steri trac detection system 0

2
ChemDAQ Steri-Trac Toxic Gas Detection System ChemDAQ’s Steri‐Trac gas monitoring system provides the optimal solution for workplace safety assurance, providing protection from toxic gases Sensor Exchange Program for Reduced Operating Costs The unique modular design of the Steri-Trac eliminates the need for in-house calibration. Using the sensor exchange program (SXP) you can receive a factory calibrated sensor every 4 months which you swap on site in a matter of seconds. User friendly Interfaces Steri-trac’s fixed detectors are designed to be used with the DAQ system to display gas concentrations, trends and exposures. Using a traffic light LED alarm system, it’s easy to interpret gas exposure in each area. Flexibility and Investment Protection The Steri-Trac toxic gas detector is a fully functioning, micro-processor controlled unit with tri-colour display and digital communications with the ability to report gas concentration in parts per million. Take Personnel Protection to the Next Level Mounted at entrance ways into areas where toxic gases are used or stored, it duplicates gas readings and alarms from its companion area detector, providing indication that it’s safe prior to entrance KEYBENEFITS DAQ software provides trend analysis and provides early warning of a situation to enable maximum protection Sensor change out reduces downtime and costs Modular design for custom configuration meets facilities specific needs ChemDAQs chemical filter eliminates nuisance alarms The only fixed detector to monitor Peracetic acid accurately at ppm levels GASESANALYSED ETO PAA O 3 H 2 O 2

Upload: a1-cbiss

Post on 12-Apr-2017

171 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Steri trac detection system 0

ChemDAQ Steri-TracToxic Gas Detection System

ChemDAQ’s Steri‐Trac gas monitoring system provides the optimal solution for workplace safety assurance, providing protection from toxic gases

Sensor Exchange Program for Reduced Operating CostsThe unique modular design of the Steri-Trac eliminates the need for in-house calibration. Using the sensor exchange program (SXP) you can receive a factory calibrated sensor every 4 months which you swap on site in a matter of seconds.

User friendly Interfaces Steri-trac’s fixed detectors are designed to be used with the DAQ system to display gas concentrations, trends and exposures. Using a traffic light LED alarm system, it’s easy to interpret gas exposure in each area.

Flexibility and Investment ProtectionThe Steri-Trac toxic gas detector is a fully functioning, micro-processor controlled unit with tri-colour display and digital communications with the ability to report gas concentration in parts per million.

Take Personnel Protection to the Next Level Mounted at entrance ways into areas where toxic gases are used or stored, it duplicates gas readings and alarms from its companion area detector, providing indication that it’s safe prior to entrance

KEYBENEFITS

• DAQ software provides trend

analysis and provides early warning

of a situation to enable maximum

protection

• Sensor change out reduces downtime

and costs

• Modular design for custom

configuration meets facilities specific

needs

• ChemDAQs chemical filter eliminates

nuisance alarms

“ The only fixed detector to monitor Peracetic acid accurately at ppm levels

GASESANALYSED

ETOPAA O3 H2O2

Page 2: Steri trac detection system 0

TECHNICALSPECIFICATION

GENERAL• Fixed detection of Ethylene Oxide, Hydrogen Peroxide &

Oxide

• Sample draw allows for EPA emissions of <1ppm / 24 hour

• Single,. multi point or multi gas installations

• Wide variety of applications including; pharmaceutical &

hospitals, drinks manufacturing and bio-decontamination

PHYSICAL• Dimensions: 10” (W) x 6” (H) x 5” (D)

• Display: Tricolour, 4 digit, 1.25”, 5 x 8 Matrix LED

PERFORMANCE • Alarms: 2 - User adjustable alarm levels, HIGH & LOW

• Output: Linear

• Calibration intervals: 4 months (factory calibrated)

OUTPUTS• Installation requirements: Wall mounted, 110 VAC, 1amp

• Communications: Digital RS485

• External contact relays: 2 relays open/closed available

• Monitors supported: Up to 8 as standard

CERTIFICATIONS• Meets OSHA regulations for monitoring toxic gases: CFR

1910.1047, CFR 1910.1000

• Meets Canadian Standards Associations for monitoring

toxic gases: CSA Z314.9-09

• Meets OSHA record-keeping requirements 30 years of

exposure data: CFR 1910.1020

• Meets the Joint Comission Standard EC 02.02.01 and

AAMI ST:41/2005 recommendations

Cert No. 996ISO 9001

11 Ark Royal Way, Lairdside Technology Park, Tranmere, Wirral, CH41 9HTT: +44(0)151 666 8300 F: +44(0)151 666 8329 E: [email protected] W: www.a1-cbiss.com Buy Online: shop.a1-cbiss.com

Ethylene Oxide Hydrogen Peroxide Ozone Peracetic Acid

Resolution (ppm) 0.1ppm 0.1 ppm 0.002ppm 0-0.1ppm

Sensor range 0-50ppm 0-10 ppm 0-5ppm 0-3ppm

Response time < 140 sec (T90) < 150 sec (T90) < 150 sec (T90)

Temperature / humidity

-20 to 50oc15 to 95%

non-condensing

0 to 50oc15 to 95%

non-condensing

-20 to + 50oc15 to 95%

non-condensing

MEASURINGRANGES