm 23041-4-8018393Length 2000 mm 23041-4-8018394
Dimensional Drawing (dimensions in mm) and Terminal Diagram
12345
160
40
Pt100
1000/1500/2000 mm
25 x
1,5
40 x
2
Heater
Temperature ControllerConnectionr
10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 19
Type 60 Probe Tube (Cement Flue Gas Probe)
Use For sampling hot cement flue gas with high dust content for CO, CO2, NO and SO2 analysis in “dry” cement production process (e.g. recuperator furnaces, furnaces with grate coolers), at rotary kiln inlet and/or after precalcination. Process Gas Sampling Conditions Pressure
pabs = max. 200 kPa (2 bar) Temperature
max. 1,200 °C for length = 3,500 mm max. 1,250 °C for length = 3,000 mm max. 1,300 °C for length = 2,500 mm max. 1,350 °C for length = 2,000 mm max. 1,400 °C for length = 1,500 mm
Flow Rate
max. 250 l/h Dust Content
max. 2,000 g/m3 with periodical compressed air cleaning in conjunction with FE2 filter unit
Dead Time (Tt) for Length = 3,500 mm
approx. 40 s at 60 l/h, approx. 9 s at 250 l/h 90% Time (T90) for Length = 3,500 mm
approx. 82 s at 60 l/h, approx. 17 s at 250 l/h Pressure Drop
approx. 10 hPa Materials of the Gas-carrying Parts Special steel, Material No. 1.4541 Mounting A flanged probe connector (internal diameter 200 mm) should be installed by the customer. The mounting flange is essential for installing the probe tube.
Cooling Water Supply Quality
Drinking water quality with approx. 30 % anti-freeze if required
Outlet Temperature
min. 50 °C (or > dew point of the cement flue gas), max. 85 °C
Pressure
pabs = max. 400 kPa (4 bar) Quantity Circulated
approx. 3 m3/h (closed system) Type 60 S Probe Tube Model with reinforced protection tube and flange DN 150, PN 16 Ordering Information
Catalog No. Length 1500 mm 23038-4-0856831Length 2000 mm 23038-4-0856832Length 2500 mm 23038-4-0856833Length 3000 mm 23038-4-0856834Length 3500 mm 23038-4-0856835Length 1500 mm, reinforced protection tube 23038-4-8018478Length 2000 mm, reinforced protection tube 23038-4-8018479Length 2500 mm, reinforced protection tube 23038-4-8018480Length 3000 mm, reinforced protection tube 23038-4-8018481Length 3500 mm, reinforced protection tube 23038-4-8018482Type 60 S, length 1500 mm 23038-4-0730642Type 60 S, length 2000 mm 23038-4-0730643Type 60 S, length 2500 mm 23038-4-0730644Type 60 S, length 3000 mm 23038-4-0730645Type 60 S, length 3500 mm 23038-4-0730646
Dimensional Drawing (dimensions in mm)
L1
L2
90
35
A
A
~60
3
3
1 2
~70
ø60,
3
ø89
L3
1 Sample gas inlet ports
2 Thickened shank of probe tube
3 Cooling water connections
L1 Probe tube length L2 Length of
thickened shank:800 mm for L1 ≤ 2,000 mm 1,300 mm for L1 ≥ 2,500 mm
L3 200 mm with reinforced protection tube
20 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004
Probe 2
Use Wherever large volumes of gas are sampled, gas pressure is raised and there is a high moisture or dust content in the gas. Process Gas Sampling Conditions Pressure
pabs = max. 400 kPa (4 bar) Temperature
max. 450 °C (probe tube unheated) max. 450 °C (probe tube partially heated) max. 130 °C (probe tube heated
Flow Rate
max. 2,000 l/h Dust Content
max. 20 g/m3 (application recommendation), max. 100 g/m3 with purging
Purging Pressure
max. 600 kPa (6 bar) Heating Temperature of Inlet Filter
≥ 200 °C (at process gas temperature of 20 °C) 90% Time (T90)
< 2 s at 250 l/h Filter Fineness
0.5 µm Material of the Gas-carrying Parts Support Tube, Heating Rod, Connecting Pipe
Special steel, Material No. 1.4571 Inlet Filter
Aluminum oxide Seals
Ceramic fiber Flange
Special steel, Material No. 1.4581 Mounting Flange
DN 65, PN 6, Form B, to DIN 2573 position of holes: 45°, 135°, 225°, 315°
Mounting Angle
any; 10° to 35° downwards from the horizontal recommend-ed (take direction of flow into account), see dimensional diagram
Sample Gas Outlet G 1/4 internal thread DIN ISO 228/1
Test Gas/Purge Air Inlet
G 3/8 internal thread DIN ISO 228/1 Electrical Connections
Harting connector, IP 55 Probe Tube Heating Heat Output (unregulated)
Nominal length 1,200 mm: 650 W Nominal length 1,700 mm: 950 W partially heated (for heating of the wall bushing): 255 W
Nominal Voltage
230 V/115 V AC, 50 to 60 Hz Ordering Information
Catalog No. 1200 mm, 220 V, 650 W 23022-4-07683011700 mm, 220 V, 950 W 23022-4-07683021200 mm, 115 V, 650 W 23022-4-07683031700 mm, 115 V, 950 W 23022-4-07683041200 mm, 220 V, 255 W, partially heated 23022-4-07683051700 mm, 220 V, 255 W, partially heated 23022-4-07683061200 mm, 115 V, 255 W, partially heated 23022-4-07683071700 mm, 115 V, 255 W, partially heated 23022-4-0768308 Dimensional Drawing and Connection Diagram
Z
260
Z
180
650
4
1
α
1 = N2 = L1
1 Filter mounting 2 Gas sampling probe tube 3 Sample gas outlet or test gas/purge air inlet 4 Harting connector
10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 21
Probe 1, Probe 11
Probe 1 Use in interior air, garage and tunnel monitoring for the gases carbon monoxide (CO) and oxygen (SO2) Technical Data Sampling Pressure
Atmospheric pressure Gas Temperature
–25 to +50 °C Sample Gas Flow Rate
max. 120 l/h Pressure Drop with New Filter
≤ 10 hPa at 120 l/h Dust Content in the Air
approx. 100 mg/m3
Sample Gas Connection Port
G 1/8 internal thread DIN ISO 228/1 or nozzle for flexible tube of internal diameter 4 mm
Materials
Silicone carbide (filter), aluminum (end seals), PVC (nozzle) Transport and Storage Temperature
–25 to +70 °C Mounting
On wall or suspended from ceiling Ordering Information
Catalog No. Room air probe 1 23011-5-5862922 Dimensional Drawing (dimensions in mm)
Probe 11 Use in interior air, garage and tunnel monitoring, including corrosive atmospheres. The materials used ensure that with the gases normally occurring no adsorption occurs. Technical Data Sampling Pressure
Atmospheric pressure Gas Temperature
–25 to +50 °C Sample Gas Flow Rate
max. 70 l/h Pressure Drop with New Filter
≤ 10 hPa (mbar) at 70 l/h Service Life of Filter (estimated)
filter constant (2500) d (days) =
flow rate (l/h) x dust concentration (mg/m3) Sample Gas Connection Port
Bolted connection for flexible pipe with internal diameter 4 mm and external diameter 7 mm
Materials
Polycarbonate (probe head), aluminum oxide (filter) Transport and Storage Temperature
–25 to +70 °C Mounting
On wall or suspended from ceiling Ordering Information
Catalog No. Room air probe 11 23021-4-0859585 Dimensional Drawing (dimensions in mm)
578,4
104
140
83
~~
~ ~~ ~~ ~
85 mm wide
22 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004
Ceramic Filter
Use For filtering dust from gases in the gas conditioning systems of analyzer systems when the gas sampling probe does not have a filter. The ceramic filter is positioned upstream the other devices in the sample gas conditioning chain. The ceramic filter can be heated with the annular heating element if condensation from the sample gas is likely. Technical Data Flow Rate (Air)
max. 300 l/h Gas Pressure
pabs = 20 kPa to 1.7 MPa (0.2 to 17 bar) at 20 °C filter temperature, max. 200 kPa (2 bar) at 180 °C
Pressure Drop
0.1 to 0.85 hPa at 30 to 250 l/h (air) Dynamic Characteristics at
Flow rate (air) 30 l/h 60 l/h 90 l/h Dead time Tt 3.7 s 2.3 s 1.7 s 90% time T90 11.5 s 5.5 s 3.9 s
Filter Fineness
0.5 µm Construction, Mounting Materials of the Gas-carrying Parts
Housing: steel, St 37 K or special steel, Material No. 1.4571; Filter: aluminum oxide; Seals: FPM, Cu, Al2O3 fiber
Gas Connection Ports
G 1/4 internal thread DIN ISO 228/1 or pipe connectors for pipe of internal diameter 4 mm
Filter Surface
approx. 212 cm2 Weight
approx. 2.1 kg Filter Heating
Annular heating element, see page 4 Mounting
On wall with angle bracket Ordering Information
Catalog No. Ceramic filter, steel version 23031-5-8018177Ceramic filter, stainless steel version 23031-5-8018178
Dimensional Drawing (dimensions in mm)
10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 23
Process Pressure Reducers
Use If the pressure of the process gas is too high for the analyzer, a pressure reducer is required to lower the pressure. The type of pressure reducer used is determined by the properties of the process gas and the required sample gas pressure.
Technical Data
Casing: Brass Casing: Brass Casing: Special steel Casing: Special steel
Construction Single-stage with one manometer (outlet)
Single-stage with two manometers (inlet /outlet)
Single-stage without manometer
Single-stage with two manometers (inlet /outlet)
Inlet pressure max. 1.5 MPa (15 bar) max. 3.5 MPa (35 bar) max. 3.5 MPa (35 bar) max. 3.5 MPa (35 bar)
Outlet pressure (adjustable)
0.8 to 10 kPa (8 to 100 mbar)
10 to 40 kPa (0.1 to 0.4 bar)
10 to 70 kPa (0.1 to 0.7 bar)
10 to 70 kPa (0.1 to 0.7 bar)
Safety valve set to –
50 kPa (0.5 bar)
80 kPa (0.8 bar)
80 kPa (0.8 bar)
Manometer display Inlet pressure
–
0 to 4 MPa (0 to 40 bar)
–
0 to 4 MPa (0 to 40 bar)
Manometer display Outlet pressure
0 to 250 hPa (0 to 250 mbar)
0 to 60 kPa (0 to 0.6 bar)
–
0 to 150 kPa (0 to 1.5 bar)
Flow rate 25 to 250 l/h 25 to 250 l/h 25 to 250 l/h 25 to 250 l/h
Material Diaphragm Buna rubber Buna rubber Special steel Special steel
Material Gasket Buna rubber Buna rubber EPDM EPDM
Material Housing Brass Special steel; gas-carrying parts: brass
Special steel Special steel
Weight approx. 1.25 kg approx. 2.65 kg approx. 2.35 kg approx. 2.35 kg
Gas temperature and ambient temperature
max. 50 °C max. 50 °C max. 50 °C max. 50 °C
Catalog No. 23035-5-5864958 23035-5-5864951 23422-4-0730699 23422-4-0730700
Supplied: Pressure reducer with angle bracket
24 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004
Process Pressure Reducers
Dimensional Drawings (dimensions in mm) Pressure reducer single-stage with one manometer (outlet)(Catalog No. 23035-5-5864958)
Pressure reducer single-stage without manometer (Catalog No. 23422-4-0730699)
79
105
Ø4
GasTech
approx. 94(depending on handwheel position)
approx. 30° (variable)
Pressure reducer single-stage with two manometers (inlet/outlet) (Catalog No. 23035-5-5864951)
Pressure reducer single-stage with two manometers (inlet/outlet) (Catalog No. 23422-4-0730700)
118
Ø4
GasTech148
approx. 94(depending on handwheel position)
approx. 30° (variable)
10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 25
Heated Sample Gas Lines
Heated Sample Gas Line CGWB 13
Parallel strip heater (holding temperature max. 200 °C)
Pt 100 resistance thermometer for control system
Flexible sample gas pipe, replaceable (option)
Self-regulating anti-frost heater (option)
Self-regulating heater (option, holding temperature max. 100 °C)
Heated sample gas line TBL 01
Fixed resistor heating conductor (holding temperature max. 200 °C)
Pt 100 resistance thermometer for control system
Self-regulating heater (option, holding temperature max. 100 °C)
Heated sample gas line TBL 01-E
Competitive version
Pipe ends with polyamide hard caps Selection Table for Heated Sample Gas Lines
CGWB 13 TBL 01 TBL 01-E
Heater
Self-regulating strip heater (30 °C) x – –
Self-regulating strip heater (100 °C) x x –
Fixed resistor heating conductor (200 °C) – x x
Parallel strip heater x – –
Outer covering Polyvinyl chloride Polyamide Polyamide
Options
Line can be shortened x – –
Gas-carrying line can be replaced x – –
Small bending radius – x x
Gas connections Various options Various options Stainless steel
Electrical connections (heater and Pt 100) Power fixed, Pt 100 plug Power and Pt 100 plug Power and Pt 100 fixed
Maximum line length 130 m (60 m) 100 m (40 m) 40 m
Applications
CO, O2, CO2 Anti-frost heater – –
CO, O2, CO2 Holding temp. 100 °C Holding temp. 100 °C –
SO2, NOx, CnHm Holding temp. 200 °C Holding temp. 200 °C Holding temp. 200 °C
26 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004
CGWB 13 Heated Sample Gas Line
Technical Data
Frost protected line Self-regulating strip heater Regulated parallel strip heater
Outer covering PVC PVC PVC
Medium-carrying pipe PTFE PTFE PTFE
Integral pipe-end connectors 1) POM cap POM cap Metal cap and metal cap with flange
Heater self-regulating strip heater self-regulating strip heater CPD heating element, regulated
Power supply 230 V or 115 V, 50 to 60 Hz 230 V or 115 V, 50 to 60 Hz 230 V or 115 V, 50 to 60 Hz
Heat output 16 W/m 38 W/m 60 W/m
Current consumption Making/holding current = 65/25 mA
0.15 A/m 0.3 A/m
Holding temperature see diagram see diagram 200 °C
Ambient temperature Tamb –30 to +60 °C –30 to +60 °C –30 to +60 °C
External diameter 30 mm 43 mm 40 mm; with replaceable flexible pipe: 45 mm
Flexible sample gas pipe 6 x 4 x 1 mm or 8 x 6 x 1 mm 6 x 4 x 1 mm or 8 x 6 x 1 mm 6 x 4 x 1 mm or 8 x 6 x 1 mm
Maximum line length 130 m 90 m 60 m with replaceable flexible pipe: 25 m, 60 m if laid straight
Line duct PG 42 PG 42 PG 42
Minimum bending radius 300 mm 300 mm 300 mm
Weight approx. 1 kg/m approx. 1 kg/m approx. 1 kg/m
Connection lead for heater (fixed)
3 x 2.5 mm2; length 2.5 m 3 x 2.5 mm2; length 2.5 m 3 x 2.5 mm2; length 2.5 m
Connection lead for Pt 100 (with plug)
– – LIYCI 2 x 0.14 mm2; length 5 m
1) For line lengths > 10 m, only the analyzer side is assembled,
an assembly set for the probe side is supplied. Holding Temperature Diagrams
50
40
30
20
- 30 - 20 - 10 + 10 + 20 + 30 + 400
Frost protected line, 6 x 4 x 1 mm PTFE, 16 W/m, self-regulating
Med
ium
pip
e te
mpe
ratu
re °C
Ambient temperature °C
100
90
80
70
- 30 - 20 - 10 + 10 + 20 + 30 + 400
Ambient temperature °C
Med
ium
pip
e te
mpe
ratu
re °
C
Self-regulating strip heater, 6 x 4 x 1 mm PTFE
Dimensional Drawing (dimensions in mm)
Probe Side Analyzer Side
∅ 6
24
11
∅ 3
0
300
L = m
50 6049
4256
∅ 6
10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 27
CGWB 13 Heated Sample Gas Line
Ordering Information
Catalog No. 23281-0- 0 0 0 0 0 0 0 ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓Version controlled (Pt 100) 2 self-regulated 16 W/m, 30°C (antifreeze) 3 self-regulated 38 W/m, 100°C 7 Connection probe side with metal cap for PFE 2 2 probe: metal cap / analyzer: flange 3 end of line: plastics 4 Gas conducting line PTFE-tube 6x4x1 mm 1 PTFE-tube 8x6x1 mm 2 Power supply 230 V, 50/60 Hz 1 115 V, 50/60 Hz 2 Controlling and monitoring without Pt 100 0 with 1 x Pt 100 1 ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓
Catalog No. 23281-0- 0 0 0 0 0 0 0
28 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004
TBL 01, TBL 01-E Heated Sample Gas Lines
Technical Data
TBL 01 Self-regulating strip heater 30 W/m
TBL 01 Fixed resistor heating conductor 90 W/m
TBL 01-E Fixed resistor heating conductor 90 W/m
Outer covering Polyamide corrugated tubing Polyamide corrugated tubing Polyamide corrugated tubing
Medium-carrying pipe PTFE PTFE PTFE
Integral pipe end connectors Hard caps, metal caps Hard caps, metal caps and metal cap with flange
Polyamide hard caps
Power supply 230 V or 115 V, 50 to 60 Hz 230 V or 115 V, 50 to 60 Hz 230 V or 115 V, 50 to 60 Hz
Heat output 30 W/m 90 W/m 90 W/m
Current consumption I = 0.15 A/m (starting current), I = 65 mA (holding current)
I = f(length) I = f(length)
Holding temperature see diagram on page 26 200 °C 200 °C
Ambient temperature –30 to +60 °C –30 to +60 °C –30 to +60 °C
External diameter 43 mm 43 mm 43 mm
Flexible sample gas pipe 6 x 4 x 1 mm or 8 x 6 x 1 mm 6 x 4 x 1 mm or 8 x 6 x 1 mm 6 x 4 x 1 mm or 8 x 6 x 1 mm
Flexible test gas pipe – 6 x 4 x 1 mm –
Maximum line length 100 m 40 m 40 m
Line duct PG 48 PG 48 –
Minimum bending radius 200 mm 200 mm 200 mm
Weight approx. 1 kg/m approx. 1 kg/m approx. 1 kg/m
Electrical connection for heater and Pt 100
Screw fitting (fixed for hard cap)
Plugs (see dimensional drawing; fixed for hard cap)
fixed
TBL 01 Dimensional Drawing (dimensions in mm)
300
A,B
49 50 60
2442
56
BA
24
11
PG7
PG11
80 250
11∅ 6
∅ 6
∅ 6
L = m
Probe sideAnalyzer side
TBL 01-E Dimensional Drawing (dimensions in mm)
120
∅ 6
10001000
∅ 6
SS
120
53
SS
Length max. 40 m
Sensor Location
End Cap PA
10/23-5.10 EN March 2004 System Components for Sample Gas Sampling 29
TBL 01, TBL 01-E Heated Sample Gas Lines
TBL 01 Ordering Information
Catalog No. 23283-0- 0 0 0 0 0 0 ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓Design Resistor heating line 90 W/m 1 self regulating heating line 30 W/m 2 Gas conducting line with PTFE-tube 6x4x1 mm 1 with PTFE-tube 8x6x1 mm 3 Material of sheathing with plastic hose of PA (outside and indoors installation) 1 Connections End of line with flange connection 1 End of line: probe: metal cap / analyzer: flange 5 End of line with flanges 6 End of line with plastic tops 7 End of line with 2 metal tops 8 Control / Monitoring without Pt 100 0 with 1 x Pt 100 1 Power supply 230 V, 50 ... 60 Hz 1 115 V, 50 ... 60 Hz 2 ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓ ↓
Catalog No. 23283-0- 0 0 0 0 0 0 TBL 01-E Ordering Information
Catalog No.
Resistor heating line 90 W/m, PTFE 6x4x1, PA caps, 230V 23284-4-0730695Resistor heating line 90 W/m, PTFE 8x6x1, PA caps, 230V 23284-4-0730696Resistor heating line 90 W/m, PTFE 6x4x1, PA caps, 115V 23284-4-0730697Resistor heating line 90 W/m, PTFE 8x6x1, PA caps, 115V 23284-4-0730698
30 System Components for Sample Gas Sampling 10/23-5.10 EN March 2004
Electronic Temperature Controller and Switching Relay
Temperature Controller Control Range –200 to +850 °C
Measurement Input Pt 100
Outputs 2 x Relay 3 A, 250 V AC
Power Supply 110 to 240 V AC +10/–15 %, 48 to 63 Hz
Dimensions 48 x 48 mm, C-rail housing
Weight 140 g
Catalog No. 23345-5-8328859 Switching Relay Load Voltage Urms = 24 to 265 V
Max. Load Current Irms = 30 A (Ambient temperature = 25 °C)
Weight 200 g
Catalog No. 23345-5-8328731
Connection Diagram
The IndustrialIT word mark and all mentioned product names in the form XXXXXIT are registered or pending trademarks of ABB. ABB has Sales & Customer Support expertise in over 100 countries worldwide. www.abb.com
The Company’s policy is one of continuous product improvement and the right is reserved to modify the information contained herein without notice.
Printed in the Fed. Rep. of Germany (03.04)
© ABB 2004