strong case - synthes.vo.llnwd.netsynthes.vo.llnwd.net/o16/llnwmb8/us mobile/synthes... ·...

8
We’ve built a STRONG CASE for changing the way you sterilize. Reusable Sterilization Container System

Upload: others

Post on 08-Jul-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: STRONG CASE - synthes.vo.llnwd.netsynthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes... · validation testing, but also ensuring the sets are within the cleared indications of the

We’ve built a

STRONG CASE for changing the way you sterilize.

Reusable Sterilization Container System

Page 2: STRONG CASE - synthes.vo.llnwd.netsynthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes... · validation testing, but also ensuring the sets are within the cleared indications of the

Retention Plate

Filter

Optional Protective Plate

Lid

In a healthcare environment focused on cost optimization, waste reduction and infection control, the DePuy Synthes Companies of Johnson & Johnson now provide an effective solution that will meet your facility’s sterilization needs while helping to minimize the impact on budget and the environment.

This system offers an assortment of rigid, reusable containers designed to sterilize medical devices and maintain that sterility for up to 180 days. And it is the ONLY container system validated to provide effective sterilization for the DePuy Synthes Companies (Trauma & CMF) instruments and implants.

Introducing the DePuy SynthesREUSABLE STERILIZATION CONTAINER SYSTEM

Offset perforated retention plate provide no direct path through the lid.

Single use disposable fi lter to provide 100% barrier effi cacy with each use.

Silicone gasket secured with a stainless steel dual action latching mechanism.

Containers are available with either perforated or solid bottom.

Discover the solution that will make a MATERIAL DIFFERENCE.

2

Reusable Sterilization Container SystemReusable Sterilization Container SystemReusable Sterilization Container SystemReusable Sterilization Container System

Page

Page 3: STRONG CASE - synthes.vo.llnwd.netsynthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes... · validation testing, but also ensuring the sets are within the cleared indications of the

Benefi ts and Cost Savings

Compatibility with DePuy Synthes Sets*

Blue Wrap Waste Reduction

Introducing the DePuy SynthesREUSABLE STERILIZATION CONTAINER SYSTEM

System is comprised of an anodized

aluminum rivet-less case assembly.

Extended sizes available for large orthopeadic sets.

Discover the solution that will make a MATERIAL DIFFERENCE.

*For DePuy Synthes Companies (Trauma & CMF) instrument and implant sets.

3Page

Page 4: STRONG CASE - synthes.vo.llnwd.netsynthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes... · validation testing, but also ensuring the sets are within the cleared indications of the

Compatibility with DePuy Synthes SetsSelecting a compatible container for an orthopaedic instrument/implant set involves not only validation testing, but also ensuring the sets are within the cleared indications of the container. The number of indicated lumens, materials, and the Volume-to-Vent ratio are some of the key considerations when choosing a proper container.

The DePuy Synthes Reusable Sterilization Container System reduces the ambiguity of set sterility by ensuring compatibility with all DePuy Synthes Trauma and DePuy Synthes CMF sets.

Benefi ts and Cost SavingsBlue wrap can rip, which requires the set to be re-sterilized. When tears in blue wrap are discovered at the last minute, re-sterilization could mean more anesthesia time for the patient and additional cost and time added to the case.

Over time the cost to purchase blue wrap, plus the labor time needed to wrap sets by hospital personnel can add up to signifi cant costs for the hospital. Switching to the DePuy Synthes Reusable Sterile Container System can reduce overall hospital costs.

ROI Analysis for the Reusable Sterilization

Container System

Figure 1: Sample cost calculator to guide facility’s specifi c analysis on possible cost reduction using the Reusable Sterilization Container System.

This cost analysis is only for those sets that will be put into containers. *Ask your DePuy Synthes Sales Consultant for a detailed cost analysis for your facility.

Estimated 5 year savings Estimated 10 year savings

Return On Investment (ROI)

ROI on capital outlay in years*

Capital Outlay for Extended PriceDePuy Synthes Containers

Current Spend on Disposables (sterilization wrap, autoclave tape, waste removal)

Disposable Cost for DePuy Synthes Containers (arrows, data cards, fi lters)

Costs to consider when comparing Reusable Sterilization Containers vs. Blue Wrap

Total Annual Cost for DePuy Synthes Disposables

4

Reusable Sterilization Container SystemReusable Sterilization Container SystemReusable Sterilization Container SystemReusable Sterilization Container System

Page

Quantity X Average Container Price

Cost/Processing Cycle X Number of Annual Cycles

Price/Box X Number of Boxes

Total Annual Usage Cost on Current Disposables

Page 5: STRONG CASE - synthes.vo.llnwd.netsynthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes... · validation testing, but also ensuring the sets are within the cleared indications of the

5Page

Blue Wrap Waste ReductionReusable Sterilization Containers provide an economical solution by reducing the volume of blue wrap disposed of in landfi lls or incinerated; while also allowing facility’s to achieve cost savings.

Did you know?…

• Approximately 19% of the waste stream generated by surgical services is blue sterile wrap.1

• Over the course of about three years, by switching over from blue wrap to sterilization containers,2 one hospital estimated a reduction in waste of 342.6 cubic yards of landfi ll, more than enough to approximately fi ll a football fi eld!

• There are over 5000 hospitals in the US and Canada.

Do the math…

The opportunity to decrease waste (and waste related costs) exists with the DePuy Synthes Reusable Sterilization Container System.

A full portfolio of sizes is available in both perforated and non-perforated confi gurations to suit your facility’s sterilization needs.

1. “Reusable totes, blue wrap recycling and composting”. Environnmental best practices for health care facilities, November 2002 JHACO environment of care standards 1.3 2.3 4.0”

2. “Going from blue to green”. www.surgicalproductsmag.com April 2011

Page 6: STRONG CASE - synthes.vo.llnwd.netsynthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes... · validation testing, but also ensuring the sets are within the cleared indications of the

PRODUCT PART NUMBERS

Descriptions and Dimensions

DePuy Synthes DePuy Synthes Volume to Vent Weight Outer Dimensions Inner Dimensions Part Number Part Description Ratio (in3/in2) (lbs.) (L x W x H), in. (L x W x H), in.

PERFORATED BOTTOM CONTAINERS

62.006.001 Full-One Level, Perforated Base 26 8.5 23.1 x 12.4 x 4.5 21.0 x 11.4 x 4.2

62.006.002 Full-Two Level, Perforated Base 31 8.8 23.1 x 12.4 x 5.3 21.0 x 11.4 x 5.1

62.006.003 Full-Three Level, Perforated Base 42 9.4 23.1 x 12.4 x 7.0 21.0 x 11.4 x 6.8

62.009.004 Extended-Four Level, Perforated Base 56 9.7 25.2 x 12.4 x 8.5 23.0 x 11.4 x 8.4

62.009.005 Extended-Five Level, Perforated Base 63 10.3 25.2 x 12.4 x 9.5 23.0 x 11.4 x 9.4

SOLID BOTTOM CONTAINERS

62.016.001 Full-One Level, Solid Base 52 7.8 23.1 x 12.4 x 4.5 21.0 x 11.4 x 4.2

62.016.002 Full-Two Level, Solid Base 62 7.9 23.1 x 12.4 x 5.3 21.0 x 11.4 x 5.1

62.016.003 Full-Three Level, Solid Base 83 8.5 23.1 x 12.4 x 7.0 21.0 x 11.4 x 6.8

62.019.004 Extended-Four Level, Solid Base 112 9.1 25.2 x 12.4 x 8.5 23.0 x 11.4 x 8.4

62.019.005 Extended-Five Level, Solid Base 126 10.0 25.2 x 12.4 x 9.5 23.0 x 11.4 x 9.4

LIFTING PLATFORMS

62.006.010 Lifting Platform for Full Container Not Applicable 2.6 20.5 x 10.6 x 1.3 Not Applicable

62.009.010 Lifting Platform for Extended Container Not Applicable 2.9 22.5 x 10.6 x 1.3 Not Applicable

6

Reusable Sterilization Container SystemReusable Sterilization Container SystemReusable Sterilization Container SystemReusable Sterilization Container System

Page

Page 7: STRONG CASE - synthes.vo.llnwd.netsynthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes... · validation testing, but also ensuring the sets are within the cleared indications of the

Additional Items

Part Number DePuy Synthes Part Description

CONSUMABLES

62.010.002 Filters (1,000 pcs)

62.010.004 Tamper Evident Arrows (1,000 pcs)

62.010.005 Data Cards (500 pcs)

REPLACEMENT PARTSLids

62.006.020 Lid for Full Sterilization Container

62.009.021 Lid for Extended Sterilization Container

Bases

62.006.031 Perforated Base for Full-One Level Container

62.006.032 Perforated Base for Full-Two Level Container

62.006.033 Perforated Base for Full-Three Level Container

62.009.034 Perforated Base for Extended-Four Level Container

62.009.035 Perforated Base for Extended-Five Level Container

62.016.031 Solid Base for Full-One Level Container

62.016.032 Solid Base for Full-Two Level Container

62.016.033 Solid Base for Full-Three Level Container

62.019.034 Solid Base for Extended-Four Level Container

62.019.035 Solid Base for Extended-Five Level Container

Filter Retention Plates

62.010.001 Optional Protective Plate

62.010.003 Filter Retention Plate, Top

62.010.006 Filter Retention Plate, Bottom

7Page

Page 8: STRONG CASE - synthes.vo.llnwd.netsynthes.vo.llnwd.net/o16/LLNWMB8/US Mobile/Synthes... · validation testing, but also ensuring the sets are within the cleared indications of the

www.depuysynthes.com

©DePuy Synthes 2015. All rights reserved. DSUS/TRM/0115/0472 1/15 JC

Manufactured by (United States):

Synthes USA Products, LLC1302 Wrights Lane EastWest Chester, PA 19380Telephone: (610) 719-5000To order: (800) 523-0322Fax: (610) 251-9056

Legal manufacturer (Canada):

Synthes (Canada) Ltd.2566 Meadowpine BoulevardMississauga, Ontario L5N 6P9Telephone: (905) 567-0440To order: (800) 668-1119Fax: (905) 567-3185

This is the only container system validated for the DePuy Synthes Companies (Trauma & CMF) instruments and implant sets that provides:

Benefi ts and Cost Savings

Compatibility with DePuy Synthes Sets*

Blue Wrap Waste Reduction

To learn more about the DePuy Synthes Reusable Sterilization Container System, contact your DePuy Synthes Consultant or visit: www.depuysynthes.com.

Indications (See User’s Manual for complete indications.)

The DePuy Synthes Reusable Sterilization Container System is a device intended to be used to enclose other medical devices to be sterilized by a healthcare provider. It allows sterilization of the enclosed medical devices and maintains sterility of the devices until used for a maximum of 180 days.

DePuy Synthes containers are suitable for dynamic air removal (pre-vacuum) steam sterilization when used according to the instructions for use.

Reusable lifting platforms are intended to hold enclosed medical devices above the fi lter areas of a perforated bottom container during sterilization and storage of the container.

Data cards are used to record information regarding a specifi c sterilization process load. Filter media allows ingress and egress of sterilant while providing a microbial barrier. Tamper evident arrows provide a visual indication that the container system has not been inadvertently opened prior to use. Each arrow contains a modality-specifi c (pre-vacuum steam) external process indicator that serves as a visual indication that the system has been exposed to a specifi c sterilization cycle parameter. Data cards, fi lters, and tamper evident arrows are single use only.

WARNING: In the USA, this product has labeling limitations. See User’s Manual for complete information.

CAUTION: Some products listed in this brochure may not have been licensed in accordance with Canadian law and may not be for sale in Canada. Please contact your sales consultant for items approved in Canada.

Dynamic Air Removal (Pre-Vacuum) Steam

Sterilization Method

Exposure Temperature: 270°F (132°C)

Pre-Conditioning Pulses: 3

Exposure Time: 4 Minutes

Dry Time Cycle: 30 Minutes

Minimum Cool Time: 60 minutes (may vary according to load contents)

Stacking Not Permitted

Cycle Parameters

25 lbs.

(Container plus contents)

Total System Weight

Orthopedic Medical Devices including Lumen (Cannulated) Devices;

Devices or Device Confi gurations with conjoined surfaces which meet, touch or unite; Mated Surfaces.

Materials: Intrinsically stable metals. Composites, thermoplastics and thermosetting polymers with constant use temperatures above 135°C.

Types of Medical Devices and Materials Validated for Use with DePuy Synthes Container

*For DePuy Synthes Companies (Trauma & CMF) instrument and implant sets.

Reusable Sterilization Container SystemReusable Sterilization Container SystemReusable Sterilization Container SystemReusable Sterilization Container SystemReusable Sterilization Container System