Strain Gauge Trainer
ST2304
Operating Manual
Ver 1.1
An ISO 9001 : 2000 company
94-101, Electronic Complex
Pardesipura, Indore- 452010, IndiaTel : 91-731- 2570301/02,4211100 Fax: 91- 731- 2555643e mail : [email protected]
Website : www.scientech.bz
Toll free : 1800-103-5050
-
7/29/2019 Strain Gauge Trainer
2/22
-
7/29/2019 Strain Gauge Trainer
3/22
ST2304
Strain Gauge Trainer
ST2304
Table of Contents
1. Introduction 4
2. Features 4
3. Technical Specifications 5
4. Operating Instructions & Panel Control Description 6
5. Functional Description of Blocks 7
6. Theory 10
7. Experiments Experiment 1 15Study of Strain Measurement using Strain Gauges and Cantilever
assembly
Experiment 2 18Determination of Linear Range of operation of Strain Measurement
Experiment 3 20Determination Sensitivity of Trainer
8. Glossary 21
9. Warranty 22
10. List of Accessories 22
RoHS Compliance
Scientech Products are RoHS Complied.
RoHS Directive concerns with the restrictive use of Hazardous substances (Pb, Cd, Cr, Hg, Br compounds) in
electric and electronic equipments.
Scientech products are Lead Free and Environment Friendly.
It is mandatory that service engineers use lead free solder wire and use the soldering irons upto (25 W) that
reach a temperature of 450C at the tip as the melting temperature of the unleaded solder is higher than the
leaded solder.
Scientech Technologies Pvt. Ltd. 3
-
7/29/2019 Strain Gauge Trainer
4/22
ST2304
Introduction
Strain Gauge Trainer ST2304 provides a study of the Strain Gauge and itsapplication for the measurement of strain. It helps to study bridge configuration of
Strain Gauge and the signal conditioning circuits required to measure strain. It usescantilever to produce strain on Strain Gauges. The Strain Gauges are firmlycemented to the cantilever at the point where the strain is to be measured. Weightsare placed on the free end of the cantilever. Strain developed changes the resistanceof Strain Gauge which is detected by full bridge configuration.
The seven-segment LED display shows strain in micro strain units. Differentweights are provided to perform linearity and sensitivity experiments. Model
ST2304 is fully covered self contained single box with USB interface and easy to
use design detailed experiment manual is supplied with interactive real timesoftware. The manual includes theory of the subject and experiments block
description, operating instruction etc.
Strain Gauge Trainer ST2304 can be interfaced with PC using real time
software. *Note : USB interface is optional.
Features
Self-contained and easy to operate.
Sensitive, Linear, Stable & Accurate. Functional blocks indicated on board Mimic. Test-points to observe the Input Output of each block. Onboard Gain Adjustment. Onboard Offset null Adjustment. Built-in DC Power Supplies. 3 digits LED display. Onboard Cantilever Arrangement.
High Repeatability and Reliability. *USB interface with Real-time Software (PC) for step by stepapproach. * only in USB interface model.
Scientech Technologies Pvt. Ltd. 4
-
7/29/2019 Strain Gauge Trainer
5/22
ST2304
Technical Specifications
Strain Gauge (350) : 4 Nos.
Gauge Factor : 2:1Maximum Bearable Weight : 500 gms.
Cantilever Material : Stainless Steel
Cantilever Width : 2.5 cm
Cantilever Thickness : 0.16 cm
Cantilever Length : 20 cm
Bridge Voltage : +8 V DC
Bridge Configuration : Full Bridge
Display : 3 Digit LED
Test Points : 8 in numbers
Power Requirement : 230V10%, 50 Hz
Dimensions (mm) : W420 H100 D255
Weight : 3.5 Kgs. (approximately)
*Baud Rate : 4800
*ADC Resolution : 8 Bit
Scientech Technologies Pvt. Ltd. 5
-
7/29/2019 Strain Gauge Trainer
6/22
ST2304
Operating Instructions and Panel Control Description
The ST2304 Strain Gauge Trainer is equipped with a built in DC-Power supply.
Attach the three pin mains cord to the three pin socket and connect the other end to a
stable 230V AC supply. A fuse (100mA/250V) rating is connected in series withmains supply. When the On/Off switch of the trainer is turned On the power
LED indication will glow, indicating that the trainer is On.
The Maximum weight that can be placed over the cantilever is 500 gms. Strain due
to weights as low as 5 gms can be detected. Do not press the cantilever by hand or
by other means as it may damage spring action of the cantilever. After placing the
weight, make the cantilever stable by hand to get steady reading on the display.
Place weights at the centre of the tray to get exact strain values.
Test points are given on the mimic board to measure the input output voltages of
different stages of instrumentation system of the strain gauge. Real time software is
installed to make the operation of the trainer more interactive. Ground of the Strain
Gauge Bridge is isolated from the circuit ground. So whenever voltages of thebridge arms are to be measured, the bridge ground should be used.
Gain of instrumentation amplifier can be adjusted with the Gain Adjustpreset given
in instrumentation amplifier block. Don't disturb the preset as it is factory calibrated
otherwise display reading will not be accurate.
Offset Null Adjustpreset is used to make the display to read 000, when no weight is
placed on the cantilever. Gain Adjustpreset of Low Noise Amplifier can be used to
increase the span limit i.e. to increase the display reading for maximum weight 500
gms. 3 Digit LED display will indicate strain in -strain.
Note : Do not disturb the preset settings except offset null adjust prior to initiatingstrain
measurement. Offset null is to be done every time before strain measurement. For lowerweights like 5, 10, 15 gm subtract No Loaddisplay reading (in case display reading is
not 000) from the display reading for 5, 10, 15 gm to get the result close to theoretical
strain. The LSB of display may fluctuate for weight placed on the cantilever. For
accurate strain measurement, take middle value of fluctuating LSB.
For software installation, open the CD provided with the trainer. Open Setup2304
folder, double click on setup.exe, and follow the instruction. Click on check box
of install USB driver option and click on finish.
Scientech Technologies Pvt. Ltd. 6
-
7/29/2019 Strain Gauge Trainer
7/22
ST2304
Functional Description of Blocks
1. Cantilever :One end of the cantilever is fixed while the other end is free to move. Weights
are placed on free end which causes bending of the cantilever and strain isproduced on the fixed end. Four strain gauges are pasted on fixed end tomeasure strain. Two strain gauges are pasted above the cantilever and othertwo are pasted below the cantilever. Weights up to 500 gms can be placed on
the cantilever. Strain due to weights as low as 5 gms can be detected.
Figure 1
Figure 2
Scientech Technologies Pvt. Ltd. 7
-
7/29/2019 Strain Gauge Trainer
8/22
ST2304
Figure 3
2. Strain Gauge Bridge :Four strain gauges are connected in the four arms of the Wheatstone bridge. +8
V DC is used to excite the bridge and output is taken from the remaining arm.When no weight is placed on the cantilever, strain developed is zero.Resistances of all four strain gauges are equal so bridge is balanced and outputof bridge is zero. Whenever weight is placed, strain developed at fixed end
causes resistance of strain gauge to vary which disturbs the balanced conditionof bridge and output is produced which is amplified by instrumentationamplifier and low noise amplifier and given to display.
3. Instrumentation Amplifier :It consists of two stages. The First stage is the buffer stage and the second
stage is the differential amplifier. Buffer stage is used to provide high input
impedance to amplifier. Differential amplifier amplifies weak difference signal
and rejects common mode signals such as 50 Hz hum and noise. The output of
Scientech Technologies Pvt. Ltd. 8
-
7/29/2019 Strain Gauge Trainer
9/22
ST2304
this block is single ended. Instrumentation amplifier is used to amplify weak
signals and have a high value of CMRR, voltage gain and low value of noise,
offset voltage, offset drift etc. Gain of instrumentation amplifier is set by Gain
Adjust preset.
4. Low Noise Amplifier :It is a low noise, low drift amplifier. This gives additional current gain to the
output of the Instrumentation amplifier. The auxiliary output is in mV as
indicated by the display. The output can be used as input to some recording
stage to record the data.
5. Display :It is 3 Digit LED display. It shows strain developed at the fixed end of the
cantilever in -strain.
Scientech Technologies Pvt. Ltd. 9
-
7/29/2019 Strain Gauge Trainer
10/22
ST2304
Theory
Strain Gauge :
If a metal conductor is stretched or compressed, its resistance changes on account of the
fact that both the length and diameter of the conductor change. There is also a change inthe value of resistivity of the conductor when it is strained and this property is called
piezoresistive effect. This is the principle of strain gauge. Strain gauge is a device the
electrical resistance of which varies in proportion to the amount of strain in the device.
The most widely used gauge is the bonded metallic strain gauge.
A strain gauge of length L, area A, and diameter D when unstrained has resistance R
= (L)/ A
When a gauge is subjected to positive strain, its length increases while its area of
cross section decreases, resistance of gauge increases with positive strain.
Poisson ratio =lateral strain
=D / D
longitudinal strain L / L
= strain = L/L
Gauge Factor =R
/R
L / L
Strain definition :
Strain is the amount of deformation of a body due to an applied force. More
specifically, strain () is defined as the fractional change in length, as shown below.
Figure 4
=LL
Strain can be positive (tensile) or negative (compressive). Although dimensionless,
strain is sometimes expressed in units such as in/in or mm/mm. In practice, the
magnitude of measured strain is very small. Therefore, strain is often expressed as
micro strain (-strain), which is x 10-6.Types of Strain gauges :
1. Unbonded metal strain gauges.2. Bonded metal wire strain gauges.3. Bonded metal foil strain gauges.4. Vacuum deposited thin metal film strain gauges.5. Sputter deposited thin film metal strain gauges.Scientech Technologies Pvt. Ltd. 10
-
7/29/2019 Strain Gauge Trainer
11/22
ST2304
6. Bonded semiconductor strain gauges.7. Diffused metal strain gaugeBonded metallic (Foil type) strain gauges are commonly used due to their
advantages over other strain gauges thus it is discussed in detail below.The metallic strain gauge consists of a very fine wire or, more commonly, metallic foil
arranged in a grid pattern. The grid pattern maximizes the amount of metallic wire or
foil subject to strain in the parallel direction as shown below. The cross sectional area of
the grid is minimized to reduce the effect of Shear strain and Poisson Strain. The grid is
bonded to a thin backing, called the carrier, which is attached directly to the test
specimen. Therefore, the strain experienced by the test specimen is transferred directly
to the strain gauge, which responds with a linear change in electrical resistance. Strain
gauges are available commercially with nominal resistance values from 30 to 3000,with 120, 350 and 1000 being the most common values.
Bonded Metallic Strain Gauge
Figure 5
It is very important that the strain gauge be properly mounted on to the test
specimen so that the strain is accurately transferred from the test specimen, through
the adhesive and strain gauge backing, to the foil itself. A fundamental parameter ofthe strain gauge is its sensitivity to strain, expressed quantitatively as the gauge
factor (GF). Gauge factor is defined as the ratio of fractional change in electrical
resistance to the fractional change in length.
GF =R/ RR/ RL / L
The Gauge Factor for metallic strain gauges is typically around 2.
Scientech Technologies Pvt. Ltd. 11
-
7/29/2019 Strain Gauge Trainer
12/22
ST2304
Strain Measurement Techniques :
In practice, the strain measurements rarely involve quantities larger than a few milli-
strain ( 10-3). Therefore, to measure the strain requires accurate measurement ofvery small changes in resistance. To measure such small changes in resistance, straingauges are almost always used in a bridge configuration with a voltage excitationsource. The general Wheatstone bridge, illustrated below, consists of four resistive
arms with an excitation voltage, VEX, that is applied across the bridge.
Figure 6
The output voltage of the bridge, Vo, will be equal to :
R3 R2 Vo = VEX
R4 R1 3 R2From this equation, it is apparent that when R1/R2 = R4/R3, the voltage output Vo will
be zero. Under these conditions, the bridge is said to be balanced. Any change inresistance in any arm of the bridge will result in a nonzero output voltage. Therefore, if
we replace R4 in Figure shown above with an active strain gauge, any changes in the
strain gauge resistance will unbalance the bridge and produce a nonzero output voltage.If the nominal resistance of the strain gauge is designated as RG, then thestrain-induced change in resistance, DR, can be expressed as DR = RG.*F.*.Assuming that R1 = R2 and R3 = RG, the bridge equation above can be rewritten toexpress Vo/ VEX as a function of strain as shown below. Note the presence of the
1/(1+GF/2) term that indicates the nonlinearity of the quarter bridge output withrespect to strain ( = strain).
Figure 7
Scientech Technologies Pvt. Ltd. 12
-
7/29/2019 Strain Gauge Trainer
13/22
ST2304
V GF 1
OV 4
1 GFEX
2Ideally, it is required that the resistance of the strain gauge to change only in response to
applied strain. However, strain gauge material, as well as the specimen material to which
the gauge is applied, will also respond to changes in temperature. Strain gauge
manufacturers attempt to minimize sensitivity to temperature by processing the gauge
material to compensate for the thermal expansion of the specimen material for which the
gauge is intended. While compensated gauges reduce the thermal sensitivity, they do not
totally remove it. The sensitivity of the bridge to strain can be doubled by making both
gauges active in a half-bridge configuration. It can be further increased by making all
four resistances of the arms of the bridge by active strain gauges in a full-bridge
configuration. The full-bridge circuit is shown below.
Figure 8V
O GF
VEX
When no strain is applied the output of the Wheatstone bridge circuit is zero. Inpractice however, resistance tolerances and strain induced by gauge application willgenerate some initial offset voltage. This initial offset voltage is typically handled intwo ways. First, a special offset-nulling or balancing circuit to adjust the resistancein the bridge to rebalance the bridge to zero output can be used. Alternatively, theinitial unstrained output of the circuit can be measured and compensate in finalmeasurement.
Scientech Technologies Pvt. Ltd. 13
-
7/29/2019 Strain Gauge Trainer
14/22
ST2304
Signal conditioning stages for strain gauge :
Amplification :
Strain gauges typically provide small signal levels. It is therefore important to have
accurate instrumentation to amplify the signal before it is given to next stage i.eADC, display etc.
Excitation :
Strain gauges require voltage excitation to generate a voltage representing strain.
This voltage source should be constant and at a level recommended by the strain
gauge manufacturer.
Bridge Completion :
Strain gauges are offered in several different configurations: quarter-bridge, half-
bridge, and full bridge. For quarter and half-bridge strain gauges, instrumentation
should provide bridge completion, adding the necessary resistors to complete a
Wheatstone bridge.
1. Full-bridge strain gauge : The entire Wheatstone bridge is provided with thestrain gauge. Instrumentation only needs to provide the excitation inputs.
2. Half-bridge strain gauge : Half of the Wheatstone bridge is provided with thestrain gauge. Instrumentation needs to provide two of the four resistors to
complete the Wheatstone bridge. This is known as half-bridge completion.
3. Quarter-bridge strain gauge : Quarter of the Wheatstone bridge is provided withthe strain gauge. Instrumentation needs to provide three of the four resistors to
complete the Wheatstone bridge. This is known as quarter-bridge completion.
Linearization/Strain Gauge Conversion :
While strain gauges are close to linear, they do stray from linear at large strains. In
addition, it will need some hardware or software to convert the voltage output of the
strain gauge into a strain measurement. The conversion formula depends on the type
of strain gauge used. Half and full-bridge strain gauges offer more accurate
conversion formulas.
Offset Nulling Circuitry :
A strain gauge application will have some position that will be identified as the rest
position (a reference position). The strain gauge should produce 0 volts at this
position. Offset nulling circuitry is used to produce 0V at rest position.
Formulas :
Ohm's Law :
Voltage = Current x Resistance (written V = IR)
Resistance = Resistivity x Length / Cross-Sectional SurfaceArea R = L / A
Stress = Strain x Modulus of Elasticity (written s = E )
Strain = (6 x Force (N) x Length) / (Width x Thickness2
x Young's modulus)
Modulus of Elasticity of stainless Steel = 10 x 106
psi
Young's Modulus E = 200 109 N/m2 for steel
Scientech Technologies Pvt. Ltd. 14
-
7/29/2019 Strain Gauge Trainer
15/22
Experiment 1
Objective :Study of Strain measurement using strain gauges and cantilever
assembly Procedure :1. Switch On the trainer.2. Observe reading of the display. It should be 000.3. If the display reading is not 000 then adjust offset null.4. *Connect USB cable between ST2304 Trainer and PC.5. *Open the Real time software and select port where the USB cable is connected.
If USB port connects beyond com10, it will not be showing in drop down
list. Go on Device manager, change its property, and assign USB port
between com 2 to com 9.
6. *Click on start button.7. *Place a weight of 400 gms on the cantilever's free end.8. *Match Calibrated value C with reference Value R by using + and button.9. *Press set Button.10. Place a weight of 50 gms on the cantilever's free end and observe the display
reading. It indicates strain developed on cantilever in strain
11. Calculate theoretical strain by the formula.(6 F L) / (W T2Y)
F = mg
Where, F = force (N),
m = mass (Kg),
g = acceleration due to gravity (9.8m/Sq. sec),
L = Length (m), w = width (m),
T = thickness (m),
Y = Young's modulus (N/Sq. m) = 200109
N/m2
for stainless steel.
Or
12 *Select the weight (which you place on the cantilever) on real time software, itwill show theoretical strain (Strain) and then click on get button to comparetheoretical and practical strains. There will be 3-4% variation between
theoretical strain and practical strain on software window.
15
-
7/29/2019 Strain Gauge Trainer
16/22
ST2304
13 Repeat the above maintained for different weights as shown in the following
table and complete the table.
Sr.
Weight (gm)
Theoretical Display
No. Strain (10-6) Reading ( l0-6)1 50
2 100
3 150
4 200
5 250
6 300
7 350
8 400
9 450
10 500
14. *After taking readings, click on Arrange Button.
16
-
7/29/2019 Strain Gauge Trainer
17/22
15. *Select theoretical Value and click on plot to get the plot between Theoretical
Strain and Weight it will appear to be linear.
16. *Select Practical Value and click on plot to get the plot between PracticalStrain and Weight.
17. *To analyze the difference between Practical strain and theoretical strain selectboth and click on plot button.
For lower weights like 5, 10, 15 gm subtract no load display reading (in casedisplay reading is not 000) from the display reading for 5, 10, 15 gm to get the
result close to theoretical strain. The LSB of display may fluctuate for anyweight placed on the cantilever. For accurate strain measurement, take middlevalue of fluctuating LSB. Real time software works only for weight above 50
gms.
* only for usb interface model.
17
-
7/29/2019 Strain Gauge Trainer
18/22
ST2304
Experiment 2
Objective :
Determining Linear Range of operation
Procedure :
1. Switch On the trainer.2. Observe display reading. It should be 000.3. If display reading is not 000, then adjust Offset Null Adjust preset slowly.4. Place weight of 50 gms on cantilever's free end.5. Note the display reading.6. Calculate theoretical strain by the formula.
(6 F L) / (W T2Y)
F = mgWhere,
F = force (N),
m =mass (Kg),
g = acceleration due to gravity (9.8 m/Sq.
see), L = Length (m),
w = width (m),
T = thickness (m),
Y = Young's modulus (N/ Sq. m) = 200 109 N/m2 for stainless steel.
7. Complete the following table by repeating steps 4, 5 and 6 for weights shownin the table.
Sr.
Weight (gm)
Theoretical Strain Display Reading
No. (10-6) ( l0-6)1 50
2 100
3 150
4 200
5 250
6 3007 350
8 400
9 450
10 500
Scientech Technologies Pvt. Ltd. 18
-
7/29/2019 Strain Gauge Trainer
19/22
ST2304
8. Plot the graph between weight (gm) and display reading (-strain).
Figure 9
9. Observe the graph. and point out the linear portion of the graph. This is thelinear range of operation.
10. Calculate :% Linearity = (Maximum deviation of display reading from theoretical) strain
100 Theoretical strain for 500 gm
Scientech Technologies Pvt. Ltd. 19
-
7/29/2019 Strain Gauge Trainer
20/22
ST2304
Experiment 3
Objective :
Determining sensitivity of
Trainer Theory :
Sensitivity :
The ratio of the change in auxillary output to a change in the value of the measurand
(strain). Sensitivity is the smallest change in strain, which the trainer is able to
detect. Strain is directly proportional to weight.
Sensitivity S =Auxillary Output
mV/gm Weight
Procedure :
1. Switch On the trainer.2. Measure the auxillary output.3. Adjust Offset Null Adjust preset slowly to get 0 mV at auxillary output terminal.4. Place weight of 5 gm on cantilever and measure the auxillary output voltage
by multimeter in 200 mV range.
5. Repeat the above step by placing the weights of 10gm, 20 gms etc.6. Calculate :
S =Auxillary Output
for above specified weights.Weight
=. mV/gm7. Compare value of sensitivity for different weights.
Scientech Technologies Pvt. Ltd. 20
-
7/29/2019 Strain Gauge Trainer
21/22
ST2304
Glossary
End Points :
The outputs at the specified upper and lower limits of the range. Unless otherwise
specified, end points are averaged during anyone calibration.Excitation :
The external electrical voltage and/or current applied to a transducer for its proper
operation. Usually expressed as range(s) of voltage and/or current values.
Linearity :
The closeness of a calibration curve to a specified straight line. Linearity is expressed as
the maximum deviation of any calibration point on a specified straight line, during
anyone calibration cycle. It is expressed as "within +/percent of full range output."
Measurand :
A physical quantity, property or condition which is measurand. The term measurand
is preferred to input, parameter to be measured, physical phenomenon,
stimulus, and variable.
Range :
The measurand values, over which a transducer is intended to measure, specified by
their upper and lower limits.
Repeatability :
The ability of a transducer to reproduce output readings when the same measurand
value is applied to it consecutively, under the same conditions, and in the same
direction. Repeatability is expressed as the maximum difference between output
readings; it is expressed as "within percent of full-scale output."
Two calibration cycles are used to determine repeatability unless otherwise specified.
Resolution :
The magnitude of output step changes as the measurand is continuously varied over
the range. Resolution is best specified as average and maximum resolution; it is
usually expressed in percent of full-scale output. In the sense of the smallest
detectable change in measurand, use threshold.
Sensitivity :
The ratio of the change in transducer output to a change in the value of the
measurand. In the sense of the smallest detectable change in measurand, threshold is
used.
Span :
The algebraic difference between the limits of the range.
Transducer :
A device which provides a usable output in response to a specified measurand. The
term transducer is usually preferred to sensor.
Scientech Technologies Pvt. Ltd. 21
-
7/29/2019 Strain Gauge Trainer
22/22
ST2304
Warranty
1. We guarantee the product against all manufacturing defects for 24 monthsfrom the date of sale by us or through our dealers. Consumables like dry cell
etc. are not covered under warranty.2. The guarantee will become void, if
a) The product is not operated as per the instruction given in the operatingmanual.
b) The agreed payment terms and other conditions of sale are not followed.c) The customer resells the instrument to another party.d) Any attempt is made to service and modify the instrument.
3. The non-working of the product is to be communicated to us immediatelygiving full details of the complaints and defects noticed specifically
mentioning the type, serial number of the product and date of purchase etc.
4. The repair work will be carried out, provided the product is dispatched securelypacked and insured. The transportation charges shall be borne by the customer.
List of Accessories
1. Weight 5 gm ........................................................................................ 2 Nos.
2. Weight 10 gm ....................................................................................... 2 Nos.
3. Weight 20 gm ....................................................................................... 1 No.
4. Weight 50 gm ....................................................................................... 3 Nos.
5. Weight 100 gm ..................................................................................... 2 Nos.
6. Weight 200 gm ..................................................................................... 1 No.
7. Mains Cord ........................................................................................... 1 No.
8. e-Manual (PC Software inclusive).........................................................1 No.
9. Aligner-911 .......................................................................................... 1 No.
10. * USB Cable ......................................................................................... 1 No.
* Only in USB interface model.
Updated 01-07-2009
Scientech Technologies Pvt. Ltd. 22