Page 1: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

Surgical Technique

Page 2: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

Nota Bene: The technique description herein is made available to the healthcare professional to illustrate the authors’ suggested treatment for the uncomplicated procedure. In the final analysis, the preferred treatment is that which addresses the needs of the patient.

Design surgeon list

Smith & Nephew would like to thank the following surgeons who were instrumental in the design and refinement of the SMF™ Short Monolithic Femoral Hip System.

Kristaps Keggi, MD Yale Medical Center New Haven, CT Waterbury Hospital Waterbury, CT

Prof Francesco Biggi, MD Ospedale San Martino di Belluno Belluno, Italy Prof Rolf Haaker, MD St Vincenz Hospital Brakel, Germany Prof Norbert Linder, MD Brüderkrankenhaus St Josef Paderborn, Germany

Todd Swanson, MD Centennial Hospital Las Vegas, NV

Prof Francesco Falez, MD Ospedale Santo Spirito Rome, Italy

Prof Thomas Hess, MD Dreifaltigkeits-Hospital Lippstadt, Germany

Page 3: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

3

Table of contents

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

Preoperative planning . . . . . . . . . . . . . . . . . . . 5

Surgical technique . . . . . . . . . . . . . . . . . . . . . . 7

Stem specifications . . . . . . . . . . . . . . . . . . . . 13

Instrument layout . . . . . . . . . . . . . . . . . . . . . . 15

Catalog information . . . . . . . . . . . . . . . . . . . . 17

Important medical information. . . . . . . . . . . . 21

Appendix . . . . . . . . . . . . . . . . . . . . . . . . . . . .24

Page 4: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

4

The SMF™ system is a total hip system consisting of proximally porous coated titanium alloy (Ti-6Al-4V) tapered femoral hip stems with optimized and easy to use instrumentation. It offers the surgeon an improved bone conserving implant that is compatible with high demand femoral head bearing options from Smith & Nephew. The SMF hip stem is shorter than conventional hip stems, which allows the stem to conserve more of the patient’s femur than other prostheses. The SMF hip stem requires a slightly higher neck resection. The proximal porous coating of the stem consists of STIKTITE™ coating, a commercially pure Titanium coating, and is designed to have a press-fit relative to the broach of approximately 1/4mm. The stems are proportionally sized and shaped in sizes -1 through 6 (core sizes), with optional larger sizes 7 through 9 (extended sizes).

The SMF stem components are indicated for uncemented use in individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma, inflammatory joint disease such as rheumatoid arthritis, or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses such as osteoarthritis; avascular necrosis; traumatic arthritis; slipped capital epiphysis; fused hip; fracture of the pelvis; diastrophic variant; nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; femoral osteotomy, or Girdlestone resection; fracture dislocation of the hip; and correction of deformity.

Introduction

Page 5: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

5

The goal of preoperative planning is to determine the correct stem size, level of the femoral neck cut and proper head, and stem offset combination. Preoperative templating requires at least an anteroposterior (AP) radiograph of the pelvis and a lateral radiograph of the affected hip. If the opposite hip is unaffected by disease, it can often provide accurate sizing information for the femoral stem.

To determine if a patient has a leg length discrepancy, the AP radiograph should be used. Draw a line tangential to both of the Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial cortex of bone on both femurs. If the patient’s legs are of equal length, the line that has been drawn will contact both femurs at the same level. If the patient’s legs are of unequal length, the lines will contact the femurs at different levels along the femur, such as the bottom of the lesser trochanter. The distance between the line that has been drawn and the reference point on both femurs is measured. The difference in these measurements indicates the patient’s leg length discrepancy.

Nota Bene: The technique description herein is made available to the healthcare professional to illustrate the authors’ suggested treatment for the uncomplicated procedure. In the final analysis, the preferred treatment is that which addresses the needs of the patient.

Note: Using this method of templating for leg length discrepancy assumes the patient has a normal, symmetrical pelvis and has neutral limb positioning.

When determining which size SMF™ stem to use, the AP radiograph should be templated. (Make sure you are looking at a true AP X-Ray. If needed, template off the contralateral ‘normal’ hip.) Using the AP radiograph, place the femoral templates over the proximal femur of both the affected and unaffected hips. The junction of the lateral femoral neck and greater trochanter serves as a good reference point for placement of the X-Ray templates. Place a mark at this junction and in the center of the femoral head. Align the lateral shoulder of the prosthesis with the mark at the junction. Find the appropriate stem that fits and fills the proximal femur. Find the appropriate head option by choosing a head whose length and angle match the center of the femoral head (Figure 2).

Note: The neck resection as indicated on the X-Ray template indicates a resection approximately 1cm more proximal than femoral resections commonly used in conventional THAs.

Preoperative planning

Anteroposterior radiograph demonstrating proper X-Ray templating

Figure 2

Anteroposterior radiograph demonstrating leg length inequality

Figure 1

Page 6: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

6

Note: The SMF™ stem was designed such that the lateral aspect of the stem should be in direct contact with lateral aspect of the femur as shown in Figure 3. The design of the implant is such that the distal medial aspect of the stem does not contact the medial side of the femoral cortex. The stem, when placed in the neutral position, appears to be in slight varus relative to the proximal femoral canal axis (see Figure 4). The current X-Ray templates provide both an extended lateral cortex line and femoral canal axis line to aid in positioning the femoral implant relative to the patient X-Ray during preoperative planning.

If the implant is positioned such that the distal tip is centered in the femoral canal, the lateral aspect of the stem will not be against the lateral cortex, and there is a strong possibility that the stem will rotate until it makes contact with the lateral cortex.

Lateral cortex

Femoral axis

Preoperative planning

Figure 3

Figure 4

Page 7: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

7

Surgical technique

The SMF™ stem hip may be implanted through all typical surgical approaches for total hip arthroplasty.

Before surgery, review instrument sets to ensure all instruments are present and working properly.

Femoral osteotomy

Note: The SMF hip stem is designed with an approximately 5–10mm higher neck cut than typical stems, preoperative templating of the proper neck resection level to achieve appropriate leg length is important.

Assemble the osteotomy guide to the broach anteversion handle. Reference marks are provided on the guide in the medial arc for referencing the lesser trochanter and above the proximal shoulder for referencing the tip of greater trochanter. Align the distal axis of the guide with the center of femur and use the reference marks to position the guide to the location identified during templating. The point of the femoral neck resection can be marked with electrocautery corresponding to both the preoperative templating and the intraoperative measurement.

Osteotomize the femoral neck.

Figure 5

Page 8: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

8

Prepare acetabulum

If acetabular reconstruction is required, prepare the acetabulum using the surgical technique for the intended Smith & Nephew acetabular component.

Femoral canal preparation

Use the box osteotome and canal finder for initial entry into femoral canal.

Sink the canal finder to the proof mark as shown in Figure 7.

Note: In order to achieve proper stem alignment, it is important to stay lateral with the box osteotome and the canal finder. Care should be taken to ensure that the initial reaming tract is in neutral alignment with the femoral axis, ie along the lateral side of the canal (Figure 7).

Figure 7a

Figure 7b

Figure 6

Page 9: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

9

Notch

Broach assembly/disassembly

Smith & Nephew makes available a variety of broach handles to accommodate the various surgical approaches to the hip. All of the broach handles listed utilize the same connection to Smith & Nephew standard broach post and are assembled and disassembled as described below.

Important: Broach handles, as well as all surgical instruments used for the implantation of the SMF™ system, must be cleaned and sterilized before every case in accordance with facility policy and regulations.

Assemble the broach to the broach handle by placing the broach post in the clamp as shown in the figure to the right. Verify that the notch in the broach post is properly aligned with the handle (Figure 12). Use the thumb to lock the clamp onto the broach (Figure 14). A modular anteversion handle can be assembled to the broach handle to provide version control. Disassemble the broach from the broach handle by lifting the lever to release the handle from the broach post.

Caution: Failure to properly align the broach before closing the handle will damage the handle. Damaging the handle may cause extended surgical time or may force the user to use another broach handle that is not as well suited for the surgical approach.

Straight Broach Handle 7136-4021

Figure 8

Single Offset Broach Handle 7136-5703

Figure 9Double Offset Broach Handle 7136-0090 (Right)

Figure 11

Double Offset Broach Handle 7136-0089 (Left)

Figure 10

Figure 12

Figure 13

Figure 14

Surgical technique

Page 10: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

10

Femoral broaching

Start the broaching procedure with the smallest available starter broach. Sequential broaching should then be carried out to the templated stem size. Note that the starter, size -1, size 0 and size 1 broaches all have more aggressive teeth to aid in the initial broaching process. Taking care to preserve the greater trochanter, the starter broach can be used to rasp laterally beneath the greater trochanter. The lateralization of the broaches is important in achieving the aforementioned neutral alignment as shown in Figure 15a. Be sure to check the stability of the broach rotationally, medially and laterally. When broaching keep the version constant. Stop broaching only when the desired level or stability is achieved. It is important to maintain broach rotation due to the rectangular geometry of the implant. For proper sizing and alignment of the broach, it is often necessary to remove bone in the posterior-lateral area of the neck as shown in Figure 15b.

Note: Care should be taken not to force a broach that is too large into the femur. Consideration should be given to using a stem one size smaller than the size templated if the final broach is difficult to seat. This helps avoid intraoperative fractures of the femur.

Note: Optional use of the broach anteversion handle may be helpful for broach/stem extraction when using the double offset broach handles.

Calcar preparation (optional)

With the final broach fully seated, remove the broach handle. Place the calcar reamer over the post of the broach and machine the femoral neck, ensuring alignment to avoid femoral fracture (Figure 16).

Figure 15a

Figure 16

Figure 15b

Posterior lateral neck removal

Page 11: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

11

Initial Trial reduction

Broach trial necks, both standard and high offset, are provided to aid in trial reduction before the final stem implant is inserted.

Place the broach trial neck as determined by preoperative templating onto the broach post. Select the trial femoral head of desired diameter and neck length and place onto the trial neck (Figure 17). Reduce the hip and re-measure leg length. Compare to previous measurements recorded from preoperative templating or leg length before dislocation. Adjustments in neck length and/or offset can be made at this time. If trialing for a Unipolar or Bipolar, trial according to the appropriate technique for the selected device.

Trial reduction Reduce the hip and evaluate in the following ways:

Soft tissue tension Some shuck is normal when applying a longitudinal distraction force to the hip. Shuck should not be excessive and the hip should not dislocate in straight traction (Figure 18).

Anterior stability Place the leg in full adduction and hyperextension, while exerting an external rotation force. If the hip cannot be fully extended, it may be too tight. If it dislocates easily, it is too loose and impingement must be addressed or component malposition exists (Figure 19).

Posterior stability Place the leg in adduction and 90° flexion. Gradually rotate internally. The hip should be stable with 45° of internal rotation. If it dislocates with minimal internal rotation, it is too loose and impingement must be addressed or component malposition exists (Figure 20).

Sleep position Place the leg in the ‘sleep position’ with the operated leg semiflexed, adducted and internally rotated over the other leg. Apply axial force to try and dislocate the hip. This position represents a dangerously unstable position that may be adopted by a patient sleeping on their nonoperated side. Once a well-balanced hip has been created with the appropriate trial head and neck, remove the broach (Figure 21).

Figure 17a

Figure 17b

Soft tissue tension

Figure 18

Sleep position

Figure 21

Surgical technique

Anterior stability

Figure 19

Posterior stability

Figure 20

Page 12: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

12

Stem Insertion

Assemble the stem punch to the broach handle by placing the stem punch post in the clamp (Figure 22). Use the thumb to lock the clamp onto the stem inserter/extractor.

Seat the stem in the femur by hand and use the punch to impact the stem in the driver dimple with gentle mallet blows to seat the stem to the level of the neck resection (Figure 23). Check the stem stability. If the stem has stopped moving with gentle mallet blows and is not completely seated, remove the stem and repeat the same size broaching steps.

Proceed to Optional Final Trial Reduction with the trial femoral heads.

Caution: Do not use excessive force to seat the stem as this may lead to femoral fractures.

Stem Removal

To remove a fixed neck SMF™ Stem, utilize a flexible osteotome as you would in any stem removal procedure. Place the stem punch under the medial taper as shown in Figure 24. Located there is a depression, into which the stem punch will fit. Once seated, strike the broach handle with several sharp impacts. If the SMF stem does not move, consider performing an extended trochanteric osteotomy.

Figure 22a Figure 22b

Figure 23a Figure 23b

Figure 24

Page 13: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

13

Stem specifications

Page 14: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

14

Stem size A/P width M/L stemStem length

Neck offset option Neck angle Neck offset

Neck height

Neck length

-1 18mm 31mm 74mmSTD 131° 31mm 21mm 22mmHO 125° 39mm 21mm 28mm

0 19mm 33mm 73mmSTD 131° 31mm 21mm 22mmHO 125° 39mm 22mm 28mm

1 19mm 34mm 73mmSTD 131° 31mm 22mm 22mmHO 125° 39mm 22mm 28mm

2 19mm 36mm 76mmSTD 131° 33mm 24mm 23mmHO 125° 40mm 24mm 29mm

3 19mm 38mm 80mmSTD 131° 34mm 25mm 25mmHO 125° 41mm 25mm 31mm

4 20mm 40mm 83mmSTD 131° 35mm 26mm 26mmHO 125° 43mm 27mm 32mm

5 20mm 41mm 87mmSTD 131° 36mm 27mm 27mmHO 125° 44mm 27mm 33mm

6 20mm 43mm 90mmSTD 131° 37mm 28mm 28mmHO 125° 45mm 29mm 34mm

7 20mm 45mm 94mmSTD 131° 38mm 30mm 29mmHO 125° 46mm 30mm 35mm

8 21mm 47mm 97mmSTD 131° 39mm 31mm 30mmHO 125° 47mm 31mm 36mm

9 21mm 48mm 101mmSTD 131° 41mm 32mm 31mmHO 125° 48mm 32mm 37mm

All neck offsets, heights and lengths are measured to a +0 head neck length.

Page 15: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

15

Instrument tray layout

Instrument Set 7135-2596Tray 1

Page 16: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

16

Instrument Set 7135-2597 Tray 2

Instrument tray layout

Page 17: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

17

Catalog information

SMF™ stem sizes and neck options

SMF Stem Fixed NeckDescription Cat No

Size -1 Standard Offset 7135-2510Size -1 High Offset 7135-2511Size 0 Standard Offset 7135-2512Size 0 High Offset 7135-2513Size 1 Standard Offset 7135-2700Size 1 High Offset 7135-2701Size 2 Standard Offset 7135-2702Size 2 High Offset 7135-2703Size 3 Standard Offset 7135-2704Size 3 High Offset 7135-2705Size 4 Standard Offset 7135-2706Size 4 High Offset 7135-2707Size 5 Standard Offset 7135-2708Size 5 High Offset 7135-2709Size 6 Standard Offset 7135-2710Size 6 High Offset 7135-2711Size 7 Standard Offset 7135-2712Size 7 High Offset 7135-2713Size 8 Standard Offset 7135-2714Size 8 High Offset 7135-2715Size 9 Standard Offset 7135-2716Size 9 High Offset 7135-2717

Page 18: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

18

Titanium Modular 12/14 Taper Sleeve*Neck Length-4 7134-4245+0 7134-4247+4 7134-4248+8 7134-4249

OXINIUM Modular Femoral Heads*40mm 44mm7134-2340 7134-2344

Femoral head options OXINIUM™ Oxidized Zirconium femoral heads 12/14 taperNeck length 22mm 28mm 32mm 36mm-3 —— 7134-2803 7134-3203 7134-3603+0 7134-2200 7134-2800 7134-3200 7134-3600+4 7134-2204 7134-2804 7134-3204 7134-3604+8 7134-2208 7134-2808 7134-3208 7134-3608 +12 —— 7134-2812 7134-3212 7134-3612+16 —— 7134-2816 7134-3216 ——

Biolox® forte Ceramic Femoral Heads 12/14 TaperNeck Length 28mm 32mm 36mmS/+0 7133-0280 7133-0320 7133-2084M/+4 7133-0284 7133-0324 7133-2085L/+8 7133-0288 7133-0328 7133-2086

Biolox® delta Ceramic Femoral Heads 12/14 Taper Neck Length 32mm 36mm 40mmS/+0 7653-9160 7653-9165 7134-6004M/+4 7653-9161 7653-9166 7134-6005L/+8 7653-9162 7653-9167 7134-6006

CoCr Femoral Heads 12/14 Taper – Cobalt Chromium – ASTM F 799Neck Length 22mm 26mm 28mm 32mm 36mm-3 —— —— 7130-2803 7130-3203 7130-3603+0 7130-2200 7130-2600 7130-2800 7130-3200 7130-3600+4 7130-2204 7130-2604 7130-2804 7130-3204 7130-3604+8 7130-2208 7130-2608 7130-2808 7130-3208 7130-3608+12 7130-2212 7130-2612 7130-2812 7130-3212 7130-3612+16 —— —— 7130-2816 7130-3216 ——

CoCr Modular Femoral Heads – Cobalt Chromium – ASTM F 799*40mm 44mm7134-2640 7134-2644

* Must Use Titanium Modular 12/14 Taper Sleeve with 40mm and 44mm OXINIUM and CoCr Modular Femoral Heads

Page 19: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

19

Osteotomy guide 7136-0086Box osteotome 7136-4002Canal finder 119657Broach handle, straight MIS 7136-4021Starter Broach 7136-5600Broaches size 1 to 6 (Sizes 7 to 9 optional)

Size -1 7136-0566Size 0 7136-0567Size 1 7135-0571 Size 2 7136-0472 Size 3 7136-0473 Size 4 7136-0474 Size 5 7136-0475 Size 6 7136-0476 Size 7 (optional) 7136-0477 Size 8 (optional) 7136-0478 Size 9 (optional) 7136-0479

Trial necks (used with broaches)Standard 7136-0480High Offset 7136-0481

Calcar planer 7136-5702Stem Punch 7136-0565Femoral head/neck impactor 7136-0093Anteversion handle 7136-401228mm, 32mm and 36mm trial heads

28mm -3 7136-970828mm 0 7136-970928mm +4 7136-971028mm +8 7136-971128mm +12 7136-971228mm +16 7136-971332mm -3 7136-971432mm 0 7136-971532mm +4 7136-971632mm +8 7136-971732mm +12 7136-971832mm +16 7136-971936mm -3 7136-972036mm 0 7136-972136mm +4 7136-972236mm +8 7136-972336mm +12 7136-9724

SMF™ Hip System instruments

Page 20: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

20

Accessory Instruments (Included in R3™ Cup and/or REFLECTION™ Cup Instrument Sets)

3.5mm Hex Screwdriver Shaft 7136-2293Ratcheting Handle 7136-2294T-handle 7136-4006

Optional InstrumentsLeft/right double offset broach handles

Left 7136-0089Right 7136-0090

Femoral Head Reducer 7136-3076Replacement Tip for Head Impactor 7136-0484

SMF™ Hip System instruments

Page 21: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

21

Total Hip SystemsImportant Medical InformationImportant NoteTotal hip arthroplasty (THA) has become a successful procedure for relieving pain and restoring motion in patients who are disabled from hip arthropathy. The goals of total hip replacement are to decrease pain, increase function, and increase mobility. Caution: Federal Law (USA) restricts the subject total hip arthroplasty devices to sale by or on the order of a physician.

MaterialsFemoral components are cobalt chromium alloy, titanium 6Al-4V alloy or stainless steel. Femoral heads are cobalt chromium alloy, OXINIUM™ oxidized zirconium, BIOLOX® forte alumina ceramic, BIOLOX delta alumina/zirconia ceramic, zirconia ceramic or stainless steel (SS). Acetabular liners are ultra-high molecular weight polyethylene (UHMWPE), cobalt chromium (CoCr) alloy, BIOLOX forte alumina ceramic, or BIOLOX delta alumina/zirconia ceramic. In the US, refer to the separate package insert provided with the ceramic acetabular liners. All poly acetabular components are UHMWPE. Acetabular shells are titanium 6Al-4V alloy or cobalt chromium (CoCr). The component material is provided on the outside carton label. Note: BIOLOX delta ceramic liners are not available for use in the US.

Some of the alloys needed to produce orthopedic implants contain some metallic components that may be carcinogenic in tissue cultures or intact organism under very unique circumstances. Questions have been raised in the scientific literature as to whether or not these alloys may be carcinogenic in implant recipients. Studies conducted to evaluate this issue have not identified conclusive evidence of such phenomenon, in spite of the millions of implants in use.

Description of SystemThe Total Hip System consists of femoral components, modular necks, proximal sleeves, taper sleeves, acetabular components, fixation screws and pegs, hole covers, centralizers, and femoral heads. Components may be grit blasted, porous coated, hydroxyapatite (HA) coated, or HA porous coated. All implantable devices are designed for single use only.

Femoral ComponentsFemoral components are available in a variety of sizes. Porous coated components are coated for biological ingrowth and are intended to be used without cement. Modular femoral components are available with an oval taper to accept Smith & Nephew, Inc. CoCr modular necks and/or a Morse type taper to accept proximal sleeves. Non-porous femoral components can feature PMMA centralizers that help produce a uniform thickness of cement.

Femoral components, which are available with 10/12, 12/14, and 14/16 tapers, mate and lock directly with Smith & Nephew, Inc. femoral heads having the same sized taper. Certain femoral heads may require taper sleeves for attachment to the femoral stem taper. See chart in the Taper Sleeve section for details.

Taper SleevesA taper sleeve may be required when mating femoral stems with specific types of femoral heads. See the following Sleeve Compatibility Chart for the appropriate combinations. Failure to utilize the proper sleeve – head combination may lead to implant failure resulting in revision surgery. Never place more than one taper sleeve on a femoral component, as this combination will increase stresses on implant and may lead to failure resulting in revision surgery.

Compatible Sleeve Combinations

Femoral Stem Taper

Sleeves to be used with the following femoral heads:

40, 44mm Modular OXINIUM heads

40, 44mm Modular CoCr heads

Modular CoCr heads*

BH Modular CoCr heads**

TANDEM™ CoCr & OXINIUM Unipolar heads

14/16 OXINIUM and CoCr heads

12/14 A A B B C -

14/16 - - - - D

no sleeve

req.

10/12 - - - - E F

Sleeve Material Description/Part Numbers

A Ti-6Al-4VTi 12/14 Modular Sleeve-4: 71344245, +0: 71344247, +4: 71344248, +8: 71344250

B CoCrCoCr 12/14 Modular Sleeve-4: 74222100, +0: 74222200, +4: 74222300, +8: 74222401

C Ti-6Al-4V

12/14 TANDEM Unipolar Sleeve-3: 71326603, +0: 71326600, +4: 71326604, +8: 71326608, +12: 71326613

D Ti-6Al-4V14/16 TANDEM Unipolar Sleeve+0: 126600, +4: 126604, +8: 126608, +12: 126613

E Ti-6Al-4V10/12 TANDEM Unipolar Sleeve+4: MH0304, +8: MH0308, +12: MH0312, +16: MH0317

F Ti-6Al-4V10/12 to 14/16 Taper Conversion Sleeve+0: MH0001, +12: MH0003

*Modular CoCr heads are intended for hemi-arthroplasty use in the US In the US, refer to the separate package insert provided with these components.**BH Modular Heads are not available for use in the US.

Modular NecksModular necks are made from CoCr alloy and are available in a variety of configurations. The modular neck mates and locks with the oval taper of a modular femoral component on one end and the taper of a 12/14 femoral head on the other end.

Femoral HeadsCobalt chromium, stainless steel, oxidized zirconium, and ceramic heads are available in multiple neck lengths for proper anatomic and musculature fit. Heads may be available in 10/12, 14/16, and 12/14 tapers. Certain modular heads and unipolar heads may require taper sleeves for attachment to the femoral stem taper. See Compatible Sleeve Combination Charts in the Taper Sleeves section for details. Heads are highly polished for reduced friction and wear.

The following BIOLOX forte ceramic heads and BIOLOX delta ceramic heads are available for use only with 12/14 taper femoral components:

BIOLOX forte Ceramic Heads

Head Diameter

Neck Length

71332800 71330280* 526969 28mm S/+0

71332804 71330284* 526970 28mm M/+4

71332808 71330288* 526971 28mm L/+8

71333200 71330320** 526914 32mm S/+0

71333204 71330324** 526915 32mm M/+4

71333208 71330328** 526916 32mm L/+8

71331047 71332084*** 76539150 36mm S/+0

71331048 71332085*** 76539151 36mm M/+4

71331049 71332086*** 76539152 36mm L/+8

*Used with REFLECTION BIOLOX forte Ceramic Acetabular Liners in the US**Used with REFLECTION BIOLOX forte Ceramic Acetabular Liners and R3 BIOLOX forte Ceramic Acetabular Liners in the US.***Used with R3 BIOLOX forte Ceramic Acetabular Liners in the US.In the US, refer to the separate package insert provided with the ceramic acetabular liners.

BIOLOX delta Ceramic Heads

Head Diameter

Neck Length

71346001 28mm S/+0

71346002 28mm M/+4

71346003 28mm L/+8

76539160 32mm S/+0

76539161 32mm M/+4

76539162 32mm L/+8

76539165 36mm S/+0

76539166 36mm M/+4

76539167 36mm L/+8

76539153† 36mm XL/+12

71346004 40mm S/+0

71346005 40mm M/+4

71346006 40mm L/+8

71330029 44mm S/+0

71330031 44mm M/+4

71330032 44mm L/+8

†Not available for use in the US

The following CoCr BIRMINGHAM HIP™ (BH) modular heads†† should be used only with BIRMINGHAM HIP acetabular cups and R3 metal acetabular liners. In the US, refer to the separate package insert provided with these components.

74222138 BIRMINGHAM HIP Modular Head 38mm

74222140 BIRMINGHAM HIP Modular Head 40mm

74222142 BIRMINGHAM HIP Modular Head 42mm

74222144 BIRMINGHAM HIP Modular Head 44mm

74222146 BIRMINGHAM HIP Modular Head 46mm

74222148 BIRMINGHAM HIP Modular Head 48mm

74222150 BIRMINGHAM HIP Modular Head 50mm

74222152 BIRMINGHAM HIP Modular Head 52mm

74222154 BIRMINGHAM HIP Modular Head 54mm

74222156 BIRMINGHAM HIP Modular Head 56mm

74222158 BIRMINGHAM HIP Modular Head 58mm

††BH Modular Heads are not available for use in the US

Acetabular ComponentsAcetabular components can be one-piece all polyethylene, or two-piece, consisting of a titanium shell and either a UHMWPE liner, BIOLOX forte ceramic liner, BIOLOX delta ceramic liner or CoCr metal liner. For BIOLOX forte ceramic liners available for use with the REFLECTION™ Ceramic Acetabular System in the US, refer to the separate package insert provided with these components. See Warnings and Precautions for specific information on screws, pegs and hole covers use. Acetabular reinforcement and reconstruction rings are used with an all polyethylene acetabular component. Note: BIOLOX delta ceramic liners are not available for use in the US. For R3 metal liners available for use with the BIRMINGHAM HIP Resurfacing (BHR) System in the US, refer to the separate package insert provided with these components.

Note: 10 Mrad cross-linked UHMWPE acetabular liners may be used with metal (CoCr and SS), oxidized zirconium, BIOLOX forte ceramic heads or BIOLOX delta ceramic heads. Stainless Steel (SS) heads are not available for use in the US.

Femoral components and femoral heads are designed for use with any Smith & Nephew polyethylene acetabular component or polyethylene-liner, metal-backed acetabular component having an appropriately-sized inside diameter. Acetabular liners are designed for use only with acetabular shells from the same product family (ie REFLECTION liners can only be used with REFLECTION shells; R3 liners can only be used with R3 shells). The use of other combinations may cause implant failure and lead to revision surgery.

INDICATIONSHip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.

Hip components are also indicated for inflammatory degenerative joint disease including rheumatoid arthritis, arthritis secondary to a variety of diseases and anomalies, and congenital dysplasia; treatments of nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; endoprosthesis, femoral osteotomy, or Girdlestone resection; fracture-dislocation of the hip; and correction of deformity.

Total hip systems may be indicated for use (i) with bone cement , (ii) without bone cement , or (iii) for use with or without cement. Reference product labeling and literature for specific applications.

The REDAPT™ revision hip system (formerly MDF) is intended to be used without cement. In the EU, REDAPT hip system is indicated for revision surgery only.

The R3 Acetabular System is for single use only and is intended for cementless use.

Acetabular reinforcement and reconstruction rings are intended to be used in primary and revision surgeries where the acetabulum has the deficiencies of the acetabular roof, anterior or posterior pillar, medial wall deficiency, and/or protrusion as a result of the indications listed previously.

Some of the diagnoses listed above may increase the risk of complications and reduce the chance of a satisfactory result. Specifically, an increased risk of complications for revision surgery for any reason has been documented in the literature. Patient selection factors such as age, weight, and activity level can negatively affect implant longevity and increase the risk of revision surgery. Literature has shown a higher likelihood of revision in younger, heavier, or more active patients. Specifically, the risk of complications is greater in obese and morbidly obese patients.

CONTRAINDICATIONS 1. Conditions that would eliminate or tend to eliminate adequate

implant support or prevent the use of an appropriately- sized implant, eg:a. blood supply limitations;b. insufficient quantity or quality of bone support, eg,

osteoporosis, or metabolic disorders which may impair bone formation, and osteomalacia; and

c. infections, osteolysis, or other conditions which lead to increased bone resorption.

2. Mental or neurological conditions which tend to impair the patient’s ability or willingness to restrict activities.

3. Physical conditions or activities which tend to place extreme loads on implants, eg, Charcot joints, muscle deficiencies, multiple joint disabilities, etc.

4. Skeletal immaturity.5. The alumina ceramic liner is contraindicated for use with any

product other than the metal shell with the correlating inner taper geometry and the appropriate sized alumina ceramic head. The alumina ceramic liner should only be used with the alumina ceramic head. In the US, refer to the separate package insert provided with the ceramic acetabular liners.

6. In revision surgery, inadequate proximal implant support is contraindicated. The literature shows an increased risk of implant failure in revision cases where proximal support is not achieved, poor bone quality exists, and smaller sized implants are utilized. The lower the implant fixation point in the femur (distance from the head center) the greater the risk of implant fracture and/or re-revision.

7. Morbid obesity.

Page 22: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

22

Contraindications may be relative or absolute and must be carefully weighed against the patient’s entire evaluation and the prognosis for possible alternative procedures such as non-operative treatment, arthrodesis, femoral osteotomy, pelvic osteotomy, resection arthroplasty, hemiarthroplasty and others.

Conditions presenting increased risk of failure include: osteoporosis, metabolic disorders which may impair bone formation, and osteomalacia.

Possible Adverse Effects of the Device (In Primary and Revision THA)

1. Wear of the polyethylene, metal, and ceramic articulating surfaces of acetabular components may occur. Higher rates of wear may be initiated by the presence of particles of cement, metal, or other debris which can develop during or as a result of the surgical procedure and cause abrasion of the articulating surfaces. Higher rates of wear may shorten the useful life of the prosthesis and lead to earlier revision surgery to replace the worn prosthetic components.

2. With all joint replacements, asymptomatic, localized, progressive bone resorption (osteolysis) may occur around the prosthetic components as a consequence of foreign-body reaction to particulate wear debris. Particles are generated by interaction between components, as well as between the components and bone, primarily through wear mechanisms of adhesion, abrasion, and fatigue. Secondarily, particles may also be generated by third-body particles lodged in the polyethylene, metal, or ceramic articular surfaces. Osteolysis can lead to future complications necessitating the removal or replacement of prosthetic components.

3. Failure to observe the Warnings and Precautions, trauma, strenuous activity, implant alignment, patient non-compliance, involuntary muscular disorders, duration of service increase the risk of loosening, bending, cracking, or fracture of implant components, which may lead to revision surgery.

4. Improper neck selection, positioning, looseness of acetabular or femoral components, extraneous bone, penetration of the femoral prosthesis through the shaft of the femur, fracture of the acetabulum, intrapelvic protrusion of acetabular component, femoral impingement, periarticular calcification, and/or excessive reaming may increase the risk of dislocations, subluxation, decreased range of motion, or lengthening or shortening of the femur, which may lead to revision surgery.

5. Congenital deformity, improper implant selection, improper broaching or reaming, osteoporosis, bone defects due to misdirected reaming, trauma, strenuous activity, improper implant alignment or placement, patient non-compliance, etc. can increase risk of femoral or pelvic fractures.

6. Failure of the implant porous coating/substrate interface or hydroxyapatite coating/porous coating bonding may result in bead separation or delamination, which may lead to increased third body wear resulting in revision surgery.

7. Implant migration or subsidence resulting in revision surgery has occurred in conjunction with compaction grafting procedures usually resulting from insufficient graft material, improper cement techniques, and/or varus stem alignment.

8. Implant loosening or fracture, particularly of smaller sized or high offset implants, is more likely to occur in patients who are young, physically active, and/or heavy, which may lead to implant failure and revision surgery.

9. Temporary or permanent device related noise such as clicking, squeaking, popping, grating, or grinding, which may lead to implant failure and revision surgery.

10. Although rare, metal sensitivity reactions and/or allergic reactions to foreign materials have been reported in patients following joint replacement, which have required device removal.

Potential Complications Associated with any Total Hip Arthroplasty surgery, Primary and Revision.

1. Infection, both early, post-operative superficial and early, post-operative deep wound infection and late periprosthetic infection.

2. Neuropathies; femoral, sciatic, peroneal nerve, and lateral femoral cutaneous neuropathies have been reported. Temporary or permanent nerve damage resulting in pain or numbness of the affected limb.

3. Wound hematoma, thromboembolic disease including venous thrombosis, or pulmonary embolus. Some studies have shown an increased risk of thromboembolic disease including venous thrombosis with cemented THA as compared to uncemented THA.

4. Myositis ossificans, especially in males with hypertrophic arthritis, limited pre-operative range of motion and/or previous myositis. Also, periarticular calcification with or without impediment to joint mobility can cause decreased range of motion.

5. Trochanteric nonunion usually associated with early weight bearing and/or improper fixation of the trochanter, when a transtrochanteric surgical approach is used.

6. Damage to blood vessels.7. Accidental patient burns from cautery device. 8. Delayed wound healing.9. Aggravated problems of the affected limb or contralateral

extremity caused by leg length discrepancy, excess femoral medialization, or muscle deficiency.

Warnings and Precautions

Preoperative1. The patient should be warned of surgical risks, and made

aware of possible adverse effects. 2. The patient should be warned that the device does not

replace normal healthy bone, that the implant can break or become damaged as a result of trauma or activity including heavy labor for occupation or recreation.

3. The patient should be warned that the implant has a finite expected service life and may need to be replaced in the future. Patients should be warned that the longevity of the implant may depend on their weight and level of activity.

4. The patient should be warned of the brittle nature of the ceramic components and the possibility of failure of the device leading to additional surgery in the future.

5. Improper selection, placement, positioning, and fixation of the implant components may result in unusual stress conditions and subsequent early failure/fracture of the components.

6. The surgeon should be thoroughly familiar with the implants, instruments, and surgical procedure prior to performing surgery. Certain insertion techniques may be different than those known for conventional hip systems, and are specifically designed to avoid potential implant failures.

7. Do not mix components from different manufacturers unless specially approved by the manufacturer of the components. Failure to comply may result in implant failure and revision surgery. For purposes of product inter-compatibility, products manufactured and labeled by entities formerly known as Plus Endoprothetik, Intraplant, Precision Implants and Plus Orthopedics (now Smith & Nephew Orthopaedics AG) may be considered as the same manufacturer, Smith & Nephew unless otherwise stated. Additional Warnings and Precautions may be included in component literature.

8. Use extreme care in handling and storage of implant components. Cutting, bending, or scratching the surface of components can significantly reduce the strength, fatigue resistance, and/or wear characteristics of the implant system. These, in turn, may induce internal stresses that are not obvious to the eye and may lead to fracture of the component. Implants and instruments should be protected from corrosive environments such as salt air during storage. Do not allow the porous surfaces to come in contact with cloth or other fiber-releasing materials, as this may compromise fixation and lead to failure.

9. Allergies and other reactions to device materials, although infrequent, should be considered, tested for (if appropriate), and ruled out preoperatively. A reaction may lead to revision surgery.

10. Fixation and expected longevity of components expected to be left in place at revision surgery should be thoroughly assessed. Damage to and/or disruption of the implant during revision surgery may lead to implant failure.

11. The surgeon should be familiar with the appropriate surgical technique. Refer to medical or manufacturer literature for specific product information, which is available upon request. Failure to follow the appropriate surgical technique may result in implant failure or revision surgery.

12. Intraoperative fracture or breaking of instruments can occur. Instruments which have experienced extensive use or excessive force are susceptible to fracture. Instruments should be examined for wear or damage and proper operation prior to surgery. Failure to do so may result in injury to the surgical team and/or the patient. Single use devices should not be reused due to risks of breakage, failure, or patient infection and revision surgery.

13. Do not cold water quench ceramic components and never sterilize ceramic heads while fixed on the stem taper, as this may induce additional stresses on component which may cause implant failure, resulting in revision surgery. (See sterilization section, below.)

14. OXINIUM oxidized zirconium femoral heads and cobalt chrome femoral heads are designed to articulate with conventional UHMWPE or XLPE bearing surfaces. BIOLOX forte femoral heads and BIOLOX delta femoral heads articulate with conventional UHMWPE or XLPE bearing surfaces, BIOLOX forte ceramic liners, or BIOLOX delta ceramic liners. BHR resurfacing heads and BH cobalt chrome modular heads articulate with BH acetabular cups or R3 metal liners. OXINIUM oxidized zirconium femoral heads, cobalt chrome femoral heads, BIOLOX forte ceramic femoral heads and BIOLOX delta ceramic femoral heads should never articulate against metal bearing surfaces because severe wear of the metal bearing surfaces may occur. OXINIUM oxidized zirconium femoral heads and cobalt chrome femoral heads should never articulate against BIOLOX delta or BIOLOX forte ceramic liners because severe wear of the bearing surfaces may occur. Note: BIOLOX delta ceramic liners and BIRMINGHAM HIP CoCr modular heads are not available for use in the US.

15. Select only Smith & Nephew femoral components for use with Smith & Nephew ceramic heads. The taper on the stem/neck is machined to tightly mate and lock with the ceramic head. An improperly dimensioned taper could result in disassociation or fracture of the ceramic head, resulting in revision surgery.

16. Do not use Smith & Nephew 36mm -3 heads with SL-PLUS™ Hip Stems and SLR-PLUS Hip Stems or any of the +16 heads with any PLUS Hip Stem. Use of these unapproved combinations may result in implant failure and revision surgery.

17. If a computer assisted surgery system is used, consult the applicable software and hardware reference manuals provided by the manufacturer to ensure proper operation of this equipment.

Intraoperative1. The general principles of patient selection and sound

surgical judgment apply. The correct selection of the implant is extremely important. The appropriate type and size should be selected for patients with consideration of anatomical and biomechanical factors such as patient age, activity levels, weight, bone and muscle conditions, any prior surgery and anticipated future surgeries, etc. Generally, the largest cross-section component which will allow adequate bone support to be maintained is preferred. Failure to use the optimum-sized component may result in loosening, bending, cracking, or fracture of the component and/or bone, resulting in revision surgery.

2. Correct selection of the neck length and cup, and stem positioning, are important. Muscle looseness and/or malpositioning of components may result in loosening, subluxation, dislocation, and/or fracture of components. Increased neck length and varus positioning will increase stresses which must be borne by the stem. The component should be firmly seated with the component insertion instruments and stability verified. Failure to do so may result in implant failure and revision surgery.

3. Care should be taken not to scratch, bend (with the exception of the Reconstruction Rings) or cut implant components during surgery for the reasons stated in Number Eight of the “Pre-Operative” section of “Warnings and Precautions.”

4. A +12 mm or +16 mm femoral head should not be used with any Small taper stems. These unapproved combinations will increase stresses which must be borne by the stem and may result in implant failure and revision surgery.

5. Modular heads, modular necks, modular sleeves and femoral components should be from the same manufacturer unless specially approved by the manufacturer of the components to prevent mismatch. Failure to comply may result in implant failure and revision surgery.

6. Stainless steel heads and stainless steel stems should only be used together. Neither should be used with other metal components. These unapproved metal combinations may corrode causing implant failure and revision surgery.

7. Use only REFLECTION Liners with REFLECTION Shells. Use only R3 Liners with R3 Shells. Failure to comply may result in implant failure and revision surgery.

8. Clean and dry all taper connections prior to impacting for assembly. The modular femoral head, neck and/or sleeve components must be firmly seated on the femoral component to prevent disassociation, excess fretting wear, implant failure, and revision surgery.

9. Take care, when positioning and drilling screw and peg holes, to avoid penetration of the inner cortex of the pelvis, penetration of the sciatic nerve, or damage to other vital neurovascular structures. Perforation of the pelvis with screws that are too long can rupture blood vessels causing the patient to hemorrhage. Do not place a screw in the center hole of the acetabular prosthesis. Placement of drills and screws in the anterior or medial portions of the prosthesis is associated with a high risk of potentially fatal vascular injury. Bone screws must be completely seated in the holes of the shell to allow proper locking for the acetabular component liner. If the tapered pegs need to be removed from the shell after impaction of the pegs, do not reuse the pegs or the peg shell holes. Use new pegs and different shell holes, or a new shell if necessary. Failure to comply may result in implant failure and revision surgery.

10. REFLECTION three hole (SP3), multi-hole (SPM), peripheral hole (SPR), INTERFIT™ and R3 Shells accept both REFLECTION spherical head screws and Universal cancellous bone screws. REFLECTION INTERFIT Shells accept the Modified REFLECTION screw hole covers. REFLECTION Peripheral Hole Screws should only be used with REFLECTION SPR Shells. Locking Head Pegs and REFLECTION Locking Head Screw Hole Covers are only for use with REFLECTION SP3. The threaded center hole in REFLECTION Shells only accepts threaded hole covers, not screws or pegs. The INTERFIT threaded hole cover is only for use with REFLECTION INTERFIT, Spiked and No Hole Shells. The REFLECTION threaded hole cover can be used with all REFLECTION and R3 shells. The R3 screw hole cover can be used with R3 and REFLECTION Three Hole shells. Refer to product literature for proper adjunctive fixation and hole cover usage. Failure to comply may result in implant failure and revision surgery.

11. Modular components must be assembled securely to prevent disassociation. Prior to seating modular components, surgical debris including blood, bone, tissue, and bone cement must be cleaned from the surfaces. Debris may inhibit the component locking mechanism leading to implant failure and revision surgery.

12. If the shell is to be cemented in place, remove extraneous cement with a plastic sculps tool to ensure proper locking of the liner. During liner insertion, make sure soft tissue does not interfere with the shell/liner interface. Chilling the liner reduces the impaction force required to seat the liner.

13. Avoid repeated assembly and disassembly of the modular components which could compromise the critical locking action of the locking mechanism, resulting in component failure and revision surgery.

14. Care should be taken to ensure proper cement mixing, an adequate cement mantle, and the complete support of all parts of the device embedded in bone cement, to prevent stress concentration which may lead to failure of the procedure. Specific cement mixing and handling instructions can be found in cement product labeling. During curing of the cement, care should be taken to prevent movement of the implant components. Failure to do so may cause implant failure and revision surgery.

15. If the head is removed from a femoral component that will be left in place at revision surgery, it is recommended that a metal head be used. Do not assemble a ceramic head on a used taper, as the ceramic head may fracture from irregularities on the femoral component taper.

16. If components are to be left in place at revision surgery, they should first be thoroughly checked for cracks, scratches, looseness, and other signs of damage, and replace if necessary. The head/neck component should be changed only when clinically necessary. Failure to comply may result in implant failure and revision surgery.

17. Once removed from the patient, implants previously implanted should never be reused, since internal stresses which are not visible may lead to early bending or fracture of these components, Reuse may also increase the risk of patient infection and result in revision surgery.

18. With the congenitally dislocated hip, special care should be taken to prevent sciatic nerve palsy. Also, note that the femoral canal is often very small and straight and may require an extra-small straight femoral prosthesis; however, a regular-sized prosthesis should be used when possible. Note that the true acetabulum is rudimentary and shallow. A false acetabulum should not ordinarily be utilized as a cup placement site for anatomical and biomechanical reasons.

19. With rheumatoid arthritis, especially for those patients on steroids, bone may be extremely osteoporotic. Care should be taken to prevent excessive penetration of the acetabular floor or fracture of the medial acetabular wall, femur, or greater trochanter.

20. Revision procedures for previous arthroplasty are technically demanding and difficult to exercise, with higher complication rates, as shown in literature. Increased operative time, increased blood loss, increased incidence of pulmonary embolus and wound hematoma, and a higher risk of infection can be expected with revision procedures. Common errors include misplacement of the incision, inadequate exposure or mobilization of the femur, inadequate removal of ectopic bone, improper positioning of components, or inadequate proximal support of the femoral component. Studies have indicated a higher risk of implant fatigue fracture in cases with inadequate proximal bone stock or where extended trochanteric osteotomies have been performed. In these cases, it is imperative that adjunctive reinforcement procedures such as bone grafting, cortical strut allografts, cables, and trochanteric plates are utilized to provide adequate proximal support to the femoral component. The use of larger prostheses may also reduce the risk of avoiding prosthetic fatigue fracture. Although these adjunctive reinforcement procedures may minimize the risk of implant failure, they do not ensure a predictable clinical result.

Page 23: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

23

21. Prior to closure, the surgical site should be thoroughly cleaned of cement, bone chips, ectopic bone, or other foreign matter. Ectopic bone and/or bone spurs may lead to dislocation or painful or restricted motion.

22.Range of motion should be thoroughly assessed for early impingement or joint instability. Postoperative instability (i.e. dislocation) is a leading complication associated with revision surgery and may result in additional surgery.

23.Proper positioning of the components is important to minimize impingement which could lead to early failure, premature wear, device related noise, and/or dislocation, all which may lead to revision surgery.

24. To minimize the risk of acetabular shell loosening in uncemented applications, surgeons should consider using orthopedic bone fixation devices such as bone screws, spikes, pegs, fins, or other bone fixation devices. To minimize loosening risk for cemented acetabular shells, care should be taken to prevent movement of the implant components while cement is curing.

25. Physicians should consider component malposition, component placement, and the effect on range of motion and stability when using modular heads (with sleeves or skirts) and overhang liners.

26. For computer assisted surgery systems, it is extremely important to correctly select input parameters (eg bony landmarks). Operators of this equipment should be familiar with the anatomy relevant to the procedure. Failure to provide proper input could cause problems such as violation of critical anatomical structures and malpositioned implants, which may lead to revision surgery.

27. Trial instrumentation may be provided for intraoperative assessment of final implant fit. Do NOT implant trial components.

28. Do not implant HA-coated devices in bone cement.29. Inappropriate use of taper sleeves may lead to implant

failure resulting in revision surgery. Select the appropriate sleeve based on the Compatible Sleeve Combinations Charts located in the Description of System section of this document.

Postoperative1. Postoperative directions and warnings to patients by

physicians, and patient care, are extremely important. Gradual weight bearing is begun after surgery in ordinary total hip arthroplasty. However, with trochanter osteotomy or certain complex cases, weight-bearing status should be individualized with the non or partial weight-bearing period extended.

2. Patients should be warned against unassisted activity, particularly use of toilet facilities and other activities requiring excessive motion of the hip, which may result in subluxation or dislocation.

3. Use extreme care in patient handling. Support should be provided to the operative leg when moving the patient. While placing the patient on bedpans, changing dressings, and clothing, and similar activities, precautions should be taken to avoid placing excessive load on the operative part of the body.

4. Postoperative therapy as prescribed by the physician should be structured to regain muscle strength around the hip and a gradual increase of activities.

5. Continued periodic x-rays as prescribed by the physician are recommended for close comparison with immediate postoperative conditions to detect long-term evidence of changes in position, loosening, bending and/or cracking of components or bone loss. With evidence of these conditions, patients should be closely observed, the possibilities of further deterioration evaluated, and the benefits of early revision considered.

6. Prophylactic antibiotics should be recommended to the patient similar to those suggested by the American Heart Association for conditions or situations that may result in bacteremia.

7. Normal daily activity may be resumed at the physician’s direction. Patients should be directed to seek medical opinion before entering potentially adverse environments that could affect the performance of the implant, such as electromagnetic or magnetic fields, including a magnetic resonance environment.

8. The patient should be advised to report any pain, decrease in range of motion, swelling, fever, squeaking, clicking, popping, grating, or grinding noises and unusual incidences. Patient reports of squeaking, clicking, popping, grating, or grinding should be carefully evaluated as they may indicate position changes in the components compromising the durability of the implants.

9. Post-operative subluxation may result in higher wear and implant damage.

Packaging and LabelingImplants should only be accepted if received by the hospital or surgeon with the factory packaging and labeling intact. If the sterile barrier has been broken or past the expiration date, return the component to Smith & Nephew, Inc. Use of damaged or expired product increases the risk of infection which may lead to revision surgery.

Healthcare provider should have a full understanding of the product labeling information including but not limited to the following: Instructions For Use (IFU) document and manufacturer provided surgical techniques and other relevant product materials.

SterilizationImplant components are supplied sterile to a Sterility Assurance Level (SAL) of 10-6. Implant components are supplied in protective packaging. Inspect packages for punctures or other damage prior to surgery. The method of sterilization is noted on the package label.

Hip implant components, disposable or single-use instruments should not be reused due to risks of breakage, failure, or patient infection. Contact your local Smith & Nephew, Inc. Sales Representative regarding procedures to return components. If not specifically labeled sterile, instruments are supplied non-sterile and must be cleaned and sterilized prior to surgery. Please see also the document, “Recommendations for decontamination and sterilization of Smith & Nephew orthopaedic devices”, which is available from customer service or via Smith & Nephew website, for further information on cleaning instructions and validated sterilization procedures.

Recommended Steam Sterilization Cycle Parameters• Dynamic Air Removal (Prevacuum) Steam Cycle: 132ºC (270ºF)

- 4 minutes or 135ºC (275ºF) - 3 minutes; Minimum drying time: wrapped instruments – 15 minutes, containerized devices - 30 minutes.

• Gravity Displacement Steam Cycle: 132°C (270°F) – 15 minutes for wrapped instruments and 30 minutes for containerized devices; Minimum drying time - 30 minutes.

• Immediate Use Steam Sterilization or Flash Steam Cycle (Reusable instruments only): 132°C (270°F) for 15 minutes in a Gravity Displacement Cycle or 4 minutes in a Dynamic Air Removal (Prevacuum) Cycle.

• United Kingdom Steam Cycle: 134°C (273°F) - 3 minutes; Minimum vacuum drying - 30 minutes. (Note: The procedure outlined in HTM 2010 should be followed).

Containment devices should be wrapped with a central supply wrap (CSR) or placed in a reusable rigid container for sterilization. Note to US Customers: FDA cleared sterilizers and wraps are to be used in your sterilization processes.

Magnetic Resonance Imaging (MRI) SafetySmith & Nephew hip systems have not been evaluated for safety and compatibility in the MR environment. Hip system components have not been tested for heating or migration in the MR environment.

Retrieval and Analysis of Removed ImplantsThe most important part of surgical implant retrieval is preventing damage that would render scientific examination useless. Special care should be given to protect the implant from damage during handling and shipment. Specifically, for conventional polyethylene or XLPE, use alternative sterilization method other than steam autoclave. Follow internal hospital procedures for the retrieval and analysis of implants removed during surgery. When handling removed implants, use precautions to prevent spread of bloodborne pathogens.

If the implant will be returned to Smith & Nephew, Inc. for analysis, contact Customer Service using the phone numbers outlined in the Information section.

INFORMATIONFor further information, please contact Customer Service at (800) 238-7538 for calls within the continental USA and (901) 396-2121 for all international calls.

Manufacturing facilities and EC representative:

Smith & Nephew Inc.1450 Brooks RoadMemphis, TN 38116 USATel.: 901-396-2121www.smithnephew.com

Smith & Nephew Orthopaedics GmbHAlemannenstrasse 1478532 Tuttlingen, GermanyTel.: 07462/208-0Fax: 07462/208-135

Explanation of symbols used in labeling:

H2O2 hydrogen peroxide sterilization

ID Inner Diameter

OD Outer Diameter

S/+0 Short

M/+4 Medium

L/+8 Long

SO Standard Offset

H or HO High Offset

For cemented use only

For uncemented use only

™Trademark of Smith & Nephew. Certain Marks Reg. U.S. Pat. & TM Off. All trademarks acknowledged.

81078790 Rev. 0 2012-05

Page 24: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

24

Four modular neck options are available: standard offset, high offset, left anteverted and right anteverted.

• Neutral necks (0° anteversion) are available in standard and high offset.

• The standard neutral neck has a neck shaft angle of 131° (Figure 25).

• The high offset neutral neck provides 8mm of lateralization. The high offset neck has a neck shaft angle of 125° and 137° when inserted in the reverse/valgus orientation (Figure 26).

• Left anteverted necks shift the femoral head, 10°, anteriorly relative to the stem on a left hip and posteriorly relative to the stem on a right hip. Right anteverted necks shift the femoral head center based on the operative leg in an equivalant manner (Figure 27).

• The anteverted necks are available in one offset option providing 6mm of lateralization relative to the standard offset neck.

• The anteverted neck has a neck shaft angle of 125° and an angle of 137° when inserted in the reverse/valgus orientation.

AppendixSummary of modular neck options This material is supplied as a reference for those performing a revision on a modular neck SMF™ System Implant (Also reference table on page 26)

Standard

Figure 25

High offset

Figure 26a

High offset reverse/valgus

Figure 26b

Right

Figure 27a

Right reverse/valgus

Figure 27b

Left

Figure 27c

Left reverse/valgus

Figure 27d

Page 25: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

25

The high offset and anteverted neck are asymetrical and different – offset, leg length and valgus angles can be achieved depending upon orientation of the neck as depicted in the chart on page 26.

To insure that the desired orientation is achieved, the face of the taper of the modular necks are etched with ‘HO’ for High Offset, ‘L’ for Left Anteverted, ‘R’ for Right Anteverted and an arrow which aids in duplicating orientation of the trial modular neck with the implanted modular neck.

When the arrow points top, in the direction of the greater trochanter, the necks provide the options as labeled. If the arrow points to the distal femur, then different head center locations are possible as highlighted on page 26.

Additionally, color coding of the anteverted neck implant packaging and corresponding plastic trial necks has been provided as follows (Figure 31).

High offset neck in varus orientation

Figure 28

High offset neck in reverse/valgus orientation

Figure 30

Laser etch on Top of high offset and left and right anteverted necks

Figure 29

Summary of modular neck options

Sample of Color Coding on Package Labels for Left and Right Anteverted Modular Necks

Figure 31

Page 26: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

26

Neck orientationarrow up – varus;arrow down – reverse/valgus

Effect (relative to standard offset neck with +0 head)

Offset Leg length

125°

+8mm 0mm

137°

0mm +8mm

Version‘+’ anteversion‘-’ retroversion

Neck orientationarrow up – varusarrow down – reverse/valgus

Effect (relative to standard offset neck with +0 head)

Offset Leg length

+10° anteversion

125°

+6mm 0

137°

0mm +6mm

-10° retroversion

125°

+6mm 0mm

137°

0mm +6mm

Right hipVersion‘+’ anteversion‘-’ retroversion

Neck orientationarrow up – varusarrow down – reverse/valgus

Effect (relative to standard offset neck with +0 head)

Offset Leg length

+10° anteversion

125°

+6mm 0

137°

0mm +6mm

-10° retroversion

125°

+6mm 0mm

137°

0mm +6mm

The table below summarizes the effects of modular neck type, operative side (patient left or right hip) and modular neck orientation.

Page 27: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

27

Neck assembly

Ensure all tapers are clean and dry prior to assembly, especially the implant pocket as debris in the taper could prevent proper seating. Before assembling, check the information laser etched on top of the neck to confirm that the desired modular neck option is being inserted. Additionally, you may refer to the Modular Neck Option Chart on page 26 to make certain the chosen neck will provide the neck angle and anterior/posterior offset desired.

Choose the appropriate femoral neck implant and place it into the stem pocket, making sure that the neck is appropriately oriented.

Optional final trial reduction A final trial reduction may be performed at this time.

After assessment of joint stability and final head size is chosen, make certain to remove the trial head.

Femoral head assembly Ensure all tapers are clean and dry prior to assembly as debris in the taper could prevent proper seating. Assemble the chosen head on the neck. Impact with multiple firm mallet blows and check that the construct is stable (Figure 32).

Implant neck removal To remove an assembled modular neck, seat the modular neck removal tool as shown to the right and impact firmly (Figure 33).

Caution: Do not use excessive force to remove neck as this may result in femoral fracture or damage to implants and instrumentation.

Caution: It is recommended that modular necks not be reused due to the potential for taper damage when removing an impacted modular neck.

Figure 32

To reassemble modular neck

Figure 33

Page 28: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

28

Stem removal

Assemble the stem inserter/extractor to the broach handle by placing the stem inserter/extractor post in the clamp. Use the thumb to lock the clamp onto the stem inserter/extractor (Figure 34).

Slide the plastic taper plug provided in the instrument set into position with the flange of the plastic taper plug engaged in the slot of the stem inserter/extractor. Make certain the hole in the plastic plug is in line with the hole in the inserter/extractor (Figure 35).

Figure 34a Figure 34b

Figure 35a Figure 35b

Page 29: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

29

Figure 36

Figure 37

Place the inserter/extractor broach handle assembly onto the stem (Figure 36) and insert the screw provided in the instrument case through the hole in the stem inserter/extractor and thread the screw all the way down. Use a 3.5mm hex driver* to tighten further if desired. With firm mallet blows remove the stem (Figure 37).

* Reference straight screwdriver shaft (Cat No 7136-2293) and ratchet driver (Cat No 71362-294) or T-handle (Cat No 7136-4006) to tighten the insertion screw.

Caution: Do not use excessive force to remove stem.

Page 30: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

30

Notes

Page 31: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial
Page 32: Surgical Technique - Smith & Nephew · Design surgeon list ... Ischia or both of the obturator foramens as shown in Figure 1. This line should extend out until it contacts the medial

Smith & Nephew, Inc.7135 Goodlett Farms ParkwayCordova, TN 38016USA

Telephone: 1-901-396-2121Information: 1-800-821-5700Orders and Inquiries: 1-800-238-7538

™ Trademark of Smith & Nephew. Certain marks Reg. US Pat. & TM Off. All Trademarks acknowledged.

www.smith-nephew.com

©2013 Smith & Nephew, Inc.All rights reserved.7138-1715 REV0 03/13