system and performance audit for non methane …€¦ · system and performance audit for non...
TRANSCRIPT
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
WCC-VOC REPORT 2013/1
Report to the World Meteorological Organization
SYSTEM AND PERFORMANCE AUDIT FOR NON METHANE VOLATILE ORGANIC COMPOUNDS
Global GAW Station – Cape Verde Atmospheric Observatory Mindelo, Cape Verde
Status November 2009
Elisabeth Weiss, Stephan Thiel, Rainer Steinbrecher
WMO World Calibration Centre for VOC
Karlsruhe Institute of Technology
KIT/IMK-IFU, Garmisch-Partenkirchen, Germany
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
Karlsruhe Institute of Technology
Contact information:
Campus Alpine
WCC-VOC
Kreuzeckbahnstr. 19
82467 Garmisch-Partenkirchen
Germany
Phone: +49 88 21 18 32 17 (Rainer Steinbrecher)
+49 88 21 18 32 38 (Elisabeth Weiss)
Fax: +49 88 21 73 57 3
E-mail: [email protected]
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
Contents
1 SUMMARY AND RECOMMENDATIONS ........................................................ 1
1.1 General ................................................................................................................... 11.2 System Audit of the Observatory ......................................................................... 11.3 Audit of the VOC Measurements .......................................................................... 11.4 Recommendations ................................................................................................ 1
1.4.1 Recommendation 1 () ................................................................................... 1
1.5 Conclusions ........................................................................................................... 11.6 Summary Ranking of the VOC Audit at the Station ...... ..................................... 2
2 INTRODUCTION ...................................................................................... 33 SYSTEM AND PERFORMANCE AUDIT FOR VOLATILE ORGANIC COMPOUNDS
(VOC) .................................................................................................. 43.1 Description of the Site .......................................................................................... 43.2 Description of the Observatory ............................................................................ 53.3 Staff / Operators .................................................................................................... 73.4 Monitoring Set-up and Procedures ...................................................................... 8
3.4.1 Air Inlet System for VOC ..................................................................................... 8
3.4.2 Gas-chromatographic System ............................................................................. 8
3.5 Operation and Maintenance ................................................................................ 113.5.1 General ............................................................................................................. 11
3.5.2 Sampling and Calibration .................................................................................. 12
3.5.3 Zero and Repeatability Checks, Target Gas ..................................................... 12
3.5.4 Maintenance ..................................................................................................... 13
3.5.5 Corrective Actions ............................................................................................. 13
3.6 Standards ............................................................................................................. 133.6.1 Regulators and Connections ............................................................................. 13
3.6.2 Zero Air ............................................................................................................. 13
3.6.3 Laboratory Standards ....................................................................................... 13
3.6.4 Working Standards ........................................................................................... 15
3.6.5 Calibration of o-VOC and DMS ......................................................................... 15
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
3.7 Data Acquisition and Processing ....................................................................... 153.7.1 General ............................................................................................................. 15
3.7.2 Chromatogram Evaluation ................................................................................ 16
3.8 Data Management and Submission ................................................................... 163.9 Documentation .................................................................................................... 17
3.9.1 Technical and QA/QC ....................................................................................... 17
3.9.2 Reports of Results ............................................................................................ 17
3.10 Intercomparison of VOC Standards ................................................................... 183.10.1 Experimental Procedure ................................................................................... 18
3.10.2 Results of the VOC Intercomparison ................................................................. 18
4 REFERENCES ...................................................................................... 235 APPENDIX ........................................................................................... 23
5.1 Individual AnaIysis Results ................................................................................ 235.2 WCC VOC target components ............................................................................ 265.3 WCC VOC Reference ........................................................................................... 265.4 List of Abbreviations and Acronyms ................................................................. 28
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 1
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
1 SUMMARY AND RECOMMENDATIONS
1.1 General A system and performance audit for VOCs was conducted at the Cape Verde Atmospheric
Observatory (CVO) (co-organised by the Chemistry Department, University of York and by
the World Calibration Centre for VOC, WCC-VOC). The audits were conducted in
November 2006 and 2009, respectively. This report summarizes the result of both audits.
The observatory is operated as a global GAW station and analyses air samples directly
collected at the station tower for - amongst others - VOC. The audits were conducted
according to the WMO/GAW guidelines and SOPs for audits as existing. The audit directly
involved the responsible site research scientist, the station manager and the station
director of the observatory. This report is for distribution to the station director, to the GAW
Country Contact, to the WMO/GAW secretariat as well as to the QA/SAC-Germany.
1.2 System Audit of the Observatory
1.3 Audit of the VOC Measurements
1.4 Recommendations
1.4.1 Recommendation 1 () • Important spare parts for the VOC analysis system (e.g. PTV injector) should be
made available on site, including manuals for e.g. Valco valves.
• The technical staff should join GAWTEC-training courses to strengthen the
background knowledge of parameters measured.
1.5 Conclusions The station staff is encouraged to continue further the excellent operation.
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 2
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
1.6 Summary Ranking of the VOC Audit at the Station ......
Audit Aspect
Adequacy
Comment (0 = inadequate () through 5 = adequate
(■■■■■■))
Access ■■■■■■
Facilities ■■■■■■
Laboratory and office space ■■■■■■
Air conditioning ■■■■■■
Power supply ■■■■■■
General management and operation ■■■■■■
Organisation ■■■■■■
Competence staff ■■■■■■
Air inlet system ■■■■■■
Instrumentation ■■■■■■
Trace gases ■■■■■■
Instrumental performance VOC ■■■■■■
Standards ■■■■■■
Data management ■■■■■■
Data acquisition ■■■■■■
Data processing ■■■■■■
Data submission ■■■■■■
Documentation ■■■■■■
Log books and internal instructions ■■■■■■
Web site ■■■■■■
GAWSIS ■■■■■■
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 3
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
2 INTRODUCTION
The Cape Verde Atmospheric Observatory (CVO) is managed under the supervision of the
University of York and is operated as a global GAW Station. This station is part of
England's contribution to the World Meteorological Organization Global Atmosphere Watch
(WMO/GAW) programme.
The tasks for the GAW-VOC network have been defined in WMO Report 171 (2007). The
Data Quality Objectives (DQO) and quality control measures set out therein are used for
QA/QC measures as agreed upon by the GAW Scientific Advisory Group for reactive
Gases (SAG RG).
In agreement with the station manager of the CVAO a system and performance audit was
conducted at the station by the WCC-VOC in November 2009.
The audits were performed according the SOP for audits
(http://www.empa.ch/plugin/template/empa/*/55558) adjusted for specific requirements to
VOC analysis in air samples. As a central part of the audit procedure, a VOC
intercomparison based on three travelling standards (TS) of the WCC-VOC was
conducted. Moreover, the whole measurement set-up as well as data processing and
quality assurance measures were reviewed.
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 4
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
3 SYSTEM AND PERFORMANCE AUDIT FOR VOLATILE ORGANIC COMPOUNDS
(VOC)
3.1 Description of the Site The following information about the CVO (site description as well as the description of the
observatory) is mainly taken from the corresponding web site web sites
(http://www.ncas.ac.uk/index.php/en/cvao-home) and the GAW Station-information-system
GawSis (http://gaw.empa.ch/gawsis).
Name: Cape Verde Atmospheric Observatory (CVO)
Function: Global station in WMO RA I – Africa
Location: 16.848 °N, 24.871 °W, (10 m asl) Time Zone: UTC-1
The Republic of Cape Verde is an island country, spanning an archipelago of 10 islands
located in the central Atlantic Ocean, 570 kilometres off the coast of Western Africa. The
islands of Cape Verde as well as its position in front of the African coast are shown in Fig.
3-1. The Cape Verde Atmospheric Observatory is located on the Island São Vicente at
Calhau on the NW facing coast line, with the prevailing trade winds blowing directly off the
ocean. The position of the CVO is also marked in Fig. 3-1.
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 5
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
3.2 Description of the Observatory The Cape Verde Atmospheric Observatory (CVO) was built in 2006. The station is built on
volcanic rock, ca. 50 m from the sea shore with no vegetation around. There are no
habitations within a 2 km radius; the nearest public road is 1.2 km downwind. The
prevailing wind is from the NE from the open ocean bringing air masses from the tropical
Atlantic and from the African continent. The main laboratory is housed in a converted 40ft
shipping container, air conditioned with instruments to carry out the measurements. A
second 20ft container is available for ancillary use. The station can be accessed by an
unpaved dead-end road. Local traffic is negligible. The site covers an area of about 1200
m² which is fenced and guarded (24 h during weekend, 6 pm to 6 am during the week).
The station has a 10 m and a 30 m tower with scaffolds, a storage container and an (Fig.
3-2).
Location of the Cape Verde Atmospheric Observatory within the Islands of Cape Verde
Fig. 3-1: Map showing the location of Cape Verde Islands and the location of the Cape Verde Atmospheric Observatory. (http://ncasweb.leeds.ac.uk/capeverde/index.php?option=com_content&view=article&id=48&Itemid=55)
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 6
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
The station is part of a bilateral German-UK initiative for long-term observations, ground-
and ocean-based, in the Eastern North Atlantic Ocean region which is quite tropical. It
links several programmes such as the international programme SOLAS, the EU-funded
TENATSO (Tropical Eastern North Atlantic Time-Series Observatory) project, and the
German SOPRAN (Surface Ocean Processes in the Anthropocene) project.
The CVO was set up to advance understanding of climatically-significant interactions
between the atmosphere and ocean. Table 3-1 gives an overview of parameters
assessed at CVO including the timescale of their measurements.
Table 3-1: Measurement program at the Cape Verde Atmospheric Observatory.
Measurement Species Instrument Timescale of measurements
Ozone UV Absorption TEI 49c October 2006 – present
CO UV Fluorescence, Aerolaser 5001 October 2006 - present
NO, NOx, NOy Chemiluminescence, Air Quality Design Inc October 2006 - present
Meteorological data at 4m, 10m and 30m Automatic Weather Station October 2006 - present
NMHCs, DMS, OVOCs Dual channel GC-FID October 2006 - present
Halocarbons GC-MS May 2007 - present
Chemical characterisation of aerosol PM10, PM2.5, PM1
HiVOL Sampler, Digitel DHA-80 subsequent AAS, PIXE, thermographic methods, CE-MS, GC-MS….
November 2006 - present
Fig. 3-2: Picture of the Cape Verde Atmospheric Observatory Cape Verde, dominated by the 30m tower for air sampling. (http://ncasweb.leeds.ac.uk/capeverde/)
30 m tower
10 m tower
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 7
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
Measurement Species Instrument Timescale of measurements
Physical characterisation of aerosol DMPS (3-900 nm) November 2006 - present Physical characterisation of aerosol 5 stage Berner impactor May 2007 - present
BrO/IO/NO LP-DOAS November 2006 - June 2007
BrO/IO/NO MAX-DOAS December 2006 - present
JO1D Radiometer Jan-Feb 2007 May 2007-present
CH4, CO2, N2O, CO, SF6 ratio of O2/N2, and Ar/N2 isotope ratios of 13C/12C and 18O/16O of CO2 and O2
Flask sampling January 2007 - present
Cloud cover Ceilometer January 2008 - present Leachable Fe, Mn, Al, NO3, PO4, amino acids, total metal concentration
Filters, XRF November 2007 - present
Solar radiation Spectral Radiometer April 2008 - present
O2, N2, SF6, CH4, N2O Gas Chromatograph June 2008 - present
CO2 NDIR June 2008 - present
O2 O2 Paramagnetic instrument June 2008 - present
Comment The Cape Verde Atmospheric Observatory fulfils the infrastructural requirements
needed for a global GAW station. The GAW laboratory and the adjacent rooms are
clean, orderly and in good shape. They provide ample space for the instruments and
related infrastructure. All support devices, such as gas cylinders, are kept in the same
environment. There is currently no need for suggesting modifications.
3.3 Staff / Operators The responsibility for the trace gas monitoring programme at the CVO GAW station lies in
the hands of Lucy Carpenter and Alastair Lewis who have the function of the station
directors. Luis Silva Mendes Neves works as a station manager. Katie Read as the Site
Research Scientist served as station contact for the audits. Contacts and positions of
these people who where all involved in the audit are summarised in Table 3-2. The audits
were performed by Rainer Steinbrecher and Stephan Thiel from the WCC-VOC.
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 8
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
Table 3-2: Staff responsibility for the VOC-measurements at the Cape Verde Atmospheric Observatory
Name Position and duty Telephone E-mail
Katie Read Responsible Site Research Scientist +44 190 443 2565 [email protected]
Luis Silva Mendes Neves Station Manager +238 93 2822 [email protected]
Lucy Carpenter & Alastair Lewis Station director +44 190 444 2526 [email protected];
Luis Mendes gained his professional experience within 4 years of work at CAO, dated by
the time of the audit. During this time his knowledge has been supported by Katie Read as
well as the station directors working with VOC since several years.
Comment It is recommended, that Luis Mendes is going to join a GAWTEC-training course to
become more familiar with GAW.
3.4 Monitoring Set-up and Procedures
3.4.1 Air Inlet System for VOC The air-inlet for gas sampling is installed at 10 m above the ground and free accessible at
360° with no other tubing around. The tubing is made out of stainless steel. Prior to VOC
analysis the sampling air passes a TEFLON in-line filter (SAVILLEX; i.d. 5 cm). The
ambient air is sucked through the manifold with a KNF Neuberger pump at a flow rate of
100 ml/min. Further downstream the ambient air is dried to a dew point of -28 °C using two
condensation traps. The condensation traps are changed on a weekly basis.
3.4.2 Gas-chromatographic System The gas-chromatographic (GC) system for the analysis of NMHC is carried out with an
Agilent 5890, which was installed in 2006. The GC is equipped with two flame ionisation
detectors (FID) from Agilent. The detectors work at a temperature of 250 °C with no
additional make-up gas.
For the separation of VOCs in air samples, two capillary column are installed which are
used in parallel without a pre-column. The specifications of the two separation columns are
listed in Table 3-3.
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 9
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
Table 3-3: Used separation columns for the VOC analysis at CVO and their characteristics.
Type / Manufacturer
Installed since
Dimension (length, i. d.) Characterisation / Packing
“Al2O3 PLOT Column” Agilent 2005 50 m
0.53 µm Al2O3/KCL-deactivated PLOT column > separates hydrocarbons mainly in
terms of increasing boiling point
“LOWOX Column” Agilent 2005 10 m
0.53 µm High polar column, separates o-VOC mainly in terms of increasing polarity
For both columns the same temperature program is applied during VOC analysis and
detailed in Table 3-4.
Table 3-4: Temperature program for both columns used in VOC analysis at CVO.
Column Temperature program
PLOT Column & LOWOX Column
40 °C for 18.5 min 40 °C to 110 °C at a rate of 13 °C/min 110 °C to 200 °C at a rate of 8 °C/min
200 °C for 25 min 200 °C to 40 °C in one step > 40 °C = standby temp.
As carrier gas Helium (He) is in operation (purity of 99.999 %) and is treated with an
“Agilent purifier” before it enters the GC. The carrier gas flow rate is 22.3 ml/min per
column. The injection of the gas-sample onto the separation columns is achieved by a
“Programmed Temperature Vaporizing” (PTV) Inlet. Sketches of the instrument set-up
both for sampling and analysis mode are shown in Fig. 3-3.
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 10
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
Fig. 3-3: Sketch of the VOC-measurement set up during sampling and analysis mode used at CVAO.
For the gas sampling, two switching valves (VALCO) with each 6 ports are employed
(indicated in Fig. 3-3 as “autosampler valve” and “sampling valve”). The valves are
controlled by a VALCO-valve units. The tubing diameter of the whole set-up in general is
1/16” and all valves are generally heated to 40 °C. Fig. 3-4 shows the sampling valve
configuration during sampling mode respectively during analysis mode.
Fig. 3-4: Configuration of the sampling valve in sampling mode (left) and in analysis mode (right) which is applied during VOC-analysis at CVAO.
The switching of all the valves as well as the start and stop signals for analysis are
handled with electronic signals from a control software (Agilent). The GC-settings are
Sampling Mode
Helium out
Helium in
Sample in
Sample out
Inject or bot t om
Inject or t op
Sampling Mode
Helium out
Helium in
Sample in
Sample out
Inject or bot t om
Inject or t op
Aut osampler Aut osamplervalve
Condensat ion t rap
MFC4
Inject or boxMFC2
Sampling mode
Pump
MFC3Dilut ant flow
N2 in
Analysis mode
Inject or box
He in
2-posit ion Sampling valve
2-posit ion Sampling valve
FID
FID Comput erColumns
Aut osampler Aut osamplervalve
Condensat ion t rap
MFC4
Inject or boxMFC2
Sampling mode
Pump
MFC3Dilut ant flow
N2 in
Analysis mode
Inject or box
He in
2-posit ion Sampling valve
2-posit ion Sampling valve
FID
FID Comput erColumns
Helium in
Analysis Mode
Sample in
Sample out
Inject or bot t om
Inject or t op
Helium out
Helium in
Analysis Mode
Sample in
Sample out
Inject or bot t om
Inject or t op
Helium out
Analysis Mode
Sample in
Sample out
Inject or bot t om
Inject or t op
Helium out
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 11
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
coordinated by the software “CHEMSTATION” (Agilent). This software also allows a fully
automated run of the whole procedure of VOC analysis. Fig. 3-5 shows an example
chromatogram resulting from the analysis of a laboratory VOC standard with the above
described system set-up and parameters.
Fig. 3-5: Example chromatogram of a VOC standard analysed at CVO.
Comment The GC system represents state-of-the-art instrumentation and is well suited for high-
quality VOC measurements. This is proofed by the chromatogram shown in Fig. 3-5
with its good separation of the GAW NMHC target components (see Table 5-4 in the
appendix).
3.5 Operation and Maintenance
3.5.1 General The Cape Verde Atmospheric Observatory is usually manned by one permanent
employee. This person is mostly present during working hours from 9:00 am to 5:00 pm on
working days.
m in20 25 30 35 40
pA
10
20
30
40
50
60
F ID1 B , (A P R06_4\28040025.D)
20
.11
7 -
m
eth
an
e
20
.56
3 -
e
tha
ne
21
.08
8 -
e
the
ne
21
.86
1 -
p
rop
an
e
22
.95
9 2
3.1
91
23
.40
6 -
p
rop
en
e 2
3.6
92
24
.05
7 -
is
o-b
uta
ne
24
.31
7 -
n
-bu
tan
24
.50
4 2
4.5
86
-
ace
tyle
ne
24
.67
9
25
.24
2
25
.71
1 2
5.7
91
-
tra
ns-
2-b
ute
ne
25
.93
7 2
6.0
06
-
1-b
ute
ne
26
.27
8 -
is
o-b
ute
ne
26
.50
0 -
ci
s-b
ute
ne
26
.91
1 -
cy
clo
pe
nta
ne
27
.08
2 -
is
o-p
en
tan
e 2
7.3
97
-
n-p
en
tan
e 2
7.4
71
28
.05
2 2
8.1
48
28
.59
9 -
cy
clo
pe
nte
ne
28
.70
9 -
tr
an
s-2
-pe
nte
ne
29
.01
0 -
2
-me
thyl
-2-b
ute
ne
29
.22
4 -
2
-me
thyl
1-b
ute
ne
29
.31
0 -
1
-pe
nte
ne 2
9.4
95
-
cis
2 p
en
ten
e 2
9.8
08
30
.05
4 3
0.1
23
30
.23
6 -
2
,2 d
ime
thyl
bu
tan
e 3
0.4
26
-
2,3
dim
eth
yl b
uta
ne
30
.54
5 -
2
-me
thyl
pe
nta
ne 3
0.9
12
-
he
xan
e 3
1.2
55
-
iso
pre
ne
31
.91
2 32
.11
0 3
2.1
93
32
.45
8 3
2.5
92
32
.70
7 3
2.8
78
33
.19
5 3
3.4
28
33
.58
2 -
2
,3 d
ime
thyl
pe
nta
ne
33
.84
2 3
3.9
32
-
2-m
eth
yl h
exa
ne
34
.20
2 -
3
me
thyl
he
xan
e 3
4.3
94
-
he
pta
ne
34
.70
5 3
4.9
40
35
.09
3 3
5.2
00
35
.45
4 -
b
en
zen
e 3
5.7
66
35
.93
0 3
6.0
87
36
.29
1 3
6.5
40
36
.69
2 3
6.8
14
-
2,2
,4 t
rim
eth
yl p
en
tan
e
37
.29
4 3
7.4
06
37
.59
8 -
2
,3,4
tri
me
thyl
pe
nta
ne
37
.70
8 -
3
me
thyl
he
pta
ne
38
.39
9
38
.99
3 -
n
-oct
an
e 3
9.1
53
40
.53
5 -
to
lue
ne
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 12
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
At the time of the audit there was no opportunity for remote instrument control.
3.5.2 Sampling and Calibration All tubing, gas cylinder lines and connections are regularly checked for leaks. At the time
of the audit all the tubings and connections were in orderly and tight manner.
3.5.2.1 Sequence of Injections
Standards, blanks and samples are usually analysed in the following order during normal
operation at CVO:
• 2 Blanks
• 2 laboratory standards
• 2 Blanks
• 60 air samples
• 2 blanks
• 60 air samples
Ambient air samples are taken every hour for 10 min. The calibration of the instrument with
the laboratory standard works automatically and is performed every analytical cycle of 120
air samples (approx. every 5 days). The laboratory standard for this has been in use since
April 2006. For calibration procedures a laboratory standard is in operation, a working
standard is not in use at CVO.
3.5.3 Zero and Repeatability Checks, Target Gas A blank check is automatically performed at CVAO every 2.5 days, every 60 samples.
Pure nitrogen is used for this and is provided from a compressed gas cylinder.
The repeatability of measurements is checked regularly, analyzing 3 identical laboratory
standard samples. The last repeatability check yielded a standard deviation smaller than
5 % for the relative peak area (with n = 3). The obtained results are within the control limits
of the DQOs for precision of 5 %.
To observe the linearity of the detector response as well as the consistency of the
response curve an abundance range test is performed once a year. For this test 6
concentration levels of a standard are measured in dependence of the sample volume.
The last abundance range test performed resulted in a linear regression coefficient (r2) of
1.00 for a 6 level calibration after a linear fit was done. This result lies well in the range of
the control limits were a r² down to 0.95 is acceptable.
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 13
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
Comment Results of the different checks indicate a very good analytical system performance.
The recommended checks are carried out on a regular basis.
3.5.4 Maintenance Maintenance of the VOC analyzing system is normally conducted as a bimonthly
inspection. Apart from this, the inlet filter is changed on a monthly basis and the
condensation trap on a weekly basis. Gas cylinders are changed on demand and the
instrument is checked daily for characteristic parameters. Other maintenance operations
are only carried out upon demand (e.g. indicated by identified leakages, retention time
shifts, peak shape alterations and so on).
3.5.5 Corrective Actions In case of instrument drift no corrective action are made at CVO. In case of instability,
baseline noise and/or peak shape degradation the instrument is checked to identify the
reason for the instabilities. Data obtained with an obvious malfunction of the instrument are
flagged. After instrumental malfunction or repair, the routine operation is continued after
additional checks for consistency of the response factors and new calibration of the
instrument.
3.6 Standards
3.6.1 Regulators and Connections The pressure regulators used on the high-pressure cylinders are two-stage and out of
stainless steel. The tubing from the cylinders to the valve consist of stainless steel 1/8”.
3.6.2 Zero Air For Blank checks pure Nitrogen (purity 99.999 %) is used as Zero Air at CVAO, supplied in
compressed gas cyclinders.
3.6.3 Laboratory Standards The laboratory standard is stored in the laboratory beneath the GC. The standard is
manufactured by Apel Riemer (S/N: CC 236351) containing 54 NMHC compounds with a
mol faction range from 0.5 to 12.00 µmol/mol and a specified uncertainty of + 5-7 % as
shown in Table 3-5. Since November 2008 a NPL Ozone precursor mix calibration
standard D838940 was used which was then replaced in February 2012 with D860619.
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 14
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
Acetylene was not stable in the Apel Riemer calibration standard and data had to be
adjusted accordingly.
Table 3-5: Laboratory Standard from Apel Riemer (S/N: CC 236351)
Compound Concentration [ppmv]
Uncertainty [%] Compound Concentration
[ppmv] Uncertainty
[%] ethene 5.35 7 3-methylpentane 1.13 5
acetylene 6.48 7 2-methyl-1-pentene 0.83 5 ethane 11.83 7 hexane 2.85 5
propene 2.19 5 t-2-hexene 0.46 5 propane 10.56 5 c-2-hexene 0.87 5 propyne 4.86 5 methylcyclopentane 0.89 5
i-butane 4.73 5 2,4-dimethylpentane
0.9 5
i-butene 3.14 5 benzene 1.86 5 1-butene 2.34 5 cyclohexane 0.5 5
1,3-butadiene 2.28 5 2-methylhexane 0.9 5
butane 9.05 5 2,3-dimethylpentane
0.44 5
t-2-butene 1.06 5 cyclohexene 0.82 5 c-2-butene 2.26 5 3-methylhexane 0.85 5
1,2-butadiene 5.39 5 1-heptene 2.13 5 i-pentane 6.87 5 heptane 4.02 5 1-pentene 1.1 5 methylcyclohexane 0.94 5
2-methyl-1-butene 1.12 5 2,3,4-trimethylpentane
0.45 5
pentane 8.13 5 toluene 2.72 5 Isoprene 4.42 5 2-methylheptane 0.46 5
t-2-pentene 0.88 5 4-methylheptane 1.11 5 c-2-pentene 2.11 5 3-methylheptane 0.89 5
2-methyl-2-butene 0.92 5 octane 0.49 5 2,2-
dimethylbutane 2.16 5 ethylbenzene 0.9 5
cyclopentene 0.89 5 m-xylene 1.32 5 cyclopentane 0.94 5 p-xylene 0.52 5
2,3-dimethylbutane
1.65 5 styrene 0.49 5
2-methylpentane 0.91 5 o-xylene 0.47 5
The stability of this standard is guaranteed for one year by the manufacturer. It has been
compared with other standards from Apel Riemer on a regular basis (every 1 – 2 years) as
well as its regularly re-calibrated by the CCL.
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 15
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
3.6.4 Working Standards A working standard is currently not in use at the Cape Verde Atmospheric Observatory.
3.6.5 Calibration of o-VOC and DMS VOC with oxygen (o-VOC) and dimethylsulfid (DMS) are calibrated by means of a
permeation device. The permeation tubes containing the VOCs are kept in heated blocks
(40 °C). The containers with the tubes are flushed at a constant flow rate 100ml/min. The
weight loss of the tubes is determined every two-three months and the emission rates are
calculated. 5-point calibrations are carried out to determine detector response with toluene
as a check of the technique. The toluene response of the FID from the permeation system
is then compared to that of the laboratory standard. At present no reference standards are
available as a CCL is not established yet.
3.7 Data Acquisition and Processing
3.7.1 General Data acquisition of the gas chromatographic signals and parameters is carried out using
the “Agilent” software “CHEMSTATION A.10.02”. As instrument parameters the Sample
flows in and out and the cold trap temperatures are recorded. The analysis of the
measured sample as well as the calibration is also automatically carried out by the
software “CHEMSTATION”.
The quality of measurements is normally assessed by manual visualisation by the operator
of the GC-System. The instrument logbook is also considered as part of the data validation
process as possible bad data resulting from instrument failure can be backtracked there.
Data filtering is based on the laboratory temperature. Outliners are then manually identified
and flagged in the database. As a reference for this “flagging” the detection level of the GC
(< 5 pptv) is used.
Time series plots for different time intervals are also available for long-term observation of
the system performance at the CVO.
The final data validation is done by Katie Read in York.
Comment Data acquisition of system parameters and data processing reflects the state-of-the art
at CVO.
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 16
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
3.7.2 Chromatogram Evaluation The chromatogram evaluation in the first step is performed with “CHEMSTATION”. Each
peak in the chromatogram is labelled with the retention time and its name. GAW VOC
target compounds analysed show a very good separation from other compounds (Fig. 5-
1). If partial overlap occurs, an automated vertical split of the peaks to the baseline is
carried out. The peak integration of all peaks is always performed manually. Major
interferences have so far only been observed with n -pentane, and some other non GAW
target compound (see Fig. 3-5).
The chromatograms itself as well as the analysis report is stored. The report includes
additional information of the peak area of each detected peak and its height. All these
parameters are used for regular data quality control and the chromatograms are regularly
inspected by an operator. Typical values for chromatogram-characteristics are listed in
Table 3-6.
Table 3-6: Typical chromatogram characteristics obtained from the analysis of VOC-flask-samples at CAL Boulder.
Chromatogram characteristic Typical value
Peak Area [pA x min] 2 - 100
Peak Hight [pA] 20-25
Baseline noise level [pA] 0.02
The calculation of the reported mole fractions is based on the peak area of the detected
analyt. A baseline reset either manually or automatically has not yet been necessary.
Column regeneration is performed every 30 runs through blank runs (see section 3.5.2.1).
Comment The chromatogram evaluation is performed in a highly sophisticated way at CVO.
3.8 Data Management and Submission There are provisions for redundant, off-site data storage at the Cape Verde Atmospheric
Observatory. A computer back up on a hard drive is performed every week with two
copies. One is send to York and one stays at the site. The VOC data sent to the World
Data Center of Greenhouse Gases (WDCGG)and available from the corresponding
website (http://ds.data.jma.go.jp/gmd/wdcgg/).
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 17
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
Comment The data backup policy and data submission to the WDCGG of the CVO is exemplary.
3.9 Documentation
3.9.1 Technical and QA/QC As part of the system audit, the documentation for QA/QC-procedures was reviewed.
Instrument manuals are available at the laboratory.
Field logbooks were kept in the laboratory and were found to be in orderly manner. The
books were kept both hand-written and electronically. At the time of the audit, they were
found to be up-to-date. Instrument logbooks were also available in the laboratory. Now
central electronic logbook where all maintenance is recorded is maintained and can be
accessed from anywhere using a username and password.
Comment It is highly recommended to have the WMO GAW Measurement Guidelines also
available at the station as the CVAO works as a global GAW station.
Additional to the field logbooks there are standard operating procedures (SOP) available
for the use of the instruments in the laboratory as well as a check list for inspection which
is controlled every day.
Comment The existence of SOP’s for the correct measurement is quite important and well done
by the lab.
At the CVO QC data forms or control charts as well as QC and field shields were not used
by the time of the audit. QA/QC at the time of the audit consisted of thorough manual
checking of the data, removal of outliers, checking and application of blank data, removal
of data where instrument was suspect. For checking detector response stability the
benzene and toluene peaks on both columns are checked for changes in sample volume
and consistency of the system. As a site QA coordinator works Alistair Lewis and he
normally reviews data summaries.
Comment QA/QC Procedures at CVO is in orderly manner.
3.9.2 Reports of Results
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 18
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
A list of publications is available on the corresponding Gawsis Web-page
(http://gaw.empa.ch/gawsis/) or from the CVO web-site
(http://www.ncas.ac.uk/index.php/en/cvao-publications).
3.10 Intercomparison of VOC Standards
3.10.1 Experimental Procedure Three travelling standards of the WCC-VOC and dedicated pressure regulators were sent
to CVO. They were stored at the same location as the in-house standards to avoid any
temperature influences. Flushing and leak checks were performed after connecting the
standards to the analytical system.
The GC used for analysis of VOC wasn’t modified prior to inter-comparison experiments.
The inter-comparison experiment itself involved the laboratory standard of the station as
well as the three WCC-VOC travelling standards listed in Table 5-6 (part of the appendix).
Each measurement of each standard was repeated five times. All data processing was
performed by the programme manager at the station. Determined mole fractions for each
individual component according to the analysed standard were submitted to the WCC-
VOC including their measurement uncertainties. For the determination of the
measurement uncertainties the error of the quoted response, the repeatability of
calibration measurements within triplicate samples, the repeatability of sample
measurements within 5 samples, the stability of the response, changing blank peak values
as well as the error on peak integration were taken into account by the CVO.
3.10.2 Results of the VOC Intercomparison The individual results reported were evaluated using the WCC-VOC tertiary standard and
the two laboratory standards calibrated to the WCC-VOC tertiary standard.
Prior to sending the travelling standards to the laboratory the WCC-VOC recalibrates all
the standards with its in-house GC system and repeated after the standards return from
the station. All calibrations at WCC-VOC are repeated five times and the mean of these
values is taken as the target concentration. The precision is calculated from the standard
deviation of five measurements multiplied by two (coverage factor 2) to achieve an
extended uncertainty range of approx. 95%. The absolute error of the target concentration
of a compound in the laboratory standards is then calculated taking into account
instrument precision and the specified error of the respective compound in the secondary
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 19
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
standard. The mean value of the calibrations for each target compound performed before
the audit and after the audit is then taken as a reference concentration with its respective
error for evaluating the analysis results of the audited laboratory.
It is remarked that only the GAW-NMHC-target components (see section 5.2) as defined in
the WMO Report 171 (http://www.wmo.int/pages/prog/arep/gaw/gaw-reports.html) are
evaluated even if the laboratory reported concentrations for more than those components.
The following diagrams show the results of the evaluation of the results reported after the
audit at CVO. The black dots with the respective error bars indicate the difference from the
reported concentration to the one determined at the WCC-VOC in percent. The bright
green lines stand for the standard’s extended uncertainty (approx. 2σ). As marked in the
graphs the red lines represent the DQOs for the GAW-VOC network.
Results of the direct analysis of the travelling standards
Fig. 3-6: Comparison of the results of WCC-VOC Standard D292363 analysed by CVAO referred to the guaranteed concentrations of the standard.
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 20
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
Fig. 3-6 shows the results of the evaluation of the reported results from the analysis of
travelling standard D292363 at the Cape Verde Atmospheric Observatory.
The analysis of the travelling standard D292363 at CAL-GMD produced quite good results.
The determined concentrations of all components meet the respective DQOs for accuracy.
Apart from Toluene all the components even matched the standards concentrations within
the error limits of the target concentrations. These results indicate that the system is
excellently suitable for VOC measurements in the 2 nmol/mol range. Only the reported
errors for i-pentane, n-pentane and isoprene seem to be relatively high compared to the
ones of the other components. The reason for this may be some small partial overlap with
other compounds eluting nearby (see Fig. 3-5).
Fig. 3-7 displays the results of the evaluation of the reported results from the analysis of
travelling standard D336442 at CVO.
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 21
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
The results of the analysis of travelling standard D336442 are nearly as good as the ones
from standard D292363. All determined concentrations for the GAW target components
easily meet the DQOs. The concentrations for all these components also nearly match the
standards concentrations within their error limits. It has to be mentioned that the mol
fractions of the components in this standard are up to 15 nmol/mol. This is about seven
times the concentration compared to the other standard D292363. The good results for the
analysis of this standard indicate that the system is suited to work quite well also for high
concentration air samples.
The results of the analysis of travelling standard D336417 are shown in Fig. 3-8. It has to
be noted that this standard contains isoprene below the detection limit of the WCC-VOC
system (0.02 nmol/mol) and is therefore missing in the following graphic.
Fig. 3-7: Comparison of the results of WCC-VOC Standard D336442 analysed by CVAO referred to the recalibrated concentrations of the standard by the WCC-VOC.
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 22
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
The correct determination of the target mol fractions of travelling standard D336417 is
challenging as they are quite low (0.1 to 1.1 nmol/mol). However, CVO performed quite
well even for this standard. Only the determined concentration for n-butane and toluene
are outside the DQOs. For all other components the determined concentrations match the
standards concentrations quite within the DQOs or even within the error limits. It is
remarked that the reported errors in this analysis are not remarkably higher than in the
analyses of the other (higher concentrated) standards.
Comment Inter-comparison experiments for the analytical system reveal a very good
performance. Only the repeatability of the measurements of the lager NMHCs (mainly i-
pentane, n-pentane and isoprene) could be a little better.
Fig. 3-8: Comparison f the results of WCC-VOC Standard D336417 analysed by CVAO referred to the recalibrated concentrations of the standard by the WCC-VOC.
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 23
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
4 REFERENCES
WMO, Global Atmosphere Watch, (2007), Report No 171: A WMO/GAW Expert Workshop on Global Long-Term Measurement of Volatile Organic Compounds (VOCs).
5 APPENDIX
5.1 Individual AnaIysis Results Table 5-1: Results of the analysis of travelling standard D292363
reported by CVAO to the WCC-VOC.
Analyt Result (nmol/mol)
Measurement uncertainty (95% confidence)
(nmol/mol) ethane 2.71 0.17 ethene 2.69 0.33
propane 2.67 0.17 propene 2.61 0.18
iso-butane 2.68 0.21 n-butane 2.6 0.18 acetylene 2.63 0.13
trans-2-butene 2.59 0.22 1-butene 2.53 0.24
cis-butene 2.53 0.18 iso-pentane 2.57 0.29 n-pentane 2.6 0.49
hexane 2.58 0.48 isoprene 2.58 0.42 benzene 2.63 0.17 toluene 2.38 0.23
Table 5-2: Results of the analysis of travelling standard D336442 reported by CVAO to the WCC-VOC.
Analyt Result (nmol/mol)
Measurement uncertainty (95% confidence)
nmol/mol ethane 12.7 0.79 ethene 7.23 0.89
propane 11.97 0.77 propene 2.27 0.16
iso-butane 6.18 0.49 n-butane 10.66 0.76 acetylene 7.7 0.40 cis-butene 3.54 0.25
iso-pentane 7.8 0.88 n-pentane 9.34 1.75
hexane 3.24 0.61 isoprene 5.39 0.88 benzene 2.23 0.18 toluene 2.34 0.18
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 24
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
Table 5-3: Results of the analysis of travelling standard D336417 reported by CVAO to the WCC-VOC.
Analyt Result (nmol/mol)
Measurement uncertainty (95% confidence)
(nmol/mol) ethane 1.13 0.07 ethene 0.81 0.10
propane 0.49 0.03 propene 0.27 0.02
iso-butane 0.54 0.05 n-butane 1.1 0.08 acetylene 0.9 0.05
trans-2-butene 0.11 0.01 1-butene 0.14 0.01
iso-pentane 1.5 0.18 n-pentane 0.44 0.08
hexane 0.16 0.03 benzene 0.36 0.02 toluene 0.59 0.10
Fig. 5-1: Ambient air chromatogram at CVO
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 25
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
Fig. 5-2: Ambient air mixing ratios of selected VOC at CVO in 2007. High VOC levels during a dust episodes.
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 26
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
5.2 WCC VOC target components The following table gives an overview for the VOC GAW-aim components with the agreed
data quality objectives for both accuracy and precision of the analysis. The agreed list of
these components would normally be longer but up to now it has only been possible to get
reliable, stable standards for these components. The other molecules are even more
reactive and furthermore the analysis of these components is a bit more difficult than the
ones listed in Table 5-4. Until it is not possible to analyze the other components as reliable
as the ones listed in Table 5-4 the WCC-VOC isn’t able to evaluate the measurement of
more of those compounds listed in the following table. The analysis of these components
respectively the results of their analysis is mainly taken into account when an audit is
appraised and the final evaluation of the audited station / laboratory is set. Table 5-4: VOC GAW target components with the agreed
data quality objectives for both accuracy and precision of the analysis.
Compound Accuracy Precision
Ethane 10% 5%
Ethyne 15% 5%
Propane 10% 5%
Iso-/n-Butane 10% 5%
Iso-/n-Pentane 10% 5%
Isoprene 20% 15%
Benzene 15% 10%
Toluene 15% 10% Mol fraction
< 0.1 nmol mol-1 ± 20 pmol mol-1 ± 15 pmol mol-1
5.3 WCC VOC Reference The WCC-VOC has one laboratory standard containing VOC mol fractions in Nitrogen.
The accurate concentrations of the single VOC’s are certified by NPL which works as a
central calibration laboratory. The composition of the standard as well as its uncertainty is
listed in Table 5-5. Compounds in bolt are part of the GAW aim components for VOC
analysis. This laboratory standard is itself used as a TS of the WCC-VOC.
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 27
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
Table 5-5: NPL-certified D292363 VOC standard at WCC-VOC. The cylinder contains VOC’s in pure nitrogen. The measurement of this standard is the basis for the recalibration of the travelling standards D336442 and D33617.
Compound Nominal
value /ppb
Uncertainty 2σ/ppb
Overall uncertainty 2σ
of analysis/ ppb ethan 2,7 0,05 0,08 ethene 2,67 0,05 0,07 ethyne 2,66 0,05 0,05
propane 2,67 0,05 0,11 propene 2,63 0,05 0,07 i-butane 2,68 0,05 0,05 1-butene 2,56 0,05 0,06
1,3-butadiene 2,63 0,05 0,05 n-butane 2,6 0,05 0,05
trans-butene 2,6 0,05 0,05 cis-2-buten 2,56 0,05 0,05 i-pentane 2,59 0,05 0,05 1-pentene 2,55 0,05 0,05 n-pentane 2,63 0,05 0,05 isoprene 2,6 0,05 0,05
trans-pentene 2,5 0,05 0,05 2-methyl-pentane 2,59 0,05 0,05
hexane 2,6 0,05 0,05 benzene 2,62 0,05 0,09
2,2,4-trimethylpentane 2,61 0,05 0,05 heptane 2,56 0,05 0,06 toluene 2,59 0,05 0,25 octane 2,59 0,05 0,05 nonane 2,49 0,05 0,06 α-pinene 2,01 0,06 0,06
Additional to this standard there are two more standards sent to the labs prior to an audit.
These two additional standards are regularly recalibrated by comparison with the analysis
results of the laboratory standard. Table 5-6 lists the nomenclature of all travelling
standards used at WCC-VOC as well as the approximated mol fractions of the VOC’s in
these standards.
VOC- Audit: Regional GAW Station – Atmospheric Observatory Cape Verde, November 2009 28
________________________________________________________________________________________________
WMO Global Atmosphere Watch World Calibration Centre for VOC’s
Table 5-6: Travelling standards of WCC-VOC used for measurements at the audit.
Cylinder Number Mol fractions of VOC’s [nmol/mol]
D292363 2,00 – 2,70
D336442 13,50 – 2,00
D336417 1,25 – 0,20
5.4 List of Abbreviations and Acronyms CVAO Cape Verde Atmospheric Observatory
DQO Data quality objective
FID Flame ionization detector
GAW Global atmospheric watch
GC Gas chromatograph
GMD Global monitoring division
KIT Karlsruhe institute of technology
m.a.s.l meter about sea level
MG-VOC Measurement guidelines for volatile organic compounds
N North
NMHC Non-methane hydro carbons
NPL National Physical Laboratory UK
o.d. outer diameter
QA/SAC Quality assurance / Science activity centre
SAG RG Scientific advisory group for reactive gases
TS Travelling standard
UBA Umweltbundesamt
W West
w/w weight per weight
WCC-VOC World calibration centre for volatile organic compounds
WDCGG World data centre for green house gases
WMO World meteorological organisation