Validation | Slide 1 of 48 August 2006 Validation Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4.

Upload: others

Post on 02-Apr-2020

21 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 1 of 48 August 2006

Validation

Supplementary Training Modules on Good Manufacturing Practice

WHO Technical Report Series, No. 937, 2006. Annex 4.

Page 2: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 2 of 48 August 2006

Validation

!  Part 1. General overview on qualification and validation

!  Part 2. Qualification of HVAC and water systems

!  Part 3. Cleaning validation

!  Part 4. Analytical method validation

!  Part 5. Computerized system validation

!  Part 6. Qualification of systems and equipment

!  Part 7. Non sterile product process validation

Page 3: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 3 of 48 August 2006

Qualification of HVAC and water systems

Part 2

Supplementary Training Modules on Good Manufacturing Practice

WHO Technical Report Series, No. 937, 2006. Annex 4. Appendix 1 and 2

Page 4: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 4 of 48 August 2006

HVAC

Objectives

! To understand key issues in –  commissioning, –  qualification and –  maintenance of

HVAC and Water systems

8.

Page 5: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 5 of 48 August 2006

●  Description of design, installation and functions

●  Specifications, requirements

●  Manuals

●  Operating procedures

●  Instructions for performance control, monitoring and records

●  Maintenance instructions and records

●  Training of personnel

―  programme and records

Documentation requirements to assist in commissioning, qualification and maintenance

HVAC

Page 6: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 6 of 48 August 2006

HVAC

Commissioning !  Precursor to qualification

!  Includes setting up, balancing, adjustment and testing of entire HVAC system to ensure it meets requirements in URS and capacity

!  Acceptable tolerances for parameters

!  Training of personnel

8.1.1, 8.1.4, 8.1.5

Page 7: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 7 of 48 August 2006

HVAC

Commissioning (2) Records and data maintained include:

!  Installation records – documented evidence of measure capacities of the system

! Data: design and measurement for, e.g. air flow, system pressures

! O&M manuals, schematic drawings, protocols, reports

8.1.2, 8.1.3, 8.1.6

Page 8: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 8 of 48 August 2006

HVAC

Qualification !  Validation is an extensive exercise

! Qualification of the HVAC system is one component in the overall approach that covers premises, systems/utilities, equipment, processes, etc.

!  See also full guidelines on "Validation" in WHO TRS, No. 937, 2005, Annex 4

! Risk based approach for HVAC qualification 8.2.1

Page 9: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 9 of 48 August 2006

HVAC

Qualification (2)

! Described in a Validation Master Plan (VMP)

!  VMP to include the nature and extent of tests, and protocols

! DQ, IQ, OQ, and PQ

! Risk analysis to determine critical and non-critical parameters, components, subsystems and controls

8.2.2 – 8.2.5

Page 10: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 10 of 48 August 2006

HVAC

Qualification (3) ! Direct impact components and critical parameters should be

included

! Non-critical systems and components are subjected to Good Engineering Practices (GEP)

!  Acceptance criteria and limits defined in design stage

! Design conditions, normal operating ranges, operating ranges, alert and action limits

8.2.5 – 8.2.11

Page 11: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 11 of 48 August 2006

HVAC

!  Design conditions and normal operating ranges set to achievable limits

!  OOS results recorded

8.2.12 – 8.2.15

Page 12: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 12 of 48 August 2006

HVAC

Qualification – examples of aspects to consider

! DQ – Design of the system, URS

–  (e.g. components, type of air treatment needed, materials of construction)

!  IQ – Verify installation

–  E.g. relevant components, ducting, filters, controls, monitors, sensors, etc.

–  Includes calibration where relevant

Page 13: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 13 of 48 August 2006

HVAC

Qualification (4) Typical parameters to be included in qualification (based on risk

assessment):

!  Temperature

!  Relative humidity

!  Supply, return and exhaust air quantities

!  Room air change rates

!  Room pressures (pressure differentials)

8.2.17

Page 14: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 14 of 48 August 2006

HVAC

Qualification (5) Typical parameters to be included in qualification (based on risk

assessment) (2):

!  Room clean-up rate

!  Particulate matter, microbial matter (viable and non-viable)

!  HEPA filter penetration tests

!  Containment system velocity

!  Warning/alarm systems

8.2.17

Page 15: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 15 of 48 August 2006

HVAC

Qualification (6)

Conduct of the tests:

!  Time intervals and procedure to be defined by the manufacturer

!  Influenced by the type of facility and level of protection

!  See also ISO 14644 for methods of testing

!  Requalification, and change control

8.2.18 – 8.2.20, 8.2.9

Page 16: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 16 of 48 August 2006

HVAC

Qualification (7)

!  Tests performed according to protocols and procedures for the tests

! Results recorded and presented in report (source data kept)

!  Traceability, e.g. devices and standards used, calibration records; and conditions specified

Page 17: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 17 of 48 August 2006

HVAC

Schedule of tests to demonstrate continuing compliance

*Test procedure as per ISO 14644 8. Table 3

Test procedure* and key aspects

Maximum time interval

Objective Test Parameter

Particle counter. Readings and positions

6 months or 12 months depending on Class

Verifies cleanliness Particle count test

Measure pressure difference

12 months Absence of cross-contamination

Air pressure difference

Measure supply and return air, calculate air change rate

12 months Verify air change rates

Airflow volume

Velocity measurement

12 months Verify unidirectional airflow and or containment condition

Airflow velocity

Page 18: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 18 of 48 August 2006

HVAC

Recommended optional strategic tests

*Test procedure as per ISO 14644 8. Table 3

Test procedure* and key aspects

Maximum time interval

Objective Test Parameter

Filter media and filter seal integrity

12 months Verify filter integrity Filter leakage

Airflow direction and pressure differential

12 months Verify absence of cross-contamination

Containment leakage

Time taken maximum 15 minutes

12 months Verify clean-up time Recovery (time)

Airflow direction, documented evidence

12 months Verify required airflow patterns

Airflow visualization

Page 19: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 19 of 48 August 2006

Cleanroom monitoring programme (1)

!  Routine monitoring programme as part of quality assurance

!  Additional monitoring and triggers, e.g.

1. Shutdown

2. Replacement of filter elements

3. Maintenance of air-handling systems

4. Exceeding of established limits

HVAC

Page 20: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 20 of 48 August 2006

Cleanroom monitoring programme (2) Particles and Microbiological contaminants

●  Number of points/locations for monitoring determined, specified, documented in procedure and or protocol

●  Sufficient time for exposure, and suitable sample size ●  Identification and marking of sampling points ●  Definition of transport, storage, and incubation conditions ●  Results to reflect the procedure/protocol followed ●  Define alert and action limits as a function of cleanliness zone/

class

HVAC

See also ISO 14644

Page 21: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 21 of 48 August 2006

air

Example of a sampling point

Cleanroom monitoring programme (3) Cleanrooms should be monitored for microorganisms and particles

HVAC

Page 22: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 22 of 48 August 2006

Definition of Conditions

air

as built air air

at rest in operation

HVAC

Page 23: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 23 of 48 August 2006

Qualification – examples of aspects to consider in qualification (OQ, PQ)

Test

Differential pressure on filters

Turbulent / mixed airflow Description Uni-directional

airflow / LAF

Room differential pressure

Airflow velocity / uniformity

Airflow volume / rate

Parallelism

Airflow pattern

2 2

N/A 2, 3

2, 3 Optional

2 2

2 N/A

2 3

1 := As built (ideally used to perform IQ)

2 = At rest (ideally used to perform OQ)

3 = Operational (ideally used to perform PQ)

HVAC

Page 24: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 24 of 48 August 2006

Test Turbulent / mixed airflow Description Uni-directional

airflow / LAF

Recovery time

Room classification (airborne particle)

Temperature, humidity

N/A 2

2 2,3

N/A 2,3

1 := As built (ideally used to perform IQ)

2 = At rest (ideally used to perform OQ)

3 = Operational (ideally used to perform PQ)

HVAC

Qualification – examples of aspects to consider in qualification (OQ, PQ)

Page 25: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 25 of 48 August 2006

HVAC

Maintenance !  Procedure, programme and records for planned, preventative

maintenance –  e.g. Cleaning of filters, calibration of devices

!  Appropriate training for personnel

!  Change of HEPA filters by suitably trained persons

!  Impact of maintenance on: –  Product quality – Qualification

8.3.1 – 8.3.5

Page 26: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 26 of 48 August 2006

●  Verification of design documentation, including ➤  description of installation and functions ➤  specification of the requirements

● Operating procedures ● Maintenance instructions ● Maintenance records ●  Training logs ●  Environmental records ● Discussion on actions if OOS values ● On site verification (walking around the site)

Inspecting the air-handling system

HVAC

Page 27: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 27 of 48 August 2006

Air-handling systems:

●  Play a major role in the quality of pharmaceuticals ●  Should be designed properly, by professionals ●  Should be treated as a critical system

Conclusion

HVAC

Page 28: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 28 of 48 August 2006

This series of explanations will now be followed by:

●  Group discussion, with a simple exercise ●  Short test

Further proceedings

HVAC

Page 29: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 29 of 48 August 2006

Group Session

HVAC

Page 30: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 30 of 48 August 2006

Group Session – modified layout

MAL = Material Air Lock PAL = Personnel Air Lock

HVAC

Page 31: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 31 of 48 August 2006

Commissioning, Qualification and

validation of Water systems

Supplementary Training Modules on Good Manufacturing Practice

WHO Technical Report Series No 929, 2005. Annex 3

Page 32: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 32 of 48 August 2006

HVAC

Objectives

! To understand key issues in –  commissioning, –  qualification and –  maintenance of

HVAC and Water systems

7.

Page 33: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 33 of 48 August 2006

Objectives To discuss the operational considerations of water systems

including:

●  Start up, commissioning and qualification

●  Monitoring

●  Maintenance

●  System reviews

Water for Pharmaceutical Use

7.

Page 34: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 34 of 48 August 2006

Start up and commissioning

!  Precursor to qualification and validation

!  Should be planned, well defined, well documented

!  Includes setting to work

!  Includes system set-up

!  Includes recording of system performance parameters

!  Controls loop tuning

Water for Pharmaceutical Use

7.1

Page 35: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 35 of 48 August 2006

Qualification

!  WPU systems are "direct impact systems"

!  Therefore stages to be considered in qualification should include DQ, IQ, OQ, PQ

!  DQ: Design review influenced by source water and required water quality

!  IQ: Installation verification of the system

Water for Pharmaceutical Use

7.2

Page 36: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 36 of 48 August 2006

Qualification

!  OQ: operational qualification

!  Presentation focusing on PQ

!  PQ demonstrates consistent and reliable performance of the system

!  Three phase approach recommended over extended period – proves reliability and robustness

Water for Pharmaceutical Use

7.2

Page 37: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 37 of 48 August 2006

Phase 1 (1)

!  A test period of 2–4 weeks - monitoring the system intensively

!  System to operate continuously without failure or performance deviation

The following should be included in the testing approach:

!  Undertake chemical and microbiological testing in accordance with a defined plan

Water for Pharmaceutical Use

7.2

Page 38: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 38 of 48 August 2006

Phase 1 (2)

!  Sample daily: –  incoming feed-water –  after each step in the purification process –  each point of use and at other defined sample points

!  Develop: –  appropriate operating ranges –  and finalize operating, cleaning, sanitizing and maintenance

procedures

Water for Pharmaceutical Use

7.2

Page 39: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 39 of 48 August 2006

Phase 1 (3)

!  Demonstrate production and delivery of product water of the required quality and quantity

!  Use and refine the standard operating procedures (SOPs) for operation, maintenance, sanitization and troubleshooting

!  Verify provisional alert and action levels

!  Develop and refine test-failure procedure

Water for Pharmaceutical Use

7.2

Page 40: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 40 of 48 August 2006

Phase 2 (1)

!  A further test period of 2–4 weeks – further intensive monitoring the system

!  Deploying all the refined SOPs after the satisfactory completion of phase 1

!  Sampling scheme generally the same as in phase 1

!  Water can be used for manufacturing purposes during this phase

Water for Pharmaceutical Use

7.2

Page 41: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 41 of 48 August 2006

Phase 2 (2)

Demonstrate:

!  Consistent operation within established ranges

!  Consistent production and delivery of water of the required quantity and quality when the system is operated in accordance with the SOPs.

Water for Pharmaceutical Use

7.2

Page 42: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 42 of 48 August 2006

Phase 3

!  Over one year after the satisfactory completion of phase 2

!  Water can be used for manufacturing purposes during this phase

Demonstrate: –  extended reliable performance –  that seasonal variations are evaluated

!  Sample locations, sampling frequencies and tests should be reduced to the normal routine pattern based on established procedures proven during phases 1 and 2

Water for Pharmaceutical Use

7.2

Page 43: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 43 of 48 August 2006

Ongoing system monitoring

!  After Phase 3 – system review needed

!  Based on review including results, establish a routine monitoring plan

!  Monitoring to include a combination of on-line monitoring and off- line sample testing

!  Data analysed for trends

Water for Pharmaceutical Use

7.3

Page 44: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 44 of 48 August 2006

Ongoing system monitoring (2)

!  Monitoring parameters to include: –  flow, pressure, temperature, conductivity, TOC

!  Samples taken: –  From points of use, and specific sample points –  In a similar way how water is used in service

!  Tests to include physical, chemical and microbial attributes

Water for Pharmaceutical Use

7.3

Page 45: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 45 of 48 August 2006

Maintenance

A controlled, documented maintenance programme covering:

!  Defined frequency with plan and instructions

!  Calibration programme

!  SOPs for tasks

!  Control of approved spares

!  Record and review of problems and faults during maintenance

Water for Pharmaceutical Use

7.4

Page 46: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 46 of 48 August 2006

System review

!  WPU (PW, HPW and WFI) systems to be reviewed at appropriate regular intervals

!  Review team includes engineering, QA, operations and maintenance

Water for Pharmaceutical Use

7.5

Page 47: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 47 of 48 August 2006

System review (2)

!  The review to cover, e.g. –  changes made since the last review; –  system performance; –  reliability; –  quality trends; –  failure events; –  investigations; –  out-of-specifications results from monitoring; –  changes to the installation; –  updated installation documentation; –  log books; and –  the status of the current SOP lists

Water for Pharmaceutical Use

7.5

Page 48: Supplementary Training Modules on Good Manufacturing Practice · Supplementary Training Modules on Good Manufacturing Practice WHO Technical Report Series, No. 937, 2006. Annex 4

Validation | Slide 48 of 48 August 2006

Validation

!  Group session