Superior X-Ray Tube Co.
Since 1987, Superior X-Ray Tube Company has been manufacturing quality x-ray tubes. We take pride inour diverse product line which is used in a wide range of applications, including, medical, dental, thicknessgauging, food inspection, bone densitometry, Non-Destructive Testing (NDT), and X-Ray Fluorescence
(XRF) Spectroscopy. We have a global customer base including Japan, China, Korea, Europe, Canada, andthe USA.
Superior X-Ray manufactures a wide variety of x-ray tubes with focal spots ranging from 50 microns to 2.8
mm and with operating voltages from 35kV up to 130 kV using standard tungsten target material. All of ourtubes come with a 15 month pro-rated warranty covering defects due to materials or workmanship when
used within the tubes specified operating limits.
Other target materials and custom tube designs are available in minimum order sizes of 100 pieces. Pleasecontact us to discuss custom designs.
We would be happy to assist you in meeting your current and future x-ray tube needs. Our companyinformation is as follows:
Superior X-ray Tube Company Phone: 815-338-4424 Fax: 815-338-9452
1220 Claussen Drive email: [email protected]
Woodstock, IL 60098 Website: www.superiorxraytube.com
U.S.A
Thank you for taking the time to review our catalog, we look forward to hearing from you.
Sincerely,
Mark McDonnellPresident
Superior X-ray Tube Company
-
8/18/2019 Superior X-ray Tube Co.
3/58
-
8/18/2019 Superior X-ray Tube Co.
4/58
Pa
GLASS WINDOW TUBES
SXR-80PEANUT SERIES
SEE PAGES 1 THRU 6
SXR-15SEE PAGES 7 AND 8
SXR-15PTSEE PAGES 9 AND 10
SXR-90-15EE PAGES 11 THRU 16
-
8/18/2019 Superior X-ray Tube Co.
5/58
Pag
GLASS WINDOW TUBES
SXR-100-15SEE PAGES 17 AND 18
SXR-100-25SEE PAGES 19 AND 20
SXR-100-20 SERIESEE PAGES 21 THRU 24
SXR-100-45EE PAGES 25 THRU 28
-
8/18/2019 Superior X-ray Tube Co.
6/58
Pag
BERYLLIUM WINDOW TUBES
SXR-50BWPEANUT SERIES
EE PAGES 29 THRU 34
SXR-212BWSEE PAGES 41 AND 42
SXR-80BWPEANUT SERIES
EE PAGES 35 THRU 40
-
8/18/2019 Superior X-ray Tube Co.
7/58
Pa
BERYLLIUM WINDOW TUBES
SXR-35-05BWSEE PAGES 45 AND 46
SXR-130-05BWSEE PAGES 49 AND 50
SXR-80-05BWSEE PAGES 47 AND 48
-
8/18/2019 Superior X-ray Tube Co.
8/58
Pag
Superior X-ray Tube Company
Customer Tube Request
In order to select an x-ray tube that will work properly in your application, please provide us with as much
information as possible.
kVp: _____________What peak or continuous voltage will be applied to the tube?
Generator Type: _____________
Self-Rectified, Full Wave Rectified, or Constant Potential (DC)
Tube Operation: _____________Will the voltage be applied to (1) the anode with the cathode cup grounded, or (2) the cathode cup with the anod
grounded, or (3) bipolar operation with ½ the voltage on the anode and ½ the voltage on the cathode?
mA: _____________
What is the maximum tube current at kVp ?
Duty Cycle: _____________ (On Time) _____________ (Off Time)
What is the exposure time (i.e. time the tube is on) and the time between exposures?
Focal Spot: _____________What is the desired focal spot size?
Anode Target Angle: _____________
Desired X-ray Beam Coverage Area: _____________When the x-ray beam reaches the person/object under study, the area receiving x-ray energy is referred to as the
“coverage area”. The angle of the anode and the distance from the tube target to the person/object determines thmaximum x-ray tube coverage. The coverage may be reduced by filters or collimators inside the x-ray tube head
Tube Cooling Method: ____________________________________________
Oil, Air, Silicone Insulating Material, etc.?Sealed or open oil bath?
Any additional forced cooling (fins, radiators, heat exchangers)?
Physical Tube Constraints: _________________________________________Please detail any constraints on the physical dimensions of the tube.
Type of Application: ______________________________________________
Dental, Medical, Industrial?
Other Requirements (For example, do you require a shield, flange, lead wires, or special anode/cathodebasing?): __________________________________________________________________________________
__________________________________________________________________________________________
-
8/18/2019 Superior X-ray Tube Co.
9/58
Pa
Operating Data
Available focal spot size (mm):
0.20
Available target angle:14°
Anode construction:
Vacuum cast copper with tungsten target
Cathode construction:Vacuum tube nickel with tungsten filament
Max. Tube operating voltage:80 kVp Full wave rectified
Max. Continuous tube currentat 80 kVp:0.60 mA
Inherent filtration/Window material:
1.5 mm/Glass
Cooling Method:High dielectric transformer oil
SXR-80-14-2P0.20 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
10/58
Pa
SXR-80-14-2P
-
8/18/2019 Superior X-ray Tube Co.
11/58
Pa
Operating Data
Available focal spot size (mm):1.0
Available target angle:
14°
Anode construction:Vacuum cast copper with tungsten target
Cathode construction:
Vacuum tube nickel with tungsten filament
Max. Tube operating voltage:
80 kVp Full wave rectified
Max. Continuous tube currentat 80 kVp:
0.75 mA
Inherent filtration/Window material:1.5 mm/Glass
Cooling Method:
High dielectric transformer oil
SXR-80-10P1.0 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
12/58
Pa
SXR-80-14-10P
-
8/18/2019 Superior X-ray Tube Co.
13/58
Pa
Operating Data
Available focal spot size (mm):
0.50
Available target angle:14°
Anode construction:
Vacuum cast copper with tungsten target
Cathode construction:Vacuum tube nickel with tungsten filament
Max. Tube operating voltage:
80 kVp Full wave rectified
Max. Continuous tube current
at 80 kVp:0.70 mA
Inherent filtration/Window material:
1.5 mm/Glass
Cooling Method:High dielectric transformer oil
SXR-80-20-5P0.5 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
14/58
Pa
SXR-80-20-5P
-
8/18/2019 Superior X-ray Tube Co.
15/58
Pa
Operating Data
Available focal spot size (mm):1.0
Available target angle:
15°
Anode construction:Vacuum cast copper withtungsten target
Cathode construction:Vacuum tube nickel with tungstenfilament.
Max. Tube operating voltage:
100 kVp / 110 kVpInverse full wave and self rectified.
Max. Continuous tube currentat 100 kVp:1.0 mA
Inherent filtration/Window material:
1.5 mm/Glass
Cooling Method:High dielectric transformer oil
SXR-151.0 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
16/58
Pa
SXR-15
-
8/18/2019 Superior X-ray Tube Co.
17/58
Pa
Operating Data
Available focal spot sizes (mm):
1.0
Available target angles:15°
Anode construction:
Vacuum cast copper with tungsten target
Cathode construction:Vacuum tube nickel with tungsten filament
Max. Tube operating voltage:100 kVp / 110 kVp
Inverse full wave and self rectified
Max. Continuous tube currentat 100 kVp:
1.0 mA
Inherent filtration/Window material:1.5 mm/Glass
Cooling Method:
High dielectric transformer oil
SXR-15PT1.0 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
18/58
Pag
SXR-15PT
-
8/18/2019 Superior X-ray Tube Co.
19/58
Pag
Operating Data
Available focal spot sizes (mm):
0.5
Available target angles:15°
Anode construction:
Vacuum cast copper with tungsten target
Cathode construction:Vacuum tube nickel with tungsten filament
Max. Tube operating voltage:
100 kVp Full wave rectified
Max. Continuous tube currentat 100 kVp:
1.0 mA
Inherent filtration/Window material:1.5 mm/Glass
Cooling Method:
High dielectric transformer oil
SXR-90-15FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
20/58
Pag
SXR-90-15
-
8/18/2019 Superior X-ray Tube Co.
21/58
Pag
Operating Data
Available focal spot sizes (mm):0.5
Available target angles:
15°
Anode construction:Vacuum cast copper with tungsten target
Cathode construction:
Vacuum tube nickel with tungsten filament
Max. Tube operating voltage:
100 kVp Full wave rectified
Max. Continuous tube currentat 100 kVp:
1.0 mA
Inherent filtration/Window material:1.5 mm/Glass
Cooling Method:
High dielectric transformer oil
SXR-90-15M0.5 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
22/58
Pag
SXR-90-15M
-
8/18/2019 Superior X-ray Tube Co.
23/58
Pag
Operating Data
Available focal spot sizes (mm):0.5
Available target angles:
15°
Anode construction:Vacuum cast copper with tungsten target
Cathode construction:
Vacuum tube nickel with tungsten filament
Max. Tube operating voltage:
100 kVp Full wave rectified
Max. Continuous tube currentat 100 kVp:
1.0 mA
Inherent filtration/Window material:
1.5 mm/Glass
Cooling Method:
High dielectric transformer oil
SXR-90-15M20.5 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
24/58
Pag
SXR-90-15M2
-
8/18/2019 Superior X-ray Tube Co.
25/58
Pag
Operating Data
Available focal spot sizes (mm):
1.0
Available target angles:15°
Anode construction:
Vacuum cast copper with tungsten target
Cathode construction:Vacuum tube nickel with tungsten filament
Max. Tube operating voltage:100 kVp Full wave rectified
Max. Continuous tube currentat 100 kVp:1.5 mA
Inherent filtration/Window material:
1.5 mm/Glass
Cooling Method:High dielectric transformer oil
SXR-100-151.0 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
26/58
Pag
SXR-100-15 – 1.0 FOCAL SPOT
60Hz
SXR-100-15 WITH SHIELD
-
8/18/2019 Superior X-ray Tube Co.
27/58
Pag
Operating Data
Available focal spot sizes (mm):1.0
Available target angles:
25°
Anode construction:Vacuum cast copper with tungsten targe
Cathode construction:
Vacuum tube nickel with tungsten filame
Max. Tube operating voltage:
100 kVp Full wave rectified
Max. Continuous tube currentat 100 kVp:
1.5 mA
Inherent filtration/Window material:1.5 mm/Glass
Cooling Method:
High dielectric transformer oil
SXR-100-251.0 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
28/58
Pag
SXR-100-25 – 1.0 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
29/58
Pag
Operating Data
Available focal spot sizes (mm):
2.8
Available target angles:20°
Anode construction:
Vacuum cast copper with tungsten target
Cathode construction:Vacuum tube nickel with tungsten filament
Max. Tube operating voltage:125 kvp Full wave rectified
Max. Continuous tube currentat 100 kVp:3.0 mA
Inherent filtration/Window material:
1.5 mm/Glass
Cooling Method:High dielectric transformer oil
SXR-100-202.8 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
30/58
Pag
SXR-100-20
-
8/18/2019 Superior X-ray Tube Co.
31/58
Pag
Operating Data
Available focal spot sizes (mm):
2.0
Available target angles:20°
Anode construction:
Vacuum cast copper with tungsten target
Cathode construction:Vacuum tube nickel with tungsten filament
Max. Tube operating voltage:125 kvp Full wave rectified
Max. Continuous tube currentat 100 kVp:3.0 mA
Inherent filtration/Window material:
1.5 mm/Glass
Cooling Method:High dielectric transformer oil
SXR-100-20-22.0 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
32/58
Pag
SXR-100-20-2
SXR-100-20-2 WITH SHIELD
-
8/18/2019 Superior X-ray Tube Co.
33/58
Pag
Operating Data
Available focal spot sizes (mm):2.0
Available target angles:
45°
Anode construction:Vacuum cast copper with tungsten target
Cathode construction:
Vacuum tube nickel with tungsten filamen
Max. Tube operating voltage:125 kVp Full wave rectified
Max. Continuous tube currentat 125 kVp:3.0 mA
Inherent filtration/Window material:
1.5 mm/Glass
Cooling Method:High dielectric transformer oil
SXR-100-45 2.0 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
34/58
Pag
SXR-100-45
-
8/18/2019 Superior X-ray Tube Co.
35/58
Pag
Operating Data
Available focal spot sizes (mm):1.5
Available target angles:
45°
Anode construction:Vacuum cast copper with tungsten target
Cathode construction:
Vacuum tube nickel with tungsten filament
Max. Tube operating voltage:
125 kVp Full wave rectified
Max. Continuous tube currentat 125 kVp:
1.5 mA
Inherent filtration/Window material:1.5 mm/Glass
Cooling Method:
High dielectric transformer oi
SXR-100-45M1.5 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
36/58
Pag
SXR-100-45M
-
8/18/2019 Superior X-ray Tube Co.
37/58
Pag
Operating Data
Available focal spot size (mm):
0.2
Available target angle:14°
Anode construction:
Vacuum cast copper with tungsten target
Cathode construction:Vacuum tube nickel with tungsten filament
Max. Tube operating voltage:
50 kVp Full wave rectified
Max. Continuous tube current at 50 kVp:
1.0 mA
Inherent filtration/Window material:.25 mm/Beryllium
Cooling Method:
High dielectric transformer oil
Note:
See Next Page for View A-A
(Mounting flange)
SXR-50-14-2BWP0.2 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
38/58
Pag
-
SXR-50-14-2BWP
View A-A (Mounting Flange)
-
8/18/2019 Superior X-ray Tube Co.
39/58
Pag
Operating Data
Available focal spot size (mm):1.0
Available target angle:
14°
Anode construction:Vacuum cast copper with tungsten target
Cathode construction:
Vacuum tube nickel with tungsten filament
Max. Tube operating voltage:
50 kVp Full wave rectified
Max. Continuous tube current at 50 kVp:1.2 mA
Inherent filtration/Window material:
.25 mm/Beryllium
Cooling Method:High dielectric transformer oil
Note:
See Next Page for View A-A(Mounting flange)
SXR-50-14-10BW1.0 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
40/58
Pag
SXR-50-14-10BWP
View A-A (Mounting Flange)
-
8/18/2019 Superior X-ray Tube Co.
41/58
Pag
Operating Data
Available focal spot size (mm):0.5
Available target angle:
20°
Anode construction:Vacuum cast copper with tungsten target
Cathode construction:
Vacuum tube nickel with tungsten filament
Max. Tube operating voltage:
50 kVp Full wave rectified
Max. Continuous tube current at 50 kVp:1.2 mA
Inherent filtration/Window material:
.25 mm/Beryllium
Cooling Method:High dielectric transformer oil
Note:
See Next Page for View A-A
(Mounting flange)
SXR-50-20-5BWP0.5 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
42/58
Pag
SXR-50-20-5BWP
View A-A (Mounting Flange)
-
8/18/2019 Superior X-ray Tube Co.
43/58
Pag
Operating Data
Available focal spot sizes (mm):0.2
Available target angles:
14°
Anode construction:Vacuum cast copper with tungsten targ
Cathode construction:
Vacuum tube nickel with tungsten filam
Max. Tube operating voltage:80 kVp Full wave rectified
Max. Continuous tube currentat 80 kVp:
0.60 mA
Inherent filtration/Window material:.25 mm/Beryllium
Cooling Method:
High dielectric transformer oil
Note:
See Next Page for View A-A
(Mounting flange)
SXR-80-14-2BWP0.2 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
44/58
Pag
SXR-80-14-2BWP
View A-A (Mounting Flange)
-
8/18/2019 Superior X-ray Tube Co.
45/58
Pag
Operating Data
Available focal spot sizes (mm):
1.0
Available target angles:14°
Anode construction:
Vacuum cast copper with tungsten targ
Cathode construction:Vacuum tube nickel with tungsten filam
Max. Tube operating voltage:
80 kVp Full wave rectified
Max. Continuous tube currentat 80 kVp:
0.75 mA
Inherent filtration/Window material:.25 mm/Beryllium
Cooling Method:
High dielectric transformer oil
Note:
See Next Page for View A-A
(Mounting flange)
SXR-80-14-10BWP1.0 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
46/58
Pag
SXR-80-14-10BWP
View A-A (Mounting Flange)
-
8/18/2019 Superior X-ray Tube Co.
47/58
Pag
Operating Data
Available focal spot size (mm):0.5
Available target angle:
20°
Cathode construction:Vacuum tube nickel with tungstenfilament
Anode construction:Vacuum cast copper with tungsten tar
Max. Tube operating voltage:80 kVp Full wave rectified
Max. Continuous tube currentat 80 kVp:0.70 mA
Inherent filtration/Window material:
.25 mm/Beryllium
Cooling Method:High dielectric transformer oil
Note:
See Next Page for View A-A(Mounting flange)
SXR-80-20-5BWP0.5 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
48/58
Pag
SXR-80-20-5BWP
View A-A (Mounting Flange)
-
8/18/2019 Superior X-ray Tube Co.
49/58
Pag
Operating Data
Available focal spot sizes (mm):
0.5
Available target angles:20°
Anode construction:
Vacuum cast copper with tungsten target
Cathode construction:Vacuum tube nickel with tungsten filament
Max. Tube operating voltage:
60 kVp Full wave rectified
Max. Continuous tube current
at 60 kVp:1.5 mA
Inherent filtration/Window material:
.75 mm/Beryllium
Cooling Method:High dielectric transformer oil
Note:See Next Page for View A-A(Mounting flange)
SXR-212BW0.5 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
50/58
Pag
SXR-212BW
View A-A (Mounting Flange)
Type 1 Flange
Type 2 Flange
-
8/18/2019 Superior X-ray Tube Co.
51/58
Pag
Operating Data
Available focal spot sizes (mm):
0.5
Available target angles:20°
Anode construction:
Vacuum cast copper with tungsten target
Cathode construction:Vacuum tube nickel with tungsten filament
Max. Tube operating voltage:
130 kVp Full wave rectified
Max. Continuous tube current
at 130 kVp:1.0 mA
Inherent filtration/Window material:
.75 mm/Beryllium
Cooling Method:High dielectric transformer oil
Note:See Next Page for View A-A(Mounting flange)
SXR-130-20-5BW0.5 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
52/58
Pag
SXR-130-20-5BW
-
8/18/2019 Superior X-ray Tube Co.
53/58
Pag
Operating Data
Available focal spot sizes (mm):
0.05
Available target angles:25°
Anode construction:
Vacuum cast copper with tungsten target
Cathode construction:Vacuum tube nickel with tungsten filament
Max. Tube operating voltage:
35 kVp Full wave rectified
Max. Continuous tube current
at 35 kVp:300 uA
Inherent filtration/Window material:
.75 mm/Beryllium
Cooling Method:Forced Air
See Next Page for View A-A
(Mounting flange)
SXR-35-05BW0.05 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
54/58
Pag
SXR-35-05BW
View A-A (Mounting Flange)
Type 2 FlangeType 1 Flange
-
8/18/2019 Superior X-ray Tube Co.
55/58
Pag
Operating Data
Available focal spot sizes (mm):0.05
Available target angles:
16°
Anode construction:Vacuum cast copper with tungsten target
Cathode construction:
Vacuum tube nickel with tungsten filament
Max. Tube operating voltage:80 kVp Full wave rectified
Max. Continuous tube currentat 80 kVp:
300 uA
Inherent filtration/Window material:.75 mm/Beryllium
Cooling Method:
High dielectric transformer oil
Note:
See Next Page for View A-A(Mounting flange)
SXR-80-05BW0.05 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
56/58
Pag
SXR-80-05BW
View A-A (Mounting Flange)
Type 2 FlangeType 1 Flange
-
8/18/2019 Superior X-ray Tube Co.
57/58
Pag
Operating Data
Available focal spot sizes (mm):0.05
Available target angles:
20°
Anode construction:Vacuum cast copper with tungstentarget
Cathode construction:Vacuum tube nickel with tungstenfilament
Max. Tube operating voltage:
130 kVp Full wave rectified
Max. Continuous tube currentat 130 kVp:
300 uA
Inherent filtration/Window mater.75 mm/Beryllium
Cooling Method:
High dielectric transformer oil
Note:
See Next Page for View A-A
(Mounting flange)
SXR-130-05BW0.05 FOCAL SPOT
-
8/18/2019 Superior X-ray Tube Co.
58/58
Pag
View A-A (Mounting Flange)
T pe 2 Flange