sterilization standards update: strategies for compliance

41
Sterilization requirements: Strategies for Compliance presented by John Beasley Owner & Sr. Consultant MedTech Review

Upload: medtech-review-llc

Post on 22-Nov-2014

6.341 views

Category:

Documents


0 download

DESCRIPTION

My presentation for MEDTEC China 2010

TRANSCRIPT

Page 1: Sterilization Standards Update:  Strategies for Compliance

Sterilization requirements: Strategies

for Compliance

presented by John Beasley

Owner & Sr. Consultant MedTech Review

Page 2: Sterilization Standards Update:  Strategies for Compliance

Topics

•  General Requirements •  Radiation Sterilization

–  Impact to Subcontractors –  Impact to Manufacturers

•  EO Sterilization –  Impact to Subcontractors –  Impact to Manufacturers

•  Summary

Page 3: Sterilization Standards Update:  Strategies for Compliance

Introduction

•  Presentation covers –  Gap Analysis between

current version and previous version of standards

–  Compliance information

–  References

•  Presentation does not cover –  All differences – only

the most significant –  The steps for

sterilization or validation of sterilization processes

Copyright  laws  prevent  MedTech  Review  and  MEDTEC  China  Conferences  from  providing  copies  of  standards  to  

conference  par;cipants.  

Page 4: Sterilization Standards Update:  Strategies for Compliance

Sterilization Processing by IR or EO

Page 5: Sterilization Standards Update:  Strategies for Compliance

General Requirements

•  Aligned somewhat with the requirements of ISO 13485:2003 – Document and Record Retention

•  Procedures for development, validation, routine control and release (ISO 13485:2003, Clause 4)

– Management Responsibility •  Authority, responsibility, including contractors (ISO

13485:2003, Clause 5)

Page 6: Sterilization Standards Update:  Strategies for Compliance

General Requirements

– Product Realization •  Procedures for purchasing, identification and

traceability, and calibration are required (ISO 13485:2003, Clause 7)

– Measurement and Improvement •  Controls for non-conforming material and

corrective actions (ISO 13485:2003, Clause 8)

Page 7: Sterilization Standards Update:  Strategies for Compliance

Sterilization by Radiation

•  ISO 11137:2006 – Part 1

•  Requirements for development, validation, and routine control of a sterilization process for medical devices (approved 15 April 2006)

– Part 2 •  Establishing the sterilization dose (approved 15

April 2006) – Part 3

•  Guidance on dosimetric aspects (approved 15 April 2006)

Page 8: Sterilization Standards Update:  Strategies for Compliance

Sterilization by Radiation

•  ISO 11137:2006 is now mandatory and fully implemented in the EU (as of April 2009) – The following standards are now OBSOLETE

•  EN 552 •  ISO 11137:1995 •  AAMI TIR 27:2001 •  ISO/TIR 13409:1996 •  TIR 15844:1998

Page 9: Sterilization Standards Update:  Strategies for Compliance

Certification

•  Contract sterilization companies who carry out irradiation processing can include EN ISO 11137-1:2006 within the scope of their ISO 13485 registration.

•  EN ISO 11137 Parts 2 and 3 cannot be included.

Page 10: Sterilization Standards Update:  Strategies for Compliance

ISO 11137:2006 – Main Changes

•  Clause 3 – Definitions – Processing category

•  Group of different products that can be sterilized together, based on

– Composition – Density – Dose requirements

Page 11: Sterilization Standards Update:  Strategies for Compliance

ISO 11137:2006 – Main Changes

– Product family •  Group of different products that can be given the

same sterilization dose based on the nature of the – Raw materials – Components – Manufacturing processes –  Equipment –  Environment –  Location

It  would  be  helpful  for  the  manufacturer  to  develop  a  process  map  or  decision  tree  to  consistently  address  the  requirements  and  document  the  

ra8onale  for  groupings.  

Page 12: Sterilization Standards Update:  Strategies for Compliance

ISO 11137:2006 – Main Changes

•  Clause 5 – Sterilizing Agent Characterization – New requirements for environmental

considerations

Are  there  environmental  impacts  such  as  discharges  into  the  air  or  water  that  may  adversely  impact  the  environment?    How  are  these  controlled?  

Page 13: Sterilization Standards Update:  Strategies for Compliance

ISO 11137:2006 – Main Changes

•  Clause 6 – Process and Equipment Characterization – Requirements have been expanded –

irradiator and its method of operation shall be specified.

Can  be  addressed  under  vendor  contract  /  purchasing  control  /  internal  control  procedures  such  as  process  control,  equipment  valida8on  and  equipment  change  

control.  

Page 14: Sterilization Standards Update:  Strategies for Compliance

ISO 11137:2006 – Main Changes

•  Clause 7 – Product Definition – New requirements emphasize control of product

AND packaging AND control of bioburden – New requirements – defining product families in

establishing dose and dose audit based on •  Bioburden •  Representative product •  Master product •  Equivalent product •  Simulated product

Product  families  must  be  established  with  documented  criteria,  maintained  (i.e.,  reviewed)  and  risks  in  using  product  families  must  be  addressed    

Page 15: Sterilization Standards Update:  Strategies for Compliance

ISO 11137:2006 – Main Changes

•  Clause 8 – Process Definitions –  Incorporates Method 1, Method 2 and Vdmax

Method (15 and 25) – New Means for dose transference

SOPs  need  to  include  requirements  for  establishment  of  max  dose  and  steriliza8on  dose;  manufacturers  need  to  document  the  appropriate  method  used  and  process  

for  valida8on.  

Page 16: Sterilization Standards Update:  Strategies for Compliance

ISO 11137:2006 – Main Changes

•  Clause 9 – Validation –  IQ / OQ / PQ requirements have been

expanded – Dose mapping (both container and product) – Process specifications established directly

from validation; requirements detailed – Defined sample sizes required

Page 17: Sterilization Standards Update:  Strategies for Compliance

ISO 11137:2006 – Main Changes

•  Clause 12 – Maintaining process effectiveness – Standard defines a more flexible approach for

maintaining effectiveness of process •  Bioburden specifications (number and type)

–  1 month, 3 month, each batch

•  Dose audits – frequency; rationale •  Failure Investigation and actions

Page 18: Sterilization Standards Update:  Strategies for Compliance

Impact to Sterilization Contractor

•  Requirements for environmental considerations

“The  poten8al  effect  on  the  environment  of  the  opera8on  of  the  radia8on  steriliza8on  process  shall  be  assessed  and  measures  to  protect  the  environment  shall  be  iden8fied.    

This  assessment,  including  poten8al  impact  (if  any)  shall  be  documented  and  measures  for  control  (if  iden8fied),  shall  be  specified  and  implemented.  

Page 19: Sterilization Standards Update:  Strategies for Compliance

Impact to Sterilization Contractor

•  Think about all of the possible environmental impacts that might occur in the event of a failure of the radiation control system. – Maintenance of the source material and its

storage (pool water levels, alarms, testing) – Control of access to the cell – Care and maintenance of the source racks/

carriers – Control room alarms – Auditing for safety

These  items  can  be  reviewed  and  covered  during  preven8ve  

maintenance  ac8vi8es  and  supplier  audits.  

Page 20: Sterilization Standards Update:  Strategies for Compliance

Impact to Sterilization Contractor

•  Process interruptions –  What happens when there is an interruption in the

exposure? –  What is the impact of stopping and restarting the

cycle and the timing? How are under-dosing and overdosing prevented?

–  What is the additional exposure to the product resulting from the travel time of the source material rising up and lowering down, possibly multiple times?

–  Is there a risk of product or package degradation caused by the changes in temperature while in the cell due to the interruptions?

Page 21: Sterilization Standards Update:  Strategies for Compliance

Impact to Sterilization Contractor

•  Partial containers – dose mapping

“If  par8ally-­‐filled  irradia8on  containers  are  to  be  used,  the  effect  of  par8al  filling  on  dose  distribu8on  within  irradia8on  containers  or  in  other  irradia8on  containers  present,  shall  be  determined  and  recorded.    When  a  carrier  is  not  completely  full,  the  density  of  the  load  becomes  inconsistent,  and  therefore  the  delivered  dose  can  vary.  By  properly  mapping  par8al  carrier  loads,  equivalent  dose  distribu8on  can  be  assured.”  

Page 22: Sterilization Standards Update:  Strategies for Compliance

Impact to Sterilization Contractor

•  Procedures for review and product release •  11.1 – Any specific periodic tests, calibrations,

maintenance tasks and necessary re-qualifications need to be documented.

•  11.2 – Procedures for review of records and product release from sterilization need to be written.

When  a  steriliza8on  process  is  found  to  not  meet  specifica8ons,  taking  into  account  the  uncertainty  of  the  measurement  system(s),  then  the  product  shall  be  

handled  as  nonconforming.  

Page 23: Sterilization Standards Update:  Strategies for Compliance

Impact to Sterilization Contractor

•  Each of the previous items must be audited in addition to – Personnel training – Dosimetery – Radiation cell refueling

Page 24: Sterilization Standards Update:  Strategies for Compliance

Impact to Manufacturer

•  VDMax 25

– Bioburden testing unchanged (<1000 CFU per device); verification does set to 10-1; sterility testing on 10 units

•  VDMax 15 – Bioburden control limits extremely low (<1.5

CFU per device) •  ISO 15843 – Dose audit frequency

– Quarterly -> Semi-annually -> Annually

Page 25: Sterilization Standards Update:  Strategies for Compliance

Regarding China

•  GB18280-2007 - Sterilization of health care products requirements for validation and routine control of radiation sterilization –  idt ISO 11137:2006

Page 26: Sterilization Standards Update:  Strategies for Compliance

Sterilization by EO

•  ISO 11135-1:2007 – Requirements for development, validation,

and routine control of a sterilization process for medical devices (approved 01 May 2007)

•  ISO 11135-2:2008 – Guidance on the application of ISO

11135-1:2007

Page 27: Sterilization Standards Update:  Strategies for Compliance

Sterilization by EO

•  ISO 11135 is now mandatory and fully implemented in the EU (as of May 2010) – The following standards are now OBSOLETE

•  EN 550:1994 •  ISO 11135:1994

Page 28: Sterilization Standards Update:  Strategies for Compliance

Certification

•  Contract sterilization companies who carry out EO processing can include EN ISO 11135-1:2007 within the scope of their ISO 13485 registration.

•  EN ISO 11135-2:2008 cannot be included.

Page 29: Sterilization Standards Update:  Strategies for Compliance

Sterilization by EO

•  Cycle Development – Several methods for cycle development

offered in Annex – Use of developmental chambers as well as

production chambers for process development expanded

– Chamber and process equivalency – Clarification of microbiological and physical

PQ provided

Page 30: Sterilization Standards Update:  Strategies for Compliance

Sterilization by EO

•  Main Changes – Clause 5 – Sterilizing Agent Characterization

•  New requirements for microbicidal effectiveness if EO outside range of widely recognized compositions or if novel diluents are to be used

•  Effects of EO on the product to be documented •  Environmental considerations

Page 31: Sterilization Standards Update:  Strategies for Compliance

Sterilization by EO

– Clause 6 – Process and Equipment Characterization •  Generally relates to the requirements of the

contractor; responsibility lies with the manufacturer to ensure contractor sterilizer complies with these specific requirements

•  Process Characterization is a result of the OQ •  Equipment Characterization is a result of the IQ

Page 32: Sterilization Standards Update:  Strategies for Compliance

Sterilization by EO

– Clause 7 – Product Definition •  Product and packaging must meet the specified

requirements for safety, quality, and performance following the sterilization process at the most challenging process parameters

– Useful Standards •  ISO 10993 (relevant parts) for biological safety and

EO residuals following exposure •  AAMI TIR 28:2001; Product Adoption and Process

Equivalency for EO Sterilization

Page 33: Sterilization Standards Update:  Strategies for Compliance

Sterilization by EO •  ISO 11737-1:2006 specifies requirements and provides

guidance for the enumeration and microbial characterization of the population of viable micro-organisms on or in a medical device, component, raw material or package.

•  ISO 17664:2004 specifies the information to be provided by the medical device manufacturer on the processing of medical devices claimed to be resterilizable, and medical devices intended to be sterilized by the processor.

•  ISO 11607-1:2006 specifies the requirements and test methods for materials, preformed sterile barrier systems, sterile barrier systems and packaging systems that are intended to maintain sterility of terminally sterilized medical devices until the point of use.

Page 34: Sterilization Standards Update:  Strategies for Compliance

Sterilization by EO

– Clause 8 – Process Definition •  Process parameters clearly defined •  Appropriateness of BI and process challenge

device to be defined and documented to demonstrate and support SAL

Page 35: Sterilization Standards Update:  Strategies for Compliance

Sterilization by EO

– Clause 9 – Validation •  The number of sensors in PQ (temperature,

humidity, BIs) in the load is based on product load volume as opposed to usable chamber volume.

•  The number of sensors in IQ/OQ is based on chamber volume.

– Clause 10 – Routine Monitoring and Control •  New requirements for additional levels of reviews

when parametric release is to be used

Page 36: Sterilization Standards Update:  Strategies for Compliance

Sterilization by EO

– Clause 11 – Product Release •  Conventional release and parametric release are

more harmonized – Clause 12 – Maintaining Process

Effectiveness •  Justification for requalification requirements /

intervals of requalification

Page 37: Sterilization Standards Update:  Strategies for Compliance

Sterilization by EO

– Requalification Review to include •  Verify appropriateness of BIs •  Verify loading patterns remain unchanged •  Verify no changes to design, materials, load

configuration or manufacturing process •  Verify no change in bioburden or characterization has

occurred •  Verify temperature distribution and chamber operation

remain unchanged •  Review of sterilization process history demonstrates

repeatability •  Review of preventive maintenance programs

demonstrate no change to sterilizing equipment Recommended  reduce  microbial  performance  qualifica;on  studies  are  performed  at  least  every  two  years  to  verify  the  documented  paperwork  review  has  captured  

any  changes  in  the  product  or  steriliza;on  review  

Page 38: Sterilization Standards Update:  Strategies for Compliance

Regarding China

•  GB 18279-2000 ethylene oxide sterilization of medical equipment validation and routine control –  idt ISO 11135:1994

Page 39: Sterilization Standards Update:  Strategies for Compliance

Summary

•  While techniques of sterilizing medical devices have not changed much over the past 25 years, the levels at which sterilization processes are monitored and controlled have changed significantly.

Page 40: Sterilization Standards Update:  Strategies for Compliance

Summary

•  Sterilization Validation is DYNAMIC –  ISO 14937:2009 specifies the elements of a

Quality Management System which are necessary to assure the appropriate characterization of the sterilizing agent, development, validation and routine monitoring and control of a sterilization process. •  It provides a standardized procedure for validating

new sterilization technologies.

Page 41: Sterilization Standards Update:  Strategies for Compliance

Thank You

John Beasley Owner & Sr. Consultant MedTech Review, LLC

www.medtechreview.com Email: [email protected]

SKYPE: medtechreview