system characteristics. recap system characteristics
TRANSCRIPT
System Characteristics
Class 3
Recap
SYSTEM CHARACTERISTICS
Medical Measurement Constraints
The amplitude and frequency ranges for each parameter
are the major factors that aff ect the design of all
instrument components
Proper measurand-sensor interface cannot be obtained
Medical variables are seldom deterministic
External energy must be minimized to avoid any damage
Nearly all biomedical measurements depend on some form of energy
being applied to the living tissue or to the sensor, for example:
X-ray, ultrasonic imaging and electromagnetic or Doppler
ultrasonic
Blood flow-meters depend on externally applied energy interacting
with living tissue
Safe level of applied energy is an important consideration
Equipment must be reliable
Additional Medical Measurement Constraints
ReliableSimple to operateWithstand physical abuse and exposure to corrosive chemicalsElectrical safety (minimize electric shock hazard)
Classification of Biomedical Instruments
Quantity being sensed:Pressure, flow, temperature, potential, etc.
Advantage: easy comparison of different methods for measuring any quantity
Principle of transduction:Resistive, capacitive, inductive, ultrasonic or electrochemicalAdvantages: a. different applications of each principle can be used to strengthen understanding of each conceptb. newer applications readily apparent
Measurement technique for each physiological system:
Cardiovascular, pulmonary, nervous, endocrineAdvantage: isolates all important measurements for specialistsDisadvantage: considerable overlap of quantities sensed and the principles of transduction used
Classification of Biomedical Instruments...
Clinical medicine specialties:Pediatrics, Obstetrics, Cardiology, Radiology, etc.Advantage: valuable for medical personnel interested in specialized instruments
Interfering and Modifying Inputs
Desired input: the measurand that the
instrument is designed to isolate and measure
Interfering inputs: quantities that inadvertently
affect the instrument as a consequence of the
principles used to acquire & process the desired
inputs
Modifying inputs: undesired quantities that
indirectly affect the output by altering the
performance of the instrument itself
Modifying inputs can affect processing of either desired or
interfering inputs
Some undesirable quantities can act as both a modifying
input and an interfering input
Example
A simplified ECG recording system provides a good example:
In this recording system:
The desired input is: Vecg –electrocardiographic voltage between 2 electrodes (RA & LA)
The interfering inputs are:
50 Hz or 60 Hz (power-line) noise voltage induced in the shaded loop by ac magnetic fields
Also the difference between the currents running through each of the electrodes to the patient and to the ground causes a voltage on Zbody
Example...
In ECG, the example of a modifying input is the orientation of the patient cables. If the plane of the cable is parallel to the ac magnetic field, magnetically introduced interference is zero. If the plane of the cables is perpendicular to the ac magnetic field, magnetically introduced interference is maximal.
Time–dependent changes in electrode impedance
Electrode motion
Elimination of Interfering and Modifying Inputs
To reduce or eliminate the effects of most interfering and modifying inputs we have two alternatives:
1. Alter the design of essential instrument components (preferred, but hard to achieve)
2. Add new components to offset the undesired inputs
Sensor characteristics
Static characteristicsThe properties of the system after all transient effects have settled to their final or steady stateAccuracy, Discrimination, Precision, Errors, Drift, Sensitivity, Linearity, Hysteresis (backslash)
Dynamic characteristicsThe properties of the system transient response to an inputZero order systemsFirst order systemsSecond order systems
Accuracy & discrimination
Accuracy is the capacity of a measuring instrument to give RESULTS close to the TRUE VALUE of the measured quantity
Accuracy is related to the bias of a set of measurements
(IN)Accuracy is measured by the absolute and relative errors
More about errors in a later
Discrimination is the minimal change of the input necessary to produce a detectable change at the output
Discrimination is also known as RESOLUTION When the increment is from zero, it is called THRESHOLD
Precision
The capacity of a measuring instrument to give the same
reading when repetitively measuring the same quantity
under the same prescribed conditions
Precision implies agreement between successive readings, NOT
closeness to the true value
Precision is related to the variance of a set of measurements
Precision is a necessary but not suffi cient condition for accuracy
Two terms closely related to precision
Repeatability
The precision of a set of measurements taken over a short time
interval
Reproducibility
The precision of a set of measurements BUT
taken over a long time interval or
Performed by different operators or
with different instruments or
in different laboratories
Example
Shooting dartsDiscrimination
The size of the hole produced by a dartWhich shooter is more accurate?Which shooter is more precise?
Shooter A Shooter B
Accuracy and ErrorsSystematic errors
Result from a variety of factorsInterfering or modifying variables (i.e., temperature)
Drift (i.e., changes in chemical structure or mechanical stresses)
The measurement process changes the measurand (i.e., loading errors)
The transmission process changes the signal (i.e., attenuation)
Human observers (i.e., parallax errors)Systematic errors can be corrected with COMPENSATION methods (i.e. feedback, fi ltering)
Random errorsAlso called NOISE: a signal that carries no information True random errors (white noise) follow a Gaussian distributionSources of randomness:
Repeatability of the measurand itself (i.e., height of a rough surface)Environmental noise (i.e., background noise picked by a microphone)Transmission noise (i.e., 60Hz hum)
Signal to noise ratio (SNR) should be >>1With knowledge of the signal characteristics it may be possible to interpret a signal with a low SNR (i.e., understanding speech in a loud environment)
Example: systematic and random errors
More static characteristics
Input range
The maximum and minimum value of the physical variable that can be measured (i.e., -40F/100F in a thermometer)
Output range can be defi ned similarly
Sensitivity
The slope of the calibration curve y=f(x)
An ideal sensor will have a large and constant sensitivity
Sensitivity-related errors: saturation and “dead-bands”
Linearity
The closeness of the calibration curve to a specifi ed straight line (i.e., theoretical behavior, least-squares fi t)
Monotonicity
A monotonic curve is one in which the dependent variable always increases or decreases as the independent variable increases
Hysteresis
The diff erence between two output values that correspond to the same input depending on the trajectory followed by the sensor (i.e., magnetization in ferromagnetic materials)
Backslash: hysteresis caused by looseness in a mechanical joint